3
Function
The radiation thermometer CellaTemp
PK detects the infrared radiation emit-
ted by an object and converts it into
an electric signal. The detected tem-
perature is displayed and transmittedto the analogue output for further pro-
cessing.
A unique combination of analogue
and digital linearization features
provides the CellaTemp PK with a
high-resolution signal processing unit.
Therefore, even with wide measuring
ranges, the sensor has a very high
temperature resolution while its noise
equivalent temperature difference is
extremely low. The pyrometer thus
supplies stable measurement read-ings even when the response times
are extremely short and the measured
temperatures are very low.
A microprocessor with high clock
speed allows for short response times
starting at 2ms.
The very bright display unit is equipped
with state-of-the-art technology. Even
from a great distance it is easy to read
and it is energy-efcient at the same
time.
Analogue output
The analogue output supplies a signal
linear to the temperature; 0/4 20 mA
are optionally available. The range
setting can be congured with the
control keys.
The analogue output deactivates and
a warning appears on the display
when the internal temperature reach-es 75 C.
M30 thread
Robust lens with
anti-refex coating
7- segment
LED display
3 control keys
LED for C / F
Plug M12
LED status switching
Switching output
The presence of a hot object in the
sensors eld of vision triggers a
switching operation when a pre-de-
ned temperature threshold is ex-
ceeded. An LED indicates this switch-ing status.
Optionally, the output can operate as
a normally closed or normally open
contact.
This congurable switch on/switch off
delay allows for a suppression of short
interference pulses and for an ideal
adaptation of the switching output to
the response time of a PLC.
Optical system
The optical system of the CellaTemp
PK is equipped with lenses with an-
ti-reective coatings. This makes the
lenses very durable, easy to clean
and perfectly suitable for rough indus-
trial use.
The excellent imaging characteristics
of the precision lenses provide a high
optical resolution and a minimum sen-
sitivity to stray light. This keeps themeasurement reading stable even
when object sizes and distances
change.
Test funct ion
A functional test of the device and
the signal processing unit can be
performed at any time via an exter-
nal control signal or via the operating
menu.
The analogue output generates an
electric current of 20.5 mA and theswitching function is triggered.
Service function
The service function is used during
setup or running operation to apply
a simulated temperature value that is
displayed and transmitted via the ana-
logue output. This feature checks the
correct functioning and range setting
for the downstream signal processing
units (display, controller, PLC) quicklyand safely even without a hot object.
Diagnostic function
The diagnostic function ensures a
high operational reliability. Incorrect
supply voltages, reverse polarities of
connections, overloads at the switch-
ing output, unacceptable ambient
temperatures or out-of-range object
temperatures appear on the displayas error messages.
Laser module
During setup of the CellaTemp PKF
26/36 the laser module PS 01/P is
mounted to the end of the bre optic
cable to adjust the sensors focussing
range and to perform an alignment to
the target.
-
8/10/2019 CellaTemp PK 201410 En
4/12
4
Technical data
Resolution
Current output
0.2 K + 0.03 % of selected range
Display
0.1 K for T < 200 C
1 K for T 200 C
Voltage supply
18 ... 32 V DC
Current consumption
50 mA at 24 V DC
without load current
Ambient temperature
0 - 65 C
Storage temperature
-20 - +80 C
Permissible humidity95 % r.H. max.
