anexo cy688 series oxygen analyzer instruction manual
TRANSCRIPT
CY688 Series Oxygen Analyzers
Instruction Manual
To learn more information, call or E-mail today! Shanghai ChangAi Electronics Co., Ltd
Oxygen Analyzer Instruction Manual
Contents
1. Introduction ................................................................................................................................................................ 1
1.1 Instrument ................................................................................................................................................................ 1
1.2 Sensor ....................................................................................................................................................................... 1
1.3 Application Range .................................................................................................................................................. 1
2. Technical Specifications ......................................................................................................................................... 1
3. Installing the Analyzer ............................................................................................................................................. 3
3.1 Open the package and check with eyes ........................................................................................................... 3
3.2 Instrument Installation .......................................................................................................................................... 3
3.3 Connection Ports.................................................................................................................................................... 4
3.4 Sensor Installation ................................................................................................................................................. 5
3.5 The Connection of Sampling Gas Paths .......................................................................................................... 5
4. Debugging .................................................................................................................................................................. 7
4.1 Turn on the Instrument.......................................................................................................................................... 7
4.2 Program Design ...................................................................................................................................................... 7
4.2.1 Set Alarm Points .................................................................................................................................................. 7
4.2.2 Set Alarm Mode.................................................................................................................................................... 8
4.2.3Return to Standard Measuring Mode .............................................................................................................. 9
4.2.4 Set Output Value................................................................................................................................................ 10
4.2.5 Calibration............................................................................................................................................................11
5. Setting Alarm Value................................................................................................................................................ 12
6. Spare Parts and Maintenance .............................................................................................................................. 13
7. Transportation and Storage ................................................................................................................................. 13
7.1 Transportation ....................................................................................................................................................... 13
7.2 Storage .................................................................................................................................................................... 13
8. A complete set of the instrument ....................................................................................................................... 14
Oxygen Analyzer Instruction Manual
Notices
1. Please read this manual prior to using the analyzer to insure safe and proper operation.
2. Installation and maintenance should be undertaken by the qualified person.
3. To reduce the electrical noise, all signal circuits should be far away from cable with strong electricity
and relay with big power.
4. Make sure the voltage of power supply is complied with the analyzer’s voltage, and electrical-wire is
connected correctly, the power can be switch on.
5. DO NOT use this analyzer in that occasion where exists flammable or explosive gas.
6. The contents of this instruction may be modified without further notification.
Oxygen Analyzer Instruction Manual
CY688 Series Oxygen Analyzers Instruction Manual
1. Introduction
1.1 Instrument
CY688 series oxygen analyzers are a kind of intelligentized instruments in which the microprocessor plays a key
role. An analyzer has a programmable alarm points, one programmable electric current (4-20mA) output port, and
one RS232 data communication port (optional part). According to the different sensors mounted in the oxygen
analyzer, it can accurately measure the oxygen concentration from 100ppm to 100% in nitrogen and oxygen
mixture gas.
1.2 Sensor
The sensor used in the oxygen analyzer is the imported solid ion current oxygen sensor, which has such
remarkable characteristics as good interchangeability, high test precision, good stability, long calibration cycle,
small volume, low power consumption and temperature compensation. The sensor can be classified as the
following models according to the measuring range.
Model Measuring Range PCH 100ppm -25.00%
PCL 10.00-99.99%
1.3 Application Range
CY688 series oxygen analyzers are suitable for measuring the oxygen concentration in nitrogen and oxygen
mixture gas or the air. And they can be used in several industries widely. For example, in air separation plant; in
medical and health industry (such as in high pressure oxygen warehouses, baby incubators, life-support machines,
anaesthesia machines); in chemistry and smelting industries (such as in the protection gas of semiconductors and
in the gas sent into high temperature smelting furnaces). And it can also be used to measure the oxygen
concentration in the process of raising creatures and plants, processing and storing vegetables and food; Besides,
the oxygen analyzer may be used to measure the oxygen concentration of the air in such places as home, naval
vessels, underground control centers, tunnels, deep wells, civil air defense projects, city galleries.
