analysis of the mechanical characteristics of facade …...analysis of the characteristics of facade...
TRANSCRIPT
Analysis of the mechanical characteristics of facade render samples
retrieved in-situ
António Armando Ortiz Soares
Extended abstract
Júri
Presidente: Prof. Augusto Martins Gomes
Orientadora: Profª. Inês dos Santos Flores Barbosa Colen
Orientador: Prof. Jorge Manuel Caliço Lopes de Brito
Vogal: Doutora Maria do Rosário da Silva Veiga
Março de 2011
Analysis of the characteristics of facade samples retrieved in-situ
1
1. Introduction and objectives
Coating mortars (renders) perform the function of protecting the walls from the degradation agents to which
they are subjected. It becomes necessary, then, to evaluate their service performance, so as to understand if
they can carry out the function attributed to them. For this purpose, the chemical, physical and mechanical
properties of the render under analysis must be studied.
In what concerns the mechanical characteristics of the renders, there are still some difficulties in their
assessment, for which reason the present work focuses on this matter, with particular detail on the evaluation
of the compressive strength of samples retrieved in-situ. In this manner, this work allows to deepen the
knowledge on the parameters to be considered in the compressive strength tests of render cores, with the
main objective of relating it with the compressive strength of normalised test specimens
2. Mechanical characteristics of coating renders
Several authors have tried to evaluate the compressive strength of coating mortars from samples retrieved in-
situ. In order to study the compressive strength of irregular samples of historic mortars retrieved in-situ, a
project was developed by LNEC (Magalhães & Veiga, 2006) in order to create a simple adaptation of the
method of determining the compressive strength defined by EN 1015-11 (CEN, 1999), to allow compressive
strength test of irregular samples in old buildings. The adaptation in question consists of producing a
confinement mortar, with a strength superior to that of the sample being tested, applied over the two parallel
faces of the irregular sample (Figure 1).
Figure 1 - Detail of samples prepared with the confinement mortar over the two more parallel faces of the samples
(Magalhães & Veiga, 2006)
As a means to evaluating the method, reference specimens were used, with dimensions 20 x 40 x 80 [mm3],
resulting from the cutting of specimens with 40 x 40 x 160 [mm3], to which the confinement mortar was
applied for the compressive strength test. The values of the compressive strength obtained with these
specimens were compared to those resulting from the testing of normalised specimens, and a high correlation
coefficient (R2 ≈ 0.96) was achieved for a linear relation.
Flores-Colen (2009) also studied the compressive strength of samples of render by comparing the
compressive strength of square cores with 50 mm length and 15 mm height retrieved in-situ, resulting from
the pull-off adhesion test with the compressive strength obtained through the testing of normalised
specimens of the same product, relating the obtained strengths through a power trend line with a correlation
coefficient R2 ≈ 0.78.
However, the results of the compressive strength of cores after the adhesion tests displayed variation
coefficients between 25-51% in some products, which the author attributes to the state of the core after the
process of heating to remove the metallic disk, to the control of the compressive strength testing machine, to
eventual variations of the thickness of the cores or to the sample’s state of degradation.
Extended abstract
2
The same author studied the possibility of directly relating the value of the compressive and tensile strength
with the adhesion strength. This study resulted in the graphic presented in Figure 2, where satisfactory
correlation coefficients are found for the relationship between adhesion strength and strength of normalised
specimens of the same products.
Legend: Rt - flexural tensile strength of normalised specimens; Rc - compressive strength of normalised specimens; Rc
ad -
compressive strength of samples after the adhesion test; fu - adhesion strength
Figure 2 - Relationship between adhesion strength and strength of the products tested (translated from Flores-Colen, 2009)
In the same study, mechanical tests were performed in campaigns conducted in-situ. It was then possible to
compare the results obtained in laboratory with real cases. In the case of the relationship between pull-off
adhesion strength and compressive strength extrapolated from the values of cohesive fracture (related to the
material’s straight tensile strength), a correlation coefficient of R2
= 0.95 was obtained between the
relationships established in laboratory and those verified in-situ.
