1 Oculus Pentacam Wet-lab Mark Purdy Purdy Consulting, LLC [email protected] OCULUS head quarter, Wetzlar/Germany Applications of the OCULUS Pentacam® for cataract and refractive surgery and general screening

Upload: lykhuong

Post on 14-Jul-2018

239 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

1

Oculus Pentacam Wet-lab

Mark Purdy Purdy Consulting LLC

purdyophthalmicsgmailcom

OCULUS head quarter WetzlarGermany

Applications of the OCULUS Pentacamreg for cataract and

refractive surgery and general screening

2

Scheimpflug Image + Iris image amp Keratometry overlay

+ 3D Anterior Chamber Analyzer

+ Pachymetry Map absolute amp relative

+ Topography Maps (ant amp post)

+ Elevation Maps (ant amp post)

+ 3D Cataract Analyzer amp PNS

+ 3D pIOL simulation amp aging prediction

+ BelinAmbrosio Enhanced Ectasia III

+ Holladay Report amp EKR Detail Report

+ Tomography amp Corneal Optical Density

+ Cataract Pre-Op-Display amp Corneal Wavefront

+ Indice Report amp Help Topic

+ Corneal Ring Display

+ various compare displays_______________

= Pentacamreg 100 Scheimpflug images within one scan in less than 2 seconds

Pentacamreg the comprehensive analyzer

Scheimpflug Example

Schematic course of a scan

3

Facts about the imaging and calculation process

bull Center of the cornea is finely measured during the rotating scan

bull Eye motion during imaging process is detected with the pupil camera

bull Detected eye motion is corrected during the calculation process

bull 3D model of the anterior segment is calculated via ray-tracing

bull The optical distortions are individually corrected

Overview of all captured Scheimpflug images

Stop calculation

Help topic

Click with the mouse button to the yellow

triangle

4

Quality specifications QS

bull parameter bull value bull borders bull classification bull explantion bull reasons

QS should be checked after every examination

General amp Glaucoma screening

Kammervolumen

Glaucoma Screening

anterior chamber bdquonormal ldquo eye

anterior chamber narrow angle glaucoma

2 68 mm

Automatic Calculation of

bull Anterior chamber depth (ACD) internalexternal

bull Anterior chamber angle (ACA) in every Scheimpflug image

bull Chamber volume (ACV)

160 mm

364deg 382deg

201deg

196deg

ACA

ACA

Chamber volume

12

Chamber volume

5

Glaucoma clinical example

post-pre Iridectomy

post-op pre-op difference

bullScheimpflug image overview to detect abnormalities during calculation -gt time saving bullIndices report bullObjective and automatic calculation of ACD ACV ACA in up to 100 locations bull ACV turns out to be best predictor for narrow angle Glaucoma and post-op iridotomy and is not provided by anterior chamber OCTrsquos

bullPentacam anterior chamber parameters in young and middle-aged Chinese Lam AK Tse JS Clin Exp Optom 2013 Jan96(1)85-91 doi 101111j1444-0938201200795x Epub 2012 Sep 7

bullAssessment of the anterior chamber parameters after laser iridotomy in primary angle close suspect using Pentacam and gonioscopy Alireza et al Int journal of Ophthalmology 20136(5)680-684 bullComparison of scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles DS Grewal GS Brar R Jain and SPS Grewal Eye (2011) 1ndash9 amp 2011 Macmillan Publishers Limited All rights reserved 0950-222X11

Benefit

Corneal Refractive Surgery

- screening for ectasia - Corneal ringsIntacs

6

What about this case

With courtesy from Renato Ambrosio

What about this case

With courtesy from Renato Ambrosio

BelinAmbrosio Display III

Now with a database for hyperopic and myopic patients

Tomographic Normal values for Corneal Elevation and pachymetry in a Hyperopic Population

Joan Kim et all Clinical amp Experimental Ophthalmology 2011 22

Includes a new parameter ARTmax = TpPPi(max) Tp = thinnest pachymetry PPi(max) = pachymetry progression index

Novel pachymetric Parameters based on Corneal Tomography for diagnosing Keratoconus

Renato Ambrosio et all Journal of Refractive Surgery posted online July 29 2011

7

What about these cases

With courtesy from Renato Ambrosio

Keratoconus

What about these cases

With courtesy from Renato Ambrosio

keratoconus

no keratoconus

bullquick visual inspection comprehensive but intuitive

bullquick decision ndash for myopic and hyperopic patients

bullfor prepost surgery KeratoconusEctasia detection

bullcombination of 5 singl definite parameters into one final overall index

bullminimizes the ldquofalse negativesrdquo and ldquofalse positivesrdquo

bullhelpful for further diagnosis

BelinAmbrosio Display III benefit

8

Cataract pre-op screening (includes the general screening)