2. Technical Specifications
Monitor: A four-digit dynamic numerical LED monitor
Visible display window: 72mm×25mm.
Page 1 www.ci-ele.com
Oxygen Analyzer Instruction Manual
Measuring Range
According to different measuring ranges, the instrument model, corresponding sensor model and measuring range
are listed in the following table.
Model Sensor Model Measuring Range
688F PCH 100ppm –25.00%
688L PCL 10.00-99.99%
NOTE: The above oxygen analyzers of different models have different measuring ranges; when the oxygen
concentration is bigger than or is equal to 1%, the display resolution is 0.01%. When the display value is beyond
the measuring range of the oxygen analyzer, the degree of accuracy of measuring oxygen concentration won’t get
guaranteed.
The Service Life of Oxygen Sensors
The normal service life is 20000 hours
Sampling Air
The airflow loop between the gas to be measured and the oxygen sensor should be shortened as far as possible in
order to measure accurately; the conduit of the sampling air should not have any leakage. In order to reduce the
influence of any reverse diffusion, the vent hole of the oxygen sensor should be equipped with a vent pipe of at
least 300mm. The gas flow velocity is controlled between 100 ml/min and 300 ml/min; the temperature range of the
sampling gas is from -5 ℃ to +50 ℃ and there’s no dew; (the sensor may be customized to use in the
temperature range of -10℃—+300℃.) The pressure of the sampling gas is from 86kPa to 106kPa.
Analog Output
The instrument has an analog output port 4-20mA. The corresponding output range may be set arbitrarily in the
measuring range of the oxygen sensor
Alarm Output
The instrument has one programmable alarm points, and its alarm points and alarm mode may be set arbitrarily in
the measuring range. The fixed lag value of the alarm points is ±2%, and the instrument has a separate alarm light.
The contact power capacity of the switches is AC 125V 0.3A, DC 110V 0.3A, and DC 30V 1A.
Data Interface (Optional part): a RS232 data communication port.
Page 2 www.ci-ele.com
Oxygen Analyzer Instruction Manual
Power Supply
85V-265V.AC, 50/60Hz, the power loss is smaller than 10VA.
Dimensions: shown in figure 1 and figure 2.
Alarm2
SS-0128
MENU
Alarm1
General Co.,Ltd
ppm
CYC
%
P1001 APTITUDE ANALYZERGeneral
20.62
Fig. 1 Oxygen analyzer Dimension
Fig. 2 View and sizes of the sensor
3. Installing the Analyzer
3.1 Open the package and check with eyes
Please open the package correctly, and long-piece cutters are prohibited when opening the package so as not to
damage the instrument. Check the instrument and make sure that the appearance of the instrument is in good
condition, the oxygen sensor hasn’t been deformed, and that the connection lead is perfect. Then go into the next
installation step.
3.2 Instrument Installation
The instrument should be installed in the place where the ambient temperature is between 0℃ and 50 ℃, the
environment humidity is smaller than the 80%RH, and the oxygen analyzer won’t receive strong electromagnetic
Page 3 www.ci-ele.com
Oxygen Analyzer Instruction Manual
disturbance, strong vibration, and collision. After the installation place of the dial plate have been determined, at
first open a 92mm×92mm foursquare installation hole on the dial plate in advance. Then take down the installation
supports of the instrument and slip the instrument from the front into the installation hole. Install the supports of the
instrument again; at last fasten the installment bolts on the supports. Concrete installation dimensions are shown in
figure 3.
