aqua-lab cs1 prep · 3.5 gpm cat pump 1/4” hose barb chemical inlet (uses #8-32 metering tips)...
TRANSCRIPT
-
© Hydra-Flex, Inc. 2017
AQUA-LAB™ CS1 PREPCHEMICAL DISPENSING SYSTEM
User ManualREV A
4000134 revA0817© Hydra-Flex, Inc. 2017
-
© Hydra-Flex, Inc. 2017
TABLE OF CONTENTS
FOR ADDITIONAL SUPPORT CALL
952-808-3640OR VISIT US ON THE WEB:
www.hydraflexinc.com
SPECIFICATIONS 1
ILLUSTRATION 1
INSTALLATION AND SET-UP 2
Initial Injector Setup 2
Foaming Prep Setup 3
Optimizing the System 4
Chemical Usage Measuring 4
TROUBLESHOOTING 5
APPENDIX A 7
APPENDIX B 8
APPENDIX C 9
APPENDIX D 10
-
© Hydra-Flex, Inc. 2017
SPECIFICATIONS
ILLUSTRATION
Page | 1
SPECIFICATIONSOperating Water Pressure 200 PSI (factory set)
Pneumatics Operating Pressure 100 PSI max. feed
Maximum Water Source Temperature 140°F
Operating Ambient Temperature 40-120°
Signal Independent on/off switch. No signal from car wash controller required
Electrical Supply120/208-230 VAC (1-Phase) – 8 ft. cord included
3/4 hp pump 10 amps @ 120 volts or 5 amps @ 208 volts
Water Supply
50 micron filtration or better recommended
1/2” OD Poly Flow or 3/8” ID Hose
Inlet pressure: Flooded – 80 PSI
If pump is fed from tank, water level must be ~28’ above pump
Air Supply3/8” feed per system
20 CFM @ 100 PSI
Solution Outlet Lines1/2” Poly lines (3/8” ID)
Max. pressure of outlet line 66 PSI
Dimensions 20”L x 12” W x 16” H
Weight ~ 45 lbs.
3/8” Push-to-ConnectFoaming Air Outlet
3/8” Push-to-ConnectAir Inlet
Foaming Air Regulatorwith Pressure Gauge
AdjustablePressure Switch
1/2” Push-to-ConnectWater Inlet(or 3/8” NPT)
3.5 GPM CAT Pump
1/4” Hose Barb Chemical Inlet(Uses #8-32 Metering Tips)
Bug Prep Assembly
1/2” Push-to-ConnectSolution Outlet
IntegratedMotor Control
Air Operated ValveManifold Body
-
© Hydra-Flex, Inc. 2017 Page | 2
INSTALLATION AND SET-UP1. Unpack from crate and inspect for damages.
2. Anchor mounting bracket to the wall in the desired location using three ¼” wall anchors
• Make sure that the mounting bracket is level.
3. Hang CS1-PREP on mounting bracket.
4. Plumb water inlet using attached ½” push to connect fitting or remove and use the 3/8” NPT to adapt to the fitting of your choice.
5. Plumb the air inlet using 3/8” poly flow hose ( 60 PSI is required to open the valve properly).
6. Plumb the solution outlet to the desired location using ½” poly flow tubing or ½” ID hose of your choice.
7. (Optional) Plumb the air outlet tubing using 3/8” poly flow tubing to a convenient location as near the gun as possible, and tee into the solution line. Install check valves on both lines.
8. Connect the gun to the end of the solution line, and install the supplied nozzle.
9. Plug the 120 VAC cord into a dedicated 20 amp plug.
10. Turn the on/off switch to “on.” The system should turn on and pressurize the outlet tubing to the set point of the pressure switch and turn off.
11. Have someone use the gun and monitor the system to make sure that the pump is not cycling unnec-essarily. If it is turning on and off rapidly, address one of the three below issues.
• Increase the pressure set point of the switch by turning clockwise.
• Increase the size of the nozzle on the gun.
• Decrease the size of the injector on the system.
12. Make sure that the pressure gauge at the manifold is set at 200 PSI by adjusting the pressure regulator.
• Make sure to engage the lock nut after adjustment.
