background information · 2018. 6. 22. · mercaptan (ch 3 sh)). the current alberta ambient air...

17
1 | Page H 2 S and TRS Measurements Continuous and Integrated Methods Prepared by Sanjay Prasad, Yu-Mei Hsu, Michael Martineau, Kendra Thomas Wood Buffalo Environmental Association April 25, 2018 1. Background information Hydrogen sulfide (H2S) is a colourless gas which has the odour of rotten eggs. H2S is considered an odour nuisance at low levels, and can result in discomforting physiological symptoms of headache and nausea. Total Reduced Sulfur (TRS) is a collection of H2S and a number of other sulfur compounds (e.g Carbonyl Sulphide (COS), Carbon Disulphide (CS2) and Methyl Mercaptan (CH3SH)). The current Alberta Ambient Air Quality Objectives for H2S are: 1-hour average of 14 μg m -3 (or 10 ppb) 24-hour average of 4 μg m -3 (or 3 ppb). The principle of operation accepted by the Alberta Air Monitoring Directive (AMD) for continuous ambient air H2S and TRS monitoring methods is Ultraviolet (UV) pulsed fluorescence: H2S and TRS analyzers are basically Sulfur dioxide (SO2) analyzers with a SO2 scrubber and a converter to convert either H2S or TRS to SO2 for analysis. Some analyzers are capable of cycling between SO2, H2S, and TRS measurements by switching between sample pathways with and without SO2 scrubbers and converters. After sulfur compounds are converted, the SO2 is measured proportional to UV absorption and fluorescence, which is related back to the H2S or TRS concentration for detection and reporting (USEPA 2017). 2. Current Method for Continuous Measurements The common method used for a continuous ambient H2S or TRS monitor is based on UV pulsed fluorescence which is also required by the Alberta AMD (Alberta Environment and Parks 2017). 2.1 Principle of the Method H2S/TRS is measured using thermal oxidation to convert H2S or TRS molecules to SO2 and the converted SO2 molecules are then analyzed using continuous fluorescent SO2 analyzers. H2S H2S molecules will be oxidized to SO2 in the presence of Oxygen and heat. This is accomplished by diverting the sample flow through the H2S converter, after a scrubber is used to remove hydrocarbons. The converter has a stainless-steel body heated to at least 320 °C. Prior to the converter, the sample flow passes through a scrubber to remove all SO2, and to allow only the H2S molecules to pass through and enter the H2S converter. The H2S molecules are then converted to SO2 as Equation 1.

Upload: others

Post on 01-Feb-2021

0 views

Category:

Documents


0 download

TRANSCRIPT

  • 1 | P a g e

    H2S and TRS Measurements – Continuous and Integrated Methods

    Prepared by Sanjay Prasad, Yu-Mei Hsu, Michael Martineau, Kendra Thomas

    Wood Buffalo Environmental Association

    April 25, 2018

    1. Background information

    Hydrogen sulfide (H2S) is a colourless gas which has the odour of rotten eggs. H2S is considered

    an odour nuisance at low levels, and can result in discomforting physiological symptoms of

    headache and nausea. Total Reduced Sulfur (TRS) is a collection of H2S and a number of other

    sulfur compounds (e.g Carbonyl Sulphide (COS), Carbon Disulphide (CS2) and Methyl

    Mercaptan (CH3SH)). The current Alberta Ambient Air Quality Objectives for H2S are:

    1-hour average of 14 μg m-3 (or 10 ppb)

    24-hour average of 4 μg m-3 (or 3 ppb).

    The principle of operation accepted by the Alberta Air Monitoring Directive (AMD) for

    continuous ambient air H2S and TRS monitoring methods is Ultraviolet (UV) pulsed

    fluorescence: H2S and TRS analyzers are basically Sulfur dioxide (SO2) analyzers with a SO2

    scrubber and a converter to convert either H2S or TRS to SO2 for analysis. Some analyzers are

    capable of cycling between SO2, H2S, and TRS measurements by switching between sample

    pathways with and without SO2 scrubbers and converters. After sulfur compounds are converted,

    the SO2 is measured proportional to UV absorption and fluorescence, which is related back to the

    H2S or TRS concentration for detection and reporting (USEPA 2017).

    2. Current Method for Continuous Measurements

    The common method used for a continuous ambient H2S or TRS monitor is based on UV pulsed

    fluorescence which is also required by the Alberta AMD (Alberta Environment and Parks 2017).

    2.1 Principle of the Method

    H2S/TRS is measured using thermal oxidation to convert H2S or TRS molecules to SO2 and the

    converted SO2 molecules are then analyzed using continuous fluorescent SO2 analyzers.

    H2S

    H2S molecules will be oxidized to SO2 in the presence of Oxygen and heat. This is accomplished

    by diverting the sample flow through the H2S converter, after a scrubber is used to remove

    hydrocarbons. The converter has a stainless-steel body heated to at least 320 °C. Prior to the

    converter, the sample flow passes through a scrubber to remove all SO2, and to allow only the

    H2S molecules to pass through and enter the H2S converter. The H2S molecules are then

    converted to SO2 as Equation 1.

  • 2 | P a g e

    H2S + O2 + heat → SO2 + H2O Equation 1

    The converted SO2 molecules then return to the standard SO2 analyzer for detection and are

    reported as H2S (Alberta Environment and Parks 2011a, Alberta Environment and Parks 2011b).

    TRS

    TRS compounds include a variety of airborne compounds which contain Sulfur. Some of the

    common compounds found in Alberta are: Carbonyl Sulphide (COS), Carbon Disulphide (CS2)

    and Methyl Mercaptan (CH3SH). Similar to H2S, TRS molecules will be oxidized to SO2 in the

    presence of Oxygen and heat. The difference between H2S and TRS is the TRS conversion

    requires a higher temperature. This is accomplished by diverting the sample flow through a TRS

    converter, after the flow has passed through the hydrocarbon scrubber. The TRS converter is a

    Quartz tube heated to a minimum of 800 °C. Prior to the converter, the sample flow must pass

    through a scrubber to remove all SO2, and to allow only the TRS molecules to pass through to

    enter the TRS converter. The TRS molecules are then converted to SO2 as Equation 2.

