Analysis of Permanent Gases:

More Challenging Than You Might Think

Mark Sinnott

October 9, 2018

October 9, 2018 Agilent J&W DB-HeavyWAX1

Page 2: Analysis of Permanent Gases: More Challenging Than You

Analysis of Permanent Gases

➢Sample Introduction – Syringes and valves

➢General discussion on PLOT columns

➢ The molesieve column is at the heart of permanent gas separations

➢Techniques when CO2 + C2’s, C3’s, etc. is also needed

➢ Column Isolation

➢ Parallel columns

➢ Cryogenic separations

➢ Unique selectivity packed columns

➢Techniques for low level detection of hydrogen

October 9, 2018 Agilent J&W DB-HeavyWAX2

Page 3: Analysis of Permanent Gases: More Challenging Than You

Gas Sampling Valve

October 9, 2018 Agilent J&W DB-HeavyWAX3

Page 4: Analysis of Permanent Gases: More Challenging Than You

Syringes for Gas injection

October 9, 2018 Agilent J&W DB-HeavyWAX4

http://pconlab.net/Hamilton-Syringes.html

Page 5: Analysis of Permanent Gases: More Challenging Than You

Gas Tight Syringes

October 9, 2018 Agilent J&W DB-HeavyWAX5

Page 6: Analysis of Permanent Gases: More Challenging Than You

PLOT columns

October 9, 2018 Agilent J&W DB-HeavyWAX6

Column Type Phase TypeChromatographic

ProcessStationary Phases

WCOT LiquidGas - Liquid

partitioning

Polysiloxanes

PEG

PLOT SolidGas - Solid

adsorbtion

Porous Polymers

Al2O3, Zeolites, etc

WCOT = Wall Coated Open Tubular

PLOT = Porous Layer Open Tubular

Page 7: Analysis of Permanent Gases: More Challenging Than You

Why use a PLOT column?

➢Highly retentive

➢ (can resolve gases at non-cryo temperatures)

➢Unique selectivity

➢Permits higher initial oven temperatures than WCOT

October 9, 2018 Agilent J&W DB-HeavyWAX7

Page 8: Analysis of Permanent Gases: More Challenging Than You

October 9, 2018

"Solid"

Porous

Layer

Fused Silica

Tubing

Ideal for the analysis of gases due to

their increased retention (k) and unique

selectivity (a) compared to WCOT

PLOT Columns

8

Page 9: Analysis of Permanent Gases: More Challenging Than You

WCOT Ethylene Analysis

Page 10: Analysis of Permanent Gases: More Challenging Than You

PLOT Ethylene Analysis

Page 11: Analysis of Permanent Gases: More Challenging Than You

JW Column PLOT column Portfolio- DB, HP, CP11

HP-PLOT Al2O3

GS-Alumina

GS-Alumina KCl

HP-PLOT Al2O3 S

HP-PLOT Al2O3 M

CP- Al2O3/KCL

CP- Al2O3/Na2SO4

CP-SilcaPLOT

CP-CarboBOND

CP-CarboPLOT P7

GS-CarbonPLOT

CP-PoraPLOT Amines

CP-PoraPLOT S

CP-PoraBOND Q

CP-PoraPLOT Q

CP-PoraPLOT Q-HT

HP-PLOT Q

CP-PoraBOND U

HP-PLOT U

GS-Q

CP-PoraPLOT U

HP-PLOT Molesieve

CP-Molsieve 5A

GS-GasPro

ShinCarbon ST

Select Permanent Gas Column

Page 12: Analysis of Permanent Gases: More Challenging Than You

Dial 1-816-650-0621 for e-Seminar Audio

Selectivity Interactions in PLOT Phases

Shape / Size Surface

ZeolitesBonded Carbon /

Molecular SievesPorous Polymers Bonded Silica Al2O3

MolesieveMolesieve

CarbonPLOTPLOT-Q/U

GasPro

SilicaPLOTAlumina

Graphic of type KFI molecular sieve obtained

from the International Zeolite Association web

site (www.izo-online.org).

Flow

Zeolite "Backbone"

Size Selectivity

Gas Flow d- d- d+

d+

neutral

+ + + + + + + + + + + + +

+ + + + + + + + + + + + +

Vapor pressure always plays a leading role in solute interactions

Page 13: Analysis of Permanent Gases: More Challenging Than You

Molesieve Column

October 9, 2018 Agilent J&W DB-HeavyWAX13

➢Separation based mostly on molecular size and shape

➢Excellent at what it does but very limited

➢H2, O2, N2, CH4, CO

➢Noble Gases

➢NO2

➢SF6

➢Limitations…cannot use for…

➢CO2

➢Water

➢C2 HC’s and larger

Page 14: Analysis of Permanent Gases: More Challenging Than You

Molesieve Separations

October 9, 2018 Agilent J&W DB-HeavyWAX14

Page 15: Analysis of Permanent Gases: More Challenging Than You

Molesieve Separations

October 9, 2018 Agilent J&W DB-HeavyWAX15

Page 16: Analysis of Permanent Gases: More Challenging Than You

Tips for using a Molesieve column

➢ Molesive material is commonly used as a moisture trap

➢ Samples must be reasonably dry

➢ Separation will change as column absorbs water over time

➢ Resolution loss between CH4 and CO indicates when column should be reconditioned (300 ᵒ c 8 hours +)

