chapter 2 titrimetric, spectrophotometric...
TRANSCRIPT
18
SECTION 2.0
DRUG PROFILE AND LITERATURE SURVEY
2.0.1 DRUG PROFILE
Ofloxacin (OFX) is chemically known as (RS)-7-fluoro-2-methyl-6-(4-
methylpiperazine-1-yl)-10-oxo-4-oxa-1-azatricyclo[7.3.1.0]trideca-5(13),6,8,11-tetraene-
11-carboxylic acid. Its empirical formula is C18H20FN3O4, with a molecular weight of
361.4. The structural formula is:
O
NN
NCH
3
F
O
OH
O
CH3
OFX is a pale yellow or bright yellow crystalline powder. It is slightly soluble in
water, methanol and methylene chloride, soluble in glacial acetic acid.
OFX is a new synthetic second generation fluoroquinolone antibiotic with a broad
spectrum of activity against gram-positive and gram-negative bacteria [1]. OFX is widely
used in the treatment of respiratory tract and urinary tract infections [2]. OFX possesses
two relevant ionizable functional groups: a basic piperazinyl group and a carboxylic
group. The carboxylic group and the carbonyl groups are required for antimicrobial
activity. OFX is official in the United States Pharmacopoeia (USP) [3] and the British
Pharmacopoeia (BP) [4], which recommends non-aqueous titrimetry with acetous
perchloric acid towards potentiometric end point [3, 4]. BP also recommends a High
performance liquid chromatography (HPLC) [3] technique for the assay of OFX in bulk
and dosage forms.
19
2.0.2.0 LITERATURE SURVEY - ANALYTICAL FRAMEWORK
2.0.2.1 Titrimetric methods
Other than official methods [3, 4] no titimetric methods were found in the
literature for the determination of OFX in pharmaceuticals. The official methods involve
the titration of the drug with 0.1 M perchloric acid towards potentiometric end point.
Only one visual titrimetric procedure was found in the literature for the assay of OFX in
bulk drug as well as in its formulations [5]. The method was based on the bromination of
the drug in HCl medium followed by the determination of unreacted NBS iodometrically.
The method was applicable over an amount of 1-8 mg OFX.
2.0.2.2 Spectrophotometric methods
Several visible spectrophotometric methods [5-12] employing different reactions
have been reported for the assay of OFX either alone or with some other therapeutic
agents. Vinay et al.,[5] developed two methods based on bromination of the drug,
followed by the determination of unreacted NBS by reacting with fixed amount of indigo
carmine or metanil yellow. The Sastry et al., [6] have reported a method based on
coupling reaction using Ce(IV). The method involved the measurement of absorbance of
oxidative coupled product formed between Ce(IV) and 3-methyl-2benzothiazoline
hydrazone. Absorbance of the colored product was measured at 630 nm and applicable
over a concentration range of 1-10 µg ml-1
. OFX was determined by complex formation
reaction using Fe(III) nitrate nonahydrate and absorbance of the amber colored
chromogen was measured at 370 nm [7]. By using the similar reaction, Mathur et al, [8]
developed a method using FeCl3 in HCl medium. The method was found to be less
sensitive and applicable over a concentration range of 20-160 µg ml-1
. A method based
on the measurement of 2:1 complex formed between OFX and iron(III) in H2SO4
medium, followed by the measurement of absorbance of the complex at 420 nm [9]. The
method was applicable to pharmaceutical formulations as well as in spiked human urine
over a concentration range of 10-120 µg ml-1
. Issa et al., [10] have developed three
methods based on the formation of chloroform extractable ion-pair complexes between
OFX and bromophenol blue (BPB), bromothymol blue (BTB) and bromocresol purple
(BCP). The methods were applicable over a concentration of 5-25, 2-15 and 2-20 µg ml-1
20
for BPB, BTB and BCP methods. Another two extractive spectrophotometric methods
were developed by Sastry et al., [11]. The methods were based on formation of
chloroform soluble ion-pair complex between Tropaelin 000 (TP 000) and Supracene
violet 3B (SV 3B). Absorbance of the ion-pair complexes were measured at 485 and 575
nm for TP 000 and SV 3B methods. Patel et al., [12] have developed a method based on
the formation of pink colored chromogen having maximum absorbance at 552.5 nm, by
the reaction between OFX and citric acid-acetic anhydride. The method was applicable
over a concentration range of 5-55 µg ml-1
.
Vinay et al., [9] developed two UV-spectrophotometric methods in
pharmaceutical formulations and in spiked human urine. The methods were based on the
the measurement of the absorbance of the drug in 0.1 M HCl at 293 nm and in 0.1 M
NaOH at 287 nm. Another UV-spectrophotometric method [14] has been reported by for
the assay of OFX in pharmaceuticals and spiked human urine.
2.0.2.3 Chromatographic techniques
High performance Liquid Chromatography (HPLC), despite its versatility, speed,
sensitivity and specificity, has sparsely been employed for the assay of OFX. The
technique has been applied for the assay of OFX in pharmaceuticals [14-18]. A method
was found in the literature using octadecyl silane chemically bonded to porous silica
particles (Waters Spherisorb, 5 µ ODS 1, 4.6 x 150 mm) column with a 0.05 M phosphate
buffer-acetonitrile (65:35, v/v) as mobile phase and the detection was made at 210 nm.
The method was rectilinear over a concentration range of 24-120 µg ml-1
[14]. Kalta et
al, [15] developed a method using acetonitrile: potassium dihydrogen phosphate buffer
(80:20) with 0.5% v/v of triethylamine and the pH was adjusted to 2.5 by adding
orthophosphoric acid. Phenomenex C18 column with 0.24% sodium lauryl sulphate with
a mobile phase consisting of acetonitrile: acetic acid (pH-4.0) (58:40:02) and detection
made at 295 nm. The method was less sensitive and found rectilinear over a range of 160-
240 µg ml-1
[16]. A method was developed by Dharuman et al., [17] using Kromasil C8
(5µ, 15 cm × 4.6 mm id) column with mobile phase consisting of 0.5%v/v triethylamine
buffer of pH 3.0 and acetonitrile in the ratio of 73: 27 and detection was made at 303
nm.C-18 column (RP-18, 5µ) coupled with a guard column of same material, in isocratic
mode with mobile phase mixture of acetonitrile: water: tri ethylamine (25:75:1) followed
21
by UV-detection at 300 nm [18].
HPLC method was also used for the assay of OFX in blood serum [19-21], blood
plasma [22, 23], waste waters [24, 25] and urine [26, 27]. Ligand exchange HPLC has
recently been applied for the direct determination of OFX enantiomers in human urine
[28]. A stability indicating high-performance thin layer chromatographic [29] assay has
also been described for OFX.
2.0.2.4 Other techniques
Various other techniques have been reported for the determination of OFX in pure
drug, pharmaceutical dosage forms and in biological fluids, and they include
spectrofluorimetry [30-34], potentiometry and conductometry [35], polarography and
voltammetry [36], differential pulse polarography [37], ion-selective electrode based
potentiometry [38], adsorptive stripping voltammetry [39], voltammetry, linear sweep
voltammetry and electrochemical impedance spectroscopy [40], chemiluminescence
spectrometry [41-45] and capillary electrophoresis [46-48].
Many of the above techniques are deficient on simplicity, cost-effectiveness and
easy accessibility. The reported HPLC methods for OFX in dosage forms [14-18] either
require an internal standard [14, 17] or are applicable for combined dosage forms [14-18].
Assays of OFX in urine [26-28] are generally complex and lengthy, mostly because of
the sample preparation, like solid-phase extraction or liquid-liquid extraction.
Titrimetry and visible spectrophotometry may serve as useful alternatives to many
of the aforesaid sophisticated techniques because of their cost-effectiveness, ease of
operation, sensitivity, remarkable accuracy and precision and wide applicability. The
reported titrimetric method have disadvantages of narrow dynamic range, use of unstable
reagent, use of hazardous chemical etc. The reported spectrophotometric methods, though
sensitive, suffer from one or the other disadvantage such as use of boiling [12] or
extraction step [10, 11], strict pH control [7, 10, 11 ], use of organic solvent/expensive
chemical [6] and measurement of less sensitive species [8].
Realizing the importance of titrimetry in pharmaceutical analysis, and
inadequacies of the currently available visible spectrophotometric methods, the author
has made an attempt to develop new titrimetric and visible spectrophotometric methods
for the determination of OFX. The methods use cerium(IV) as the oxidimetric reagent,
22
and 0-dianisidine and p-tolidine as auxiliary reagents. In addition, extraction-free
spectrophotometric procedures using anionic dyes as ion-pair reagents have also been
developed. Using a different column and mobile phase, a new and sensitive HPLC
method with a wide linear dynamic range was also developed. The details concerning the
method development and method validation are presented in the following sections of
this chapter.