13. At the end of each day make sure that the system switch is turned “off.” This will ensure that the system doesn’t cycle on and off during the night, and will not cause a flood if there is a hose or fitting failure downstream of the system.
INITIAL INJECTOR SETUP(Based on field experience this is HFI’s recommended staring point)
1. All HFI prep systems ship with a ¾ GPM injector and a #8 25° nozzle to be attached to the prep gun of your choice. If a higher or lower flow rate or spray angle is desired, refer to the chart and select the injector / nozzle combination that best meets your needs.
2. Connect solution lines to the bug prep assembly using the attached push to connect fitting.
• Do not over-tighten poly fittings or they may crack.
3. Connect 1/4” poly line from chemical container to the injector.
INJECTORSIZE
INJECTORCOLOR
PREP GUNNOZZLE
SIZE
PREP GUNNOZZLEORIFICE
APPROX. BACK PRESSURE
0.029” White 3 0.043” 40
0.029” White 4 0.052” 15
0.040” Yellow 5 0.057” 50
0.040” Yellow 6 0.062” 25
0.040” Yellow 7 0.067” 10
0.051” Tan 7 0.067” 50
0.051” Tan 8 0.072” 40
0.051” Tan 10 0.080” 15
0.057” Red 8 0.072” 60
0.057” Red 10 0.080” 45
0.057” Red 15 0.096” 18
-
© Hydra-Flex, Inc. 2017 Page | 3
• Ensure a foot valve or similar check valve/filter is installed on line.
• These must be present or metering tips may clog.
4. Metering tips will need to be installed to set dilution ratio (see Appendix A for ratio charts to determine tip).
WARNING: If the spray nozzle on the gun is too small for the injector chosen, the back pressure put on the injector may cause the injector not to function and chemical will not be pulled. The back pressure should not exceed 66 PSI. Back pressure gauges are available for purchase from HFI (P/N 1001105). Water will continue to flow as normal.
INJECTOR VACUUM CHECK (FOR TROUBLESHOOTING INJECTORS)1. At the Chem-Flex injector, remove the chemical feed line from the injector hose barb.
2. Attach the tubing of the vacuum gauge to the Chem-Flex hose barb.
3. With the gun open and the system running, an injector that is working properly will have a reading greater than or equal to (≥) 20 in Hg.
4. If injector is not functioning:
• Check metering tip for clogs (can be tested with no metering tip to ensure injector is performing).
• Try a smaller injector (this will produce less flow and thus less back pressure).
• Remove a nozzle(s) at the arch, allowing water to free flow (this will reduce back pressure).
There is a variation of performance in the injectors that comes from slight variations in the dimensions of the parts and in assembly that are unavoidable. It is common to see the resultant vacuum range from 20 in Hg all the way up to 28. There is also variation in the through hole size on the meter tips from the manufacture (within their manufacturing tolerances). Using the same tip color from site to site is a good starting point. However with the potential for variation from part to part it is reasonable to still need to do some adjustments from there
FOAMING PREP SETUPIf you would like to foam your prep solution:
1. Nozzle selection must be based on specific set up, chemical, and quantity / makeup of the foam you desire.
2. The air outlet line should be run from the outlet on the CS1 to as close to the gun as feasible for your specific set up.
3. HFI recommends using a high flow check valves on both the air and chemical line to prevent any back feed up either line.
4. For the thickest and highest performance of foam, a “foam generator” is recommended at the outlet of the prep gun. HFI part number 1001751.
5. Occasionally, if the air regulator is set too high, you may need to lower the regulators to its lowest setting and then turn it back up to the desired pressure.
-
© Hydra-Flex, Inc. 2017 Page | 4
What do metering tips do?• Increases or decreases the amount of
chemical in the solution.
What do nozzles do?• Determines the pattern and back pressure
of the solution.
