Applications of CCD Array Spectrometers in Undergraduate
Education at EIU
Jonathan Blitz and Daniel Sheeran
Department of Chemistry
![Page 2: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/2.jpg)
The traditional educational spectrophotometer
Single beam
Single channel
![Page 3: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/3.jpg)
Multichannel Spectrometer
![Page 4: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/4.jpg)
Spectrophotometry – the mundaneMalachite Green Absorbance Spectrum from Multichannel Spectrometer
-0.05
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.45
400 450 500 550 600 650 700 750 800
Wavelength (nm)
Ab
sorb
ance Output from Spec-20:
Absorbance = 0.390
• Entire spectrum shows λmax and choice of analysis wavelength
• 2 spectrometers for lab of 24 students is sufficient
![Page 5: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/5.jpg)
The multichannel spectrometer is easily configured for emission spectrometry using a fiber optic probe for light input.
This has been used as a portable fluorescence emission spectrometer.
![Page 6: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/6.jpg)
Fluorescence
A typical luminescence spectrometer Our luminescence spectrometer
Handheld UV Source
Multichannel Spectrometer
Emission Monochromator Mulitchannel Detector
![Page 7: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/7.jpg)
Fluorescence Emission Spectrum
![Page 8: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/8.jpg)
0
500
1000
1500
2000
2500
3000
3500
4000
200 250 300 350 400 450 500 550 600 650
Wavelength (nm)
Em
iss
ion
Inte
ns
ity
0
0.5
1
1.5
2
2.5
3
Ab
so
rba
nc
e
B. Fluorescence emission spectrum of quinine
C. Absorbance spectrum of quinine
A. Hand held UV source
Absorbance and Fluorescence Emission Spectra of Quinine in
Water
![Page 9: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/9.jpg)
Fluorescence Emission Spectra
0
500
1000
1500
2000
2500
3000
3500
4000
350 400 450 500 550 600 650 700 750
Wavelength (nm)
Em
issi
on
Inte
nsi
ty (
Co
un
ts)
Quinine
Rhodamine BFluorescein
Eosin Yellow
![Page 10: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/10.jpg)
The multichannel spectrometer has also been used as a detector for qualitative analysis in a traditional flame test …
![Page 11: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/11.jpg)
Fiber Optic Collection of Na Flame Emission Spectrum
![Page 12: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/12.jpg)
Flame Emission Spectra of the Alkalis
0
1000
2000
3000
4000
400 500 600 700 800 900 1000Wavelength (nm)
Em
iss
ion
Inte
ns
ity
Li
Na
K
Cs
Rb
![Page 13: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/13.jpg)
Qualitative Analysis
“Group IV” Cations: Na+, K+, Ca2+, Mg2+, NH4+
Visual Flame Test:
• Na: persistent yellow
• K: transient red or lavender
• Ca: unreliable, transient red
Na masks K when both are present (Co blue filter)
False + and false – for Ca common with visual flame test
![Page 14: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/14.jpg)
Qualitative Analysis
Spectrometer Flame Test:
•Na and K easily detected and distinguished
•Ca not detected, chemical test only
0
1000
2000
3000
4000
400 450 500 550 600 650 700 750 800 850
Wavelength, nm
Em
iss
ion
Inte
ns
ity
KNa
background Ca
![Page 15: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/15.jpg)
Qualitative Analysis
Spectrometer Flame Test Student Results:
•Na > 95% correct on 1st try
•K > 95% correct on 1st try
•Ca > 98% correct on 1st try
Fewer false +, fewer false –
No blue filter for visual observation of K
![Page 16: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/16.jpg)
Quantitative Analysis
• Emission intensity is transient and variable
• Use another alkali as an internal standard
• Sample introduction with a nichrome wire does not provide adequate quantitative results
• Aspirate a sample using a continuous sprayer into the air intake of a Bunsen burner for improved sample introduction
![Page 17: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/17.jpg)
Compressed Air
Spectrometer
Fiber optic
![Page 18: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/18.jpg)
![Page 19: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/19.jpg)
Na Calibration (Li Internal Standard)
R2 = 0.996
0
0.5
1
1.5
2
2.5
0 100 200 300 400 500 600
ppm Na
Na
/Li
Em
iss
ion
In
ten
sit
y R
ati
oK Calibration (Li Internal Standard)
R2 = 0.995
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
0 100 200 300 400 500 600
ppm K
K/L
i E
mis
sio
n I
nte
ns
ity
Ra
tio
Cs Calibration (Li Internal Standard)
R2 = 0.999
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 100 200 300 400 500 600
ppm Cs
Cs
/Li
Em
iss
ion
In
ten
sit
y R
ati
o
Rb Calibration (Li Internal Standard)
R2 = 0.995
0
0.5
1
1.5
2
2.5
3
0 100 200 300 400 500 600
ppm Rb
Rb
/Li
Em
iss
ion
In
ten
sit
y R
ati
o
Li Calibration (K Internal Standard)
R2 = 0.994
0
0.1
0.2
0.3
0.4
0.5
0.6
0 50 100 150 200 250 300 350 400 450
ppm Li
Li/
K E
mis
sio
n I
nte
ns
ity
Ra
tio
![Page 20: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/20.jpg)
K unknown = 200.0 ppm Measured = 202.5 ppm
Student results not that good, but calibrations are OK
Detection limits using spray: 30-40 ppm for all alkalis
Detection limits much lower using nichrome wire
![Page 21: Applications of CCD Array Spectrometers in Undergraduate Education at EIU Jonathan Blitz and Daniel Sheeran Department of Chemistry](https://reader036.vdocument.in/reader036/viewer/2022062515/56649cea5503460f949b5aaf/html5/thumbnails/21.jpg)
Conclusions
• A multichannel spectrometer is useful for teaching spectrophotometry
• Using a fiber optic probe the spectrometer was configured for the following tasks:
As a fluorescence emission spectrometer
For qualitative analysis of alkalis
For quantitative analysis of alkalis using an internal standard calibration