  • C7SM-1.01 2/70

    Service Manual of Cardio 2000

    Revision History

    Edition Date Page Comment

    Ver 1.0 2019-10-18 Initial Release

  • C7SM-1.01 3/70

    Service Manual of Cardio 2000

    - TABLE OF CONTENTS –

    Section 1. General Information ............................................................................................... 5

    1-1. Introduction ...................................................................................................................................................................... 5

    1-2. Features of the Equipment ...................................................................................................................................... 5

    1-3. Equipment Description .............................................................................................................................................. 6

    1-4. System Installation .....................................................................................................................................................15

    1-5. Terms of Warranty .......................................................................................................................................................16

    1-6. Specification .................................................................................................................................................................16

    Section 2. Troubleshooting .....................................................................................................18

    2-1. Power .................................................................................................................................................................................18

    2-2. Wave Printing ................................................................................................................................................................19

    2-3. Diagnosis .........................................................................................................................................................................22

    2-4. Filter ....................................................................................................................................................................................23

    2-5. Trouble Shooting Guide ..........................................................................................................................................25

    Section 3. Adjustment .............................................................................................................31

    3-1. LCD Backlight................................................................................................................................................................31

    3-2. Lead Fault ........................................................................................................................................................................32

    3-3. Filter Setting ....................................................................................................................................................................33

    Section 4. Block Diagram/Description ................................................................................34

    4-1. Circuit Block Diagram ..............................................................................................................................................34

    4-2. Equipment Overview ................................................................................................................................................35

    4-3. Size and Weight ...........................................................................................................................................................38

    Section 5. Disassembly and Assembly ................................................................................39

    5-1. Module Description ...................................................................................................................................................39

    5-2. Internal Block Diagram & Cable ........................................................................................................................46

    5-3. Disassembly ...................................................................................................................................................................49

    5-4. Assembly .........................................................................................................................................................................59

  • C7SM-1.01 4/70

    Service Manual of Cardio 2000

    Section 6. Replaceable Parts List and Accessory .............................................................60

    6-1. Replaceable Part List ................................................................................................................................................60

    Section 7. Connector Pin Assignment .................................................................................63

    7-1. Patient Cable Port ......................................................................................................................................................63

    Section 8. PCB Part Disposition ..............................................................................................64

    8-1. Base (Power+Analog) Board ................................................................................................................................64

    8-2. CDUI Board .....................................................................................................................................................................65

    8-3. CPU Board .......................................................................................................................................................................66

    Section 9. Software Upgrade .................................................................................................67

    9-1. Requirement of Upgrading....................................................................................................................................67

    9-2 Procedure of Upgrading (Cardio7 UI screen, upgrade method is same) .................................68

  • C7SM-1.01 5/70

    Service Manual of Cardio 2000

    Section 1. General Information

    1-1. Introduction

    CardioCare2000 is an electrocardiogram (ECG) device capable of measuring and recording 12

    channels of cardiac signals of patients. The device enables operators to record cardiac signals,

    review them using various viewing modes, and perform automatic detection of cardiac conditions.

    In addition, the device allows operators to enter patient information for it to be printed alongside

    cardiac signals, so that operators can sort and manage charts conveniently. It is also possible to

    manage the digital files by transferring the stored data to a networked PC.

    Furthermore, user convenience has been further enhanced by addition of a functionality that

    can initiate ECG recording & saving, filter enhancing, parameter sorting and automatically

    detecting cardiac signal all at once by pressing one key.

    A battery (optional) can be added to the device so that the device can be operated

    conveniently during bedside visits or emergency conditions.

    1-2. Features of the Equipment

    - Print 12-channel (ch) cardiac signals with reports in: 3ch+1 rhythm, 6ch+1 rhythm, and 12ch

    rhythm patterns on A4 or Letter size printing paper.

    - Record and print 12 channel rhythms continuously and simultaneously.

    - Calculate heart rate, P-R-T axis and PR, QRS, QT and QTc widths from recorded signals

    automatically and print them on a report for use in rhythm analysis.

    - Get diagnostic reports using automatic detection functionality.

    - Able to modify filter setting, signal sensitivity, printing speed, channel view settings and rhythm

    settings, and print on previously recorded EKG signals to aid data analysis.

    - Able to attach a battery so that the device can become portable.

    - More than 200 patient‟s data can be saved and transferred to a PC through LAN.

    - It offers various protocols to make it possible to link with hospitals‟ computing networks. Also File

    Databases are strengthened.

  • C7SM-1.01 6/70

    Service Manual of Cardio 2000

    1-3. Equipment Description

    The CardioCare2000 system consists of the items below. Unpack the package and check the

    items below are included. Also, be sure to check for any damage to the body and accessories.

    Basic Components and Accessories

    ① CardioCare2000 Body (1 EA) – Dimension: 290(W) 300(D) 97.5(H)mm

    ② Patient Cable (1 EA) – Length: 3,700mm (Max)

    ③ Limb Electrodes (1 SET)

    ④ Chest Electrodes (1 SET)

    ⑤ ECG Paper (1 EA)

    ⑥ Power Cable (1 EA) – Length: 3,000mm (Max)

  • C7SM-1.01 7/70

    Service Manual of Cardio 2000

    Optional Components

    ① Battery (1 EA) – Replaceable and Rechargeable, Lithium ion, 11.1V, 2600mA

    ② Hanger (1 EA)

    ③ Cart (1 EA)

    CAUTIONS

    You may have distortion or signal noise when you use nonstandard or other brand

    accessories. We strongly recommend you use only the authorized accessories which we

    supply.