Thus, the direct relationship between adhesion strength and compressive strength seems to be a good method
to evaluate the mechanical characteristics of the coating mortars. However, since fracture in the pull-off
adhesion test may be adhesive or cohesive on the substrate, part of the information pertaining to the internal
structure of the coating mortar is lost, since only the lower strength limit being known. As such, this method
is only possible when there are straight cohesive ruptures in the mortar, which occur mostly in “weak”
mortars.
Due to the few existing studies on the evaluation of the strength of mortars from samples of render, studies
related to the behaviour of concrete are referred to as a means to understanding some of the parameters
which may affect the value of compressive strength in test cores.
When a cylindrical concrete specimen is subjected to compression, it tends to expand sideways. However,
there is a friction force at the surface of contact between the plates of the testing machine and the specimen,
leading to lateral compression forces, which are responsible for the formation of a confinement area which
originates a cone at the moment of rupture, as can be seen in Figure 3. When that friction force is eliminated,
the lateral compression forces disappear, and a cracking rupture is reached. However, it is difficult to
eliminate the friction force, for which reason it is regarded as viable to consider the lateral restriction along a
certain length and consequent confinement area (Kim & Yi, 2002; Kim et al., 1998).
Analysis of the characteristics of facade samples retrieved in-situ
3
Figure 3 - Cylindrical concrete specimen subjected to a compression load (Kim et al., 1998; Kim & Yi, 2002)
Chung (1979), cited by Mohiuddin (1995), studied the effect of the h/d1 ratio on the compressive strength of
concrete cores and concluded that confinement, exerted by the plates of the compression machine on the
sides of the core, creates a triaxial state of tension in the concrete that originates an increase in the
compressive strength as the core’s h/d ratio decreases.
In the beginning of the 20th century, Connerman (1925) showed that the compressive strength decreases with
the increase of the h/d ratio of the specimen being studied, tending to infinite, for values of h/d lower than
0,5. The same author also defined an interval for the h/d ratio in cylinders, between 1,5 and 2,5, where the
difference in strengths is not significant, having as reference a specimen with h/d ratio = 2.
More recently, other researchers, while studying the effect of the relationship between height and diameter or
lateral dimension, also reached curves with power trend lines, as is the case of Leonhardt & Mönning (1977),
cited by Costa & Appleton (2008), as represented in Figure 4 where the reference specimen is the cube, for
which reason the reference value (of one) occurs for the geometric relation h/d = 1.
Figure 4 - Relation between h/d and the compressive strength of concrete specimens (translated from Leonhardt & Mönning,
1997, cited by Costa & Appleton, 2008)
When references are made between specimens produced with concrete identical to that used on site, their
strength will be different from the one determined in-situ, even if the possibility of achieving perfect cores,
identical in size to the test specimens, is considered, due to divergences in compacting and curing
(Indelicato, 1997).
Therefore, the need to estimate the strength of normalised specimens from cores retrieved in-situ emerged.
Thus, the concept of estimating the actual or in-situ strength referred to cubic specimens was used when
values of strength resulting from cores are used for structural calculations to represent the strength of the
existing structure, and the concept of potential strength estimate is used when there are doubts about the
quality of the concrete used (Concrete Society, 1976).
1 The h/d ratio defines the slenderness of the test specimen through the relation between height (h) and diameter or lateral dimension
(d)
Extended abstract
4
Thus, Equation 1 can be used (Concrete Society, 1976; BSI, 1983) to estimate concrete’s strength from cores
retrieved in-situ.
Eq. 1
Where:
Estimated strength [N/mm2];
Measured strength of the concrete core [N/mm2];
h/d ratio between the height (h) and the lateral dimension or diameter of the concrete core;
Experimental constant which depends on the type of estimate (D = 2.5 for the estimate of the actual
strength of a cubic specimen from a core retrieved horizontally in relation to concreting).