    TRS + O2 + heat → SO2 + H2O Equation 2

    The converted SO2 molecules then return to the standard SO2 analyzer for detection and are reported as

    TRS (Alberta Environment and Parks 2011a, Alberta Environment and Parks 2011b).

    The primary detection principle utilized in SO2 analyzers is UV fluorescence. Where a valence

    electron absorbs UV at a wavelength of 214 nm and enters an excited state (*). When the

    electron drops back into its usual state, a photon is emitted at a lower energy wavelength of 390

    nm as Equation 3.

    SO2 + hv1 → SO2* → SO2 + hv2 Equation 3

    * = Excited State

    hv1 = Exposure Light at 214 nm

    hv2 = Emitted Light at 390 nm

    The light emitted by the decaying SO2 electrons is filtered and channeled into a photo multiplier

    tube and amplified and converted to an electrical signal (Alberta Environment and Parks 2011c).

    2.2 Operations/Diagram

    The continuous H2S/TRS monitor measures H2S/TRS concentrations in ambient air by thermal

    conversion of H2S/TRS to SO2 with a molybdenum catalytic converter and UV fluorescence of

    the SO2 gas (Sumner et al. 2005). H2S measurements are made using a low temperature

    converter operating typically between 300 to 350 °C. For TRS measurements, the conversion of

    sulfur compounds to SO2 occurs at a much higher temperature typically between 800 and 1000

  • 3 | P a g e

    °C. To maximize converter performance, it is important to operate the analyzers with converters

    in the appropriate temperature ranges as specified by manufacturers (Alberta Environment and

    Parks 2017).

    Figure 1 illustrates a pneumatic diagram of Teledyne Instruments T101 H2S Analyzer. The air

    sample is drawn into the monitor through the sample gas inlet and then flows through a sample

    filter and a hydrocarbon scrubber (hydrocarbon kicker) to remove particles and hydrocarbons,

    respectively, from the air sample. A H2S/SO2 mode valve is used to control the air flow for SO2

    measurement or H2S measurement (Table 1).

    The air sample then flows into the sample chamber, where pulsating UV Light excites the SO2

    molecules. The condensing lens focuses the pulsating UV light into the mirror assembly. The

    mirror assembly contains four selective mirrors that reflect only the wavelengths which excite

    SO2 molecules. As the excited SO2 molecules decay to lower energy states they emit UV light

    that is proportional to the SO2 concentration. The bandpass filter allows only the wavelengths

    emitted by the excited SO2 molecules to reach the photomultiplier tube (PMT). The PMT detects

    the UV light emission from the decaying SO2 molecules (Thermo Scientific 2015).

    Figure 1. Pneumatic Diagram of Teledyne Instruments T101 Analyzer (Adapted from Teledyne

    Monitor Labs (2016))

  • 4 | P a g e

    Table 1. H2S – SO2 Switching Valve Operating States (Adapted from Teledyne Monitor Labs

    (2016))

    GAS MODE CONDITION OF H2S –SO2 SWITCHING

    VALVE

    VALVE PORT

    CONNECTION

    H2S Open to SO2 Scrubber and Molybdenum

    Converter COM → NO

    SO2 Open to directly to Sample Chamber. Bypasses

    SO2 Scrubber and Molybdenum Converter COM → NC

    H2S –SO2 Switches between above two states every 10

    minutes.

    -

    2.3 Interferences

    It should be noted the Ultraviolet (UV) pulsed fluorescence method for measuring H2S/TRS is

    subject to interference from a number of sources, including SO2, hydrocarbons (Luke 1997) and

    other species which absorb at about 190 - 230 nm and fluoresces at 340 - 410 nm. Obviously,

    since the analyzer measures H2S by converting it to SO2, the most significant interfering gas for

    this measurement would be ambient SO2 which is present in the sample gas. A SO2 chemical

    scrubber is applied to remove SO2 from the sample gas before the H2S → SO2 conversion takes

    place.

    The second most common source of interference is from other gases which fluoresce in a similar

    fashion to SO2 when exposed to UV Light. The most significant of these gases is a class of

    hydrocarbons called aromatic hydrocarbons (Luke 1997) of which xylene and naphthalene are

    two prominent examples. A hydrocarbon scrubber (kicker) mechanism removes aromatic

    hydrocarbons present in the sample gas before it the reaches the sample chamber.

    Because ozone absorbs UV Light over a relatively broad spectrum it could cause a measurement

    offset by absorbing some of the UV given off by the decaying SO2* (* = excited state) in the

    sample chamber. This can be prevented from occurring by having a very short light path between

    the area where the SO2* fluorescence occurs and the PMT detector (Teledyne Monitor Labs

    2016).

    While the decay of SO2* to SO2 happens quickly, it is not instantaneous. Since it is not

    instantaneous, it is possible for the extra energy possessed by the excited electron of the SO2*

    molecule to be given off as kinetic energy during a collision with another molecule. This

    process, in effect, heats the other molecule slightly and allows the excited electron to move into a

    lower energy orbit without emitting a photon. The most significant interferents in this regard are

    nitric oxide (NO), carbon dioxide (CO2), water vapor (H2O) and molecular oxygen (O2). In

    ambient applications the quenching effect of these gases is negligible (Teledyne Monitor Labs

    2016).

  • 5 | P a g e

    To monitor H2S/TRS concentrations in ambient air, the H2S/TRS is oxidized to SO2 by a

    Molybdenum converter. However, the converter unit does not oxidize all H2S/TRS to SO2, rather

    it uses a conversion efficiency determined by the manufacturer. The conversion efficiency varies

    from instrument to instrument, and is typically above 80% (Bluhme et al. 2016).

