  • 1. ENGLISH

2. 3Congratulations on your purchase of a Kodak dental systems. Thank you for yourconfidence in our products and we will do all in our power to ensure yourcomplete satisfaction.You now have the Installation Guide for the Kodak 8000 / Kodak 8000C digitalpanoramic and cephalometric unit. We recommend that you thoroughlyfamiliarize yourself with this Guide in order to make the most effective use of yoursystem.The information contained in this Manual may be subject to modificationwithout notice, justification or notification to the persons concerned.No part of this Guide may be reproduced without the express permission ofEastman Kodak Company. ENGLISHThe brand names and logos reproduced in this Guide are copyright.RVG RadioVisioGraphie, Digipan, Trophy, Trophy Radiologie and the TR logo are registered trademarks of Eastman Kodak Company. Microsoft, MS-DOS and Windows are trade marks or registeredtrade marks of the Microsoft Corporation. All other names or products referred to in this documentare only used for the purpose of identification and may be the trade marks or registered trade marksof their respective owners.The RVG technology is the subject of an international patent registered by Eastman KodakCompany.This document was originally written in French.Revision date: 10/2004The Kodak 8000 unit complies with Directive 93/42/CEE relating to medical device. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 3. 5Contents NOTES FOR THE USER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 WARNINGS AND SAFETY INSTRUCTIONS . . . . . . . . . . . . . . . . . . 8 RESPONSIBILITY AND WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . 10 SECTION 1: DIMENSIONAL DIAGRAMSENGLISHGeneral arrangement views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Overall height of the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Width and depth of the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Dimensions of the column and attaching parts . . . . . . . . . . . . . . . . . . . . . . 12Minimum/maxium height of the chin rest . . . . . . . . . . . . . . . . . . . . . . . . . . 13Installation drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Standard installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Installation at an angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Installing the cephalometric arm on the right side . . . . . . . . . . . . . . . . . . . 16Installing the cephalometric arm on the left side . . . . . . . . . . . . . . . . . . . . . 17Optional base plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 SECTION 2: PACKAGINGDescription of the components . . . . . . . . . . . . . . . . . . . . . . . . . . 21Box # 1: Head Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Box # 2: Packaging for the Column Assembly . . . . . . . . . . . . . . . . . . . . . . . 24Box # 3 (Optional): Cephalostat unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 SECTION 3: CHECKING THE PREMISES BEFORE INSTALLATIONStandards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Compliance with standards: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Position of the unit in the room: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Minimum width of the door: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Minimum height of the ceiling: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Strength of the wall: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Load-bearing capacity of the floor: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Required electrical specifications: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Electrical power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 SECTION 4: INSTALLATION PROCEDURETools to be supplied by the installer: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Procedure for opening the cartons . . . . . . . . . . . . . . . . . . . . . . . 31Box # 1: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Box # 2: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Box # 3: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 4. 6 Mechanical and electrical installation . . . . . . . . . . . . . . . . . . . . . . 34 Installing the wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Installing the computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Minimum configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Installing the data acquisition board in the computer . . . . . . . . 45 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Step-by-step instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Installing / updating the software . . . . . . . . . . . . . . . . . . . . . . . . . 46 Configuring the installation hardware . . . . . . . . . . . . . . . . . . . . . 51 Post-installation checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Fitting the covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57SECTION 5:TECHNICAL SPECIFICATIONS General technical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . 61 International standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 Conforming to IEC Standards 601.2.7 and 601-1 . . . . . . . . . . . . . . . . . . . 62 Characteristics of the X-ray radiogenic assembly . . . . . . . . . . . . . . 64 Conforming to IEC Standard 601.1.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Conforming to IEC Standard 601.2.28 . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Characteristics of the radiogenic unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65SECTION 6:ANNUAL MAINTENANCE Annual maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 5. Notes for the User 7Notes for the User X-rays are not innocuous and can be dangerous if not used properly. Therefore, the instructions and warnings contained in this Manual must be carefully followed. As a manufacturer of radiology units which conform to the most stringent radiological protection standards in force throughout the world, Eastman Kodak Companys know-how guarantees the maximum degree of protection against radiation hazards. Nonetheless, you are handling a radiology unit which is specially designed to emit a dose of x-rays in order to carry out a medical diagnosis with the aid of a Eastman Kodak Company digital imaging system (exclusive patent). ENGLISH The room in which your radiology unit is to be installed must comply with all official regulations applicable to protection against radiation. Your dealer will be pleased to assist with the initial use of your radiology unit and will supply any relevant information you may require. The "WARNING" and "IONIZING RADIATION" symbols signify: "WARNING: IONIZING RADIATION". To install, operate and service the unit, follow the instructions contained in this Manual. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 6. 