Another way of relating the strength of a normalized concrete specimen with a non normalized one is
through Equation 2, obtained through the theory basis of the non linear mechanical fracture of concrete,
presented by Kim et al. (1998) for cylindrical specimens and not taking into account the concrete’s strength
class, nor any restrictions on the maximum dimension of the aggregate used.
Eq. 2
Where:
Compressive strength of normalised cylinders [N/mm
2];
Compressive strength of a non normalised cylinder [N/mm2];
Cylinder height[mm];
Cylinder diameter [mm];
Ratio between height (h) and lateral dimension or diameter (d).
3. Characterisation of the experimental work
In the experimental phase, the following were applied over bricks: traditional multi-layer renders
(spatterdash, under coat and final coat), traditional control renders (a layer with the composition of the under
coat of a multi-layer render) and industrial renders with the thicknesses presented in Table 1.
Table 1 - Thicknesses of render
Type of
render
Traditional render (mm) Industrial
render
(mm) Spatterdash
Under
coat
Final
coat
Type 1 3 15 6 -
Type 2 3 25 6 -
Type 3 - - - 15
Type 4 - - - 30
Type 5 - 15 - -
Type 6 - 25 - -
For the production of traditional render, river sand was chosen as aggregate, and CEM II/B-L 32.5 N cement
was used as binder, presenting the compositions indicated in Table.
Analysis of the characteristics of facade samples retrieved in-situ
5
Table 2 - Characteristics of each layer of the traditional render used
Layers Ratio (volume) Water/cement ratio
Spatterdash 1:2 0.70
Under coat 1:4 1.00
Final coat 1:4.5 1.06
In the case of industrial render, a mineral regularization render was used, with the characteristics presented in
Table 3, to display a compressive strength value (5.06 N/mm2) close to the average indicated by producers of
general use industrial mortars (Flores-Colen, 2009).
Table 3 - Characteristics of the industrial mortar used, adapted from Flores-Colen (2009)
Components Cement (type I)
[%]
Water repellent
[%]
Water retainer and plasticiser
(cellulose ether) [%] Water/product ratio
Type Binder Admixture Admixture Characteristic
Quantity 15 - 25 0.10 – 0.50 0.05 – 0.10 0.17
Reference specimens (Figures 5 and 6) with the same characteristics of the renders applied (as a way of
simulating cores) and normalised cores, were used.
Figure 5 - Moulds with the under coat in fresh state and
spatterdash in hardened state
Figure 6 - Final aspect of a reference specimen
Upon application of the plaster and production of the specimens, the samples were stored in a curing
chamber of controlled environment up to the test date, with a temperature of 20 ± 2 ºC and relative humidity
of 65 ± 5 %, according to EN 1015 - 11 (CEN, 1999). For the initial curing, polyethylene bags were resorted
to for the durations indicated in Table 4.
Table 4 - Initial curing time in polyethylene bags
Type of
product or
render layer
Types 1 and 2 in brick or
reference specimens Types 3 and 4
in brick or
reference
specimens
Types 5 and 6
in brick or
reference
specimens
Specimens with
normalised size
Spatterdash Under
coat
Final
coat
Traditional
mortar
Industrial
mortar
Initial
curing
(days)
14 11 3 11 11 11 11
Compressive strength tests were performed at 7, 14, 28 and 90 days. The flexural strength test was
performed to obtain 2 half normalised specimens (NS) for compressive strength tests for each NS produced,
and the pull-off adhesion test to obtain cores for the compressive strength test, using a heating process to
Extended abstract
6
remove the pull-off metallic disk. Table 5 presents the general distributions of tests by age, according to the
type of section, specimen and products used.
Table 5 - General distribution of the compressive strength tests by test age
Average number of compressive
strength tests
by test age
CB RS NS
Total Tr
Id
Tr
Id T
r I
d
Section Ml S
l M
l S
l
Round d= 50 mm 12 6 12 8 8 8 - - 54
Square 40 x 40 mm2 8 4 8 8 7 8 6 6 55
Square 50 x 50 mm2 4 2 4 8 8 8 - - 34
Total 24 12 24 24 23 24 6 6 143
Legend: CB - cores; RS - reference specimens; NS - normalised specimens; Tr - traditional render; Id - industrial render; Sl - single-
layer render; Ml - multi-layer render
Some of the tests had to be repeated, and it was not possible to use a few of the cores or comparison
specimens for the compressive strength test, due to their being damaged during the process of removal of the
metallic disk (in the case of the cores) or unmolding (in the case of the reference specimens). Thus, the total
quantity for the 4 testing ages was 678 compressive strength tests, from which 649 were valid, a total of 334
cores having been produces, from which a total of 237 valid adhesion tests were obtained.