Fig. 3 Installation sketch map of the instrument
and the sizes of the dial plate hole
3.3 Connection Ports
The junction hole of the instrument and sensor, the analog output port 4-20mA, the data communication port,
on alarming switches, and the power supply all lie in the back of the instrument; each connecting wire must be
linked according to the nameplate on the back of the instrument. Various ports and port names are listed in the
following table:
Port Port Name Port Port Name 1 4-20mA anode 9 2 4-20mA cathode 10 BLUE (connecting blue wire) 3 RS232 GND 11 BROWN (connecting brown wire) 4 RS232 TX 12 RED (connecting red wire) 5 RS232RX 13 WHITE (connecting white wire) 6 85-265V.AC power supply L 14 Alarm public contact COM 7 85-265V.ACpower supply N 15 Alarm constant closed contact CN 8 FG 16 Alarm constant open contact ON
The oxygen sensor and the instrument are connected by a 4-core cable. One end of the cable is connected to
SENSOR mark on the back of the instrument; the other end is connected with the oxygen sensor in the analzyer.
Note that the direction of the locating slot in the socket must be in accordance with that of the locating raised
Page 4 www.ci-ele.com
Oxygen Analyzer Instruction Manual
platform; otherwise the instrument will be damaged. The standard length of the connection cable is 30cm, and the
cable within 10 meters may be customized according to users’ need. It should be paid attention to that the cable
length has been determined by debugging the oxygen sensor before leaving factory. So the length can’t be
changed arbitrarily, otherwise measuring errors will come into being. In order to reduce the interference of
electronic noises, the electric cable should be far away from such disturbance sources as electric cables with big
electric current, high power relays, and electric motors. You’d better screen the electric cable. The connection ports
on the back of the instrument are shown in figure 4.
Fig.4 The back view of the instrument
3.4 Sensor Installation
The oxygen sensor should be mounted in the place where the ambient temperature is between 0℃ and 50 ℃, the
environment humidity is less than the 80%RH, and the oxygen sensor won’t receive strong electromagnetic
disturbance, strong vibration, and collision. The oxygen sensor is a solid cylinder on whose foundation there are
three installation holes.
3.5 The Connection of Sampling Gas Paths
The sampling gas sent to the oxygen sensor must be clean and not corrosive. The sample gas with impurity and
corrosive substances can be sent to the oxygen sensor only after having been processed with a filter or chemical
absorbent. The oxygen sensor can’t work in the atmosphere with increasing or reducing gas pressure, and it
Page 5 www.ci-ele.com
Oxygen Analyzer Instruction Manual
should work in stable air pressure from 86kpa to 106kpa, otherwise measuring errors will come into being. The
pipeline connection must be extremely good when low concentration oxygen is measured. The part between the
conduit of ppm measuring components and the air intake of the oxygen sensor must be metal tubing or rigid plastic
tubing. If the flowmeter and air bleed pump are mounted in the air intake of the oxygen sensor, they must be
suitable for processing ppm concentration. The airstream loop between the gas to be measured and the oxygen
sensor should be shortened as far as possible in order to measure accurately and reduce lag time. The gas flow
velocity should be controlled between 100 ml/min and 300ml/min. If the gas path is too long, which results from the
limitation of scene installation conditions, a three-way valve and an exhaust pipe whose length must be more than
300mm may be installed near the air inlet of the oxygen sensor for increasing the gas flow velocity and gas
replacement speed of the gas path. It’s very important to install an exhaust pipe of at least 300mm at the
connection part of the oxygen sensor’s air outlet for preventing the air from reverse pervasion. Because reverse
pervasion may occur even if the sampling gas flows outwards, which results from the difference of oxygen partial
pressure between the sampling gas and the air. All the connecting pipes should use the standard metal pipe whose
diameter is 6mm as far as possible, and the transition tie-in is of sealing rigid connection. Gas paths are connected
according to figure 5, and the connection of the gas paths is finished after all the gas paths have been checked and
been assured that there’s no air leakage.
Fig.
FIG.6 Sampling gas path connection sketch map
Page 6 www.ci-ele.com
Oxygen Analyzer Instruction Manual
4. Debugging
4.1 Turn on the Instrument
When all the parts of the analyzer have been linked, the power may be turned on. After the power has been turned
on, the instrument self-checks the monitor at first, and all the digital tubes and LBD are lightened, twinkling. The
monitor will display the model of the analyzer for 8 seconds, and then the sensor begins to be preheated; when
d288 is counted down to d001, the oxygen concentration of the sampling gas that is supplied to the oxygen sensor
will be read. The analyzer has been calibrated before transportation and can be used at any moment. Concrete
display contents are shown in figure 6.
