b-1c tube failure metallurgical assessment by betz
TRANSCRIPT
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
1/13
GE Power & WaterWater & Process Technologies
Metallurgy Services 9669 Grogan’s Mill Road The Woodlands, Texas 77380281-367-6201281-363-7794 Fax
METALLURGICAL LAB REPORT
Representative: Robert Stewart Plant: Petrojam Ltd.Location: Kingston. JamaicaUnit: B1C, 150 psigReport No.: 2014-0030Date: March 13, 2014
BACKGROUND
Two tube sections from the B1C Boiler at the subject account were submitted for failure
analysis. When the tubing was removed on January 13, 2014, the time in service was five
years.
RESULTS
Figure 1 is a photograph showing the submitted tubing. The samples were not labeled
and, for the purposes of this report, were arbitrarily identified as Tubes A and B. Tube A
exhibited perforation (Figures 2 and 3). The external surface of Tube B was covered with
tan and gray deposit, no failure was observed (Figure 4).
The tubes were split lengthwise to facilitate examination of the internal surface. The
internal surface were covered with light-gray, white and brown deposit (Figures 5 and 6).
Test sections were removed from both tubes to determine the deposit weight density
(DWD) values. Micrometer measurements taken before and after the deposit removal
step were used to measure the deposit thickness (Figures 7 and 8). The following results
were obtained:
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
2/13
Page 2 of 13
DWD (g/ft 2)
Maximum Deposit
Thickness (in.)
Tube A 33 0.006
Tube B 41 0.007
Scanning electron microscope-energy dispersive x-ray analysis (SEM-EDXA) was used
to determine the elemental composition of the external and internal deposit in the
submitted tubes. The semi-quantitative results (Table 1) indicate the external deposits
were composed of mostly iron, calcium and sulfur species. Minor amounts of vanadium,
sodium, chlorine, and silicon were also detected. The presence of sulfur, chlorine,
sodium, and vanadium are known to cause fireside corrosion via molten salt corrosion.
The internal deposit on the waterside (Table 2) consists primarily of iron, calcium, and
sulfur compounds. Minor amount of sodium, magnesium, chlorine, vanadium, and silicon
were also detected, along with trace amounts of other constituents.
.
The external and internal surfaces were mechanically cleaned to facilitate examination of
the underlying metal. There was evidence of corrosion pitting on the internal surface after
cleaning (Figures 9 – 11). The maximum internal surface pit depth measured in Tubes Aand B was 0.014 inches and 0.018 respectively. Figures 13 and 14 showed the external
surface of Tubes A and B after mechanical cleaning, which exhibited general/fireside
corrosion.
At the failure edge of Tube A, the wall thickness was measured to be 0.003 inches, which
represents an approximate 95% loss of metal thickness as compared to the tube nominal
thickness of 0.055 inches. The combined wall loss of Tube B due to external surface
corrosion and internal pitting was 0.030 inches, or 57%, when compared to the maximum
thickness of 0.070 inches.
Transverse sections were cut out from selected areas of the tubes and prepared for
metallographic inspection. Figures 15 – 19 show the microstructure along the internal and
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
3/13
Page 3 of 13
external surface of the tubing, which exhibited general corrosion/corrosion pitting. The
internal pitting appeared consistent with dissolved oxygen corrosion in the tubing. The
mid-wall microstructures of the tubing consisted of ferrite/pearlite, normal for carbon
steel boiler tubing in the as-manufactured condition. No overheating was observed in the
tubing.
CONCLUSIONS
1. The failure of Tube A was caused by combination fireside corrosion (over 95%
loss of wall thickness estimated in the failed region), and general internal
corrosion pitting within the thinned area.
2. Tube B also exhibited combined wall loss due to external surface corrosion and
internal corrosion pitting (57% wall loss, when compared to the maximum wall
thickness).
3. The maximum internal pitting was 0.014 inches and 0.018 inches respectively in
Tubes A and B. The pitting was likely caused by dissolved oxygen corrosion.
4. The composition of the fireside deposit in the thinned regions was varied, withsome vanadium, sulfur-compounds, and trace chloride salts noted to be present.
Vanadium, sulfur, sodium, and chlorides can combine to produce aggressive
molten salts, to flux protective iron oxide scales on the tube fireside surfaces,
resulting in accelerated metal loss.
