biomedical engineering company of australia consulting engineers biomec australia 1 as/nzs 3003:2011...

98
Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton and Lawrie Knuckey February 2012 Seminar

Upload: ahmad-holby

Post on 30-Mar-2015

226 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 1

AS/NZS 3003:2011Electrical installations

Patient areasStuart Clifton and Lawrie

Knuckey

February 2012 Seminar

Page 2: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 2

Clinical engineering framework• Physiological effects of electric current on living tissue:

– Nerve and muscle stimulation– Heating – Deposition of metal ions (current with any d.c. component)

• Pathological effects of electric current on living tissue:– Cardiac arrest– Respiratory arrest– Burning– Denervation– Ulceration (d.c. burns)

• Risk of death/severity of injury – major factors:– Magnitude of the current– Duration of current flow– Current path

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 3: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 3

Clinical engineering framework• Macroshock

– Current passing through the body other than via direct cardiac contact.

• Microshock– Current passing through the body via direct electrical contact with the

heart.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 4: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 4

Clinical engineering framework• Macroshock

– Often apparent from patient reaction ranging from sensory stimulation to violent muscular contraction.

– Burning, denervation and possibly ventricular fibrillation may occur depending on the magnitude and path of the current. (Cardiac standstill and massive burning may occur in high voltage shocks e.g. contact with 12 kV conductor; lightning strike.)

– The possibility of macroelectrocution is exacerbated with the application of medical electrical equipment when:

• the skin resistance is reduced or bypassed by electrodes, electrode paste or gel, or invasive conductors, or

• the subject is strapped to the equipment, restrained or unconscious, thus inhibiting escape through voluntary or involuntary action.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 5: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 5

Clinical engineering framework• Microelectrocution

– Risk exists when an electrical conductor is placed into direct contact with a ventricular heart muscle.

– Ventricular fibrillation may be induced by current as low as 100 μa (below the threshold for feeling or visible muscle stimulation).

– May occur when an electrode wire, electrolyte-filled catheter or transducer is introduced into direct contact with ventricular heart muscle.

– Such procedures include:• cardiac catheterization, • intracardiac ECG, • cardiac pacing,• Swan Ganz procedure.

– These are referred to in AS/NZS 2500 as cardiac-type procedures.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 6: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 6

Clinical engineering framework• Micro-electrocution

– Patient must not be exposed to:• Voltages >50 mV --- need EP earthing.• Current sources >50 μA --- need intact earth wires.

– Safe equipment design needed:• Medical electrical equipment/systems must comply with standards.

– Safe installation needed:• Patient area must be wired as CPA.

– Safe equipment application needed:• Flexible protective earthing conductors must be intact.• Hospital must control leakage currents when assembling trolley-

mounted medical electrical systems.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 7: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 7

Clinical engineering frameworkSafe use of electricity in medicine is maximized by following the safety triangle:

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

SAFE EQUIPMENT

• AS/NZS 3200-series• IEC 60601-series

SAFE INSTALLATION

• AS/NZS 3000 and 3003 (wiring)

• X-ray protection installation codes

• Surgical laser installation codes

SAFE USE

• AS/NZS 2500 and AS/NZS 3551

• Preventive maintenance.• Software upgrades.• Safety testing.• Performance verification.• User education.

Page 8: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 8

Regulatory framework• Throughout Australia, state legislation requires:

– Compliance with AS/NZS 3000 (the “Wiring Rules”).– Electrical installation work must be carried out by licensed electrical

workers.• Electrical installation work includes:

– Wiring a building.– Installing fixed electrical equipment.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

We are consulting engineers – not lawyers. Legal questions may need to be discussed with a state electrical authority or a lawyer.

Page 9: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 9

Regulatory framework• Compliance with AS/NZS 3003:

– Was not a legal requirement when first published in 1974.– Became a requirement of hospital building contracts.– Became a requirement of medical and allied health professional bodies

(e.g. ACHS).– Became a legal requirement in Victoria in 1999.– Became a legal requirement throughout Australia in 2007 when it was

included as a requirement of AS/NZS 3000.• Formal inspection/testing of AS/NZS 3003 installation work – by a licensed

electrical inspector – became a legal requirement in Victoria in 1999. • Formal inspection/testing of AS/NZS 3003 installation work – by anyone

acceptable to the hospital – became a legal requirement throughout Australia in 2011 when it was included as a requirement of AS/NZS 3003.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 10: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 10