    WARNING

    How to replace battery: Please make sure you use the right battery we show here.

    Otherwise we are not liable for any damages and/or explosion/fire caused by using the

    wrong battery.

  • C7SM-1.01 8/70

    Service Manual of Cardio 2000

    Body Configuration

    ▣ Top View

    ① Handle

    ② Printer Cover

    ③ Printer Cover Switch

    ④ LCD

    ⑤ Control Panel

    ② ③

  • C7SM-1.01 9/70

    Service Manual of Cardio 2000

    ▣ Front View

    ① Printer Cover Switch

    ▣ Rear View

    ① ②

    ④ ⑤

    ① Protective Ground Terminal

    ② Power Switch

    ③ AC Power Connection Port ( Fuse : 250V , 5A ( High breaking type ))

    ④ USB Port(Use for equipment upgreades only)

    ⑤ RJ45 LAN Port

  • C7SM-1.01 10/70

    Service Manual of Cardio 2000

    ▣ Left Side View

    ① Handle

    ▣ Right Side View

    ① Patient cable connection port

    WARNING

    There is a risk of electric shock if the Rest stand of the equipment is damaged or can

    not be fixed to the product. Do not use the product and immediately ask the

    manufacturer and the seller for repair

    NOTE

    To avoid an expected electric shock, do not open the equipment cover or

    disassemble the equipment. Refer servicing to Bionet, Inc.

  • C7SM-1.01 11/70

    Service Manual of Cardio 2000

    LCD Panel

    LCD panel displays system setting status after indicating the version of the system and

    manufacturer name for 2 seconds when power is turned on.

    Displayed items on the LCD are as follows.

    ① ② ③ ④

    ⑤ ⑥ ⑦

    ① Display ECG signal level out of 5, 10, 20, aut (Auto Gain)

    ② Display printing speed out of 12.5, 25, 50

    ③ Display base line filter setting on or off

    ④ Display EMG filter setting on or off

    ⑤ Display heart rate

    ⑥ Display channel form of the output report out of 3ch+1rhy, 6ch+1rhy, 12ch rhy, 60s 1rhy.

    ⑦ Display rhythm channel setting out of I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6..

    1 0 1 2 . 5 O F F O F F

    1 0 0 6 C H + 1 R H Y a V R

    ① ②

  • C7SM-1.01 12/70

    Service Manual of Cardio 2000

    Indicator Lamp

    I Indicate the lead connection status.

    If LED is illuminated green, connection status is good .

    If LED is off, connection status is bad.

    In this case, check off which one of lead connection is in lead-off status

    through monitor mode output.

    Caution: Start printing when LED is illuminated green.

    Indicate the battery capacity status: Green and Red colors are used

    to display 2 different statuses.

    The battery indicator is red when the device is operational without having

    an A/C power supply. When an A/C power is supplied, the battery

    indicator will be either red or green depending on the charge level. No

    color will be shown, if there is no battery installed in the device.

    Indicate AC power is operated if LED is illuminated green and not in use if

    LED is off.

    Control panel

    Select signal level(mm/mV) out of 5, 10, 20, aut

    (I-aVF:10, V1-V6:5 )

    Select printing speed(mm/sec) out of 12.5, 25, 50

    Select on or off whether or not to activate the filter that eliminates base line

    drift

    Select on or off whether or not to activate the filter that eliminates EMG.

    ECG signal can be distorted by applying this filter.

    This key is excepted to use when network function is added ahead.

  • C7SM-1.01 13/70

    Service Manual of Cardio 2000

    Select channel form of the output report out of 3ch+1rhy, 6ch+1rhy, 12ch

    rhy, 60s 1rhy

    Select rhythm channel

    Call up the menu to enter patient ID, name, age, sex, height, weight

    Display real-time ECGs from the patient continuously or monitoring the

    patient ECGs for long time,

    Record ECGs for 10 seconds by pressing key and prints in A4 sized report

    form after calculating the measurement parameter of the recorded ECGs

    Print the same report as previously printed one in record mode or prints

    after changing system setting such as filter, signal level, output speed,

    channel form, rhythm cannel of previously recorded ECGs data in record

    mode.

    Exit during printing or recording of ECGs.

    Except that, call up the menu or select menu.

    Move the selected focus to the left in menu mode

    Move the selected focus to the right in menu mode

  • C7SM-1.01 14/70

    Service Manual of Cardio 2000

    Power

    When AC power is supplied to the equipment, AC power indicator lamp at the front side is

    illuminated green and battery is being charged automatically.

    If no battery is installed, battery indicator lamp will be off. Once a battery is plugged in and

    the device is powered on, it will change its power mode to supply power to the battery. If the

    device is running on the power from the battery, battery charge indicator lamp will light up

    red color. If the battery is running low on power, three continuous buzzer alarms will sound. If

    this occurs, turn off the device and connect the A/C power before turning the device back

    on.

    - Required time for charging : more than 4 hours

    - Continuous battery operating time : 1 hour

    Caution

    For the protection of environment, do NOT throw wasted battery away. Please inquire

    at the biomedical engineering lab of the hospital and dispose according to the

    appropriate procedure to the authorized place(according to national regulations).

    Battery Charge

    Indicator Lamp

    AC Power

    Indicator Lamp

  • C7SM-1.01 15/70

    Service Manual of Cardio 2000

    1-4. System Installation

    Precautions for Installation

    While installing CardioCare2000, please pay attention to the following items:

    - Use the equipment between the ambient temperature 5 to 40℃ and humidity 10 - 90%.

