chapter 16 infrared absorption spectroscopy an ir spectrum contains information about the functional...
TRANSCRIPT
Chapter 16
Infrared Absorption Spectroscopy
An IR spectrum contains information about the functional groups in a molecule, and this is used to uniquely identify the compound.
Wavenumbers (cm-1) are used since they are directly proportional to energy –
e.g. convert 2.5 µm to wavenumbers (cm-1)
λ
hchνE
400 nm 780 nm 2500 nm
or 4000 cm-1
50,000 nm
or 200 cm-1
UV Vis Near IR Mid IR Far IR
Typical Infrared Spectrum
Higher energy vibrations
C-H
“overtones”
C=O C-C
Mechanical Model of a Stretching Vibration in a Diatomic Molecule
μ
k
2π
1ν
Treats the vibrating bond like a spring with a given “stiffness” --
n = frequency of vibration
k = force constant (spring “stiffness”)
µ = “reduced mass”
21
21
mm
mm
Quantum Mechanical Treatment of Normal Modes
1. Only certain vibrations are “allowed”
2. The vibrational quantum number v = 0, 1, 2, 3……
3. Only transitions between adjacent energy levels are possible, i.e. v = 1
4. The frequency of the photon has to equal the frequency of the vibration(see next slide)
5. The molecule must have a change in “dipole moment” as a result of the vibration
Dipole moment = charge difference X separation distance
Homonuclear diatomics never have an Infrared spectrum. Why?
+-
+ + +
- -
+
--
+
IR Absorption if the vibration results in a change in dipole moment
Frequency of absorbed light frequency of vibrationFrequency of absorbed light frequency of vibration
So
S1
v = 1
Common Types of Vibrations (“Normal Modes”)
Higher energy modes
Lower energy modes
Example – predict the normal modes of CO2
The number of normal modes = 3N-6
For linear molecules it’s 3N-5
Which normal modes are Infrared active?
Example 16-1, p.436
Calculate the approximate wavenumber and wavelength of the fundamental absorption peak due to the stretching vibration of a C=O group (k = 1.0 x 103 N/m)
Force Constants increase with Bond Strength
Bond Type*
Force Constant,
k (N/m)
Wave-number (cm-1)
Bond Energy (kJ/mol)
C-C 5 x 102 800-1200 347
C=C 10 x 102 1600 620
CΞC 15 x 102 2100 812
Bond Type*
Reduced mass (kg)
Wave-number (cm-1)
Bond Energy (kJ/mol)
C-H 1.55 x 10-27 3000 414
C-N 1.07 x 10-26 1000-1350 276@
C-O 1.14 x 10-26 1000-1300 351@
* all bonds are stretches@ consider these as being approx. equal
(a).
(b)
μ
k
2π
1ν
(a).as then frequency
(b) When masses approx. equal, then see peak at same wavenumbers
IR Sources
1. Nichrome – Ni/Cr alloy, resistively heats up and emits IR, 1100 K
2. Globar – SiC rod, 1500 K
3. Nernst Glower – electrically heated rare earth-oxide ceramic, 2000 K
Output from a Nernst Glower
cm-1
Optics – IR spectra are measured in the mid-IR, so have to use halides such as NaCl and KBr
Never use aqueous samples, or water to clean
salt plates
400 nm 780 nm 2500 nm
or 4000 cm-1
50,000 nm
or 200 cm-1
UV Vis Near IR Mid IR Far IR
Sample Handling – a few drops of “neat” sample between “salt plates”
Solutions
Detectors for the Mid-IR – 200-4000cm-1
1. Pyroelectric – crystal of DTGS (Deuterated Triglyceine Sulfate)
Nice link:
