biomedical technology wingcmsmedical.com.br/conteudo/downloads/download.php?id=4464_75… · lissy...
TRANSCRIPT
p
BIOMEDICAL TECHNOLOGY WINGSREE CHITRA TIRUNAL INSTITUTE FOR MEDICAL SCIENCES AND TECHNOLOGY
Thiruvananthapuram, INDIA - 695 012
STUDY REPORT
Study Title: Evaluation of Coronary Stent for in vitro hemo compatibility and cyto
compatibility
Study number: TPTRU017
Report number: SRTRU003.07
Revision No: 000
Page 1 of 15
Date effective0 Issued to
Name & Address of theSponsor
Sahajanand Medical Technologies Pvt LtdSaidpura, Surat-395009
Gujarat
]
\n''tA
~ \," "~"~~ . r\
Signature of Study Director t"'~.1i! }."'\)' ).~£} J.' oJ..rt'" S'. ,.~. \.(1, Y
"
~~\\Y -1 .~
( l ~."'" '$J
. ,L,~F U .It-f "~'1 ,.J.~ ~~,.
'I(' .,,~IJ'"'\ ~ ./J)
i~r "...1)",
1 Summary of the study
The objective of this study was to evaluate and compare the
hemocompatibility of modified stents (devices) against bare metal stents
submitted by the sponsor. Tests recommended in IS010993 part4 were
selected for evaluation of hemocompatibility. Devices were exposed to
anticoagulated human blood or platelet rich plasma and analysis of
blood/PRP before and after exposure was done for various parameters, to
determine the effect of the device on blood components/systems. Devices
were analyzed for qualitative/quantitative leukocyte and platelet adhesion.
Hemocompatibility studies using whole blood showed significant hemolysis
with all types of devices including the modified stent (Supracore). Though
there is reduction in platelet count after exposure to whole blood, PRP
exposure and 1-125 PRP exposure results indicate that platelet adhesion is
significant only on bare devices. Fibrinogen consumption and related
prolongation in plasma clotting is also observed in the case of all devices as
compared to reference. In studies with platelets, some effect is seen on
platelet aggregability and secretion which is the least with polymer alone
coated samples. Overall, there is no severe effect on any of the parameters
that were studied to assess hemocompatibility, except that hemolysis is
significant.
2 Identification of study
2.1 Purpose: To evaluate the in vitro hemocompatibility of thedevices using anticoagulated human blood.
2.2 Test Item & reference Item
2.2.1 Name of the Test Item:
Stents: Co-Cr, Polymer alone coated and Supracore(6 Replicates for each device)
Signature ofStudy Director ch~ BMT WING
SCTIMST
ReportNo: SRTRU003.07 Volume No. 02STUDY Page No: 2 of 15 Date effective: 17 Feb 2007
REPORT ControlStatus:
Evaluation of Coronary Stent for in vitro hemo compatibility
2.2.2
2.2.3
2.2.4
2.2.5
2.2.6
2.2.7
Preparation of the Test Item, if applicable
Source of the Test Item: Sahajanand Medical TechnologiesPrivate Ltd
Characterization of Test Item, if applicable: NA
Name of the Reference Item: None
Preparation of the Reference Item: NA
Source of the Reference Item: NA
3 Study Sponsor & Test Facility
3.1 Name and address of the sponsor: Ms. Prathibha, Sahajanand MedicalTechnologies Pvt Ltd
3.2 Name of the study monitor, if any.
3.3 Name and address of the test facility and test sites involved:Thrombosis Research Unit
Biomedical Technology Wing,
Poojapura
Thiruvananthapuram- 695012
Kerala, INDIA.
3.4 Name and address of the study director:
Lissy K. Krishnan, Scientist F, TRU, BMT Wing, SCTIMST, Trivndrum-695012.
3.5 Name and address of the principal investigator, if any.
Krishna Prasad C, Project Scientist, TRU, BMT Wing
4 Study dates
4.1Dateof Samplesubmission:22-4-06to 1-7-06
4.2 Dates of Testing: 23-08-06 to 15-01-07
Signature ofStudy Director ~ BMT WING
SCTIMST
Report No: SRTRU003.07 Volume No. 02STUDY
Page No: 3 of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
5 Description of Methods
5.1 Interaction of devices with Whole Blood (WB):
Interactionof materialswith WB was done underagitation to analyzethe
effect of device on haematologyparameters.