    - Check the power cord is properly connected, and the probe carefully handled.

    - Do not plug multiple cords in a power outlet.

    - Install and operate unit on a flat surface.

    - If you experience noise, ground the device.

    - Do not use a power cord that may make a connection noise.

    - Device settings will be recorded in the internal memory even when it is off.

    - Prevent any shock or excessive force that may cause damage to the device.

    - Place the device away from any dust or flammable materials.

    Power Connection

    The equipment needs electrical power to operate. Plug in one end of the power cable to wall

    socket and the other to CardioCare2000.

    Patient Cable Connection

    - Connect the patient cable to the patient cable connection port on the right side of the body.

    - Connect limb electrodes to RL (N), LL (F), RA(R), and LA (L) leads of the patient cable and chest

    electrodes to V1 (C1), V2 (C2), V3 (C3), V4 (C4), V5 (C5), and V6 (C6), respectively.

    Paper Installation

    - Push the printer cover release switch to the right to open the printer door of the

    CardioCare2000. Install EKG paper with the side to be recorded appearing on top.

    Close the cover to finish the paper installation process.

    WARNING

    No modification of this equipment is allowed.

    Do not modify this equipment without authorization of the manufacturer.

    If this equipment is modified, appropriate inspection and testing must be conducted to

    ensure continued safe use of equipment.

  • C7SM-1.01 16/70

    Service Manual of Cardio 2000

    1-5. Terms of Warranty

    - This product was made through strict quality control and examination. The repair and

    compensation standards follow the consumer damage compensation regulations of the Ministry of

    Finance and Economy.

    - The warranty for this product runs for 1 year. (2 years in EU) But the warranty for accessories runs for

    6 months.

    - This product is guaranteed against faulty workmanship for a period of 1 year. Under this warranty,

    we undertake to repair free of charge at our service centers.

    - When having troubles during this period, please provide us with the model name, serial number,

    date of purchase and a description of the problem you are experiencing.

    CAUTIONS

    Federal law restricts this device to sale by or on the order of a physician

    1-6. Specification

    ECG Leads Simultaneous 12 Leads Resting ECG

    Recording Channel 3CH+1RHY, 6CH+1RHY, 12CH,

    1CH Long Time (1min)

    Gain 5, 10, 20, Auto (I~aVF: 10, V1~V6: 5) mm/mV

    Printing Speed 12.5, 25, 50 mm/sec

    Sampling Rate Analysis Sampling Rate – 500sample/sec

    Digital Sampling Rate - 8,000sample/sec

    Filters AC (50/60Hz, -20dB or better),

    Muscle (25~35Hz, -3dB or better),

    Base Line Drift (off, 0.05, 0.1, 0.2Hz, -3dB or better),

    Low Pass Filter(off, 40Hz, 100Hz, 150Hz)

  • C7SM-1.01 17/70

    Service Manual of Cardio 2000

    Display 2 x 16 Char LCD Display

    Sensitivity, Speed, Filter Status, HR, Printing Form, Rhythm Lead

    User Interface Short Key

    Printer Resolution Thermal Print Head, Roll Paper

    Report Paper

    - Width: A4/Letter – 215mm (8.5”)

    - Length: A4 – 297mm (11.7”)

    Letter – 279mm (11”)

    Resolution: 8dot/mm (0.125mm pitch)

    Patient data ID, Name, Age, Gender, Height, Weight

    Basic Measurement Heart Rate: 30~300bpm (±3bpm)

    PR, QRS, QT/QTc, P-R-T axis

    Electrical Internal Noise: 20uV(p-p)max

    Input Impedance: More than 50MΩ

    Input Voltage range: ±5mV

    CMRR: Greater than 105dB

    DC offset Voltage: ≥ ±400mV

    Patient Leakage Current: < 10uA

    Frequency Response: 0.05~150 with in –3dB

    Isolated, Defibrillation and ESU Protected

    Line Power Input: 100- 240Vac, 1.5-0.75A, 50-60Hz, Output: 15Vdc, 4.2A

    Battery type Replaceable and Rechargeable, Lithium ion, 10.8V, 3350mA

    Battery Capacity 360 minutes of continuous operation without recording or

    200 ECGs in 12 channel format at 25mm/s and 10mm/mV

    with a battery charged during approximately 3 hours from total

    discharge (with display off)

    Communication LAN

    Safety Conformity Class I, Type CF Defibrillation-proof applied Part.

    Environmental Operating Temperature: 5~40℃

    Operating Humidity: 10~90%RH

    Atmospheric Pressure: 70~106KPa

    Dimensions 290(W) 300(D) 97.5(H)mm, Approx 3.0kg (Main Body)

    Standard Accessory Patient Cable (1 EA), Limbs Electrodes (1 SET),

    Chest Electrodes (1 SET), ECG Paper (1 EA),

    Power Cord (1 EA)

    Options Rechargeable Battery (1 EA),

    Cart(1 EA), Hanger (1 EA)

  • C7SM-1.01 18/70

    Service Manual of Cardio 2000

    Section 2. Troubleshooting

    2-1. Power

    Problem Solution

    Case1.

    The EKG2000

    open will turn on

    and off

    frequently.

    Case2.

    Power will not

    turn on.

    < General Checking >

    1. Check the power cable is connected with AC Power connection at the

    back of the system.

    2. Check the power source of the multi-tab with connection to the power

    cable. (If power is connected, AC power indicator is lighted.)

