article 10c1p.ir/wp-content/uploads/014-asme-sec-v-a-10.pdf01 01 article 10 leak testing t-1000...

28
ARTICLE 10 T-1000 Introduction ................................................................ 173 T-1010 Scope ...................................................................... 173 T-1020 General .................................................................... 173 T-1021 Procedure Requirements ..................................................... 173 T-1022 Referencing Code ........................................................... 173 T-1030 Equipment.................................................................. 173 T-1031 Gages...................................................................... 173 T-1040 Miscellaneous Requirements .................................................. 174 T-1041 Cleanliness ................................................................. 174 T-1042 Openings ................................................................... 174 T-1043 Temperature ................................................................ 174 T-1044 Pressure /Vacuum (Pressure Limits) ........................................... 174 T-1050 Procedure .................................................................. 174 T-1051 Preliminary Leak Test ....................................................... 174 T-1052 Test Sequence .............................................................. 174 T-1060 Calibration ................................................................. 174 T-1061 Pressure /Vacuum Gages ..................................................... 174 T-1062 Temperature Measuring Devices .............................................. 174 T-1063 Calibration Leak Standards ................................................... 174 T-1070 Test ....................................................................... 175 T-1080 Evaluation .................................................................. 175 T-1081 Acceptance Standards........................................................ 175 T-1090 Documentation .............................................................. 175 T-1091 Test Report ................................................................ 175 T-1092 Record Retention ........................................................... 175 Mandatory Appendices Appendix I Bubble Test — Direct Pressure Technique..................................... 176 I-1000 Introduction ................................................................ 176 I-1010 Scope ...................................................................... 176 I-1030 Equipment.................................................................. 176 I-1031 Gases ...................................................................... 176 I-1032 Bubble Solution............................................................. 176 I-1033 Immersion Bath............................................................. 176 I-1070 Test ....................................................................... 176 I-1071 Soak Time ................................................................. 176 I-1072 Surface Temperature......................................................... 176 I-1073 Application of Solution ...................................................... 176 I-1074 Immersion in Bath .......................................................... 176 I-1075 Lighting and Visual Aids .................................................... 176 I-1076 Indication of Leakage ....................................................... 176 I-1077 Post-Test Cleaning .......................................................... 177 168 COPYRIGHT American Society of Mechanical Engineers Licensed by Information Handling Services COPYRIGHT American Society of Mechanical Engineers Licensed by Information Handling Services

Upload: others

Post on 26-Apr-2020

16 views

Category:

Documents


1 download

TRANSCRIPT

ARTICLE 10

T-1000 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173T-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173T-1020 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173

T-1021 Procedure Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173T-1022 Referencing Code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173

T-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173T-1031 Gages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173

T-1040 Miscellaneous Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1041 Cleanliness. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1042 Openings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1043 Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1044 Pressure/Vacuum (Pressure Limits). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174

T-1050 Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1051 Preliminary Leak Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1052 Test Sequence. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174

T-1060 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1061 Pressure/Vacuum Gages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1062 Temperature Measuring Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174T-1063 Calibration Leak Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174

T-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175T-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

T-1081 Acceptance Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175T-1090 Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

T-1091 Test Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175T-1092 Record Retention. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

Mandatory AppendicesAppendix I Bubble Test — Direct Pressure Technique. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1000 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176

I-1031 Gases. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1032 Bubble Solution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1033 Immersion Bath. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176

I-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1071 Soak Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1072 Surface Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1073 Application of Solution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1074 Immersion in Bath. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1075 Lighting and Visual Aids. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1076 Indication of Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176I-1077 Post-Test Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177

168

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

I-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177I-1081 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177I-1082 Repair/Retest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177

Appendix II Bubble Test — Vacuum Box Technique. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1000 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177

II-1031 Bubble Solution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1032 Vacuum Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1033 Vacuum Source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177

II-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1071 Surface Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1072 Application of Solution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1073 Vacuum Box Placement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177II-1074 Pressure (Vacuum) Retention. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178II-1075 Vacuum Box Overlap. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178II-1076 Lighting and Visual Aids. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178II-1077 Indication of Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178II-1078 Post-Test Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178

II-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178II-1081 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178II-1082 Repair/Retest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178

Appendix III Halogen Diode Detector Probe Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178III-1000 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178III-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178

III-1011 Alkali-Ion Diode (Heated Anode) Halogen Leak Detectors. . . . . . . . . . . . . . . . . . . . . . 178III-1012 Electron Capture Halogen Leak Detectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178

III-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178III-1031 Tracer Gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178III-1032 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178III-1033 Capillary Calibration Leak Standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179

III-1060 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1061 Standard Leak Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1062 Warm Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1063 Scanning Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1064 Detection Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1065 Frequency and Sensitivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179

III-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1071 Location of Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1072 Concentration of Tracer Gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1073 Soak Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179III-1074 Scanning Distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180III-1075 Scanning Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180III-1076 Scanning Direction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180III-1077 Leakage Detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180III-1078 Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180

III-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180III-1081 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180III-1082 Repair/Retest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180

169

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

Appendix IV Helium Mass Spectrometer Test — Detector Probe Technique. . . . . . . . . . . . . . . . . . . 180IV-1000 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180IV-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180IV-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180

IV-1031 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180IV-1032 Auxiliary Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181IV-1033 Calibration Leak Standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181

IV-1060 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181IV-1061 Instrument Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181IV-1062 System Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181

IV-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181IV-1071 Location of Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181IV-1072 Concentration of Tracer Gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181IV-1073 Soak Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181IV-1074 Scanning Distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182IV-1075 Scanning Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182IV-1076 Scanning Direction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182IV-1077 Leakage Detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182IV-1078 Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182

IV-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182IV-1081 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182IV-1082 Repair/Retest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182

Appendix V Helium Mass Spectrometer Test — Tracer Probe Technique. . . . . . . . . . . . . . . . . . . . . 182V-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182V-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182

V-1031 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182V-1032 Auxiliary Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183V-1033 Calibration Leak Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

V-1060 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183V-1061 Instrument Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183V-1062 System Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

V-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184V-1071 Scanning Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184V-1072 Scanning Direction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184V-1073 Scanning Distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184V-1074 Leakage Detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184V-1075 Flow Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

V-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184V-1081 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184V-1082 Repair/Retest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

Appendix VI Pressure Change Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184VI-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184VI-1020 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185

VI-1031 Pressure Measuring Instruments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1032 Temperature Measuring Instruments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185

VI-1060 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1061 Pressure Measuring Instruments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185

170

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

VI-1062 Temperature Measuring Instruments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185

VI-1071 Pressure Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1072 Vacuum Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1073 Test Duration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1074 Small Pressurized Systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1075 Large Pressurized Systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185VI-1076 Start of Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186VI-1077 Essential Variables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186

VI-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186VI-1081 Acceptable Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186VI-1082 Rejectable Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186

Appendix VII Glossary of Terms for Leak Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186VII-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186VII-1020 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186

Appendix VIII Thermal Conductivity Detector Probe Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187VIII-1000 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187VIII-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187

VIII-1011 Thermal Conductivity Leak Detectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187VIII-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187

VIII-1031 Tracer Gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187VIII-1032 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187VIII-1033 Capillary Calibration Leak Standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187

VIII-1060 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1061 Standard Leak Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1062 Warm-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1063 Scanning Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1064 Detection Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1065 Frequency and Sensitivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188

VIII-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1071 Location of Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1072 Concentration of Tracer Gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1073 Soak Times. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1074 Scanning Distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1075 Scanning Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1076 Scanning Direction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188VIII-1077 Leakage Detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189VIII-1078 Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189

VIII-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189VIII-1081 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189VIII-1082 Repair/Retest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189

Appendix IX Helium Mass Spectrometer Test — Hood Technique. . . . . . . . . . . . . . . . . . . . . . . . . . . 190IX-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190IX-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

IX-1031 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190IX-1032 Auxiliary Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190IX-1033 Calibration Leak Standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

IX-1050 Technique. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

171

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

IX-1051 Permeation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190IX-1052 Repetitive or Similar Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

IX-1060 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190IX-1061 Instrument Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190IX-1062 System Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191

IX-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191IX-1071 Standard Technique. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191IX-1072 Alternative Technique. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192

IX-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192IX-1081 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192

IX-1082 Repair/Retest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192

Appendix X Ultrasonic Leak Detector Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193X-1000 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193X-1020 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

X-1021 Written Procedure Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193X-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

X-1031 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193X-1032 Capillary Calibration Leak Standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

X-1060 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193X-1061 Standard Leak Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193X-1062 Warm-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1063 Scanning Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1064 Frequency and Sensitivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194

X-1070 Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1071 Location of Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1072 Soak Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1073 Scanning Distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1074 Scanning Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1075 Leakage Detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194

X-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1081 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1082 Repair/Retest. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194

X-1090 Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1091 Examination Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194X-1092 Performance Demonstration Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194

TablesIII-1031 Tracer Gases. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179VIII-1031 Tracer Gases. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187X-1021 Requirements of an Ultrasonic Leak Testing Procedure. . . . . . . . . . . . . . . . . . . . . . . . . 193

Nonmandatory AppendixAppendix A Supplementary Leak Testing Formula Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195A-10 Applicability of the Formulas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195

172

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

01

ARTICLE 10LEAK TESTING

T-1000 INTRODUCTION

T-1010 SCOPE

This Article describes methods and requirements forthe performance of leak testing.

(a) When a leak testing method or technique ofArticle 10 is specified by a referencing Code Section,the leak test method or technique shall be used togetherwith Article 1, General Requirements.

(b) Definition of terms used in this Article are inMandatory Appendix VII of this Article.

