Normal Range:

• Axial lengthBiggest source of error in IOL power calculations.

• 96% of axial lengths fall within the range 21.0 to 25.5 mm

• 60% of AL is between 22.5 and 24.5 mm

Page 3: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

• Ultrasounds are calibrated with average velocities for normal length eyes. These velocities are incorrect for short eyes, causing significant measurement errors.

• Accurate measurement of axial length in hyperopic eyes is especially important since any error is greatly magnified in proportion to the length of the eye.

• Immersion or LASER biometry can provide superior results in these cases.

Page 4: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

• In a hyperopic eye with axial length of 20 millimeters each mm error in measurement cause 3.75 D post operative refractive error.

Page 5: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences
Page 6: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

• Difference in axial length of more than 0.3mm requires confirmation

• Most individuals have similar axial lengths in each eye except:

– unilateral refractive error

– coloboma

– Staphyloma

Page 7: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences
Page 8: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences
Page 9: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

MYOPIA

Page 10: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

• In the axial myopia, the presence of a staphyloma should be considered, especially if there is difficulty obtaining a distinct retinal spike during A-scan.

• an immersion A/Bscan approach to axial length measurement in the setting of a posterior staphyloma

Page 11: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

• nearly all eyes with pathologic myopia have some form of posterior staphyloma.

• For highly myopic eyes (axial > 28mm) a B-scan should be carried out to determine the presence or otherwise of staphylomata

Page 12: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

Long eye, foveal deep

Page 13: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

Very long eye, posterior pole staphyloma

Page 14: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

Buphthalmic globe: very long eye (37·5 mm): deep anterior chamber.

Page 15: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

Megalocornea: average length eye (23·1 mm): deep anterior chamber (5·2 mm).

Page 16: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

• The Ascan vector is adjusted to pass through the middle of the cornea as well as the middle of the anterior and posterior lens echoes to assure that the vector will intersect the retina in the region of the fovea.

• IOL Master permits accurate measurements when posterior staphylomata are present. Since the patient fixates along the direction of the measuring beam.

Page 17: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences
Page 18: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

IOL POWER CALCULATION FORMULA

Page 19: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

Calculation formula:

• The SRK/T and Holladay formulae worked best overall.

• SRK I and II were found to be least accurate

• No statistical difference was found between SRK/T, Hoffer Q and the Holladay formulae

Page 20: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

Axial Length Formula

<22 mm Hoffer Q or SRK/T

22 – 24.5 mm SRK/T or Holloday, Haigis

> 24.6 SRK/T

Page 21: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences
Page 22: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

When using IOL Master do not forget

• For AL < 22 do immersion A scan

• For AL > 30 do A and B scan

Page 23: Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

Silicone oil filled vitreous cavity: eye measures 49 mm on B-scan, but actual axial length is 34·3 mm.