    3. Check the power switch.

    4. Check any of two fuses on the Power Module (SMPS) and on fuse on the

    Base board is cut.

    If it is, replace Power board.

    Case1:

    Check if the soldering of J1 on the Base board (SMPS + Base board

    Connector) is poor. Re-solder the cable.

    ● When the EKG2000 is turned on, the pin no. 8‟s voltage has to be always

    14.6V.

    1) Is the pin no. 1/2‟s voltage is 14.6 Yes replace the main board.

    2) Is the pin no 1/2‟s voltage is 14.6 No replace the power board.

    ● When the REC key is pressed,(during printer is working)

    1) Pin no. 12‟s voltage has to be changed from 0V to 24V

    If not, faulty power board, replace the power board.

    2) Pin no. 15‟s voltage has to be changed from 3V to 0V.

    If not, faulty CPU, replace the main board.

  • C7SM-1.01 19/70

    Service Manual of Cardio 2000

    2-2. Wave Printing

    Problem Solution

    1-1. The wave pattern changes from the flat line

    on the left to the wave forms on the right.

    1. If the normal wave patterns show up to the

    right of chart as shown in problem1, 2,

    diagnosis output print too early.

    2. It takes some time to show the normal wave

    patterns.

    3. The normal wave patterns will show if output

    is requested after the system is stabilized

    by showing the HR variance of ±5 on the LCD

    display.

    4. Check connection of the limbs electrode

    and the chest electrode to the body parts.

    5. If used on the dry skin or in the winter

    season, connectivity may be degraded. In

    that case, used water or ECG gel. The

    electrode must be washed after using ECG

    gel. Otherwise, the electrode will be rusted

    and corroded.

    1-2. Some wave patterns are not shown in V1~6.

    * Connection location of the limbs electrode

    and the chest electrode

    2. The diagnosis is not shown in print.

    1. Press the menu key, showing the overlapped

    circle and triangle, at the bottom of the

    Keypad.

    2. Press the right arrow button twice to select

    PRN and then press the menu key.

    3. Press the left arrow button once to select

    DIAG and then press the menu key.

    4. Press ESC key from the menu to enter the

    basic display and then press '0' to printout.

  • C7SM-1.01 20/70

    Service Manual of Cardio 2000

    3-1. There are too many noises in the wave

    pattern.

    3-2. The noises periodically occur throughout

    the wave patterns.

    3-3. The thick noise is made on the wave

    patterns.

    3-4. The wave output looks like the noise and

    are not properly shaped.

    1. This problem occurs when the filter value is

    not corrected.

    2. Press the menu key, showing the overlapped

    circle and triangle, at the bottom of the

    Keypad.

    3. Press the right arrow key once to select FLT

    and then press the menu key.

    4. Check if the filter value shows BASE: ON,

    MUSC: OFF, AC: 60 Hz, LDF: 150Hz and, if not,

    correct it.

    5. Since the noise occurs as in problems 2 and

    3 if the AC filter is turned off or set at

    different frequency from the power source,

    the system must be set at 50Hz or 60Hz to

    correspond to the input power. (It is different

    each country) The AC filter status is shown at

    the left bottom of the printout as

    'CUT OFF (ON) AC'.

    6. Problems 3.4 occurs when the BASE is set at

    OFF. Change it to ON.

  • C7SM-1.01 21/70

    Service Manual of Cardio 2000

    4. The wave output looks like the noise and are

    not properly shaped.

    1. This problem occurs when the chest

    electrode is not properly in contact with the

    body.

    2. Check with simulator there is any cut on the

    patient cable.

    3. Check there is any damage to the chest

    electrode.

    4. Test it again after applying alcohol, water or

    ECG gel to the electrode.

    * If the problem can not be solved by the

    above procedures, please contact us.

    5. Heavy noise on the wave forms

    1. The noise occurs as there is no signal from

    the limbs electrode.

    2. If the polar plate in the limbs electrode is

    blacked or copper colored, or if ECG gel

    residue remains, the signal may not be

    detected. Replace it.

    3. If discoloring is too noticeable in the limbs

    electrode, replace it.

    6. The wave pattern shows the straight line

    among V1~6.

    1. If some wave patterns are too small as

    shown in the problem 6, the problem is most

    likely to be with the patient cable, limbs

    electrode or chest electrode.

    2. If the problem is not with the accessories,

    then the main board is the problem.

    3. The main board must be shipped to us for

    repair since the condenser of the main

    board must be replaced.

    7-1. When diagnosis output key is pressed, the

    measurement goes on past 10S.

    7-2. Operation stops at processing, print after

    10S measurement.

    1. Replace the main board.

    * If the above procedure does not solve the

    problem, please contact us

    8. The backlight on the screen is on, but there is

    no display.

    1. Check the Key cable which connects Key

    board and the main board and replace the

    main board.

  • C7SM-1.01 22/70

    Service Manual of Cardio 2000

    2-3. Diagnosis

    Problem Resolution

    1. The diagnosis shows LVH.

    1. If disease is diagnosed, abnormal ECG always

    appears.

    2. LVH occurs when the sum of one of V1 or V2

    and one of V5 or V6 is 3.5 or higher.

    3. Upgrade to the latest version of the software.

    2. Diagnosis shows MI.

    1. If MI is diagnosed, it is likely that the print button

    was pressed before the wave patterns are

    stabilized.