    Sumner et al. (2005) have assessed the effect of potential interferant gases (Table 2) on the

    response of the SO2 analyzer ( Teledyne Model 101E) with seven (7) gases in zero air and a 100-

    ppb H2S standard. No interference effect was observed in the Model 101E response to SO2, a

    blend of C1 to C6 alkanes, and ammonia. The Model 101E showed an interference effect for

    carbonyl sulfide in zero air of 20% and in 100-ppb H2S of 6%. Carbon disulfide resulted in an

    interference effect of 6% in zero air and 9% in the 100-ppb H2S matrix. The interference effect

    of methyl mercaptan on the Model 101E was 33% in both zero air and 100-ppb H2S. Dimethyl

    sulfide resulted in a 12% interference effect in both matrices.

    Table 2. Interference Effect Evaluation (adapted from Sumner et al. (2005))

    Interferant Approximate

    concentration (ppb)

    Interference Effect (%)

    Zero Air Matrix 100-ppb H2S Matrix

    Sulfur dioxide 100 0 0

    Carbonyl sulfide 100 20 6

    Carbon disulfide 100 6 9

    Methyl mercaptan 100 33 33

    Dimethyl sulfide 100 12 12

    Hydrocarbon blend 500 (total) 0 0

    Ammonia 500 0 0

    3. Designations – US, Australia, New Zealand, Europe

    H2S or TRS are not criteria pollutants in the United States of America, therefore the United

    States Environmental Protection Agency (USEPA) does not have a Federal Reference Method

    (FRM) for H2S or TRS (USEPA 2017).

    4. AMD and Manufacturer Performance Specifications

    In Alberta, as required by the Alberta AMD, Chapter 4: Monitoring Requirements and

    Equipment Technical Specifications, the operating principle for measurement of H2S and TRS is

    UV pulsed fluorescence. The minimum performance specifications required by the AMD are

    listed in Table 3. The manufacturer performance specifications of two analyzers including

    Teledyne T101 and Thermo Scientific Model 450i are listed in Table 4 and Table 5, respectively.

  • 6 | P a g e

    Table 3. Alberta AMD performance specifications for H2S and TRS continuous analyzers

    (adapted from Alberta Environment and Parks (2017))

    Criteria Performance Specification

    Required Operating Range* (Full Scale): 0.1 ppm, 0.5 ppm or 1.0 ppm

    Zero Noise: 0.5 ppb RMS

    Lower Detection Limit: 1.0 ppb

    Zero Drift (24-hr): 1.0 ppb

    Span Drift (24-hr) 1% of full scale

    Linearity: 1% of full scale

    Precision: 1.0 ppb or 1% of reading

    Rise time: Maximum of 120 seconds

    Fall time: Maximum of 120 seconds * The typical operating range is 0.1 ppm with 0.5 ppm and 1.0 ppm used in unique circumstances such as

    emergency monitoring. RMS is the root mean square of differences.

  • 7 | P a g e

    Table 4. Teledyne Model T101 Basic Unit Specifications (adapted from Teledyne Monitor Labs

    (2016))

    Parameter Description

    Ranges

    H2S: Min 0-50 ppb Full scale; Max 0-10 ppm Full scale

    SO2: Up to 0-20 ppm Full scale

    (selectable, independent ranges and auto ranging supported)

    Measurement units ppb, ppm, µg/m3, mg/m3 (selectable)

    Zero Noise

  • 8 | P a g e

    Table 5. Thermo Scientific Model 450i Specifications (adapted from Thermo Scientific (2018))

    Parameter Description

    Preset Ranges

    0-0.05, 0.1, 0.2, 0.5, 1, 2, 5, and 10 ppm,

    0-0.2, 0.5, 1, 2, 5, 10, 20, and 25 mg/m3

    Extended Ranges

    0-0.5, 1, 2, 5, 10, 20, 50, and 100 ppm,

    0-2, 5, 10, 20, 50, 100, 200, and 250 mg/m3

    Custom Ranges 0-0.05 to 100 ppm, 0-0.2 to 250 mg/m3

    Zero Noise

    Manual SO2 or

    Combine Sulfur

    Automatic mode

    SO2 or H2S

    1.0 ppb 3.0 ppb (10 second averaging time)

    0.5 ppb 1.5 ppb (60 second averaging time)

    0.25 ppb 0.75 ppb (300 second averaging time)

    Lower Detectable

    Limit

    Manual SO2 or

    Combine Sulfur

    Automatic mode

    SO2 or H2S

    2.0 ppb 6.0 ppb (10 second averaging time)

    1.0 ppb 2.0 ppb (60 second averaging time)

    0.5 ppb 1.5 ppb (300 second averaging time)

    Zero Drift (24 hour) Less than 1 ppb

    Span drift (24 hour) ± 1% Full Scale

    Response Time 80 seconds (10 second average time)

    110 seconds (60 second average time)

    320 seconds (300 seconds average time)

    Precision 1% of reading or 1 ppb (whichever is greater)

    Linearity ±1% full scale 80% H2S to SO2 (Note: Various other Sulfur compounds can be converted

    at varying %)

    Operating

    Temperature 20⁰C to 30⁰C

    Power Requirements 100 VAC, 115 VAC, 220-240 VAC ±10% @ 300W

    Size and Weight 16.75"(W) x 8.62"(H) x 23"(D), 48 lbs. (21.8 kg)

    Outputs

    Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power

    Fail Indication (standard). 0-20 or 4-20 mA Isolated Current Output

    (optional)

    Inputs 16 Digital Inputs (standard), 8 0-10 Vdc Analog Inputs (optional)

    Available Options Teflon particulate filter, Rack mounts, Rear extenders

  • 9 | P a g e

    5. Other Methods for H2S and RSC Measurements

    Both passive and active measurements have been developed for H2S and Reduced Sulfur

    Compounds (RSCs) measurements. A general outline of the analytical protocols commonly

    employed in RSC analysis is illustrated in Figure 2.