5.1.a Exposure of devices to WB
Blood from human volunteer was collected into the anticoagulant, CPO.
Replicates of test devices were placed in separate polystyrene vials and were
immersed in phosphate buffered saline for 5 min before they were exposed to
blood. To each vial that contains the test devices, blood was added and an initial
sample was collected immediately for analysis. The remaining blood was
exposed to the respective device for 30 min under agitation at 75 :t 5 rpm using
an Environ shaker thermo stated at 35 :t 2°C (WPTRU012). The blood sample
and the device were then analyzed. Six numbers of empty polystyrene vials
were exposed with blood as reference.
5.1.b. Consumption of platelets and leukocytes (WBC) by cell counts
The count reduction in blood was estimated by detecting the counts in
initial and 30 min samples collected from step 5.1.a., using Haematology
Analyzer Sysmex K 4500 as per WPTRU015. The equipment calibration was
verified using traceable control.
5.1.c. Plasma coagulation (Fibrinogen and Partial Thromboplastin timeassay)
The initial and 30 min samples collected from step 5.l.a, were centrifuged
at 4000 rpm for 15 min as per WPTRU006 and platelet poor plasma (PPP) was
aspirated. Fibrinogen was measured in each PPP sample by using a reagent kit
obtained from Oiagnostica Stago (France) on Start 4, coagulation analyzer by
method WPTRU026. Partial thromboplastin time in each PPP sample was
detected using a reagent kit obtained from Oiagnostica Stago (France) oh Start
4, coagulation analyzer as per WPTRU023.
Signature ofStudy Director d~A~~
BMT WINGSCTIMST
ReportNo: SRTRU003.07 Volume No. 02STUDY
Page No: 4 of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
5.1.d. Percentage Hemolysis
The total hemoglobin in the initial samples was measured using automatic
haematology analyzer (Sysmex K 4500) as per WPTRU015. The free
hemoglobin liberated in to plasma (separated as in step 5.l.c) after 30 min
exposure was measured in each sample using Diode Array Spectrophotometer
as per WPTRU022 and the percentage hemolysis was calculated using the
formula (Free Hb/Total Hb) x 100.
5.1.e. Qualitative Analysis of Leukocyte Adhesion
The devices exposed to blood were rinsed immediately with PBS to
remove blood completely, fixed with 1% glutaraldehyde for 1h and the cells
attached to the devices were identified using Geimsa stain and were analyzed as
per work procedure WPTRU028. The devices were cut open and the luminal side
was analyzed under light microscope (Leica IMR) for presence of leukocytes.
Representative photomigraphs are taken and documented.
5.2 Interaction of Device with Platelet Rich Plasma (PRP)
Interaction of device with platelet rich plasma was done under dynamic
condition to understand the effect of device on platelets and deposition of
platelets to the device.
5.2.a. Exposure of Devices to PRP
Blood from human volunteer was collected into the anti coagulant; CPO.
Blood was centrifuged at 2500 rpm for 5 min as per the method described in
WPTRU005. The platelet rich plasma (PRP) was collected. The test devices that
were expanded into the lumen of silicone tube (done by sponsor) was connected
and perfused with phosphate buffered saline for 5 min. After draining the PBS,
PRP was perfused as per WPTRU010 at a flow rate of 50 ml min-1 and within 1
min an initial sample was collected for analysis. After perfusion for 30 min, the
PRP samples were collected for analysis. Empty silicone tubes of the same
dimension without devices were perfused with PRP as reference. Devices were
Signature ofStudy Director ~ BMT WING
SCTIMST
ReportNo: SRTRU003.07 Volume No. 02STUDY
Page No: 5 of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
rinsed thoroughly and were processed for analysis using scanning electron
microscope.