(c) The test methods or techniques of these methodscan be used for the location of leaks or the measurementof leakage rates.

The Nonmandatory Standards of Article 27 may beused in the preparation of leak test specifications. Thespecific test method(s) or technique(s) and Glossary ofTerms of the methods in this Article are described inMandatory Appendices I through X and NonmandatoryAppendix A as follows:

Appendix I — Bubble Test — Direct Pressure Tech-nique

Appendix II — Bubble Test — Vacuum Box Tech-nique

Appendix III — Halogen Diode Detector Probe TestAppendix IV — Helium Mass Spectrometer Test —

Detector Probe TechniqueAppendix V — Helium Mass Spectrometer Test —

Tracer Probe TechniqueAppendix VI — Pressure Change TestAppendix VII — Glossary of TermsAppendix VIII — Thermal Conductivity Detector

Probe TestAppendix IX — Helium Mass Spectrometer Test —

Hood TechniqueAppendix X — Ultrasonic Leak Detector TestAppendix A — Supplementary Leak Testing Formula

Symbols

T-1020 GENERAL

T-1021 Procedure Requirements

Leak testing shall be performed in accordance with

173

a written procedure. Each procedure shall include atleast the following information, as applicable:

(a) extent of the examination;(b) type of equipment to be used for detecting leaks

or measuring leakage rates;(c) surface cleanliness preparation and type of equip-

ment used;(d) method or technique of the test that will be

performed;(e) temperature, pressure, gas, and percent concentra-

tion to be used.

T-1022 Referencing Code

For the leak testing method(s) or technique(s) speci-fied by the referencing Code, the referencing CodeSection shall then be consulted for the following:

(a) personnel qualification/certification(b) technique(s) /calibration standards(c) extent of examination(d) acceptable test sensitivity or leakage rate(e) report requirements(f) retention of records

T-1030 EQUIPMENT

T-1031 Gages

(a) Gage Range.When dial indicating and recordingpressure gage(s) are used in leak testing, they shouldpreferably have the dial(s) graduated over a range ofapproximately double the intended maximum pressure,but in no case shall the range be less than 11⁄2 normore than four times that pressure. These range limitsdo not apply to dial indicating and recording vacuumgages. Range requirements for other types of gagesgiven in an applicable Mandatory Appendix shall beas required by that Appendix.

(b) Gage Location.When components are to bepressure/vacuum leak tested, the dial indicating gage(s)shall be connected to the component or to the componentfrom a remote location, with the gage(s) readily visible

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

T-1031 2001 SECTION V T-1063.2

to the operator controlling the pressure/vacuum through-out the duration of pressurizing, evacuating, testing,and depressurizing or venting of the component. Forlarge vessels or systems where one or more gagesare specified or required, a recording type gage isrecommended, and it may be substituted for one ofthe two or more indicating type gages.

(c) When other types of gage(s) are required by anapplicable Mandatory Appendix, they may be used inconjunction with or in place of dial indicating orrecording type gages.

T-1040 MISCELLANEOUSREQUIREMENTS

T-1041 Cleanliness

The surface areas to be tested shall be free of oil,grease, paint, or other contaminants that might maska leak. If liquids are used to clean the component orif a hydrostatic or hydropneumatic test is performedbefore leak testing, the component shall be dry beforeleak testing.

T-1042 Openings

All openings shall be sealed using plugs, covers,sealing wax, cement, or other suitable material that canbe readily and completely removed after completionof the test. Sealing materials shall be tracer gas free.

T-1043 Temperature

The minimum metal temperature for all componentsduring a test shall be as specified in the applicableMandatory Appendix of this Article or in the referencingCode Section for the hydrostatic, hydropneumatic, orpneumatic test of the pressure component or parts. Theminimum or maximum temperature during the test shallnot exceed that temperature compatible with the leaktesting method or technique used.

T-1044 Pressure/Vacuum (Pressure Limits)

Unless specified in the applicable Mandatory Appen-dix of this Article or by the referencing Code Section,components that are to be pressure-leak tested shallnot be tested at a pressure exceeding 25% of the DesignPressure.

174

T-1050 PROCEDURE

T-1051 Preliminary Leak Test

Prior to employing a sensitive leak testing method,it may be expedient to perform a preliminary test todetect and eliminate gross leaks. This shall be donein a manner that will not seal or mask leaks duringthe specified test.

T-1052 Test Sequence

It is recommended that leak testing be performedbefore hydrostatic or hydropneumatic testing.

T-1060 CALIBRATION

T-1061 Pressure/Vacuum Gages

(a) All dial indicating and recording type gages usedshall be calibrated against a standard deadweight tester,a calibrated master gage, or a mercury column, andrecalibrated at least once a year, when in use, unlessspecified differently by the referencing Code Sectionor Mandatory Appendix. All gages used shall provideresults accurate to within the Manufacturer’s listedaccuracy and shall be recalibrated at any time thatthere is reason to believe they are in error.

(b) When other than dial indicating or recordingtype gages are required by an applicable MandatoryAppendix, they shall be calibrated as required by thatMandatory Appendix or referencing Code Section.

T-1062 Temperature Measuring Devices

When temperature measurement is required by thereferencing Code Section or Mandatory Appendix, thedevice(s) shall be calibrated in accordance with therequirements of that Code Section or Mandatory Ap-pendix.

T-1063 Calibration Leak Standards

T-1063.1 Permeation Type Leak Standard.Thisstandard shall be a calibrated permeation type leakthrough fused glass or quartz. The standard shall havea helium leakage rate in the range of 1 × 10−6 to 1× 10−10 std cm3/sec. (1 × 10−7 to 1 × 10−11 Pa m3/s)

T-1063.2 Capillary Type Leak Standard.This stan-dard shall be a calibrated capillary type leak througha tube. The standard shall have a leakage rate equalto or smaller than the required test sensitivity times

01

01

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

T-1063.2 ARTICLE 10 — LEAK TESTING T-1092

the actual percent test concentration of the selectedtracer gas.

T-1070 TEST

See applicable Mandatory Appendix of this Article.

T-1080 EVALUATION

T-1081 Acceptance Standards

Unless otherwise specified in the referencing CodeSection, the acceptance criteria given for each methodor technique of that method shall apply. The supplemen-tal leak testing formulas for calculating leakage ratesfor the method or technique used are stated in theMandatory Appendices of this Article.

T-1090 DOCUMENTATION

T-1091 Test Report

The test report shall contain, as a minimum, thefollowing information as applicable to the method ortechnique:

175

(a) date of test;(b) certification level and name of operator;(c) test procedure (number) and revision number;(d) test method or technique;(e) test results;(f) component identification;(g) test instrument, standard leak, and material identi-

fication;(h) test conditions, test pressure, tracer gas, and gas

concentration;(i) gage(s) — manufacturer, model, range, and identi-

fication number;(j) temperature measuring device(s) and identifica-

tion number(s);(k) sketch showing method or technique setup.

T-1092 Record Retention

The test report shall be maintained in accordancewith the requirements of the referencing CodeSection.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

ARTICLE 10MANDATORY APPENDICES

APPENDIX I — BUBBLE TEST —DIRECT PRESSURE TECHNIQUE

I-1000 INTRODUCTION

I-1010 SCOPE

The objective of the direct pressure technique ofbubble leak testing is to locate leaks in a pressurizedcomponent by the application of a solution or byimmersion in liquid that will form bubbles as leakagegas passes through it.

I-1030 EQUIPMENT

I-1031 Gases

Unless otherwise specified, the test gas will normallybe air; however, inert gases may be used.

NOTE: When inert gas is used, safety aspects of oxygen deficientatmosphere should be considered.

I-1032 Bubble Solution

(a) The bubble forming solution shall produce a filmthat does not break away from the area to be tested,and the bubbles formed shall not break rapidly due toair drying or low surface tension. Household soap ordetergents are not permitted as substitutes for bubbletesting solutions.

(b) The bubble forming solution shall be compatiblewith the temperature of the test conditions.

I-1033 Immersion Bath

(a) Water or another compatible solution shall beused for the bath.

(b) The immersion solution shall be compatible withthe temperature of the test conditions.

176

I-1070 TEST

I-1071 Soak Time

Prior to examination the test pressure shall be heldfor a minimum of 15 min.

I-1072 Surface Temperature

As a standard technique, the temperature of thesurface of the part to be examined shall not be below40°F (4°C) nor above 125°F (52°C) throughout theexamination. Local heating or cooling is permittedprovided temperatures remain within the range of 40°F(4°C) to 125°F (52°C) during examination. Where itis impractical to comply with the foregoing temperaturelimitations, other temperatures may be used providedthat the procedure is demonstrated.

I-1073 Application of Solution

The bubble forming solution shall be applied to thesurface to be tested by flowing, spraying, or brushingthe solution over the examination area. The number ofbubbles produced in the solution by application shouldbe minimized to reduce the problem of masking bubblescaused by leakage.

I-1074 Immersion in Bath

The area of interest shall be placed below the surfaceof the bath in an easily observable position.

I-1075 Lighting and Visual Aids

When performing the test, the requirements of Article9, T-952 and T-953 shall apply.

I-1076 Indication of Leakage

The presence of continuous bubble growth on thesurface of the material indicates leakage through anorifice passage(s) in the region under examination.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

I-1077 ARTICLE 10 — MANDATORY APPENDICES II-1073

I-1077 Post-Test Cleaning

After testing, surface cleaning may be required forproduct serviceability.

