Angular Response Measurements of a Bentham DMc150 Spectroradiometer

Ewan Eadie1, Hannah Oliver1, Julie Smyth2, Harry Moseley1

1Photobiology Unit, Ninewells Hospital and Medical School, Dundee2 Medical Physics, Ninewells Hospital and Medical School, Dundee

Page 2: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Spectroradiometers & Radiometers

Page 3: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Calibration

Correct for errors Human: Poor calibration procedures, improper maintenance, stray light from poor technique, inaccurate lamp orientationEquipment: Non-linearity, aging, directional response, temperature

Evaluate remaining uncertainty

Page 4: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Spectroradiometer Response

Depends on:Direction of incident radiationTemperature of Photomultiplier TubeUniformity of irradiation

Spectroradiometer used for a range of different source geometriesIdeal Cosine response

Page 5: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Responsef2 error: quantifies the quality of the spectroradiometers directional evaluation of the incident radiation

Pye, S.D. and Martin, C.J.: A study of the directional response of ultraviolet radiometers.

Page 6: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

f2

*2f

( ) ( )( )

( ) εεε

εεϕε

ϕε

ε

dff

RR

freading

reading

∫=

=

−=

=

309.1

0

22

2

2sin(%)

1cos0

,,

Page 7: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Background

Yearly calibration of our Bentham DMc150 spectroradiometerTwo calibrated lamps

Bentham CL3 30W deuterium lampBentham CL2 100W quartz halogen lamp

Two diffusersPTFE diffuserQuartz diffuser

Page 8: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

UV Radiometer Calibration

Deuterium Lamp

Quartz Halogen Lamp

Lamp

Diffuser

PTFE Quartz

200 < λ

< 325

325 < λ

< 600 600 < λ

< 800

-

Page 9: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Response of PTFE and Quartz Diffusers

0

0.2

0.4

0.6

0.8

1

1.2

0 10 20 30 40 50 60 70 80Angle (degrees)

Rela

tive

Inte

nsity

Cosine

PTFE 500nm

PTFE 600nm

PTFE 700nm

PTFE 800nm

Quartz 500nm

Quartz 600nm

Quartz 700nm

Quartz 800nm

Page 10: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Response Measurements

Bentham DMc150 SpectroradiometerPTFE/Quartz diffuserQuartz Halogen LampAngular “Jig”

Distance = 50cmVertical AlignmentHorizontal

±0° - 75°

Vertical ±0° - 70°

Page 11: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Response Measurements

Bentham DMc150 SpectroradiometerPTFE/Quartz diffuserQuartz Halogen LampAngular “Jig”

Distance = 50cmVertical AlignmentHorizontal

±0° - 75°

Vertical ±0° - 70°

Page 12: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Horizontal Measurements

Page 13: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Vertical Measurements

Page 14: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Results

4 readings for each angleHorizontal ±εVertical ±ε

Average R(ε)Relative Response: ( )

( )0=εε

RR

Page 15: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Response of PTFE and Quartz Diffusers

0.000

0.200

0.400

0.600

0.800

1.000

1.200

0 10 20 30 40 50 60 70 80

Angle (degrees)

Rel

ativ

e R

espo

nse

CosinePTFE 500nmPTFE 600nmPTFE 700nmPTFE 750nmPTFE 800nmQuartz 500nmQuartz 600nmQuartz 700nmQuartz 750nmQuartz 800nm

Page 16: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

UV Radiometer Calibration

Deuterium Lamp

Quartz Halogen Lamp

Lamp

Diffuser

PTFE Quartz

200 < λ

< 325

325 < λ

< 600 600 < λ

< 800

-

Page 17: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Response of PTFE and Quartz Diffusers

0.000

0.200

0.400

0.600

0.800

1.000

1.200

0 10 20 30 40 50 60 70 80

Angle (degrees)

Rel

ativ

e R

espo

nse

CosinePTFE 500nmPTFE 600nmPTFE 700nmPTFE 750nmPTFE 800nmQuartz 500nmQuartz 600nmQuartz 700nmQuartz 750nmQuartz 800nm

Page 18: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

PTFE f2f2 error

0

5

10

15

20

25

30

35

(%)

250nm300nm350nm400nm450nm500nm550nm600nm650nm700nm750nm800nm

Page 19: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Quartz f2f2 error

0

5

10

15

20

25

30

35

(%)

250nm300nm350nm400nm450nm500nm550nm600nm650nm700nm750nm800nm

Page 20: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Correction Factor CF(ε)

True irradiance underestimatedApply a correction factorDepends on

Wavelength (PTFE)Average between 250nm and 500nm

Range of anglesSource geometry

Page 21: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Correction Factor CF(ε)

Source Range of Angles

Bank of 6 x 180cm lamps ±

to 70°

Single 180cm lamp ±

to 70°

Ninewells UVA1 Bank 8 x 60cm ±

to 35°

Dr Honle Column ±

to 60°

Single 60cm lamp ±

to 35°

Dr Honle Dermalight Ultra 1 ±

to 20°

Deuterium Lamp n/a

QHT lamp n/a

Dr Honle lamp n/a

Page 22: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Angular Correction Factor CF(ε)

Source CF(θ)

Bank of 6 x 180cm lamps 1.07

Single 180cm lamp 1.04

Ninewells UVA1 Bank 8 x 60cm

1.03

Dr Honle Column 1.03

Single 60cm lamp 1.02

Dr Honle Dermalight Ultra 1 1.01

Deuterium Lamp n/a

QHT lamp n/a

Dr Honle lamp n/a

Page 23: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Conclusions

250nm to 750nm: PTFE diffuser better represents a cosine responseResponse with PTFE is wavelength dependentAngular correction factor, CF(ε), included in UV radiometer calibrationsUncertainty in CF(ε) = 0.4%

Page 24: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Conclusions

250nm to 750nm: PTFE diffuser better represents a cosine responseResponse with PTFE is wavelength dependentAngular correction factor, CF(ε), included in UV radiometer calibrationsUncertainty in CF(ε) = 0.4%

Page 25: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Conclusions

250nm to 750nm: PTFE diffuser better represents a cosine responseResponse with PTFE is wavelength dependentAngular correction factor, CF(ε), included in UV radiometer calibrationsUncertainty in CF(ε) = 0.4%

Page 26: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Conclusions

250nm to 750nm: PTFE diffuser better represents a cosine responseResponse with PTFE is wavelength dependentAngular correction factor, CF(ε), included in UV radiometer calibrationsUncertainty in CF(ε) = 0.4%

Page 27: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

References

1. Methods of characterizing the performance of radiometers and photometers. CIE No 53. 1982

2. Methods of characterizing illuminance meters and luminance meters. Performance, characteristics and specifications. CIE No 69. 1987

3. Pye, S. D. and Martin, C. J. A study of the directional response of ultraviolet radiometers: I. Practical evaluation and implications for ultraviolet measurement standards. Phys. Med. Biol. 2000

4. Martin, C. J. and Pye, S. D. A study of the directional response of ultraviolet radiometers: II. Implications for ultraviolet phototherapy derived from computer simulations. Phys. Med. Biol. 2000

Page 28: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing

Acknowledgements

Dr Harry MoseleyProfessor James FergusonPhotobiology Unit, Ninewells Hospital and

Medical School, DundeeChristine Wall, Teresa Goodman, Laura

Crane and Paul Miller