. ver:E.00.00 Agilent Technologies UK Ltd 610 Wharfedale Road, Winnersh Triangle, Wokingham, BERKSHIRE RG41 5TP AGILENT TECHNOLOGIES INTERNAL ASSESSMENT PROGRAM : AQ-SSU-12/95 Tel +44(0)118 9276201 Fax: 0118 9276855 Certification Cover Sheet Accredited Calibration Order Number: 1-1234567890-1 of Page 1 2 Manufacturer: Agilent Technologies, Inc. Description: System Multimeter Model Number: 3458A Options Installed: 001,002,H01 Serial Number: 1234A56789 Customer Asset No. Customer: Location of Calibration: Agilent Technologies Sales & Services GmbH & Co.KG Agilent Technologies UK Ltd Herrenberger Strasse 130 610 Wharfedale Road, Winnersh Triangle, 71034 Boeblingen Wokingham, BERKSHIRE RG41 5TP GERMANY Tel +44(0)118 9276201 Fax: 0118 9276855 Procedure: LF1002UK 2 Mar 2011 Date of Calibration: Customer PO Number: (23 +/-1) °C Temperature: Humidity: (45 +/-10)% RH THIS CERTIFICATE COVER SHEET IS COMPLEMENTARY TO THE ACCREDITED CALIBRATION CERTIFICATE. Conformance with the requirements of ISO/IECgilent is accredited to comply with ISO/IEC 17025:2005 and our quality system is registered to ISO9001:2008. The product was tested in accordance with procedures assessed by the accreditation body and the Remarks or Special Requirements: This certificate is composed of 2 pages containing a summary of calibration information. Based on the manufacturer recommended or user requested calibration interval, the next calibration is due on 2 Mar 2012 . The User should determine the suitability of this instrument for its intended use. 2 Mar 2011 Issue date: Andy Clark - North Europe Support Mgr

Upload: others

Post on 15-May-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Certification Cover Sheet - Keysight · CERTIFICATE NUMBER. 1-1234567890-1A. Page 2 of 6 pages ... Any quoted uncertainty refers only to the measured value and does not carry any

.

ver:E.00.00

Agilent Technologies UK Ltd

610 Wharfedale Road, Winnersh Triangle,

Wokingham, BERKSHIRE RG41 5TP AGILENT TECHNOLOGIES INTERNAL ASSESSMENT PROGRAM : AQ-SSU-12/95 Tel +44(0)118 9276201 Fax: 0118 9276855

Certification Cover Sheet Accredited Calibration

Order Number: 1-1234567890-1

ofPage 1 2

Manufacturer: Agilent Technologies, Inc. Description: System MultimeterModel Number: 3458A Options Installed: 001,002,H01Serial Number: 1234A56789 Customer Asset No. Customer: Location of Calibration:

Agilent Technologies Sales & Services GmbH & Co.KG Agilent Technologies UK LtdHerrenberger Strasse 130 610 Wharfedale Road, Winnersh Triangle,71034 Boeblingen Wokingham, BERKSHIRE RG41 5TPGERMANY Tel +44(0)118 9276201 Fax: 0118 9276855

Procedure: LF1002UK2 Mar 2011Date of Calibration: Customer PO Number:(23 +/-1) °CTemperature: Humidity: (45 +/-10)% RH

THIS CERTIFICATE COVER SHEET IS COMPLEMENTARY TO THE ACCREDITED CALIBRATION CERTIFICATE. Conformance with the requirements of ISO/IECgilent is accredited to comply with ISO/IEC 17025:2005 and our quality system is registered to ISO9001:2008. The product was tested in accordance with procedures assessed by the accreditation body and the

Remarks or Special Requirements:

This certificate is composed of 2 pages containing a summary of calibration information.

Based on the manufacturer recommended or user requested calibration interval, the next calibration is due on 2 Mar 2012 . The Usershould determine the suitability of this instrument for its intended use.