I-1080 EVALUATION

I-1081 Leakage

Unless otherwise specified by the referencing CodeSection, the area under test is acceptable when nocontinuous bubble formation is observed.

I-1082 Repair/Retest

When leakage is observed, the location of the leak(s)shall be marked. The component shall then be depressur-ized, and the leak(s) repaired as required by the referenc-ing Code Section. After repairs have been made, therepaired area or areas shall be retested in accordancewith the requirements of this Appendix.

APPENDIX II — BUBBLE TEST —VACUUM BOX TECHNIQUE

II-1000 INTRODUCTION

II-1010 SCOPE

The objective of the vacuum box technique of bubbleleak testing is to locate leaks in a pressure boundarythat cannot be directly pressurized. This is accomplishedby applying a solution to a local area of the pressureboundary surface and creating a differential pressureacross that local area of the boundary causing theformation of bubbles as leakage gas passes throughthe solution.

II-1030 EQUIPMENT

II-1031 Bubble Solution

(a) The bubble forming solution shall produce a filmthat does not break away from the area to be tested,and the bubbles formed shall not break rapidly due toair drying or low surface tension. The number ofbubbles contained in the solution should be minimizedto reduce the problem of discriminating between existingbubbles and those caused by leakage.

(b) Soaps or detergents designed specifically forcleaning shall not be used for the bubble formingsolution.

177

(c) The bubble forming solution shall be compatiblewith the temperature conditions of the test.

II-1032 Vacuum Box

The vacuum box used shall be of convenient size[e.g., 6 in. (152 mm) wide by 30 in. (762 mm) long]and contain a window in the side opposite the openbottom. The open bottom edge shall be equipped witha suitable gasket to form a seal against the test surface.Suitable connections, valves, lighting, and gage shallbe provided. The gage shall have a range of 0 psi(0 kPa) to 15 psi (103 kPa), or equivalent pressureunits such as 0 in. Hg to 30 in. Hg. The gage rangelimit requirements of T-1031(a) do not apply.

II-1033 Vacuum Source

The required vacuum can be developed in the boxby any convenient method (e.g., air ejector, vacuumpump, or motor intake manifold). The gage shall registera partial vacuum of at least 2 psi (4.1 in. Hg) (13.8 kPa)below atmospheric pressure or the partial vacuum re-quired by the referencing Code Section.

II-1070 TEST

II-1071 Surface Temperature

As a standard technique, the temperature of thesurface of the part to be examined shall not be below40°F (4°C) nor above 125°F (52°C) throughout theexamination. Local heating or cooling is permittedprovided temperatures remain in the range of 40°F(4°C) to 125°F (52°C) during the examination. Where itis impractical to comply with the foregoing temperaturelimitations, other temperatures may be used providedthat the procedure is demonstrated.

II-1072 Application of Solution

The bubble forming solution shall be applied to thesurface to be tested by flowing, spraying, or brushingthe solution over the examination area before placementof the vacuum box.

II-1073 Vacuum Box Placement

The vacuum box shall be placed over the solutioncoated section of the test surface and the box evacuatedto the required partial vacuum.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

II-1074 2001 SECTION V III-1032

II-1074 Pressure (Vacuum) Retention

The required partial vacuum (differential pressure)shall be maintained for at least 10 sec examination time.

II-1075 Vacuum Box Overlap

An overlap of 2 in. (51 mm) minimum for adjacentplacement of the vacuum box shall be used for eachsubsequent examination.

II-1076 Lighting and Visual Aids

When performing the test, the requirements of Article9, T-952 and T-953 shall apply.

II-1077 Indication of Leakage

The presence of continuous bubble growth on thesurface of the material or weld seam indicates leakagethrough an orifice passage(s) in the region under exami-nation.

II-1078 Post-Test Cleaning

After testing, cleaning may be required for productserviceability.

II-1080 EVALUATION

II-1081 Leakage

Unless otherwise specified by the referencing CodeSection, the area under test is acceptable when nocontinuous bubble formation is observed.

II-1082 Repair/Retest

When leakage is observed, the location of the leak(s)shall be marked. The vacuum box shall then be ventedand the leak(s) repaired as required by the referencingCode Section. After repairs have been made, the repairedarea or areas shall be retested in accordance with therequirements of this Appendix.

APPENDIX III — HALOGEN DIODEDETECTOR PROBE TEST

III-1000 INTRODUCTION

The more sophisticated electronic halogen leak detec-tors have very high sensitivity. These instruments makepossible the detection of halogen gas flow from the

178

lower pressure side of a very small opening in anenvelope or barrier separating two regions at differentpressures.

III-1010 SCOPE

The halogen detector probe test method is a semiquan-titative method used to detect and locate leaks, andshall not be considered quantitative.

III-1011 Alkali-Ion Diode (Heated Anode)Halogen Leak Detectors

The alkali-ion diode halogen detector probe instru-ment uses the principle of a heated platinum element(anode) and an ion collector plate (cathode), wherehalogen vapor is ionized by the anode, and the ionsare collected by the cathode. A current proportional tothe rate of ion formation is indicated on a meter.

III-1012 Electron Capture Halogen LeakDetectors

The electron capture halogen detector probe instru-ment uses the principle of the affinity of certain molecu-lar compounds for low energy free electrons usuallyproduced by ionization of gas flow through an elementwith a weak radioactive tritium source. When the gasflow contains halides, electron capture occurs causinga reduction in the concentration of halogen ions presentas indicated on a meter. Non-electron capturing nitrogenor argon is used as background gas.

III-1030 EQUIPMENT

III-1031 Tracer Gas

Gases that may be used are shown in Table III-1031.

III-1031.1 For Alkali-Ion Diode. Halogen leak de-tectors, select a tracer gas from Table III-1031 thatwill produce the necessary test sensitivity.

III-1031.2 For Electron Capture. Halogen leakdetectors, sulfur hexafluoride, SF6, is the recommendedtracer gas.

III-1032 Instrument

An electronic leak detector as described in III-1011or III-1012 shall be used. Leakage shall be indicatedby one or more of the following signaling devices.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

01

01

III-1032 ARTICLE 10 — MANDATORY APPENDICES III-1073

TABLE III-1031TRACER GASES

Commercial Chemical Designation ChemicalDesignation Symbol

Refrigerant-11 Trichloromonofluoromethane CCl3FRefrigerant-12 Dichlorodifluoromethane CCl2F2

Refrigerant-21 Dichloromonofluoromethane CHCl2FRefrigerant-22 Chlorodifluoromethane CHCIF2

Refrigerant-114 Dichlorotetrafluoroethane C2Cl2F4

Refrigerant-134a Tetrafluoroethane C2H2F4

Methylene Chloride Dichloromethane CH2Cl2Sulfur Hexafluoride Sulfur Hexafluoride SF6

(a) Meter — a meter on the test instrument, or aprobe, or both.

(b) Audio Devices— a speaker or set of headphonesthat emits audible indications.

(c) Indicator Light — a visible indicator light.

III-1033 Capillary Calibration Leak Standard

A capillary type leak standard per T-1063.2 using100% tracer gas as selected per III-1031.

III-1060 CALIBRATION

III-1061 Standard Leak Size

The maximum leakage rateQ for the leak standarddescribed in III-1033 containing 100% tracer concentra-tion for use in III-1063 shall be calculated as follows:

Q p Qs%TG100

where Qs is 1 × 10−4 std cm3/s (1 × 10−5 Pa m3/s),unless specified otherwise by the referencing CodeSection, and %TG is the concentration of the tracergas (in %) that is to be used for the test (See III-1072).

III-1062 Warm Up

The detector shall be turned on and allowed to warmup for the minimum time specified by the instrumentmanufacturer prior to calibrating with the leak standard.

III-1063 Scanning Rate

The instrument shall be calibrated by passing theprobe tip across the orifice of the leak standard in III-

179

1061. The probe tip shall be kept within1⁄8 in. (3.2 mm)of the orifice of the leak standard. The scanning rateshall not exceed that which can detect leakage rateQfrom the leak standard. The meter deflection shall benoted or the audible alarm or indicator light set forthis scanning rate.

III-1064 Detection Time

The time required to detect leakage from the leakstandard is the detection time and it should be observedduring system calibration. It is usually desirable tokeep this time as short as possible to reduce the timerequired to pinpoint detected leakage.

III-1065 Frequency and Sensitivity

Unless otherwise specified by the referencing CodeSection, the sensitivity of the detector shall be deter-mined before and after testing and at intervals of notmore than 4 hr during testing. During any calibrationcheck, if the meter deflection, audible alarm, or indicatorlight indicates that the detector cannot detect leakagefrom the leak standard of III-1061, the instrument shallbe recalibrated and areas tested after the last satisfactorycalibration check shall be retested.

III-1070 TEST

III-1071 Location of Test

(a) The test area shall be free of contaminants thatcould interfere with the test or give erroneous results.

(b) The component to be tested shall, if possible,be protected from drafts or located in an area wheredrafts will not reduce the required sensitivity of the test.

III-1072 Concentration of Tracer Gas

The concentration of the tracer gas shall be at least10% by volume at the test pressure, unless otherwisespecified by the referencing Code Section.

III-1073 Soak Time

Prior to examination, the test pressure shall be held aminimum of 30 min. When demonstrated, the minimumallowable soak time may be less than that specifiedabove due to the immediate dispersion of the halogengas when:

(a) a special temporary device (such as a leech box)is used on open components to test short segments;

01

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

01

III-1073 2001 SECTION V IV-1031

(b) components are partially evacuated prior to initialpressurization with halogen gas.

