captair 321 smart & midcap · captair smart 321 erlab’s innovative technology keeps you safe...
TRANSCRIPT
SaferSimpler
Captair 321 Smart & MidcapDuctless filtering fume hoods
Productdatasheet
Safer to operate- ExclusiveErlabfiltrationtechnologycombiningactivatedcarbonandHEPA/ULPAtoadapttothemanipulation - MeetsAFNORNFX15211/ANSIZ9.5-2012filtrationefficiencystandard(class1*and2)- Realtimesensorstodetectmainfiltersaturationwithsolvents,acidsorformaldehyde*- Safetyfilterincaseofmainfiltersaturation - Airfacevelocitypermanentmonitoring - ErlabSafetyProgram:applicationanalysisandvalidation,usageframeworkcertification,usagefollow-up- Connecteddeviceallowingreceptionofsafetynotificationsandusestatus*
Simpler to useRealtimestatuscommunicationbylightandsoundpulses*:-Airfacevelocitydecrease- Mainfiltersaturation- Fanfailure- Excessscheduledworkingtime
Flexibility- Modularfiltrationcolumnadaptingtoapplicationchanges- Easyandfastrelocation
Savings- Noductworkcost- Annualenergycost<100€- Comparedtoanextractionfumecupboard,energysavings compensatefilterreplacementcost
Environment- Nochemicalreleaseintotheatmosphere- Lowenergyconsumption
*:NotavailableonCaptair321Midcap
Powerful light guided
communication
Ask for the highest level of
filtration performances
Connect your equipment and benefit from the
remote access thanks to our mobile solutions
Dowload our eGuard application
694 mm 694 mm
510 mm 510 mm
27 mm
470 mm
700 mm
Captair 321 Smart & MidcapDuctless Filtering Fume Hoods
Heightsaccordingtothefiltrationcolumnconfiguration
Type1Cor1P 1110mm
Please add 150mm between the last filter and the ceiling to allow a good air recirculationandtoreplacefilterseasilyType2Cor1P1Cor1C1P* 1205mm
Type1P2Cor1P1C1P* 1285mm
Tempered glass Trespa® Top LabPLUS 304 L stainless steel
Work surfaces with built in spill tray
Benchcap: fixed work bench Mobicap : mobile rolling cart
800 mm
715mm
780 mm
365mm180 mm
160 mm295 mm
245 mm
1110 mm min
1285 mm max425 mm
620 mm
74 mm
835 mm (without work surface)
822 mm min (with work surface)
FILTRATION TECHNOLOGY
ABOVE
2020 mm min 2195 mm max
2000 mm min2175 mm max
FILTRATION TECHNOLOGY
ABOVE
FILTRATION TECHNOLOGY
ABOVE
2020 mm min 2195 mm max
2000 mm min2175 mm max
FILTRATION TECHNOLOGY
ABOVE
FILTRATION TECHNOLOGY
ABOVE
321Captair SmartHotte à filtration sans raccordementErlab’s innovative technology keeps you safe
Communicationinterface*
Simple communication by audible and light pulses: unit running time, air face velocity, automatic filter saturation detection, ventilation settings, fan failure alarm
Filtration technology 1 column that can be configured to handle liquids, powders, or both
Carbon filtration for gases and vapours Following filtration column configuration (see table above)
Particulate filtration for powders Following filtration column configuration (see table above)
eGuard* APP for remote control to monitor the hood, change the settings, and deliver safety alerts immediately to your devices (mobile, tablet and PC)
Internal lighting* LED lighting > 650 Lux
Anemometer* Air face velocity alarm
Anemometer** Air face velocity indicator
Chemical Listing List of approved chemicals
Safety Standards AFNOR NF X 15-211:2009: France - BS 7989: EnglandDIN 12 927:Germany - EN 1822:1998 (HEPA H14 & ULPA U17 Filters) - CE Marking
Air Flow 220 m3/h - 135 CFM
Air Face Velocity 0.4 to 0.6 m/s - 79 fpm to 118 fpm
Voltage/Fequency 220 V / 50-60 Hz
Power consumption 65 W
Sash openings Oblong
Structure Corrosion resistant electro-galvanized steel coated with anti-acid polymer
