calibration for lhaaso_wfcta yong zhang, ll ma on behalf of the lhaaso collaboration 32 nd...
TRANSCRIPT
Calibration for LHAASO_WFCTA
Yong Zhang, LL Ma on behalf of the LHAASO collaboration
32nd International Cosmic Ray Conference, Beijing 2011
Large High Altitude Air Shower Observatory Wide Field of view Cherenkov Telescope Array —LHAASO_WFCTA
outline• Introduction• Calibration
• Photometric calibration– Using Hybrid Photo Diode (HPD)
– Using Nitrogen Laser
• Weather calibration– Using Nitrogen Laser
– Using Infrared detector
– Using Star light
• summary
Introduction
Light collector: 5m2 spherical mirrors with reflectivity 82%Camera: 16 × 1 6 PMTsPixel size: 1 °× 1 °FOV: 14 °× 1 6°Electronics: DC coupling, FADC 10bits 50M HzPhysics Goal: to study the energy spectrum & compositions of cosmic rays(1013—1015eV) .
Prototype of Cherenkov telescope
Prototypes @ YBJ
ARGO-YBJ HALL
• Light source(1) calibration: using calibrated Hybrid Photo Diode (HPD) to measure light flux from UVLED(355nm): IHires = #photons/mm2
• WFCT_Probe : two PMTs (XP3062), measuring the flux from the same source. CHires=k*IHires (k=QE*G*APMT) • Light source(2) calibration: IYBJ =CYBJ/Chires*IHires
• absolute gain: G= CFADC/(IYBJ*APMT) (FADC count/pe)• CR measurement: in observations , #photons=CCR/G
Number of photons is then measured.
Photometric calibration(1) ——Using HPD
PC
Inverse-polarity
Amplifier
PulseGenerator
UV LEDtrigger
HPD
WFCTA_Probe
UV light355nm
This work is done at Hires lab
WFCTA_Probe
UV LED
Mirror
WFCTA clustertrigger
This work is done at YBJ
Calibration result
Resolution: -- HPD: 4.8 -- CRTNT Probe: 5% => ILED : 6.9%
Photometric calibration(1) ——Using HPD
Calibration results of the two prototypes
The laser calibration system (shown in figure 1) includes: 1 、 Nitrogen laser : parameters are shown in Table 1.2 、 theodolite : Resolution is 0.26 second of arc3 、 Pyroelectric energy meter+radiomter : Calibration Accuracy is ± 3%4 、 Sky windows: 1m×1m5 、 Up/down flat : controlled by motorThis laser calibration system is built in a container and is able to controlled remotely by login a local PC104.
Up/down flat
N2 laser
Figure 1 : The mechanical structure of laser calibration system
Theodolite
Sky window
Container
Photometric calibration(2) ——Using Nitrogen laser
feature parameters
WavelengthSpectral bandwidthPulse width (FWHM)Pulse energyEnergy stability Peak powerAverage powerBeam size Beam divergence (full angle)Repetition rate
337.1nm0.1nm<3.5 ns170 μJ3% std. dev. (at 10 Hz)45kW3mW (at 20 Hz)3 .7mm5 . 8 mrad1 to 20 Hz
Table 1: Parameters of nitrogen laser
2.52km Laser Detector
θ1 θ2
TM1 TM2
SM
Figure 2: Geometry of laser calibration system
● This system had been installed at ARGO-YBJ site from March 2011.
● This system is located 2.52km apart from two telescopes station.
● The light received by the telescope is proportional to the energy of the laser pulse
● The absolute laser energy can be measured accurately by Pyroelectric energy
meter.
Photometric calibration(2) ——Using Nitrogen laser
Figure 3: Image of laser track with 65◦ in elevation
We tested this laser calibration system on April 2 and8, 2011. Figure 3 shows the example image of laser track.
Photometric calibration(2) ——Using Nitrogen laser
LaserDetector
θ1θ2
TM1TA1
TM2TA2
SMSA
Figure 4: Geometry of laser calibration system
● This system is located 176m and 71m apart from the two prototypes
of Cherenkov telescope respectively.
● Backscattering light by molecules and by aerosols is received.
●We will measure the daily variation of atmosphere using this system
from next observation season.
Weather calibration(1) ——Using Nitrogen laser
• Monitor clouds. • scan the whole sky once/15min
Figure 5: The infrared temperature of the whole sky
Figure 6: The distributions of the infrared temperature
Weather calibration(2) ——Using Infrared detector
Cloudy condition
Good weather condition
• The telescope can observe the night
sky background(NSB).
• A clear correlation between the star
light and the FADC counts recorded
by the telescope can be seen clearly
• The correlation is disappeared under
the bad weather condition.
Advantages:• The flux of star is very stable • Almost have the same path
with Cherenkov photons
Weather calibration(3) ——Using Star light
NSB measured by one PMT in one night
NSB measured by all PMTs of one cluster in one night
steps of the weather selection
• 1: on hourly scale: – A linear fit between the flues of
the star light and the FADC counts is done. If the differences between the FADC counts and the fitted value are larger than 4RMS, the points are subtracted as bad weather conditions
• 2: on the whole night scale:– The selection is based on the
correlation coefficient between the FADC counts and the fluxes of the star light.
The distribution of the correlation coefficient of the 39 days of Nov. and Dec. 2009.
Good weather condition
Summary
• photometric calibration using HPD had been done,
Resolution is 7%.
• The laser calibration system had been installed at
ARGO-YBJ site from March 2011. This system will be
operated from next observation season
• 133 nights are calibrated using stars light. 99 nights is
good weather, the value of correlation coefficient are
larger than 0.8