calibration fundamentals
DESCRIPTION
Calibration FundamentalsTRANSCRIPT
Biometrix Corporation (800)-890-8909 1
Calibration Fundamentals
Biometrix Corporation (800)-890-8909 2
Objectives
Understand why we calibrate Understand 21 CFR Part 820 Introduce the basis for the basic SI units Explain the standards used in the calibration
chain. Explain uncertainty and the 4:1 ratio.
Biometrix Corporation (800)-890-8909 3
Metrology
What is Metrology? The science that deals with measurement
What does it mean to calibrate? To check, adjust, or determine display accuracy by
comparison with a standard (the graduations of a quantitative measuring instrument)
To make corrections in; adjust a set of operations that establish, under specified
conditions, the relationship between the values of quantities indicated by a measuring instrument or measuring system … and the corresponding values realized by standards
Biometrix Corporation (800)-890-8909 4
Metrology
Why Calibrate? The quality of drug product is dependent upon
certain conditions of manufacture, including temperature, pressure, and pH. In order to ensure consistent and correct drug manufacture, instrumentation relating to the drug production must be calibrated.
In order to accurately describe the processes by which drugs are created, and gauge the effect of any deviations on the drug process
Biometrix Corporation (800)-890-8909 5
Metrology
..and because the government requires it. 21 CFR Part 820.72 governs Inspection,
Measuring and Test Equipment.
Biometrix Corporation (800)-890-8909 6
21 CFR Part 820 (a) Control of inspection, measuring, and test equipment. Each
manufacturer shall ensure that all inspection, measuring, and test equipment, including mechanical, automated, or electronic inspection and test equipment, is suitable for its intended purposes and is capable of producing valid results. Each manufacturer shall establish and maintain procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained. The procedures shall include provisions for handling, preservation, and storage of equipment, so that its accuracy and fitness for use are maintained. These activities shall be documented.
Biometrix Corporation (800)-890-8909 7
21 CFR Part 820
Calibration. Calibration procedures shall include specific directions and limits for accuracy and precision. When accuracy and precision limits are not met, there shall be provisions for remedial action to reestablish the limits and to evaluate whether there was any adverse effect on the device`s quality. These activities shall be documented.
Biometrix Corporation (800)-890-8909 8
21 CFR Part 820
Calibration standards. Calibration standards used for inspection, measuring, and test equipment shall be traceable to national or international standards. If national or international standards are not practical or available, the manufacturer shall use an independent reproducible standard. If no applicable standard exists, the manufacturer shall establish and maintain an in-house standard.
Biometrix Corporation (800)-890-8909 9
21 CFR Part 820
Calibration records. The equipment identification, calibration dates, the individual performing each calibration, and the next calibration date shall be documented. These records shall be displayed on or near each piece of equipment or shall be readily available to the personnel using such equipment and to the individuals responsible for calibrating the equipment.
Biometrix Corporation (800)-890-8909 10
N.I.S.T. As a non-regulatory agency of the United States Department of
Commerce’s Technology Administration, the National Institute of Standards (NIST) develops and promotes measurement, standards, and technology to enhance productivity, facilitate trade, and improve the quality of life. As part of this mission, NIST scientists and engineers continually refine the science of measurement, making possible the ultraprecise engineering and manufacturing required for today’s most advanced technologies. They also are directly involved in standards development and testing done by the private sector and government agencies.
NIST was originally called the National Bureau of Standards (NBS), a name that it had from 1901 until 1988
Biometrix Corporation (800)-890-8909 11
Definitions
Unit (of measurement) Particular quantity, defined and adopted by
convention, with which other quantities of the same kind are compared in order to express their magnitudes relative to that quantity.
Value (of a quantity) Magnitude of a particular quantity generally
expressed as a unit of measurement multiplied by a number.
Biometrix Corporation (800)-890-8909 12
Definitions
MeasurementSet of operations having the object of determining a value of a quantity.
Error (of measurement) Result of a measurement minus a true value of
the measured.