APPLICATION OPTIMIZATION(REPEAT FOR EACH APPLICATION)View Performance at gun with decision maker
• Application too wet• Increase foaming air pressure• Reduce injector size (decreases water)• Increase metering tip (increases
chemical)
• Application too dry• Decrease foaming air pressure• Increase injector size (increases water)• Decrease metering tip (decreases
chemical)
• Nozzle sputters• Decrease foaming air pressure• Decrease number of nozzle(s) and/or
size used on arch• Increase injector size (increases water)
• Too much chemical used• Decrease metering tip• Decrease metering tip and injector
size (to maintain desired ratio)
OPTIMIZING THE SYSTEMConsistently achieve the desired cleaning and presentation/performance using the least amount of chemical and water.
INJECTORS VS. METERING TIPS VS. NOZZLESThe key to optimizing the system is through trial and error. Don’t be afraid to try these steps to achieve your ideal performance
What do injectors do?• Increases or decreases the amount of water in the solution.
• No chemical • Check vacuum/backpressure of
injector for clogging (see page X for Injector Vacuum Check Instructions or page X for troubleshooting)
• Check foot valve• Check metering tip
• Nozzle fan pattern not filled• Reduce nozzle size• Increase injector size (increases water)
• Water not present at all nozzles on arch• Verify check valves are functioning• Verify nozzles are not plugged• Reduce number of nozzles• Reduce nozzle size• Increase injector size (increases water)
CHEMICAL USAGE MEASURINGVERIFY TITRATION OF CHEMICALS BEFORE PROCEEDING
1. Set up lab scale with small bucket of chemical to be measured.
2. Put the suction line into the bucket.
3. Run the application being tested to “prime” the line. (All air bubbles must be removed for accuracy.)
4. Record the Initial Weight from the scale. (Tarring the scale with weight on the scale can affect accuracy.)
5. Run the application for 1 vehicle (or manually for the same amount of time it would be on for 1 vehicle).
6. Record the Final Weight from the scale.
0
8
16
24
-
© Hydra-Flex, Inc. 2017 Page | 5
RECOMMENDED MAINTENANCETHE RECOMMENDED SERVICE AND MAINTENANCE ON THE AQUA-LAB SYSTEM ARE AS FOLLOWS. Semi-Annually
• Check and replace injector metering tips.• Clean / check air and water valves.• Inspect and replace chemical lines as needed.• Ensure lines are tightly secured to injector hose barbs, clip 1” off old hose as needed that was
stretched by hose barb.
Annually• Clean water regulator.• Inspect motor starter for corrosion, if identified order replacement/spare parts.
1-3 Years• Inspect and replace injectors.• Replace / rebuild water valves.• Replace main pressure regulator.
TROUBLESHOOTINGPROBLEM POTENTIAL CAUSED SOLUTIONS
Pump Operates, but delivers little or no water
Pump not primed See priming instructions.
Inlet restriction Check all in-line filters and inlet plumbing for restrictions.
Inadequate water supply Check pressure on inlet side of pump to be sure positive pressure is maintained
Undersized piping Replace with larger piping
Leak on the inlet side Make sure connections are tight
Worn or defective pump parts Replace worn parts or entire pump, clean parts if required
Pump check valves clogged Clean and check all 6 pump check valves
Incorrect motor rotation Reverse motor rotation by interchanging any two leads
7. Subtract the Initial Weight from the Final Weight to determine the weight of used product.
8. Divide the Per Car Weight in grams by the specific gravity of the chemical to determine the milliliters of chemical used per vehicle.