2 Mar 2011Issue date: Andy Clark - North Europe Support Mgr

nesaydin
Text Box
THIS CERTIFICATE COVER SHEET IS COMPLEMENTARY TO THE ACCREDITED CALIBRATION CERTIFICATE. Conformance with the requirements of ISO/IEC 17025:2005 for testing and calibration laboratories is proven by the accreditation of our calibration laboratory. Our quality management system is ISO 9001:2008 certified.
nesaydin
Text Box
This certificate cover sheet is composed of two pages and forms an integral part of the accredited calibration certificate. The calibration document is retained for a period of six years.
Page 2: Certification Cover Sheet - Keysight · CERTIFICATE NUMBER. 1-1234567890-1A. Page 2 of 6 pages ... Any quoted uncertainty refers only to the measured value and does not carry any

.

ver:E.00.00

Agilent Technologies UK Ltd

610 Wharfedale Road, Winnersh Triangle,

Wokingham, BERKSHIRE RG41 5TP AGILENT TECHNOLOGIES INTERNAL ASSESSMENT PROGRAM : AQ-SSU-12/95 Tel +44(0)118 9276201 Fax: 0118 9276855

Certification Cover Sheet Accredited Calibration

Order Number: 1-1234567890-1

ofPage 2 2

Traceability Information:

Technician ID Number: 123456Traceability is to national standards Euramet members (NPL, PTB, LNE, etc), or other recognized standards laboratories.

Supporting documentation relative to traceability is available for review by appointment.

This certificate shall not be reproduced, except in full, without prior written approval of the laboratory.

Calibration Equipment Used:

Certificate Number:Model Number: Model Description: Trace Number: Cal Due Date:

4000A Calibrator / Transconductance AM UK5802 10 Jun 2011 1-2803087793-1100M Resistor Box UK7709 7 Sep 2011 1-2630462444-13458A System Multimeter UK11880 17 Apr 2011 1-2981577422-111667A Power Splitter UK8490 28 Jul 2012 1-2163489050-14200 Datron 4200 UK5906 11 Mar 2011 1-2828232811-11613 100 Ohm.Resistor. UK4216 29 May 2011 1-2311243546-15685A 10 Ohm Std Resistor UK7677 19 Apr 2011 1-2252952521-1ZOD80 1MOhm Resistor UK4276 18 May 2012 1-2518547562-180010 100kOhm Resistor UK4275 29 May 2011 1-2311237589-11614 1kOhm Resistor UK4338 11 Jul 2011 1-2350731803-11G Resistance Std UK3464 19 Aug 2011 1-2162055572-1732A DC Ref Std UK8510 28 Apr 2011 1-2730351669-15615 10MOhm Std Resistor UK11892 18 May 2012 1-2476830285-15685B 10kOhm Std Resistor UK12795 11 Aug 2011 1-2064606753-111050A 1V, 50 Ohm Thermal Converter UK3075 26 Aug 2013 1-2633902970-1RS3 Standard Resistor UK5561 29 May 2011 1-2311243667-13325A Function Generator UK7740 18 Aug 2012 1-3073411537-111049A 3V Thermal Converter UK4003 4 Aug 2012 1-2183410267-15685A 10 Ohm Std Resistor UK5564 19 Apr 2011 1-2257663805-1

Page 3: Certification Cover Sheet - Keysight · CERTIFICATE NUMBER. 1-1234567890-1A. Page 2 of 6 pages ... Any quoted uncertainty refers only to the measured value and does not carry any

Date Of Issue: 2 March 2011

Instrument Description: Digital Multimeter

Instrument Manufacturer: Hewlett-Packard

Model Number: 3458A

Serial Number: 1234A56789

Our Reference Number: 1-1234567890-1

Customer: Agilent Technologies GmbH, Böblingen, Deutschland

1-1234567890-1A

NUMBER

Fax: +44 (0)118 927 6855

CERTIFICATE

0147

Page 1 of 6 pages

Frank Shelton

APPROVED SIGNATORY

CERTIFICATE OF CALIBRATION

sA

Agilent Technologies UK Ltd.610 Wharfedale Road

Winnersh Triangle

RG41 5TP

WOKINGHAM

Berkshire

Tel: +44 (0)118 927 6201

Permanent lab Mobile lab

Test location

Customer's building

Customer: Agilent Technologies GmbH, Böblingen, Deutschland

Date Of Calibration: 2 March 2011

Ambient Temperature: 23 ± 1 ° C Relative Humidity: 45 ± 10 %

REMARKS:

This certificate records the on-receipt calibration status of the instrument, the item was returned in this condition.If any measured value exceeds the specification, "Out-of-spec" is shown in the Status column.