Corneal Optical Density

Corneal optical density average or maximum

Detailed optical densitometry evaluation in fixed layers and zones The table chart contains average values only The diagrams are displaying the max and avg values depending on the selected column or line

Corneal optical densitometry in a single slice or in a layer defined through 2 slices

Corneal Optical Density

post PKP

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 2: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

2

Scheimpflug Image + Iris image amp Keratometry overlay

+ 3D Anterior Chamber Analyzer

+ Pachymetry Map absolute amp relative

+ Topography Maps (ant amp post)

+ Elevation Maps (ant amp post)

+ 3D Cataract Analyzer amp PNS

+ 3D pIOL simulation amp aging prediction

+ BelinAmbrosio Enhanced Ectasia III

+ Holladay Report amp EKR Detail Report

+ Tomography amp Corneal Optical Density

+ Cataract Pre-Op-Display amp Corneal Wavefront

+ Indice Report amp Help Topic

+ Corneal Ring Display

+ various compare displays_______________

= Pentacamreg 100 Scheimpflug images within one scan in less than 2 seconds

Pentacamreg the comprehensive analyzer

Scheimpflug Example

Schematic course of a scan

3

Facts about the imaging and calculation process

bull Center of the cornea is finely measured during the rotating scan

bull Eye motion during imaging process is detected with the pupil camera

bull Detected eye motion is corrected during the calculation process

bull 3D model of the anterior segment is calculated via ray-tracing

bull The optical distortions are individually corrected

Overview of all captured Scheimpflug images

Stop calculation

Help topic

Click with the mouse button to the yellow

triangle

4

Quality specifications QS

bull parameter bull value bull borders bull classification bull explantion bull reasons

QS should be checked after every examination

General amp Glaucoma screening

Kammervolumen

Glaucoma Screening

anterior chamber bdquonormal ldquo eye

anterior chamber narrow angle glaucoma

2 68 mm

Automatic Calculation of

bull Anterior chamber depth (ACD) internalexternal

bull Anterior chamber angle (ACA) in every Scheimpflug image

bull Chamber volume (ACV)

160 mm

364deg 382deg

201deg

196deg

ACA

ACA

Chamber volume

12

Chamber volume

5

Glaucoma clinical example

post-pre Iridectomy

post-op pre-op difference

bullScheimpflug image overview to detect abnormalities during calculation -gt time saving bullIndices report bullObjective and automatic calculation of ACD ACV ACA in up to 100 locations bull ACV turns out to be best predictor for narrow angle Glaucoma and post-op iridotomy and is not provided by anterior chamber OCTrsquos

bullPentacam anterior chamber parameters in young and middle-aged Chinese Lam AK Tse JS Clin Exp Optom 2013 Jan96(1)85-91 doi 101111j1444-0938201200795x Epub 2012 Sep 7

bullAssessment of the anterior chamber parameters after laser iridotomy in primary angle close suspect using Pentacam and gonioscopy Alireza et al Int journal of Ophthalmology 20136(5)680-684 bullComparison of scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles DS Grewal GS Brar R Jain and SPS Grewal Eye (2011) 1ndash9 amp 2011 Macmillan Publishers Limited All rights reserved 0950-222X11

Benefit

Corneal Refractive Surgery

- screening for ectasia - Corneal ringsIntacs

6

What about this case

With courtesy from Renato Ambrosio

What about this case

With courtesy from Renato Ambrosio

BelinAmbrosio Display III

Now with a database for hyperopic and myopic patients

Tomographic Normal values for Corneal Elevation and pachymetry in a Hyperopic Population

Joan Kim et all Clinical amp Experimental Ophthalmology 2011 22

Includes a new parameter ARTmax = TpPPi(max) Tp = thinnest pachymetry PPi(max) = pachymetry progression index

Novel pachymetric Parameters based on Corneal Tomography for diagnosing Keratoconus