9. Repeat for each chemical application.
Pump won’t start or run at full speed
Blown fuse or circuit breaker Replaced fuses or close circuit breaker
Defective motor starter contactor Replaced motor starter contactor
Incorrect motor voltage Voltage must be within 10% of motor rated voltage
Defective motor Replace motor
Pump components damaged Replace worn part or entire pump
Excessive Noise from Pump
Pump not secured firmly Secure properly
Restricted inlet Clean or correct restriction
Cavitation (sounds like marbles in pump) Increase inlet size
Not primed Re-prime pump
Water regulator fluttering Try to adjust regulator down and then back up or replace regulator
Pump check valves clogged Clean and check all 6 pump check valves
-
© Hydra-Flex, Inc. 2017 Page | 6
PROBLEM POTENTIAL CAUSED SOLUTIONS
Pump LeaksWorn mechanical seal Replace shaft seal
Worn o-ring seals Replace
Pump cycles on and offPressure switch set too low Increase set point of pressure switch
Foam generator clogging Clean foam generator
Injector is not drawing chemical - Passes Vacuum Pressure check test Clogged chemical feed
Check chemical hose, foot valve, meter-ing tip and hose barb for debris or clogs
Injector is not drawing chemical - Fails Vacuum Pressure check test
Too much back pressure on injector
Perform back pressure check test out-lined in Section IV if this manual. If the result is higher than 66 psi then Increase arch nozzle size or quantity, use larger tubing, or use a smaller flow injector
Clogged injector check valveBlow compressed air through the chem-ical hose barb on the injector to remove debris
Clogged injector nozzle Remove injector and blow out any debris with compressed air
Defective Injector If Vacuum check fails but Back Pressure is less than 66 psi, replace the injector
System won’t regulate up to 200 psi
Debris in regulator Remove regulator and clean out debris
Opening to many valves at onceSystem is limited to opening up to 8 valves at once or pressure will be greatly affected by so much demand
Pump not primed Follow priming instructions
Defective Check Valve Replace check valve if broken
Defective Regulator Replace Regulator
Defective Pump Replace Pump
Motor rotation incorrect Verify rotation and adjust wiring to correct
Pump check valves clogged Clean and check all 6 pump check valves.
Flow at arch is too low
Incorrect Injector Flow Rate Selection Replace with desired injector size
System pressure too low Ensure system pressure is set at 200 psi
Foam generator plugged Ensure cleaned and clear
Downstream pluming restrictive
Perform Back pressure test outlined in Section IV of this manual, if over 66 psi, increase tube size and reduce elbows, turns or other restrictive plumbing
No flow from injector
Solenoid valve malfunction
Ensure valve is receiving the correct electrical signal and voltage
Disassemble valve and clean out debris (See valve replacement instructions)
Clogged Injector Remove injector and blow out debris with compressed air
No water supply Check that the system has a supply of water
Valve stuck open – Staying open when signal is off
Debris in valve seat Remove and clean valve (See valve replacement instructions)
O-ring failure Replace O-rings inside valve
-
© Hydra-Flex, Inc. 2017 Page | 7
APPENDIX A: CHEM-FLEX INJECTORS - CHEMICAL DILUTION RATIOS
3.00
”1:
251
1: 5
031:
754
1: 1
006
1: 1
509
1: 2
012
1: 2
263
1: 2
416
1: 2
460
1: 2
628
1: 4
526
1: 5
532
1: 8
047
1: 1
0059
1: 1
2070
1: 1
5088
2.00
”1:
181
1: 3
631:
544
1: 7
261:
108
91:
145
11:
163
31:
140
21:
133
71:
151
01:
326
61:
399
11:
580
61:
725
71:
870
81:
108
85
1.00
”1:
104
1: 2
081:
311
1: 4
151:
623
1: 8
311:
934
1: 8
351:
778
1: 8
981:
186
91:
228
41:
332
21:
415
31:
498
31:
622
9
0.75
”1:
82
1: 1
651:
247
1: 3
291:
494
1: 6
591:
741
1: 6
421:
668
1: 7
551:
148
31:
181
21:
263
61:
329
51:
395
41:
494
2
0.50
”1:
59
1: 1
191:
178
1: 2
381:
357
1: 4
751:
535
1: 4
871:
496
1: 5
581:
106
91:
130
71:
190
11:
237
61:
285
21:
356
4
0.25
”1:
34
1: 6
81:
102
1: 1
361:
204
1: 2
721:
306
1: 2
821:
304
1: 3
441:
612
1: 7
481:
108
81:
136
01:
163
21:
204
0
#8
-32
MET
ERIN
G T
IPS
Flow
Rat
e(G
PM) a
t 200
PSI
0.25
0.50
0.75
1.00
1.50
2.00
2.25
3.00
3.25
3.75
4.50
5.50
8.0
10.0
12.0
15.0
Inje
ctor
Col
orW
hite
Yello
wTa
nRe
dO
rang
eG
ray
Blue
Ligh
t Bl
ueLi
ght
Gre
enPi
nkPu
rple
Dar
k G
reen
Blac
kBl
ack
Blac
kBl
ack
Noz
zle
Size
0.02
9”
(0.7
mm
)0.