If a measured value plus its uncertainty exceeds the specification, "Guardband" is shown.Note: Measurements shown in italics are not UKAS accredited but are included for completeness of the

calibration report. Where the uncertainty is not stated, it has not yet been determined.

The testpoints are those established by the manufacturing division's standards laboratory for the

special "Gold Cal". Results of the calibration status at shipment are available as an ASCII format datafile.

TEST DESCRIPTION:The 90-day specification has been used to define the absolute test limits in this certificate. They include the

original manufacturer's calibration uncertainty for NIST traceability as published in the product's

Calibration Manual 03458-90017:Dec2000.

A minimum warm-up period of 4 hours was allowed before testing. In accordance with themanufacturer's recommended practice, the appropriate "ACAL" routines were invoked before testing each

function. The Firmware version reported by REV? was 8.2

The value for CALNUM? reported by the 3458A PRIOR to testing was 55

and at COMPLETION of testing was 55

This certificate is issued in accordance with the laboratory accreditation requirements of the United Kingdom Accreditation Service. It provides traceability of

measurement to recognised national standards, and to the units of measurement realised at the National Physical Laboratory or other recognised national

standards laboratories. This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory.

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.

The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of confidence of approximately 95%.

Permanent lab Mobile lab

Test location

Customer's building

standards laboratories. This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory.

Permanent lab Mobile lab

Test location

Customer's building

nesaydin
Stamp
Page 4: Certification Cover Sheet - Keysight · CERTIFICATE NUMBER. 1-1234567890-1A. Page 2 of 6 pages ... Any quoted uncertainty refers only to the measured value and does not carry any

DC Voltage Accuracy

Last calibration internal temperature (Cal? 59) = 36.97 ºC

Present internal temperature (Temp?) = 35.00 ºC

3458A set-up: NPLC 500, NDIG 8

Zero Offsets

A low thermal short was connected to the input terminals.

Applied Range Specification Measured Value Uncertainty Status

(applied ± limit) (if applicable)

0 mV 100 mV ± 0.9 µV -0.00009 mV ± 0.4 µV

0 V 1 V ± 1.0 µV -0.00000087 V ± 0.4 µV Guardband

0 V 10 V ± 2.0 µV -0.0000006 V ± 0.4 µV

CERTIFICATE OF CALIBRATION

UKAS Accredited Calibration Laboratory No. 0147

Applicable Specification :

sACERTIFICATE

NUMBER

1-1234567890-1A

Page 2 of 6 pages

Standard

Option 002

Gain

The following measurements used the multimeter's "Math Null" function to exclude thermal emf's

and other internal offsets from the measured value.

Applied Range Specification Measured Value Uncertainty Status

(applied ± limit) (if applicable)

100 mV 100 mV ± 0.85 µV 99.99999 mV ± 0.25 µV-100 mV 100 mV ± 0.85 µV -100.00006 mV ± 0.25 µV

1 V 1 V ± 5.4 µV 0.99999901 V ± 1.0 µV

-1 V 1 V ± 5.4 µV -0.99999950 V ± 1.0 µV

10 V 10 V ± 46.5 µV 9.9999946 V ± 10.0 µV-10 V 10 V ± 46.5 µV -9.9999942 V ± 10.0 µV

100 V 100 V ± 680 µV 99.999843 V ± 100 µV

-100 V 100 V ± 680 µV -99.999838 V ± 100 µV

1000 V 1000 V ± 18.6 mV 1000.00199 V ± 1.2 mV-1000 V 1000 V ± 18.6 mV -1000.00255 V ± 1.2 mV

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.