4. Analysis of results
Due to the need of using the pull-off adhesion test to extract render cores, the hypothesis of comparing
adhesive strength to compressive strength, and considering the existence of different geometric shapes of
cores (round and square), it was considered important to evaluate the influence of the crosss-section in the
pull-off adhesion test. In this respect, a reference (Figure 7) between the adhesive strength of round cores
with 50 mm of diameter and square ones (with 40 and 50 mm), obtained from the same brick and with the
same type of rupture (cohesive fracture in the substrate), was made, in order to restrict the variables. It is
concluded that the shape of the cross-section does not significantly influence the results as long as the area is
equivalent.
Square section Relationship Correlation coefficient
40 x 40 mm2
50 x 50 mm2
Legend: fuRound - adhesive strength of cores with round section; fu
Square - adhesive strength of cores with square; T17 - designation of
the set of cores from the same brick excluded from the relation
Figure 7 - Relationship between adhesive strength of round and square cores of industrial render with cohesive fracture in
the substrate
Analysis of the characteristics of facade samples retrieved in-situ
7
At 14 days, for the industrial render, it was possible to obtain a linear relation between the adhesive strength
(fu) and the compressive strength of cores (RcCB
) with R2 = 0.86 which is close to that found by Flores-Colen
(2009) for the relationship between adhesive strength (fu) and compressive strength of normalised specimens
(RcNS
), as seen in Figure 8.
Reference Relation Correlation coefficient
Present study, at day 7
Present study, at day 14 0
Flores-Colen (2009) 0
Flores-Colen (2009) 0
Legend: fu - adhesive strength; RcCB - compressive strength of cores; Rc
NS - compressive strength of normalised specimens; RtNS -
flexural strength; T,45,2 e T,45,5 - designation of the cores tested at 14 days excluded from the relationship
Figure 8 - Relationship between adhesive strength and compressive strength of industrial render
In order to analyze the viability of producing reference specimens (RS), with the same product and similar
size to those of cores extracted from render applied on brick (CB) for comparison in the compressive
strength test, the specimens were organised by groups of thickness and lateral size, so as to assess the
relationship between cores (CB) and reference specimens (RS) for each type of product and section (square
or round). Relationships with good correlation coefficients were obtained, which indicates the potential for
using reference specimens (RS) to study of compressive strength, especially for traditional multi-layer
render.
Product Relationship Correlation coefficient
All
Types 1 and 2
Types 5 and 6
Types 3 and 4
Legend: RcCB - compressive strength of cores; Rc
RS - compressive strength of reference specimens; Types 1 and 2 - traditional
multi-layer render with a under coat 15 and 25 mm thick respectively; Types 5 and 6 - traditional single-layer render 15 and 25 mm
thick respectively; Types 3 and 4 - industrial render 15 and 30 mm thick respectively
Figure 9 - Relationship between compressive strength of cores and square reference specimens
0
10
20
30
0 10 20 30 40
RcRS[N/m
m2 ]
RcCB [N/mm2]
Types 1 e 2
Types 5 e 6
Types 3 e 4
Linear (All)
Linear (Types 1 e 2)
Linear (Types 5 e 6)
Linear (Types 3 e 4)
Linear (y=x)
Extended abstract
8
To study the possibility of evaluating the compressive strength of traditional multi-layer render through that
obtained for traditional single-layer render (produced with the same mix and size used at the under coat of
the multi-layer render), a comparison between the strength obtained with two products both for cores (CB)
(Figure 10) and reference specimens (RS) was made.