III-1074 Scanning Distance

After the required soak time per III-1073, the detectorprobe tip shall be passed over the test surface. Theprobe tip shall be kept within1⁄8 in. (3.2 mm) of thetest surface during scanning. If a shorter distance isused during calibration, then that distance shall not beexceeded during the examination scanning.

III-1075 Scanning Rate

The maximum scanning rate shall be as determinedin III-1063.

III-1076 Scanning Direction

The examination scan should commence in the upper-most portion of the system being leak tested whileprogressively scanning downward.

III-1077 Leakage Detection

Leakage shall be indicated and detected accordingto III-1032.

III-1078 Application

The following are two examples of applications thatmay be used (note that other types of applications maybe used).

III-1078.1 Tube Examination. To detect leakagethrough the tube walls when testing a tubular heatexchanger, the detector probe tip should be inserted intoeach tube end and held for the time period established bydemonstration. The examination scan should commencein the uppermost portion of the tubesheet tube rowswhile progressively scanning downward.

III-1078.2 Tube-to-Tubesheet Joint Examination.Tube-to-tubesheet joints may be tested by the encapsula-tor method. The encapsulator may be a funnel typewith the small end attached to the probe tip end andthe large end placed over the tube-to-tubesheet joint.If the encapsulator is used, the detection time is deter-mined by placing the encapsulator over the orifice onthe leak standard and noting the time required for anindicated instrument response.

180

III-1080 EVALUATION

III-1081 Leakage

Unless otherwise specified by the referencing CodeSection, the area tested is acceptable when no leakageis detected that exceeds the allowable rate of 1 × 10−4

std cm3/s (1 × 10−5 Pa m3/s).

III-1082 Repair/Retest

When unacceptable leakage is detected, the locationof the leak(s) shall be marked. The component shallthen be depressurized, and the leak(s) repaired asrequired by the referencing Code Section. After repairshave been made, the repaired area or areas shall beretested in accordance with the requirements of thisAppendix.

APPENDIX IV — HELIUM MASSSPECTROMETER TEST —

DETECTOR PROBE TECHNIQUE

IV-1000 INTRODUCTION

IV-1010 SCOPE

This technique describes the use of the helium massspectrometer to detect minute traces of helium gas inpressurized components. The high sensitivity of thisleak detector makes possible the detection of heliumgas flow from the lower pressure side of a very smallopening in an envelope or barrier separating two regionsat different pressures, or the determination of the pres-ence of helium in any gaseous mixture. The detectorprobe is a semiquantitative technique used to detect andlocate leaks, and shall not be considered quantitative.

IV-1030 EQUIPMENT

IV-1031 Instrument

A helium mass spectrometer leak detector capableof sensing and measuring minute traces of helium shallbe used. Leakage shall be indicated by one or moreof the following signaling devices.

(a) Meter — a meter on, or attached to, the testinstrument.

(b) Audio Devices— a speaker or set of headphonesthat emits audible indications.

(c) Indicator Light — a visible indicator light.

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

01

IV-1032 ARTICLE 10 — MANDATORY APPENDICES IV-1073

IV-1032 Auxiliary Equipment

(a) Transformer.A constant voltage transformer shallbe used in conjunction with the instrument when linevoltage is subject to variations.

(b) Detector Probe.All areas to be examined shallbe scanned for leaks using a detector probe (sniffer)connected to the instrument through flexible tubing ora hose. To reduce instrument response and clean uptime, the tubing or hose length shall be less than 15ft (4.6 m), unless the test setup is specifically designedto attain the reduced response and clean up time forlonger tubing or hose lengths.

IV-1033 Calibration Leak Standards

Calibration leak standards may be either a permeationor capillary type standard per T-1063.1 and T-1063.2.The type of leak standard used shall be established bythe instrument or system sensitivity requirement, or asspecified by the referencing Code Section.

IV-1060 CALIBRATION

IV-1061 Instrument Calibration

IV-1061.1 Warm Up. The instrument shall be turnedon and allowed to warm up for the minimum timespecified by the instrument manufacturer prior to cali-brating with the calibrated leak standard.

IV-1061.2 Calibration. Calibrate the helium massspectrometer per the instruments manufacturer’s opera-tion and maintenance manual, using a permeation typeleak standard as stated in T-1063.1 to establish thatthe instrument is at optimum or adequate sensitivity.The instrument shall have a sensitivity of at least 1 ×10−9 std cm3/s (1 × 10−10 Pa m3/s) for helium.

IV-1062 System Calibration

IV-1062.1 Standard Leak Size.The maximum leak-age rateQ for the leak standard described in IV-1033,containing 100% helium concentration for use in IV-1062.2, shall be calculated as follows:

Q p Qs%TG100

where Qs is 1 × 10−4 std cm3/s (1 × 10−5 Pa m3/s),unless specified otherwise by the referencing CodeSection, and%TG is the concentration of the tracer

181

gas (in %) that is to be used for the test (See IV-1072).

IV-1062.2 Scanning Rate.After connecting the de-tector probe to the instrument, the system shall becalibrated by passing the detector probe tip across theorifice of the leak standard in IV-1062.1. The probetip shall be kept within1⁄8 in. (3.2 mm) of the orificeof the leak standard. The scanning rate shall not exceedthat which can detect leakage rateQ from the leakstandard. The meter deflection shall be noted the audiblealarm or indicator light set for this scanning rate.

IV-1062.3 Detection Time. The time required todetect leakage from the leak standard is the detectiontime, and it should be observed during system calibra-tion. It is usually desirable to keep this time as shortas possible to reduce the time required to pinpointdetected leakage.

IV-1062.4 Frequency and Sensitivity.Unless other-wise specified by the referencing Code Section, thesystem sensitivity shall be determined before and aftertesting and at intervals of not more than 4 hr duringthe test. During any calibration check, if the meterdeflection, audible alarm, or visible light indicates thatthe system cannot detect leakage per IV-1062.2, thesystem, and if necessary, the instrument, shall be recali-brated and all areas tested after the last satisfactorycalibration check shall be retested.

IV-1070 TEST

IV-1071 Location of Test

The component to be tested shall, if possible, beprotected from drafts or located in an area where draftswill not reduce the required sensitivity of the test.

IV-1072 Concentration of Tracer Gas

The concentration of the helium tracer gas shall beat least 10% by volume at the test pressure, unlessotherwise specified by the referencing Code Section.

IV-1073 Soak Time

Prior to testing, the test pressure shall be held aminimum of 30 min. The minimum allowable soaktime may be less than that specified above due to theimmediate dispersion of the helium gas when:

(a) a special temporary device (such as a leech box)is used on open components to test short segments;

01

01

01

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

01

IV-1073 2001 SECTION V V-1031

(b) components are partially evacuated prior to initialpressurization with helium gas.

IV-1074 Scanning Distance

After the required soak time per IV-1073, the detectorprobe tip shall be passed over the test surface. Theprobe tip shall be kept within1⁄8 in. (3.2 mm) of thetest surface during scanning. If a shorter distance isused during system calibration, then that distance shallnot be exceeded during test scanning.

IV-1075 Scanning Rate

The maximum scanning rate shall be as determinedin IV-1062.2.

IV-1076 Scanning Direction

The examination scan should commence in the lower-most portion of the system being tested while progres-sively scanning upward.

IV-1077 Leakage Detection

Leakage shall be indicated and detected accordingto IV-1031.

IV-1078 Application

The following are two examples of applications thatmay be used (note that other types of applications maybe used).

IV-1078.1 Tube Examination. To detect leakagethrough the tube walls when testing a tubular heatexchanger, the detector probe tip should be inserted intoeach tube end and held for the time period established bydemonstration. The examination scan should commencein the lowermost portion of the tubesheet tube rowswhile progressively scanning upward.

IV-1078.2 Tube-to-Tubesheet Joint Examination.Tube-to-tubesheet joints may be tested by the encapsula-tor method. The encapsulator may be a funnel typewith the small end attached to the probe tip end andthe large end placed over the tube-to-tubesheet joint.If the encapsulator is used, the detection time is deter-mined by placing the encapsulator over the orifice onthe leak standard and noting the time required for anindicated instrument response.

182

IV-1080 EVALUATION

IV-1081 Leakage

Unless otherwise specified by the referencing CodeSection, the area tested is acceptable when no leakageis detected that exceeds the allowable rate of 1 × 10−4

std cm3/s (1 × 10−5 Pa m3/s).

IV-1082 Repair /Retest

When unacceptable leakage is detected, the locationof the leak(s) shall be marked. The component shallthen be depressurized, and the leak(s) repaired asrequired by the referencing Code Section. After repairshave been made, the repaired area or areas shall beretested in accordance with the requirements of thisAppendix.

APPENDIX V — HELIUM MASSSPECTROMETER TEST — TRACER

PROBE TECHNIQUE

V-1010 SCOPE

This technique describes the use of the helium massspectrometer to detect minute traces of helium gas inevacuated components.

The high sensitivity of this leak detector, when tracerprobe testing, makes possible the detection and locationof helium gas flow from the higher pressure side ofvery small openings through the evacuated envelopeor barrier separating the two regions at different pres-sures. This is a semiquantitative technique and shallnot be considered quantitative.

V-1030 EQUIPMENT

V-1031 Instrument

A helium mass spectrometer leak detector capableof sensing and measuring minute traces of helium shallbe used. Leakage shall be indicated by one or moreof the following signaling devices.

(a) Meter — a meter on or attached to the testinstrument.