8 Warnings and safety instructionsWarnings and safety instructions CAUTION Laser radiation. Do not stare into beam. Class 2 laser product Maximum output power: 1 mW, 650 nm IEC60825-1 :1993+A1 :1997+A2 :2001 This unit emits laser radiation. For maximum safety, advise the patient not to look at the beam. Before illuminating the beams, lower the Frankfurt plane beam to the lowest level. While making adjustments, ensure that the beam is not directed into the patients eyes. DANGER This is an electrical unit. Do not expose it to water spray. Such action may cause an electric shock or a malfunction of the unit. WARNING The user is responsible for the operation and maintenance of this unit. This unit must only be operated by legally qualified persons. The cover of the unit must not be opened by the operator. When necessary, inspection and maintenance operations should only be carried out by an approved Eastman Kodak Company technician. WARNING This unit must be installed in x-ray room which complies with current installation standards. From this location, visual or audio communication must be maintained with the patient, together with access to the control interface during exposure. WARNING The Kodak 8000 unit has been designed to comply with European and international medical standards. In order to ensure conformity, the PCI interface board must be installed in a computer configuration which conforms to IEC 950 Standard. The PC and the peripheral equipment connected to it must not be located in the immediate vicinity of the patient, thus at least 1.5 m from the panoramic unit. For details of the data processing system (PC and screen), refer to the installation manuals for your computer. Sufficient clear space must be left around the CPU in order to ensure that it is properly ventilated. In order to obtain maximum image quality and visual comfort, the screen should be positioned so as to avoid direct light reflections (internal or external lighting). WARNING Do not operate the unit if there is the threat of an earthquake. Following an earthquake, ensure that the unit is operating satisfactorily before using it again. Failure to observe this precaution may expose patients to hazards.SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 7. Warnings and safety instructions9WARNING X-ray equipment can be hazardous to patients and the operator if the exposuresafety factors and operating instructions are not observed.WARNING Do not place objects within the field of operation of the unit.WARNING Ensure that the patient and the operator are wearing aprons to protect themagainst X-rays.Ensure that any parts of the unit which may come into contact with the patient and theENGLISHoperator have been disinfected after each patient has been exposed to X-rays.While adjusting the height of the unit, the operator must ensure that the patient is kept clearof the mechanism.When the unit is not in use, ensure that the ON/OFF switch is set to OFF (O).If the unit develops a fault, switch it off (O) and display a notice Unserviceable.WARNING To dispose of the unit and/or its components, contact our representatives or ourCompany.WARNING Do not rotate the arm unit by hand as this may damage the unit.WARNING The operator must ask the patient to refrain from moving during the entire periodof exposure.WARNING The operator must ask the patient to refrain from moving until the ARM UNIT hasstopped moving and throughout the RESET movement.WARNING Equipment not suitable for use in the presence of a flammable anaestheticmixture with air or with oxygen or nitrous oxide.SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 8. 10 Responsibility and Warranty Responsibility and Warranty Eastman Kodak Company guarantees for a period of two years starting from the invoice date of the product, that it does not contain any defects, in the materials used or at manufacturing level, and this only when used and maintained correctly. Only the damage affecting the products themselves will be taken into account, the client can in no case claim damage compensation if it does not purely concern the products themselves provided that this exclusion of responsibility does not infringe any legal measures. This warranty does not cover damages and faults due to accident, incorrect usage, improper use, negligence or wear and tear due to normal use. Eastman Kodak Company can therefore not be held liable for the consequences resulting from the non-application of the instructions contained in the installation and user manuals, namely bodily harm, profit loss, operational interruptions, data loss, loss of a financial nature or any direct or indirect damage. If, on the other hand, the product is proven to be faulty with regards to materials or manufacturing, the responsibility will lie with Eastman Kodak Company but will be limited to the repair of the aforementioned at its own facilities or, at the discretion of Eastman Kodak Company, of its replacement. Regardless of the solution chosen by Eastman Kodak Company, the client will need to return the faulty product at his own expense. It has clearly been established that except for the repair of the product where it will assume the cost of spare parts and labour, the Eastman Kodak Company will only take on the costs of return to the delivery address initially communicated by the client. Eastman Kodak Company will have the possibility to examine any presumed defect. Clients are not entitled to delay the payment of their bills, nor to any financial deductions by way of the products reputed to be faulty. Due to the continuous development of its products, Eastman Kodak Company reserves the right to amend at any time the manual and products mentioned therein, without justification or having to notify the people involved. Our products are guaranteed ex-works in their original packaging. Eastman Kodak Company cannot be held liable for damages resulting from transportation conditions. Consumable spares, software and accessories are excluded from the warranty. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 9. SECTION 1 Dimensional diagrams 11SECTION 1DIMENSIONAL DIAGRAMSGeneral arrangement viewsCAUTION If the Kodak 8000 unit is planned to be upgraded with a digital cephalometric ENGLISHattachment, please refer to the depth dimension mentioned in italics on the diagrams. Thefixing hole of the screw will indeed be farthest from the wall to allow the mounting of thecephalometric attachment.NOTE The standard version of the unit has an overall height of 2315 mm (91.25 inches).A unit with an overall height of 2200 mm (86.6 inches) is available by special order.Overall height of the unitSM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 10. SECTION 112 Dimentional diagramsWidth and depth of the unitDimensions of the column and attaching parts SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 11. SECTION 1Dimensional diagrams13 Minimum/maximum height of the chin rest ENGLISHInstallation drawing Standard Installation A/ Access to the right of the unitNOTEThese installationdrawings show therecommendedconfigurations toallow optimal accessto the unit. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 12. SECTION 114 Dimentional diagrams B/ Access to the left of the unitInstallation at an angleA/ Access to the right of the unit SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 13. SECTION 1Dimensional diagrams15B/ Access to the left of the unit ENGLISH SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 14. SECTION 116Dimentional diagramsInstalling the cephalometric arm on the right sideWarningThe dimensions shown are the recommended dimensions for optimal installationand easy access to the unit for both maintenance and day-to-day use.These dimensions are valid for an optimized installation, i.e., one in which thearm is being installed in the same side as the access to the Kodak 8000 unit inplace. For example, for a Kodak 8000 unit with right-hand access, we recommendthat the cephalometric arm be installed on the right-hand side. All access to theunit can then be made on the same side, which will save space in the room.If you are installing the arm on a different side from that used to access thepanoramic unit, more space will be required in the room. It will be the installersresponsibility to ensure that the equipment is properly installed.NOTE Provide for adequate space to the rear of the cephalostat sensor to ensure easyaccess for