23
SECTION 2.1
TITRIMETRIC ASSAY OF OFLOXACIN IN PHARMACEUTICALS USING
CERIUM(IV) AS AN OXIDIMETRIC REAGENT
2.1.1.0 INTRODUCTION
Cerium(IV) sulfate is a chemical compound which is frequently used as an
oxidising agent in titration. The orange-colored cerium(IV) ion is reduced to the colorless
cerium(III) ion. The potential for reduction is high, about +1.44V. It is a very safe
oxidizer, one of the most nontoxic ones [49].
Ce4+ + e-Ce3+
It can be used only in acid solution, best in 0.5 M or higher concentrations.
Solutions of cerium(IV) sulphate (Ce(IV)SO4) in dilute H2SO4 are stable even at boiling
temperature [50].
Ce(IV) has been used for the titrimetric analysis of some pharmaceuticals [51-53],
ascorbic acid [54] etc. It has also been used for the spectrophotometric assay of
simvastatin [55], olanzapine [56, 57], lansoprazole [58], ciprofloxacin [59], propranolol
[60], atenolol [61], methylthiouracil [62], antiamoebics and anthelmintics [63] and
diuretics [64], methdilazine[65], ganciclovir [66] and carbamazepine [67].
Although, Ce(IV) has been extensively used for the titrimetric and
spectrophotometric assay of several pharmaceuticals, there is no report found for the
assay of OFX using cerium(IV). In this section, titrimetric assays of OFX are described
based on the oxidation of drug by Ce(IV). In direct titrimetry, the acidified solution of
OFX is titrated directly with Ce(IV) using ferroin as indicator, and indirect titrimetry
involves the addition of known excess of Ce(IV) to an acidified solution of OFX followed
by the determination of unreacted oxidant by back titration with ammonium ferrous
sulphate (FAS) using the same ferroin indicator. The details of the method development
and validation are presented in this Section 2.1.
24
2.1.2.0 EXPERIMENTAL
2.1.2.1 Reagents and Standards
All chemicals and reagents used were of analytical or pharmaceutical grade and
distilled water was used throughout the experiment.
Cerium(IV) sulphate (0.025 M): The solution was prepared by dissolving 5.05 g of the
chemical (LOBA Chemie, Mumbai,India, assay 99.9%) in 14 ml of conc. H2SO4 and
diluting to volume in a 500 ml calibrated flask with water, mixed well, filtered using
glass wool, standardized [68] and used in both the methods.
FAS (0.025 M): The solution of FAS (0.05 M) was prepared by dissolving 4.9 g of the
chemical (LOBA Chemie, Mumbai, India, assay 99-101%) in water containing 5 ml of
2M H2SO4 and diluted to volume in a 500 ml calibrated flask with water and used in
indirect titrimetry.
Sulphuric acid (2 M): Prepared by appropriate dilution of the conc. H2SO4 (S. D. Fine
Chem, Mumbai, India, sp. gr. 1.84) with water.
Ferroin indicator: Prepared by dissolving 0.742 g of 1,10-phenanthroline monohydrate
in 50 ml of 0.025 M ferrous sulphate solution (0.348 g of ferrous sulphate heptahydrate in
50 ml water).
Preparation of Standard OFX Solution
Pure OFX (Pharmaceutical grade) sample was kindly provided by Cipla India Ltd,
Mumbai, India, and was used as received. Standard OFX solution (1.5 mg ml-1
) was
prepared by dissolving calculated quantity of pure drug in 0.1 M H2SO4.
Two brands of tablets containing OFX, Zenflox-400 (Mankind Pharma Pvt Ltd.,
New Delhi, India) and Ofloxin-400 (J. B. Chemicals and Pharmaceuticals Ltd, Mumbai,
India), used in the investigation were purchased from local commercial sources.
2.1.3.0 ASSAY PROCEDURES
2.1.3.1 Direct titrimetry
A 10 ml aliquot of pure drug solution containing 1.5-15 mg of OFX was
accurately measured and transferred into a 100 ml titration flask. Five ml of 2 M H2SO4
was added, and titrated with 0.025 M Ce(IV)SO4 using 1 drops of ferroin as indicator
until the sky blue color appeared.
25
2.1.3.2 Indirect titrimetry
A 10 ml aliquot of standard solution containing 1.5-15 mg of OFX was measured
accurately and taken in a 100 ml titration flask, and five ml of 2 M H2SO4 was added to
the flask. By means of a pipette, 10 ml of 0.025 M Ce(IV) was added, content mixed and
the flask kept aside for 15 min. Then, the unreacted Ce(IV) was titrated against 0.025 M
FAS using ferroin as indicator to a first appearance of orange red color. A blank titration
was performed with 10 ml water in place of OFX solution.
The amount of OFX in the aliquot was computed from the formula:
Amount (mg) = VMwS/n
where V = ml of Ce(IV) reacted
Mw = relative molecular mass of drug
S = strength of Ce(IV), moles l-1
n = number of moles of Ce(IV) reacting with per mole of OFX.
2.1.3.3 Procedure for tablets
Ten tablets were accurately weighed and powdered. A portion equivalent to 150
mg OFX was accurately weighed and transferred into a 100 ml calibrated flask, 60 ml of
0.1 M H2SO4 was added to the flask and the content was shaken thoroughly for 15-20
min to extract the drug into the liquid phase; the volume was finally diluted to the mark
with the same acid, mixed well and filtered using a Whatman No. 42 filter paper. The
filtrate equivalent to 1.5 mg ml-1
OFX was assayed by both the methods.
2.1.3.4 Placebo blank and synthetic mixture analysis
A placebo blank of commonly employed tablet excipients, consisting of talc (50
mg), starch (50 mg), acacia (20 mg), methyl cellulose (30 mg), sodium citrate (20 mg),
magnesium stearate (20 mg) and sodium alginate (2 mg) was made and its solution was
prepared by taking about 30 mg as described under Section 2.1.3.3.
To 50 mg placebo blank of the composition described above, 50 mg and 100 mg
of OFX was added separately and homogenized, transferred to two separate 50 ml
calibrated flask and the solution was prepared as described under Section 2.1.3.3.
2.1.4.0 RESULTS AND DISCUSSION
From the preliminary experiments, OFX was found to undergo oxidation with
Ce(IV). Based on this observation, two titrimetric methods were developed. In direct
26
titrimetry, OFX was titrated directly with Ce(IV) using ferroin as indicator, and indirect
titrimetry involved back titration of the unreacted Ce(IV) with FAS using the same
indicator.
2.1.4.1 Optimization of variables
The experimental variables which provided accurate and precise results were optimized.
In both the methods, the reactions were carried out in H2SO4 medium, since in
this medium the reactions were found to be stoichiometric. In direct titrimetry,
reproducible and regular stoichiometry was obtained when 0.33-1.67 M H2SO4
concentration was maintained. Hence, 5 ml of 2 M H2SO4 solution in a total volume of 15
ml (0.67 M overall) was found to be the most suitable concentration for a quantitative
reaction between OFX and Ce(IV). In indirect titrimetry, the reaction stoichiometry was
unaffected when 0.2-1.0 M H2SO4 was maintained. Hence, 5 ml of 2 M H2SO4 in a total
volume of 25 ml (0.4 M overall) was used. At the laboratory temperature (30 ± 2o C), the
time required for complete oxidation was found to be 15 min and the contact time up to
60 min, a small amount of Ce(IV) is consumed but without producing a definite
stoichiometry. Hence, the reaction time of 15 min was fixed in method B. In both the
methods, the reaction stoichiometry of OFX: Ce(IV) was calculated to be 1:5 in the range
of 1.5-15 mg in both the methods.
2.1.4.2 Method validation
The proposed methods were validated for precision, accuracy, robustness
and ruggedness and specificity, according to the International Conference on
Harmonization (ICH) [69] guidelines.
Quantitative data
Over the range investigated 1.5-15 mg, a fixed stoichiometry of 1:5 (OFX:
Ce(IV)) in both the methods was obtained which served as the basis for calculations.
Precision and accuracy
Intra-day precision and accuracy of the proposed methods were evaluated by
replicate analysis (n=7) of standards at three different levels on the same day. Inter-day
precision and accuracy were determined by assaying the standards at the same amount
levels on five consecutive days. Precision and accuracy were based on the calculated
27
relative standard deviation (RSD, %) and relative error (RE, %) of the found amount
compared to the taken one, respectively (Table 2.1.1).
Table 2.1.1 Results of intra-day and inter-day accuracy and precision study
Method
OFX
taken,
mg
Intra-day accuracy and
precision
(n=7)
Inter-day accuracy and
precision
(n=7)
OFX found %RE %RSD OFX
found %RE %RSD
Direct
titrimetry
6.00 5.92 1.32 2.40 6.09 1.50 2.51
9.00 8.86 1.56 1.20 9.18 2.00 1.67
12.0 12.10 0.83 2.15 12.20 1.67 1.97
Indirect
titrimetry
6.00 6.19 3.17 2.40 5.80 3.26 2.57
9.00 8.74 2.89 1.46 9.27 2.95 1.73
12.0 12.18 1.50 1.20 12.28 2.32 1.54
RE: Relative error and RSD: Relative standard deviation.