    Figure 2. General protocols of Reduced Sulfur Compounds (RSCs) analysis in air matrices. The

    abbreviations used in figure are as follows: LC-AFS = liquid chromatography-atomic

    fluorescence spectrometer, IC = ion-chromatography, SPME = solid phase microextraction, FPD

    = flame photometric detector, PFPD = pulsed flame photometric detector, MS = mass

    spectrometer, AED = atomic emission detector, SCD = sulfur chemiluminescence detector, and

    TD = thermal desorber. (adapted from Pandey and Kim (2009).)

    5.1 H2S Passive Measurement

    Many H2S passive samplers are made commercially available (Shooter et al. 1995, Tang et al.

    2002, Venturi et al. 2016) for the remote locations with no power accesses. The H2S passive

    collection media are coated with chemicals (e.g., Silver nitrate, zinc acetate) to collect H2S for

    sample analysis (Tang et al. 2002, Pavilonis et al. 2013).

    Sampling stage Preconcentration stageDetection stage

    (Detector type)

    LC-AFS

    IC

    Conductomery

    Potentiometry

    Coulometry

    Voltametry

    FPD

    PFPD

    MS

    AED

    SCD

    Bag or Canister

    Non-GC methods

    GC method

    Alkaline aqueous Phase

    Diffusion scrubbers

    SPME

    Tube

    TD

    Cryogenic trappingSolid sorbentsSorption on metal

    surfaces

  • 10 | P a g e

    5.2 Gas Chromatograph (GC) Methods

    For RSC analysis in the air, the gas chromatography (GC) technique has been and is the most

    common methodology because of its excellent separation capability and quantitative recovery

    (Helmig 1999, Willig et al. 2004). The GC-based analysis commonly involves a sample

    collection stage, injection, and separation on a chromatographic column. The final detection can

    then be carried out through various Sulfur-selective (or universal) detectors.

    Direct chromatographic analysis or direct injection (DI) of RSCs into a GC injector is highly

    recommended when the concentration of samples is in the detectable range of a given GC-setup.

    The use of the DI method can reduce possible loss (or gain) due to contact with different surface

    types, and reduces analysis time by eliminating time-consuming procedures such as

    supplementary sample treatments by which contamination or loss of analytes can occur (Kim

    2006, Pandey and Kim 2008, Pandey and Kim 2009). However, the application of the DI

    approach is often limited, as most detectors of the GC method are not sensitive enough to cover

    ambient samples which are typically below a few ppb in concentration. Therefore, research has

    sought to improve and develop GC methods based on Sulfur-specific detectors, including

    electron capture detector (ECD), flame photometric detector (FPD), pulsed flame photometric

    detector (PFPD), sulfur chemiluminescence detector (SCD), atomic emission detection (AED),

    Hall electrical conductivity detector (HECD), and photo ionization detector (PID) with combined

    application of a mass spectrometer (MS) and GC system (Nielsen and Jonsson 2002, Dincer et

    al. 2006, Oostdijk et al. 2007, Ras et al. 2008). USEPA CFR Promulgated Test Method 15

    (USEPA 2017) has employed an online GC-FPD method for monitoring hydrogen sulfide,

    carbonyl sulfide, and carbon disulfide emissions from stationary sources

    To allow the analysis of increasingly smaller quantities of target compounds, the range of

    instrumental detectability needs to be improved. As a means to extend the detectability of a given

    GC system, one can increase the total amount of analytes injected by adopting some

    preconcentration (or sample enrichment) stages: (1) sorption on certain metal surfaces (Braman

    et al. 1978, Barnard et al. 1982, Ferek et al. 1986), (2) sorption on solid adsorbents (Maier and

    Fieber 1988, Sunesson et al. 1995, Inomata et al. 1999), and (3) cryogenic trapping (Kim 2005,

    Kim 2005, Ras et al. 2008, Pandey and Kim 2009). However, the analysis of low-level sulfur

    species through such modifications can be subject to positive blanks (e.g., memory effect) or

    sorptive loss in the chromatographic system (Sulyok et al. 2001, Sulyok et al. 2002).

    Sampling and Preconcentration Strategies. RSCs from air can be collected in vessels such as

    glass bulbs, canister bags, polymer bags, and Tedlar film bags. Considering the highly reactive

    nature of Sulfur compounds, the sampling vessels should be inert enough to reduce adsorptive

    loss (Sulyok et al. 2001). Moreover, careful attention should be given to tubing and connecting

    materials used for the sampling of Sulfur compounds. The reason for this is certain materials

    can act as significant sources of bias in the determination of RSC concentrations (Sulyok et al.

    2001, Sulyok et al. 2002, Kim et al. 2006, Winkel and Tangerman 2008).

    ASTM Method D5504-01(ASTM International 2001) use

    canister samples for H2S collection

    and require samples to be analyzed by GC-PFPD or GC-SCD within 24 hours. ASTM Method

  • 11 | P a g e

    D5504-01 has been demonstrated that H2S in passivated canisters does not degrade over 24

    hours; therefore, detailed holding time tests are not necessary. However, the acceptability of

    this holding time should be verified by analyzing an ambient air sample several times over the

    period (at least 24 hours) following sample collection. The H2S concentration at 24 hours

    following sample collection should be within 15% of the initial measured value (ASTM

    International 2001, Battelle 2005).

