b0774d01

Upload: mythos1976

Post on 03-Apr-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 B0774d01

    1/15

    Spark Optical Emission Spectroscopy

    Dr Heath Bagshaw CMA, Trinity College Dublin

    Analytical

    Workshop 2012

  • 7/28/2019 B0774d01

    2/15

    Why Use Spark OES Analysis?

    Quick, reliable and reproducible analysis technique

    Can analyse wide range of elements

    Usually used for metals\alloy analysis, such as :-

    Steel, cast iron and high alloyed steels

    Non-ferrous metals and their alloysAI: wrought alloys, casting alloys, etc.

    Cu: bronze, brass, cupronickel, etc.

    Mg, Zn alloys, solders

    Nitrogen in steel

    P in aluminiumUltra low carbon analysis

    Elements such as Se, La, Te, etc.

    Good limit of detection for most materials - generally less then 50 ppm although 10

    ppm detection limits are typical.

    Higher resolution spectrometers (using PMTs) have limits of detection of

  • 7/28/2019 B0774d01

    3/15

    Analytical

    Workshop 2012

    When an element is burned it burns with a distinct colour

    Colours - Flame Testing

    Barium Strontium Potassium

    These colours are due to the emission spectrum of the element.

  • 7/28/2019 B0774d01

    4/15

    Basic Theory

    Analytical

    Workshop 2012

  • 7/28/2019 B0774d01

    5/15

    The emitted radiation\light is split using a prism or diffraction grating to produce a

    spectrum.

    More Basic Theory

    Analytical

    Workshop 2012

  • 7/28/2019 B0774d01

    6/15

    modern equipment uses a slightly different set up:-

    A diffraction grating is used to produce the spectrum and mirrors focus the radiation

    onto detectors.

    Basics cont

    Analytical

    Workshop 2012

  • 7/28/2019 B0774d01

    7/15

    Analytical

    Workshop 2012

    Emission Spectra

    In the case of optical emission spectroscopy the EM radiation is in the visible\optical

    region of the EMS which we see as different colours depending on the

    wavelength\energy of the emitted photon.

    The energy of a photon and its wavelength are related by the following equation:-

    where h is Planck's constant (6.626 10-34 joules) and c is the

    speed of light (2.998 108 m/s).

    hcE=

  • 7/28/2019 B0774d01

    8/15

    Analytical

    Workshop 2012

    Examples of Emission Spectra

    Emission spectrum of Fe

    Emission spectrum of H

  • 7/28/2019 B0774d01

    9/15

    Emitted Light

    Polychromatic light

    ComputerCalculation of results

    Comparison to Standards

    Data output and storage

    Analytical

    Workshop 2012

    The System

    CCD measures thelight intensity at

    each wavelength

    Diffraction Grating,

    Produces a

    dispersion spectrum

    Samplesparked to

    produce a

    burn

  • 7/28/2019 B0774d01

    10/15

    Entrance Slit

    Grating

    130 800 nm

    Analytical

    Workshop 2012

    The Spectrometer (Schematic) The diffraction grating Is like a concave mirror with lines ruled on it.

    Modern Gratings have up to 4000 rules/mm.

    The incoming light is reflected and diffracted, producing the spectrum

  • 7/28/2019 B0774d01

    11/15

    Analytical

    Workshop 2012

    The Spectrometer (Real)

    Entrance Slit

    130 800 nm

    The CCD chips simultaneously detect all incident light and determine the intensity of

    each wavelength.

  • 7/28/2019 B0774d01

    12/15

    The Sample

    Generally speaking the sample is metallic, flat and freshly ground.

    The sample is clamped in place and sparked to generate a spectrum

    this leaves burn marks on the sample:-

    If the sample does not burn correctly spurious results can be produced.

    Poor burning can be caused by poor grinding and rough\uneven surfaces.

    Analytical

    Workshop 2012

    Burn marks

    Sample

  • 7/28/2019 B0774d01

    13/15

    Analytical

    Workshop 2012

    Analysis The Intensity of an emission line (colour) is proportional to concentration allows

    measurement of how much of each element is present.

    A number of standards are run first to set up a calibration curve, these take into

    account any matrix matching difficulties (i.e. overlap of elements in some materials).

    Take the example of Fe

    Requires six standards to correctly calibrate the instrument

    for Fe analysis.

  • 7/28/2019 B0774d01

    14/15

    AnalyticalWorkshop 2012

    Analysis Continued

    Once calibration is completed numerous samples can be analysed.

    The sample is simply clamped into place, sparked and a spectrum collected.

    The spectrometer collects the intensity of light at all wavelengths and compares this

    to the values for the calibration standard. This gives an accurate value of the elements

    present in the sample.

    Multiple sparks are collected until concordant results are obtained within an

    acceptable standard deviation.

    Further samples of the same alloy type can then be analysed.

    Different alloys require re-calibration before analysis can occur.

  • 7/28/2019 B0774d01

    15/15

    System Types

    Spark Optical Emission Spectrometers come in all shapes and sizes for a multitude of

    uses.

    Foundry\steel making applications usually floor standing to analyse composition of

    metals during production quality control.

    Portable units often used in manufacturing plants to assess

    metal composition.

    Portable units to analyse large immovable materials

    Hand held units to assist in scrap sorting

    The smaller the unit the less accuracy In the measurements

    so convenience over quality still useful as analysis can be

    Performed anywhere.