Age Relationship Correlation coefficient
All
7 days
14 days
28 days
0
90 days
All w/o 7 days
0
Legend: RcMl - compressive strength of traditional multi-layer render; Rc
Sl - compressive strength of traditional single-layer render
Figure 10 - Relationship between compressive strength of cores of traditional multi- and single-layer render
Analysing the relationship between the strength obtained with single- and multi-layer render cores, it was
found that the 7-day relationship has a slope of around 0.87 (which is close to y=x), with a correlation
coefficient R2 = 0.76 (Figure 10), while at the remaining ages, with similar correlation coefficients, the slope
is close to 3, except for the results obtained at 28 days which present a slope of 4.2 with R2 ≈ 0.38.
The traditional multi-layer render tested at 7 days was produced without applying of the final coat,
possessing only spatterdash and under coat. This particularity of the cores tested at 7 days, together with the
fact that this age in which the slope closest to y=x was obtained, with a good R2, may indicate that the
spatterdash (the layer with the strongest ratio of cement) does not significantly influence compressive
strength.
Analysing the relation between the strength obtained with reference specimens of traditional single- and
multi-layer renders, it was also found that the trend line with the slope closest to y=x is that determined from
the results obtained at 7 days, for the same reasons referred in this type of relationship in cores.
However, while good correlations were obtained analysing the relationship between single- and multi-layer
renders for cores (CB) and reference specimens (RS) treated separately, low correlation coefficients were
obtained while trying to relate cores of traditional multi-layer render with reference specimens of traditional
single-layer render.
5
10
15
20
25
30
35
40
5 10 15 20 25
RcSl[N/m
m2 ]
RcMl [N/mm2]
7 days
14 days
28 days
90 days
Linear (All)
Linear (7 days)
Linear (14 days)
Linear (28 days)
Linear (90 days)
Linear (y=x)
Linear (All w/o 7 days)
Analysis of the characteristics of facade samples retrieved in-situ
9
In evaluating compressive strength as a function of the h/d ratio, it was chosen, based on studies already
conducted2, to use a strength index (I
S), which translates the increase in strength in each item in relation to
the strength obtained for NS of the same product, indicated in Equation 3.
Eq. 3
Where:
IS Strength index;
RcCBorRS(Measured)
Measured compressive strength of cores or reference specimens [N/mm2];
RcNS(Measured)
Average of the compressive strengths of normalised specimens [N/mm2].
Thus, it is convenient to consider the compressive strength obtained in the normalised specimens (Table 6).
Table 6 - Measured compressive strength of normalised specimens (NS)
Compressive strength of
normalised specimens (NS)
[N/mm2]
Age [days]
7 14 28 90
Traditional render 8.8 10.6 10.8 13.1
Industrial render 4.7 5.2 7.6 9.3
In the present study, as indicated in Figure 11, it was possible to obtain a relation of the same kind of those
observed by Gonnerman (1925) and by ASTM (1992), cited by Tokyay & Özdemir (1997).
Type of plaster Relation Correlation coefficient
All types
Id Industrial
Tr Multi-layer
Single-layer
Legend: IS - strength index; h/d - ratio between height (h) and diameter (d); Id - industrial render; Tr - traditional render
Figure 11 - General relationship between h/d and strength index (IS)
2 Tokyay & Özdemir (1997) and ASTM (1992) cited by the same author; Kim et al. (1998); Kim & Yi (2002); Gonnerman (1925);
Leonhardt & Mönning (1997) cited by Costa & Appleton (2008)
0
2
4
6
8
10
12
0 0,1 0,2 0,3 0,4 0,5 0,6 0,7 0,8 0,9 1 1,1
IR
h/d
Traditional multi-layer
Traditional single-layer
Industrial
Potencial (All types)
Potencial (Traditional multi-layer)
Potencial (Traditional single-layer)
Potencial (Industrial)
Extended abstract
10
However, bearing in mind the differences between the correction factors obtained in the various studies
(including the present work), it is concluded that in all of them it is possible to describe the relationship
between h/d and IS through a power trendline.