maintenance purposes.NOTE The Kodak 8000C unit cannot be installed in a corner. A/ Overall height of the unit NOTE These installation drawings show the recommended configurations to allow optimal access to the unit.SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 15. SECTION 1 Dimensional diagrams17B/ Width and depth of the unit ENGLISHInstalling the cephalometric arm on the left sideA/ Overall height of the unitSM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 16. SECTION 118 Dimentional diagramsB/ Width and depth of the unit Optional base plateA mounting base is available as an option for installation without drilling into thefloor. This base plate can be used with the Kodak 8000 unit or the Kodak 8000Cunit.For installation of a Kodak 8000 unit, wall mounting the unit is STRONGLYRECOMMENDED in addition to use of the base plate to ensure optimalstability.For installation of a Kodak 8000C unit, wall mounting is REQUIRED, even ifthe unit is mounted on a base plate as well. Dimensions: 900 x 600 Gross weight: 62 kg Net weight: 50 kg Reference:CG351 SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 17. SECTION 1Dimentional diagrams19 Dimensions of the base plate of the Kodak 8000 unit ENGLISH Overall height of the unit including the base plate SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 18. SECTION 120 Dimentional diagrams Dimensions of the column and attaching parts SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 19. SECTION 2Packaging21SECTION 2PACKAGINGStandard packing Box # 1: Head assembly Box # 2: Column assemblyENGLISHDimensions (cm)71 (D) x 123 (L) x 125 (H) 35 (D) x 240 (L) x 47 (H)Weight 120 kg80 kgCephalometric option Box # 3: Cephalostat unitDimensions (cm) 90 (D) x 1640 (L) x 610 (H)Weight80 kgDescription of the components Box # 1: Head Assembly The box contains various items: x Separate components y One box containing the technical accessories z One box of AccessoriesSM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 20. SECTION 222 PackagingSeparate componentsProductCode DescriptionQuantity CG 333 Panoramic head assembly1 HY 273 Upper column cover 1 LV 098 Handswitch 1Mounting hook for HY 1731handswitch CG 329 Connection box 1Extension for the connection CP 9061boxTechnical accessoriesProductCode DescriptionQuantity CJ 648 PCI board1 LO 096 Masking floppy disk1 SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 21. SECTION 2Packaging23LOSoftware 1150KDIS and driversLO 073Test report1Centering screen tool forENGLISHOT 001 1maintenanceBall phantom for technicalOT 011tests and for preventive and 1NOTEA 3-meter mains power corrective maintenancecord with a Europeanplug is also suppliedwith the unit.ZN 013Alen keys2ZN 014 One box of accessoriesProduct CodeDescription QuantityJR 265Panoramic chin rest1(supplied fitted to the unit)JR 263Chin rest for maxillary sinus1sinus maxillairesNasal support for TMJJR 264 1programHY 048Standard bite block6SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 22. SECTION 224 PackagingBite block for edentulous HT 182 2patients FC 020 Bite block support2Plastic hygienic protective BD 1031sleeves (box of 500 items)Box # 2: Packaging for the Column Assembly Product Code Description Quantity CG 298 Column assembly1 CG 332 Wall mounting bracket1 CG 492 Wall-bracket cover 1 SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 23. SECTION 2Packaging25Box # 3 (Optional): Cephalostat unitSeparate components ProductCode Description Quantity /Cephalostat arm1ENGLISH /Cephalostat head 1 HY 306Cover of the secondary arm1 CG 382 + Ear rods2 CG 383Accessories ProductCode Description Quantity CP 278 SUB-D9/25-pin adapter 1Cephalometric masking / 1floppy disk /Bag of screws1SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 24. SECTION 3Checking the premises before installation27SECTION 3CHECKING THE PREMISESBEFORE INSTALLATIONStandardsENGLISH Compliance with standards: Check that the room in which the unit is to be installed complies with the electrical standards and the radiological protection standards in force in the country concerned.Mechanical Position of the unit in the room: Prior to placing the order and before installation, carefully check the position of the panoramic unit in the room. We recommend that the unit should be installed along a wall, but a minimum amount of space must be provided to allow easy access to the patient or by the maintenance technician. The wall bracket allows an optimal installation in a corner of the room. Minimum width of the door: The width of the door must be at least 80 cm (31.5). Minimum height of the ceiling: The overall height of the unit is 230 cm (90.5). We recommend that the minimum height of the ceiling should be 240 cm (94.5). If space is at a premium, the height of the column can be adjusted at the factory if specifically requested on the order. Strength of the wall: The wall must be capable of withstanding an extraction force of 150 kPa at each point of attachment. Depending on the type of wall, it is the responsibility of the installer to choose a fixing system which is capable of withstanding an extraction force of 150 kPa applied to each fixing. Load-bearing capacity of the floor: The load-bearing capacity of the floor should be 500 kg/m2. The floor must be solid and flat.SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 25. SECTION 328 Checking premises before installation ElectricalRequired electrical specifications:Nominal voltage*Maximum lineMinimumMaximum (no load)current110 V 99 V 143 V 20 A130 V 99 V 143 V 20 A230 V 207 V264 V 10 A240 V 207 V264 V 10 A* The units are supplied for operation at either 110 / 130 V 60 Hz or 230 / 240 V 50 Hz or 230 /240 V 60 Hz. The frequency and the operating voltage cannot be modified on site. They must beselected when the order is placed.CAUTION The power supply line must be equipped with a connection box that ensures aconstant connection. It should not be possible to connect the unit to the power supply withoutusing a tool, and the unit must be protected against any accidental disconnection.CAUTION If other units are installed on the same line, interference and voltage fluctuationsmay cause the radiological unit to operate abnormally.We strongly recommend that a separate electrical line should be dedicated tosupply power to the Kodak 8000 unit. This line should be protected by a circuitbreaker with a maximum current of: - 16 A at 230 / 240 V - 20 A at 110 / 130 Vand a differential of 30 mA. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 26. SECTION 3Checking the premises before installation 29Electrical power supplyEvery radiological unit must be connected to a reliable electrical power supply with astandardized grounding connection.A single-phase alternating current power supply is required and the characteristics of theelectrical installation should be as follows: Supply voltage 230V / 240 V230V / 240 V 110 / 130 V Frequency50 Hz 60 Hz60 Hz Electrical supply6 kW6 kW 6 kW ENGLISH Line current required16 A16 A 20A Cable cross-sectionFor 30 m: For 30 m:For 10 m: according to length2.5 mm2 2.5 mm22.5 mm2 Max. line impedance0.5 0.5 0.12 Differential circuit breaker 30 mA 30 mA30 mA Maximum current to trip16 A16 A 20 A circuit breaker Characteristics of 250V250V 250V emergency switch 3 A UL listed 3 A UL listed6 A UL listed Characteristics of the 60 W60 W60 W warning lamps Contactor16 A UL listed16 A UL listed 20 A UL listed- The power supply line must be protected by a differential circuit breaker which trips atmaximum current (refer to the characteristics above). - Two emergency switchs shall be installed in order to simultaneously isolate the current to the active conductors of the radiological installation via a conductor to the exclusion of any electrical equipment (refer to the characteristics above). These emergency switch shall be located near the control panel and near the radiological unit so that operator or patient can quickly cut the power supply if necessary. The device shall have an actuator colored red,and shall maintain the equipment in the off (open) condition until a deliberate action, different from that used to actuate it, is performed. Use the devices below: Emergency switch - manufacturer : Cutler-Hammer - ref: E22AT111 or E22AT112 Contactor - manufacturer : Telemecanique - ref: LC1 D4011 F6 (110/130 V - 60 Hz) P5 (230/240 V - 50 Hz) U6 (230/240 V - 60 Hz) If you can not use the devices, you shall use an equivalent Emergency switch and contactor UL listed with the same characteristics. For additional information, please refer to UL 2601.1 clause 22.7. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 27. SECTION 330 Checking premises before installation - A red warning lamp must be fitted near the switch to indicate that the unit is live.It should be installed outside the point of access to the radiological room (one lamp should be fitted at each point of access). - A yellow warning lamp may be connected to the unit, it lights up during ready state and during X-ray transmission. - A door safety switch may be connected to the unit. This safety switch should trip if the door remains open, thus deactivating the handswitch.- The handswitch may be installed at a distance of up to 10 meters from the unit.Mains distribution board110 /130 V230 / 240 V60 Hz 50 / 60 HzMaximum currentdifferential circuit breakerIdiff = 30 mA Idiff = 30 mAImax = 20 A Imax = 16 AContactor20 A16 AEmergency stop buttonNear the switch buttonEmergency stop buttonNear the unitRed POWER ON lampPower supply socket20 A socket16 A socket SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 28. SECTION 4 Installation procedure31SECTION 4INSTALLATION PROCEDURE Tools to be supplied by the installer: - Power drillNOTE - Screws and boltsAll screw fittingsENGLISHhave metric threads. - Wall plug - Spirit level - Tape rule - Metric Allen keys - A 22-mm wrenchProcedure for opening the boxes Box # 1: 1. Remove the upper cardboard from the packaging (a) and lift off the packagingcarton (b).(a) (b) 2. Release the upper cover (c) with the aid of the two screws. Remove the cover.Unscrew the protection from the patient handle (d). (c) (d)SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 29. SECTION 432 Installation procedure3. Release the cover of the sliding unit (e) with the aid of the 2 screws. Remove it by turning it so as to avoid damage to the temple support arms (f ).(f)(e)4. Release the 4 screws (g) securing the metal transport frame to the pallet. Place the transport trolley (with the panoramic head) at the side of the pallet. This may allow the head to be transported upstairs while ensuring that it is protected.WARNING Do not use the patient handle (1) to manoeuvre the unit. Grip only the upper partof the tubes (2).(2) (g)(1)Box # 2:1. Open the carton from the side. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 30. SECTION 4Installation procedure 332. Remove the column from its box, using 2 persons - one at each end of the column.ENGLISHPlace the column on the floor. You can now proceed with the installation of thewall bracket. Wire the electrical connection board.Box # 3:1. Remove the upper cover from the packaging.2. Remove the Accessories box and the various covers.3. Remove the exterior carton from the top.4. Remove the cephalometric unit with its base.5. Retrieve the cephalometric arm.SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 31. SECTION 434 Installation procedure Mechanical and electrical installation 1. While the column is still on the floor, in a horizontal position, fit the wallmounting bracket, without tightening the screws.Choose the second fixing hole, farthest from the wall. CAUTION If the unit is to receive the cephalometric upgrade, EITHER NOW OR AT A LATER DATE, choose the mounting hole furthest from the wall. 2. Install the wiring in the base of the column, in accordance with the CAUTIONinstructions specified above (refer to the diagram on the next page). Do not connect the unit to the mains supply at this stage of the2.1 Connect the power cord to the unit. installation procedure.The unit must have a continuous power supply. It should not be possibleto connect the unit to the power supply without using a tool, and theunit must be protected against any accidental disconnection.Depending on the standards in force in your country, it is recommendedthat the following functions should be provided: NOTE Safety lighting For remote control of the X-rays, install a 12 mm Connect the "X-ray emissions" safety lamp (max. 60 W). diameter conduit with pull-in This lamp illuminates when the unit is emitting X-rays. tape (or cable guide) between the base of the unit and the location of Door safety switch: the X-ray shutter release;If this function is used, remove the strap (A) which is installed on the The same applies to the unit as the default configuration. door safety switch contact Wire in a Door safety dry contact and the X-ray emissions safety lamp.(characteristics of the contact: 250 V min., 0.5 A min.). This contact should be closed when the X-ray protection facilities are in place; otherwise it should be open. If the contact is open, the x-ray handswitch release is disabled.2.2 Install the X-ray handswitch release cables directly into the base of thecolumn or use the extension kit supplied with the unit.2.3 The procedure for wiring the RS232 link to the Com 1 or Com 2 porton the PC is described in the Section "Installing the PCI acquisitionboard in the computer". CAUTION Modifications of the RS 232 and RS 422 cables are under the responsibility of the Kodak dental systems certified dealer. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 32. SECTION 4 Installation procedure352.4 The procedure for wiring the RS422 link for the digital acquisition system tothe PCI data acquisition board supplied with the unit is described in theSection "Installing the data acquisition board in the computer".ENGLISHPower cord Safety Door safety X-raySUB-D SUB-D lamp contact handswitch 9-pin 9-pin 60 W (max) 0.5A femalemaleor strap+ extensionkitTo COM 1To PCI or COM 2acquisition on the PC boardDo not position the cover on the base of the column.CAUTION OPTIONAL BASE PLATE If a mounting base is to be used, it should beinstalled at this stage. However, mounting the column on the wall is STRONGLYrecommended in order to ensure OPTIMUM STABILITY and to insulate the unit fromvibration.CAUTION If the unit is aimed at receiving a cephalometric attachment, fixing the column tothe wall is IMPERATIVE. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 33. SECTION 436 Installation procedure3. Place the column in a vertical position, taking care not to trap any cables between the column and the floor. Hold the mounting bracket against the wall at the top of the column and mark the position of the fixing holes. Move the column to one side as you drill the holes. Insert the wall plugs. Then place the column in position and attach the mounting bracket to the wall. Check that the colomn is properly aligned in both directions. (h)4. Mark the position of the 3 fixing holes at the base of the column and move the column to one side while you drill the holes. To do this, slacken the rotation screw on the wall mounting bracket. Insert the plugs and tighten. Check again that the column is vertical, then screw the mounting bracket securely to the wall.5.Unscrew the panoramic head from its transport trolley (i).Grasp the unit on each side by the handle (j) and the plate (k).Locate the panoramic head on the column by positioning it on the hook ofthe sliding unit (l). Fit the 2 upper screws but do not tighten them.WARNING Do not use the patient handle to manoeuvre the unit. Grip only the upper part ofthe tubes (j). (i)(k) (j) (l) SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 34. SECTION 4 Installation procedure 37CAUTION 6.Then release the arm unit by withdrawing the transit locking screw (m)This locking screwwhich passes through the CJ 653 board. Attach the screw to the chassis, somust be usedthat it can be quickly retrieved if the unit is dismantled at a later date.whenever the unitis transported ormoved. If the armunit is not immobilisedwith the locking screw,the unit may sufferserious damage(m) ENGLISH7.Rotate the arm unit by hand.You can now locate the 2 screws at the base and then tighten them.8.Tighten the 4 screws of the head assembly, while relieving the weight of thehead at the front to facilitate tightening.Position the cover of the sliding unit (n). Attach the switch to this cover. (n)The other covers are installed at a later stage, after checking the settings ofthe unit.You can now install the wiring connections. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 35. SECTION 438 Installation procedure Installing the optional cephalostatInstalling the optional cephalostatInstall the arm on the sliding unit on the rear of the column and insert the 6hexagon socket cap screws with the washers.NOTE If the arm cannot be mounted on the six pins, enlarge the mounting holes in thecephalostat arm. A tighter hold will make any adjustments faster. NOTE Bring the cables out through the top of the arm.Installing the cephalostat headATTENTION The cables are inside the elbow tube. Be careful not to damage them wheninstalling the cephalostat head.TIP: Screw in the height adjusting screw as tightly as possible.Use adhesive tape to bind together the cables that emerge from the tube, and slide them intothe interior. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 36. SECTION 4 Installation procedure 391.Place the cephalostat head onto the elbow tube so that the tube protrudes by 40 mm.40 mm ENGLISH2.Use a spirit level to verify that the assembly is horizontal and make adjustments ifnecessary. Tighten the four Phillips head screws. WARNING Pay careful attention to the electronics components.3.Use a spirit level to verify that the assembly is vertical. If necessary, adjust thescrews on each side of the collars on the primary arm.4.Put the head clamp in a lateral position. Attach the head clamp arms. CAUTION The two arms are different. Be careful to mount the arm with the ball on the cephalostat receiver side and the arm with the washer on the generator side. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 37. SECTION 440 Installation procedure5. Make an initial adjustment to the height of the cephalostat unit by rotating the adjusting ring on the top of the upper collar. Measure the distances between the floor and the middle of the primary collimator and between the floor and the center of the balls on the head clamp. If these heights are not the same, re-adjust the cephalostat height using the adjusting ring.6. Pass the cables through the elbow tube and: a) Connect cable # 44 to the LAMBDA power supply and add a ferrite. b) Connect cable # 46 to J6 (CJ 697) and add a ferrite. c) Connect cable # 48 to J2 (CJ 697) and add a ferrite. (b)(c) (a)7. Remove the red transport collar from the cephalostat. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 38. SECTION 4Installation procedure 41Installing the wiringGeneral connectionsCAUTIONEnsure that the unit isswitched off and disconnectedfrom the power supply.ENGLISH Cable # 23 to J 9of the CJ 653 boardENGLISH (O)(P) (Q) (R) Cable # 36 on J5 of the CJ 699 board (O)Blue cable (Q) and Cable # 37 on J8 of the brown cord (R) to the CJ 699 board (P)mains input terminals.Their position isdependent on the wiringfor the filter output.+ yellow cord (Ground)to the screw at the side of the filter CAUTION Never connect one of theDigital sensor to PC mains wires to the centralconnection terminal of the filter ground SUB-D9-pin male and female Cable # 10 to J 15Cable # 12 to J 18of the CJ 664 board of the CJ 664 board SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 39. SECTION 442 Installation procedureAdditional connections for the optional cephalostatCAUTION This connection should only be made when installing a Kodak 8000 unit withcephalostat. CAUTION Ensure that the unit is switchedCable # 39/40 to J9 off and disconnected on the CJ 699 board from the power supply.Cable # 44 to J12 on the CJ 699 board CAUTION The CJ 664 board must have EPROMs that are Version 2.02 or later. Use care during installation. CAUTION Attach the cable near the J20 so as to minimize sporadic breakdowns. Digital sensor to PC connection SUB-D9-pinmale and femaleCordon # 46 sur J 20EPROMS de la carte CJ 664 (V. 2. xx)(v. 4)*Cable # 38 to J17Cable # 11 to J16 on the CJ 664 boardon the CJ 664 board CAUTION* If your CJ 664 is earlier than Version 4, connect the CJ 715 adapter circuit to J3 and connect cable # 46 as well as the cable that was at J3. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 40. SECTION 4 Installation procedure43Installing the computer Environment For details of the computer system (PC and screen), refer to the installation manuals of your microcomputer. Sufficient clear space must be left around the CPU to ensure that it is properly ventilated. In order to obtain maximum image quality and visual comfort, the screen should be positioned so as to avoid direct light reflections (internal or external lighting). The imaging system must be inspected and checked in accordance with the standards currently in force.ENGLISH The Kodak 8000 unit has been designed to comply with European and international medical standards, therefore the computer configuration should comply with IEC Standard 950. The PC and the peripheral equipment connected to it must not be located in the immediate vicinity of the patient, thus at least 1.5 m from the panoramic unit. Check that the electrical system complies with the standards currently in force in the country of installation: in France this is Standard NFC 15-100. Minimum configuration CAUTION The Kodak dental imaging software MUST be installed on a PC with at least 256 MB of RAM. It is therefore COMPULSORY to check the PC configurations and, if necessary, to proceed to the update of the computer. Microprocessor Pentium III or equivalent StandardsIEC 950 Hard disk40 GB Graphics cardResolution 1024 x 768 - 24 bits RAM (Random Access Memory) 256 MB Operating system Windows 2000 and XP Type of PC Multimedia Available hardware slots One PCI slot + One RS 232serial communication port DriversCD-ROM + floppy disk Data backup system High capacity magnetic mediumSM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 41. SECTION 444 Installation procedure X ray switchPC ( IEC 950 )(not supplied)X ray safety light RS 232 port(not supplied) (COM 1 / 2) X ray safety switch RS 422 port (not supplied)(PCI board) Mains cable SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 42. SECTION 4Installation procedure 45Installing the PCI acquisition board in the computerDescriptionCAUTION The Windows 2000 and Windows XP operating systems are supported. Access inBefore carrying out any