    2. Check HR before printout and print it again.

    3. Upgrade to the latest version of the software.

  • C7SM-1.01 23/70

    Service Manual of Cardio 2000

    2-4. Filter

    Base line Filter

    Base line drift is caused by a patient‟s breathing. ECG signal is overlapped on a large parabola. If

    you want to set Base line drift, select „ON‟. Otherwise, select OFF. The set value appears on the

    bottom part of the printed form indicating 0.1HZ- when set to ON, and 0Hz- when set to OFF.

    AC Filter

    In addition to ECG, other signals such as power noise, base line drift by respiration and

    electromyogram may be recorded in the output signals.

    There are three types of the noise filter; 60Hz, 50Hz and OFF. When set to OFF, noise is not

    eliminated. Set to 50Hz, it indicates to filter 50Hz power noise. Set to 60Hz, it means filtering 60Hz

    power noise.

    When the system is powered by battery, power noise is not generated, so you may have clear

    ECG signal if set to OFF. In Europe where the power noise is 50Hz, the system is set to 50Hz. In other

    nations including Korea and the U.S., it is set to 60Hz.

    The status of the set filter can be checked on the bottom part of the printout, indicating AC 60Hz,

    AC 50Hz or AC OFF.

  • C7SM-1.01 24/70

    Service Manual of Cardio 2000

    Low Pass Filter

    When ECG is not clear even after all three filters are set to ON, the use of low frequency filter will

    produce clear ECG signal. The low frequency filter can be set to 40Hz, 100Hz, and 150Hz and OFF.

    Set to 40Hz, it eliminates all signals over 40Hz. The set value is indicated on the lower part of the

    printed form. If you set the filter to OFF, -250Hz is indicated. If set to 40Hz, -40Hz is indicated. Set to

    100 Hz, -100Hz is indicated. If set to 150Hz, -150Hz is indicated.

    FREQUENCY 40HZ100HZ 150HZ 250HZ

  • C7SM-1.01 25/70

    Service Manual of Cardio 2000

    2-5. Trouble Shooting Guide

    1) Power is off

    The power is off

    Does the power input cable break

    down?

    Change the power input cable

    Is the fuse blown?

    Change the fuseAgain blown

    fuse

    Does the power switch cable short-

    circuit?

    Change the power switch cable

    Does the main power

    connector break down?

    Change the main power connector

    Normal operation

    UseNo

    Yes

    Yes

    Yes

    Yes No

    Yes

    B

    A

    B

    Does the main power module produce the

    power properly?

    Repair or change the main power

    moduleNo

    Yes

    Ask the manufacturer

    about it, then take action

    No

    Yes

    A

    A

    A

    A

    A

    No

    No

    No

  • C7SM-1.01 26/70

    Service Manual of Cardio 2000

    2) LCD is blank although the power is on

    LCD is blank although the power is on

    Broken 20 pin shield cable?

    Change the 20 pin shield cable

    Does the LCD cable have a poor condition of connection or short-

    circuit?

    Check the condition of the connection and then change if it's not good

    Is the LCD module

    unsatisfactory?

    Change the LCD Module

    Normal operation

    Use

    No

    Yes

    Yes

    Yes

    No

    Yes

    B

    A

    B

    Ask the manufacturer

    about it, then take action

    A

    A

    A

    No

    No

  • C7SM-1.01 27/70

    Service Manual of Cardio 2000

    3) Printer is not working properly

    When the printer is not working

    properly

    Does the paper come out while printing?

    Does the motor work properly?

    Does the signal come out on the paper while printing?

    Normal operation

    Use

    No

    No

    Yes

    B

    A

    B

    Ask the manufacturer

    about it, then take action

    Have you connected the motor power

    cable correctly?

    Change the motorNo

    Connet the motor power cable

    correctlyNo

    Have you connected the

    TPH powor cable correctly

    Connect the THP power cable

    correctlyNo

    Does you place on the

    printer correctly?

    No

    Make the paper flat and place it facing up in order to let printer

    cover shut?

    Is the motor power cable satisfactory?

    Change the motor power cable

    Is the TPH power cable is satisfactory?

    Change the TPH power cable

    Yes

    Yes

    Yes

    Yes

    Yes

    No

    No

    Yes

    Yes

    A

    A

    A

    A

    A

  • C7SM-1.01 28/70

    Service Manual of Cardio 2000

    4) EKG wave does not come out or it's very small

    EKG wave does not come out or

    small

    Is the patient cable

    connecter correctly?

    Reconnect the patient cable

    carefully

    Is the patient cable short-circuited?

    Is the electrode

    connected to the patient

    cable correctly?

    Normal operation

    UseNo

    No

    Yes

    Yes

    B

    A

    B

    Is the electrode

    connected to the patient

    cable correctly?

    A

    Change the patient cable

    A

    Cleanse the electrode with

    alcohol or water, then it use

    Reconnected the electrode to the

    patient cable properly

    Ask the manufacturer

    about it, then take action

    No A

    A

    Yes

    No

    Yes

    Control the volume by setting

    a gain on the main board

    Is the EKG signal adjusted

    properly?No

    Yes

    A

  • C7SM-1.01 29/70

    Service Manual of Cardio 2000

    5) USB memory does not working properly

    EKG files do not write to USB memory

    properly

    Is USB memory formatted in FAT file system?

    Is USB memory space full ?

    Does the USB sig cable have a poor condition of connection?

    Is USB sig cable broken?

    Format USB memory in FAT

    file system

    Remove all files in USB memory

    Check the condition of the connection and Re-

    connect the cable.