In the analyses made of this parameter, good correlation coefficients were registered, since no significant
differences were found in the trend lines with changes to the geometry of the cross-section (square or round),
the type of specimen (core or comparison specimen), age and type of product.
It was observed in the same analysis that, for values of h/d < 0,4, independently of the type of product or the
type of specimen (core (CB) or comparison specimen (RS)), the value of compressive strength (Rc) increases
significantly, displaying high oscillations in the value of Rc for small variations of h/d. Thus, the use of
specimens with h/d ≥ 0,4 is advised in the study of compressive strength of CB or RS. Thus, if a specimen
with for an example 50 mm of diameter is used, it is suggested that it is at least 20 mm of thick.
Higher values of Rc were observed in normalised specimens with (h/d = 1) than in CB or RS (with h/d < 1),
when Rc tends to decrease with the increase in h/d. This may be related to the different methods of
compacting used according to the type of specimen produced.
Studying the expressions used to estimate the compressive strength (Rc) of concrete, applied in the estimate
of Rc of cores of mortar, it was found that they do not supply reliable results. However, it is considered
necessary to ascertain if this occurred due to their application to the study of mortar or to the fact that the
specimens of mortar used display a ratio of h/d < 1, while the expressions were deducted for specimens with
h/d ratio between 1 and 2.
5. Conclusions
According to the study conducted, it was possible to establish which parameters affect compressive strength
of cores of render retrieved from building façades. These are succinctly presented in Table 7.
Table 7 - Parameters which may or may not affect compressive strength
Parameters Degree of
influence
Ratio between height and lateral dimension (h/d) (strength increases with decrease in h/d) (++)
Number of layers (depends on the constitution of each layer) (+)
Spatterdash (layer in the traditional multi-layer render with the strongest ratio in binder) (--)
Under coat (main layer of the traditional multi-layer render) or type of product in the case of industrial
render (++)
Final coat (layer in the traditional multi-layer renderer with the weakest ratio in binder) (++)
Age (++)
Geometry of the section (--)
Test machine (++)
Heat for removal of the metallic disk (-)
Existence of remains of brick in the surface of the brick (-)(r.d.)
Legend: (++) - clearly affects; (+) - affects; (-) - does not affect; (--) - clearly does not affect; (r.d.) - requires further development
The comparison of the compressive strength of cores or reference specimens with that of normalised
specimens through a strength index indicating relative strength, as occurs in concrete, proved to be a good
way of evaluating strength. Thus, and considering the values of IS close to 1 obtained as a function of the h/d
ratio, shapes and sizes of the cores to use, according to the thickness and type of render to apply, are
suggested in Table 8.
Analysis of the characteristics of facade samples retrieved in-situ
11
Table 8 - Recommended cross-sections of cores
Recommended cross-sections of cores
Render thickness [mm]
Traditional (multi-layer) Industrial
Minimum Maximum Minimum Maximum
Square 40 x 40 mm
2 20 31 19 31
50 x 50 mm2 25 38 24 39
Round D = 50 mm 26 40 26 39
Considering the different methods of evaluating the compressive strength of cores retrieved from rendered
façades, and their potential, Table 9 was prepared, wherein the methods advised for each type of render to
analyse are indicated.
Table 9 - Methods to estimate compressive strength of cores retrieved in-situ
Types of render Methods
RS NS fu MCo CBSl
Render of weak mortars, or of advanced state of degradation - - - ‡
(n.d.) -
Traditional multi-layer independently of the type of fracture in the pull-off test † - - -
Industrial
render
Cohesive fracture in the pull-off test ◘ - -
Adhesive or cohesive fracture in the pull-off test - - -
Legend: RS - comparison with reference specimens with the same number and constitution of layers; NS - comparison with
normalised specimens; fu - relationship with adhesive strength; MCo - use of confinement mortars (Magalhães & Veiga, 2006); CBSl
- comparison with single-layer; - advised method; - method which may be used; † - the finishing layer conditions the
compressive strength; ◘ - only if the h/d ratio is high (greater than 0.4); ‡ - weak mortars can sustain damage upon removal of the
metallic disk of the pull-off test; (n.d.) - not deal with in this study
While reference specimens are the most indicated to evaluate the compressive strength only of traditional
multi-layer render, they may contribute to future studies as a means of defining aspects or parameters which
affect compressive strength without the need of cores.
