appendix vi vernier lab technology - cwu
TRANSCRIPT
(Last Accessed 9-5-2019)
APPENDIX VI
Vernier Lab Technology
Part I – LabQuest Mini Software for basic data collection
Plug in a sensor and start collecting data
Adjusting X-axis parameters (Time Based and Keyboard Entry experiments)
Part II – Available Sensors and Operating Features
Sensor I – Vernier Stainless Steel Temperature Probe®
Sensor II – Vernier Gas Pressure Sensor®
Sensor III – Vernier SpectroVis Plus®
Sensor IV – Vernier pH Sensor®
Sensor V – Vernier Voltage Probe®
(Last Accessed 9-5-2019)
Part I – LabQuest Mini Software for basic data collection
Plug in sensor and start collecting data
*Choose any sensor (temperature, gas pressure, pH, and/or voltage), plug into any of three interface channels (CH1, CH2, and/or CH3), and press collect data. For SpectroVis Plus, see Sensor III, below.* 1. For example, connect Vernier’s
Stainless Steel Temperature
Sensor® (temperature sensor) to
the interface (CH 1, CH2, or CH3).
2. Open the data-collection
software.
3. The software will recognize the
temperature sensor and load a
default data collection setup with
time on the x-axis.
4. To begin collecting data press
the button and every
second for 180 data points, as
default setting.
5. When done collecting data,
press the button and a
graph should appear.
Adjusting X-axis Parameters
Should you need to acquire sensor input on y-axis
and adjusting X-axis (time based and keyboard entry):
Time-based Experiments: 1. Select Experiment and then Data Collection.
2. The default for all experiments is set at 2
samples/second for 180 seconds, shown in Figure
3. Adjust the time variable with Duration: and/or
Sampling Rate: and/or Continuous Data Collection to
meet your experimental parameters. When you have
made your choice, click .
4. You will now see the main Logger Pro window, Figure
1, and your x-axis adjustments. Click to begin
your experiment, click once your experiment
is complete, and then save your data immediately.
Figure 1: Temperature sensor data collection interface versus time (180 data points).
Figure 2: Data Collection window with x-axis time variable options for your experiment
(Last Accessed 9-5-2019)
Keyboard Entry Experiments (Events with Entry): 1. Select Experiment and then Data Collection.
2. The default, as before, is a time based experiment
with 2 samples/second for 180 seconds, shown in
Figure 2.
3. To change your experiment from a Time Based
experiment to a numerical keyboard entry, choose
from the drop down menu under Mode and select
Events With Entry, as shown in Figure 3.
4. Label your Column Name with the correct descriptors
and numerical Units that you plan to observe and
keyboard enter into the experiment for data
collection. (ex. If you are performing a titration
experiment with milliliter aliquots of 0.1OM NaOH, it
is important to place the following descriptors and
units in the fields shown in Figure 4.) Once you have
the correct descriptors, then click .
5. You will now see the main Logger Pro window and can
start collecting data by clicking .
6. Once the sensor you have chosen stabilizes in the
lower left hand window, click and an Events
With Entry window, Figure 5, will open and prompt
you to enter a numerical keyboard value. Enter your
numerical value and click .
7. As soon as your experiment is complete, click
and then save your data immediately.
Figure 3: Keyboard enter numbers for your experiment.
Figure 4: Describe data on the x-axis with correct units
Figure 5: Logger Pro window showing Events With Entry for you to enter numerical keyboard values.
(Last Accessed 9-5-2019)
Part II – Available Sensors and Operating Features
Sensor I: Vernier Stainless Steel Temperature Sensor® Data Collection with temperature sensor–
*NOTE, IN MOST CASES THE TEMPERATURE SENSOR DOES NOT NEED TO BE CALIBRATED DUE TO EXTREME CALIBRATIONS PREFORMED BY MANUFACTURERS.*
1. Connect Vernier’s Stainless
Steel Temperature Sensor®
(temperature sensor) to the
interface.
2. Open the data-collection
software.
3. The software will recognize
the temperature sensor and
load a default data collection
setup with time on the x-axis.
4. To begin collecting data
press the button.
5. When done collecting data,
press the button and a
graph should appear. Save
your data immediately.
Accuracy, precision, and extreme environments –
• May be used when trying to obtain an accurate ΔT.