5. Microstructural analysis indicated that no overheating occurred in the tubing.
Michael AdeosunMetallurgical Engineer
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
4/13
Page 4 of 13
Element
Tube A ODWhite(wt %)
Tube A ODBrown(wt %)
Tube B ODTop
(wt %)
Tube B ODBottom(wt %)
Iron 93.4 38.7 41.5 96.4
Calcium – 27.1 26.2 –
Sodium 3.1 1.1 0.5 –
Vanadium – 3.0 2.9 –
Sulfur 2.6 26.3 26.2 2.8
Silicon – 2.4 0.9 –
Chlorine 0.6 – – –
Chromium – – – 0.5
Aluminum 0.4 1.4 1.8 0.4Data is normalized with carbon and oxygen excluded
TABLE 1. SEMI-QUANTITATIVE ELEMENTAL SEM-EDXA OF THETUBES EXTERNAL SURFACE DEPOSIT
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
5/13
Page 5 of 13
Element
Tube A IDWhite(wt %)
Tube A IDBrown(wt %)
Tube B IDBrown(wt %)
Tube B IDGray
(wt %)
Calcium 31.7 9.2 0.7 0.5
Iron 30.0 80.1 94.5 93.1
Sulfur 27.9 3.2 – 0.8
Vanadium 3.4 3.8 – –
Silicon 2.3 1.9 3.2 4.5
Nickel 1.9 – – –
Aluminum 1.6 1.3 0.1 0.5
Chromium – – – 0.7
Sodium 0.5 – 0.6 –
Magnesium 0.3 0.6 0.5 – Data is normalized with carbon and oxygen excluded.
TABLE 2. SEMI-QUANTITATIVE ELEMENTAL SEM-EDXA OF THETUBES INTERNAL SURFACE DEPOSIT
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
6/13
Page 6 of 13
Figure 1. Photograph showing thesubmitted tube samples as-received forevaluation.
Figure 2. Photograph showing theexternal surface of Tube A, at the failure.
Figure 3. Photograph showing a crosssection of Tube A, at the failure (arrow).
GE Power & Water Water & Process Technologies
Tube A
Tube B
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
7/13
Page 7 of 13
Figure 4. Photograph showing theexternal surface of Tube B.
Figure 5. Photograph showing theinternal surface of Tube A at the failure(arrow).
Figure 6. Photograph showing theinternal surface of Tube B.
GE Power & Water Water & Process Technologies
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
8/13
Page 8 of 13
Figure 7. Photograph of the internalsurface of DWD test section, Tube A.
Figure 8. Photograph of the internalsurface of DWD test section, Tube B.
Figure 9. Photograph of the internalsurface of DWD test section, Tube A,after mechanical cleaning.
GE Power & Water Water & Process Technologies
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
9/13
Page 9 of 13
Figure 10. Photograph of the internalsurface of DWD test section, Tube B,after mechanical cleaning.
Figure 11. Photograph showing theexternal surface of Tube B, aftermechanical cleaning.
Figure 12. Photograph showing theinternal surface of Tube B, aftercleaning.
GE Power & Water Water & Process Technologies
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
10/13
Page 10 of 13
Figure 13. Photograph showing theexternal surface of Tube A, at the failure(arrow), after cleaning
Figure 14.. Photograph showing theexternal surface of Tube B, aftermechanical cleaning.
Figure 15. Photomicrograph showing thesteel microstructure at the failure edge ofTube A. Nital Etch. 50x.
GE Power & Water Water & Process Technologies
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
11/13
Page 11 of 13
Figure 16. Photomicrograph showing theexternal and internal surfaces of Tube A,adjacent to the failure. Nital Etch. 100x.
Figure 17. Photomicrograph showingthe mid-wall steel microstructure ofTube A. Nital Etch. 500x.
Figure 18. Photomicrograph showing theinternal surface steel microstructureobserved on Tube B. Nital Etch. 50x.
GE Power & Water Water & Process Technologies
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
12/13
Page 12 of 13
Figure 19. Photomicrograph showing themid-wall steel microstructure of Tube B.Nital Etch, 500x.
GE Power & Water Water & Process Technologies
-
8/18/2019 B-1C Tube Failure Metallurgical Assessment by BETz
13/13
Page 13 of 13
Notes