Regulatory framework• The law does not in itself require electrical installations to be upgraded to

comply with changes in AS/NZS 3003, but … – New electrical installation work must comply with current standards.– Alterations* to existing electrical installations are deemed to be new

electrical installations.– Alterations* to locations used for medical electrical equipment can only

be carried out if the location is a BPA or a CPA.*Alterations from new socket-outlets to a new or replacement CT scanner. – AS/NZS 3003:2011 also requires some upgrading of patient areas when

certain alterations are carried out.• Upgrading old facilities:

– Is a requirement of AS/NZS 2500.– May be a duty of care under occupational health and safety legislation

or common law.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 11: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 11

Regulatory framework• Implementation of AS/NZS 3003:

– AS/NZS 3003 is ignored in significant parts of the medical and allied health sectors.

– Broad acceptance of legal requirements is inhibited by: • Excessive requirements in the 2011 edition of the standard.• Lack of action by state electrical authorities.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 12: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 12

Regulatory framework• Energy Safe Victoria guidelines:

– Even plug-in appliances are “fixed” electrical equipment if:a) >18 kg without wheels; orb) fixed in position; orc) recommended to be fixed in position; ord) provided with means for fixing; ore) part of a number of discrete components that function as a group.

– Fixed means secured, attached or connected by conduit, metal pipe-work, duct, bolt, screw, clip, other fastening device or fixed wiring.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 13: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 13

Regulatory framework• Energy Safe Victoria guidelines:

– Even plug-in appliances are “fixed” electrical equipment if:a) >18 kg without wheels; orb) fixed in position; orc) recommended to be fixed in position; or **d) provided with means for fixing; ore) part of a number of discrete components that function as a group.

– Fixed means secured, attached or connected by conduit, metal pipe-work, duct, bolt, screw, clip, other fastening device or fixed wiring.

**Includes plug-in medical electrical equipment such as:Dental chairs.Mammography units.Wall-mounted X-ray viewing boxes.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 14: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 14

Regulatory framework• Energy Safe Victoria guidelines:

– Even plug-in appliances are “fixed” electrical equipment if:a) >18 kg without wheels; orb) fixed in position; orc) recommended to be fixed in position; ord) provided with means for fixing; ore) part of a number of discrete components that function as a group.**

– Fixed means secured, attached or connected by conduit, metal pipe-work, duct, bolt, screw, clip, other fastening device or fixed wiring.

**Seems to include multi-bed patient monitoring systems!

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• B

PAs (body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 15: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 15

New edition of AS/NZS 3003• New standard still applies in patient areas:

(Wherever mains powered medical electrical equipment is used.)• Patient areas must still be wired as:

BPAs (body-protected electrical areas), or CPAs (cardiac-protected electrical areas)

• LPDs (leakage protective devices) still required for:– Most socket-outlets.– Some high risk medical electrical equipment.

• LPDs are still:– Type 1 RCDs (residual current devices), or– Isolation transformers with line isolation/overload monitors.

• Socket-outlet colours unchanged except for some generator-backed circuits.• Requirements for EP (equipotential) earth wiring in CPAs similar to 2013 edition –

but with better drawings.

• new standard:– Still

applies wherever mains powered medical electrical equipment is used.

– Still requires patient areas to be wired as:• BPAs

(body-protected electrical areas), or

• CPAs (cardiac-protected electrical areas)

• BPAs and CPAs

Page 16: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 16

Summary of major changes• Each requirement is followed by a compliance test.• New definition of essential supply.• Certain patient areas must only be BPAs.• Certain patient areas must be CPAs.• LPD protection and EP earthing of socket-outlets outside patient areas:

– Has been extended from 2 m to 5 m from each entry.– Is now required with or without a door.

• Final sub-circuits may only supply one BPA and its ensuite.• Final sub-circuits may no longer supply socket-outlets for cleaning

purposes and other socket-outlets.• Socket-outlets requiring LPD protection clearly specified:

Page 17: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 17

Summary of major changes• UPS status indicator now required in some patient areas.

• RCDs must now be:

– readily accessible,– not installed under benches, and– not positioned near some light switches.

• Inaccessible socket-outlets must now have readily accessible isolating switches.

• Socket-outlets for cleaning purposes now only required near BPAs.

• EP terminals no longer required in CPAs.