Robustness and ruggedness
Method robustness was tested by making small incremental change in H2SO4
concentration in both the methods, and in method B, even reaction time was slightly
altered. To check the ruggedness, analysis was performed by four different analysts; and
on three different burettes by the same analyst. The robustness and the ruggedness were
checked at three different drug levels. The intermediate precision, expressed as percent
RSD, which is a measure of robustness and ruggedness was within the acceptable limits
as shown in the Table. 2.1.2.
Selectivity
The proposed methods were tested for selectivity by placebo blank and synthetic
mixture analyses. A convenient aliquot of the placebo blank solution was subjected to
analysis according to the recommended procedures. In both the cases, there was no
interference by the inactive ingredients as shown by near zero ml consumption of Ce(IV).
A separate experiment was performed with the synthetic mixture. The analysis of
synthetic mixture solution yielded percent recoveries which ranged of 95.71- 104.3
with standard deviation of 1.17 – 1.36 in both the cases. The results of this study are
presented in Table 2.1.3 indicating that the inactive ingredients did not interfere in the
assay. These results further demonstrate the accuracy as well as the precision of the
proposed methods.
28
Table 2.1.3 Results of recovery of the drug from synthetic mixture
*Mean value of five determinations
Application to tablets
In order to evaluate the analytical applicability of the proposed methods to the
quantification of OFX in commercial tablets, the results obtained by the proposed
methods were compared to those of the reference method [4] by applying Student’s t-test
for accuracy and F-test for precision. The reference method involved the titration of OFX
in anhydrous acetic acid with acetous perchloric acid to potentiometric end point
detection. The results (Table 2.1.4) show that the Student’s t- and F-values at 95 %
confidence level are less than the theoretical values, which confirmed that there is a good
agreement between the results obtained by the proposed methods and the reference
method with respect to accuracy and precision.
Table 2.1.2 Results of robustness and ruggedness expressed as intermediate
precision (%RSD)
Method
OFX
taken,
mg
Method robustness
Method ruggedness Parameter altered
H2SO4 , ml*
% RSD,
(n = 3)
Reaction
time**
,
min
Inter-
analysts’
% RSD,
(n = 4)
Inter-
burettes’
% RSD,
(n = 3)
Direct
titrimetry
6.00 1.19 - 0.87 1.21
9.00 1.07 - 0.82 1.14
12.0 1.02 - 0.71 1.31
Indirect
titrimetry
6.00 1.16 1.09 1.03 1.25
9.00 1.07 1.12 0.96 1.29
12.00 0.89 0.98 0.83 1.35 * H2SO4 volumes used were 4.8, 5.0 and 5.2 ml of 2 M in both the methods. **Reaction times altered were 13, 15 and 17 min in method B.
Method OFX in synthetic
mixture taken, mg
OFX recovered*
(Percent ± SD)
Direct titrimetry
6.00 103.5 ± 1.23
9.00 102.9 ± 1.17
12.00 104.3 ± 1.21
Indirect titrimetry
6.00 95.97 ± 1.29
9.00 96.78 ±1.24
12.00 95.71 ± 1.36
29
Recovery studies
The accuracy and validity of the proposed methods were further ascertained by
performing recovery studies. Pre-analysed tablet powder was spiked with pure OFX at
three concentration levels (50, 100 and 150 % of that in tablet powder) and the total was
found by the proposed methods. In both the cases, the added OFX recovery percentage
values ranged of 94.58 – 101.0 % with standard deviation of 1.13 – 1.32 (Table 2.1.5)
indicating that the recovery was good, and that the co formulated substance did not
interfere in the determination.
Table 2.1.4 Results of analysis of tablets by the proposed methods and statistical
comparison with reference method
Tablet brand
name
Label
claim,
mg/tablet
Found* (Percent of label claim ± SD)
(n=5)
Reference
method
Direct titrimetry Indirect titrimetry
Zenflox** 400
98.12 ± 1.18
96.58 ± 1.24 t = 2.01
F = 1.10
96.50 ± 1.38 t = 2.00
F = 1.37
Ofloxin***
400
99.14 ± 1.07
98.42 ± 1.11
t = 1.04
F = 1.08
97.13 ± 1.36 t = 2.57
F = 1.62 *Mean value of five determinations. ** Mankind Pharma Pvt Ltd., New Delhi, India;*** J. B. Chemicals and Pharmaceuticals Ltd,
Mumbai, India.
The value of t (tabulated) at 95 % confidence level and for four degrees of freedom is 2.77.
The value of F (tabulated) at 95 % confidence level and for four degrees of freedom is 6.39.
Table 2.1.5 Results of recovery study by standard addition method
Method Tablet
studied
OFX in
tablet,
mg
Pure
OFX
added,
mg
Total
found, mg
Pure OFX
recovered*
Percent ± SD
Direct
titrimetry
Ofloxin-
400
5.91 3.00 8.94 101.0 ± 1.32
5.91 6.00 11.81 98.33 ± 1.23
5.91 9.00 14.42 94.58 ± 1.15
Indirect
titrimetry
Ofloxin-
400
3.89 2.00 5.81 96.15 ± 1.21
3.89 4.00 7.77 97.08 ± 1.13
3.89 6.00 9.85 99.38 ± 1.28 *Mean value of three measurements.
30
SECTION 2.2
SPECTROPHOTOMETRIC DETERMINATION OF OFLOXACIN IN
PHARMACEUTICALS BY REDOX REACTION USING CERIUM(IV)
2.2.1.0 INTRODUCTION
The chemistry and applicability of Ce(IV) has been reviewed in Section 2.1.1.0.
In this Section, 2.2, two spectrophotometric methods (method A and method B) are
described based on the oxidation of OFX by a measured excess of cerium(IV) in H2SO4
medium. The unreacted oxidant was determined by reacting with either p-toluidine (p-
TD) and measuring the absorbance at 525 nm (method A) or o-dianisidine (o-DA) and
measuring the absorbance at 470 nm (method B). The proposed methods were validated
for linearity, sensitivity, selectivity, robustness and ruggedness besides intra-day and
inter-day precision and accuracy.
2.2.2.0 EXPERIMENTAL
2.2.2.1 Apparatus
All absorption measurements were made using a Systronics model 106 digital
spectrophotometer (Systronica Ltd, Ahmedabad, India) with 1 cm path length quartz
cells.
2.2.2.2 Reagents and standards
All chemicals used were of analytical reagent grade. Distilled water was used
throughout the investigation.
Cerium(IV) (6000 and 300 µg ml-1
): An approximately 0.025 M cerium (IV) sulphate was
prepared by dissolving 5.1 g of chemical (LOBA Chemie, Mumbai,India, 99.9 % pure) in
0.5 M H2SO4 with the aid of heat, and filtered using Whatmen No 42 filter paper, and
diluted to 500 ml in a calibrated flask with the same acid and standardized against standard
0.025 M ferrous ammonium sulphate in the presence of ferroin as indicator [64]. The
solution was then diluted appropriately with 0.5 M H2SO4 to get working concentrations of
6000 µg ml-1
and 300 µg ml-1
Ce(IV) for use in method A and method B, respectively.
Acetic acid (1:1 and 3:2 v/v): In two different 100 ml calibrated flasks 50 ml and 60 ml of
glacial acetic acid (S.D. Fine Chem Ltd, Mumbai, India, purity 100%) were diluted upto
the mark with water to get 1:1 and 3:2 acetic acid, respectively.
31
p-Toluidine, (p-TD) (1 %): Prepared by dissolving 1 g of the chemical (Merk, Mumbai,
India, 99 % assay) in 1:1 acetic acid and diluted to 100 ml with the same solvent in a
calibrated flask.
o-Dianisidine, o-DA (0.01 %): The solution was prepared by dissolving accurately
weighed 10 mg of the chemical (Loba chemie, Mumbai, India, 99 % assay) in 100 ml of
3:2 acetic acid.
Sulphuric acid (10 M, 5 M and 0.1 M): Prepared by diluting the concentrated acid
(Merck, Mumbai, India, Sp. gr. 1.84, 98%) appropriately with water to get 10 M, 5 M and
0.1 M acid solutions.
Preparation of standard OFX solutions (300 and 20 µg ml-1
)
A stock standard solution equivalent to 600 µg ml-1
was prepared by dissolving 60
mg of pure OFX (99.8% pure) in 0.1 M H2SO4 and diluting to the mark in a 100 ml
calibrated flask with the same acid. The stock solution was diluted stepwise with the same
acid to obtain a working concentration of 300 and 20 µg ml-1
for method A and method B,
respectively.
Two brands of tablets, Zenflox-400 (Mankind Pharma Pvt Ltd., New Delhi, India)
and Ofloxin-400 (J. B. Chemicals and Pharmaceuticals Ltd, Mumbai, India), used in the
investigation were purchased from local market and used in the investigation.