Renato Ambrosio et all Journal of Refractive Surgery posted online July 29 2011

7

What about these cases

With courtesy from Renato Ambrosio

Keratoconus

What about these cases

With courtesy from Renato Ambrosio

keratoconus

no keratoconus

bullquick visual inspection comprehensive but intuitive

bullquick decision ndash for myopic and hyperopic patients

bullfor prepost surgery KeratoconusEctasia detection

bullcombination of 5 singl definite parameters into one final overall index

bullminimizes the ldquofalse negativesrdquo and ldquofalse positivesrdquo

bullhelpful for further diagnosis

BelinAmbrosio Display III benefit

8

Cataract pre-op screening (includes the general screening)

Corneal Optical Density

Corneal optical density average or maximum

Detailed optical densitometry evaluation in fixed layers and zones The table chart contains average values only The diagrams are displaying the max and avg values depending on the selected column or line

Corneal optical densitometry in a single slice or in a layer defined through 2 slices

Corneal Optical Density

post PKP

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 3: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

3

Facts about the imaging and calculation process

bull Center of the cornea is finely measured during the rotating scan

bull Eye motion during imaging process is detected with the pupil camera

bull Detected eye motion is corrected during the calculation process

bull 3D model of the anterior segment is calculated via ray-tracing

bull The optical distortions are individually corrected

Overview of all captured Scheimpflug images

Stop calculation

Help topic

Click with the mouse button to the yellow

triangle

4

Quality specifications QS

bull parameter bull value bull borders bull classification bull explantion bull reasons

QS should be checked after every examination

General amp Glaucoma screening

Kammervolumen

Glaucoma Screening

anterior chamber bdquonormal ldquo eye

anterior chamber narrow angle glaucoma

2 68 mm

Automatic Calculation of

bull Anterior chamber depth (ACD) internalexternal

bull Anterior chamber angle (ACA) in every Scheimpflug image

bull Chamber volume (ACV)

160 mm

364deg 382deg

201deg

196deg

ACA

ACA

Chamber volume

12

Chamber volume

5

Glaucoma clinical example

post-pre Iridectomy

post-op pre-op difference

bullScheimpflug image overview to detect abnormalities during calculation -gt time saving bullIndices report bullObjective and automatic calculation of ACD ACV ACA in up to 100 locations bull ACV turns out to be best predictor for narrow angle Glaucoma and post-op iridotomy and is not provided by anterior chamber OCTrsquos

bullPentacam anterior chamber parameters in young and middle-aged Chinese Lam AK Tse JS Clin Exp Optom 2013 Jan96(1)85-91 doi 101111j1444-0938201200795x Epub 2012 Sep 7

bullAssessment of the anterior chamber parameters after laser iridotomy in primary angle close suspect using Pentacam and gonioscopy Alireza et al Int journal of Ophthalmology 20136(5)680-684 bullComparison of scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles DS Grewal GS Brar R Jain and SPS Grewal Eye (2011) 1ndash9 amp 2011 Macmillan Publishers Limited All rights reserved 0950-222X11

Benefit

Corneal Refractive Surgery

- screening for ectasia - Corneal ringsIntacs

6

What about this case

With courtesy from Renato Ambrosio

What about this case

With courtesy from Renato Ambrosio

BelinAmbrosio Display III

Now with a database for hyperopic and myopic patients

Tomographic Normal values for Corneal Elevation and pachymetry in a Hyperopic Population

Joan Kim et all Clinical amp Experimental Ophthalmology 2011 22

Includes a new parameter ARTmax = TpPPi(max) Tp = thinnest pachymetry PPi(max) = pachymetry progression index

Novel pachymetric Parameters based on Corneal Tomography for diagnosing Keratoconus

Renato Ambrosio et all Journal of Refractive Surgery posted online July 29 2011

7

What about these cases

With courtesy from Renato Ambrosio

Keratoconus

What about these cases

With courtesy from Renato Ambrosio

keratoconus

no keratoconus

bullquick visual inspection comprehensive but intuitive

bullquick decision ndash for myopic and hyperopic patients

bullfor prepost surgery KeratoconusEctasia detection

bullcombination of 5 singl definite parameters into one final overall index

bullminimizes the ldquofalse negativesrdquo and ldquofalse positivesrdquo

bullhelpful for further diagnosis

BelinAmbrosio Display III benefit

8

Cataract pre-op screening (includes the general screening)

Corneal Optical Density

Corneal optical density average or maximum

Detailed optical densitometry evaluation in fixed layers and zones The table chart contains average values only The diagrams are displaying the max and avg values depending on the selected column or line

Corneal optical densitometry in a single slice or in a layer defined through 2 slices

Corneal Optical Density

post PKP

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 4: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

4

Quality specifications QS

bull parameter bull value bull borders bull classification bull explantion bull reasons