Biometrix Corporation (800)-890-8909 13
Definitions
Accuracy Conformity to fact. Precision; exactness. The ability of a measurement to match the
actual value of the quantity being measured
Biometrix Corporation (800)-890-8909 14
Definitions
Traceability Property of the result of a measurement or the
value of a standard whereby it can be related to stated references, usually national or international standards, through an unbroken chain of comparisons all having stated uncertainties.
Includes conductivity solutions, pH solutions, viscosity solutions, and turbidity solutions.
Biometrix Corporation (800)-890-8909 15
Definitions
Standard Reference material (SRM) Reference material, accompanied by a certificate,
one or more of whose property values are certified by a procedure which establishes traceability to an accurate realization of the unit in which the property values are expressed, and for which each certified value is accompanied by an uncertainty at a stated level of confidence.
Biometrix Corporation (800)-890-8909 16
Definitions
Reference Material (RM) Material or substance one or more of whose property
values are sufficiently homogeneous and well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to materials.
A reference material may be in the form of a pure or mixed gas, liquid or solid. Examples are water for the calibration of viscometers, sapphire as a heat-capacity calibrant in calorimetry, and solutions used for calibration in chemical analyses.
Biometrix Corporation (800)-890-8909 17
Definitions
Working Standard Standard that is used routinely to calibrate or
check material measures, measuring instruments or reference materials.
A working standard is usually calibrated against a reference standard.
Biometrix Corporation (800)-890-8909 18
Definitions
Reference Standard Standard, generally having the highest
metrological quality at a given location or in a given organization, from which measurements made there are derived.
Biometrix Corporation (800)-890-8909 19
Definitions
Primary Standard Standard that is designated or widely
acknowledged as having the highest metrological qualities and whose value is accepted without reference to other standards of the same quantity.
Biometrix Corporation (800)-890-8909 20
Definitions
Resolution A measure of the smallest portion of the signal
that can be observed. For example, a thermometer with a display that reads to three decimal places would have a resolution of 0.001ºC. In general, the resolution of an instrument has a better rating than its accuracy.
Biometrix Corporation (800)-890-8909 21
Definitions
Repeatability closeness of the agreement between the results
of successive measurements of the measurand (eg mass value) carried out under the same conditions of measurement. Conditions include: same procedure, observer, instrument, conditions, location; and carried out over a short period of time.
Biometrix Corporation (800)-890-8909 22
Definitions
Reproducibility closeness of the agreement between the results
of measurements of the measurand (eg mass value) carried out under changed conditions of measurement. Includes changing some of those conditions which are held constant for ‘repeatability’, and may refer to measurements carried out over a long period of time.
Biometrix Corporation (800)-890-8909 23
Definitions
Uncertainty parameter, associated with the result of a
measurement, that characterizes the dispersion of values that could reasonably be attributed to the measurand
Biometrix Corporation (800)-890-8909 24
Definitions
Standard Material measure, measuring instrument, reference
material or measuring system intended to define, realize, conserve or reproduce a unit or one or more values of a quantity to serve as a reference.
Examples kg mass standard 100 Ω standard resistor standard ammeter cesium frequency standard
Biometrix Corporation (800)-890-8909 25
International System of Units (SI) The SI is founded on seven SI base units for seven base
quantities assumed to be mutually independent. Length Mass Time Temperature Quantity of Substance Luminous Intensity Electric Current
Biometrix Corporation (800)-890-8909 26
Length
Meter The meter is the length of the path traveled by
light in vacuum during a time interval of 1/299 792 458 of a second.
Symbol = m It used to be one ten millionth of the distance of
the meridian through Paris.
Biometrix Corporation (800)-890-8909 27
Mass
Kilogram The kilogram is the unit of mass; it is equal to the
mass of the international prototype of the kilogram.
The international prototype is a large chunk of platinum-iridium in Paris.