    A major difficulty in the sampling of RSCs is interference caused by atmospheric oxidants such

    as SO2, ozone, and NOx. To help overcome these difficulties, many substances were developed

    as scrubbers, including PTFE, Tygon, glass fiber filters, chromosorb, anakrom, and glass beads

    with a coating of chemicals (ex., Na2CO3, MnO2, KOH, and NaOH) (Braman et al. 1978,

    Andreae et al. 1985, Saltzman and Cooper 1988, Bates et al. 1990, Ayers et al. 1991, Kittler et

    al. 1992, Watanabe et al. 1995).

    Another considerable difficulty in determining RSCs is relative humidity. Water significantly

    lowers the capacity of adsorbents and can clog cryogenic traps (Haberhauer-Troyer et al. 1999).

    Moreover, it can cause baseline perturbations and retention time-shifts in chromatography to

    deteriorate detection (Haberhauer-Troyer et al. 1999). To overcome the humidity problem, a

    number of approaches have been developed by applying dryers (Juhani and Hannu 1988,

    Hofmann et al. 1992, Haberhauer-Troyer et al. 1999).

    Sorption on Metal Surfaces. The extent of RSC sorption can be affected greatly by the type of

    metals (mainly gold, palladium, and platinum) used for such a reaction (Braman et al. 1978,

    Barnard et al. 1982, Ferek et al. 1986, Kagel and Farwell 1986). These metallic materials are

    also found in modified forms such as glass or quartz tubes filled with gold wool, gold plated

    sand, or metal foils (Braman et al. 1978, Ferek et al. 1986, Davison and Allen 1994, Swan and

    Ivey 1994, Davison et al. 1996, Curran et al. 1998).

    Sorption on Solid Sorbents. Solid sorbent surfaces are regarded as the most general tools for

    the preconcentration of volatile species, e.g., activated charcoal, silica gel, aluminum oxide,

    graphitized carbon black, molecular sieves, and porous sorbents (Davison et al. 1996, Wylie

    and Mora 1996, Lewis et al. 1997, De Bruyn et al. 1998). Porous sorbents, such as Tenax have

    been the most popular choice for Sulfur species (Pio et al. 1996, De Bruyn et al. 1998, Ras et al.

    2008), but the breakthrough volume of Tenax is not altered by changes in humidity because of

    its hydrophobic property (Steinhanses and Schoene 1990). The trapping efficiency of Tenax

    tubes is often limited for organosulfur compounds with low boiling points (Tangerman 1986,

    Shooter et al. 1992). The capacity of a sorbent can increase significantly by adding a

    cryofocusing system (Kroupa et al. 2004). Many studies have been carried out to determine the

    affinity and/or breakthrough volume of different sorbents (Przyjazny 1985, Devai and Delaune

    1996, Devai and DeLaune 1997) for the adsorption capacity. The application of various solid

    sorbents (i.e., silica gel, molecular sieve, and Carbosieve SIII) has been validated by

    measurements of RSC concentrations or fluxes under various environmental settings (Delaune

    et al. 2002).

  • 12 | P a g e

    Cryogenic Trapping. Trapping efficiency of adsorbent traps increases exponentially with a

    decrease in temperature (Pollmann et al. 2006). The cryogenic trapping technique can increase

    the capacity of sorptive materials effectively. The cryogenic traps (i.e., sampling loops) are

    generally composed of PTFE, borosilicate glass, and quartz tubes (Wardencki 1998). For

    sampling of RSCs, the cryogenic traps are normally immersed in a strong cryogen such as

    liquid nitrogen or in liquid argon (Simo et al. 1993). Cryogenic traps can also be used with a

    number of sorbents as packing materials to increase sorption efficiency. Commonly used

    packing materials include glass-fiber wool, glass beads, Tenax, Porapack Q, activated carbon,

    and carbopack. Based on the technique of cryogenic trapping, and GC-PHD, an automated

    Sulfur instrument was developed to analyze RSCs continuously (Von Hobe et al. 2008).

    Solid-Phase Microextraction (SPME). The solid-phase microextraction (SPME) method, a

    potential solvent-free sample preparation technique, allows a single-step treatment for

    sampling, isolation, and enrichment (Xiong et al. 2003, Ouyang and Pawliszyn 2006, Ouyang

    and Pawliszyn 2006, Demeestere et al. 2007). For air matrices, the SPME fiber can be used to

    extract analytes either by direct exposure to raw samples or by the use of the headspace method

    on pretreated samples (Arthur and Pawliszyn 1990, Li et al. 2001, Ouyang and Pawliszyn

    2006).

    In addition to GC methods, there are alternative methods which include liquid chromatograph

    (LC), ion chromatograph (IC), and sensor-based methods. The reliability of these alternative

    methods for practical application of real environmental samples has not been sufficiently

    researched. Methodologies for online monitoring at ambient concentration levels would help

    overcome the time-consuming and tedious processes of sampling, enrichment, and offline

    analytical protocols in the laboratory (Pandey and Kim 2009).

  • 13 | P a g e

    6. References

    Alberta Environment and Parks (2011a). Standard Operating Procedure for Hydrogen Sulphide

    (H2S) and Total Reduced Sulphur (TRS) detectors. Alberta, Government of Alberta.

    http://environmentalmonitoring.alberta.ca/wp-content/uploads/2015/09/SOP-010.pdf.

    Alberta Environment and Parks (2011b). Standard Operating Procedure for Hydrogen Sulphide

    (H2S) and Total Reduced Sulphur (TRS) converter. Alberta, Government of Alberta.

    http://environmentalmonitoring.alberta.ca/wp-content/uploads/2015/09/SOP-001.pdf.

    Alberta Environment and Parks (2011c). Standard Operating Procedure for Measurement of SO2

    by SO2 fluorescence. Alberta, Government of Alberta.

    http://environmentalmonitoring.alberta.ca/wp-content/uploads/2015/09/SOP-021.pdf.

    Alberta Environment and Parks (2017). Continuous Analyzers Hydrogen Sulphide (H2S) and

    Total Reduced Sulphur (TRS). Alberta, Government of Alberta.

    http://aep.alberta.ca/air/legislation-and-policy/air-monitoring-

    directive/documents/HydrogenSulphide-TRS-Jun26-2017.pdf.