(non-condensing)
Protection
IP65 according to DIN 40050
Protection class III
Housing material
V2A (1.4305)
Connectivity
M12 thread 5-pin connectionA-coding (DIN EN 61076-2-101)
Weight
approx. 0.4 kg
Approvals
EMC
EN 61000-6-4
EN 61000-6-2
Shock resistance(EN 60068-2-27)30 g (11 mg)
Vibration resistance(EN 60068-2-6)
5 g (10...2000 Hz)
Outputs
Analogue output
0/4 - 20 mA linear according to
NAMUR 43, scalable
Load max. 500
Switching output
PNP open collector active from
supply voltage (positive)NC or NO contact
Current rating150 mA
timed safety shutdown at
overload 250 mA
Reverse polarity p rotection
for supply voltage
for analogue output
for switching output
Test input
Digital inputTyp 3 (IEC 61131-2)
Low level 5 V DC
High level 11 V DC
Load current 11.6 mA at 30 V DC
Display
LED display
4 x 7 segments, red
Height of digits 8 mm
1 x LEDDisplay of switching status,
overload and incorrect connection
of supply voltage
2 x LED
Display of temperature unit
(C or F)
Operating elements
3 control keys
Fibre optics cable forCellaTemp PKF
Other lengths up to 50 m on request
Ambient temperature-40 ... +250 C
Material
Nickel-plated brass
Type Length Weight
LWL-2HT 2 m 0.08 kg
LWL-5HT 5 m 0.19 kg
LWL-10HT 10 m 0.38 kg
Adjustable parameters
Analogue output
0/4 - 20 mA
Scaling of analogue output
Switching output
Switching point and reset point
Switching function:
NC / NO contactSwitch on/switch off delay
General parameters
Quick emissivity setting via
temperature display
Damping time constant
Hold time of peak picker
Reset to factory default setting
Key lock
Display
Temperature unit C / F
Temperature simulation
Test function
Error diagnostic
Output overload
Sensor excess temperature
Out-of-range (high/low)
Incorrect supply voltage connection
Incorrect supply voltage
Pin assignment
! It is imperative to use
a shielded cable
Shipment includes
Radiation thermometer
Operating manual
2 Mounting nuts
Additionally with CellaTemp PKF
Measuring head
Laser module PS 01/P
Fibre optic cable
(state required length)
! Order the connecting cable
VK 02/L separately in the
desired length.
-
8/10/2019 CellaTemp PK 201410 En
5/12
5
Models and technical data
TypeTemp.
range
Spectral
range
Focus
rangeSpot size Precision*
Response
time t 98
Repeat-
ability
Temperature
coefcient
Compact pyrometer with xed focus lens
PK 11 AF 1 0 - 1000 C32 - 1832 F
8 - 14 m
0.3 m 11 mm
0.75% of temp.
reading [C]
plus 2.0 K
60 ms
1 K
0.1 K/K
(fr T < 250 C)
0.04 %/K
(fr T > 250 C)
PK 12 AF 1-30 - 300 C
-22 - 572 F0.3 m 19 mm 90 ms
PK 14 AF 1 0 - 500 C
32 - 932 F1.0 m 0.5 m 60 ms
PK 18 AF 1 0 - 500 C
32 - 932 F0.3 m 11 mm 60 ms
PK 21 AF 1250 - 1600 C
482 - 2912 F
1.0 -
1.7 m1.5 m 12 mm
0.3% of temp.
reading [C]
plus 2.5 K
2 ms for
T > 600 C
1 K 0.07 %/K
PK 29 AF 1150 - 800 C
302 - 1472 F
1.8 -
2.2 m0.3 m 7.5 mm
0.5% of temp.
reading [C]
plus 2.5 K
2 ms for
T > 300 C
PK 31 AF 1500 - 2500 C
932 - 4532 F
0.78 -
1.06 m1.5 m 9 mm
0.2% of temp.
reading [C]
plus 2.5 K
2 ms for
T > 900 C1 K 0.07 %/K
PK 41 AF 1 300 - 1300 C572 - 2372 F4.6 -
4.9 m0.4 m
12 mm
0.5% of temp.
reading [C]plus 2.5 K
100 ms
2 K
0.04 %/K
PK 42 AF 1500 - 2500 C
932 - 4532 F 8.5 mm
1.0% of temp.
reading [C]4 K
PK 51 AF 1400 - 1400 C
752 - 2552 F
3.8 -
4.0 m0.4 m 12 mm
1.0% of temp.
reading [C] 100 ms 2 K 0.04 %/K
Pyrometer with bre optics cable and focusable measuring head
PKF 26 AF 1 300 - 1600 C572 - 2912 F
1.0 -1.7 m
0.2 m ...