    Exchange the USB sig cable

    No

    No

    No

    Yes

    Yes

    Yes

    Yes

    No

    Ask the manufacturer about it, then take action

  • C7SM-1.01 30/70

    Service Manual of Cardio 2000

    6) USB barcode reader does not working properly

    Does USB Barcode reader not work

    properly?

    Does the USB sig cable have a poor condition of connection?

    Is USB sig cable broken?

    Check the condition of the connection and Re-

    connect the cable.

    Exchange the USB sig cable

    No

    Yes

    Yes

    No

    Ask the manufacturer about it, then take action

  • C7SM-1.01 31/70

    Service Manual of Cardio 2000

    Section 3. Adjustment

    3-1. LCD Backlight

    The LED backlight is powered by two resistors as shown in the following figure.

  • C7SM-1.01 32/70

    Service Manual of Cardio 2000

    3-2. Lead Fault

    NOTE

    Ready LED is turned off when the lead fault occurs.

  • C7SM-1.01 33/70

    Service Manual of Cardio 2000

    3-3. Filter Setting

    AC power noise, base line drift caused by breathing , EMG signal can be recorded along with

    the output ECG signal. To set the AC noise filter, move to “FLT” menu to call up “AC” menu and

    press key. key toggles between off, 50hz, 60hz. Off indicates that filter does not

    suppress ac interference. As for 50Hz or 60Hz, filter suppresses 50Hz or 60Hz AC power noise. In case

    of battery power, ECG signal is clear only with Off setting because AC noise interference hardly

    exist. The selection is 50Hz for most of Europe and 60Hz for USA and other countries. The selected

    setting is shown on the bottom line of the output form indicated AC 60Hz, AC 50Hz or AC OFF

    Base line drift is caused by patient breath. Base line drift filter can be set by pressing key.

    BASE key switches on or off. off, 0.05, 0.1 or 0.2Hz indicates base lien filter is applied or not.

    The selected setting is shown on the bottom line of the output form indicated by „0.1hz-„ as for filter

    on and „0hz-„ as for filter off.

    EMG signal is induced from muscle and internal organs of the patient.

    For clear ECG signal output, it is necessary to eliminate EMG signal interference.

    EMG filter can be set by pressing MUSC key. MUSC key switches on or off.

    On or off indicates EMG filter is applied or not. The selected setting is shown on the bottom line of

    the output form indicated „EMG:On‟ or „EMG:Off

    Low pass filter can be selected to optimize ECG signal after applying above three filters and if

    ECG signal is still interfered. To set the low pass filter, move the menu to “FLT” to call up “LPF” menu

    and press key. key toggles between off, 40hz, 100hz, 150hz. As for 40Hz, it indicates

    filter suppresses higher frequency than 40Hz. The selected setting is shown on the bottom line of the

    output form indicated by „Off„ as for off, by‟-40hz‟ as for 40hz, by‟-100hz‟ as for 100hz, by ‟-150hz‟ as

    for 150hz.

    To apply filter could result in ECG distortion. So, filter should be appropriately applied to reduce

    the fluctuation without affecting the ECG signal when output signal is badly interfered only.

    For your reference, base line file and ac filter should be always active and EMG filter can be

    applied appropriately. It is recommended for LPF to set 150Hz during diagnosis.

  • C7SM-1.01 34/70

    Service Manual of Cardio 2000

    Section 4. Block Diagram/Description

    4-1. Circuit Block Diagram

    Basic Module

    CPU Module

    Analog Module

    Character LCD

    User Interface Module

    CPU

    DDR32G x 2

    EMMC4G

    USB OTGConnector

    PMIC

    JTAGConnector

    PRINTER ASS’Y

    MCU3.3V, 5V Power

    Isolator

    Shield Driver

    ECG AFEPACE Detec-tor

    Membrane Keys Isolator

    ThermalPrinterHeader

    StepMotor

    PaperSensor

    Analog Module Connector

    Analog Module Connector

    User Interface Module Connector

    Printer Connector

    USB 1 Port

    LAN PORT(RJ45)

    Touchcontroller

    BacklightPower

    RS232Transceiver

    EthernetTransceiver

    PaperSensorDriver

    5V, 3.3V, 4.2V

    Power

    MotorDriver

    TPH SignalTransceiver

    Charger Power MCU

    System Power Switch

    Coin Battery Buzzer

    User Interface Module ConnectorPSU ConnectorBattery Connector

    Ready LED

    AC / Battery Indicator LED

    24VPrinter Power

    PrinterMCU

    Battery Cable PSU CableUser Interface Signal Cable Analog Board Cable

    Printer Cable

    Character LCD Pin Header

    Character LCD Pin Header

    Patient Connector

    Power SupplyUnit

    Battery

    Uartport

  • C7SM-1.01 35/70

    Service Manual of Cardio 2000

    4-2. Equipment Overview

    1) Top

    A : Printer Assembly

    B : LCD Assembly

    C : Control Panel

    2) Bottom

    A : Battery Cover

    A C

    B

    A

  • C7SM-1.01 36/70

    Service Manual of Cardio 2000

    3) Front

    A : Printer Cover Switch

    B : Top Cover Assembly

    C : Bottom Cover Assembly

    4) Rear

    A : Protective Ground Terminal

    B : Power Switch

    C : AC Power Connection Port

    D : USB Port (Use for equipment upgrade only)