2.2.3.0 ASSAY PROCEDURES
2.2.3.1 Method A (using p-TD)
Into a series of 10.0 ml calibrated flasks, 0.0, 1.0, 2.0, 3.0 and 4.0 ml of 300 µg ml-1
standard OFX solution were placed by means of a microburette, and the total volume was
adjusted to 4 ml by adding requisite quantity of 0.1 M H2SO4. To each flask were added 2
ml of 5 M H2SO4 and 1 ml of 6000 µg ml-1
cerium(IV) sulphate, the last being measured
accurately by means of a microbuttete. The content was mixed and let stand for 5 min with
occasional shaking. Finally, 1 ml of 1% p-TD solution was added and diluted to the mark
with 1 M H2SO4, mixed and absorbance of each solution measured at 525 nm against a
water blank.
2.2.3.2 Method B (using 0-DA)
Different aliquots containing 0.0, 0.5, 1.0, 1.5 and 2.0 ml of 20 µg ml-1
of drug
solution were accurately measured and transferred into a series of 10 ml calibrated flasks
32
and the total volume was brought to 2 ml with 0.1 M H2SO4. To each flask were added 5
ml of 10 M H2SO4 and 1 ml of 300 µg ml-1
cerium(IV) sulphate, the last being measured
accurately. The content was mixed and kept aside for 20 min before adding 1 ml of 0.01 %
o-DA to each flask. The volume was brought to the mark with 10 M H2SO4 immediately,
mixed and absorbance measured at 470 nm against water.
In either method, a calibration graph was prepared by plotting the measured
absorbance vs concentration (µg ml-1
) of OFX, and the concentration of the unknown was
read from the calibration graph or computed from the regression equation derived using the
absorbance-concentration data.
2.2.3.3 Procedure for tablets
Twenty tablets were weighed and pulversized. A quantity of tablet powder
containing 60 mg of OFX was transferred into a 100 ml calibrated flask. The content was
shaken with about 70 ml of 0.1 M H2SO4 for 20 min. The mixture was diluted to the mark
with the same acid. It was filtered using Whatman No 42 filter paper. First 10 ml portion
of the filtrate was discarded and the resulting solution (600 µg ml-1
) was diluted to 300
µg ml-1
and used in method A. The tablet extract was diluted to get 20 µg ml-1
OFX and
subjected to analysis by following the procedure described under method B.
2.2.3.4 Placebo blank and synthetic mixtures analysis
Twenty mg of the placebo blank was extracted and its solution was prepared as described
under tablets; and a convenient aliquot was subjected to analysis.
A synthetic mixture was prepared by adding pure OFX (100 mg) to the above
mentioned placebo blank and the mixture was homogenized. Synthetic mixture
containing 60 mg of OFX was weighed and its solution was prepared as described under
section 2.2.3.3. Three different aliquots were subjected to analysis by the general
procedure described in method A after diluting to 300 µg ml-1
with 0.1 M H2SO4. The
stock solution was diluted to get 20 µg ml-1
and used for the assay in method B. The
concentration of OFX was found from the calibration graph or from the regression
equation.
2.2.4.0 RESULTS AND DISCUSSION
Ce(IV) has the ability to rapidly oxidize OFX and to form colored products with
two amines viz. p-toluidine and o-dianisidine [70, 71], as these reactions were
33
successfully employed to develop two indirect procedures for the determination of OFX
in pharmaceuticals. The methods are based on the oxidation of OFX in H2SO4 medium
by a measured excess of Ce(IV) followed by the determination of the residual oxidant by
reacting with p-toluidine or o-dianisidine and measuring the absorbance of the resultant
colored oxidation product at 525 or 470 nm. The possible reaction scheme is shown in
Scheme 2.2.1. The measured absorbance was found to decrease linearly with increasing
concentration of OFX, thus forming the basis for the quantification of drug.
OFX + Ce(IV)H+
Oxidation product
of OFXUnreacted Ce(IV)
+
Unreacted Ce(IV)
NH2
H2N
O
O+
NH2+
H2N
O
O
+
Unreacted Ce(IV)
+
NH2
CH3
p-TD
o-DAOxidation product of o-DA
( measured at 470 nm)
CH2
NH2+
Oxidation product of p-TD
(measured at 525 nm) Method A
H+
Method B
H+
Scheme 2.2.1 Possible reaction scheme for the reaction between Ce(IV) and p-TD and o-
DA in H2SO4 medium.
When a fixed concentration of Ce(IV)SO4 is reacted with increasing concentrations
of OFX, there occurred a concomitant decrease in the concentration of Ce(IV). When
treated with a fixed quantity of p-TD or o-DA, this resulted in decreasing concentrations of
the resultant colored product showing a proportional decrease in absorbance at the
respective analytical wavelength. The decrease in absorbance of the resultant color
compared to the reagent blank was found to be proportional to the concentration of OFX,
and this formed the basis of determination.
34
2.2.4.1 Optimization of variables
Various experimental variables involved in the oxidation of OFX by Ce(IV) and the
latter’s determination with p-TD or o-DA were optimized. Preliminary experiments with p-
TD or o-DA were performed to determine the upper Beer’s law limits for Ce(IV) and these
were found to be 600 µg ml-1
and 30 µg ml-1
with p-TD and o-DA, respectively, under the
optimized experimental conditions. Hence, different concentrations of OFX were reacted
with 1 ml of 6000 µg ml-1
or 300 µg ml-1
Ce(IV) in acid medium before determining the
unreacted oxidant with either p-TD or o-DA. This enabled to evaluate the concentration
ranges over which OFX could be determined by either method.
For both steps involved in the assay, i.e., the oxidation of OFX by Ce(IV) and the
latter’s determination of with p-TD or o-DA, H2SO4 medium was selected after preliminary
experiments. In method A, 2 ml of 5 M H2SO4 in a total volume of 6 ml (overall acidity
~1.7 M) and in method B, 5 ml of 10 M H2SO4 in a total volume of 7 ml (overall ~7 M)
was adequate for the oxidation in 5 min and any delay upto 20 min had no effect in method
A whereas in method B, 10 min was required for complete oxidation of OFX. At lower
acid concentrations oxidation reaction was found to be incomplete. In method A, the
purple color formed (within 5 min) on adding p-TD and peaking at 525 nm was stable upto
90 min; where as in method B, 10 min standing time was necessary for the formation of red
colored product (λmax 470 nm) with a stability period of 20 min. The colored products
formed with p-TD and o-DA required higher H2SO4 concentration for stability, and hence,
1 M and 10 M H2SO4 were used as diluent after adding p-TD and o-DA to the residual
Ce(IV) respectively.
For each system, two blanks were prepared. The first blank, which contained all
reactants except OFX, gave maximum absorbance (intercept of the calibration graph). The
second blank was prepared in the absence of OFX and Ce(IV)SO4 to assess the
contribution of other reactants to the signal (absorbance) of the system. Since the second
blank had insignificant absorbance in both methods, measurements were made against
water.
2.2.4.2 Method validation
The proposed methods were validated for linearity, sensitivity, precision, accuracy,
robustness, ruggedness, selectivity and recovery.
35
Linearity and sensitivity
A linear correlation was found between absorbance at λmax and concentration of
OFX in the ranges given in Table 2.2.1. The graphs are described by the regression
equation:
Y = a + bX
(where Y = absorbance of 1-cm layer of solution; a = intercept; b = slope and X =
concentration in µg ml-1
). Here, Beer’s law is obeyed in the inverse manner. Regression
analysis of the Beer’s law data using the method of least squares was made to evaluate
the slope (b), intercept (a) and correlation coefficient (r) for each system and the values
are presented in Table 2.2.1. A plot of log absorbance and log concentration, yielded
straight lines with slopes equal to 1.00 and 1.01 for method A and method B,
respectively, further establishing the linear relationship between the two variables. The
optical characteristics such as Beer’s law limits, molar absorptivity and Sandell
sensitivity values of both the methods are also given in Table 2.2.1. The limits of
detection (LOD) and quantitation (LOQ) calculated according to ICH guidelines [69]
using the formulae: LOD = 3.3 S/b and LOQ = 10 S/b, (where S is the standard deviation
of blank absorbance values, and b is the slope of the calibration plot) are also presented in
Table 2.2.1. The high values of ε and low values of Sandell sensitivity and LOD indicate
the high sensitivity of the proposed methods.
Accuracy and precision
The assays described under “general procedures” were repeated seven times
within the day to determine the repeatability (intra-day precision) and five times on
different days to determine the intermediate precision (inter-day precision) of the
methods. These assays were performed for three levels of analyte. The results of this
study are summarized in Table 2.2.2. The percentage relative standard deviation (%RSD)
values were ≤ 7.69% (intra-day) and ≤ 4.75% (inter-day) indicating good precision of the
methods. Accuracy was evaluated as percentage relative error (RE) between the
measured mean concentrations and taken concentrations for OFX. Bias {bias % =
[(Concentration found - known concentration) x 100 / known concentration]} was
calculated at each concentration and these results are also presented in Table 2.2.2.