• The EPJ (equipotential junction) in a CPA may now have conveniently located nodes.

Page 18: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 18

Summary of major changes• New commissioning and certification requirements.

• New requirements for special patient areas:

– Home installations for haemodialysis.– Home installations for other medical electrical equipment. – Self harm areas (e.g. psychiatric and prison wards).

• New requirements for alterations, additions and repairs to electrical installations in patient areas:

Page 19: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 19

Summary of major changes• New limitations on magnetic fields:

– Maximum field strength specified for some (not all) patient areas intended for ECG monitoring and recording.

– Requirements apply prior to commencing work, and on completion of work.

– Requirements apply to new patient areas and to alterations.

• New marking requirements for BPA and CPA area signs.• Full technical requirements for routine inspection and testing.

Page 20: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 20

New title Contention about the institutions covered by previous editions:

– Hospital. – Dental surgery.– Day surgery.– Imaging practice.– Physiotherapy practice.– Doctors’ room.– Nursing home.

– Other aged care facilities.

AS/NZS 3003 should apply in patient areas of any institution.

?

Page 21: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 21

New title1976 Electrical installations in electromedical treatment areas

1985 Electrical installations — Patient treatment areas of hospitals and medical and dental practices

1999 Electrical installations — Patient treatment areas of hospitals and medical and dental practices

2003 Electrical installations — Patient areas of hospitals, medical and dental practices and dialyzing locations

2011 Electrical installations — Patient areas

Page 22: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 22

Patient areas2003 Locations in hospitals and medical and dental practices in which

it is intended that mains powered medical electrical equipment will be used. Locations in patients’ homes and other facilities intended for dialysis are also included.

2011 Locations where it is intended that mains powered medical electrical equipment will be used on a patient. This does not include areas such as corridors and lifts where medical equipment is only used in an emergency or for transporting patients.

Page 23: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 23

Patient areas2003 Locations in hospitals and medical and dental practices in which

it is intended that mains powered medical electrical equipment will be used. Locations in patients’ homes and other facilities intended for dialysis are also included.

2011 Locations where it is intended that mains powered medical electrical equipment will be used on a patient. This does not include areas such as corridors and lifts where medical equipment is only used in an emergency or for transporting patients.

The standards applies wherever mains powered medical electrical equipment is used.

Page 24: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 24

Patient areas2003 Locations in hospitals and medical and dental practices in which

it is intended that mains powered medical electrical equipment will be used. Locations in patients’ homes and other facilities intended for dialysis are also included.

2011 Locations where it is intended that mains powered medical electrical equipment will be used on a patient. This does not include areas such as corridors and lifts where medical equipment is only used in an emergency or for transporting patients.

Discourages application of the standard in areas used just for charging/storing medical electrical equipment (e.g. beds).

Page 25: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 25

Patient areas2003 Locations in hospitals and medical and dental practices in which

it is intended that mains powered medical electrical equipment will be used. Locations in patients’ homes and other facilities intended for dialysis are also included.

2011 Locations where it is intended that mains powered medical electrical equipment will be used on a patient. This does not include areas such as corridors and lifts where medical equipment is only used in an emergency or for transporting patients.

The standard only applies in areas intended for the regular use of medical electrical equipment.

Page 26: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 26

Patient areas2003 Locations … in which it is intended that mains powered medical

electrical equipment will be used …2011 Locations where it is intended that mains powered medical electrical

equipment will be used …

• Under either standard:– Where a room is intended for medical electrical equipment that will

be connected to a patient in another room, both rooms are patient areas. For example:

A&E/ICU/CCU/ward/ staff base with patient monitor.Cardiac cath lab recording room with physiological recorder.Sleep lab recording room.CT control room with injector control unit.

Page 27: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 27

Patient areas• Electrical equipment supplied by socket-outlets outside patient areas:

– LPD protection required: Socket-outlets outside BPAs or CPAs used to supply medical electrical equipment in the patient area.

– EP earthing required:Socket-outlets outside CPAs used to supply:• medical electrical equipment in the CPA, or• any electrical equipment accessible in the patient environment.