(b) Audio Devices— a speaker or set of headphonesthat emits audible indications.

(c) Indicator Light — a visible indicator light.

01

01

01

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

01

01

01

V-1032 ARTICLE 10 — MANDATORY APPENDICES V-1062.5

V-1032 Auxiliary Equipment

(a) Transformer.A constant voltage transformer shallbe used in conjunction with the instrument when linevoltage is subject to variations.

(b) Auxiliary Pump System.When the size of thetest system necessitates the use of an auxiliary vacuumpump system, the ultimate absolute pressure and pumpspeed capability of that system shall be sufficient toattain required test sensitivity and response time.

(c) Manifold. A system of pipes and valves withproper connections for the instrument gages, auxiliarypump, calibration leak standard, and test component.

(d) Tracer Probe.Tubing connected to a source of100% helium with a valved fine opening at the otherend for directing a fine stream of helium gas.

(e) Vacuum Gage(s).The range of vacuum gage(s)capable of measuring the absolute pressure at whichthe evacuated system is being tested. The gage(s) forlarge systems shall be located on the system as far aspossible from the inlet to the pump system.

V-1033 Calibration Leak Standard

A capillary type leak standard per T-1063.2 with amaximum helium leakage rate of 1 × 10−5 std cm3/s(1 × 10−6 Pa m3/s) shall be used unless otherwisespecified by the referencing Code Section.

V-1060 CALIBRATION

V-1061 Instrument Calibration

V-1061.1 Warm Up. The instrument shall be turnedon and allowed to warm up for the minimum timespecified by the instrument manufacturer prior to cali-brating with the calibration leak standard.

V-1061.2 Calibration. Calibrate the helium massspectrometer per the instruments manufacturer’s opera-tion and maintenance manual, using a permeation typeleak standard as stated in T-1063.1 to establish thatthe instrument is at optimum or adequate sensitivity.The instrument shall have a sensitivity of at least 1 ×10−9 std cm3/s (1 × 10−10 Pa m3/s) for helium.

V-1062 System Calibration

V-1062.1 Standard Leak Size.The calibrated leakstandard, as stated in V-1033, shall be attached to thecomponent as far as possible from the instrumentconnection to the component. The leak standard shallremain open during system calibration.

183

V-1062.2 Scanning Rate.With the component evacu-ated to an absolute pressure sufficient for connectionof the helium mass spectrometer to the system, thesystem shall be calibrated for the test by passing thetracer probe tip across the orifice of the leak standard.The probe tip shall be kept within1⁄4 in. (6 mm) ofthe orifice of the leak standard. For a known flow ratefrom the tracer probe of 100% helium, the scanningrate shall not exceed that which can detect leakagethrough the calibration leak standard into the test system.

V-1062.3 Detection Time. The time required todetect leakage from the leak standard is the detectiontime, and it should be observed during system calibra-tion. It is desirable to keep this time as short as possibleto reduce the time required to pinpoint detected leakage.

V-1062.4 Frequency and Sensitivity.Unless other-wise specified by the referencing Code Section, thesystem sensitivity shall be determined before and aftertesting and at intervals of not more than 4 hr duringtesting. During any calibration check, if the meterdeflection, audible alarm, or visible light indicates thatthe system cannot detect leakage per V-1062.2, thesystem, and if necessary, the instrument, shall be recali-brated and all areas tested after the last satisfactorycalibration check shall be retested.

V-1062.5 Hood Technique.A calibrated leakCLstandard as per T-1063.1 with 100% helium shall beattached, where feasible, to the component as far aspossible from the instrument connection to the compo-nent. The calibrated leak standard shall remain openduring system calibration until the response time hasbeen determined.

(a) Evacuation.With the component evacuated toan absolute pressure sufficient for connection of thehelium mass spectrometer to the system, the calibratedleak standard shall be opened to the system. Thecalibrated leak standard shall remain open until theinstrument signal becomes stable, and the response timehas been determined.

(b) Response Time.The time is recorded when thecalibrated leak standard is opened to the componentand when the increase in output signal becomes stable.The elapsed time between the two readings is theresponse time. The stable instrument reading shall benoted and recorded asM1 in divisions.

(c) Background Reading.1 BackgroundM2 in divi-sions is established after determining response time.The calibration leak standard is closed to the system

1 System background noise. For definition of symbols, see Nonmanda-tory Appendix A.

01

01

01

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

V-1062.5 2001 SECTION V VI-1010

and the instrument reading shall be recorded when itbecomes stable.

(d) Preliminary Calibration.The preliminary systemsensitivity shall be calculated as follows:

S1 pCL

M1 − M2p std cm3 / sec / div

The calibration shall be repeated when there is anychange in the leak detector setup (e.g., a change inthe portion of helium bypassed to the auxiliary pump,if used) or any change in the calibrated leak. Thecalibrated leak standard shall be isolated from thesystem upon completing the preliminary system sensitiv-ity calibration.

(e) Final Calibration. Upon completing the test ofthe system, and with the component still under thehood, the instrument output readingM3 shall be deter-mined with the calibrated leak closed. Again, the cali-brated leak shall be opened into the system beingtested. The increase in instrument outputM4 shallbe used in calculating the final system sensitivity asfollows:

S2 pCL

M4 − M3std cm3 / sec / div

If the final system sensitivityS2 has decreased belowthe initial system sensitivityS1 by more than 35%, theinstrument shall be cleaned and/or repaired, recalibratedand the component or system retested.

(f) Measured Leakage Rate.The measured leakagerate of the component shall be determined as follows:

Q1 p S2(M3 − M2) std cm3 / sec

(g) Actual Leakage Rate.Calculation of actual leak-age rate (corrected for tracer gas concentration used):

Q2 pQ1 × 100

% Hestd cm3 / sec

V-1070 TEST

V-1071 Scanning Rate

The maximum scanning rate shall be as determinedin V-1062.2.

184

V-1072 Scanning Direction

The examination scan should commence in the upper-most portion of the system being tested while progres-sively scanning downward.

V-1073 Scanning Distance

The tracer probe tip shall be kept within1⁄4 in.(6 mm) of the test surface during scanning. If a shorterdistance is used during system calibration, then thatdistance shall not be exceeded during the examinationscanning.

V-1074 Leakage Detection

Leakage shall be indicated and detected accordingto V-1031.

V-1075 Flow Rate

The minimum flow rate shall be as set in V-1062.2.

V-1080 EVALUATION

V-1081 Leakage

Unless otherwise specified by the referencing CodeSection, the area tested is acceptable when no leakageis detected that exceeds the allowable rate of 1 × 10−5

std cm3/s (1 × 10−6 Pa m3/s).

V-1082 Repair /Retest

When unacceptable leakage is detected, the locationof the leak(s) shall be marked. The component shallthen be vented, and the leak(s) repaired as requiredby the referencing Code Section. After repairs havebeen made, the repaired area or areas shall be retestedin accordance with the requirements of this Appendix.

APPENDIX VI — PRESSURECHANGE TEST

VI-1010 SCOPE

This test method describes the techniques for de-termining the leakage rate of the boundaries of a closedcomponent or system at a specific pressure or vacuum.Pressure hold, absolute pressure, maintenance of pres-sure, pressure loss, pressure decay, pressure rise, andvacuum retention are examples of techniques that maybe used whenever pressure change testing is specified

01

01

01

01

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

VI-1010 ARTICLE 10 — MANDATORY APPENDICES VI-1075

as a means of determining leakage rates. The testsspecify a maximum allowable change in either pressureper unit of time, percentage volume, or mass changeper unit of time.

VI-1020 GENERAL

Each of the tests shall be performed by either pneu-matic pressurization or evacuation of a closed compo-nent or system to a specific pressure or vacuum. Temper-ature, pressure, or vacuum are systematically recordedfor a specified period of time. Analysis of data deter-mines the component or system acceptability with re-spect to leakage rate or pressure change per unit of time.

VI-1030 EQUIPMENT

VI-1031 Pressure Measuring Instruments

(a) Gage Range.Dial indicating and recording typegages shall meet the requirements of T-1031(a). Liquidmanometers or quartz Bourdon tube gages may be usedover their entire range.

(b) Gage Location.The location of the gage(s) shallbe that stated in T-1031(b).

(c) Types of Gages.Regular or absolute gages maybe used in pressure change testing. When greater accu-racy is required, quartz Bourdon tube gages or liquidmanometers may be used. The gage(s) used shall havean accuracy, resolution, and repeatability compatiblewith the acceptance criteria.

VI-1032 Temperature Measuring Instruments

Dry bulb or dew point temperature measuring instru-ments, when used, shall have accuracy, repeatability, andresolution compatible with the leakage rate acceptancecriteria.

VI-1060 CALIBRATION

VI-1061 Pressure Measuring Instruments

All dial indicating, recording, and quartz Bourdontube gages shall be calibrated per T-1061(b). Thescale of liquid manometers shall be calibrated againststandards that have known relationships to nationalstandards, where such standards exist.

185

VI-1062 Temperature Measuring Instruments

Calibration for dry bulb and dew point temperaturemeasuring instruments shall be against standards thathave known relationships to national standards, wheresuch standards exist.

VI-1070 TEST

VI-1071 Pressure Application

Components that are to be tested above atmosphericpressure shall be pressurized per T-1044.

VI-1072 Vacuum Application

Components that are to be tested under vacuum shallbe evacuated to at least 2 psi (4.1 in. Hg) (13.8 kPa)below atmospheric pressure or as required by the refer-encing Code Section.