36
Percent relative error (%RE) values of ≤ 4% demonstrates the good accuracy of the
proposed methods.
Table 2.2.1 Sensitivity and regression parameters
Parameter Method A Method B
λmax, nm 525 470
Linear range, µg ml-1
0.0 – 120 0.0 – 4.0
Molar absorptivity(ε), l mol-1
cm-1
2.43 × 103 5.99 × 10
4
Sandell sensitivity*, µg cm
-2 0.1485 0.006
Limit of detection (LOD), µg ml-1
0.63 0.02
Limit of quantification (LOQ), µg ml-1
1.92 0.07
Regression equation, Y**
Intercept (a) 0.6260 0.6446
Slope (b) -0.0051 -0.1579
Standard deviation of a (Sa) 0.3654 0.3436
Standard deviation of b (Sb) 0.0033 0.0973
Variance (Sa2) 0.1339 0.1181
Regression coefficient (r) -0.9973 -0.9995 *Limit of determination as the weight in µg per ml of solution, which corresponds to an
absorbance of A = 0.001measured in a cuvette of cross-sectional area 1 cm2 and l = 1 cm.
**Y=a+bX, Where Y is the absorbance, X is concentration in µg ml-1, a is intercept and b is
slope.
Table 2.2.2 Results of intra-day and inter-day accuracy and precision study
% RE: Percentage relative error; %RSD: Percentage relative standard deviation
Robustness and ruggedness
The robustness of the methods was evaluated by making small incremental
changes in the volume of reagents and contact time after adding the reagents, and the
effect of the changes was studied on the absorbance of the coloured systems. The changes
had negligible influence on the results as revealed by small intermediate precision values
Method
OFX
taken,
µµµµg ml-1
Intra-day accuracy and
precision
(n=7)
Inter-day accuracy and
precision
(n=5)
OFX found,
µµµµg ml-1
%RE %RSD
OFX
found,
µµµµg ml-1
%RE %RSD
A
30.0
45.0
60.0
31.05
46.22
61.09
3.50
2.70
1.82
7.69
4.00
1.85
30.77
46.37
59.96
2.58
3.04
3.75
3.28
4.75
3.46
B
1.0
2.0
3.0
1.02
2.08
3.11
2.00
4.00
3.67
2.58
2.88
1.33
1.04
2.06
3.09
4.00
3.14
2.86
4.58
3.94
3.08
37
expressed as %RSD (≤ 3.62%). Method ruggedness was demonstrated having the
analysis done by four analysts, and also by a single analyst performing analysis on four
different instruments in the same laboratory. Intermediate precision values (%RSD) in
both instances were in the range 0.86 – 4.64% indicating acceptable ruggedness. The
results are presented in Table 2.2.3.
Table 2.2.3 Results of robustness and ruggedness expressed as intermediate
precision (% RSD).
Method OFX
taken*
Robustness Ruggedness
Parameter altered Inter-analysts,
(%RSD)
(n=4)
Inter-
instruments,
(%RSD)
(n=4)
Volume of
Reagent*
(%RSD)
Reaction
time**
A
30.0
60.0
90.0
2.67
3.42
1.83
3.58
2.16
2.78
1.62
1.35
0.96
3.58
4.64
2.85
B
1.0
2.0
3.0
2.45
3.62
2.76
1.64
2.37
1.85
1.26
0.86
1.35
2.72
3.14
4.35 *Volumes of reagents added were 1 ± 0.2 ml. ** In method A, reaction times studied were 5±1
min and in method B reaction times were 10±1 min.
Selectivity
To determine the selectivity of the methods, the placebo solution was subjected to
analysis according to the recommended procedures. The absorbance readings for placebo
blank were nearly zero for both the methods, inferring no interference from the placebo.
The analysis of synthetic mixture solution yielded percent recoveries in the range
98.65 - 102.6 % with standard deviation of 1.10 - 1.24. The results of this study indicate
that the inactive ingredients did not interfere in the assay. This showed a high selectivity
of the methods.
Application to tablets
The described procedures were successfully applied to the determination of OFX
in its pharmaceutical formulations. The results obtained (Table 2.2.4) were statistically
compared with the BP method [4]. The method involved the titration of OFX in
anhydrous acetic acid with acetous perchloric acid to a potentiometric end point. The
results obtained by the proposed methods agreed well with those of reference method and
with the label claim. The results were also compared statistically by a Student’s t-test for
accuracy and by a variance F-test for precision with those of the reference method at 95%
38
confidence level as summarized in Table 2.2.4. The results showed that the calculated t-
and F-values did not exceed the tabulated values inferring that proposed methods are as
accurate and precise as the reference method.
Recovery studies
To further assess the accuracy of the proposed methods, recovery experiment was
performed by applying the standard-addition technique. The recovery was assessed by
determining the agreement between the measured standard concentration and added
known concentration to the sample. The test was done by spiking the pre-analysed tablet
powder with pure OFX at three different levels (50, 100 and 150 % of the content present
in the tablet powder (taken) and the total was found by the proposed methods. Each test
was repeated three times. From this test the percentage recovery values were found in the
range of 97.58 – 104.8 with standard deviation values from 0.33 – 1.50%. Closeness of
the results to 100% showed the fairly good accuracy of the method. These results are
shown in Table 2.2.5.
Table 2.2.4 Results of analysis of tablets by the proposed methods and statistical
comparison of the results with the reference method.
Brand
name Nominal
amount,
(mg/tablet)
Found* (Percent of label claim ± SD)
Reference
method
Proposed methods
Method A Method B
Zenflox** 400
103.5±1.28
101.6±2.14 t = 1.76
F = 2.79
102.1±2.65 t = 1.13
F = 1.13
Ofloxin***
400
101.3±1.08
100.8±2.58 t = 0.43
F = 5.71
99.67±2.06 t = 1.64
F = 3.64 *Average of five determinations. ** Mankind Pharma Pvt Ltd., New Delhi, India;*** J. B. Chemicals and Pharmaceuticals Ltd,
Mumbai, India.
Tabulated t value at the 95% confidence level is 2.77. Tabulated F value at the 95% confidence
level is 6.39.
39
Table 2.2.5 Results of recovery study using standard addition method
Method A Method B
OFX in
tablet
extract,
µµµµg ml-1
Pure
OFX
added,
µµµµg ml-1
Total
OFX
found,
µµµµg ml-1
Pure OFX
recovered
(Percent±SD*)
OFX in
tablet
extract,
µµµµg ml-1
Pure
OFX
added,
µµµµg ml-1
Total
OFX
found,
µµµµg ml-1
Pure OFX
recovered
(Percent±SD*)
30.48
30.48
30.48
15.0
30.0
45.0
45.28
60.51
75.32
98.65±0.86
100.1±1.06
99.65±0.56
1.02
1.02
1.02
0.5
1.0
1.5
1.51
2.04
2.59
97.58±0.77
102.3±0.89
104.5±1.50
30.24
30.24
30.24
15.0
30.0
45.0
45.44
60.42
76.59
101.3±0.89
100.6±0.33
103.0±0.65
1.00
1.00
1.00
0.5
1.0
1.5
1.52
2.02
2.57
104.0±1.25
101.6±1.08
104.8±1.11 * Mean value of three determinations
40
SECTION 2.3
SIMPLE AND SENSITIVE SPECTROPHOTOMETRIC ASSAY OF
OFLOXACIN IN PHARMACEUTICALS BASED ON ION-PAIR REACTION
2.3.1.0 INTRODUCTION
Ion-association is a chemical reaction whereby ions of opposite electrical charge
come together in solution to form a distinct chemical entity. Ions of opposite charge are
naturally attracted to each other by the electrostatic force. This is described by Coulomb's
law.
F is the force of attraction, q1 and q2 are the magnitudes of the electrical charges,
ε is the dielectric constant of the medium and r is the distance between the ions [72].
Sulphonphthalein dyes are commonly used as anionic dyes to form ion-pair complex with
the nitrogenous compound present in positively charged protonated forms.
Ion-pair complexation reaction has been applied for the spectrophotometric
determination of many organic compounds [73-76] to mention a few. In response to the
problem resulting from extraction of the ion pair few articles were published for the
analysis of pharmaceutical compounds namely β-blockers [77], acebutalol hydrochloride
[78], benzalkonium chloride [79], quetiapine fumarate [80], tramadol hydrochloride
[81], pheniramine maleate [82], dothiepin hydrochloride [83], domperidone [84] and
rizatriptan benzoate [85] etc, through ion pair formation without extraction.