040”
(1.0
mm
)0.
051”
(1.3
mm
)0.
057”
(1.4
mm
)0.
070”
(1.8
mm
)0.
083”
(2.1
mm
)0.
086”
(2.2
mm
)0.
095”
(2.4
mm
)0.
098”
(2.5
mm
)0.
106”
(2.7
mm
)0.
117”
(2.9
mm
)0.
125”
(3.2
mm
)0.
161”
(4.1
mm
)0.
117”
(4.5
mm
)0.
186”
(4.7
mm
)0.
207”
(5.3
mm
)
CO
PPER
1: 5
71:
104
1: 1
551:
195
1: 2
811:
406
1: 4
681:
598
1: 6
291:
739
1: 8
811:
107
41:
201
91:
231
81:
332
41:
425
9
PUM
PKIN
1: 4
31:
82
1: 1
191:
126
1: 2
381:
348
1: 3
981:
519
1: 5
541:
647
1: 7
741:
946
1: 1
465
1: 1
832
1: 2
310
1: 3
042
BURG
UN
DY
1: 3
41:
67
1: 9
71:
111
1: 2
071:
304
1: 3
471:
461
1: 4
951:
575
1: 6
901:
845
1: 1
244
1: 1
670
1: 1
975
1: 2
469
LIM
E1:
28
1: 5
71:
81
1: 1
001:
183
1: 2
701:
307
1: 4
141:
447
1: 5
181:
622
1: 7
641:
988
1: 1
262
1: 1
481
1: 1
958
TAN
1: 2
81:
57
1: 8
11:
100
1: 1
831:
270
1: 3
071:
414
1: 4
471:
518
1: 6
221:
764
1: 8
261:
104
21:
128
61:
155
7
ORA
NG
E1:
23
1: 4
41:
64
1: 7
81:
137
1: 1
961:
215
1: 2
931:
314
1: 3
651:
437
1: 5
361:
751
1: 8
741:
113
61:
126
2
TURQ
UO
ISE
1: 1
71:
31
1: 4
51:
55
1: 9
11:
126
1: 1
341:
185
1: 1
971:
230
1: 2
741:
336
1: 5
611:
752
1: 8
851:
109
9
PIN
K1:
14
1: 2
41:
35
1: 4
21:
68
1: 9
31:
98
1: 1
301:
143
1: 1
591:
188
1: 2
241:
372
1: 4
731:
570
1: 7
28
LIG
HT
BLU
E1:
11
1: 1
71:
24
1: 3
11:
47
1: 6
41:
66
1: 9
31:
98
1: 1
141:
136
1: 1
661:
318
1: 4
061:
483
1: 5
72
BRO
WN
1: 1
01:
15
1: 2
21:
28
1: 4
31:
58
1: 5
91:
84
1: 8
81:
103
1: 1
221:
150
1: 2
701:
338
1: 4
081:
507
RED
1: 1
21:
17
1: 2
31:
34
1: 4
51:
46
1: 6
51:
69
1: 8
01:
95
1: 1
161:
202
1: 2
541:
306
1: 3
75
WH
ITE
1: 1
21:
16
1: 2
21:
31
1: 4
21:
43
1: 6
11:
64
1: 7
41:
88
1: 1
081:
180
1: 2
261:
272
1: 3
43
GRE
EN1:
11
1: 1
41:
20
1: 2
81:
37
1: 3
81:
53
1: 5
51:
65
1: 7
61:
94
1: 1
601:
201
1: 2
411:
300
BLU
E1:
10
1: 1
21:
17
1: 2
31:
30
1: 3
11:
44
1: 4
61:
53
1: 6
31:
77
1: 1
221:
152
1: 1
851:
224
YELL
OW
1: 9
1: 1
21:
16
1: 2
01:
22
1: 3
01:
31
1: 3
61:
42
1: 5
21:
80
1: 1
001:
120
1: 1
48
BLA
CK
1: 1
01:
13
1: 1
61:
17
1: 2
31:
24
1: 2
81:
33
1: 4
01:
61
1: 7
71:
92
1: 1
15
PURP
LE1:
6.6
1: 8
.31:
91:
10
1: 1
31:
13
1: 1
51:
17
1: 2
11:
31
1: 3
91:
46
1: 5
7
GRA
Y1:
5.3
1: 6
.71:
6.9
1: 7
.61:
9.9
1: 1
01:
11.