5.2.b. Platelet Count ReductionThe count reduction was estimated by detecting the counts in PRP
samples collected from step 5.2.a, using Haematology Analyzer Sysmex K 4500
as per WPTRU015. The equipment calibration was verified using traceable
control.
5.2. c. Platelet Function
The amplitude of platelet aggregation in response to agonists, ADP and collagen
were determined as per the work procedure WPTRU017, in samples from step 5.
2.a.
5.2. d. Platelet SecretionThe PRP samples collected in step 5.2.a were centrifuged and the PPP
was obtained using the procedure WPTRU006. The platelet factor 4 (PF4) was
analyzed with commercially available ELISA kit (Hyphen Biomed, France) as per
the method WPTRU019. The difference in PF4 between the samples obtained at
1 min and 30 min was estimated to detect platelet activation induced by the
device.
5.2.e. Analysis of Platelet Adhesion
The devices exposed to PRP were rinsed with PBS to remove PRP
completely, fixed with 2% glutaraldehyde and dehydrated as per WPTRU030.
The devices were cut open and the luminal side was analyzed for platelet
deposition. Before analysis test devices were critical point dried, gold sputter
coated and were analyzed under ESEM as per WPSEM001.
Deviation: In the original study protocol, use of high voltage-high vacuum SEM
analysis was suggested. However, there was charging and cracking in the case
of the entire polymer coated devices. Therefore, samples were analyzed using
environmental SEM at low voltage.
Signature ofStudy Director ~ BMT WING
SCTIMST
Report No: SRTRU003.07 Volume No. 02STUDY
Page No: 6 of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
5.3. Interaction of Devices with Radio labeled Platelet Rich Plasma (1251-PRP)
Interaction of devices with radio labeled platelets was done to quantify the
number of platelets adhered to the device during the exposure period (Resmi
K.R., Nissey Varghese and Lissy K. Krishnan. (2004) Procedure for quantification
of platelet adhesion to biomaterials by Radioscintigraphy. Thrombosis Research
114; 121-128).
5.3.a. Exposure of devices with radio labeled PRP
Blood from human volunteer was collected into the anticoagulant; CPO.
Blood was centrifuged to obtain platelet rich plasma (PRP) at 2500 rpm for 5 min
as per the method described in WPTRU005. Platelets from PRP was collected,
washed and labeled with 1-125,as described in work procedure WPTRU008. The
labeled platelets were re-suspended in the same donor's platelet poor plasma
and the count was adjusted to get 2x108 to 2.5x1 08 per ml PRP. The test devices
were exposed to 1-125labeled platelets for 30 min under agitation at 75 :t 5 rpm
usingan Environshaker thermo stated at 35 :t 2°C (WPTRU012).
Deviation: Most of the stent had come out of the silicone tube by the time we
received the samples, probably because they were not tight fit into the lumen,
when the sponsor prepared the samples. Therefore, the exposure of 1-125
platelets by perfusion that was proposed in the Study Plan was not done;
instead, devices were exposed under aqitation.
5.3.b. Radioscintigraphy
The exposed devices were rinsed with PBS to remove the PRP completely
and fixed with glutaraldehyde. The devices were then dried and phosphor
imaged to detect the number of platelets deposited on the device as per the
WPTRU029. Aliquots of radio labeled platelets with known number of platelets
were imaged to make the calibration curve. The intensity of radio images of each
device was used for estimation of platelet adhesion on to it based on the
calibration curve.
Signature ofStudy Director ~ BMT WING
SCTIMST
Report No: SRTRU003.07 Volume No. 02STUDY
Page No: '7 of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
6 Study Result
For all quantitative data percentage difference between the 1 min (initial)
value and 30 min (final) value is determined and average deviation (n=6) of each
type sample is calculated. The results are presented for all types of materials in
table format for each parameter. For percentage hemolyis and quantification of
platelet adhesion by radioscintigraphy the values shown are that was detected
after 30 min exposure.
6.1. Whole Blood Exposure
6.1. a. Table 1 Consumption of Platelets and WBC during whole blood
exposure
Observation: No significant consumption in WBC is observed. Platelet
consumption is significant in the case of all samples, as compared to reference.