EXAMINATION LIGHT MUST BE PLUGGED IN HERE

Page 28: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 28

Patient areas• Within a patient area:

Page 29: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 29

“Accessible” & “readily accessible”• Readily accessible now means:

– Capable of being reached quickly and without climbing over or removing obstructions, standing on a chair or using a ladder, and

– Not more than 2 m above the ground, floor or platform.• RCDs must now be readily accessible and not mounted under benches, in

cupboards or cabinets, or behind heavy equipment.• Socket-outlets require LPD protection if they are accessible – not just

readily accessible – without a tool.• Socket-outlets that are not readily accessible must have readily accessible

isolating switches or a dedicated LPD. • Permanently wired appliances protected by LPDs that also protect socket-

outlets or other appliances must have readily accessible isolating switches.

Page 30: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 30

“Group” of socket-outlets• Group of socket-outlets now means:

One or more socket-outlets contained within one surround.

• No longer a group:

Page 31: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 31

Extent of the patient area

2003 The boundary of the cardiac-protected electrical area extends to the walls enclosing the patient location(s) or, if such walls are incomplete (e.g. entryway without a door) to the boundary formed if such walls or projections were extended.

Page 32: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 32

Extent of the patient area

2003 Socket-outlets that are:• located within 2 m of a patient area, and• in free communication with the patient area (i.e. no door),… must be:• protected by an LPD, and• connected to thearthing system (if CPA).

Page 33: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 33

Extent of the patient areaFirst public draft: Where the walls of a BPA are incomplete (e.g. entryway without a door) and the opening is greater than 2 m, the adjacent area is part of the BPA.

Page 34: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 34

0 1 2

Extent of the patient area

Page 35: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 35

Extent of the patient area

Page 36: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 36

Extent of the patient area

Page 37: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 37

Extent of the patient area

2011 The boundary of the cardiac-protected electrical area extends to the walls enclosing the patient location(s) or, if such walls are incomplete (e.g. an entryway without a door) to the boundary formed if such walls or projections were extended.

In this context, the following are not considered walls or boundaries: partitions, dividers, screens, benches, moveable walls or curtains.

Page 38: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 38

Extent of the patient area

2003 Socket-outlets that are:• located within 2 m of a patient area, and• in free communication with the patient area (i.e. no door),… must be:• Protected by an LPD, and• Connected to the EP earthing system (if CPA).

Page 39: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 39

Extent of the patient area

2011 Socket-outlets that are:• located within 5 m of a patient area (door or no door), • in any room, corridor, etc, opening directly off the patient area,… must be:• Protected by an LPD, and• Connected to the EP earthing system (if CPA).

Page 40: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 40

Extent of the patient area

2011 The boundary of the cardiac-protected electrical area extends to the walls enclosing the patient location(s) or, if such walls are incomplete (e.g. an entryway without a door) to the boundary formed if such walls or projections were extended.In this context, the following are not considered walls or boundaries: partitions, dividers, screens, benches, moveable walls or curtains.

Page 41: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 41

Extent of the patient area

2011 The boundary of the cardiac-protected electrical area extends to the walls enclosing the patient location(s) or, if such walls are incomplete (e.g. an entryway without a door) to the boundary formed if such walls or projections were extended.In this context, the following are not considered walls or boundaries: partitions, dividers, screens, benches, moveable walls or curtains, and entryways wider than 10% of the perimeter of the patient area. (Proposed ESV ruling.)

Page 42: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 42

Extent of the patient area

2011 The boundary of the cardiac-protected electrical area extends to the walls enclosing the patient location(s) or, if such walls are incomplete (e.g. an entryway without a door) to the boundary formed if such walls or projections were extended.In this context, the following are not considered walls or boundaries: partitions, dividers, screens, benches, moveable walls or curtains, and entryways wider than 10% of the perimeter of the patient area. (Proposed ESV ruling.)

Page 43: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 43

Socket-outlets near patient areas

Page 44: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 44

Socket-outlets near patient areas

STAFF BASEBUSY CORRIDORBEDROOM

< 5m

xxxxxxxxxxxx

Page 45: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 45

Socket-outlets near patient areas

xxxxxxxxxxxx

xxxxxxxxxxxx

xxxxxxxxxxxx

xxxxxxxxxxxx

Page 46: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 46

Socket-outlets near patient areas

Page 47: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 47

New essential supply definition • Generator-backed circuits that are subject to load shedding are no longer

considered as essential supplies.

• Many generator-backed outlets are no longer on essential supplies

• ESV has issued exemptions allowing additional red socket-outlets on circuits subject to load shedding under programmable load shedding.