Fig.8 The model of the analyzer、preheat in manner of cutting down, oxygen concentration display process
4.2 Program Design
4.2.1 Set Alarm Points
When the analyzer is in standard measuring mode, press the “MENU” key for 4 seconds and the analyzer enters
the menu mode. And clicking the up key or the down key can choose the menu options. The menu options are AL
(set alarm value), E (set alarm mode), UP (output upper value, LO (set output lower value), CAL (calibration), and
Page 7 www.ci-ele.com
Oxygen Analyzer Instruction Manual
go—(return to the measuring mode).
When AL is chosen, press the “MENU” key for one second and the analyzer enters the setting mode of alarm value.
Clicking the “CYC” key (to shift digit) may change the twinkling digit of current unit or value, and Clicking the up key
or down key may change the unit or value of current twinkling digit. After the value and unit have been adjusted,
press the “MENU” key for one second, current value and unit are stored, and the analyzer returns to the menu
mode.The program design process of alarm points is shown in figure 7.
Fig. 7 Program design of alarm points
4.2.2 Set Alarm Mode
In the menu mode, clicking the up key or the down key can choose the menu options. When E is chosen (set alarm
mode), press the “menu” key for one second and the analyzer enters the setting mode of alarm pattern. Clicking
Page 8 www.ci-ele.com
Oxygen Analyzer Instruction Manual
the up key, or the down key, or the “CYC” key can change current alarm mode. O represents no alarm, H
represents high alarm and L represents low alarm. After the alarm mode has been adjusted, press the “MENU” key
for one second and current alarm pattern is stored; the analyzer returns to the menu mode automatically. The
program design process of alarm mode is shown in figure 8.
Fig. 8 Program design of alarm mode
4.2.3Return to Standard Measuring Mode
In the menu mode, clicking the up key or the down key can choose the menu options. When the “go—” menu is
chosen, click the “MENU” key, the analyzer exits the menu mode and returns to standard measuring mode. The
program operating process is shown in figure 9.
Page 9 www.ci-ele.com
Oxygen Analyzer Instruction Manual
Fig.9 Return to standard measuring mode
4.2.4 Set Output Value
When the analyzer is in the menu mode, clicking the up key or the down key may choose the UP menu (set output
upper value). Press the “MENU” key for one second, and the analyzer enters the mode of setting output upper limit.
Click the “CYC” key (to shift digit) to change the twinkling digit of current unit or value; clicking the up key or the
down key may change the unit or value of current twinkling digit. After the value and unit have been adjusted, press
the “menu” key for one second, current value and unit are stored, and the analyzer returns to the menu mode. The
setting output value is in the range of 0.00-99.99%. The same method is used to adjust the unit or value of LO
(output lower value). Note that the value of UP cannot be smaller than that of LO; otherwise the output will be
deviant and the unit of the measured value in standard measuring mode will twinkle. Set the output value, for
example, set the value of UP as 95.00% and the value of LO as 20.00%. When the oxygen concentration is smaller
than or equal to 20.00%, the corresponding output of 4-20mA is 4mA. When the oxygen concentration is bigger
than or equal to 95.00%, the corresponding output of 4-20mA is 20mA. When the oxygen concentration is between
the upper limit value and the lower limit value, the output is linear with the input correspondingly. The program
design process of the output range is shown in figure 10.
Page 10 www.ci-ele.com
Oxygen Analyzer Instruction Manual
Fig.10 Set Output range
4.2.5 Calibration
When the analyzer is in the menu mode, clicking the up key or the down key can choose the menu options. When
CAL is chosen, press the “MENU” key for one second, and the analyzer enters the calibration mode. The analyzer
will display “PASS” for 4 seconds, reminding the user to input the password. In this mode, the calibration right
password should be input at first and only if the password is right, can calibration begin, otherwise the analyzer will
return to the menu mode until the password is right. The fixed password of the analyzer is 0918. The input method
of the password is as the following: click “CYC” to change the twinkling digit of current numerical value; clicking the
up key or the down key may change the value of current twinkling digit. When the value has been adjusted as 0918,
press the “MENU” key for one second, and the analyzer enters the calibration state.