• Acid concentration and maximum exposure time for Vernier’s stainless steel temperature sensor: 1 M HCl 20 min
2 M HCl 10 min
3 M HCl 5 min
1 M H2SO4 48 hours
2 M H2SO4 20 min
3 M H2SO4 10 min
1 M HNO3 48 hours
2 M HNO3 48 hours
3 M HNO3 48 hours
1 M CH3COOH 48 hours
2 M CH3COOH 48 hours
3 M CH3COOH 48 hours
1 M H3PO4 48 hours
2 M H3PO4 48 hours
3 M H3PO4 48 hours
Figure 6: Temperature probe data collection interface.
(Last Accessed 9-5-2019)
Calibration of temperature sensor–
Obtain the Vernier Stainless Steel Temperature Sensor® (temperature sensor), three containers for water
baths at different temperatures (cool, warm and hot) and a glass thermometer. Before you retrieving your
water baths, check to see the temperature range of your experiment so that the cool water bath is slightly
lower and the hot water bath is slightly higher than the experimental range. You want to make sure that
the calibration “contains” your experimental data between the extrema of the calibration points.
Select Experiment and then Calibrate and a window like Figure 2 should appear. Select the Stainless Steel
Temperature Probe. In the sensor setting
window and under the Calibrate tab, shown
above, press Calibrate Now. For each of the
three water samples, place the temperature
sensor alongside the thermometer in the sample
and allow both readings to stabilize. Enter the
temperature reading measured by the
thermometer and press Keep. Repeat this
process for each of the three water samples.
Click after the third water bath reading.
To save the calibration for later:
• Go to the calibration dialog box and select Calibration Storage. Set the Calibration Retrieval
Preference to Experiment file and save it.
After saving, when you open that experiment file, the stored calibration will be used, instead of the normal
calibration stored on the computer for this probe rather than the normal calibration that is stored on the
computer.
For more information about Vernier’s Stainless Steel Temperature Sensor®, please refer to their website:
http://www.vernier.com/products/sensors/temperature-sensors/tmp-bta/
(Section Last Accessed 11-29-2016)
Figure 2: Temperature probe calibration interface.
(Last Accessed 9-5-2019)
Sensor II: Vernier Gas Pressure Sensor®
Data Collection with gas pressure sensor –
*NOTE GAS PRESSURE PROBE DOES NOT NEED TO BE CALIBRATED DUE TO EXTREME CALIBRATIONS PREFORMED BY MANUFACTURERS, JUST SELECT ONE OF THE PREPROGRAMMED CALIBRATIONS.*
Connect the Vernier Gas Pressure
Sensor® (gas sensor) to the interface.
1. Open the data-collection software.
2. The software will identify the gas
sensor and load a default data-
collection setup with time on the x-
axis.
3. Begin collection of data by
selecting the button and a
graph should be generated.
4. When done collecting data, press
the button. Save your data
immediately.
Choosing units of pressure –
Pressure can be measured in many different units. Atmospheres is most commonly used. Below is a list of simple conversions of 1 atmosphere to other pressure units of measurement. To switch between units, select the Data tab and select Column Options and a screen should pop up that allows for you to switch between different pressure units.
1 atmosphere is = 101.325 kPa = 760 mm Hg = 29.92 in. of Hg (at 0°C) = 14.70 psi = 1013 millibar
For more information about Vernier’s Gas Pressure Sensor®, please refer to their website:
http://www.vernier.com/products/sensors/pressure-sensors/gps-bta/
(Section Last Accessed 11-29-2016)
Figure 2: Pressure sensor unit choice interface.
Figure 1: Pressure sensor data collection interface.
(Last Accessed 9-5-2019)
Sensor III: Vernier SpectroVis® Plus
Data Collection with SpectroVis Plus– *NOTE Vernier’s SpectroVis Plus® MUST BE CALIBRATED PRIOR TO USE, REQUIRING A MINIMUM OF 90 SECOND WARM UP TIME.* Step to prepare Vernier’s SpectroVis Plus® for use are as follow:
1. Connect the Vernier SpectroVis Plus® (spectrometer) directly to the USB port of the computer. 2. Start the data-collection software (Logger Pro) and choose New from File menu. The software will
identify the sensor and load a default data-collection setup, Figure 1.
3. The default data type is absorbance. If you want to measure the absorbance of a solution, proceed directly to the Calibration section below. If you want to change the units of measurement to %Transmittance or Fluorescence, or Intensity, then choose Experiment, Change Units, Spectrometer 1: and then select the data type you wish to measure, Figure 2.
4. Do not collect any spectral data until calibration is complete. Go to the next section to learn more about calibration.