    E : RJ45 LAN Port

    A

    C

    B

    C

    A B

    D E

  • C7SM-1.01 37/70

    Service Manual of Cardio 2000

    5) Right Side

    A : Patient Cable Connection Port

    6) Left Side

    A: Handle

    A

    A

  • C7SM-1.01 38/70

    Service Manual of Cardio 2000

    4-3. Size and Weight

    Size: 290(W) X 300(H) X 89.5(D) mm

    Weight: approx. 3 Kg

  • C7SM-1.01 39/70

    Service Manual of Cardio 2000

    Section 5. Disassembly and Assembly

    5-1. Module Description

    Base Board

    CPU Board

    Power Board

    (SMPS)

    UI Board

  • C7SM-1.01 40/70

    Service Manual of Cardio 2000

    1) UI Board

    A : Main Cable (Input Voltage)

    B : Membrane Keys Connector

    C : LCD Connector (LCD)

    D : FFC Cable (40 pin)

    It is linked to the Base Board with a Cable and consists of Membrane keys, Power Indicator LED, Battery

    Indicator LED, Ready LED.

    - Membrane Keys

    - Power Indicator LED : Show the state of connection with AC adapter and a green light indicator

    when plugging AC adapter into device.

    - Battery Indicator LED : A light indicator lets you know that there is battery and show your current

    battery charge strength. A red light indicator lets you know that the battery is charging and a

    green light indicator lets you know the batter is fully charged.

    - Ready LED : It Indicate the lead connection status.

    A

    B

    C

    D

  • C7SM-1.01 41/70

    Service Manual of Cardio 2000

    2) Base (Power + Analog) Board

    A : USB Port

    B : LAN Port

    C : Switch Connector

    D : Input DC Connector (SMPS Output)

    E : Voltage Regulator (3 ea)

    F : LCD Data Connector

    G : Motor Driver

    H : Main Connector

    I : Transformer

    J : Isolation IC (2 ea)

    K : Patient Cable Connector

    L : MCU (3 ea)

    The digital part consists of MCU, DC/DC Converter, DC/AC Controller and chip, which are linked

    to other external modules.

    A

    B

    C D E F

    G

    H

    J I K

    L

  • C7SM-1.01 42/70

    Service Manual of Cardio 2000

    3) CPU Board

    A : MCU

    B : DDR3 SDRAM

    C : Power Management

    D : eMMC Flash

  • C7SM-1.01 43/70

    Service Manual of Cardio 2000

    4) Power Module (MDS-065APS15-BA)

    It receives 100-240V AC, 50-60Hz signal or battery and supplies 15V to the Base board.

    5) Printer

    It consists of the TPH (Thermal Printer Head), motor, and paper sensor. And it is linked to the Base

    board with 3 pin, 6 pin, and 14 pin cable and approves 24V AC power which will be approved only

    if the printer assembly is working.

  • C7SM-1.01 44/70

    Service Manual of Cardio 2000

    6) Battery

    It is linked to the Digital board with a 3 pin cable. It is 10.8V, 3350mA/h and consists of three 4.2V,

    3350mA/h Li-ion Rechargeable battery. It is packed in parallel with 2 sets which are connected with

    3 cells in series and 1 spacer.

    The cables of battery consist of 3 pins connector which are VCC, GND, GND (Battery Check) as

    follow picture. GND (Battery Check) cable should be connected to the board, because the cable

    is for checking the status of battery.

    GND

    GND (Battery Check) VCC

  • C7SM-1.01 45/70

    Service Manual of Cardio 2000

    7) LCD

    Character LCD : It connects to CDUI Board LCD with 14 pin Pin Header. It consists of LCD.

  • C7SM-1.01 46/70

    Service Manual of Cardio 2000

    5-2. Internal Block Diagram & Cable

    Power

    Managing

    MCU

    Lithium Ion

    Battery

    Power

    Supply

    Unit

    System

    5V Regulator

    System

    3.3V Regulator

    PMIC

    USB

    Battery

    Green

    LED

    Battery

    Orange

    LED

    AC

    Green

    LED

    Power

    Button

    24V

    Printer Power

    PM

    3.3V LDO

    Printer

    MCU

    VSYS(15V, 9V~12.6V)

    VLINE(15V)

    VBAT(9V~12.6V)

    SYS_5V

    SYS_3.3V

    System PowerControl

    PowerState

    Signals

    BatteryLevel

    BatteryTemp.

    User interface

    PrinterPowerControl

    VPM_3.3V

    Battery

    Charger

    LCD

    Power

    CPU

    VCPU

    Analog Isolator

    Charger State/Control

    BatteryConnection

    System

    Power

    Switch

    VSYS_SW(15V, 9V~12.6V)

  • C7SM-1.01 47/70

    Service Manual of Cardio 2000

    2) Cable

    LCD Assembly

    40Pin FFC Cable

    Battery Cable Assembly

  • C7SM-1.01 48/70

    Service Manual of Cardio 2000

    QN-TPH-Cable

    Motor Assembly

    Power Switch Cable Assembly

  • C7SM-1.01 49/70

    Service Manual of Cardio 2000

    5-3. Disassembly

    Step1. Removing the Top cover

    a. Removing the 5 screws from the bottom cover.

  • C7SM-1.01 50/70

    Service Manual of Cardio 2000

    b. Removing the Top cover.

    1) Open the printer cover by pushing slide knob.

    2) Disconnect 40 pin FFC cable and Main Cable.

    3) Remove Top cover as shown below.

    • Pull 40 pin FFC Cable from the connector of Base board.