POWERAVAILABLE

NEW SOCKET-OUTLET EXISTING SOCKET-OUTLETS

POWERAVAILABLE

POWERAVAILABLE

Page 48: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 48

Classification of patient areas Some locations must now be CPAs:• Cardiac catheter laboratories and control rooms.• Cardiac ICU.• CCU.• ICU with regular thermodilution Swann-Ganz monitoring.• Neo natal ICU (Level 3).• Operating theatres for cardiac and thoracic surgery.• Other locations intended for regular or routine cardiac-type procedures.

(No mention of wards)

Page 49: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 49

Classification of patient areas Some locations must only be BPAs (unless intended for cardiac-type procedures):

• Accident and emergency wards.• Blood collection areas.• Chiropractic and physiotherapy rooms.• CT rooms.• Day procedure theatres.• Delivery suites.• Dental surgeries.• Endoscopy theatres and procedure rooms.• General medical and surgical wards.• High dependency wards.• MRI rooms.• Nurseries.• Operating theatres for non-cardiac surgery • Patient ensuites• Patient examination rooms in outpatients.

• Plasmapherisis areas.• Plaster rooms.• Recovery areas or wards associated with

operating theatres and imaging rooms.• Respiratory function laboratories.• Resuscitation bays.• Stress test rooms.• Treatment rooms.• Ultrasound rooms.• X-ray rooms (Except cardiac cath labs!)• Doctors consulting room.• Nursing homes. (Patient areas only!)

Page 50: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 50

?

Classification of patient areas Some locations must only be BPAs:• Patient ensuite bathrooms,

shower rooms, toilets …

Page 51: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 51

Classification of patient areas BPAs and CPAs still not allowed in the same room.

Page 52: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 52

UPS status indicators• UPS status indicator now required in patient areas with:

– socket-outlets or permanently wired equipment supplied from a UPS, and

– continuous patient observation, … including: – operating theatres, and nursing stations or staff bases in:

recovery wards, intensive care units, isolation rooms, andother special care areas.

Page 53: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 53

UPS status indicators• UPS status indicator requirements:

– Must be marked “UPS STATUS”.– Must be connected to the UPS electrical supply system.– Loss of supply to the UPS must initiate synchronized, pulsating,

audible and visual alarms.– Audible alarm must continue until muted.– Visual alarm must continue until the supply is restored to the UPS or

the UPS is exhausted.

• New standard is not retrospective:

– Existing installations do not have to be upgraded, but …– new work (one new socket-outlet) must comply with this requirement.

Page 54: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 54

Socket-outlet protection• Socket-outlets requiring LPD protection* clearly specified:

*Except socket-outlets mounted on – and forming part of – a medical electrical

system.

Page 55: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 55

Socket-outlet protection• Intention was to treat 10 amp IEC sockets

on endoscopy systems as general purpose outlets.

… except for unintended consequence:

*Except socket-outlets mounted on – and forming part of – a medical electrical

system.

Page 56: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 56

Socket-outlet protection• Socket-outlets requiring LPD protection* clearly specified.

--- another unintended consequence:

xxxxxxxxxxxx

xxxxxxxxxxxx

xxxxxxxxxxxx

Special purpose industrial outlet

in Pan Room

Page 57: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 57

Magnetic field requirements7.1 GENERAL• Health care facilities must ensure magnetic fields do not exceed safe levels

in areas used for sensitive monitoring or recording equipment.• Maximum field strength and test methods are specified for locations

intended for ECG monitoring and recording.• For electrophysiological measurements such as EEG, EMG and ENG:

– Substantially more demanding requirements apply.– Maximum field strength and test methods are not specified.– Facility management must ensure magnetic fields are reduced to

acceptable levels when specifying the wiring of such locations.

Page 58: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 58

Magnetic field requirements7.2 REQUIREMENTS• The following patient areas “where ECG monitoring and recording is

expected to be carried out” must be arranged so that:– Total magnetic field flux density does not exceed 3 microtesla (μT).– No individual harmonic exceeds 1.6 μT. – Limits apply between 800 and 1200 mm above floor, over the whole

area except within 300 mm of the walls.• Nominated patient areas:

– Accident and emergency departments.– Cardiac catheterisation laboratories.– Coronary care units and intensive care units.– Outpatient departments for diagnostic ECG.– Resuscitation units.– Stress testing units.