Thus, the use of reference specimens for the study of compressive strength presents the following
advantages:
Variables introduced by the base (such as water absorption) are eliminated;
A better control of the thickness of layers is achieved (than in the case of render applied on brick,
where the latter’s irregularities may affect the control of the thickness);
Specimens do not need to go through the process of removal of the pull-off adhesive test’s metallic
disk (through exposure to heat);
Specimens do not require correction of the surface of contact with the plates of the compression
machine;
Greater ease of transport and storage in laboratory due to the lower weight, and due to occupying
less space than bricks or others.
Due to the elimination of a number of variables such as contact with the brick’s surface, or because the
process of compacting of coating mortars applied on a surface differs from that used in reference specimens,
these do not rigorously represent the cores retrieved in-situ.
Extended abstract
12
References
ASTM (1992) - “ASTM C42 - Standard method of obtaining and testing drilled cores and sawn beams of
concrete. Philadelphia: American Society for Testing and Materials.
BSI (1983) - “Testing concrete - Part 120. Method for determination of the compressive strength of concrete
cores”. BS 1881. London: British Standards Institution.
CEN (1999) - “Methods of test for mortar for masonry - Part 11: Determination of flexural and compressive
strength of hardened mortar”. EN 1015-11:1999 (Ed.1). Brussels: Comité Européen de Normalisation.
Chung, H. (1979) - “Effect of length/diameter ratio on compressive strength of drilled concrete core - a semi
rational approach”. Cement, Concrete and Aggregates 1 (2), pp. 68-70.
Concrete Society (1976) - “Concrete core testing for strength - technical report. 11”. London: The Concrete
Society.
Costa, A.; Appleton, J. (2008) - “Estruturas de betão I. Parte II - Materiais”: IST, 65 p.
Flores-Colen, I. (2009) - “Metodologia de avaliação do desempenho em serviço de fachadas na óptica da
manutenção predictiva”. PhD Thesis in Civil Engineering: IST, 487 p.
Gonnerman, H. (1925) - “Effect of size and shape of test specimen on compressive strength of concrete”.
ASTM Proc. 25, Part II, pp. 237-250.
Indelicato, F. (1997) - “Estimate of concrete cube strength by means of different diameter cores: A statistical
approach”. Materials and Structures 33 (3), pp. 131-138.
Kim, J.; Yi, S. (2002) - “Application of size effect to compressive strength of concrete members”.
SĀDHANĀ - Academy Proceedings in Engineering Sciences 27 (4), pp. 467-484.
Kim, J.; Yi, S.; Park, C.; Eo, S. (1998) - “Size effect on compressive strength of plain and spirally reinforced
concrete cylinders”. ACI Structural Journal 96 (1), pp. 88-96.
Leonhardt, F.; Mönning, E. (1977) - “Construções de concreto, Vol. 1: Princípios básicos do
dimensionamento de estruturas de concreto armado”. Rio de Janeiro: Interciência, 305 p.
Magalhães, A.; Veiga, M. (2006) - “Avaliação da resistência de amostras de argamassas históricas recolhidas
em obra - Metodologia e correlação de resultados”. Report 259/06 – NRI, Lisbon: National Laboratory of
Civil Engineering (LNEC), 34 p.
Mohiuddin, N. (1995) - “Use of small diameter cores in evaluation of in-situ concrete strength in Eastern
Saudi Arabia”. MSc Dissertation in Science in Civil Engineering. Dharan: King Fahd University of
Petroleum & Minerals, 95 p.
Tokyay, M.; Özdemir, M. (1997) - “Specimen shape and size effect on the compressive strength of higher
strength concrete”. Cement and Concrete Research 27 (8), pp. 1281-1289.