The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of confidence of approximately 95%.

Standard

Option 002

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.

Standard

Option 002

nesaydin
Text Box
1-1234567890-1A
Page 5: Certification Cover Sheet - Keysight · CERTIFICATE NUMBER. 1-1234567890-1A. Page 2 of 6 pages ... Any quoted uncertainty refers only to the measured value and does not carry any

A voltage of substantially sinusoidal form was applied to the 3458A input.

ACBAND 10,2E6 for frequencies upto 2MHz

otherwise: ACBAND 10,1E7

(applied ± limit) (if applicable)

10 mV 1 kHz 10 mV ± 3.12 µV 10.00001 mV ± 3.0 µV10 mV 20 kHz 10 mV ± 4.12 µV 9.99767 mV ± 3.0 µV Guardband

10 mV 100 kHz 10 mV ± 51.12 µV 9.97883 mV ± 7.0 µV

10 mV 300 kHz 10 mV ± 402.02 µV 9.82043 mV ± 20 µV

10 mV 1 MHz 10 mV ± 125.02 µV 9.97074 mV ± 47 µV10 mV 4 MHz 10 mV ± 707.02 µV 9.86228 mV

100 mV 1 kHz 100 mV ± 9.2 µV 99.9952 mV ± 6.0 µV Guardband

100 mV 20 kHz 100 mV ± 16.2 µV 99.9950 mV ± 20.0 µV Guardband

100 mV 100 kHz 100 mV ± 82.2 µV 99.9718 mV ± 10.0 µV100 mV 300 kHz 100 mV ± 310.2 µV 99.8284 mV ± 30.0 µV

100 mV 1 MHz 100 mV ± 1010.2 µV 99.4532 mV ± 400.0 µV

StatusApplied Range Specification Measured Value

CERTIFICATE OF CALIBRATION sACERTIFICATE

NUMBER

UKAS Accredited Calibration Laboratory No. 0147 1-3113179760-1A

Page 3 of 6 pages

AC Voltage Accuracy (Synchronous mode)

3458A set-up: SETACV SYNC; RES .002; LFILTER ON;

Uncertainty

100 mV 1 MHz 100 mV ± 1010.2 µV 99.4532 mV ± 400.0 µV

100 mV 4 MHz 100 mV ± 4.0702 mV 101.1678 mV

100 mV 8 MHz 100 mV ± 4.0802 mV 100.8173 mV

100 mV 10 MHz 100 mV ± 15.1002 mV 102.1516 mV

1 V 1 kHz 1 V ± 92 µV 0.999986 V ± 60 µV

1 V 20 kHz 1 V ± 162 µV 1.000002 V ± 60 µV

1 V 50 kHz 1 V ± 322 µV 1.000146 V ± 60 µV1 V 100 kHz 1 V ± 822 µV 1.000204 V ± 90 µV

1 V 300 kHz 1 V ± 3.102 mV 1.000952 V ± 0.30 mV1 V 500 kHz 1 V ± 10.102 mV 1.002121 V ± 0.30 mV

1 V 1 MHz 1 V ± 10.102 mV 1.004372 V ± 0.40 mV

1 V 4 MHz 1 V ± 40.702 mV 1.011627 V

1 V 8 MHz 1 V ± 40.802 mV 1.006734 V1 V 10 MHz 1 V ± 151.002 mV 0.983150 V

3 V 100 kHz 10 V ± 2.602 mV 2.99878 V ± 0.90 mV

3 V 2 MHz 10 V ± 120.702 mV 3.01210 V

3 V 4 MHz 10 V ± 120.702 mV 3.07623 V

3 V 8 MHz 10 V ± 120.802 mV 3.02936 V3 V 10 MHz 10 V ± 460.002 mV 2.94926 V

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.

The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of confidence of approximately 95%.