October 9, 2018 Agilent J&W DB-HeavyWAX16

CH4 CO

Page 17: Analysis of Permanent Gases: More Challenging Than You

But I also need to analyze for CO2….its getting complicated…

October 9, 2018 Agilent J&W DB-HeavyWAX17

➢ Recall that CO2 cannot be done on a molesieve column (absorbed)

➢ Alternative options:

➢ 2 unique injections on 2 columns

➢ Molesieve + PLOT-Q

➢ 1 injection on 1 column at Cryo temps (-80 ᵒc!)

➢ GasPro

➢ 1 Injection on 1 column – ShinCarbon ST (packed column)

➢ 1 injection with 2 columns + valve to “direct traffic”

➢ Column Isolation - Molesieve + PLOT-Q

➢ 1 injection with 2 columns in parallel

➢ Select Permanent Gas Column - Molesieve + PLOT-Q

…did I mention this would get complicated?? ☺

Page 18: Analysis of Permanent Gases: More Challenging Than You

PLOT-Q

October 9, 2018 Agilent J&W DB-HeavyWAX18

Page 19: Analysis of Permanent Gases: More Challenging Than You

GasPro - Cryo

October 9, 2018 Agilent J&W DB-HeavyWAX19

Page 20: Analysis of Permanent Gases: More Challenging Than You

ShinCarbon ST

October 9, 2018 Agilent J&W DB-HeavyWAX20

Argon Carrier

Helium Carrier

Page 21: Analysis of Permanent Gases: More Challenging Than You

October 9, 2018 Agilent J&W DB-HeavyWAX21

Courtesy Diablo Analytical

Valves

Page 22: Analysis of Permanent Gases: More Challenging Than You

1 Injection 2 Columns + Valve ….Column Isolation

October 9, 2018 Agilent J&W DB-HeavyWAX22

H2 O2 N2

CH4 CO

CO2 C2’s

C3’s

PLOT-Q

Molesieve

Variable Restrictor

CO / CH4 / N2 / O2 / H2

C3’s / C2’s / CO2

C3’s / C2’s / CO2CO / CH4 / N2 / O2 / H2 /

Sample

Introduction

Page 23: Analysis of Permanent Gases: More Challenging Than You

Column isolation: Setting up valve timing

October 9, 2018 Agilent J&W DB-HeavyWAX23

Page 24: Analysis of Permanent Gases: More Challenging Than You

Column isolation

October 9, 2018 Agilent J&W DB-HeavyWAX24

Page 25: Analysis of Permanent Gases: More Challenging Than You

Column Isolation – Flexibility of Elution Order….

October 9, 2018 Agilent J&W DB-HeavyWAX25

Page 26: Analysis of Permanent Gases: More Challenging Than You

1 Injection 2 Columns in Parallel…Select Permanent Gas Column

October 9, 2018 Agilent J&W DB-HeavyWAX26

Page 27: Analysis of Permanent Gases: More Challenging Than You

Select Permanent Gas Column

October 9, 2018 Agilent J&W DB-HeavyWAX27

Page 28: Analysis of Permanent Gases: More Challenging Than You

Select Permanent Gas Column

October 9, 2018 Agilent J&W DB-HeavyWAX28

Page 29: Analysis of Permanent Gases: More Challenging Than You

So you think that was complicated… what if I need to analyze for hydrogen too…

October 9, 2018 Agilent J&W DB-HeavyWAX29

➢The trouble with hydrogen

➢Must use TCD

➢Sensitivity is carrier gas dependent

➢Creates unique issues related sensitivity

➢Recall the negative hydrogen peak from earlier

Page 30: Analysis of Permanent Gases: More Challenging Than You

TCD

➢ When using He carrier sensitivity for hydrogen is on the order of only ~10%

➢ No problem right? I’ll use Argon or Nitrogen carrier and get down to ppm levels for hydrogen…

➢ This is true, however now sensitivity for all the other compounds is now very poor…a real “catch 22”…

➢ Lets see why this is the case…

October 9, 2018 Agilent J&W DB-HeavyWAX30

Page 31: Analysis of Permanent Gases: More Challenging Than You

Thermal Conductivity Detector - TCD

October 9, 2018 Agilent J&W DB-HeavyWAX31

https://ws680.nist.gov/publication/get_pdf.cfm?pub_id=907540

NameThermal Conductivity @ 400 K

(mW m-1 K-1)