  • C7SM-1.01 51/70

    Service Manual of Cardio 2000

  • C7SM-1.01 52/70

    Service Manual of Cardio 2000

    Step2. Replacing the Printer Assembly

    a. Remove the 4 screws from the bottom cover.

    b. Disconnect the 3 printer cables from the Base Board.

  • C7SM-1.01 53/70

    Service Manual of Cardio 2000

    Step3. Replacing the Base Board

    a. Remove the 9 screws from the bottom cover.

  • C7SM-1.01 54/70

    Service Manual of Cardio 2000

    b. Disconnect the power connector and the battery cable.

    c. Disconnect the power switch cable and ground cable.

  • C7SM-1.01 55/70

    Service Manual of Cardio 2000

    Step5. Replacing the Power Module (SMPS)

    a. Disassemble Ground cable of Power Board (SMPS) and AC Inlet Ground cable from

    Protection Grounding Terminal.

    b. Remove the 4 screws from SMPS Bracket.

  • C7SM-1.01 56/70

    Service Manual of Cardio 2000

    Step6. Replacing the Key Board

    a. Remove the 4 screws from Inside of Top cover

  • C7SM-1.01 57/70

    Service Manual of Cardio 2000

    b. Remove the 4 screws and disconnect the 40Pin SMD FFC cable, Main cable, Membrane

    Keys cable.

    Step7. Replacing the LCD

    a. Remove the 4 screws from LCD bracket

  • C7SM-1.01 58/70

    Service Manual of Cardio 2000

    Step8. Replacing the Battery

    a. Remove the 2 screws from Bottom cover

    b. Take out Battery from Bottom cover

    c. Remove the Battery cable connector

  • C7SM-1.01 59/70

    Service Manual of Cardio 2000

    5-4. Assembly

    Assemble the reverse way from Step.8 to Step.1

    * Note

    Please check whether two parts which is circled in red (rubber handle & printer lock) are

    connected correctly.

  • C7SM-1.01 60/70

    Service Manual of Cardio 2000

    Section 6. Replaceable Parts List and Accessory

    6-1. Replaceable Part List

    < Top Cover Ass’y>

  • C7SM-1.01 61/70

    Service Manual of Cardio 2000

  • C7SM-1.01 62/70

    Service Manual of Cardio 2000

  • C7SM-1.01 63/70

    Service Manual of Cardio 2000

    Section 7. Connector Pin Assignment

    7-1. Patient Cable Port

    Pin Description

    Pin Name

    1 V2

    2 V3

    3 V4

    4 V5

    5 V6

    6 GND

    7 N.C

    8 N.C

    9 RA

    10 LA

    11 LL

    12 V1

    13 N.C

    14 RL

    15 N.C

  • C7SM-1.01 64/70

    Service Manual of Cardio 2000

    Section 8. PCB Part Disposition

    8-1. Base (Power+Analog) Board

  • C7SM-1.01 65/70

    Service Manual of Cardio 2000

    8-2. CDUI Board

  • C7SM-1.01 66/70

    Service Manual of Cardio 2000

    8-3. CPU Board

    < Bottom>

  • C7SM-1.01 67/70

    Service Manual of Cardio 2000

    Section 9. Software Upgrade

    9-1. Requirement of Upgrading

    [ AdvancePreparation ]

    1) USB memory (*Note 1)

    2) Unzip the bionet.zip file and copy to USB memory the bionet.zip file.

    3) Copy to bionet_ekg2000\packages folder the bup file.

    * Note 1: File system of USB memory must be the FAT32 or FAT .

  • C7SM-1.01 68/70

    Service Manual of Cardio 2000

    9-2 Procedure of Upgrading

    (Cardio7 UI screen, upgrade method is same)

    1) Connect to the usb port that upgrade software files are copied to a usb memory.

    Setup Path

    - key → SYS → UPGRADE

    2-1) Enter the Password

    2-2) Run Upgrade Program that check the password.(*Note 1)

    * Note 1 :PASSWORD: Products that display the current date (MMDD) .

    - key → SYS → DATE

    - ex) Basic → Date : Aug, 25 → Password : 0825

  • C7SM-1.01 69/70

    Service Manual of Cardio 2000

    3-1) Select the upgrade file.

    3-2) Press key.

    4-1) Select the APP(version) → Press key.

    4-2) Run to upgrade.

    5) When the upgrade is complete, turn off and on.

  • C7SM-1.01 70/70

    Service Manual of Cardio 2000

    International Sales & service

    Bionet Co., Ltd. :

    5F, Shinsegae I&C Digital Center 61 Digital-ro 31 gil,

    Guro-gu, SEOUL 08375, REPUBLIC OF KOREA

    Tel : +82-2-6300-6410 / Fax : +82-2-6499-7789 / e-mail: [email protected]

    Website: www.ebionet.com

    U.S.A sales & service representative

    Bionet America, Inc. :

    2691, Dow Ave, Suite B

    Tustin, CA92780U.S.A.

    Toll Free: 1-877-924-6638 / Fax: 1-714-734-1761 / e-mail: [email protected]

    Website: www.bionetUS.com

    European sales & service representative

    MGB Endoskopische Gerä te GmbH Berlin :

    Schwarzschildstraße 6

    D-12489 Berlin, Germany

    Tel: +49(0)-30-6392-7000 / Fax: +49(0)-30-6392-7011 / e-mail: [email protected]

    Website: www.mgb-berlin.de

    Bionet Co., Ltd

    Model Name: EKG2000

    http://www.ebionet.com/http://www.bionetus.com/http://www.mgb-berlin.de/