    Andreae, M. O., et al. (1985). "Dimethyl sulfide in the marine atmosphere." Journal of

    Geophysical Research: Atmospheres 90(D7): 12891-12900.

    Arthur, C. L. and J. Pawliszyn (1990). "Solid phase microextraction with thermal desorption

    using fused silica optical fibers." Analytical Chemistry 62(19): 2145-2148.

    ASTM International (2001). ASTM D5504-01: Standard Test Method for Determination of

    Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatography and

    Chemiluminescence. West Conshohocken, PA, ASTM International. www.astm.org.

    Ayers, G. P., et al. (1991). "Coherence between seasonal cycles of dimethyl sulphide,

    methanesulphonate and sulphate in marine air." Nature 349: 404.

    Barnard, W. R., et al. (1982). "The flux of dimethylsulfide from the oceans to the atmosphere."

    Journal of Geophysical Research: Oceans 87(C11): 8787-8793.

    Bates, T. S., et al. (1990). "The biogeochemical sulfur cycle in the marine boundary layer over

    the Northeast Pacific Ocean." Journal of Atmospheric Chemistry 10(1): 59-81.

    Battelle (2005). Generic Verification Protocol for Verification of Ambient Hydrogen Sulfide

    Analyzers at an Animal Feeding Operation. Columbus, OH, Battelle.

    https://archive.epa.gov/research/nrmrl/archive-etv/web/pdf/01_vp_sulfide.pdf.

    Bluhme, A. B., et al. (2016). "Water vapor inhibits hydrogen sulfide detection in pulsed

    fluorescence sulfur monitors." Atmos. Meas. Tech. 9(6): 2669-2673.

    Braman, R. S., et al. (1978). "Preconcentration and determination of hydrogen sulfide in air by

    flame photometric detection." Analytical Chemistry 50(7): 992-996.

    Curran, M. A. J., et al. (1998). "Spatial distribution of dimethylsulfide and

    dimethylsulfoniopropionate in the Australasian sector of the Southern Ocean." Journal

    of Geophysical Research: Atmospheres 103(D13): 16677-16689.

    Davison, B., et al. (1996). "Dimethyl sulfide and its oxidation products in the atmosphere of the

    Atlantic and Southern Oceans." Atmospheric Environment 30(10): 1895-1906.

    Davison, B. M. and A. G. Allen (1994). "A method for sampling dimethylsulfide in polluted and

    remote marine atmospheres." Atmospheric Environment 28(10): 1721-1729.

    http://environmentalmonitoring.alberta.ca/wp-content/uploads/2015/09/SOP-010.pdfhttp://environmentalmonitoring.alberta.ca/wp-content/uploads/2015/09/SOP-001.pdfhttp://environmentalmonitoring.alberta.ca/wp-content/uploads/2015/09/SOP-021.pdfhttp://aep.alberta.ca/air/legislation-and-policy/air-monitoring-directive/documents/HydrogenSulphide-TRS-Jun26-2017.pdfhttp://aep.alberta.ca/air/legislation-and-policy/air-monitoring-directive/documents/HydrogenSulphide-TRS-Jun26-2017.pdfhttp://www.astm.org/https://archive.epa.gov/research/nrmrl/archive-etv/web/pdf/01_vp_sulfide.pdf

  • 14 | P a g e

    De Bruyn, W. J., et al. (1998). "Shipboard measurements of dimethyl sulfide and SO2 southwest

    of Tasmania during the First Aerosol Characterization Experiment (ACE 1)." Journal of

    Geophysical Research-Atmospheres 103(D13): 16703-16711.

    Delaune, R. D., et al. (2002). "Flux of Reduced Sulfur Gases Along a Salinity Gradient in

    Louisiana Coastal Marshes." Estuarine, Coastal and Shelf Science 54(6): 1003-1011.

    Demeestere, K., et al. (2007). "Sample preparation for the analysis of volatile organic

    compounds in air and water matrices." Journal of Chromatography A 1153(1): 130-144.

    Devai, I. and R. D. Delaune (1996). "Evaluation of various solid adsorbents for sampling trace

    levels of methanethiol." Organic Geochemistry 24(8): 941-944.

    Devai, I. and R. D. DeLaune (1997). "Trapping efficiency of various solid adsorbents for

    sampling and quantitative gas chromatographic analysis of carbonyl sulfide." Analytical

    Letters 30(1): 187-198.

    Dincer, F., et al. (2006). "Chemical characterization of odorous gases at a landfill site by gas

    chromatography–mass spectrometry." Journal of Chromatography A 1122(1): 222-229.

    Ferek, R. J., et al. (1986). "Vertical distribution of dimethylsulphide in the marine atmosphere."

    Nature 320: 514.

    Haberhauer-Troyer, C., et al. (1999). "Investigation of membrane dryers and evaluation of a new

    ozone scrubbing material for the sampling of organosulphur compounds in air." Journal

    of Chromatography A 852(2): 589-595.

    Helmig, D. (1999). "Air analysis by gas chromatography." Journal of Chromatography A 843(1):

    129-146.

    Hofmann, U., et al. (1992). "Cryogenic trapping of reduced sulfur compounds using a nafion

    drier and cotton wadding as an oxidant scavenger." Atmospheric Environment. Part A.

    General Topics 26(13): 2445-2449.

    Inomata, Y., et al. (1999). "Simultaneous measurement of volatile sulfur compounds using

    ascorbic acid for oxidant removal and gas chromatography–flame photometric

    detection." Journal of Chromatography A 864(1): 111-119.

    Juhani, K. and R. Hannu (1988). "The analysis of reduced sulphur gases in ambient air of

    workplaces." Chemosphere 17(5): 905-914.