The international prototype will soon be a sphere of pure Silicon with an exactly known number of atoms
Symbol = kg
Biometrix Corporation (800)-890-8909 28
Time
Second The second is the duration of 9 192 631 770
periods of the radiation corresponding to the transition between the two hyperfine levels of the ground state of the cesium 133 atom.
Symbol = s It used to be 1/86400 of the astronomical day.
Biometrix Corporation (800)-890-8909 29
Temperature
Kelvin The Kelvin, unit of thermodynamic temperature, is
the fraction 1/273.16 of the thermodynamic temperature of the triple point of water.
What is the triple point? The temperature and pressure at which solid, liquid, and
gaseous water coexist in equilibrium.
Symbol = K
Biometrix Corporation (800)-890-8909 30
Electric Current
Ampere The ampere is that constant current which, if
maintained in two straight parallel conductors of infinite length, of negligible circular cross-section, and placed 1 meter apart in vacuum, would produce between these conductors a force equal to 2 x 10-7 Newton per meter of length.
Biometrix Corporation (800)-890-8909 31
SI Derived Units Examples of SI derived
Area square meter m2 Volume cubic meter m3 Speed velocity meter per second m/s Acceleration meter per second squared m/s2 Wave number reciprocal meter m-1 Mass density kilogram per cubic meter kg/m3 Specific volume cubic meter per kilogram m3/kg Current density ampere per square meter A/m2 Magnetic field strength ampere per meter A/m Concentration mole per cubic meter mol/m3 Luminance candela per square meter cd/m2 Mass fraction kilogram per kilogram kg/km
Biometrix Corporation (800)-890-8909 32
CGMP Calibrations
CGMP Calibration must be traceable through all vendors and paperwork to the original NIST calibration Example: NIST calibrates primary standard for primary
standards lab Primary Standards Lab calibrates Intelligent RTD for
Biotech Manufacturing Facility Biotech Manufacturing Facility calibrates HART
temperature blocks with IRTD Calibration Group calibrates bioreactor instrumentation
using the HART temperature blocks.
Biometrix Corporation (800)-890-8909 33
4:1 Ratio
When using one instrument or standard to calibrate another, the ratio of accuracies should be maintained at no lower than a 4:1 ratio.
Example Using an IRTD with an accuracy of +/- 0.01 K to
calibrate a HART Block with an accuracy of +/- 0.05 K.
Biometrix Corporation (800)-890-8909 34
Uncertainty To calculate the uncertainty of the overall test system, the
uncertainties of the individual components are combined using a statistical approach.
Each component is represented by a standard deviation (standard uncertainty).
The deviations are combined under the rule of Propagation of Uncertainty.
This rule is mathematically expressed as the root sum squared. The square root of the sum of the squares.
When combined they are known as the
Combined Standard Uncertainty
Biometrix Corporation (800)-890-8909 35
Combined Standard Uncertainty Example
A pressure measurement system contains a pressure transducer, ma transmitter, and PLC Analog Input.
The uncertainty of the pressure transducer is +/- 0.5% The uncertainty of the ma transmitter is +/-0.2% The uncertainty of the PLC Analog Input is +/- 0.1% sqroot((0.5%)2 + (0.2%)2 + (0.1%)2 ) Combined Standard Uncertainty = 0.547%
Biometrix Corporation (800)-890-8909 36
Expanded Uncertainty
The expanded uncertainty, suggested symbol U, is obtained by multiplying the combined standard uncertainty by a coverage factor, suggested symbol k, which typically has a value between 2 and 3 (i.e., U = kuc)
For a normal distribution and k = 2 or 3, the expanded uncertainty defines an interval having a level of confidence of 95.45% or 99.73%, respectively.
Biometrix Corporation (800)-890-8909 37
Expanded Uncertainty
If we apply the 4:1 ratio on top of the expanded uncertainty, the maximum specification for accuracy with a 99.73% confidence level would be.. 0.547% Combined Standard Uncertainty K Factor of 3 for 99.73& confidence 4:1 ratio 4 x 3 x 0.547% = +/- 6.564 % Luckily most pressure transducers are
significantly more accurate.