Page 59: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 59

Magnetic field requirements• Nominated patient areas:

– Accident and emergency departments.– Cardiac catheterisation laboratories.– Coronary care units and intensive care units.– Outpatient departments for diagnostic ECG.– Resuscitation units.– Stress testing units.

• It is unclear why these requirements do not apply in locations such as:- Cardiac ultrasound rooms.- ECG rooms (other than those in outpatient departments).- ECT rooms.- Imaging rooms with ECG monitoring or synchronised imaging.- Operating theatres.- Procedure rooms intended for general anaesthesia.

Page 60: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 60

Magnetic field requirements7.3 TESTING• New building:

– Prior to new work: Contractor must confirm with facility management that electrical reticulation has been designed and located to achieve compliance.

– After work is finished: Contractor must arrange compliance testing.• Existing building:

– Prior to new work: Contractor must confirm with facility management that satisfactory testing has been carried out.

– After work is finished: Contractor must arrange for compliance testing to be repeated.

Page 61: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 61

Magnetic field requirements7.3 TESTING• Existing building:

– Prior to new work: Contractor must confirm with facility management that satisfactory testing has been carried out.

– After work is finished: Contractor must arrange for compliance testing to be repeated.

• Within existing buildings, these requirements apply to:– New patient areas – such as:

• New cardiac cath lab. New ICU.– Alterations in existing patient areas – such as:

• New $30 socket-outlet in existing ICU.• New $1,000,000 X-ray system in existing cardiac cath lab.

Page 62: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 62

Magnetic field requirements7.3 TESTING• Existing building:

– Prior to new work: Contractor must confirm with facility management that satisfactory testing has been carried out.

– After work is finished: Contractor must arrange for compliance testing to be repeated.

What about a room that has been successfully used for electrocardiography for 20 years?

Page 63: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 63

Magnetic field requirements7.3 TESTING• New building:

– Prior to new work: Contractor must confirm with facility management that electrical reticulation has been designed and located to achieve compliance.

– After work is finished: Contractor must arrange compliance testing.

What about trying out the room with an ECG machine?

Page 64: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 64

Socket-outlets for cleaning purposesSocket-outlet marked for cleaning purposes:

• Still required in each CPA.

• Can be outside a BPA if located within 15 m.

• Not required in MRI rooms.

• Still required in CPA ensuites.

• Not required in BPA ensuites if available within 15 m.

• Must still be protected by an LPD.

• Must be supplied from a different circuit than other socket-outlets.

(This circuit may supply a large number of such socket-outlets.)

Page 65: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 65

Separation of circuits• BPAs

Circuits supplying socket-outlets* must not supply electrical points in another room (except any ensuite).

• CPAsCircuits must not supply:– Socket-outlets* in more than one patient location.– Electrical points in another room.

*Except socket-outlets for cleaning purposes• Socket-outlets for cleaning purposes

– Must be supplied by dedicated circuit.– Circuit may supply many patient areas.

Page 66: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 66

0.1 EP earthing system retained• But EP terminals no longer required. • Looping allowed providing it does not depend on pressure of single screw

bearing directly on wiring.

• Cpa euite

SINGLE SCREW TYPE

NOT ALLOWED

SINGLE SCREW OK AS DOES NOT BEAR DIRECTLY ON WIRE

JOINING WIRES UNDER CRIMP THEN INTO SINGLE

SCREW CONNECTOR IS ALLOWED

Page 67: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 67

0.1 EP earthing system retained• EP junction may have multiple nodes.• Nodes must be isolated from structural metal.

EP JunctionNODE NODE

0.01 ohm connections between the EPJ and

its nodes.0.1 ohm connections to EP junction or one of

its nodes.

Distribution Board

Page 68: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 68

0.1 EP earthing system retained• RCDs mounted in structurally connected wall boxes

Page 69: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 69

0.1 EP earthing system retained• RCDs mounted in non-structurally connected wall boxes

Page 70: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 70

0.1 EP earthing system retained• Transformer isolated supply - structurally connected wall boxes

Page 71: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 71

0.1 EP earthing system retained• Transformer isolated supply – non-structurally connected wall boxes

Page 72: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 72

0.1 EP earthing system retained• But all equipotential conductor now ≥ 4mm2.

E (SOCKET-OUTLETS) E (ENCLOSURE)

EARTH WIRING TO PROTECTIVELY EARTHED

PARTS OF THE ENCLOSURE

EARTH WIRING TO SOCKET- OUTLETS

EPJ

Page 73: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 73

0.1 EP earthing system retained• Special earthing required for X-ray and imaging systems:

Page 74: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 74

0.1 EP earthing system retained• 0.1 EP earthing system still only works for building earth currents < 0.5 A.• Building earth currents > 16 A easily encountered in buildings with more

than one connection between neutral bars and earth, e.g. old buildings with metal-clad wiring.

