QA-it is the planned $ systematic action that a diagnostic x-ray facility will produe consistently high images with min exposure of patient $ healing arts personnel.
QC-it deals with the technique used in the monitoring (or testing) and maintenance of thetechnical elements or components of an X-ray system.
(CR) Computed Radiography
Digital way of doing general radiography with Conventional X-ray machines
is a process of capturing radiographic data from a conventional X-ray machine and processing the data digitally to produce crisp and high quality radiographic images
Basic Components of CR
M A TRIX LR 3300
DigitizerProcessing Server
ID Tablet
Laser Camera
Cassette withImaging Plate
Preview & ID Station
CR WORKFLOW
M A TRIX LR 3300
Rx
Network
Digitizer
Printing
Identification
Processing server
Exposure
Digital Radiography
Performed by digital X-ray machines with flat panel detectors
INTRODUCTION to QA QC of CR and DR
Direct digital radiography (DR) and computerized radiography (CR) systems are now being used routine for radiography Screening Programme
Routine quality control (QC) is essential to ensure that the equipment is performing as expected meets standards
Recommended routine QC tests for DR systems
Frequency Test Section
Daily Checks on acquisition and reporting monitors
Daily System check Daily Printer checks using test pattern Weekly Check of contrast-to-noise ratio Weekly Image quality tests Weekly Uniformity check Monthly AEC thickness check Monthly Test of reporting monitors
Monthly Mechanical safety and function checks As required Detector flat-field calibration
As required Repeat analysis
As required Printer checks following software upgrade
Recommended routine QC tests for CR systems
Frequency Test Section Daily Checks on acquisition and reporting
monitors Daily System check Daily Printer checks using test pattern Weekly Check of contrast-to-noise ratio Weekly Image quality tests
Weekly Uniformity check Monthly AEC thickness check Monthly Test of reporting monitors Monthly Mechanical safety and function checks
Six monthly Image plate matching and artefact check
As required Image plate erasure
As required Image plate cleaning As required Repeat analysis As required Printer checks following software upgrade
Artifacts
CR- Dust on the ip, scraches/cracks on ip,dirt on reader,roller marks
DR- faulty individual pixel,cluster of pixel, loss of resolution in one or both direction
MONITOR- CRT(distortion,possibly due to interference from other electrical devices)
PRINTER- if artifacts not seen on hard copy then it is caused by printer, bandling, streaking, fine line in direction of film appear
1111
Artifacts in computed Artifacts in computed radiography radiography
Hardware artifactsHardware artifacts
Soft ware artifactsSoft ware artifacts
1212
Hard ware artifacts Hard ware artifacts
Reasons Reasons dust particle dust particle
&scratches &scratches Damage to the Damage to the
laser beam head of laser beam head of CR systemCR system
Dust on the laser beam head appearas an irregular white lines
Remedy Head has to be cleaned by Camel hair brush by vendor
engineer
Appearance
1313
Damaged laser beam head in CR Damaged laser beam head in CR readerreader
Laser beam headHas to be replaced
Remedy
AppearanceAppears as multiple linear white lineAppears as multiple linear white line
1414
Software artifactsSoftware artifacts
Excessive edge enhancementA B
Artifact simulate acute phase of Paget's disease as a result Of excessive edge enhancement .(B) normal E.E
Computed Radiography and Quality Control Computed radiography still uses
conventional radiographic equipment so basic radiography quality control testing is still needed. This would include: Calibration of the generator Collimation Beam Alignment Linearity of exposure Spatial resolution (focal spot)
MONITOR CHECKS
Monitor checks should be performed on both the acquisition and reporting* monitors under recommended working conditions. It should be noted that cathode ray tube (CRT) and flat panel displays (FPD) may have different types of problems.
Daily checks on acquisition and reporting monitors
Method
• Check for obvious faults such as flicker, gross distortion, artefacts
• Check general condition.• Clean if necessary (follow the supplier’s
instructions).• Look at the text and lines on the screen – are
they sharp and straight?• Keep a record of all checks, note any problems
and take action to get them corrected
Monthly test of reporting monitors Using the SMPTE Test Pattern
1919
SMPTESMPTE
This organization developed a test pattern that can This organization developed a test pattern that can be used to determine grey scale image quality on be used to determine grey scale image quality on videos and computer videos and computer
display monitor.display monitor.
Other parameter which can be evaluated are-Other parameter which can be evaluated are- Spatial resolution Spatial resolution Brightness Brightness ContrastContrast AliasingAliasing
Contt………….. There should be no blurring or bleeding of bright display areas into
dark areas or aliasing of the spatial imaging patterns. All monitors used for interpretation should be tested at least
monthly. For the dynamic range test, both the 5% and 95% contrast areas
should be visible and distinctly different than the adjacent 0% and 100% areas.