61:
13
1: 1
61:
24
1: 3
01:
35
1: 4
4
OPE
N1:
4.9
1: 5
.31:
5.2
1: 6
.01:
6.1
1: 6
.11:
7.6
1: 8
.31:
10
1: 1
21:
15
1: 1
81:
23
3/7”
po
lyflo
w (L
LDP
E) t
ubin
g is
req
uire
d t
o e
nsur
e a
seal
bet
wee
n th
e tu
be
wal
l and
the
sp
iral
plu
g.
(Ass
umes
feed
pre
ssur
e of
200
PSI
(14
bar)
)
NO
TE: D
iluti
on
rati
os
giv
en a
re b
ased
on
dra
win
g w
ater
thr
oug
h th
e m
eter
ing
tip
s an
d a
re m
eant
as
a st
arti
ng p
oin
t fo
r sy
stem
co
nfig
urat
ion.
R
esul
ts a
re e
xpec
ted
to
var
y w
hen
dra
win
g c
hem
ical
s d
ue t
o d
iffer
ence
s in
vis
cosi
ty a
nd t
emp
erat
ure.
The
re m
ay b
e sl
ight
var
iati
ons
of
per
form
ance
in
inje
cto
rs a
nd m
eter
ing
tip
s th
at a
re u
navo
idab
le d
ue t
o m
anuf
actu
re t
ole
ranc
es. U
sing
the
sam
e ti
p c
olo
r fr
om
sit
e to
sit
e is
a g
oo
d s
tart
ing
po
int.
H
ow
ever
, wit
h th
e p
ote
ntia
l fo
r va
riat
ion
fro
m p
art,
to
par
t it
is r
easo
nab
le t
o s
till
need
to
do
so
me
adju
stm
ents
fro
m t
here
.
-
© Hydra-Flex, Inc. 2017 Page | 8
APPENDIX B: CHEM-FLEX INJECTOR PART NUMBERS
Max Pressure Inlet
NPT up to 1000 PSI (69 bar)
PC2 up to 500 PSI (34 bar)Max Pressure
Outlet 333 PSI (23 bar)Temperature Range 33°F - 175°F (.5°C - 79°C)
Maximum Wrench Torque 30 ft-lbs (41 N-m)
SPECIFICATIONS
INLET CONNECTION PC2 (QUICK CONNECT)
OUTLET CONNECTION 3/8” MNPT
COLOR FLOW ORIFICE FLOW RATE @ 200 PSI (14 bar) SINGLE BARB DUAL BARB
White 0.029” (0.74 mm) 0.25 GPM (1.0 LPM) 618029 -
Yellow 0.040” (1.02 mm) 0.50 GPM (1.9 LPM) 618040 629040
Tan 0.051” (1.39 mm) 0.75 GPM (2.8 LPM) 618051 629051
Red 0.057” (1.45 mm) 1.00 GPM (3.8 LPM) 618057 629057
Orange 0.070” (1.78 mm) 1.50 GPM (5.7 LPM) 618070 629070
Gray 0.083” (2.11 mm) 2.00 GPM (7.6 LPM) 618083 629083
Dark Blue 0.086” (2.18 mm) 2.25 GPM (8.5 LPM) 618086 629086
Light Blue 0.095” (2.41 mm) 3.00 GPM (11.3 LPM) 618095 629095
Light Green 0.098” (2.49 mm) 3.25 GPM (12.3 LPM) 618098 629098
Pink 0.106” (2.69 mm) 3.75 GPM (14.2 LPM) 618106 629106
Purple 0.117” (2.98 mm) 4.50 GPM (17.0 LPM) 618117 629117
Dark Green 0.125” (3.18 mm) 5.50 GPM (20.8 LPM) 618125 629125
-
© Hydra-Flex, Inc. 2017 Page | 9
16
15
14
17
4
5
21
7 11
12
10
9
8
PART NAME PART NUMBER1 Air Actuated Hydra-Cannon Valve Replacement Kit - Composite 3000931
2 10-32” X 1/4” Push-To-Connect Air Fitting For Air Actuated Valve 3000987
3 (Not Shown) Hydra-Cannon Valve O-Ring Replacement Kit 1001155
4 Solenoid Actuated Air Valve - Foaming
24 VAC - 1001428
24 VDC - 1001429
120 VAC - 1001430
5 Din I X Field Wireable Cable - With LED24 VAC/VDC - 3001157
120 VAC - 3001156
6 (Not Shown) DIN I 9’ (3 m) - No LED 3000954
7 Foaming Small Air Regulator & 0-60 PSI Gauge 3000808
8 Outlet Pressure Gauge 0-400 PSI Bottom Mount 3000491
9 Single Gray Hydra-Cannon Replacement Assembly 1001777
10 1/2” NPT Hydra-Cannon Manifold End Fitting Assembly 1001556
11 1/4” Push To Connect Tee 3000815
