    Basic Principles of

    CT Scanning

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    2/39

    The basics of CT

    CT imaging chain

    System components

    Acquisition methods Image quality

    Applications

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    3/39

    X-ray: The beginning

    X-Rays founded in 1895 by

    Wilhelm Conrad Roentgen

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    4/39

    CT: The beginning

    CT founded in 1970 by Sir Godfrey Hounsfield

    Engineer with EMI, LTD.

    first applications were in neuroradiology

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    5/39

    CT Scanner

    X-Ray modality used to the body

    in cross section

    Used to determine extent of trauma

    location and type of tumors

    status of blood vessels

    pre surgical planning

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    6/39

    CT System

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    7/39

    Basic CT scanner components

    Gantry

    X-Ray Tube

    Detector Control Console

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    8/39

    Gantry

    CT X-ray tube

    High voltage generator

    Detector array Data acquistion system

    Slip ring

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    9/39

    The CT X-ray Tube

    Anode heat capacity

    3.5 MHU up to 28 MHU

    Determines maximum mAs

    Determines volume length

    Dictates generator size

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    10/39

    Detector Elements

    Capture energy that has not been

    attenuated by the patient

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    11/39

    Control console

    Set scan parameters kVp, mA, scan time, reconstruction

    filter, etc.

    Set scan mode Surview, Axial or Spiral

    IRS (Image reconstructionSystem)

    Review and archive images

    Post-processing

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    12/39

    CT

    CT - Computed Tomography

    CAT Scan - Computerized Axial

    Tomography

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    13/39

    Scanning methods

    Surview AP,Lat

    Surview, Scanogram , Topogram.

    Conventional CT

    Axial Start/stop

    Volumetric CT Helical or spiral CT

    Continuous acquisition

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    14/39

    Digital Projection

    X-ray tube and detector remain stationary

    Patient table moves continuously

    With X-rays on

    Produces an image covering a range of

    anatomy Similar to a conventional X-ray image, e.g. flat

    plate of the abdomen

    Image used to determine scan location

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    15/39

    Axial CT

    X-ray tube and detector rotate360

    Patient table is stationary

    With X-rays on

    Produces one cross-sectionalimage

    Once this is complete patient ismoved to next position

    Process starts again at thebeginning

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    16/39

    Volume CT

    X-ray tube and detector rotate 360

    Patient table moves continuously

    With X-rays on

    Produces a helix of image information This is reconstructed into 30 to 1000 images

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    17/39

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    18/39

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    19/39

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    20/39

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    21/39

    Single Slice = One 10mm slice per rotationDual Slice = Two 5mm slices per rotation

    Quad Slice = Four 2.5mm slices per rotation

    Multislice Effectiveness

    Everything is better

    Resolution 2x 4x-8x

    Speed same same

    Volume same same

    Power same same

    Dual Quad

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    22/39

    . .

    single detector arc dual detector arc

    pre-patient collimation

    post-patient collimation

    x-ray tube focal spot

    _ __

    Mx8000

    Dual Slice

    Dual Slice Detector Optimized for2 Slice Acquisition

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    23/39

    Approximately 10%

    more efficient than

    matrix detectors

    Variable Wide Area Detector

    AsymmetrixVariable detector length Fixed detector length

    Quad Detector Technology

    variable wide area detector

    Variable slice thickness

    4 x 1mm

    4 x 5mm

    4 x 2.5mm

    2 x 0.5mm

    2 x 8mm

    2 x 10mm

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    24/39

    8 Element

    2-D array

    4 Slices

    Quad TechnologyHow it works

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    25/39

    CT

    CT attenuation information

    CT image quality

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    26/39

    Attenuation

    X-ray beam passes through patient

    Each structure attenuates X-ray beam

    differentlyAccording to individual densities

    Radiation received by detector varies

    according to these densities

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    27/39

    Density information

    Transferred from detector to CT computer

    (A to D converter)

    Reconstructed by computer into a cross-sectional image

    Displayed on screen

    Each pixel displayed on monitor has varying brightness

    The greater the attenuation, the brighter the pixel

    The less attenuation, the darker the pixel

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    28/39

    Density information

    Density values

    correspond to a

    range of numbers

    Hounsfield scale

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    29/39

    Window settings

    Window width

    Determines range of CT numbers displayed on an image

    Values above this range = white

    Values below this range = black

    Window level

    Sets the center CT number displayed on the monitor

    Determines the location on the Hounsfield scale about which the

    window width will be centered

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    30/39

    CT image quality

    Spatial resolution

    Ability to resolve small

    objects in an image

    Measured in lp/cm

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    31/39

    CT image quality

    Spatial resolution

    Ability to resolve small

    objects in an image

    Measured in lp/cm

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    32/39

    Isotropic Imaging

    True 0.5mm Isotropic imaging

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    33/39

    CT image quality

    Contrast resolution

    Ability to differentiate small density differences in an image

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    34/39

    Post Processing Options

    Visualization of

    vasculature in

    relation to pathology

    Show course of vessels Show stent placement

    Define vascular stricture

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    35/39

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    36/39

    Neuro-Angiography

    Circle of Willis

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    37/39

    Thin-SliceSpiralLungs

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    38/39

    MIP4D-Angio

    MasterCutPanoramic View

    RenalArteries

  • 7/28/2019 Basics_of_CT Scan(Lovish Vij; Eic 2 ; 101285007)

    39/39

    Ext. Spiral Acq.Spiral Acquisition

    UltraFast 0.5 sec

    Coverage 1400mm

    Pitch 1.75

    Acq. Time 41.5 secFOV 420 mm

    ST2.5 mm

    Rec. Incr. 1.6 mm

    Std Res. 8 lp/cm

    120 kV, 96 mAs

    CTDI100w 7.3 mGy

    CTDIFDAw 4.9 mGy

    Extended Spiral Acquisition