administrator mode is required in order to properly configure and finalize thetechnical work on thecomputer, ensure that it is installation procedure.switched off anddisconnected fromthe mains power supply.Step-by-step instructions1.Install the PCI board (CJ 648 v.6) in an unoccupied slot.ENGLISHCAUTIONThe board must be installedby an approved Kodak dentalsystems technician.2.Connect the RS 232 cable (COM 1 or 2) and the RS 422 cable. CAUTION When using a Kodak 8000C unit, connect the RS 422 cable to the 25-pin connector using an adapter.3.Boot the PC. Wait while Windows detects the presence of a new board. CAUTION At this stage of the installation, the CD-ROM should not be in the drive.4.Cancel the installation of the board and proceed directly with installation ofthe software, so that the drivers will be installed automatically.SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 43. SECTION 446 Installation procedureInstalling / updating the software Launching the installation programATTENTIONOnce you finish installing the components, you must install the Kodak dentalThe Kodak 8000C unit imaging software (v. 6).is compatible only withVersion 6.0x or later of theKodak dental imaging 1.Insert the CD-ROM in the CD-ROM drive.software. 2.Wait for the installation program to start. If the program does not start automatically, click on the Start button, then select Run and enter D:setup.exe on the command line (if D is the letter for the CD-ROM drive; this letter may be different on your computer). 3.Choose the installation language for the Kodak dental imaging software. 4.Accept the License Agreements. 5.Select PANO to specify the hardware for which the software will be installed.CAUTION Do not forget to select PANO.NOTE When the Kodak dental imaging software is being installed for the first time,the option Patient File is also available. Select this option. 6.Click on Next to begin installation of the software. 7.Follow the instructions and click on OK each time until the installation is complete.NOTE When you see the window that says Choose your installation operationchoose the default selection Initial installation. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 44. SECTION 4Installation procedure 478. Select your panoramic unit: Kodak 8000 or Kodak 8000C.ENGLISH9. A new configuration window is displayed. Choose : a) The operating language of the Kodak 8000 graphic interface. Specifically,this selection will determine the language used for the voice assistance. b) The number of the serial port on the PC to which the panoramic unit isconnected. c) Optionally: activation of vocal mode (voice assistance). This option can bemodified while the software is in use. (a)(b) (c)(d)CAUTION The Demonstration mode (d) is intended solely for commercial use. Activationof this mode disables any communication with the unit and, as a result, prevents imageacquisition.Do not activate this mode for an on-site installation of the Kodak 8000C unit. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 45. SECTION 448Installation procedure10. When you see the window that asks you to install the masks, follow theinstructions. CAUTION The masking files can be found on the floppy disks (pano.msk and cephalo.msk). Be sure to insert the proper disk in the drive.11. Complete the installation and restart your computer.The installation of the unit is now complete. For more information, refer tothe Kodak 8000C unit and the Kodak dental imaging software user guides.A detailed online Help function is also available at any time from the Kodakdental imaging software. NOTE If you wish to add Kodak dental systems equipment at a later date, simply restart the Kodak dental imaging software installation program, select Modify in the screen that appears and choose the appropriate option(s) from the menu. The drivers for your new equipment will then be installed. For additional information, refer to the user guides for your Kodak dental systems equipment. NOTE In order to use the software in a network, the software must be installed on every workstation. The database path must be specified in the patient file. Enter the network path in the Options menu under Preferences , Patient file. For example: //Server_name/database_path/. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 46. SECTION 4Installation procedure 49LicenseAfter installation of the Kodak dental imaging software you have 30 days toregister the software. You can use the software with all its functionality for aperiod of 30 days. Once this delay has elapsed you will not be able to acquire anynew images, the visualization and manipulation of images however will still bepossible.To enter the license number and obtain an activation key, you need to:Launch the software, select a patient at random and launch the imaging module.In the imaging module, select the ? menu and click on License The windowbelow will open, allowing you to enter the necessary information for the activationENGLISHof your software.The serial number of your software is automatically inserted in the form, as well asthe version of the software. You can also find the serial number on your CD-Rom.The installation key is automatically generated based on your serial number and thecomputer you install the software on.Please complete all the fields of the form!NOTE You can install the software on a maximum of 5 different computers using thesame serial number. Each computer will generate its own installation key that you need totransmit to the Kodak licensing department. To install additional computers (beyond 5) youwill have to order a license upgrade CD-Rom, each upgrade will allow you to install up to5 additional computers.SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 47. SECTION 450 Installation procedureNOTE When upgrading or re-installing your software you will need to activate thelicenses again and request new license numbers for those computer(s).To transmit the information you can select between transmission through theInternet or by post/fax. You will quickly receive the license number by email, faxor post, which you need to enter in the field marked by an arrow above tocompletely activate your software.If you select Send by Internet an email will be created containing the data youhave entered. You only need to send the email to:[email protected] (for the UK) or [email protected] (international)If you select Send by post a form containing the data you have entered willautomatically be printed on the default printer defined in Windows. This willallow you to:- either send the form by post to: Trophy Radiology UK LTD One Rochester CourtMedway city estate - Rochester Kent ME2 4NWUnited Kingdom or Trophy Radiologie License service 4 rue F. PelloutierCroissy-Beaubourg77437 Marne la Valle Cedex 2 France- either fax the form to+44 634 73 12 55 (for the UK)or + 33 1 64 80 84 99 (international)For all questions related to the activation and license of your software, you cancontact your Kodak certified dealer. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 48. SECTION 4 Installation procedure 51Configuring the installation hardware(for Kodak 8000C unit only)1. Start the Kodak dental imaging software.2. Start the Kodak 8000C unit in technician mode.3. Select the New panoramic image icon. ENGLISH4. Click on the Information icon, then Manufacturing parameters.5. Select the type of cephalostat installed (right-hand or left-hand arm). - For a cephalometric arm installed on the right side of the unit, select Cephalostat on the right-hand side: - For a cephalometric arm installed on the left side of the unit, select Cephalostat on the left-hand side:6. Save and exit.7. Exit the interface. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 49. SECTION 452Installation procedurePost-installation checksAligning the panoramic sensorWARNING The technician must comply with the standards currently in force while installingand using the radiology unit.Whenever necessary, he must use a shield or wear a lead-lined apron.The unit must be switched off before the covers are removed.Any work which has to be carried out on the unit while it is switched on andwhile the covers are removed must be undertaken by an approved Kodak dentalsystems technician. Unqualified persons must not approach the unit.1. Acquire an image using the ball phantom supplied with the unit. Remove the biteblock. Position the ball phantom and check that the graduated scale on the chin rest is set to the "0" position.2. Check that the sagittal-median and canine lasers intersect perpendicularly on the central ball of the ball phantom. Check that the laser of the Frankfurt plane is perpendicular to the centerline of the digital sensor. If the lasers are not correctly aligned, refer to the Service Manual.3. Acquire a panoramic image at 60 kV, 2 mA.4. On the image (refer to the diagram below), check: - the size of the balls - the circularity of the balls. The circularity of the central ball is the most important feature to be checked.= OK = NO ! = NO ! SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 50. SECTION 4Installation procedure 53 - the right / left symmetry.ENGLISH5.If the results of the checks on the negative are satisfactory, the unit is ready foruse and the installation is complete.If the results of any of these checks are unsatisfactory, refer to the ServiceManual in order to identify the causes of any problems.Aligning the cephalometric sensor and covering thefixed collimator1.Use a spirit level to verify that the receiver unit is properly horizontal andvertical and make adjustments if necessary.2.Select the Cephalostat acquisition mode.SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 51. SECTION 454Installation procedure3. From the Information menu, under Calibration, choose Acquisition without security.4. Set the Kodak 8000 unit to cephalostat mode, remove the collimation power supply from the X-ray field, attach the hood and the head clamp in lateral position and tighten the arms to about the halfway position.4. Prepare an image using the following constants: 60 kV, 4 mA, 0.5 second.6. On the image, verify that the two balls are concentric. The distances A and A must be identical, with a tolerance of 0.5 mm.AA OKNO!7. Adjust the receiver unit so that the two balls are aligned. - Horizontal adjustment: Slightly adjust the angle of the elbow tube with respect to the primary arm so that the ball is in the center of the ring (b). - Vertical adjustment: Slightly adjust the angle at which the hood is inclined with respect to the connecting arm so that the ball is in the center of the ring (c). (c) (b) SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 52. SECTION 4 Installation procedure 558. Check the position of the balls in a lateral image. Using the cephalostat measuring tool, verify that the tools are 30 mm 5 mm from the nearest edge. ENGLISH9. Check the field covered (by the fixed collimator). On the resulting radiograph, verify that the fixed collimator cannot be seen in the image. Repeat the process with a symmetric image. If a white edge appears, move the hood assembly along the horizontal tube and re-adjust the balls.10. Set your collimator to Lateral ceph.11. Switch on the Frankfort laser beam.12. Place the cephalostats ear posts in the lateral position.13. Verify that the beam is centered on the ear posts ball. If it is not properlycentered, adjust the beam on the generator.- Vertical position: Loosen the adjusting screw and adjust the mirror in orderto center the beam on the ball. Tighten the screw.- Horizontal position: Loosen the screw that holds the laser support assemblyin place. Adjust the position using the adjusting screw that acts on the screwopposite the assembly attachment.- Focus: Remove the laser tube from its base and rotate the lens until thebeam is sharply defined. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 53. SECTION 456Installation procedureAdjusting the collimators (lateral and front)1. For each image, choose the appropriate position.2. Prepare an image using the following constants: 60 kV, 4 mA, 0.5 second.3. Check the collimator and adjust the setting to ensure that there is a white area (a gradually lighter shade) on all sides of the image.WARNING As lead is a soft metal, avoid squeezing it when making any tightening adjustments.Adjusting the arm1. Select the New ceph image mode.2. From the Information menu, under Calibration, choose Acquisition without security.3. In the Arm calibration window, enter a value for the Arm centering offset in cephalometric mode of between -10 and 10, such that the locking rod is properly in place opposite the stop piece when positioning in Cephalo mode. Press the stand positioning button to rotate. Once the proper value has been determined, save and exit. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 54. SECTION 4Installation procedure57Fitting the coversMounting the standard coversYou can now install the last of the covers.1. Fit the upper cover to the panoramic unit. ENGLISH2. Install the rear panel to the sliding unit while rotating the column. (for the Kodak 8000 unit only)3. Position the upper cover on the column.4. Position the cover on the wall-bracket. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 55. SECTION 458Installation procedure5.Install the cover of the electrical junction box. CAUTION When installing a Kodak 8000C unit, mount any additional covers before powering up the unit (see below).6.Check that all the screws in the wall and floor have been tightened.7.Connect the power supply to the panoramic unit.Mounting additional covers on the optionalcephalostat1.Install the secondary arms rear panel.2.Mount the front cover to the cephalostat.3.Mount the rear cover to the cephalostat. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 56. SECTION 4Installation procedure594.Position the carbon plate on the cephalometric sensors front panel. CAUTION The plate must be positioned in the proper direction. Always verify that the logo is on the interior side closest to the arm.Positioning the plate with a Positioning the plate with aleft-side armright-side arm ENGLISH5.Check that all the screws in the wall and floor have been tightened.6.Connect the power supply to the panoramic unit.SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 57. SECTION 5Technical specifications 61SECTION 5TECHNICAL SPECIFICATIONSGeneral technical specificationsDigital Panoramic and cephalometricProduct identificationRadiologyUnit, controlled fromENGLISHa PC interface. Face-to-face positioningDigital sensorPano: Sensor CCD + optical fiber sensorCeph: CCD sensorDimensions of the matrixPano : 2 500 x 1 244 pixelsCeph : 1840 x 1360 pixelsGray scales 4,096 12 bitsImage field Pano: 5 x 126 mmCeph: 196 x 260 mmHigh tension generator140 kHz (max.)60 - 90 kV2 15 mA orLine voltage230-240 V / 50 Hz , 230-240 V / 60 Hz110 - 130 V / 60 HzFocal spot of tube0.5 mm (IEC 336)Total filtration> 2.5 mm eq. AlAnode voltage 90 kV max.Cathode current 15 mA max.Imaging proceduresPanoramicSegmented panoramicMaxillary sinusLateral TMJ, 2 sectionsLateral TMJ, 4 sectionsLateral cephalometric imageAP / PA cephalometric imageOblique cephalometric imageSubmento-vertex imageCarpus