VI-1073 Test Duration

The test pressure (or vacuum) shall be held for theduration specified by the referencing Code Section or,if not specified, it shall be sufficient to establish theleakage rate of the component system within the accu-racy or confidence limits required by the referencingCode Section. For very small components or systems,a test duration in terms of minutes may be sufficient.For large components or systems, where temperatureand water vapor corrections are necessary, a test durationin terms of many hours may be required.

VI-1074 Small Pressurized Systems

For temperature stabilization of very small pressurizedsystems, such as gasket interspaces, where only system(metal) temperature can be measured, at least 15 minshall elapse after completion of pressurization and beforestarting the test.

VI-1075 Large Pressurized Systems

For temperature stabilization of large pressurizedsystems where the internal gas temperature is measuredafter completion of pressurization, it shall be determinedthat the temperature of the internal gas has stabilizedbefore starting the test.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

VI-1076 2001 SECTION V VII-1020

VI-1076 Start of Test

At the start of the test, initial temperature and pressure(or vacuum) readings shall be taken and thereafter atregular intervals, not to exceed 60 min, until the endof the specified test duration.

VI-1077 Essential Variables

(a) When it is required to compensate for barometricpressure variations, measurement of the test pressureshall be made with either an absolute pressure gageor a regular pressure gage and a barometer.

(b) When it is required by the referencing CodeSection, or when the water vapor pressure variationcan significantly affect the test results, the internal dewpoint temperature or relative humidity shall be measured.

VI-1080 EVALUATION

VI-1081 Acceptable Test

When the pressure change or leakage rate is equalto or less than that specified by the referencing CodeSection, the test is acceptable.

VI-1082 Rejectable Test

When the pressure change or leakage rate exceedsthat specified by the referencing Code Section, theresults of the test are unsatisfactory. Leak(s) may belocated by other methods described in the MandatoryAppendices. After the cause of the excessive pressurechange or leakage rate has been determined and repairedin accordance with the referencing Code Section, theoriginal test shall be repeated.

NOTE: For more information regarding this method of testing referto the following:

(a) 10 CFR 50, Appendix J,Primary Containment Leakage Testingfor Water Cooled Power Reactors.

(b) ANSI/ANS 56.8-1981,American National Standard Contain-ment System Leakage Testing Requirements,published by the Ameri-can Nuclear Society.

APPENDIX VII — GLOSSARY OFTERMS FOR LEAK TESTING

VII-1010 SCOPE

This Mandatory Appendix is used for the purposeof establishing standard terms and definitions of termswhich appear in Article 10, Leak Testing.

186

VII-1020 GENERAL

(a) ASTM E 1316, Standard Terminology for Nonde-structive Examinations, has been adopted by the Com-mittee as SE-1316.

(b) SE-1316 Section 8 provides the definitions ofterms listed in (e).

(c) For general terms such asDiscontinuity, Evalua-tion, Flaw, Indication, Inspection,etc., refer to Article1, Mandatory Appendix I.

(d) The following SE-1316 terms are used in conjunc-tion with this Article: absolute pressure; backgroundsignal; gage pressure; gas; halogen; halogen leakdetector; hood test; leak; leakage rate; leak testing;mass spectrometer; mass spectrometer leak detector;sampling probe; standard leak; tracer gas; vacuum.

(e) The following Code terms, which are in additionto SE-1316, and are Code specific, are used in conjunc-tion with this Article.

background reading— seebackground signalin (d)calibration leak standard— seestandard leakin (d)detector probe— seesampling probein (d)dew point temperature— that temperature at which

the gas in a system would be capable of holding nomore water vapor and condensation in the form of dewwould occur

dry bulb temperature— the ambient temperature ofthe gas in a system

halogen diode detector— seehalogen leak detectorin (d)

helium mass spectrometer— seemass spectrometerand mass spectrometer leak detectorin (d)

hood technique— seehood testin (d)immersion bath— a low surface tension liquid into

which a gas containing enclosure is submerged to detectleakage which forms at the site or sites of a leak or leaks

immersion solution— see immersion bathinert gas— a gas that resists combining with other

substances. Examples are helium, neon, and argon.instrument calibration— introduction of a known

size standard leak into an isolated leak detector forthe purpose of determining the smallest size leakagerate of a particular gas at a specific pressure andtemperature that the leak detector is capable of indicatingfor a particular division on the leak indicator scale

leakage — the fluid, either liquid or gas, flowingthrough a leak and expressed in units of mass flow;i.e., pressure and volume per time

leak standard— seestandard leakin (d)quartz Bourdon tube gage— this high accuracy

gage is a servonulling differential pressure measuringelectronic instrument. The pressure transducing elementis a one piece fused quartz Bourdon element.

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

VII-1020 ARTICLE 10 — MANDATORY APPENDICES VIII-1033

regular pressure— seegage pressurein (d)sensitivity — the size of the smallest leakage rate

that can be unambiguously detected by the leak testinginstrument, method, or technique being used

soak time— the elapsed time between when thedesired differential pressure is attained on a systemand the time when the test technique is performed todetect leakage or measure leakage rate

standard dead weight tester— a device for hydrau-lically balancing the pressure on a known high accuracyweight against the reading on a pressure gage for thepurpose of calibrating the gage

system calibration— introduction of a known sizestandard leak into a test system with a leak detectorfor the purpose of determining the smallest size leakagerate of a particular gas at a specific pressure andtemperature that the leak detector as part of the testsystem is capable of indicating for a particular divisionon the leak indicator scale

thermal conductivity detector— a leak detector thatresponds to differences in the thermal conductivity ofa sampled gas and the gas used to zero it (i.e.,background atmosphere)

vacuum box— a device used to obtain a pressuredifferential across a weld that cannot be directly pressur-ized. It contains a large viewing window, special easyseating and sealing gasket, gage, and a valved connec-tion for an air ejector, vacuum pump, or intake manifold

water vapor— gaseous form of water in a system

APPENDIX VIII — THERMALCONDUCTIVITY DETECTOR

PROBE TEST

VIII-1000 INTRODUCTION

These instruments make possible the detection of atracer gas flow from the lower pressure side of a verysmall opening in an envelope or barrier separating tworegions at different pressures.

VIII-1010 SCOPE

The thermal conductivity detector probe test methodis a semiquantitative method used to detect and locateleaks, and shall not be considered quantitative.

VIII-1011 Thermal Conductivity Leak Detectors

The thermal conductivity detector probe instrumentuses the principle that the thermal conductivity of agas or gas mixture changes with any change in the

187

TABLE VIII-1031TRACER GASES

ChemicalDesignation Chemical Designation Symbol

. . . Helium He

. . . Argon Ar

. . . Carbon Dioxide CO2

Refrigerant-11 Trichloromonofluoromethane CCl3FRefrigerant-12 Dichlorodifluoromethane CCl2F2

Refrigerant-21 Dichloromonofluoromethane CHCl2FRefrigerant-22 Chlorodifluoromethane CHClF2

Refrigerant-114 Dichlorotetrafluoroethane C2Cl2F4

Refrigerant-134a Tetrafluoroethane C2H2F4

Methylene Chloride Dichloromethane CH2Cl2Sulfur Hexafluoride Sulfur Hexafluoride SF6

concentration(s) of the gas or gas mixture (i.e., theintroduction of a tracer gas in the area of a leak).

VIII-1030 EQUIPMENT

VIII-1031 Tracer Gas

In principle, any gas having a thermal conductivitydifferent from air can be used as a tracer gas. Thesensitivity achievable depends on the relative differencesof the thermal conductivity of the gases [i.e., backgroundair (air used to zero the instrument) and the sampledair (air containing the tracer gas) in the area of a leak].Table VIII-1031 lists some of the typical tracer gasesused. The tracer gas to be used shall be selected basedon the required test sensitivity.

VIII-1032 Instrument

An electronic leak detector as described in VIII-1011 shall be used. Leakage shall be indicated by oneor more of the following signaling devices:

(a) Meter.A meter on the test instrument, or a probe,or both.

(b) Audio Devices.A speaker or sets of headphonesthat emit(s) audible indications.

(c) Indicator Light. A visible indicator light.

VIII-1033 Capillary Calibration Leak Standard

A capillary type leak standard per T-1063.2 using100% tracer gas as selected per VIII-1031.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

01

01

01

01

VIII-1060 2001 SECTION V VIII-1076

VIII-1060 CALIBRATION

VIII-1061 Standard Leak Size

The maximum leakage rateQ for the leak standarddescribed in VIII-1033 containing 100% tracer concen-tration for use in VIII-1063 shall be calculated asfollows:

Q p Qs% TG100

where Qs [in std cm3/s (Pa m3/s)] is the required testsensitivity and%TG is the concentration of the tracergas (in %) that is to be used for the test. See VIII-1072.

VIII-1062 Warm-up

The detector shall be turned on and allowed to warmup for the minimum time specified by the instrumentmanufacturer prior to calibrating with the leak standard.

VIII-1063 Scanning Rate

The detector shall be calibrated by passing the probetip across the orifice of the leak standard in VIII-1061.The probe tip shall be kept within1⁄2 in. (13 mm) ofthe orifice of the leak standard. The scanning rate shallnot exceed that which can detect leakage rateQ fromthe leak standard. The meter deflection shall be notedor the audible alarm or indicator light set for thisscanning rate.

VIII-1064 Detection Time

The time required to detect leakage from the leakstandard is the detection time and it should be observedduring system calibration. It is usually desirable tokeep this time as short as possible to reduce the timerequired to pinpoint detected leakage.