In the literature survey present in section 2.0.2.2, two spectrophotometric assay of
OFX based on ion-pair complex formation reaction through extraction step [10, 11]. The
methods were based on the formation of chloroform extractable ion-pair complexes
between OFX and bromophenol blue (BPB), bromothymol blue (BTB) and bromocresol
purple (BCP) [10]. The methods were based on formation of chloroform soluble ion-pair
complex between Tropaelin 000 and Supracene violet 3B [11]. The ion-pairs formed at
specific pH were extracted into chloroform before measurement. These procedures
41
involve strict pH control besides the cumbersome extraction step. Many a time,
incomplete separation may lead to erratic results.
In this section 2.3, the author explored the utilization of simple ion-pair
extraction free technique for the determination of OFX in pure form and in tablets. The
present methods employed bromothymol blue (BTB) in method A and bromophenol blue
(BPB) in method B as chromogenic agents and the resulting ion pairs are measured
directly in dichloromethane. The methods were demonstrated to be more sensitive than
the reported extraction methods [8, 9] for OFX and offer the advantages over the existing
spectrophotometric methods in terms of simplicity, sensitivity and speed.
2.3.2.0 EXPERIMENTAL
2.3.2.1 Apparatus
The instrument is the same that was described in Section 2.2.2.1.
2.3.2.2 Reagents
All chemicals used were of analytical reagent grade and distilled water was used
throughout the study.
BTB and BPB: The solutions of 0.03 % BTB and 0.05 % BPB (both from Loba Chemie,
Mumbai, India) were prepared in dichloromethane (Merck, Mumbai, India).
Preparation of Standard OFX Solution
A stock standard solution containing 250 µg ml-1
OFX was prepared by
dissolving 25 mg of pure OFX in dichloromethane (Merck, Mumbai, India). From this
stock solution prepared 25 and 20 µg ml-1
OFX and used for the assay in Method A and
Method B, respectively.
Tablets used were the same as described in section 2.1.2.0.
2.3.3.0 ASSAY PROCEDURES
2.3.3.1 Method A (using BTB)
Aliquots of 0.25, 0.5……….4.0 ml OFX standard solution in dichloromethane
(25 µg ml-1
) were measured accurately and transferred into a series of 5 ml calibrated
flask. To each flask, 1 ml of 0.03 % BTB was added, diluted to the mark with
dichloromethane and mixed well. The absorbance of the resulting yellow color
chromogen was measured at 410 nm against the reagent blank.
42
2.3.3.2 Method B (using BPB)
Aliquots of 0.25, 0.5……….4.0 ml OFX standard solution in dichloromethane
(20 µg ml-1
) were measured accurately and transferred into a series of 5 ml calibrated
flask. To each flask, 1 ml of 0.05 % BPB was added, diluted to the mark with
dichloromethane and mixed well. The absorbance of the resulting yellow color
chromogen was measured at 410 nm against reagent blank.
In both the methods, a standard graph was prepared by plotting the increasing
absorbance values versus concentration of OFX. The concentration of the unknown was
read from the standard graph or computed from the respective regression equation
derived using the Beer’s law data.
2.3.3.3 Procedure for tablets
Ten tablets were accurately weighed and powdered. A portion equivalent to 25
mg OFX was accurately weighed and transferred into a 50 ml calibrated flask, 30 ml of
dichloromethane was added to the flask and the content was shaken thoroughly for 15-20
min to extract the drug into the liquid phase; the volume was finally diluted to the mark
with dichloromethane, mixed well and filtered using a Whatman No. 42 filter paper. An
aliquot of the filtrate (500 µg ml-1
in OFX) was further diluted to get required
concentrations and analysed for OFX following the procedures described for the
calibration curve.
2.3.3.4 Placebo blank and synthetic mixture analyses
Placebo blank solution was prepared as described under section 2.3.3.3 by taking
20 mg of placebo; and then analysed using the procedures described above.
Ten mg of placebo blank of the composition described above, 10 mg of OFX was
added and homogenized, transferred to 50 ml calibrated flask and the solution was
prepared as described under section 2.3.3.3. After dilution to 20 µg ml-1
OFX, 2.5 ml in
each case was then subjected to analysis by the procedures described above.
2.3.3.6 Procedure for stoichiometric relationship
Job’s method of continuous variations of equimolar solutions was employed
6.805 × 10-5
M each of OFX and BTB (method A) and 5.444 × 10-5
M each of OFX and
BPB (method B) solutions in dichloromethane were prepared separately. A series of
solutions was prepared in which the total volume of OFX and the dye was kept at 5 ml.
43
The drug and reagent were mixed in various complementary proportions (0:5, 1:4,
2:3,………..5:0, inclusive) and completed as directed under the recommended
procedures. The absorbance of the resultant ion-pair complex was measured at 410 nm.
2.3.4.0 RESULTS AND DISCUSSION
OFX is an amino compound containing a piperazine moiety. Hence, attempts
were made to determine it by applying ion pair technique in which basic cationic nitrogen
in OFX forms ion pair with an anionic dye, where a highly yellow colored ion-pair
complex is formed.
Absorption spectra
OFX in dichloromethane does not absorb in the visible region and the new broad
absorptions in the visible region after the addition of the dyes to the fixed concentrations
of the drug indicates the formation of ion pair complex between the drug and dye. Figure
2.3.1 shows the absorption spectra of the ion pair complex, OFX-dye (BTB and BPB) in
dichloromethane. The absorption maximum of the ion pair in dichloromethane is at 410
nm where the absorbance of the reagent blank is insignificant (0.02) in both the methods.
0
0.1
0.2
0.3
0.4
0.5
0.6
360 380 400 420 440 460 480 500 520
Wavelength, nm
Ab
so
rba
nc
e
BTB
BTB (BLANK)
BPB
BPB (BLANK)
Figure 2.3.1 Absorption spectra of ion-pair complexes [OFX (10 µg ml-1
): BTB (0.03%)
and OFX (8 µg ml-1
): BPB (0.05%)] against respective reagent blank
2.3.4.1 Optimization of variables
Optimum reaction conditions for quantitative determination of ion-pair complexes
were established via various preliminary experiments such as choice of organic solvent,
concentration of the dye and reaction time.
44
Choice of organic solvent
A number of organic solvents such as dichloromethane, acetone, methanol,
dioxane and carbon tetrachloride were examined since OFX is soluble in these solvents.
Among these solvents, dichloromethane was preferred as the most suitable solvent
because in this medium, the reagent blank gave negligible blank absorbance and the ion-
pair complex formed was found to exhibit higher sensitivity and stability. In other
solvents, the reagent blank yielded high absorbance value.
Effect of dye concentration
The influence of the concentration of BTB and BPB on the intensity of the color
developed at the selected wavelength and constant drug concentration was studied using
different amounts (0.25-3.0 ml) of 0.03% of BTB in method A and (0.25-2.0 ml) of
0.05% BPB in method B. As shown in Figure 2.3.2 the constant absorbance readings
were obtained between (0.5-3.0 ml) of 0.03% of BTB in method A and (0.5-2.0 ml) of
0.05% BPB in method B. Hence, 1 ml of each 0.03% BTB and 0.05% BPB was used for
method A and method B, respectively.
0
0.1
0.2
0.3
0.4
0.5
0.6
0 0.5 1 1.5 2 2.5 3 3.5
volume of dyes
Ab
so
rba
nc
e
BTB (0.03%)
BPB (0.05%)
Figure 2.3.2 Effect of dye concentration (10 µg ml-1
OFX in method A & 8 µg ml-1
OFX
in method B)
Effect of reaction time
The optimum reaction time for the development of color at ambient temperature
(25±2oC) was studied and it was found that complete color development was
instantaneous in both the methods. The formed color was stable for at least 60 min in
both the cases (Figure 2.3.3).
45
Figure 2.3.3 Effect of time (10 µg ml-1
OFX in method A & 8 µg ml-1
OFX in method B)
Stoichiometric ratio
Job’s method of continuous variations was applied to establish the reaction
stoichiometry. In both the cases, the plot reached a maximum value at a mole fraction of
0.5 which indicated the formation of 1:1 (OFX:Dye) complex (Fig. 2.3.4) between
positive protonated OFX+ and BTB
- or BPB
- anion as shown in Fig. 2.3.5. In OFX, the
nitrogen atom bonded to electron donating methyl group in the piperazine ring is the most
vulnerable one for protonation [86].
