b0774d01
TRANSCRIPT
-
7/28/2019 B0774d01
1/15
Spark Optical Emission Spectroscopy
Dr Heath Bagshaw CMA, Trinity College Dublin
Analytical
Workshop 2012
-
7/28/2019 B0774d01
2/15
Why Use Spark OES Analysis?
Quick, reliable and reproducible analysis technique
Can analyse wide range of elements
Usually used for metals\alloy analysis, such as :-
Steel, cast iron and high alloyed steels
Non-ferrous metals and their alloysAI: wrought alloys, casting alloys, etc.
Cu: bronze, brass, cupronickel, etc.
Mg, Zn alloys, solders
Nitrogen in steel
P in aluminiumUltra low carbon analysis
Elements such as Se, La, Te, etc.
Good limit of detection for most materials - generally less then 50 ppm although 10
ppm detection limits are typical.
Higher resolution spectrometers (using PMTs) have limits of detection of
-
7/28/2019 B0774d01
3/15
Analytical
Workshop 2012
When an element is burned it burns with a distinct colour
Colours - Flame Testing
Barium Strontium Potassium
These colours are due to the emission spectrum of the element.
-
7/28/2019 B0774d01
4/15
Basic Theory
Analytical
Workshop 2012
-
7/28/2019 B0774d01
5/15
The emitted radiation\light is split using a prism or diffraction grating to produce a
spectrum.
More Basic Theory
Analytical
Workshop 2012
-
7/28/2019 B0774d01
6/15
modern equipment uses a slightly different set up:-
A diffraction grating is used to produce the spectrum and mirrors focus the radiation
onto detectors.
Basics cont
Analytical
Workshop 2012
-
7/28/2019 B0774d01
7/15
Analytical
Workshop 2012
Emission Spectra
In the case of optical emission spectroscopy the EM radiation is in the visible\optical
region of the EMS which we see as different colours depending on the
wavelength\energy of the emitted photon.
The energy of a photon and its wavelength are related by the following equation:-
where h is Planck's constant (6.626 10-34 joules) and c is the
speed of light (2.998 108 m/s).
hcE=
-
7/28/2019 B0774d01
8/15
Analytical
Workshop 2012
Examples of Emission Spectra
Emission spectrum of Fe
Emission spectrum of H
-
7/28/2019 B0774d01
9/15
Emitted Light
Polychromatic light
ComputerCalculation of results
Comparison to Standards
Data output and storage
Analytical
Workshop 2012
The System
CCD measures thelight intensity at
each wavelength
Diffraction Grating,
Produces a
dispersion spectrum
Samplesparked to
produce a
burn
-
7/28/2019 B0774d01
10/15
Entrance Slit
Grating
130 800 nm
Analytical
Workshop 2012
The Spectrometer (Schematic) The diffraction grating Is like a concave mirror with lines ruled on it.
Modern Gratings have up to 4000 rules/mm.
The incoming light is reflected and diffracted, producing the spectrum
-
7/28/2019 B0774d01
11/15
Analytical
Workshop 2012
The Spectrometer (Real)
Entrance Slit
130 800 nm
The CCD chips simultaneously detect all incident light and determine the intensity of
each wavelength.
-
7/28/2019 B0774d01
12/15
The Sample
Generally speaking the sample is metallic, flat and freshly ground.
The sample is clamped in place and sparked to generate a spectrum
this leaves burn marks on the sample:-
If the sample does not burn correctly spurious results can be produced.
Poor burning can be caused by poor grinding and rough\uneven surfaces.
Analytical
Workshop 2012
Burn marks
Sample
-
7/28/2019 B0774d01
13/15
Analytical
Workshop 2012
Analysis The Intensity of an emission line (colour) is proportional to concentration allows
measurement of how much of each element is present.
A number of standards are run first to set up a calibration curve, these take into
account any matrix matching difficulties (i.e. overlap of elements in some materials).
Take the example of Fe
Requires six standards to correctly calibrate the instrument
for Fe analysis.
-
7/28/2019 B0774d01
14/15
AnalyticalWorkshop 2012
Analysis Continued
Once calibration is completed numerous samples can be analysed.
The sample is simply clamped into place, sparked and a spectrum collected.
The spectrometer collects the intensity of light at all wavelengths and compares this
to the values for the calibration standard. This gives an accurate value of the elements
present in the sample.
Multiple sparks are collected until concordant results are obtained within an
acceptable standard deviation.
Further samples of the same alloy type can then be analysed.
Different alloys require re-calibration before analysis can occur.
-
7/28/2019 B0774d01
15/15
System Types
Spark Optical Emission Spectrometers come in all shapes and sizes for a multitude of
uses.
Foundry\steel making applications usually floor standing to analyse composition of
metals during production quality control.
Portable units often used in manufacturing plants to assess
metal composition.
Portable units to analyse large immovable materials
Hand held units to assist in scrap sorting
The smaller the unit the less accuracy In the measurements
so convenience over quality still useful as analysis can be
Performed anywhere.