Therefore: The brightness and contrast of the monitor is adequate if the 5% squares at both ends of the gray scale are visible.
The gray scale is shown as a series of squares in the center of the image that range from black (0%) to white (100%).
It may be difficult to get the monitor to show both of these inner squares perfectly.
Most monitors do better showing the 95% square than the 5% square.
However, you might try adjusting the ambient light to improve the visibility of the 5% square.
2121
Diagnostic video monitorluminance & uniformity
For Digital Systems: • Display SMPTE test pattern to evaluate initial
quality of monitor Determine manufacturers recommended settings – ensure monitor is set accordingly
• If no guidance material available follow: – Video Monitor Manual Brightness & Contrast
Setting Procedure Diagnostic quality monitors typically include this measurement in calibration check - perform test following vendor recommendations
2222
Luminance and uniformity contd
•Use the TG-18 UNL -80 or SMPTE test Use the TG-18 UNL -80 or SMPTE test pattern ,or set an entirely white displaypattern ,or set an entirely white display
•Use a photo timer to measure luminance Use a photo timer to measure luminance -Measure maximum luminance in the -Measure maximum luminance in the
centre and four corners of displaycentre and four corners of display -Measure minimum luminance in centre -Measure minimum luminance in centre
of displayof display
2323
Diagnostic video monitorluminance & uniformity
•Determine monitor type
•Calculate % difference of each of corners to centre
2424
Monitor type Monitor type Maximum luminance Maximum luminance
Color CRT Color CRT > 80 Cd/m2> 80 Cd/m2
Color LCD Color LCD > 100Cd/m2> 100Cd/m2
Grey scale CRT Grey scale CRT secondarysecondary
> 150 Cd/m2> 150 Cd/m2
Grey scale diagnosticGrey scale diagnostic >200 Cd/m2>200 Cd/m2
Grey scale LCD Grey scale LCD 350 Cd/m2350 Cd/m2
Minimum luminanceMinimum luminance
CRT CRT <1.0 Cd/m2<1.0 Cd/m2
LCD LCD <1.5 Cd/m2<1.5 Cd/m2
% diff% diff
All All With in 15%With in 15%
2525
Image quality Image quality uniformity & artifactsuniformity & artifacts
2626
Image qualityImage quality high contrast (spatial) resolution high contrast (spatial) resolution
For photo stimulable phosphor systemsFor photo stimulable phosphor systems Determine standard set up& technical factors- do once Determine standard set up& technical factors- do once
- low KV- low KVp, p, Extended SIDExtended SID
use a lead –bar resolution test pattern phantomuse a lead –bar resolution test pattern phantom
- slightly angle phantom - slightly angle phantom Minimize moiré &line pairing pattern Minimize moiré &line pairing pattern Process with high-contrast algorithmProcess with high-contrast algorithm
2727
Image qualityImage qualitylow contrast resolution low contrast resolution (delectability)(delectability) photo stimulable phosphor systems:photo stimulable phosphor systems: erase cassetteserase cassettes Obtain several contact images of a low contrast Obtain several contact images of a low contrast
phantom phantom
process using a high –contrast algorithm & edge process using a high –contrast algorithm & edge
enhancementenhancement
algorithm. algorithm.
2828
Image quality low contrast Image quality low contrast resolution (delectability)resolution (delectability)
Weekly check of contrast-to-noise ratio (CNR)
If region of interest (ROI) facility is availableMethod• Use the image of the test object from the daily test• Draw two ROIs • Record the mean (M1) and standard deviation (SD) of the
pixel value in ROI 1.• Record the mean (M2) of the pixel value in ROI 2.• Subtract M2 from M1 and divide by SD to calculate the
signal-to-noise ratio (CNR):
CNR =M1- M2 SDCNR baseline ± 20%If this level is exceeded then take action to correct the
problem.
3030
Reader reboot Reader reboot
Reboot Reboot weekly Or as required weekly Or as required
Follow the schedule as prescribed by Follow the schedule as prescribed by the manufacturethe manufacture
Daily wiping of the insertion slot reduces the chance of dust getting into the readout chamber
3131
Daily wiping of the insertion slot reduces the chance of
dust getting into the readout chamber
Weekly uniformity check
Uniformity should be visually checked on a weekly basis using the image of the test object or plain Perspex
from the daily test. By setting a narrow grey scale window areas of non-uniformity will be seen. Magnify or zoom the image electronically and inspect it in a systematic fashion to look for artefacts such as faulty clusters of pixels or areas of unusually low noise (where the background mottle appears blurred or smoother than otherareas of the image). This is particularly important for some DR systems
Monthly AEC thickness check
If region of interest (ROI) facility is available • Measure SNR and compare with baseline values • Measure CNR and compare with baseline values Remedial levels Detector dose indicator baseline for that thickness ± 10% SNR baseline for that thickness ± 20% CNR baseline for that thickness ± 20% No disturbing artefacts should be visible. If any of the levels are exceeded then take action to correct
the problem.