12 (Not Shown) 1/4” Push To Connect To 1/8” NPT Elbow 3000803
13 (Not Shown) 1/4” Hose Barb Foaming Air Check Valve 3000819
14 20 GPM (75 LPM) Bypass Pressure Regulator - Brass 3000951
15 1/4” NPT Pressure Switch 3000969
16 2.9 GPM (11 LPM) Pump 3001031
17 Motor - 1 HP 3001031
18 (Not Shown) Contactor (120 VAC - Sprecher & Shaun) 3000865
19 (Not Shown) Timer Relay 3000644
20 (Not Shown) Complete Bug Prep Kit 1001759
21 Pressure Switch Manifold Only 1001550
22 (Not Shown) Foaming Prep Gun & Injector 1001754
23 (Not Shown) Foaming Prep Gun Only 1001751
24 (Not Shown) Dual Lumen 3/8” Spiral Hose 30’ 1001778
25 (Not Shown) Foam Generator For Foam Prep Gun 1001750
APPENDIX C: REPLACEMENT PARTS LIST - AQUA-LAB™ CS1 PREP
-
© Hydra-Flex, Inc. 2017 Page | 10
FACTORY LIMITEDHydra-Flex, Inc warrants its equipment to be free from defect in material or workmanship under proper
normal use for a period of one (1) year beginning the date of purchase.
Hydra-Flex, Inc’s liability shall be limited to repair or replacement of parts found to be defective within the
warranty period and following Hydra-Flex, Inc’s inspection. Hydra-Flex, Inc shall have the option requiring
the return of defective material to establish the purchaser’s claim. In the event of repair or replacement
this limited warranty is non-cumulative. Neither labor nor transportation charges are included in this
warranty.
This warranty is based upon the proper care and maintenance of the warranted equipment. Warranty
does not apply if the merchandise is altered or modified in any way. Warranty does not apply to any
equipment which has been subject to misuse, inappropriate use of tools, including exposure to harsh
chemicals, neglect, lack of maintenance, freezing, fluid hammer, accident, third party damage, fluid
impurities such as sand or minerals, acts of God or acts of war. Nor does it apply to any equipment which
has been repaired or altered by anyone not so authorized by Hydra-Flex, Inc. All equipment must be
properly installed in accordance with specified plumbing, electrical, and mechanical requirements. The
warranty does not apply to normal wear and tear or routine maintenance components as described in the
equipment manual.
Except as expressly stated herein, Hydra-Flex, Inc shall not be liable for damages of any kind in
connection with the purchase, maintenance, or use of this equipment including loss of profits and all
claims for consequential damages. This limited warranty is in lieu of all other warranties expressed or
implied. Hydra-Flex, Inc neither assumes nor authorizes any person to assume for it any other obligation
or liability in connection herewith. This warranty is neither assignable nor transferable.
Transportation damage claims are to be submitted to the carrier of the damaged material.
APPENDIX D: AQUA-LAB™ WARRANTY
T: 952-808-3640 • www.hydraflexinc.com • [email protected]
4000134 revA0717