However, non specific retention of platelets on the device and tubing may have
contributed to the difference seen.
6.1.b Table 2. Plasma Coagulation
Signature ofStudy Director ~ BMT WING
SCTIMST
ReportNo: SRTRU003.07 Volume No. 02STUDY
Page No: 8 of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
Sample ID Platelet Consumption WBC Consumption
(% Count Reduction) (% Count Reduction)
Co-Cr bare stents 39.54::1:9.33 2.92::1:1.78
Polymer alone 21.28::1:4.60 2.07::1:1.11
Coated
Supracore 44.19::1:14.75 4.66::1:2.62
Reference 1.84::1:1.25 1.43::1:0.60
Sample ID Plasma PTT
Fibrinogen (% Delay in Clotting time)
(% Reduction)
Co-Cr bare stents 12.28::1:9.0 14.61::1:5.89
Polymer alone 14.27::1:3.05 9.03::1:3.93
Coated
Observation: There seems to be a reduction in clottable fibrinogen after exposure
of blood for 30 min, more significantly with polymer alone coated devices.
Intrinsic coagulation seems to be affected by all types of devices because PTT is
prolonged significantly as compared to reference.
6.1.c Table 3 Percentage Hemolysis
Observation: All samples showed significant hemolysis as compared to
reference.
6.1.d Fig1. Qualitative Analysis of Leukocyte Adhesion
The representative photomicrographs are given in the figure 1. The
leukocyte adhesion is highest on bare stents. On polymer coated devices, very
few leukocytes were found. On drug loaded stents, a staining pattern was
observed which is not specific for leukocytes. A uniform patterning is seen which
may be due to retention of stain to the porous structure generated during
polymer coating for drug loading, which needs to be confirmed.
Signature ofStudy Director ~ BMT WING
SCTIMST
Report No: SRTRU003.07 Volume No. 02STUDY Page No: 9 of 15 Date effective: 17 Feb 2007
REPORT ControlStatus:
Evaluation of Coronary Stent for in vitro hemo compatibility
Supracore 8.99:t14.75 10.44:t2. 98
Reference 2.43:t3.14 - 0.04:t0.04
Sample ID Hemolysis (% after
exposure)
Co-Cr bare stents 0.30:t0.04
Polymer alone Coated 0.26:t0.05
Supracore 0.19:t0.05
Reference 0.07:t0
Bare
Supra core
Polymer
Fig.1 Light microscopic images of leukocyte adhesion: Representative Fields
shown are of different types of stents after exposure to whole blood for 30min.
The stained cells are leukocytes adhered to the stent. The uniform pattern
observed on drug loaded stents are not specific for leukocytes, and may be due
to retention of stain in the porous structure generated during polymer coating for
drug loading.
Signature ofStudy Director ~ BMT WING
SCTIMST
Report No: SRTRU003.07 Volume No. 02STUDY
Page No: 10 of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
6.2 PRP Exposure
6.2.a Table 4 Platelet Consumption during PRP exposure
Observation: Platelet consumption during exposure of devices to PRP seems to
be low and is comparable, but slightly elevated in the case of Supracore as
compared to reference.
6.2.b Table 5 Platelet Function
Observation: In the case of ADP induced aggregation, devices seem to have
affected the aggregatory response; however, the S.D. is high. Exposure of PRP
to both bare and Supracore devices have affected the collagen-induced
aggregation of exposed PRP.
Signature ofStudy Director ~ BMT WING
SCTIMST
ReportNo: SRTRU003.07 Volume No. 02STUDY
Page No: 11of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
Sample 10 Platelet Consumption
(% Count Reduction)
Co-Cr bare stents 5.68:t2.76
Polymer alone Coated 5.95:t1.45
Supracore 7.52:t4.85
Reference 3.32:t2.05
Sample 10 % Change in Aggregation % Change in
in response to ADP Aggregation in
response to
Collagen
Co-Cr bare stents 15.48:t7.29 11 .51 :t7.08
Polymer alone
Coated 10.11:t12.71 5.26:t4.34
Supracore 10.62:t8.66 11.23:t7.94
Reference 5.52:t3.86 6.72:t2.85
6.2.c Table 6 Platelet secretion
Observation: Except after exposure to polymer alone coated stent, in other
samples platelet granule secretion is high as compared to reference.