0.0 0.2 0.4 0.6 0.8 1.0
0.0
0.1
0.2
0.3
0.4
0.5
0.6
Ab
so
rba
nce
[OFX]/[OFX]+[DYE]
0.0 0.2 0.4 0.6 0.8 1.0
0.0
0.1
0.2
0.3
0.4
0.5
0.6
Absorb
ance
[OFX]/[OFX]+[DYE]
(a) (b)
Figure 2.3.4 Job’s Continuous - variations plots (a) OFX + BTB and (b) OFX + BPB
46
O
NN
NCH3
F
O
OH
O
CH3
O
NN
NCH
3
F
O
OH
O
CH3
(a)
Br
HO
Br
O
C
BrOH
BrO
SO2
BrHO
Br
O
C
BrO
Br
SO3H
BrHO
Br
O
C
BrO
Br
SO3-
+ H+
+
BrHO
Br
O
C
BrO
Br
SO3-
+ H+
+
BrHO
Br
O
C
BrO
Br
SO3-
Bromophenol blue(lactoid ring)
(quinoid ring)
1:1 complex OFX:BPB
OFX
H
(b)
Figure 2.3.5 Possible reaction schemes (a) OFX: BTB and (b) OFX: BPB
Conditional stability constants (Kf) of the ion-pair complexes
The conditional stability constants (Kf) of the ion-pair complexes for OFX were
calculated from the continuous variation data using the following equation [87]:
Kf =
A/Am
[1-A/Am]n+2 CM(n)n
where A and Am are the observed maximum absorbance and the absorbance value when
all the drug present is associated, respectively. CM is the mole concentration of drug at the
maximum absorbance and n is the stoichiometry with which dye ion associates with drug.
47
The log Kf ± CL (Confidence limit at 95 %, n =3) values for OFX-BTB and OFX-BPB
ion-pair associates were 7.339 ± 0.559 and 6.757 ± 0.671, respectively.
2.3.4.2 Method validation
The proposed methods have been validated for linearity, sensitivity, precision,
accuracy, selectivity and recovery.
Linearity and sensitivity
Under optimum conditions a linear relation was obtained between absorbance and
concentration of OFX in the range of 1.25-20.0 µg ml-1
(method A) and 1.0-16.0 µg ml-1
(method B) (Figure 2.3.6). Correlation coefficient, intercept and slope for the calibration
data are summarized in Table 2.3.1. Sensitivity parameters such as apparent molar
absorptivity and sandell sensitivity values, the limits of detection and quantification are
calculated as per the current ICH guidelines [69] are compiled in Table 2.3.1 speak of the
excellent sensitivity of the proposed method. The limits of detection (LOD) and
quantification (LOQ) were calculated according to the same guidelines.
Table 2.3.1 Sensitivity and Regression Parameters
Parameter Method A Method B
λmax, nm 410 410
Linear range, µg ml-1
1.25-20.0
1.0-16.0
Molar absorptivity(ε), l moL-1
cm-1
1.74 x 104
2.18 x 104
Sandell sensitivity*, µg cm
-2 0.021 0.017
Limit of detection (LOD), µg ml-1
0.14 0.08
Limit of quantification (LOQ), µg ml-1
0.43 0.23
Regression equation, Y**
Intercept (a) 0.002 0.0006
Slope (b) 0.046 0.061
Standard deviation of a (Sa) 0.126 0.013
Standard deviation of b (Sb) 0.007 0.001
Regression coefficient (r) 0.999 0.999 *Limit of determination as the weight in µg per ml of solution, which corresponds to an absorbance of
A = 0.001 measured in a cuvette of cross-sectional area 1 cm2 and l = 1 cm. **Y=a+bX, Where Y is the
absorbance, X is concentration in µg ml-1, a is intercept, b is slope.
48
0
0.1
0.2
0.3
0.40.5
0.6
0.7
0.8
0.9
1
0 5 10 15 20 25
Concentration of drug (µg mL-1)
Ab
so
rba
nc
e
0
0.2
0.4
0.6
0.8
1
0 4 8 12 16
Concentration of drug (µg mL-1)
Ab
so
rba
nc
e
Method A Method B
Figure 2.3.6 Calibration curves
Accuracy and precision
In order to evaluate the precision of the proposed methods, solutions containing
three different concentrations of the OFX were prepared and analyzed in seven replicates.
The results obtained from this investigation are summarized in Table 2.3.2. The low
values of the relative standard deviation (% R.S.D) and percentage relative error (% R.E)
indicate the high precision and the good accuracy of the proposed methods. The assay
procedure was repeated seven times, and percentage relative standard deviation (%
R.S.D) values were obtained within the same day to evaluate repeatability (intra-day
precision) and over five different days to evaluate intermediate precision (inter-day
precision).
Table 2.3.2 Results of intra-day and inter-day accuracy and precision study
Method OFX
taken
( µg ml-1
)
Intra-day accuracy and
precision
(n= 7)
Inter-day accuracy and
precision
(n= 7)
OFX
found
( µg ml-1
)
%RE %RSD
OFX
found
( µg ml-1
)
%RE %RSD
A
10.0 9.78 2.21 1.02 9.70 3.00 1.18
15.0 14.81 1.27 1.14 14.61 2.61 1.29
20.0 19.42 2.89 1.22 19.38 3.10 1.31
B
4.0 3.92 2.00 1.17 3.88 3.00 1.24
8.0 7.83 2.12 1.04 7.78 2.75 1.13
12.0 11.73 2.25 1.21 11.7 2.50 1.29 RE. Relative error and RSD. Relative standard deviation.
49
Robustness and ruggedness
Method robustness was evaluated by making small incremental change in dyes
concentrations (n = 3). The %RSD with the altered dyes concentration was ≤ 1.32 %. The
capacity remains unaffected by small deliberate variations. In order to demonstrate the
ruggedness of the methods, determinations at three different concentrations of the drug
were carried out by four different analysts, and also with three different instruments by a
single analyst. The inter-analysts RSD were ≤ 0.91 %, whereas the inter-instrumental
variation expressed as RSD were ≤ 1.08 %. These low values of intermediate precision
demonstrate the robustness and ruggedness of the proposed methods (Table 2.3.3).
Selectivity
The proposed methods were tested for selectivity by placebo blank and synthetic
mixture analyses. The placebo blank solution prepared as described earlier was subjected
to analysis according to the recommended procedures. The resulting absorbance readings
for both the methods were same as the reagent blank, inferring no interference from
placebo. The analysis of synthetic mixture solution prepared above yielded percent
recoveries which ranged from 98.45 to103.6 with standard deviation of 0.91–1.13 in all
the cases. The results of this study indicated that the inactive ingredients did not interfere
in the assay. These results further demonstrate the accuracy as well as the precision of the
proposed methods.
Table 2.3.3 Results of robustness and ruggedness expressed as intermediate precision
(%RSD)
Method
OFX
taken,
µg
ml-1
Method robustness
Method ruggedness Parameter altered
BTB, ml*
%RSD
(n = 3)
BPB,
ml*
%RSD
(n = 3)
Inter-analysts’
%RSD
(n = 4)
Inter-
instruments’
%RSD
(n = 3)
A 4.0 1.21 1.18 0.83 1.35
8.0 1.34 1.11 0.98 1.21
12.0 1.27 1.23 0.92 1.27
B
4.0 1.12 1.41 0.89 1.52
8.0 1.18 1.32 0.96 1.47
12.0 1.23 1.29 0.83 1.41 * Dyes (BTB and BPB) volumes used were 0.8, 1.0 and 1.2 ml
50
Application to tablets
In order to evaluate the analytical applicability of the proposed methods to the
quantification of OFX in commercial tablets, the results obtained by the proposed
methods were compared to those of the reference method [4] by applying Student’s t-test
for accuracy and F-test for precision. The reference method involved the titration of OFX
in anhydrous acetic acid with acetous perchloric acid to a potentiometric end point
detection. The results (Table 2.3.4) show that the Student’s t- and F-values at 95 %
confidence level are less than the tabulated values, which confirmed that there is a good
agreement between the results obtained by the proposed methods and the reference
method with respect to accuracy and precision.
Table 2.3.4 Results of analysis of tablets by the proposed methods and statistical comparison with the reference method
Tablet Brand
name
Label
claim,
mg/tablet
Found* (Percent of label claim ± SD)
Reference
method
Method A Method B
Zenflox**
400
99.92 ± 1.11
98.12 ± 1.19 t=2.47
F= 1.15
98.87 ± 1.15
t=1.47
F= 1.07
Ofloxin***
400
100.3 ± 1.06
98.95 ± 1.12
t=1.96
F=1.12
102.1 ± 1.17
t=2.55
F=1.21 *Mean value of five determinations. ** Mankind Pharma Pvt Ltd., New Delhi, India;*** J. B. Chemicals and Pharmaceuticals Ltd, Mumbai,India. The value of t (tabulated) at 95 % confidence level and for four degrees of freedom is 2.77. The value of F (tabulated) at 95 % confidence level and for four degrees of freedom is 6.39.
Recovery studies
The accuracy and validity of the proposed methods were further ascertained by
performing recovery studies. Pre-analysed tablet powder was spiked with pure OFX at
three concentration levels (50, 100 and 150 % of that in tablet powder) and the total was
found by the proposed methods. In all cases, the added OFX recovery percentage values
ranged of 98.60-102.9 % with standard deviation of 1.07-1.21 (Table 2.3.5) indicating
that the recovery was good, and that the co formulated substance did not interfere in the
determination.