QS should be checked after every examination

General amp Glaucoma screening

Kammervolumen

Glaucoma Screening

anterior chamber bdquonormal ldquo eye

anterior chamber narrow angle glaucoma

2 68 mm

Automatic Calculation of

bull Anterior chamber depth (ACD) internalexternal

bull Anterior chamber angle (ACA) in every Scheimpflug image

bull Chamber volume (ACV)

160 mm

364deg 382deg

201deg

196deg

ACA

ACA

Chamber volume

12

Chamber volume

5

Glaucoma clinical example

post-pre Iridectomy

post-op pre-op difference

bullScheimpflug image overview to detect abnormalities during calculation -gt time saving bullIndices report bullObjective and automatic calculation of ACD ACV ACA in up to 100 locations bull ACV turns out to be best predictor for narrow angle Glaucoma and post-op iridotomy and is not provided by anterior chamber OCTrsquos

bullPentacam anterior chamber parameters in young and middle-aged Chinese Lam AK Tse JS Clin Exp Optom 2013 Jan96(1)85-91 doi 101111j1444-0938201200795x Epub 2012 Sep 7

bullAssessment of the anterior chamber parameters after laser iridotomy in primary angle close suspect using Pentacam and gonioscopy Alireza et al Int journal of Ophthalmology 20136(5)680-684 bullComparison of scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles DS Grewal GS Brar R Jain and SPS Grewal Eye (2011) 1ndash9 amp 2011 Macmillan Publishers Limited All rights reserved 0950-222X11

Benefit

Corneal Refractive Surgery

- screening for ectasia - Corneal ringsIntacs

6

What about this case

With courtesy from Renato Ambrosio

What about this case

With courtesy from Renato Ambrosio

BelinAmbrosio Display III

Now with a database for hyperopic and myopic patients

Tomographic Normal values for Corneal Elevation and pachymetry in a Hyperopic Population

Joan Kim et all Clinical amp Experimental Ophthalmology 2011 22

Includes a new parameter ARTmax = TpPPi(max) Tp = thinnest pachymetry PPi(max) = pachymetry progression index

Novel pachymetric Parameters based on Corneal Tomography for diagnosing Keratoconus

Renato Ambrosio et all Journal of Refractive Surgery posted online July 29 2011

7

What about these cases

With courtesy from Renato Ambrosio

Keratoconus

What about these cases

With courtesy from Renato Ambrosio

keratoconus

no keratoconus

bullquick visual inspection comprehensive but intuitive

bullquick decision ndash for myopic and hyperopic patients

bullfor prepost surgery KeratoconusEctasia detection

bullcombination of 5 singl definite parameters into one final overall index

bullminimizes the ldquofalse negativesrdquo and ldquofalse positivesrdquo

bullhelpful for further diagnosis

BelinAmbrosio Display III benefit

8

Cataract pre-op screening (includes the general screening)

Corneal Optical Density

Corneal optical density average or maximum

Detailed optical densitometry evaluation in fixed layers and zones The table chart contains average values only The diagrams are displaying the max and avg values depending on the selected column or line

Corneal optical densitometry in a single slice or in a layer defined through 2 slices

Corneal Optical Density

post PKP

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 5: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

5

Glaucoma clinical example

post-pre Iridectomy

post-op pre-op difference

bullScheimpflug image overview to detect abnormalities during calculation -gt time saving bullIndices report bullObjective and automatic calculation of ACD ACV ACA in up to 100 locations bull ACV turns out to be best predictor for narrow angle Glaucoma and post-op iridotomy and is not provided by anterior chamber OCTrsquos

bullPentacam anterior chamber parameters in young and middle-aged Chinese Lam AK Tse JS Clin Exp Optom 2013 Jan96(1)85-91 doi 101111j1444-0938201200795x Epub 2012 Sep 7

bullAssessment of the anterior chamber parameters after laser iridotomy in primary angle close suspect using Pentacam and gonioscopy Alireza et al Int journal of Ophthalmology 20136(5)680-684 bullComparison of scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles DS Grewal GS Brar R Jain and SPS Grewal Eye (2011) 1ndash9 amp 2011 Macmillan Publishers Limited All rights reserved 0950-222X11

Benefit

Corneal Refractive Surgery

- screening for ectasia - Corneal ringsIntacs

6

What about this case

With courtesy from Renato Ambrosio

What about this case

With courtesy from Renato Ambrosio

BelinAmbrosio Display III

Now with a database for hyperopic and myopic patients

Tomographic Normal values for Corneal Elevation and pachymetry in a Hyperopic Population

