aula-254 gold aula-254 gold - envea...analytical technique is commonly known as cold vapor atomic...

Click here to load reader

Upload: others

Post on 02-Aug-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

  • Rinse station

    Turn-table

    Rinse pump

    Photometer

    Thermoelectricdehumidifi er

    Crossfl owreactor

    Rinse solution SnCl2Drain

    Argon

    NH3OHCl

    KMnO4

    T=98°C

    Schematic fl ow diagram: AULA-254 Gold with Automatic Sample Digester ASD

    MAIN APPLICATIONS

    Environmental monitoring: Water, soils, sludge ...

    Analysis of foodstuffs: fi sh, innards, plants ...

    Medicine: urine, blood, saliva, hair ...

    Chemical industry: process monitoring, quality control ...

    Geochemistry, mining

    Petrochemistry

    Metallurgy and material testing

    Applicable to EPA Methods: 245.1, 245.7, 1631, 7470, 7471A . . . .

    SPECIFIC FEATURES

    Fully automated fl ow analysis system

    Compact modular construction

    Integrated GoldTrap for maximum of sensitivity:

  • ENVEA (Headquarters)111 Bd Robespierre / CS 8000478304 Poissy CEDEX 4 - FRANCE +33(0)1 39 22 38 00 [email protected]

    Visit us at:www.envea.global

    +33(0)1 39 22 38 00 [email protected]

    TECHNICAL SPECIFICATIONS

    Measuring principle:UV absorption (CVAAS),wavelength = 253.7 nm

    Principle of preconcentration:Amalgamation on gold, thermal desorption by rapid heating (MI GoldTrap)

    UV source:Electrodeless low-pressure mercury lamp (EDL)

    Stabilization method: Reference beam method

    Optical cell:Fused silica (Suprasil) l = 230 mmheated, approx. 45°C

    Measuring range:• 10 ng/l to 50 μg/l (GoldTrap off)• 1 ng/l to 5 μg/l (GoldTrap on)

    Detection limit:•