Each oxygen sensor is provided with a unique corresponding Pre-Amp processing circuit board for convenient use
and maintenance. The Pre-Amp processing circuit boards of any two oxygen sensors can’t be exchanged
arbitrarily; otherwise the oxygen sensors will be damaged.
Production serial number and the characteristic values of the oxygen sensor are shown on the nameplate of the
Pre-Amp processing circuit board. After the analyzer enters the calibration state, standard gas is sent into the
Page 11 www.ci-ele.com
Oxygen Analyzer Instruction Manual
sensor for about 30 minutes until the sensor is completely stable, and then calibration begins. Click the “CYC” key
(to shift digit) to change the twinkling digit of current unit or value; clicking the up key or the down key may change
the unit or value of current twinkling digit. Input the concentration of standard gas into the analyzer. Press the
“menu” key for 2 seconds, and the analyzer will display the sensor’s output current value for 6 seconds. The unit of
electric current is μA, current value is stored and the analyzer returns to the menu mode. New analyzers have
been calibrated correctly before transportation; you’d better not modify it. When the sensor of the analyzer needs
to be replaced, the analyzer can work normally after being calibrated again. The calibration process of oxygen
sensors is shown in figure 12.
Fig.11 Calibration program design of the analyzer
5. Setting Alarm Value
The analyzer has two programmable alarm points, namely AL1(X)and AL2(X). When the measured value is
bigger than the alarm value, the corresponding alarm indicator lights Alarm1 and Alarm2 will twinkle; the buzzer will
sound, and the corresponding relay will respond. The analyzer has two ways of prompt for errors. The unit
indicator light will twinkle when the setting of the output range of the analyzer is wrong (namely the value of UP is
Page 12 www.ci-ele.com
Oxygen Analyzer Instruction Manual
smaller than that of LO). The output range should be reset; otherwise the output value of the analyzer won’t work
normally. The number displayed on the analyzer will twinkle when the oxygen concentration of the gas to be
measured is bigger than the sensor’ measuring range. You should make sure that the oxygen concentration of the
gas to be measured is the middle value of the measuring range of the oxygen sensor when buying sensors.
Prompt for monitor alarm is shown in figure 14.
FiFig.12 Prompt for monitor alarm
6. Spare Parts and Maintenance
In the whole set of the oxygen analyzer, only the oxygen sensor and its corresponding Pre-Amp processing circuit
board can be replaced, and the design life of all the other parts is 100,000 hours. The instrument should be sent
back to GENERAL Company or the local office for service if there’s something wrong with the instrument. If you
have some questions when ordering the oxygen sensor, please give the serial number and model number of the
product. And the above parameters have been given on the nameplate that lies on the right part of the instrument.
7. Transportation and Storage
7.1 Transportation
The oxygen analyzer that has been well packed may be transported by such vehicles as automobile, airplane,
steamboat and train. And the oxygen analyzer should be protected from rain and snow during transportation. Care
should be taken to put it down carefully and lightly, and avoid heavy pressure when conveying the oxygen analyzer.
7.2 Storage
Products that have been well packed should be preserved in the storehouse where the temperature is -5~40℃,
relative humidity is not bigger than 95%RH at 25℃, and there is no corrosive substances such as acidic matter
and alkaline matter.
Page 13 www.ci-ele.com
Oxygen Analyzer Instruction Manual
8. A complete set of the instrument
1. An analyzer
2. An oxygen sensor
3. A connection cable
Page 14 www.ci-ele.com
Shanghai ChangAi Electronic Science & Technology Co., Ltd
Add: 17J, 1138 ChangAn Road, Zhabei, Shanghai, P.R. China 200070
Tel:021-51068326
Fax:021-51068327
Web:www.ci-ele.com
Email:[email protected]