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.

nesaydin
Text Box
1-1234567890-1A
Page 6: Certification Cover Sheet - Keysight · CERTIFICATE NUMBER. 1-1234567890-1A. Page 2 of 6 pages ... Any quoted uncertainty refers only to the measured value and does not carry any

(applied ± limit) (if applicable)

10 V 10 Hz 10 V ± 1.120 mV 10.00002 V ± 0.60 mV

10 V 20 Hz 10 V ± 1.120 mV 10.00004 V ± 0.60 mV

10 V 40 Hz 10 V ± 0.920 mV 10.00005 V ± 0.60 mV10 V 1 kHz 10 V ± 0.920 mV 10.00017 V ± 0.60 mV

10 V 4 kHz 10 V ± 1.620 mV 9.99989 V ± 0.60 mV

10 V 8 kHz 10 V ± 1.620 mV 9.99956 V ± 0.60 mV

10 V 10 kHz 10 V ± 1.620 mV 9.99952 V ± 0.60 mV10 V 20 kHz 10 V ± 1.620 mV 9.99963 V ± 0.60 mV

10 V 50 kHz 10 V ± 3.220 mV 9.99963 V ± 0.60 mV10 V 100 kHz 10 V ± 8.220 mV 9.99652 V ± 0.90 mV

10 V 300 kHz 10 V ± 31.020 mV 9.98207 V ± 3.0 mV

10 V 500 kHz 10 V ± 101.020 mV 9.97858 V ± 7.0 mV10 V 1 MHz 10 V ± 101.020 mV 10.01006 V ± 20.0 mV

CERTIFICATE

NUMBER

Uncertainty Status

1-3113179760-1A

Page 4 of 6 pages

CERTIFICATE OF CALIBRATION

UKAS Accredited Calibration Laboratory No. 0147

Applied Range

ACV (Synchronous mode) continued

Specification Measured Value

sA

100 V 1 kHz 100 V ± 0.0222 V 100.0040 V ± 0.0060 V

100 V 20 kHz 100 V ± 0.0222 V 99.9987 V ± 0.0060 V

100 V 50 kHz 100 V ± 0.0372 V 100.0009 V ± 0.0060 V100 V 100 kHz 100 V ± 0.1222 V 99.9948 V ± 0.0090 V

700 V 1 kHz 1000 V ± 0.301 V 699.835 V ± 0.140 V Guardband

700 V 20 kHz 1000 V ± 0.441 V 699.819 V ± 0.140 V

AC Current Accuracy

The applied current was of substantially sinusoidal form.

3458A set-up: ACBAND 10,1E4; LFILTER ON; NPLC 100

Note: The 100 mA and 1 A currents were applied for 5 minutes before measurement.

Applied Range Specification Measured Value Uncertainty Status

(applied ± limit) (if applicable)

10 µA 1 kHz 10 µA ± 36.05 nA 10.0131 µA ± 3.3 nA

100 µA 1 kHz 100 µA ± 90.50 nA 99.9939 µA ± 33.0 nA

1 mA 1 kHz 1 mA ± 505.0 nA 0.999946 mA ± 330 nA10 mA 1 kHz 10 mA ± 5.05 µA 9.99959 mA ± 1.5 µA

100 mA 1 kHz 100 mA ± 50.50 µA 99.9906 mA ± 15 µA

1 A 1 kHz 1 A ± 1.205 mA 0.999889 A ± 0.200 mA

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.

The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of confidence of approximately 95%.

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.

nesaydin
Text Box
1-1234567890-1A
Page 7: Certification Cover Sheet - Keysight · CERTIFICATE NUMBER. 1-1234567890-1A. Page 2 of 6 pages ... Any quoted uncertainty refers only to the measured value and does not carry any

3458A set-up:

1 megohm and higher:

NDIG 6

Last calibration internal temperature (Cal? 60) = 36.97 ºC

Present temperature (Temp?) = 35.00 ºC

Applied Range Spec. Error Uncertainty Status

(if applicable)