VIII-1065 Frequency and Sensitivity

Unless otherwise specified by the referencing CodeSection, the sensitivity of the detector shall be deter-mined before and after testing and at intervals of notmore than 4 hr during testing. During any calibrationcheck, if the meter deflection, audible alarm, or indicatorlight indicate that the detector cannot detect leakageper VIII-1063, the instrument shall be recalibrated andareas tested after the last satisfactory calibration checkshall be retested.

188

VIII-1070 TEST

VIII-1071 Location of Test

(a) The test area shall be free of contaminants thatcould interfere with the test or give erroneous results.

(b) The component to be tested shall, if possible,be protected from drafts or located in an area wheredrafts will not reduce the required sensitivity of the test.

VIII-1072 Concentration of Tracer Gas

The concentration of the tracer gas shall be at least10% by volume at the test pressure, unless otherwisespecified by the referencing Code Section.

VIII-1073 Soak Times

Prior to examination, the test pressure shall be held aminimum of 30 min. When demonstrated, the minimumallowable soak time may be less than that specifiedabove due to the immediate dispersion of the tracergas when:

(a) a special temporary device (such as a leech box)is used on open components to test short segments;

(b) components are partially evacuated prior to initialpressurization with tracer gas.

VIII-1074 Scanning Distance

After the required soak time per VIII-1073, thedetector probe tip shall be passed over the test surface.The probe tip shall be kept within1⁄2 in. (13 mm) ofthe test surface during scanning. If a shorter distanceis used during calibration, then that distance shall notbe exceeded during the examination scanning.

VIII-1075 Scanning Rate

The maximum scanning rate shall be as determinedin VIII-1063.

VIII-1076 Scanning Direction

For tracer gases that are lighter than air, the examina-tion scan should commence in the lowermost portionof the system being tested while progressively scanningupward. For tracer gases that are heavier than air, theexamination scan should commence in the uppermostportion of the system being tested while progressivelyscanning downward.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

VIII-1077 ARTICLE 10 — MANDATORY APPENDICES VIII-1082

01 VIII-1077 Leakage Detection

Leakage shall be indicated and detected accordingto VIII-1032.

VIII-1078 Application

The following are two examples of applications thatmay be used (note that other types of applications maybe used).

VIII-1078.1 Tube Examination. To detect leakagethrough the tube walls when testing a tubular heatexchanger, the detector probe tip should be insertedinto each tube and held for the time period establishedby demonstration.

VIII-1078.2 Tube-to-Tubesheet Joint Examination.Tube-to-tubesheet joints may be tested by the encapsula-tor method. The encapsulator may be a funnel typewith the small end attached to the probe tip end andthe large end placed over the tube-to-tubesheet joint.If the encapsulator is used, the detection time is deter-

189

mined by placing the encapsulator over the orifice onthe leak standard and noting the time required for anindicated instrument response.

VIII-1080 EVALUATION

VIII-1081 Leakage

Unless otherwise specified by the referencing CodeSection, the area tested is acceptable when no leakageis detected that exceeds the maximum leakage rateQ,determined per VIII-1061.

VIII-1082 Repair/Retest

When unacceptable leakage is detected, the locationof the leak(s) shall be marked. The component shallthen be depressurized, and the leak(s) repaired asrequired by the referencing Code Section. After repairshave been made, the repaired area or areas shall beretested in accordance with the requirements of thisAppendix.

01

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01 ARTICLE 10 — APPENDIX IXHELIUM MASS SPECTROMETER TEST — HOOD

TECHNIQUE

IX-1010 SCOPE

This technique describes the use of the helium massspectrometer to respectively detect and measure minutetraces of helium gas in evacuated components.

The high sensitivity of this leak detector, when hoodtesting, makes possible the detection and measurementof total helium gas flow from the higher pressure sideof all hooded, very small openings through the evacuatedenvelope or barrier that separates the two regions atdifferent pressures. This is a quantitative measurementtechnique.

IX-1030 EQUIPMENT

IX-1031 Instrument

A helium mass spectrometer leak detector capableof sensing and measuring minute traces of helium shallbe used. Leakage shall be indicated by a meter on orattached to the test instrument.

IX-1032 Auxiliary Equipment

(a) Transformer.A constant voltage transformer shallbe used in conjunction with the instrument when linevoltage is subject to variations.

(b) Auxiliary Pump System.When the size of thetest system necessitates the use of an auxiliary vacuumpump system, the ultimate absolute pressure and pumpspeed capability of that system shall be sufficient toattain required test sensitivity and response time.

(c) Manifold. A system of pipes and valves withproper connections for the instrument gages, auxiliarypump, calibration leak standard, and test component.

(d) Hood. Any suitable envelope or container, suchas a plastic bag, with a through aperture for themanifold.

(e) Vacuum Gage(s).The range of vacuum gage(s)capable of measuring the absolute pressure at which

190

the evacuated system is being tested. The gage(s) forlarge systems shall be located on the system as far aspossible from the inlet to the pump system.

IX-1033 Calibration Leak Standard

A permeation type leak standard per T-1063.1 witha maximum helium leakage rate of 1 × 10−6 std cm3/s (1 × 10−7 Pa m3/s) shall be used, unless specifiedotherwise by the referencing Code Section.

IX-1050 TECHNIQUE

IX-1051 Permeation

When systems with long response times (i.e., lowhelium mass spectrometer throughput) are to be tested,helium permeation through nonmetallic seals can leadto false results. In cases like this, it is recommended,if possible, to locally hood test such seals or excludethem from the hood if the seals are not required tobe tested.

IX-1052 Repetitive or Similar Tests

For repetitive tests or where the test time is knownfrom previous similar tests, the preliminary calibration,per IX-1062.4, may be omitted.

IX-1060 CALIBRATION

IX-1061 Instrument Calibration

IX-1061.1 Warm Up. The instrument shall be turnedon and allowed to warm up for the minimum timespecified by the instrument manufacturer prior to cali-brating with the leak standard.

IX-1061.2 Calibration. Calibrate the helium massspectrometer per the instrument manufacturer’s opera-tion and maintenance manual using a permeation type

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

IX-1061.2 ARTICLE 10 — MANDATORY APPENDICES IX-1071.5

leak standard as stated in T-1063.1 to establish thatthe instrument is at optimum or adequate sensitivity.The instrument shall have sensitivity of at least 1 ×10−9 std cm3/s (1 × 10−10 Pa m3/s) for helium.

IX-1062 System Calibration

IX-1062.1 Standard Leak Size.A calibrated leakCL standard as per T-1063.1 with 100% helium shallbe attached, where feasible, to the component as faras possible from the instrument connection to thecomponent.

IX-1062.2 Response Time.With the componentevacuated to an absolute pressure sufficient for connec-tion of the helium mass spectrometer to the system,the system shall be calibrated by opening the leakstandard to the system. The leak standard shall remainopen until the instrument signal becomes stable.

The time shall be recorded when the leak standardis first opened to the component and again when theincrease in output signal becomes stable. The elapsedtime between the two readings is the response time. Thestable instrument reading shall be noted and recorded asM1 in divisions.

IX-1062.3 Background Reading.1 BackgroundM2

in divisions is established after determining responsetime. The leak standard shall be closed to the systemand the instrument reading shall be recorded when itbecomes stable.

IX-1062.4 Preliminary Calibration. The preliminarysystem sensitivity shall be calculated as follows:

S1 pCL

M1 − M2p std cm3 / s / div (Pa m3 / s / div)

The calibration shall be repeated when there is anychange in the leak detector setup (e.g., a change inthe portion of helium bypassed to the auxiliary pump,if used) or any change in the leak standard. Theleak standard shall be isolated from the system uponcompleting the preliminary system sensitivity cali-bration.

IX-1062.5 Final Calibration. Upon completing thetest of the system per IX-1071.4, and with the compo-nent still under the hood, the leak standard shall beagain opened into the system being tested. The increasein instrument output shall be noted and recorded as

1 System background noise. For definition of symbols, see Nonmanda-tory Appendix A.

191

M4 in divisions and used in calculating the final systemsensitivity as follows:

S2 pCL

M4 − M3p std cm3 / s / div (Pa m3 / s / div)

If the final system sensitivityS2 has decreased belowthe preliminary system sensitivityS1 by more than35%, the instrument shall be cleaned and/or repaired,recalibrated, and the component retested.

IX-1070 TEST

IX-1071 Standard Technique

IX-1071.1 Hood. For a single wall component orpart, the hood (envelope) container may be made ofa material such as plastic.

IX-1071.2 Filling of Hood with Tracer Gas. Aftercompleting preliminary calibration per IX-1062.4, thespace between the component outer surface and thehood shall be filled with helium.

IX-1071.3 Estimating or Determining Hood TracerGas Concentration. The tracer gas concentration inthe hood enclosure shall be determined or estimated.

IX-1071.4 Test Duration.After filling the hood withhelium, the instrument outputM3 in divisions shall benoted and recorded after waiting for a test time equalto the response time determined in IX-1062.2 or, ifthe output signal has not become stable, until the outputsignal stabilizes.

IX-1071.5 System Measured Leakage Rate.Aftercompleting final calibration per IX-1062.5, the systemleakage rate shall be determined as follows:

(a) For tests where no change in output signal occurs(i.e., M2 p M3), the system leakage rate shall bereported as being “below the detectable range of thesystem” and the item under test passes.