Figure 1: Start-up window for Vernier's SpectroVis Plus®.
Figure 2: Changing units of measurement for spectrometer.
(Last Accessed 9-5-2019)
Calibration of SpectroVis Plus– 1. To calibrate the SpectroVis Plus, choose
Calibrate and then Spectrophotometer 1
from the Experiment menu, Figure
3. (Note: For best results, don’t skip the
warm-up period.)
2. Fill 1cm by 1cm cuvette about 3/4 full with
distilled water (or the solvent being used in
the experiment) to serve as the “blank”. After
the Spectrophotometer has warmed up, wipe
clean and place the “blank” cuvette in the
Spectrophotometer. Align the cuvette so the
clear sides of the cuvette are facing the light
source and the detector and press, Finish
Calibration and then OK, Figure 4.
3. Your spectrometer is now calibrated and ready to collect a measurement vs. wavelength data.
Measurement versus. Wavelength: (Absorbance vs. nanometers)
1. Fill a cuvette about 3/4 full of a sample of the solution to be tested (ex. Allura Red). Place the sample
in the Spectrophotometer and press and then to end data collection, Figure 5.
Figure 5: Spectral data of Absorbance versus Wavelength (nm) for Allura Red solution.
2. To store the spectrum data, choose Store Latest Run from the Experiment menu. Save your data
immediately.
Figure 3: Warm up period is necessary for good results.
Figure 4: Finish calibration with deionized water.
(Last Accessed 9-5-2019)
Measurement vs. Concentration data: (Absorbance vs. Concentration - Beer’s Law Plots)
1. Generate a spectrum as described above in Measurement versus wavelength (380-950nm).
2. Click the Configure Spectrophotometer Data Collection button, , and an interactive window
appears, Figure 5.
Collection Mode - The three options for data collection are offered. If the measurement (Absorbance
in this example) vs. Time or vs. Concentration is selected, a wavelength or wavelengths will need to be
chosen.
Graph - The graph displays a full-spectrum analysis of the sample in the cuvette holder. By default, the
wavelength with the maximum measured value will be selected. You may wish to select a different
wavelength. See Step 3 for details.
List of wavelength options - This column lists all the available wavelengths. It becomes active when
either the Concentration or Time mode is selected.
Collection Mode - The three options for data collection are offered. If the measurement (Absorbance
in this example) vs. Time or vs. Concentration is selected, a wavelength or wavelengths will need to be
chosen.
Graph - The graph displays a full-spectrum analysis of the sample in the cuvette holder. By default, the
wavelength with the maximum measured value will be selected. You may wish to select a different
wavelength. See Step 3 for details.
List of wavelength options - This column lists all the available wavelengths. It becomes active when
either the Concentration or Time mode is selected.
Configuring the Spectrometer data-collection dialog box for Absorbance versus Concentration, Figure 6:
3. Select Absorbance (or %T) vs. Concentration as the data-collection mode. The wavelength with the
maximum value from the spectrum (λ max) will be automatically selected (ex. 501.5nm for Allura
Figure 6: Interactive “Configure Spectrometer” window for Vernier Spectrovis Plus®
(Last Accessed 9-5-2019)
Red). There are three options when choosing a wavelength (or wavelengths) for subsequent
measurements.
Option 1 - The default option is to use a single 10 nm band. This measures the average absorbance from ~5 nm on either side of the chosen wavelength. You can change the center wavelength value by clicking on the graph or by choosing a wavelength from the list.
Option 2 - If you wish to use the λ max chosen by Logger Pro, but you want the absorbance to be measured only at that one wavelength, change Single 10 nm Band to Individual Wavelengths. You may then select up to ten wavelengths to measure at the same time.
Option 3 - If you wish to measure an average over a range of contiguous wavelengths of your choice,
change Single 10 nm Band to Individual Wavelengths. Click . Select boxes in the list or drag your cursor on the graph to select up to ten contiguous wavelengths. Check Combine Contiguous Wavelengths.
4. Once you have selected the best option for choosing λ max, click to continue.
5. Click and place your first sample in the cuvette slot of the Spectrophotometer. After the
readings stabilize, click . Enter the concentration of the sample and click .
6. Place your second sample in the cuvette slot. After the readings stabilize, click . Enter the
concentration of the second sample and click .
7. Repeat Step 6 for the remaining samples. When finished, click to end data collection.
8. Click Linear Fit, , to see the best fit line equation for the standard solutions.
9. If performing a Beer’s Law experiment in determination of an unknown concentration, then place the
unknown sample in the cuvette holder. Choose Interpolation Calculator from the Analyze menu. A
helper box will appear, displaying the absorbance and concentration of the unknown. Click .