    Kagel, R. A. and S. O. Farwell (1986). "Evaluation of metallic foils for preconcentration of

    sulfur-containing gases with subsequent flash desorption/flame photometric detection."

    Analytical Chemistry 58(6): 1197-1202.

    Kim, K.-H. (2005). "Performance characterization of the GC/PFPD for H2S, CH3SH, DMS, and

    DMDS in air." Atmospheric Environment 39(12): 2235-2242.

    Kim, K.-H. (2005). "Some Insights into the Gas Chromatographic Determination of Reduced

    Sulfur Compounds (RSCs) in Air." Environmental Science & Technology 39(17): 6765-

    6769.

    Kim, K.-H. (2006). "The properties of calibration errors in the analysis of reduced sulfur

    compounds by the combination of a loop injection system and gas chromatography with

    pulsed flame photometric detection." Analytica Chimica Acta 566(1): 75-80.

    Kim, K.-H., et al. (2006). "The loss patterns of reduced sulfur compounds in contact with

    different tubing materials." Journal of Chromatography A 1132(1): 228-233.

  • 15 | P a g e

    Kittler, P., et al. (1992). "An indicating oxidant scrubber for the measurement of atmospheric

    dimethylsulphide." Atmospheric Environment. Part A. General Topics 26(14): 2661-

    2664.

    Kroupa, A. n., et al. (2004). "Breakthrough characteristics of volatile organic compounds in the

    −10 to +170°C temperature range on Tenax TA determined by microtrap technology."

    Journal of Chromatography A 1038(1): 215-223.

    Lewis, A. C., et al. (1997). "High-Speed Isothermal Analysis of Atmospheric Isoprene and DMS

    Using On-Line Two-Dimensional Gas Chromatography." Environmental Science &

    Technology 31(11): 3209-3217.

    Li, K., et al. (2001). "Solvent vapour monitoring in work space by solid phase micro extraction."

    Journal of Hazardous Materials 83(1): 83-91.

    Luke, W. T. (1997). "Evaluation of a commercial pulsed fluorescence detector for the

    measurement of low‐level SO2 concentrations during the Gas‐Phase Sulfur

    Intercomparison Experiment." Journal of Geophysical Research: Atmospheres

    102(D13): 16255-16265.

    Maier, I. and M. Fieber (1988). "Retention characteristics of volatile compounds on tenax TA."

    Journal of High Resolution Chromatography 11(8): 566-576.

    Nielsen, A. T. and S. Jonsson (2002). "Trace determination of volatile sulfur compounds by

    solid-phase microextraction and GC-MS." Analyst 127(8): 1045-1049.

    Oostdijk, J. P., et al. (2007). "Selective and sensitive trace analysis of sulfur mustard with

    thermal desorption and two-dimensional gas chromatography–mass spectrometry."

    Journal of Chromatography A 1150(1): 62-69.

    Ouyang, G. and J. Pawliszyn (2006). "Recent developments in SPME for on-site analysis and

    monitoring." TrAC Trends in Analytical Chemistry 25(7): 692-703.

    Ouyang, G. F. and J. Pawliszyn (2006). "SPME in environmental analysis." Analytical and

    Bioanalytical Chemistry 386(4): 1059-1073.

    Pandey, S. K. and K.-H. Kim (2008). "The fundamental properties of the direct injection method

    in the analysis of gaseous reduced sulfur by gas chromatography with a pulsed flame

    photometric detector." Analytica Chimica Acta 615(2): 165-173.

    Pandey, S. K. and K.-H. Kim (2009). "Comparison of different calibration approaches in the

    application of thermal desorption technique: A test on gaseous reduced sulfur

    compounds." Microchemical Journal 91(1): 40-46.

    Pandey, S. K. and K.-H. Kim (2009). "A Review of Methods for the Determination of Reduced

    Sulfur Compounds (RSCs) in Air." Environmental Science & Technology 43(9): 3020-

    3029.

    Pavilonis, B. T., et al. (2013). "Passive monitors to measure hydrogen sulfide near concentrated

    animal feeding operations." Environmental science. Processes & impacts 15(6): 1271-

    1278.

    Pio, C. A., et al. (1996). "Sulphur and nitrogen compounds in variable marine/continental air

    masses at the southwest European coast." Atmospheric Environment 30(18): 3115-

    3127.

  • 16 | P a g e

    Pollmann, J., et al. (2006). "Evaluation of solid adsorbent materials for cryogen-free trapping—

    gas chromatographic analysis of atmospheric C2–C6 non-methane hydrocarbons."

    Journal of Chromatography A 1134(1): 1-15.

    Przyjazny, A. (1985). "Preconcentration of volatile organosulphur compounds from the

    atmosphere on selected porous polymers." Journal of Chromatography A 333: 327-336.

    Ras, M. R., et al. (2008). "Determination of volatile organic sulfur compounds in the air at

    sewage management areas by thermal desorption and gas chromatography–mass

    spectrometry." Talanta 74(4): 562-569.

    Saltzman, E. S. and D. J. Cooper (1988). "Shipboard measurements of atmospheric

    dimethylsulfide and hydrogen sulfide in the Caribbean and Gulf of Mexico." Journal of

    Atmospheric Chemistry 7(2): 191-209.

    Shooter, D., et al. (1992). "The Chromatographic Analysis of Reduced Sulfur Gases in Antarctic

    Waters Following Preconcentration onto Tenax." International Journal of

    Environmental Analytical Chemistry 47(4): 239-249.

    Shooter, D., et al. (1995). "A passive sampler for hydrogen sulfide." Environmental Monitoring

    and Assessment 38(1): 11-23.

    Simo, R., et al. (1993). "Field sampling and analysis of volatile reduced sulphur compounds in

    air, water and wet sediments by cryogenic trapping and gas chromatography." Journal of

    Chromatography A 655(2): 301-307.