• Test 1 - verify protective earthing of socket-outlets and fixed Class I electrical equipment.

• Test 2 - verify EP earthing of structurally connected, non-electrical fittings in the patient environment.

• Test 3 - verify <50 mV between the EP junction and:– All socket-outlets in the CPA.– Fixed electrical equipment in the patient environment.– Structurally connected, non-electrical fittings in the patient

environment.

Page 75: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 75

Marking of BPA and CPA wall signs

• Person and organization that carried out the required certification:

• Date of certification.

• Person or organization that carried out the most recent electrical safety testing.

• Date of test.• Report number• Date by which the next test is required.

• Edition of AS/NZS 3003 against which the area was originally commissioned.

Page 76: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 76

Marking of BPA and CPA wall signs

• Person and organization that carried out the required certification of:- The original installation?- The latest installation work?

• Date of certification.

• Person or organization that carried out the most recent electrical safety testing.

• Date of test.• Report number• Date by which the next test is required.

• Edition of AS/NZS 3003 against which the area was originally commissioned.

Page 77: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 77

Marking of BPA and CPA wall signs

• Person and organization that carried out the required certification.

• Date of certification.

• Person or organization that carried out the most recent electrical safety testing.

• Date of test.• Report number• Date by which the next test is required.

The most recent testing may be just a new socket-outlet, but the new sticker obliterates what is often the only record of the last full routine test.

• Edition of AS/NZS 3003 against which the area was originally commissioned.

Page 78: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 78

Marking of BPA and CPA wall signs

• Person and organization that carried out the required certification:

• Date of certification.

• Person or organization that carried out the most recent electrical safety testing.

• Date of test.• Report number• Date by which the next test is required.

• Edition of AS/NZS 3003 against which the area was originally commissioned

Biomec:• Combines these two labels

(saved cost and effort).• Assumes this sticker relates to

certification of the original installation.

Page 79: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 79

Marking of BPA and CPA wall signs

Page 80: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 80

Statement by electrical contractorThe contractor must:

• Certify in writing that the work has been tested to – and complies with – AS/NZS 3000.

• Certify in writing that the work followed a formal specification prepared by the health care institution or practice, or… that the number of socket-outlets has been discussed with medical or nursing personnel and the likely loading on each circuit has been checked.

• Supply documentary evidence that the isolation transformer, line isolation monitor and overload monitor comply with AS/NZS 4510.

• Supply a certificate of compliance or a manufacturer’s statement that each ELV supply complies with a suitable standard.

Page 81: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 81

Special patient areas• Revised requirements for home haemodialysis:

– Room to be wired as BPA with following exceptions. – Dedicated circuit protected by LPD required for socket-outlets used for

medical electrical equipment. – Other socket-outlets to be protected by 30 ma RCD. – Socket-outlet for cleaning purpose not required. – Supply available lights not required unless RO unit is in another room. – RCD protected labeling not required.– BPA area sign not required.

• Questions now emerging from clinicians and clinical engineers about special installation requirements in:– Rental properties.– Holiday locations.

Page 82: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 82

Special patient areas• New requirements for home use of other medical electrical equipment

– No special wiring requirement if medical electrical equipment is: • Internally powered or Class II, and• Provided with Type BF or CF applied part.

– Similar requirements to those for haemodialysis if: • Class I, or • Provided with Type B applied part.

• Questions now emerging from clinicians and clinical engineers about the special installation requirements for the home use of: – Class I medical electrical equipment with Type BF or CF applied parts.– Some Class II medical electrical equipment, even with Type B applied

parts.

Page 83: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 83

Special patient areas• New requirements for BPAs for self harm patients (e.g. psychiatric and

prison wards) – RCDs may be located outside the room.– Amber supply available lights not required on socket-outlets. – BPA area signs may be located outside room entrance.

Page 84: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 84

Alterations, additions and repairs• Alterations and additions

… from new socket-outlets to new or replacement CT scanners.