http://www.doitpoms.ac.uk/tlplib/pyroelectricity/printall.php?question=2&type=1
1. DTGS maintains polarization when heated to just below the Curie Point
2. Above the Curie Point, the permanent polarization of the DTGS crystal disappears.
3. The closer to the Curie Point, the more responsive the detector
4. Much faster response so can take a spectrum much more quickly (FTIR)
Detectors for the Mid-IR – 200-4000cm-1
2. Photoconduction - Mercury-Cadmium-Tellurium (MCT)
http://www.newport.com
Liquid N2 @ 77K1. Semiconductor-based
2. Cooled to 77 K using liquid N2
3. Much faster response so can take a spectrum much more quickly (FTIR)
http://www.boselec.com/pdf/Laser_Focus_World.PDF
Definition of the Fourier Transform () -
dtetftf ti )()()( 1i f 2
The Inverse Fourier Transform -
detf ti)(2
1)()( 1
The functions f(t) and () are called "transform pairs"
f(t) ()
time (t) frequency (s-1, Hz)
e.g. FTIR
distance (cm) wavenumber (cm-1)
backgroundinterferogram
P/Po =IR spectrum
background spectrum = Po
sample spectrum = P
sampleinterferogram
e.g. FT-NMR
time (s) ppm (Hz)
Free Induction Decay (FID) Signal (time)
NMR Spectrum (ppm, Hz)
Fourier Transforms "invert dimensionality" –
signal domain FT domain Instrument
time (t) frequency (s-1, Hz) oscilloscope
mirror wavenumber (cm-1) FTIRdistance (cm)
free induction ppm (from Hertz) FT-NMRdecay (t)
Fast Fourier Transform (FFT) on computers -
• Cooley-Tukey algorithm
• very efficient on a PC
movable mirror →
D
fixed mirror
50/50 beamsplitter
IR source
mirror distance, x (cm) →
¼
interferogram
++
+--
-
/4 2 /4 3/4 4/4
If polychromatic radiation enters the interferometer, then each wavelength produces a separate interferogram. The output from the interferometer will therefore be a superposition of all wavelengths -
1 1
mirror distance (+x, cm) →
2 2
3 3
←mirror distance (-x, cm)
....4cos)(4cos)(4cos)( 332211 xAxAxAsignal atdetector S(x)
N
i
ii xA
1
4cos)(
dxA ii 4cos)(
this term contains the IR spectrum
Taking the Fourier Transform of S(x) results in the IR spectrum, A() -
dxxxSA i 4cos)()(
Depends on the distance the mirror moves in the Michelson Interferometer – the further it moves, the greater the resolution.
1
2
1
x
minimum detectable difference in wavenumber
x = distance mirror moves
e.g. What length of mirror drive in a Michelson Interferometer is required to separate 20.34 and 20.35 m?
"retardation“ = 2x
Fourier Transform Infrared (FTIR) Spectrometers
The Multiplex (Fellgett's) Advantage - entire spectrum obtained virtually instantaneously; results in a higher S/N because during the time a scanning instrument is slowly obtaining the spectrum, a multiplex instrument such as an FTIR can acquire 100’s pf spectra. Averaging these extra spectra causes the increased S/N (see next slide) and allows quantitative IR.
S/N increases as where N = number of "resolution elements"N
4000 cm-1400 cm-1
e.g. scan range 400 - 4000 cm-1 at a resolution of 2 cm-1
2 cm-1
18002
)4004000(1
1
cm
cmN "resolution elements"
so the S/N increases by IF the FTIR can obtain the entire spectrum for the same amount of time a slower, scanning instrument requires to acquire only one resolution element.
X401800
Signal-to-Noise Ratio (S/N)
Single-Beam – run background first (see next slide) and store in memory (Po)
Higher energy throughput (Jacquinot’s Advantage) and no stray light problems
Typical IR background spectrum (Po)
inexpensive (~ $25K)
range = 350 – 7800 cm-1 (29 to 1.3 m)
max resolution = 4 cm-1
scan time = as fast as 1 sec
detector = DTGS
expensive (~ $100K)
range = 10 – 50,000 cm-1 (1000 m to 200 nm)
max resolution < 0.01 cm-1
scan time = can be minutes at high-res
detector(s) = DTGS, MCT
Advantages over Scanning Instruments
1. fast scan times
2. higher S/N because of signal averaging
3. higher resolution
4. because there are no slits, there’s a higher energy throughput (Jacquinot’s Advantage); means larger signals for the same concentration (therefore lower LODs)
5. no stray light problems
Disadvantages
6. higher cost than scanning instruments
7. difficult to align the interferometer (automation helps)
8. IR optics water soluble (beamsplitter made of KBr)
Chapter 17
Applications of IR Spectrometry
1. Qualitative Analysis
"fingerprint region""group frequency region"
2. Quantitative Analysis
1. lower source powers than in UV-Vis (in FTIR Jacquinot's Advantage partially offsets)
2. detectors suffer from thermal noise (i.e. larger backgrounds or blank)
3. salt plates have a very short path length (A = bc)
Signal averaging with FTIR's has improved the S/N enough to allow more sensitive quantitative work in the IR.
Much less sensitive (i.e. higher LODs) in the IR compared to the UV-Visbecause of -
3. Diffuse Reflectance Accessory
Typically used for powdered samples, i.e. forensic drug analyses
Kubelka-Munk Units = converts the reflectance spectrum to the equivalent of an absorbance spectrum.
R’ = sample refectivity/KBr reflectivity
f(R’) = (1 - R’)2/2 R’ = k/S
k = 2.303 C
= the molar absorptivityC is the sample concentrationS = "scattering coefficient"
For a sample to follow a linear relationship between f(R’) and concentration, the following criteria must be met:
1. the sample must be diluted in a non-absorbing matrix such as KBr or KCl.2. the scattering coefficient S must remain constant over the entire spectrum.3. there must be no specular, or regular reflectance off the surface of the
sample.
4. Attenuated Total Internal Reflectance (ATR) Accessory
Total Internal Reflection and the "evanescent wave" -
Samples - conventional (solutions, liquids, etc) as well as powders, pastes, suspensions, colloids
5. Infrared Microscopy