Joan Kim et all Clinical amp Experimental Ophthalmology 2011 22

Includes a new parameter ARTmax = TpPPi(max) Tp = thinnest pachymetry PPi(max) = pachymetry progression index

Novel pachymetric Parameters based on Corneal Tomography for diagnosing Keratoconus

Renato Ambrosio et all Journal of Refractive Surgery posted online July 29 2011

7

What about these cases

With courtesy from Renato Ambrosio

Keratoconus

What about these cases

With courtesy from Renato Ambrosio

keratoconus

no keratoconus

bullquick visual inspection comprehensive but intuitive

bullquick decision ndash for myopic and hyperopic patients

bullfor prepost surgery KeratoconusEctasia detection

bullcombination of 5 singl definite parameters into one final overall index

bullminimizes the ldquofalse negativesrdquo and ldquofalse positivesrdquo

bullhelpful for further diagnosis

BelinAmbrosio Display III benefit

8

Cataract pre-op screening (includes the general screening)

Corneal Optical Density

Corneal optical density average or maximum

Detailed optical densitometry evaluation in fixed layers and zones The table chart contains average values only The diagrams are displaying the max and avg values depending on the selected column or line

Corneal optical densitometry in a single slice or in a layer defined through 2 slices

Corneal Optical Density

post PKP

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 6: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

6

What about this case

With courtesy from Renato Ambrosio

What about this case

With courtesy from Renato Ambrosio

BelinAmbrosio Display III

Now with a database for hyperopic and myopic patients

Tomographic Normal values for Corneal Elevation and pachymetry in a Hyperopic Population

Joan Kim et all Clinical amp Experimental Ophthalmology 2011 22

Includes a new parameter ARTmax = TpPPi(max) Tp = thinnest pachymetry PPi(max) = pachymetry progression index

Novel pachymetric Parameters based on Corneal Tomography for diagnosing Keratoconus

Renato Ambrosio et all Journal of Refractive Surgery posted online July 29 2011

7

What about these cases

With courtesy from Renato Ambrosio

Keratoconus

What about these cases

With courtesy from Renato Ambrosio

keratoconus

no keratoconus

bullquick visual inspection comprehensive but intuitive

bullquick decision ndash for myopic and hyperopic patients

bullfor prepost surgery KeratoconusEctasia detection

bullcombination of 5 singl definite parameters into one final overall index

bullminimizes the ldquofalse negativesrdquo and ldquofalse positivesrdquo

bullhelpful for further diagnosis

BelinAmbrosio Display III benefit

8

Cataract pre-op screening (includes the general screening)

Corneal Optical Density

Corneal optical density average or maximum

Detailed optical densitometry evaluation in fixed layers and zones The table chart contains average values only The diagrams are displaying the max and avg values depending on the selected column or line

Corneal optical densitometry in a single slice or in a layer defined through 2 slices

Corneal Optical Density

post PKP

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 7: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

7

What about these cases

With courtesy from Renato Ambrosio

Keratoconus

What about these cases

With courtesy from Renato Ambrosio

keratoconus

no keratoconus

bullquick visual inspection comprehensive but intuitive

bullquick decision ndash for myopic and hyperopic patients

bullfor prepost surgery KeratoconusEctasia detection

bullcombination of 5 singl definite parameters into one final overall index

bullminimizes the ldquofalse negativesrdquo and ldquofalse positivesrdquo

bullhelpful for further diagnosis

BelinAmbrosio Display III benefit

8

Cataract pre-op screening (includes the general screening)

Corneal Optical Density

Corneal optical density average or maximum

Detailed optical densitometry evaluation in fixed layers and zones The table chart contains average values only The diagrams are displaying the max and avg values depending on the selected column or line

Corneal optical densitometry in a single slice or in a layer defined through 2 slices

Corneal Optical Density

post PKP

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 8: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

8

Cataract pre-op screening (includes the general screening)

Corneal Optical Density

Corneal optical density average or maximum

Detailed optical densitometry evaluation in fixed layers and zones The table chart contains average values only The diagrams are displaying the max and avg values depending on the selected column or line

Corneal optical densitometry in a single slice or in a layer defined through 2 slices

Corneal Optical Density

post PKP

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 9: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