10. Select Absorbance (or %T) vs. Concentration as the data-collection mode. The wavelength with the
maximum value from the spectrum (λ max) will be automatically selected. There are three options
when choosing a wavelength (or wavelengths) for subsequent measurements.
Option 1 - The default option is to use a single 10 nm band. This measures the average absorbance from ~5 nm on either side of the chosen wavelength. You can change the center wavelength value by clicking on the graph or by choosing a wavelength from the list.
Option 2 - If you wish to use the λ max chosen by Logger Pro, but you want the absorbance to be measured only at that one wavelength, change Single 10 nm Band to Individual Wavelengths. You may then select up to ten wavelengths to measure at the same time.
Option 3 - If you wish to measure an average over a range of contiguous wavelengths of your choice,
change Single 10 nm Band to Individual Wavelengths. Click . Select boxes in the list or drag your cursor on the graph to select up to ten contiguous wavelengths. Check Combine Contiguous Wavelengths.
11. Click to continue.
(Last Accessed 9-5-2019)
12. Click . Place your first sample in the cuvette slot of the Spectrophotometer. After the readings
stabilize, click . Enter the concentration of the sample and click .
13. Place your second sample in the cuvette slot. After the readings stabilize, click . Enter the
concentration of the second sample and click .
14. Repeat Step 6 for the remaining samples. When finished, click to end data collection. Save
your data immediately.
15. Click Linear Fit, , to see the best fit line equation for the standard solutions.
16. If doing Beer’s law to determine the concentration of an unknown, place the unknown sample in the
cuvette holder. Choose Interpolation Calculator from the Analyze menu. A helper box will appear,
displaying the absorbance and concentration of the unknown. Click .
Measurement vs. Time: (Absorbance vs. Seconds – Chemical Kinetics)
1. Generate a spectrum as described above in Measurement versus Wavelength (380-950nm).
2. Click the Configure Spectrophotometer Data Collection button, .
3. Select Absorbance vs. Time as the data-collection mode. The wavelength of maximum absorbance
will be selected. Click to continue or click and select a wavelength on the graph or
in the list of wavelengths. See the previous section for more details on wavelength selection.
4. The default settings are 1 sample per second for 200 seconds. To change the data-collection
parameters for your experiment, choose Data Collection from the Experiment menu and make the
necessary time parameter changes. Click .
5. Mix the reactants. Transfer ~2 mL of the reaction mixture to a cuvette and place the cuvette in the
Spectrophotometer. Click . Click if you wish to end data collection early. Save your data
immediately.
6. Click Curve Fit, , to calculate a function for your data.
Specifications for Vernier SpectroVis Plus®:
For more information about Vernier’s SpectroVis Plus®, please refer to their website: http://www.vernier.com/products/sensors/spectrometers/visible-range/svis-pl/
(Section Last Accessed 11-29-2016)
Support for fluorescence Excitation centered at 405 and 500 nm
Light source Incandescent with LED support
Detector Linear CCD
Wavelength range 380 nm–950 nm
Wavelength reporting interval ~1 nm
Optical resolution 4.0 nm (at 656 nm) to 25 nm (at 486 nm)
Wavelength accuracy ± 3.0 nm (at 650 nm) to ± 7.0 nm (at 450 nm)
Photometric accuracy ± 13.0%
Typical scan time ~2 s
Operating temperature 15–35ºC
(Last Accessed 9-5-2019)
Sensor IV: Vernier pH Sensor®
Data Collection with pH Electrode–
*NOTE Vernier’s pH SENSOR® DOES NOT NEED TO BE CALIBRATED, BUT A TWO-POINT CALIBRATION IS HIGHLY RECOMMENDED.*
Step to prepare Vernier’s pH SENSOR® for use are as follow:
For time-based experiments:
1. Carefully remove the
storage bottle from
Vernier pH sensor®,
including the plastic cap,
and thoroughly rinse the
lower section of the probe
using distilled or
deionized water.
2. Connect the pH sensor to
the interface.
3. Open the data-collection
software.
4. The software will identify
the pH Sensor and load a
default data-collection
setup with time on the x-
axis, Figure 1.
5. Begin collection of data by selecting the button.
6. When done collecting data, press the button and a graph should appear.
Important: Do not fully submerge the sensor to the top. The handle is not waterproof.