imagePeriod of exposureAdult panoramic:13.9 sec.Pediatric panoramic:13.2 sec.Maxillary sinus:10.7 sec.Lateral TMJ, 2 sections: 4.6 sec.Lateral TMJ, 4 sections: 9.2 sec.Cephalometric images: 0.1-3.2 sec. SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 58. SECTION 562 Technical specificationsExposure mode2 programs (adult and child) 3 sizes (adjustable by the user)MagnificationPano: 1.27 Ceph: 1.15Required space Kodak 8000: 1500 (L) x 1950 (D) x 2400 (H) mm* Kodak 8000C: 2800 (L) x 1950 (D) x 2400 (H) mm*Weight Kodak 8000: 120 kg Kodak 8000C: 145 kg* A unit with short column is available by special order. International standards Electromagnetic compatibility With regard to EMC, this product complies with European Directive 89/336/ CEE. This product complies with European Standard EN 60601-1-2. Its classification is Group I, Class B. Conforming to IEC Standards 601.2.7 and 601-1 MODEL: The Kodak 8000 unit radiographic dental diagnostic unit, IEC 601.2.7, Class 1, Type B, Intermittent Operation, with protection against X-rays complying with IEC 601.1.3 AMBIENT OPERATING CONDITIONS Temperatures 5 ~35 C Relative humidity30 ~ 85% Atmospheric pressure 700 ~ 1060 hpa STORAGE CONDITIONS Temperature-10 ~ 60 C Relative humidity10 ~ 95% Atmospheric pressure 700 ~ 1060 hpa TRANSPORT CONDITIONS Temperature-10 ~ 60 C Relative humidity10 ~ 95% Atmospheric pressure 700 ~ 1060 hpa SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 59. SECTION 5 Technical specifications 63Type of electrical power supply:230 - 240 V110 -130 V50/60 Hz, single-phase 60 Hz, single-phaseAcceptable fluctuation 10% 10%Apparent resistance of the power supply circuit 0.5 (max.) 0.12 (max.)Permanent absorbed current0.45 A 0.9 ACurrent absorbed during the X-rays emission 9.5 A17 AMax. absorbed power 2.2 kVA1.9 kVABy shutter release at a max. By shutter release at a max.Protection for the power supply systemcurrent of 16 A and acurrent of 20 A and adifferential current of 30 mAdifferential current of 30 mANominal high tension90 kV90 kVMaximum corresponding tube current10 mA10 mANominal tube current15 mA15 mAMaximum corresponding high tension80 kV68 kVTube current/voltage combination for80 kV, 15 mA 85 kV, 12 mAmaximum output powerNominal power for an exposure timeat 80 kV, 15 mA: at 85 kV, 12 mA:of 0.1 s. 1200 W 1020 W ENGLISHUtilization rate in continuous mode(for example: one exposure 85 kV, 5 mA 33W.13.9 sec every 3 minutes)Utilization rate in intermittent mode(for example: one exposure 80 kV, 15 mA 93 W. 13.9 sec every 3 minutes)Selection of the load parameters:kV (in increments of 1 kV)From 60 to 90 kVmA (in increments of 25%) From 2 to 15 mACooling conditions:Maximum dissipation of heat from the X-rayradiogenic assembly into the ambient air70 W.(for a utilization rate in continuous mode):Details of the load parameters:High tension:kV 10%Current in the tube: mA 20%Exposure time: seconds (10% + 1ms)Measurement conditions:KV Indirect on the peak kilovoltmeterMA Direct measurement in the circuit using an oscilloscopeExposure time:Measurement at 75% of the kV values SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 60. SECTION 564 Technical specifications Characteristics of the X-ray radiogenic assemblyConforming to IEC Standard 601.1.3IEC Standard 601.1.3CompliantNominal value of the inherent filtration at 70 kV 2.5 mm (0.10) eq. AlNominal value of the supplementary filtrationNot applicableat 70 kVNominal value of the total filtration2.5 mm (0.10) eq. Alat 70 kVFiltration value for the enclosure of the0.2 mm (0.08") eq Alradiogenic unit (at 100 kV)Filtration value for the enclosure of the image0.2 mm (0.08") eq Alreceiver unit (at 100 kV)Filtration value for the sensor case0.8 mm (0.31") AlValue for the arms of the head clamp (at 100 kV)1.0 mm (0.039") eq AlDistance between the focal spot and the sensor522 mm ( 20.55 ". )Location of the reference axisRefer to the diagram belowSize of the beam at the sensor5 x 126 mmRadiation leakage after one hours operation(maximum utilization rate of 93 W, i.e. 90 kV , 10 mA , < 1 mGy 13.9 sec. every2 minutes 15 sec.)Conforming to IEC Standard 601.2.28The radiogenic unit comprises: a transformer and a radiogenic tube, together with their associated electronic components immersed in oil. an aluminum filter which enhances the quality of the beam and reduces the dose received by the patient. a unit which limits the size of the beam at the image receiver unit. a thermal cutout which trips at an operating temperature of 65 C 5 C. SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 61. SECTION 5Technical specifications 65 Characteristics of the radiogenic unit IEC Standard 601.2.28 Compliant ManufacturerTrophy Radiologie S.A. Degree of protection against electric shock Class I Degree of protection of the parts applied to Type B the leakage current from the patient Maximum accumulated heat110 kJ Maximum continuous heat dissipation 33 W. Nominal value of the focal spot (IEC 336/192) 0.5 mm Tolerances on the position of the focal spot+/- 2.5 mm Weight8.2 kgENGLISH Dimensions235 x 245 x 120 mm Heating and cooling curves:Accumulated heat in kJ Cooling Time in minutesSM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 62. SECTION 566 Technical specifications Beam limitation unit:ManufacturerTrophy Radiologie S.A.TypeRigidly mounted unitwith fixed windowdimensions, not removabledand integrated x-ray generatorMaximum symmetrical field of radiationComplies with IEC5 mm to 126 mm at a distanceStandard 806of 522 mm from the focal pointMaximum symmetrical field Complies with IECof radiation of 190 mm / 250 mm Standard 806in cephalometric mode at a distanceof 1720 mm from the focus Position of the reference axis:TUBEHEADASSEMBLYX-RAYGENERATOR622 REFERENCE0AXIS FOR TH00,X-RAY120,00 PRIMARY RAE DIATION BEAM0 30,TUBE 45,504 HORIZONTAL 24,50 SENSOR303235 Positioning of the reference axis for cephalometric imaging:1720 mm TUBEHEAD ASSEMBLY1 X-RAY GENERATORREFERENCE AXIS FOR THPRIMARY RADIATION BEAME X-RAY TUBE 5.0276.0HORIZONTAL129.5 SENSOR SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 63. SECTION 5 Technical specifications 67 Characteristics of the X-ray tube:Manufacturers name CEITypeOPX 105Nominal high tension90 kVNominal anodic power810 WMaximum amount of heat accumulated in the anode 30 kJNominal focusing value (IEC 336)0.5 mm ( 0.020 )Collector electrode materials TungstenAngle of the collector electrode5Inherent filtration 0.5 mm ( 0.20 " ) eq. Al ENGLISH Heating and cooling curves: SM 670KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 64. SECTION 568 Technicals specifications Load alignment chart: Anodic currentExposure time Characteristics of emissions from the filament: Anodic currentFilament current SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE 65. SECTION 6Preventive maintenance 69SECTION 6PREVENTIVE MAINTENANCEAnnual maintenanceWe recommend that a general inspection of the unit should be carried out every yearby an approved Kodak dental systems technician. The inspection should cover theENGLISHfollowing points: - Check the mechanical installation - Clean the rotation track - Check the alignment of the radiation - Make a test exposure on the ball phantom and check the image - Check the points of attachment to the floor and wall - Check the degree of legibility of the labels - Check the X-ray generator - Check the focal trough - Damage to cables, covers, oil leaks, etc.If the results of any of these checks are unsatisfactory, refer to the Service Manual inorder to rectify any problems.If there is any doubt, do not operate the unit.SM 670 KODAK 8000 & KODAK 8000C INSTALLATION GUIDE