Set up test just as done for filmSet up test just as done for filmMake Ist exposure on the thinner phantom Make Ist exposure on the thinner phantom - expose thicker phantom at the same setting - expose thicker phantom at the same setting - record the mAs - record the mAs If a printer is available , print both images and measure optical If a printer is available , print both images and measure optical densities.densities.If using the on screen tool , select same size region- of – interestIf using the on screen tool , select same size region- of – interest (ROI) at same location on both images and record average (ROI) at same location on both images and record average measurement.measurement.
DETECTOR FLAT-FIELD CALIBRATION (DR SYSTEMS ONLY) Some DR detectors may have a non-uniform response (due to
variations in sensitivity, faulty pixels etc). Also,there are non-uniformities in the x-ray beam due to the anode heel effect and x-ray beam divergence.
Some DR systems correct for these inherent non-uniformities by a process of flat-fielding. Flat-field correction maps are obtained using a standard beam attenuator (usually a Perspex block) for one or more exposure conditions
(eg different target/filter combinations and focal spot sizes).
Some system require the user to carry out this flat-fielding process periodically, and it is therefore included here although it is not strictly a QC test. On other systems this is carried out by the service engineer at routine service visits.
Method
• Carry out the flat field calibration according to manufacturer’s protocols.• Record and initial that the procedure has been performed.
IMAGE PLATE (IP) CHECKS (CR SYSTEMS ONLY)
Image plate erasure It is recommends that any cassettes left in the
x-ray room, or left un-used for a few days (eg over a weekend)
should be erased using the image reader ‘secondary erasure’ cycle before use, to remove any ‘fog’ from the IP.
Six-monthly image plate matching and artefact check This procedure should also be carried out on new image
plates. Image plates and cassettes should be matched in both their sensitivity (S value per unit exposure) and
in the mAs derived under automatic exposure control. Full size Perspex blocks are preferable for these tests
so that the whole IP can be checked for artefacts.
MethodUsing each IP in turn:• Conduct primary erasure before starting.• Place 4 cm thickness of Perspex on the table (if more convenient, other thicknesses such as 4.5 or 5 cm would also be suitable).• Place IP cassette in the bucky.• Operate the unit in a fixed kV, automatic mAs mode – select kV in normal use, • Make an exposure and record the mAs value.• Read the IP after a standard delay time (eg 1 minute) –
• Record the S value.
• View image with a narrow display window and check for artefacts.
• Repeat for each IP.Then:
• Calculate the mean (average) mAs value for all IPs.• Calculate the mean S value for all IPs.Remedial levelsmAs value mean ± 5%S value mean ± 10% (this allows for variation due to mAs and variation in IP sensitivity).
Plates showing scratches or marks should be cleaned and the test repeated. Plates with permanent scratches or marks should be removed from service
MONTHLY MECHANICAL SAFETY AND FUNCTION CHECKS
Check the safety and function of the system. It is recommended that a local checklist is drawn up for each
system to identify relevant features to be checked (eg items that are safety-critical or areas known to be prone to faults).
the local system, for example: • environmental checks (some digital systems
are particularly sensitive to environmental conditions such as temperature and humidity)
• checks relating to the reporting workstation . Keep a record of all checks, note any problems
and take action to get them corrected
PRINTER CHECKS
The printer is set up on installation when the engineer should ensure that the hard copy matches the soft copy image.
Daily printer checks using test pattern Printer checks should be carried out using
standard viewing conditions each day that the printer is used. Print
SMPTE test pattern or the manufacturer supplied test pattern and perform the following checks:
• Geometrical distortion – check image is printed without geometrical distortion; borders should be completely visible and straight lines should be straight.
• Contrast visibility – SMPTE test pattern, the 5% and 95% squares should be clearlyvisible.
• Printer artefacts – check test pattern for printer artefacts no disturbing artefacts shouldbe visible.• If a densitometer is available, measure densities and compare with baseline values. Printer checks following software upgradeAfter software changes or an upgrade, it may be advisable to print both a test pattern and a clinical image to confirm that the hard copy remains similar to the soft copy display