    Steinhanses, J. and K. Schoene (1990). "Thermal desorption—gas chromatography of some

    organophosphates and S-mustard after trapping on Tenax." Journal of Chromatography

    A 514: 273-278.

    Sulyok, M., et al. (2002). "Observation of sorptive losses of volatile sulfur compounds during

    natural gas sampling." Journal of Chromatography A 946(1): 301-305.

    Sulyok, M., et al. (2001). "Investigation of the storage stability of selected volatile sulfur

    compounds in different sampling containers." Journal of Chromatography A 917(1):

    367-374.

    Sumner, A. L., et al. (2005). Environmental Technology Verification Report: Teledyne-API

    Model 101E Ambient Hydrogen Sulfide Analyzer. Columbus, OH, Battelle.

    https://archive.epa.gov/nrmrl/archive-etv/web/pdf/01_vr_teledyne.pdf.

    Sunesson, A.-L., et al. (1995). "Evaluation of adsorbents for sampling and quantitative analysis

    of microbial volatiles using thermal desorption-gas chromatography." Journal of

    Chromatography A 699(1): 203-214.

    Swan, H. B. and J. P. Ivey (1994). "Analysis of atmospheric sulfur gases by capillary gas

    chromatography with atomic emission detection." Journal of High Resolution

    Chromatography 17(12): 814-820.

    Tang, H., et al. (2002). "A New All-Season Passive Sampling System for Monitoring H2S in

    Air." TheScientificWorldJOURNAL 2.

    Tangerman, A. (1986). "Determination of volatile sulphur compounds in air at the parts per

    trillion level by tenax trapping and gas chromatography." Journal of Chromatography A

    366: 205-216.

    https://archive.epa.gov/nrmrl/archive-etv/web/pdf/01_vr_teledyne.pdf

  • 17 | P a g e

    Teledyne Monitor Labs (2016). Operation Manual Model T101 UV Fluorescence Hydrogen

    Sulfide Analyzer. Englewood, CO, Teledyne Monitor Labs. http://www.teledyne-

    ml.com/pdfs/T101_H2S_RevB.pdf.

    Thermo Scientific (2015). Thermo Scientific Model 450i Instruction Manual Pulsed

    Fluorescence SO2-H2S-CS analyzer. Franklin, MA, Thermo Scientific.

    https://www.thermofisher.com/order/catalog/product/450I.

    Thermo Scientific (2018). "Thermo Scientific Model 450i Hydrogen Sulfide & Sulfur Dioxide

    Analyzer Pulsed Fluorescence gas analyzer." Retrieved April 18, 2018,

    https://assets.thermofisher.com/TFS-Assets/LSG/Specification-Sheets/EPM-450i-

    Datasheet.pdf.

    USEPA (2017). LIST OF DESIGNATED REFERENCE AND EQUIVALENT METHODS.

    Research Triangle Park, NC, UNITED STATES ENVIRONMENTAL PROTECTION

    AGENCY. https://www3.epa.gov/ttn/amtic/criteria.html.

    USEPA (2017). METHOD 15 -DETERMINATION OF HYDROGEN SULFIDE, CARBONYL

    SULFIDE, AND CARBON DISULFIDE EMISSIONS FROM STATIONARY

    SOURCES. Research Triangle Park, NC, UNITED STATES ENVIRONMENTAL

    PROTECTION AGENCY. https://www.epa.gov/emc/method-15-hydrogen-sulfide-

    carbonyl-sulfide-and-carbon-disulfide.

    Venturi, S., et al. (2016). "Hydrogen sulfide measurements in air by passive/diffusive samplers

    and high-frequency analyzer: A critical comparison." Applied Geochemistry 72: 51-58.

    Von Hobe, M., et al. (2008). "Automated in situ analysis of volatile sulfur gases using a Sulfur

    Gas Analyser (SUGAR) based on cryogenic trapping and gas-chromatographic

    separation." International Journal of Environmental Analytical Chemistry 88(5): 303-

    315.

    Wardencki, W. (1998). "Problems with the determination of environmental sulphur compounds

    by gas chromatography." Journal of Chromatography A 793(1): 1-19.

    Watanabe, S., et al. (1995). "Relation between the concentrations of DMS in surface seawater

    and air in the temperate North Pacific region." Journal of Atmospheric Chemistry 22(3):

    271-283.

    Willig, S., et al. (2004). "Development of a rapid and accurate method for the determination of

    key compounds of pig odor." Journal of Chromatography A 1038(1): 11-18.

    Winkel, E. G. and A. Tangerman (2008). "Appropriate sample bags and syringes for preserving

    breath samples in breath odor research: a technical note." Journal of Breath Research

    2(1): 017011.

    Wylie, D. J. and S. J. Mora (1996). "Atmospheric dimethylsulfide and sulfur species in aerosol

    and rainwater at a coastal site in New Zealand." Journal of Geophysical Research:

    Atmospheres 101(D15): 21041-21049.

    Xiong, G., et al. (2003). "On-site calibration method based on stepwise solid-phase

    microextraction." Journal of Chromatography A 999(1): 43-50.

    http://www.teledyne-ml.com/pdfs/T101_H2S_RevB.pdfhttp://www.teledyne-ml.com/pdfs/T101_H2S_RevB.pdfhttps://www.thermofisher.com/order/catalog/product/450Ihttps://assets.thermofisher.com/TFS-Assets/LSG/Specification-Sheets/EPM-450i-Datasheet.pdfhttps://assets.thermofisher.com/TFS-Assets/LSG/Specification-Sheets/EPM-450i-Datasheet.pdfhttps://www3.epa.gov/ttn/amtic/criteria.htmlhttps://www.epa.gov/emc/method-15-hydrogen-sulfide-carbonyl-sulfide-and-carbon-disulfidehttps://www.epa.gov/emc/method-15-hydrogen-sulfide-carbonyl-sulfide-and-carbon-disulfide