1.0000000 Ω 1 Ω ± 68 ppm 0.9999978 Ω -2.2 ppm ± 4.0 ppm

10.000000 Ω 10 Ω ± 23 ppm 9.999977 Ω -2.3 ppm ± 3.5 ppm

100.00000 Ω 100 Ω ± 18 ppm 100.00009 Ω 0.9 ppm ± 3.5 ppm

1.0000000 k Ω 1 k Ω ± 11.5 ppm 0.9999990 k Ω -1.0 ppm ± 3.5 ppm

10.000000 k Ω 10 k Ω ± 11.5 ppm 9.999998 k Ω -0.2 ppm ± 1.2 ppm

Measured Value

NDIG 5

100 MΩ:

2-wire ohms; NPLC 500

sACERTIFICATE

NUMBER

1-3113179760-1A

Page 5 of 6 pages

1 GΩ:

CERTIFICATE OF CALIBRATION

UKAS Accredited Calibration Laboratory No. 0147

4-wire ohms; OCOMP ON; Delay 1; NPLC 100; NDIG 7

Resistance Accuracy

Below 1 megohm:

100.00000 k Ω 100 k Ω ± 11.5 ppm 100.00003 k Ω 0.3 ppm ± 4.0 ppm

1.0000000 M Ω 1 M Ω ± 17 ppm 0.9999985 M Ω -1.5 ppm ± 6.0 ppm

10.000000 M Ω 10 M Ω ± 63 ppm 10.000000 M Ω 0.0 ppm ± 7.0 ppm

100.0000 M Ω 100 M Ω ± 513 ppm 100.0099 M Ω 99 ppm ± 60 ppm

1.00000 G Ω 1 G Ω ± 0.501 % 1.00111 G Ω 0.111 % ± 0.28 %

The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of confidence of approximately 95%.

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.

nesaydin
Text Box
1-1234567890-1A
Page 8: Certification Cover Sheet - Keysight · CERTIFICATE NUMBER. 1-1234567890-1A. Page 2 of 6 pages ... Any quoted uncertainty refers only to the measured value and does not carry any

3458A set-up: NPLC 500; NDIG 7

Note: The 100 mA and 1 A currents were applied for 5 minutes before measurement.

(applied ± limit) (if applicable)

100 nA 100 nA ± 43.5 pA 99.9997 nA ± 27 pA

-100 nA 100 nA ± 43.5 pA -99.9999 nA ± 27 pA

1 µA 1 µA ± 60 pA 0.999998 µA ± 40 pA

-1 µA 1 µA ± 60 pA -0.999996 µA ± 40 pA

10 µA 10 µA ± 0.30 nA 9.999973 µA ± 0.08 nA-10 µA 10 µA ± 0.30 nA -9.999980 µA ± 0.08 nA

100 µA 100 µA ± 2.80 nA 100.00008 µA ± 0.65 nA

-100 µA 100 µA ± 2.80 nA -100.00012 µA ± 0.65 nA

1 mA 1 mA ± 25.0 nA 0.9999999 mA ± 7.5 nA

-1 mA 1 mA ± 25.0 nA -1.0000005 mA ± 7.5 nA

Applied

CERTIFICATE

NUMBER

UKAS Accredited Calibration Laboratory No. 0147 1-3113179760-1A

Page 6 of 6 pages

DC Current Accuracy

Uncertainty StatusRange

* The published specifications for the 100nA, 1µA and 10 µA ranges are "typical" only.

sACERTIFICATE OF CALIBRATION

Specification * Measured Value

-1 mA 1 mA ± 25.0 nA -1.0000005 mA ± 7.5 nA

10 mA 10 mA ± 0.250 µA 9.999991 mA ± 0.075 µA

-10 mA 10 mA ± 0.250 µA -9.999994 mA ± 0.075 µA

100 mA 100 mA ± 4.00 µA 99.99938 mA ± 0.80 µA

-100 mA 100 mA ± 4.00 µA -99.99967 mA ± 0.80 µA

1 A 1 A ± 115 µA 0.9999447 A ± 20 µA

-1 A 1 A ± 115 µA -0.9999479 A ± 20 µA

The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of confidence of approximately 95%.

Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.Any quoted uncertainty refers only to the measured value and does not carry any implication regarding the long term stability of the instrument.