9

Patients selection for Premium IOL

Cataract Pre-Op display

bull It was developed in co-operation with Prof Naoyuki Maeda bull Its main clinical use is the pre-op assessment of corneal optical

properties for premium IOL selection using a four steps criteria bull evaluation of corneal irregularities bull corneal shape assessment bull evaluations of corneal spherical aberrations bull evaluations of the corneal astigmatism

bull An article was published in bdquoThe Highlights of Ophthalmologyldquo

Assessment of Corneal Optical Quality for Premium IOLs with Pentacamregldquo

Cataract Pre-Op display

1st Step Total HOA lt 03microm therefore suitable for multifocal IOLs

2nd Step regular corneal topography may be suitable for toric IOL

3rd Step Total Z40 (spherical aberration) = 0105microm therefore aberration free IOL

4th Step Comparison of corneal astigmatism ndash all are equal therefore suitable

for toric IOL or LRI

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 10: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

10

toric IOLs

Repeatabillity of corneal astigmatism

Repeatabillity of corneal astigmatism

ΔK IOL Master 500

COR relative

COR LOA range

051 D 5806 102 D

ΔK Pentacam HR sim K

COR relative

COR LOA range

032 D 4016 064 D

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 11: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

11

Repeatabillity of corneal astigmatism

4346

6905

5806

4836

5474

4016

8177

662

5213

4487 4299

33 3027

3604

0

10

20

30

40

50

60

70

80

90

100

rela

tive

CO

R [

]

Repeatabillity of corneal astigmatism

bull Scheimpflug imaging derived values showed best repeatability

bull Manual keratometry performed worst

bull In range evaluated no influence of value of astigmatism on COR

bull Scheimpflug Tomography offers more data combined with more reliable measurements

Total corneal astigmatism important for toric IOLlsquos

Anterior surface only Is the posterior surface really important The study

Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement

showed 10 of eyes with more than 1 D of astigmatism (candidates to

toric IOL) had a difference in magnitude gt05 D or a difference in angle gt10deg between anterior and total astigmatism So considering the posterior surface is useful for these eyes

JAU-DER HO CHING-YAO TSAI AND SHIOW-WEN LIOU copy 2009 BY ELSEVIER INC ALL RIGHTS

RESERVED 0002-939409$3600 doi101016jajo200812020

Conclusion the posterior surface should be considered in terms of angle and amount of the astigmatism

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 12: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

12

Total Corneal Refractive Power

The Pentacamreg calculates the Total Corneal Refractive Power

of the cornea considering - anterior and posterior corneal shape

- corneal thickness - Snellrsquos law of refraction

bull Pentacam provides most accurate and repeatable astigmatic measurements

bull 4 steps to find the best Premium IOL

bull Solution for toric IOLlsquos incl posterior corneal surface in terms of

bull Detection

bull Selection and providing of the best IOL calculation methode

bull Print-out for the OR

Benefit

IOL power calculation with Pentacamreg

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 13: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

13

Import of the axial length of the human eye

The Pentacamreg software allows the import of the axial

length from optical biometers

Currently the axial length from the

bull AL-Scan from NIDEK

bull Lenstar form Haag Streit

bull OA-1000 from Tomey

is supported by the Pentacamreg software

Phaco Optics and OKULIX are supporting this feature

already

Summary Pentacam for the Cataract surgery

1 pre-op screening using the Indices Display corneal optical densitometry and Scheimpflug image (PNS) and BelinAmbrosio Enhanced Ectasia

2 Premium IOL selection using the Cataract pre-op display

3 If necessary use the power distribution display for details

4 IOL power calculation using the

bull ASCRS or the manufacturers online calculators

bull ray tracing calculators (Phaco Optics OKULIX) BESSt II Shammas post hyperopic LASIK EKRlsquos and Holladay II

5 Keratometry and Topography overlay in the iris image as a print-out for the operating theather

Imaging

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 14: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

14

Pentacam Example PMD

Pentacam example cornea

Descemetrsquos Stripping with Endothelial

Keratoplasty

With courtesy from Prof Michael Belin

Pentacamreg example cornea

With courtesy from Dr T Neuhann

DLKP (deep lamelar keratoplasty)

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip

Page 15: Applications of the OCULUS Pentacam® for cataract and ... · densitometry and Scheimpflug image (PNS) and Belin/Ambrosio Enhanced Ectasia 2. Premium IOL selection using the Cataract

15

Post-op control

With courtesy from Renato Ambrosio MD

Thank you for your attention

Please visit us under wwwpentacamcom

Please visit us under wwwoculusde

Please visit us under wwwoculustv

Thank you for your attention

Hands On Portion hellip