6.2.d Fig. 2 Analysis of Platelet Adhesion
The SEM images of the devices are given in the figure 2. Platelets
adhesion is highest on bare stents. Aggregates of platelets were found especially
near the bends. On polymer coated devices, the platelet adhesion was found be
very minimal. A representative lower magnification picture (500x) and High
magnification picture (3000x) are given side by side for each type of devices.
Observation: On Bare Devices, the black colored spots on Figure A are platelets.
In the higher magnification image (B) a large aggregate of platelet and lots of
small aggregates are seen (black structures). On Polymer alone coated devices,
platelets are rarely seen. Some single platelets are seen in white color. The
difference in color is because, the image was captured in backward scatter mode
to improve the contrast between the device surface and adhered surface. Since
polymer coating was grayish, the platelets appear as white. In the higher
magnification image (D), the polymer was seen to be peeling off.
Signature ofStudy Director cl~ BMT WING
SCTIMST
ReportNo: SRTRU003.07 Volume No. 02STUDY
Page No: 12of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
Sample ID % Increase in PF4
Release on Exposure
(IUlml)
Co-Cr bare stents 26.72::1:4.53
Polymer alone Coated 7.72::1:2.49
Supracore 39.51::1:22.70
Reference 8.86::1:2.53
Bare
Polymer alone
Supracore
Fig.2 Electron micrograph of Bare, Polymer coated and drug loaded devicesexposed to platelet rich plasma. A, C and E 500x ; B, D and F 3K. Scale baris shown in each figure.
Signature ofStudy Director ~ BMT WING
SCTIMST
Report No: SRTRU003.07 Volume No. 02STUDY
Page No: 13of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
On Drug loaded devices also, the platelets appear as white spots. Platelet
adhesion is less on these devices, and in high magnification, a uniform pattern
due to coating with polymer was seen (E; F). In Figure F, platelet adhesion was
negligible, but the polymer was seen peeling off. These breaks might have
occurred during the High voltage SEM analysis.
6.2.e Quantification of platelet adhesion
Standard curve generated using log platelet count against respective
radioactivity on each aliquot is shown in Fig 3. The number of platelets deposited
on each material is given in the Table. Out of the total number of platelets in 1251-
PRP to which the materials were exposed, the percentage of cells adhered to the
material is calculatedand given in the Table 7.
.//
#
A/
Signature ofStudy Director 0~'/'~ ~
BMT WINGSCTIMST
ReportNo: SRTRU003.07 Volume No. 02STUDY Page No: 14of 15 Date effective: 17 Feb 2007
REPORT ControlStatus:
Evaluation of Coronary Stent for in vitro hemo compatibility
Fig.3. Standard Curve used for estimating adhered platelet number. Curve
showed a correlation coefficient 0.9992. Red marks are of known standard
aliquots and green are of tests (Platelets adhered to materials)
Table 7 . Number of platelets adhered on devices
Observation: Platelet deposition is highest on bare stents and minimum on
polymer alone coated stents.
7 Archive
No Archive
8 Documentation
8.1 All recordsgeneratedduringthe study
9 Authorized signatures
Signature ofStudy Director ~ BMT WING
SCTIMST
ReportNo: SRTRU003.07 Volume No. 02STUDY
Page No: 15of 15 Date effective: 17 Feb 2007REPORT Control
Status:
Evaluation of Coronary Stent for in vitro hemo compatibility
Sample ID Number of Percentage
Platelets on each adhered w.r.t total
devices platelets in PRP
Co-Cr bare stents 5415:f:1133 0.0012:f:0.0003
Polymer alone
Coated 1315:f:538 0.0003:f:0.0001
Supracore 2075:f:578 0.0005:f:0.0001
Authority I Name I Designation I Signature I Date
Study \l FdJ.Director Lissy K.Krishnan ScientistF '"
Technical/oV/oManager G.S. Bhuvaneshwar Head, BMT WIng