– All new work must comply with 2011 standard.– Patient area must be upgraded if not already BPA or CPA .– Alteration cannot proceed until routine safety test is up-to-date.– All socket-outlet colours must be upgraded if necessary to 2011

standard if total number is increased by >10%.– Earthing must be upgraded if necessary to the 2011 standard if:

• the EP junction is inaccessible for making new connections, or• new work involves installing fixed equipment rated above 2.0 kW.

• Repairs– Standard goes close to disallowing repairs to fixed medical electrical

equipment in any location not wired as a BPA or a CPA.

Page 85: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 85

Some other outstanding problems

plug-in medical electrical equipment

CPA required wherever plug-in medical electrical equipment is used for cardiac-type procedure, except X-ray and imaging systems.

Page 86: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 86

Some other outstanding problems

such as rooms – not roof spaces!

Locations that have to be wired as CPAs

Page 87: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 87

Some other outstanding problemsLocations that have to be wired as BPAs

BPAs (!) in nursing homes

Page 88: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 88

Some other outstanding problems

socket-outlets (!) in

Circuits in CPAs

Page 89: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 89

Some other outstanding problems

readily accessible (!)

Locations that have to be wired as BPAs

Page 90: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 90

Some other outstanding problems

Table 2.1 of AS/NZS 3009:1998

Socket-outlet colours

(colour coding requirements)

Page 91: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 91

Some other outstanding problems

Available: • in the hospital?• in the particular building of the

hospital?• at the switch-board supplying the

UPS?• in the room housing the UPS?

Supply to UPS that then supplies socket-outlets

Page 92: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 92

Routine in-service testing• Detailed inspection and testing requirements are included in the standard.

• As routine testing is now a requirement in all patient areas, it has become a legal requirement throughout Australia.

• Must be up-to-date before starting any alteration or addition.

• Associated report form is full of errors.

• Biomec recommends that hospitals and medical practices:

– Review procedures or contracts for routine testing of patient areas.– Adopt the new requirements (AS/NZS 3003:2012 Section 9).– Ignore or disallow use of the associated report form (Appendix G).– Disallow test stickers for alterations/additions.

Page 93: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 93

COMMON REQUIREMENTS – BPAs AND CPAs• Testing interval

– Maximum 12 months.• Inspection

– All electrical equipment and fittings.– Labelling.

• Socket-outlets – Colour. – Power-available indicator lights on RCD protected circuits.– Maximum number of protected points.– Toggle marking or “on” indicator lights.– Double-pole switches on transformer isolated supplies.– LPD protection where required.

Routine in-service testing

Page 94: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 94

COMMON REQUIREMENTS – BPAs AND CPAs• RCDs used as LPDs

– Sensitivity test – one test.– Maximum tripping time – two tests. – Test interval may be increased to 24 months based on risk

management relating to the actual RCDs reliability data; but each RCD test button must then be tested each six months.

• Transformer isolated supplies– Installation prospective hazard current – LIM alarm - resistive and capacitive, balanced and unbalanced faults.– Overload alarm

• Uninterruptible power supplies (UPSs) – Check that (someone else?) has tested the UPS status indicators.

Routine in-service testing

Page 95: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 95

Routine in-service testingADDITIONAL TESTS – BPAs

• Protective earthing– Reticulated metal service may be used as a reference in some

circumstances.– Verify earthing resistance to:

• Earth pins of socket-outlets accessible below 2.3 m.• Earth pins of socket-outlets used for electrical appliances

accessible below 2.3 m. • Class I electrical appliances accessible below 2.3 m.

Page 96: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 96

Routine in-service testingADDITIONAL TESTS – CPAs

• Earthing– EP junction/node or medical gas/suction outlet may be used as

reference in some circumstances.– Use four-wire milliohm meter to verify earthing resistance to:

• Earth pins of socket-outlets accessible below 2.3 m.• Earth pins of socket-outlets used for electrical appliances

accessible below 2.3 m. • Class I electrical appliances accessible below 2.3 m.

• Potential difference to EP junction:– At least two socket-outlets in each patient location.– Reticulated plumbing in at least one patient location.– Permanently wired electrical equipment in patient environment.

Page 97: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 97

Thank you

Page 98: Biomedical Engineering Company of Australia Consulting engineers Biomec Australia 1 AS/NZS 3003:2011 Electrical installations Patient areas Stuart Clifton

Biomedical Engineering Company of AustraliaConsulting engineers

Biomec Australia 98

Thank you