1
BioimpedanceBasics
FYS4180 – høsten 2012Ørjan G. Martinsen
![Page 2: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/2.jpg)
2
BioimpedanceBasics
Impedance:• is a measure of the impediment or opposition to the flow of alternating current
• is a complex number.
• has an “ohmic” part (free ions) and a capacitive part (bound ions).
• is the inverse of admittance (conductance and susceptance)
It:• changes with tissue type, physiology and anatomy
• can hence be used as a transducer mechanism for physiological parameters
• can be measured with two, or more electrodes
• can be measured invasively or non‐invasively
• depends on the measurement frequency
• does normally not require any expensive components
![Page 3: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/3.jpg)
3
Material propertiesSimple parallel plate:
AdC
AdG
r
0 :tyPermittivi
:tyConductivi
dAR :yResistivit
Complex values
= ’ + j ’’
ρ = ρ’ + j ρ’’
= ’ + j ’’
![Page 4: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/4.jpg)
4
Two‐electrode system
sinIm Reactance
cosRe Resistance
Impedance
iv
ivX
iv
ivR
ivZ
sinIm eSusceptanc
cosRe eConductanc
1 Admittance
vi
viB
vi
viG
ZviY
sin1sin90 ifonly 1 and
cos1cos0 ifonly 1
BX
GR
![Page 5: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/5.jpg)
5
Four‐electrode system
![Page 6: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/6.jpg)
6
Transfer impedance
i v i v
• Input: i , output: v , transfer function:
• This is transfer impedance, not impedance
• Cannot use for calculation of , or
• Except for homogenous, uniform material
A B
Ziv
![Page 7: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/7.jpg)
7
Sensitivity field calculations
• Simple and very powerful tool, but not used very much
• Example: Direct current resistance measurement with a four‐electrode system on a homogeneous medium
![Page 8: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/8.jpg)
8
Sensitivity field calculations #1
Imagine that you inject a current I between the two current electrodes, and compute the current density J1 in each small volume element in the material as a result of this current
![Page 9: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/9.jpg)
9
Sensitivity field calculations #2
Imagine that you instead inject the same current between the voltage pick-up electrodes, and again compute the resulting current density J2 in each small volume element.
![Page 10: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/10.jpg)
10
Sensitivity field calculations #3
• The sensitivity S and total measured resistance R are then (reciprocal system):
• Positive S: Increased resistivity in this area higher total measured resistance• Negative S: Increased resistivity in this area lower total measured resistance• High absolute value of S: Very sensitive to changes in resistivity
dvI
RI
SV
2
212
21 and JJJJ
![Page 11: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/11.jpg)
11
Four‐electrode system
![Page 12: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/12.jpg)
12
Closer look …
![Page 13: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/13.jpg)
13
Adding a low‐conducting slice
A slice with only 20% of the conductivity of the surrounding medium is added
![Page 14: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/14.jpg)
14
Volume impedance density
The plot now shows the sensitivity divided by the conductivity (same as multiplying with the resistivity, since this is DC).
![Page 15: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/15.jpg)
15
Two‐electrode system
![Page 16: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/16.jpg)
16
Dispersions
Counter-ions
Charging of cell membranes
Relaxation ofwater molecules
E.g. makromolecules
![Page 17: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/17.jpg)
17
Data analysis• Measured with Solartron 1260 +
1294• Four electrodes on lower leg• Exported as text file
![Page 18: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/18.jpg)
18
Bode plot of impedance
-1.00E+01
0.00E+00
1.00E+01
2.00E+01
3.00E+01
4.00E+01
5.00E+01
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06
Z'Z''
![Page 19: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/19.jpg)
19
Bode plot of admittance
-2.00E-02
-1.00E-02
0.00E+00
1.00E-02
2.00E-02
3.00E-02
4.00E-02
5.00E-02
6.00E-02
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06
Y'Y''
2222 XRXB
XRRG
![Page 20: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/20.jpg)
20
Bode plot of |Z| and
-20
-10
0
10
20
30
40
50
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06
|Z|f
![Page 21: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/21.jpg)
21
Bode plot of |Y| and
0
0.01
0.02
0.03
0.04
0.05
0.06
1.00E+00 1.00E+01 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06-1.50E+01
-1.00E+01
-5.00E+00
0.00E+00
5.00E+00
1.00E+01
1.50E+01
|Y|f
![Page 22: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/22.jpg)
22
Wessel, Nyquist, Argand, Cole, …
-6.00E+00
-4.00E+00
-2.00E+00
0.00E+00
2.00E+00
4.00E+00
1.5E+01 2.0E+01 2.5E+01 3.0E+01 3.5E+01 4.0E+01 4.5E+01
Impedance
-1.00E-02
-5.00E-03
0.00E+00
5.00E-03
1.00E-02
2.0E-02 2.5E-02 3.0E-02 3.5E-02 4.0E-02 4.5E-02 5.0E-02 5.5E-02
Admittance
![Page 23: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/23.jpg)
23
Measuring sweat activity
![Page 24: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/24.jpg)
24http://www.fys.uio.no/elg/bioimp
Elektrodermal response EDR
• Psychiatry• Neurological
diseases• Pain assessment• Lie detection• Anesthetic depth
0
2
4
6
8
10
12
14
16
18
0 10 20 30 40 50 60 70
Time [sec]
Con
duct
ance
G o
r sus
cept
ance
B S
/cm
2 ] G righ
G left
B right
B left
![Page 25: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/25.jpg)
25
Measurements under stress
0
2
4
6
8
10
12
14
16
18
20
13:33 13:34 13:35 13:36 13:36 13:37 13:38 13:39 13:40 13:40 13:41 13:42 13:43 13:44 13:44 13:45 13:46 13:47
Skin
AC
con
duct
ance
[µS/
cm2 ]
Left axilla
Right axilla
Hypothenar
Abdomen
Relaxation Relaxation
Stress‐period withcalculations
![Page 26: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/26.jpg)
26
Measurements during physical excercise
![Page 27: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/27.jpg)
27
Needle positioning –where is the needle tip?
![Page 28: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/28.jpg)
28
![Page 29: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/29.jpg)
29
• Insulated shafts
• Commonly used for EMG
• Apply a small electric current between the needle and a counter‐electrode
• Measurement sensitivity is J 2
• Can also use hollow needle
Needle positioning
![Page 30: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/30.jpg)
30
![Page 31: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/31.jpg)
31
![Page 32: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/32.jpg)
32
Collecting tissue data
![Page 33: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/33.jpg)
33
Classification – arteries and veinsone outlier removed
![Page 34: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/34.jpg)
34
Classification – muscle and fat
![Page 35: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/35.jpg)
35
Classification – subdermis and fat
![Page 36: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/36.jpg)
36
Needle arrayTo find blood vessels (artery and vein) during Emergency Cardiopulmonary Bypass (ECPB) for hypothermia and controlled reperfusion. The objective is to develop an automatic tool that finds blood vessels and
guides the cannulae into the vessels.
![Page 37: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/37.jpg)
37
![Page 38: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/38.jpg)
38
Invasive 2D imaging?
• Ongoing data analysis• No classification yet• Simple plot of phase angle
at 100 kHz
![Page 39: Bioimpedance Basics](https://reader030.vdocument.in/reader030/viewer/2022012809/61bfbf9982fd584aea4bb083/html5/thumbnails/39.jpg)
39
First invasive impedance tomography image
• No multivariate analysis• No classification only continuous color
• Simple plot of phase angle at 100 kHz
• Multivariate classification will most likely increase the image quality
• Step in vs. step out