Argon 22.4

Hydrogen 230.9

Helium 189.6

Nitrogen 32.8

Carbon Monoxide 32.3

Carbon Dioxide 25.2

Acetylene 33.3

Ethylene 34.7

Ethane 36.0

Propane 31.0

Butane 28.3

Pentane 24.9

Hexane 23.4

Page 32: Analysis of Permanent Gases: More Challenging Than You

ShinCarbon ST

October 9, 2018 Agilent J&W DB-HeavyWAX32

Argon Carrier

Helium Carrier

Page 33: Analysis of Permanent Gases: More Challenging Than You

Workarounds for low level hydrogen detection

➢ 2 unique injections:

➢ 1 for hydrogen using Argon or N2 carrier 2nd injection for all other gases using helium carrier

➢ 1 injection using column isolation and argon carrier + methanizer➢ TCD coupled to FID

➢ Methanizer placed after TCD but before FID to convert CO/CO2 to methane for enhanced detection of these gases by FID

➢ Sabatier Reaction (1897): CO2 + 4 H2 CH4 + 2H2O

➢Transformer Oil Gas Analyzer (TOGA)

October 9, 2018 Agilent J&W DB-HeavyWAX33

400 ᵒ c

Ni

Page 34: Analysis of Permanent Gases: More Challenging Than You

Conversion of CO/CO2 to methane

➢ Discovered by French chemist Paul Sabatier in 1897 (Sabatier Reaction)

October 9, 2018 Agilent J&W DB-HeavyWAX34

TCD

Methanizer

(H2 + Ni )

350 ᵒ c

FID

Page 35: Analysis of Permanent Gases: More Challenging Than You

October 9, 2018 Agilent J&W DB-HeavyWAX35

Page 36: Analysis of Permanent Gases: More Challenging Than You

Transformer Oil Gas Analyzer (TOGA) Chromatograms

October 9, 2018 Agilent J&W DB-HeavyWAX36

Page 37: Analysis of Permanent Gases: More Challenging Than You

Transformer Oil Gas Analyzer (TOGA) Chromatograms

October 9, 2018 Agilent J&W DB-HeavyWAX37

H2

O2

N2

TCD

Page 38: Analysis of Permanent Gases: More Challenging Than You

Transformer Oil Gas Analyzer (TOGA) Chromatograms

October 9, 2018 Agilent J&W DB-HeavyWAX38

Page 39: Analysis of Permanent Gases: More Challenging Than You

Summary

Analytes Column Technique

H2/O2/N2/CH4/CO Molesieve • 1 Injection 1 Column

H2/O2/N2/CH4/CO + Ar Molesieve (thick film) • 1 Injection 1 Column

H2/O2/N2/CH4/CO + CO2 + C2’s etc Molesieve + PLOT-Q

• 2 Injections 2 Columns

• 1 Injection + Valve (Column Isolation)

• 1 Injection Parallel Columns

(Select Perm Gas Column)

H2/O2/N2/CH4/CO + CO2 + C2’s etc GasPro • Cryo (-80 ᵒ c)

H2/O2/N2/CH4/CO + CO2 + C2’s etc ShinCarbon • Packed only

H2/O2/N2/CH4/CO + CO2 + C2’s etc

+ Low level HydrogenMolesieve + PLOT-Q

• Single injection onto molesieve

for H2 detection only

• Argon carrier + Methanizer (TOGA)

October 9, 2018 Agilent J&W DB-HeavyWAX39

Page 40: Analysis of Permanent Gases: More Challenging Than You

Contact Agilent Chemistries and Supplies Technical Support

1-800-227-9770 Option 3, Option 3:

Option 1 for GC/GCMS Columns and Supplies

Option 2 for LC/LCMS Columns and Supplies

Option 3 for Sample Preparation, Filtration and QuEChERS

Option 4 for Spectroscopy Supplies

Available in the USA 8-5 all time zones

[email protected]

[email protected]

[email protected]

[email protected]

Page 40Agilent Restricted

We are always here to help!

Page 41: Analysis of Permanent Gases: More Challenging Than You

Outline

October 9, 2018 Agilent J&W DB-HeavyWAX41

• Plot columns – General discussion PLOT vs WCOT

• Molesieve

• Very narrow functionality

• Great for resolving O2/N2 at non-cryo temps

• Traps CO2 and water

• The Problem with CO2 and HC larger than C1

• Option 1: Make 2 unique injections

• Option 2: Cryo + GasPro

• Option 3: Column isolation

• Option 4: ShinCarbon

• Detection issues when there is a need for H2

• Explanation of TCD (catch 22)

• Work arounds:

• 2 unique injections

• Argon carrier + Methanizer (TCD + FID in series)

Page 42: Analysis of Permanent Gases: More Challenging Than You

Flow Chart

October 9, 2018 Agilent J&W DB-HeavyWAX42

Y/N

Permanent gases Only

(No CO2 or C2’s +)

MolesieveYes

Molesieve