(b) For tests where the output signal (M3) remainson scale, the leakage rate shall be determined as follows:

Q pS2 (M3 − M2) × 100

%TGstd cm3 / s (Pa m3 / s)

where %TG is the concentration of the tracer gas (in%) in the hood. See IX-1071.3.

(c) For tests where the output signal (M3) exceedsthe detectable range of the system (i.e., output signalis off scale), the system leakage rate shall be reported

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

IX-1071.5 2001 SECTION V IX-1082

as being “greater than the detectable range of thesystem” and the item under test fails.

IX-1072 Alternative Technique

IX-1072.1 System Correction Factor.For heliummass spectrometer leak indicator meters in leakage rateunits, a System Correction Factor (SCF) may be utilizedif it is desired to utilize the acutal indicator meterleakage rate units in lieu of converting the readingsto divisions [e.g., the values ofM1, M2, M3, and M4

are directly read from the helium mass spectrometerin std cm3/s (Pa m3/s)].

IX-1072.2 Alternative Formulas. The following for-mulas shall be used in lieu of those described inIX-1062:

(a) Preliminary Calibration (per IX-1062.4).The pre-liminary system correction factor (PSCF) shall be calcu-lated as follows:

PSCFp CL / (M1 − M2)

(b) Final Calibration (per IX-1062.5).The final sys-tem correction factor (FSCF) shall be calculated asfollows:

FSCFp CL / (M4 − M3)

If the FSCF has decreased below the PSCF by more than

192

35%, the instrument shall be cleaned and/or repaired,recalibrated, and the component retested.

(c) System Measured Leakage Rate (per IX-1071.5).The system leakage rate shall be determined as follows:

Q p(FSCF × M3) × 100

%TGstd cm3 / s (Pa m3 / s)

IX-1080 EVALUATION

Unless otherwise specified by the referencing CodeSection, the component tested is acceptable when themeasured leakage rateQ is equal to or less than 1 ×10−6 std cm3/s (1 × 10−7 Pa m3/s) of helium.

IX-1081 Leakage

When the leakage rate exceeds the permissible value,all welds or other suspected areas shall be retestedusing a tracer probe technique. All leaks shall bemarked and temporarily sealed to permit completionof the tracer probe retest. The temporary seals shallbe of a type which can be readily and completelyremoved after testing has been completed.

IX-1082 Repair/Retest

The component shall then be vented and the leak(s)repaired as required by the referencing Code Section.After repairs have been made, the repaired area or areasshall be retested in accordance with the requirements ofthis Appendix.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01 ARTICLE 10 — APPENDIX XULTRASONIC LEAK DETECTOR TEST

X-1000 INTRODUCTION

This technique describes the use of an ultrasonicleak detector to detect the ultrasonic energy producedby the flow of a gas from the lower pressure side ofa very small opening in an envelope or barrier separatingtwo regions at different pressures.

(a) Due to the low sensitivity [maximum sensitivityof 10−2 std cm3/s (10−3 Pa m3/s)] of this technique, itshould not be utilized for the acceptance testing ofvessels that will contain lethal or hazardous substances.

(b) This is a semiquantitative method used to detectand locate leaks and shall not be considered quantitative.

X-1020 GENERAL

X-1021 Written Procedure Requirements

X-1021.1 Requirements.Ultrasonic leak testing shallbe performed in accordance with a written procedurewhich shall, as a minimum, contain the requirementslisted in Table X-1021. The written procedure shallestablish a single value, or range of values, for eachrequirement.

X-1021.2 Procedure Qualification.When procedurequalification is specified, a change of a requirement inTable X-1021 identified as anessentialvariable, fromthe specified value, or range of values, shall requirerequalification of the written procedure. A change ofa requirement identified as anonessentialvariable fromthe specified value, or range of values, does not requirerequalification of the written procedure. All changesof the essential and nonessential elements from thevalue, or range of values, specified by the writtenprocedure shall require revision of, or an addendumto, the written procedure.

X-1030 EQUIPMENT

X-1031 Instrument

An electronic ultrasonic leak detector capable ofdetecting acoustic energy in the range of 20 to 100

193

TABLE X-1021REQUIREMENTS OF AN ULTRASONIC LEAK TESTING

PROCEDURE

Non-Essential Essential

Requirement Variable Variable

Instrument Manufacturer & Model XLeak standard (decrease in size only) XPressurizing gas XTest pressure (decrease in pressure only) XSoak time (decrease in time only) XScanning distance (increase in distance only) XScanning rate (increase in rate only) XSignaling device XPersonnel Qualification X

kHz shall be utilized. Leakage shall be indicated byone or more of the following signaling devices:

(a) Meter — A meter on the test instrument, or aprobe, or both.

(b) Audio Device —A set of headphones that emit(s)audible indications.

X-1032 Capillary Calibration Leak Standard

A capillary type leak standard per Article 10, T-1063.2.

X-1060 CALIBRATION

X-1061 Standard Leak Size

The maximum leakage rateQ for the leak standardin X-1032 shall be 1 × 10−1 std cm3/s (1 × 10−2 Pam3/s), unless otherwise specified by the referencingCode Section.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

X-1062 2001 SECTION V X-1092

X-1062 Warm-up

The detector shall be turned on and allowed to warmup for the minimum time specified by the instrumentmanufacturer prior to calibration.

X-1063 Scanning Rate

The leak standard shall be attached to a pressureregulated gas supply and the pressure set to that to beused for the test. The detector shall be calibrated bydirecting the detector/probe towards the leak standardat the maximum scanning distance to be utilized duringtesting and noting the meter deflection and/or pitch ofthe audible signal as the detector/probe is scannedacross the leak standard. The scanning rate shall notexceed that which can detect leakage rateQ from theleak standard.

X-1064 Frequency and Sensitivity

Unless otherwise specified by the referencing CodeSection, the sensitivity of the detector shall be verifiedbefore and after testing, and at intervals of not morethan 4 hr during testing. During any verification check,should the meter deflection or audible signal indicatethat the detector/probe cannot detect leakage per X-1063, the instrument shall be recalibrated and areastested after the last satisfactory calibration check shallbe retested.

X-1070 TEST

X-1071 Location of Test

The component to be tested shall, if possible, beremoved or isolated from other equipment or structuresthat could generate ambient or system noise that candrown out leaks.

X-1072 Soak Time

Prior to testing, the test pressure shall be held aminimum of 15 min.

X-1073 Scanning Distance

After the required soak time per X-1072, the detectorshall be passed over the test surface. The scanningdistance shall not exceed that utilized to determine themaximum scanning rate in X-1063.

194

X-1074 Scanning Rate

The maximum scanning rate shall be as determinedin X-1063.

X-1075 Leakage Detection

Leakage shall be indicated and detected accordingto X-1031.

X-1080 EVALUATION

X-1081 Leakage

Unless otherwise specified by the referencing CodeSection, the area tested is acceptable when no leakageis detected that exceeds the allowable rate of 1 × 10−1

std cm3/s (1 × 10−2 Pa m3/s).

X-1082 Repair/Retest

When unacceptable leakage is detected, the locationof the leak(s) shall be marked. The component shallthen be depressurized, and the leak(s) repaired asrequired by the referencing Code Section. After repairshave been made, the repaired area or areas shall beretested in accordance with the requirements of thisAppendix.

X-1090 DOCUMENTATION

X-1091 Examination Report

The report of examination shall contain the followinginformation:

(a) procedure ID and revision;(b) manufacturer and model number of instrument;(c) calibration leak standard size and serial number;(d) pressurizing gas and test pressure;(e) soak time;(f) maximum scanning distance and rate;(g) map or record of rejectable leakage or areas

cleared;(h) identification of welds or areas scanned;(i) examination personnel identity;(j) date of examination.

X-1092 Performance Demonstration Report

Performance demonstrations, when required by thereferencing Code Section, shall be documented.

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services

01

ARTICLE 10NONMANDATORY APPENDIX

APPENDIX A — SUPPLEMENTARYLEAK TESTING FORMULA

SYMBOLS

A-10 APPLICABILITY OF THEFORMULAS

(a) The formulas in this Article provide for thecalculated leak rate(s) for the technique used.

(b) The symbols defined below are used in theformulas of the appropriate Appendix.

(1) System sensitivity calculation:S1ppreliminary sensitivity (calculation of sensi-

tivity), std cm3/s /div (Pa m3/s /div)S2pfinal sensitivity (calculation of sensitivity),

std cm3/s /div (Pa m3/s /div)(2) System measured leakage rate calculation:

Qpmeasured leakage rate of the system (cor-rected for tracer gas concentration), stdcm3/s (Pa m3/s)

(3) System Correction Factors:

195

PSCFpPreliminary System CorrectionFactorFSCFpFinal System Correction Factor

(4) Tracer gas concentration:%TG p concentration of Tracer Gas, %

(5) Calibrated standard:CLpcalibrated leak leakage rate, std cm3/s

(Pa m3/s)(6) Instrument reading sequence:

M1pmeter reading before test with calibratedleak open to the component [divisions, orstd cm3/s (Pa m3/s)]

M2pmeter reading before test with calibratedleak closed to component [divisions, or stdcm3/s (Pa m3/s)] (system background noisereading)

M3pmeter reading (registering component leak-age) with calibrated leak closed [divisions,or std cm3/s (Pa m3/s)]

M4pmeter reading (registering component leak-age) with calibrated leak open [divisions,or std cm3/s (Pa m3/s)]

COPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling ServicesCOPYRIGHT American Society of Mechanical EngineersLicensed by Information Handling Services