Accuracy and precision –
• It is necessary to remove the pH sensor in between experiments and wash the lower section with
deionized water before using this device in another experiment
• For newer pH sensors, it will take approximated 20-25 seconds for the pH reading to stabilize.
• For older pH sensors, it may take over 30 seconds for the pH reading to stabilize.
• Temperature range of pH electrode should remain constant during a pH experiment. The pH
electrode may be used between 5°C to 80°C but recalibration is necessary if temperature changes
appreciable (+/- 10°C) during experiment.
• When not being used, store probe in storage bottle. Never store probe in deionized water.
Figure 1: pH probe data collection interface versus time.
(Last Accessed 9-5-2019)
For Titration Experiments with pH Sensor:
5. Configure the experiment, as described above.
6. Select Experiment and then Data Collection.
7. You should now see a window that is shown in Figure 2.
8. Under the Mode drop down menu, choose Events with
Entry and then press .
9. You should now see a window that is shown in Figure 3.
10. Type a description of the units and concentration of your
titrating analysis solution in the Column Name: field
(ex.mL of 0.1M NaOH) and Short Name: that describes a
basic description (ex.mL of NaOH) and press .
11. Place the pH electrode into the solution that is being
titrated and press to start the experiment. Once
your pH reading has stabilized, press , type in the
total volumetric amount delivered into your titration
experiment, and then press OK, as shown in Figure 4.
12. Your data will be recorded on the spreadsheet, plotted
on the pH versus volume graph, and then you can
proceed to collect the next data point.
Figure 2: Changing x-axis to Events With Entry.
Figure 3: Changing to units of mL.
Figure 4: Hand-entered pH Titration mode
(Last Accessed 9-5-2019)
Two-Point Calibration with pH Sensor–
Obtain the Vernier’s pH Sensor®, two low
volume cuvette containers (~15mL) for two
different buffer calibration solutions (ex. 4.0
and 10.4). Select Experiment and then
Calibrate and a window like Figure 5 should
appear. Select the pH Sensor. In the sensor
setting window and under the Calibrate tab,
shown above, press Calibrate Now. For each
one of the buffer solutions, place the cleaned
pH sensor into the solution, submerge the bulb
and allow the voltage (V) reading to stabilize.
This may take 20-30 seconds. Enter the value of
the pH buffer solution into the Reading 1: Enter
Value in Data Units field and press .
Repeat this process for the second buffer
solution, entering the second value into the Reading 2: Enter Value in Data Units field and click Done. If
you are provided a third buffer, procure a clean ~15mL cuvette and test the accuracy of your calibrated
pH sensor by submersing the tip of the pH sensor into this buffer and observing the number stabilizing in
the lower left hand corner.
To save the calibration for later:
• Go to the calibration dialog box and select Calibration Storage. Set the Calibration Retrieval
Preference to Experiment file and save it.
After saving, when you open that experiment file, the stored calibration will be used, instead of the normal
calibration stored on the computer for this probe rather than the normal calibration that is stored on the
computer.
For more information about Vernier’s pH Sensor®, please refer to their website:
http://www.vernier.com/products/sensors/ph-sensors/ph-bta/
(Section Last Accessed 11-29-2016)
Figure 5: Calibration window for pH sensor.
(Last Accessed 9-5-2019)
Sensor II: Vernier Voltage Probe®
Data Collection with voltage probe –
*NOTE VOLTAGE PROBE DOES NOT NEED TO BE CALIBRATED DUE TO EXTREME CALIBRATIONS PREFORMED BY MANUFACTURERS, JUST SELECT ONE OF THE PREPROGRAMMED CALIBRATIONS.*
Connect the Vernier Voltage Probe® (voltage sensor) to the interface.
1. Open the data-collection software.
2. The software will identify the voltage sensor
and load a default data-collection setup with
time on the x-axis.
3. Begin collection of data by selecting the
button and a graph should be generated.
4. When done collecting data, press the
button. Save your data immediately.
Choosing precision of voltage –
Precision of voltage can be measured to many more decimal places past the zero OR in total significant figures. To adjust your decimal places or significant figures, select the Data tab and select Column Options and a screen should pop up. Click on Options and see the options in Figure 2, to the right.
For more information about Vernier’s Voltage Probe®, please refer to their website:
http://www.vernier.com/products/sensors/voltage-probes/vp-bta/
(Section Last Accessed 11-29-2016)
Figure 1: Voltage probe data collection interface.
Figure 2: Pressure sensor unit choice interface.