av-qa-sop-021 sop for polarimeter
TRANSCRIPT
-
7/27/2019 AV-QA-SOP-021 SOP for Polarimeter
1/3
Department: Quality AssuranceDOC. No. : AV/QA/SOP/021
Page No. : Page 1 of 3
Issue Date: 01.03.06 Revision No. & Date : (00)-01.03.06
Subject Operating Procedure for Polarimeter.
REVISIONS
Revision No.
RevisedPages
Revision Description Date of Revision
00 0 Original Issue 01.03.06
Compiled By Approved By Authorized by
Document Controller Management Representative Factory Manager
CLASSIFIED: Internal Use Only
-
7/27/2019 AV-QA-SOP-021 SOP for Polarimeter
2/3
Department: Quality AssuranceDOC. No. : AV/QA/SOP/021
Page No. : Page 2 of 3
Issue Date: 01.03.06 Revision No. & Date : (00)-01.03.06
Subject Operating Procedure for Polarimeter.
1.0 Purpose:
To ensure the proper operation of Polarimeter.
2.0 Scope :
The SOP is applicable for the operation of Polarimeter
3.0 Responsibility
HOD Quality
QC Chemist
4.0 Definition:
SOP: Standard Operating Operating Procedure.
5.0 Procedure:
1. Reading the Scale:
The scale seen in the telescope is divided at 1o intervals from 1 to 3600 adjacent to 0 and
below the main scale is the international sugar scale that may be used directly in the
evaluation of sugar solutions.
2. Use of Micro meter Drum:
1. The micrometer drum is employed to subdivide the angular and I.S.S. scale and itself
carries two scales, angular on the left subdivided of intervals of .05 and I.S.S. on the
right subdivided at intervals of 0.1.
2. The index marker is at the top of the aperture for both scales. It is noted that movement
of the drum from 0 to 1.0 cause the line seen in eyepiece to traverse exactly one scale
division respectively for the I.S.S. and angular scales.
3. Using the knurled wheel of the micrometer drum rotate it to bring the index line
coincident with a scale line. The true reading is then the scale line concerned plus the
drum reading indicated on the appropriate scale.
3. Sample Tubes & Filling:
1) To obtain accurate results, great care must be exercised in filling the tube and
ensuring that the end windows are not strained thus causing erroneous results.
2) Remove both end caps of the tube and detach the glass window which must be
carefully cleaned. The ends of the tube must now be cleaned and dried.3) Place one window over the end of the tube and replace the caps.
4) Carefully screw the cap down until the first resistance appears. Do not screw any
further. Invert the tube and ensuring that the open end is dry, pour in the sample
carefully until a meniscus appears bulging above the tubes end.
5) There should be no air bubble, if there is any air bubble that is supplied in bubble
trap.
4. Adjusting to Zero:
1. With no sample in the through set the field to a precise match observed in
telescope. Set the micrometer drum to zero and observed the scale in
telescope.
2. Grasp the knurled washer and with the other hand slacken the knurled
nut. The drum is now released from the screw.
Compiled By Approved By Authorized by
Document Controller Management Representative Factory Manager
CLASSIFIED: Internal Use Only
-
7/27/2019 AV-QA-SOP-021 SOP for Polarimeter
3/3
Department: Quality AssuranceDOC. No. : AV/QA/SOP/021
Page No. : Page 3 of 3
Issue Date: 01.03.06 Revision No. & Date : (00)-01.03.06
Subject Operating Procedure for Polarimeter.
3. Turn the screw by mean of knurled washer until the index line coincides
with 0 on the scale. Ensure that the drum still reads zero, if not bring it tothis position by moving it through the rectangular aperture in the sleeve.
4. Check the balance and that the scale and drum read zero.
5. Operation Mode:
1. Prior to test the operator ascertains that the instrument with empty
observation tube inserted is at zero position. The field of view show three
comparison fields of equally gray shade, separated from each other by
two fine vertical dividing line and the scales read zero.
2. Thereafter the solution poured into the observation tube and place on the
through.
3. Now the operator observes three field of different shade.
4. The analyzer with triple shadow effect and the graduated circle scale now
be turned by mean of the knob until the three field of comparison showagain equally gray shade.
5. The reference graticule on the scale plate now indicate the angle of
rotation both for dextro rotation and laevo rotation.
6.0 Documentation :
The results recorded by Polarimeter.
7.0 Reference:
-Instrument manual
-In house operating procedure
Compiled By Approved By Authorized by
Document Controller Management Representative Factory Manager
CLASSIFIED: Internal Use Only