180 : 10.3% of temp.
reading [C]
plus 2.5 K
2 ms forT > 600 C
2 K 0.07 %/K
PKF 36 AF 1 550 - 2500 C
1022 - 4532 F
0.78 -
1.06 m190 : 1
2 ms for
T > 900 C
*(at e= 1 and Ta = +23 C)
-
8/10/2019 CellaTemp PK 201410 En
6/12
6
Dimensions
Target d iagram1
Compact pyrometer
CellaTemp PK 11 / 18
CellaTemp PK 21CellaTemp PK 14
Pyrometer with bre optics cable
CellaTemp PK 12
Target diameter [mm]
Target distance [mm]
15 13 11 20 29 39 48 57 66 75 167 259 350 442
100 200 300 400 500 600 700 800 900 1000 5000400030002000
Target diameter [mm]
Target distance [mm]
100 200 300 400 500 1000 1500 2000 2500 3000 5000450040003500
12 12 12 12 12 12 12 20 28 36 44 52 60 68
Target diameter [mm]
Target distance [mm]
100 200 300 400 500 600 700 800 900 1000 2000 3000 4000 5000
50 100 150 200 250 300 350 400 450 500 1000 1500 2000 2500
Target diameter [mm]
Target distance [mm]
18 18,5 19 31 44 56 68 81 93 105 230 350 475 600
100 200 300 400 500 600 700 800 900 1000 2000 3000 4000 5000
Lengths of compact pyrometers
Type Length
PK 11
183 mm
PK 12
PK 14
PK 18
PK 21
208 mmPK 29
PK 31
PK 41
199 mmPK 42
PK 51
30
M
x1,5
67,9 mm
34,5 mm
-
8/10/2019 CellaTemp PK 201410 En
7/12
7
Target d iagram1
CellaTemp PK 42CellaTemp PK 41 / 51
CellaTemp PK 31
CellaTemp PKF 26 2 CellaTemp PKF 36 2
CellaTemp PK 29
Target diameter [mm]
Target distance [mm]
16 13 11 8.5 15 22 28 35 42 48 115 181 247 313
100 200 300 400 500 600 700 800 900 1000 2000 3000 4000 5000
Target diameter [mm]
Target distance [mm]
100 200 300 400 500 600 700 800 900 1000 2000 3000 4000 5000
17 16 14 12 20 28 35 43 50 58 134 209 285 360
Target diameter [mm]
Target distance [mm]
100 200 300 400 500 1000 1500 2000 2500 3000 5000450040003500
14,6 14,2 13,8 13,4 13 11 9 17 25 33 41 49 57 65
D = 180:1 2 6 11 17 22 28 33 39 44 50 56
Target distance [mm]
Target diameter [mm]
1000 2000 3000 4000 5000 6000 7000 8000 9000 100000
D = 190:1 2 5 10.5 16 21 26 31.5 37 42 47 52.5
Target distance [mm]
Target diameter [mm]
1000 2000 3000 4000 5000 6000 7000 8000 9000 100000
Target diameter [mm]
Target distance [mm]
17,5 14 10,5 21 32 42 53 63 74 84 190 280 400 505
100 200 300 400 500 600 700 800 900 1000 5000400030002000
1Taking into account the lens tolerances at 90% of the maximum amount of radiated energy.
2The specied target spot diameter is valid for the specied distance.
An additional lens can be mounted onto the pyrometers lens
system to capture smaller target spots.