Side and front panels Chemical resistant acrylic
Filtration module Polypropylene
Equipment
Work Surfaces Trespa® Top Labplus, Glass or 304L Stainless Steel
Molecode* Detection sensor for : Type S, for solvents / Type A, for acids / Type F, for formaldehyde
Benches Mobile (Mobicap) or fixed (Benchcap)
Bench equipment Technical gases outlets, water outlets, front control valves, sink, power sockets (Only compatible with Trespa® Top Labplus worktop and fixed bench)
Particulate Pre-filter Protects the main filter(s) from dustTransparentBack Panel Clear acrylic panel for easy viewing
Accessories
Captair 321 Smart & MidcapDuctless Filtering Fume Hoods
FILTRATION TECHNOLOGY
Modular design of the filtration column allows toadapt to every protectionneeds
Carbon filtration for gases and vapoursAS:For organic vapoursBE+:Polyvalent for acid + organic vapoursF:For formaldehyde vapoursK:For ammonia vapours
Particulate filtration for powdersHEPA H14:99.995 % efficiency filtration of particles over 0.1 µm in sizeULPA U17:99.999995 % efficiency filtration of particles over 0.1 µm in size
Ventilation
MolecodeAutomatic alarm to detect filter saturation
Products handled / Applications
Liquid chemicals handlings Powders handlings Liquid chemicals and powders handlings
Liquid chemicals handlings in clean room
Cus
tom
ized
filtr
atio
n co
lum
n Class I* according
to theNF X 15-211
NA
Class 2 according
to theNF X 15-211
Maximum
protection Maximum
protection Maximum
protection
P C
2C
1C 1P 1P1C
1P2C 2C1P
1C1P
Available filters :
*:NotavailableonCaptairMidcap/**:NotavailableonCaptairSmart
About Erlab
France +33 (0) 2 32 09 55 80 │ [email protected]
Germany 0800 330 47 31│ [email protected]
China +86 (0) 512 5781 4085 │ [email protected]
Italy +39 (0) 2 89 00 771 │ [email protected]
Malaysia +60 (0) 7 3 555 724 │ [email protected]
Spain +34 93 673 24 74 │ [email protected]
USA+1 800-964-4434 │ [email protected]
United-Kingdom +44 (0) 1722 341 940 │[email protected]
www.erlab.com
1 R&D laboratory AworldwideuniquelabasawarranttoErlabexpertiseinfiltration.Erlabinnovatesconstantlytodesignhighestlevelof
chemicalprotectiontechnologies.
2 Safety standards OurproductsperformancesarecertifiedbycompliancewiththeAFNORNFX15211standard.
3 Independent tests Ourfiltration technologyhasbeen testedmultiple timesbyan independent laboratoryaccording toAFNORNFX15
211requirements.Resultsprovefiltersefficiencytomaintainconcentrationinrecirculatedairatlowerthan1%oftestedchemicalsTLV.
4 Chemical listing Ouruniqueknow-howenableustocommunicateonourfiltersretentioncapacitiesover700chemicals,followingAFNOR
NFX15211requirements.
5 Erlab Safety Program Oursafetyprogramisthewarrantofyourprotectionatallstages:anyprojectbeginswithanapplicationanalysistoallow
ourlabtorecommendthesuitablesolution.Allinstalledequipementsarethenfollowed-uptoguaranteethehighestlevelofprotectionovertime.
6 Our questionnaire Allows our lab specialists to recommend adapted filtration fume hood, filter type and predictive life-time, such as
personnalisedassessment.
7 We make your fume hood uses a commitment Ourlaboratoryprovidesausagecertificateforeachfumehood,incompliancewithAFNORNFX15211requirements.
ErlabOur mission: guarantee your protection in the lab
Erlabinventedthefirstfiltrationfumehoodin1968.45yearsofexpertiseinchemicalsfiltrationenableustoguaranteeyourprotectionthroughoutthefollowingaspects:
Erlab R&D laboratory
Cap
tair
Smar
t V.0
1_10
_201
5 -T
echn
ical
dat
a ar
e no
t con
tract
ual.
Erla
b re
serv
es th
e rig
ht to
mod
ify th
e pr
esen
t doc
umen
t with
out p
rior n
otic
e.