Supplementary lenses
Pyrometer PK
Supplementary lens
PyrometerSupplemen-
tary lens
Distance
[mm]
Spot size
[mm]
PK 11 PK 11/E AF 1 45 2
PK 21 / 31 PK 21/E AF 1 520 5
-
8/10/2019 CellaTemp PK 201410 En
8/12
-
8/10/2019 CellaTemp PK 201410 En
9/12
9
Typical assemblies
Mounting assembly PK 01-007
consisting of:
Pyrometer PKAir purge PS 01/A
Supplementary tube ZA 01/Q-35
Clamping collar with angle
PS 11/K-35 AF2
Mounting assembly PK 01-006
consisting of:
Pyrometer PK Thermal insulating tube PS 01/K
Air purge PS 01/A
Tube cap ZA 01/A
Intermediate tube ZA 01/M
Dust stop ZA 01/C
Ball ange ZA 01/D
Clamp ZA 01/E
Flange ZA 01/I
Mounting assembly PK 21-001
with protective lens
consisting of: Pyrometer PK
Thermal insulating tube PS 01/K
Quartz protective lens PS 01/I AF2
Air purge PS 01/A
Tube cap ZA 01/A
Intermediate tube ZA 01/M
Dust stop ZA 01/C
Ball ange ZA 01/D
Clamp ZA 01/E
Flange ZA 01/I
Mounting assembly PK 21-002
with bayonet coupling and protective
lens
consisting of:
Pyrometer PK
Thermal insulating tube PS 01/K
Bayonet coupling PS 11/N AF4
Quartz protective lens PS 01/I AF2
Air purge PS 01/A AF2
Dust stop ZA 01/C Intermediate tube ZA 01/M
Ball ange ZA 01/D
Clamp ZA 01/E
Flange ZA 01/W
Mounting assembly PK 01-011
consisting of:
Pyrometer PK
Air purge PS 01/A
Reduction 1 1/4" to M30x1,5
Flange DN 50
Oscillating mirror PZ 20/X
An oscil lating mirror can be mounted
to the pyrometer. It allows to scan the
target object along a line.
Use the analogue output in conjuctionwith a RS-422 interface to collect tem-
perature readings together with target
position.
With the included PC software CellaS-
can the temperature prole is shown
online.
-
8/10/2019 CellaTemp PK 201410 En
10/12
10
Appl ications
Asphalt and concrete mixing Crystal growing Coke oven
Wires, tubes and rods Filaments / Metal bands Runner
Incineration plant Rolling mill Blast furnace / stove dome
Power plant Continuous casting Rotary kiln
Sinter plant Glass production Induction heating
-
8/10/2019 CellaTemp PK 201410 En
11/12
11
Other products
Portix Series
Almost indestructible, portable
pyrometers with rugged aluminum
housing for universal, rapid
measurement.
CellaTemp PA
Versatile pyrometers with focusable
lens, through-the-lens sighting/ laser
spotlight or video camera.
Mikro PV
Intensity comparison pyrometer for
ultra accurate measurement.
CellaSwitch
Compact infrared switch with LED
display and auto-diagnostics.
CellaPort Series
Portable one- and two-colour-
pyrometers with through-the-lens
sighting, laser pilot light and
USB port.
CellaTemp PA-LWL
Versatile ber optics pyrometers with
focusable head and laser spotlight.
CellaTemp PM
Ultra miniature pyrometers
for very small spaces.
CellaSwitch-LWL
Compact ber optics infrared
switch with LED display and auto-
diagnostics.
CellaTemp PZ
Probus pyrometers with focusable
lens, through the lens sighting or
laser spotlight.
Since 1968, the division MSR ITS
of KELLER HCW GmbH devel-
ops and produces precise instru-
ments and system solutions fornon-contact temperature meas-
urement.
With the very large product range
of more than 200 models MSR
ITS offers instruments and sys-
tem solutions for the different
measuring tasks.
A worldwide network of distribu-
tors and service centers provides
professional expertise and servic-
es.
Subjecttotechnicalchanges2014/10
-
8/10/2019 CellaTemp PK 201410 En
12/12
KELLER HCW GmbHDivision MSR
Carl-Keller-Strae 2 10
49479 Ibbenbren
Germany
www.keller-msr.com
Tel. +49 (0) 54 51 85 0
Fax +49 (0) 54 51 89 73 92
Distribution and Service Centres
China
www.keller-msr.cn
Tel. +86 (10) 82 86 79 20
France
www.keller-msr.fr
Tel. +33 (0) 9 51 45 30 50
India
www.earthtekniks.com
Tel. +91 (44) 27 46 37 47
Russiawww.keller-msr.ru
Tel. +7 (495) 664 27 84
Foreign Distributors