51
Table 2.3.5 Results of recovery studies by standard addition method
Method
Tablet
studied
OFX in
tablet,
µg ml-1
Pure
OFX
added,
µg ml-1
Total
found,
µg ml-1
Pure OFX
recovered*
Percent ± SD
A Ofloxin-400
5.0 2.5 7.48 99.20 ± 1.14
5.0 5.0 9.93 98.60 ± 1.07
5.0 7.5 12.45 99.33 ± 1.11
B
Ofloxin-400
4.0 2.0 6.02 101.0 ± 1.17
4.0 4.0 8.12 102.9 ± 1.09 4.0 6.0 9.97 99.50 ± 1.21
*Mean value of three measurements.
52
SECTION 2.4
HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC DETERMINATION
OF OFLOXACIN IN PHARMACEUTICALS AND HUMAN URINE
2.4.1.0 INTRODUCTION
Chromatography, literally "color writing", was first employed by Russian scientist
Michael Tsvet in 1900 for the separation of plant pigments. Chromatographic technique
was developed subsequently as a result of the work of Archer John Porter Martin and
Richard Laurence Millington Synge during the 1940s and 1950s, and their work
encouraged the rapid development of several chromatographic techniques including
HPLC [88].
The most characteristic feature of the development in the methodology of
pharmaceutical and biomedical analysis during the past 25 years is that HPLC became
undoubtedly the most important analytical method for identification and quantification of
drugs, either in their active pharmaceutical ingredient or in their formulations [89]. HPLC
offers enhanced detection sensitivity, improved accuracy, and reproducibility of drug
analysis in the course of drug research, development and quality control testing of
marketed drug products.
The literature survey on the HPLC methods for the assay of OFX presented in
Section 2.0.2.3 reveals that the currently available HPLC methods for OFX in
pharmaceutical [12-16] or urine [24, 25] are associated with one or other disadvantages.
Hence the aim of the present work was to develop and validate a simple and rapid HPLC
method for determination of OFX in its tablets and human urine without involving any
unwelcome steps like use of the internal standard, solid-phase or liquid-liquid extraction
step etc. The proposed method is simpler than all the reported methods for urine, and is
more sensitive than the reported HPLC methods for OFX in its dosage forms. The details
of method development and method validation are presented in this section, 2.4.
2.4.2.0 EXPERIMENTAL
2.4.2.1 Apparatus
Chromatographic analysis was carried out using Alliance Waters HPLC system
equipped with Alliances 2657 series low pressure quaternary pump, a programmable
53
variable wavelength UV-visible detector, Waters 2996 photodiode array detector and
auto sampler. Data were collected and processed using Waters Empower 2.0 software.
2.4.2.2 Reagents
All the reagents used were of analytical reagent grade and all solutions were
prepared in bi-distilled water.
Phosphate buffer (pH 7.0): The solution was prepared by dissolving 0.35 g of
potassium dihydrogen phosphate in 100 ml of water and the pH was brought to 7.0 by
adding 4% orthophosphoric acid using a pH meter.
Preparation of standard solutions
A 400 µg ml-1
stock standard solution of OFX was prepared by dissolving 40 mg
of the pure drug in mobile phase solution [potassium dihydrogen phosphate buffer (pH
7.0) and methanol (35:65, v/v)]. Working solutions were prepared by diluting the stock
solution with the mobile phase. The same tablets mentioned in the previous section were
also used in this work.
2.4.3.0 ASSAY PROCEDURES
2.4.3.1 Chromatographic conditions
Chromatographic assay was performed using an Intersil C8-3 (5 µm, 3.9 × 150
mm i.d.,) column. The mobile phase was composed of potassium dihydrogen phosphate
buffer (pH 7.0) and methanol (35:65, v/v). The flow rate was maintained at 1.0 ml min-1
.
The column effluent was monitored on UV detector set at 254 nm.
2.4.3.2 Procedure for preparation of calibration curve
Working solutions equivalent to 0.05 - 60.0 µg ml-1
OFX were prepared by
diluting appropriate aliquots of the stock solution. Aliquots of 10 µL were injected
(triplicate) and eluted with the mobile phase under the reported chromatographic
conditions. The average peak area versus the concentration of OFX in µg ml-1
was
plotted. Alternatively, corresponding regression equation was derived using the mean
peak area-concentration data and the concentration of the unknown was computed from
the regression equation.
54
2.4.3.3 Procedure for tablets
Twenty tablets were accurately weighed, finely pulverized and mixed using a
mortar and pestle. An amount of tablet powder equivalent to 40 mg of OFX was weighed
and transferred into a 100 ml calibrated flask, 50 ml of mobile phase was added and was
sonicated for 20 min in an ultrasonic bath to complete dissolution of the OFX, the content
was then diluted to the mark with the mobile phase, mixed well and filtered using a
Whatman No. 42 filter paper. Aliquots of this solution were successively diluted with the
mobile phase and then subjected to analysis as per the general procedure described for the
calibration curve.
2.4.3.4 Placebo blank analysis
Employing 20 mg of the placebo prepared as described in section 2.1, its solution
was prepared as described under section 2.4.3.3 and then subjected to analysis.
2.4.3.5 Procedure for synthetic mixture analysis
To assess the role of the inactive ingredients on the assay of OFX, a synthetic
mixture was separately prepared by adding 20 mg of OFX to about 20 mg placebo
mentioned above. The drug was extracted and solution prepared as described under the
section 2.4.3.3. The solution after appropriate dilution was analyzed following the
recommended procedure.
2.4.3.6 Procedure for analysis of spiked human urine
Urine sample was collected from a healthy male aged about 34 years. Urine
samples, 3 ml each were spiked with varying amounts of OFX (10, 20 and 40 µg ml-1
)
and 2 ml of methanol (urine protein precipitating agent). The content was vortexed for 30
s and the precipitated protein was separated out by centrifugation for 20 min at 4000 rpm.
The centrifugate was mixed with the mobile phase and the pH was adjusted to 7 (by
adding 4% orthophosphoric acid/1% KOH). The volume of the final solution was brought
to 10 ml by adding the mobile phase. Aliquots of 20 µL were injected (triplicate) and
eluted with the mobile phase under the reported chromatographic conditions. The
concentration of OFX was found using regression equation and the percentage recovery
of OFX was calculated.
55
2.4.4.0 RESULTS AND DISCUSSION
Method development and optimization
The proposed method permits the quantitation of OFX in commercial tablets and
in human urine. In order to obtain good linearity, sensitivity and selectivity, the method
was optimized and validated in accordance with the current ICH guidelines [64].
A well defined symmetrical peak and good results were obtained upon measuring
the response of eluent under the optimized conditions after thorough experimental trials
that could be summarized as follows:
Choice of column
Two different columns were used for performance investigations, including
hypersil BDS C8 (250 mm x 4.0 mm i.d, 5.0 µm particle size) thermo column, Intersil C8-
3 (5 µm, 3.9 × 150 mm i.d.,) column and chromatopack (250 mm x 4.6 mm i.d., 5 µm
particle size) column. The experimental studies revealed that the Intersil C8-3 column
was more suitable since it gave better sensitivity.
Choice of wavelength
Spectroscopic analysis of the compound showed that OFX has maximum UV
absorbance at 254 nm. Therefore, the chromatographic detection was performed at 254
nm.
Mobile phase composition
Chromatographic conditions were optimized by changing the mobile phase
composition and buffers used in the mobile phase. Different experiments were carried out
to optimize the mobile phase. Several modifications in the mobile phase composition
were performed in order to study the possibilities of changing the selectivity of the
chromatographic system. These modifications included the change of the type and ratio
of the organic modifier, the pH, the strength of the phosphate buffer, and the flow rate.
Precise and accurate results with maximum number of theoretical plates and good peak
were obtained when the mobile phase of potassium dihydrogen phosphate buffer (pH 7.0)
and methanol (35:65, v/v) was used.
Type of organic modifier
Methanol was replaced by acetonitrile but it did not give a good peak. Methanol
was the organic modifier of choice giving nice, elegant and highly sensitive peak.
56
Ratio of organic modifier
The effect of ratio of organic modifier on the selectivity and retention time of the
test solute was investigated using mobile phases containing 25-75% methanol. At 65%
methanol well defined peak and the highest number of theoretical plates were observed.
Effect of pH and ionic strength of buffer
The effect of pH of the mobile phase on the selectivity and retention time of the
test solute was investigated using mobile phases of pH ranging from 4.0 – 9.0. The results
revealed that pH 7.0 was most appropriate and giving well defined peak and the highest
number of theoretical plates. At lower and higher pH, non-symmetrical peak and smaller
number of theoretical plates were observed. Therefore, pH 7.0 was fixed as optimum.
The same trend was observed after making alteration in the ionic strength of the buffer
and ~0.4% phosphate buffer was used as working buffer throughout the investigation.
The effect of flow rate
The effect of flow rate on the symmetry, sensitivity and retention time of the peak
was studied and a flow rate of 1 ml min-1
was optimal for better symmetry and reasonable
retention time.
2.4.4.2 Method validation
Linearity and sensitivity
Stock standard solution of OFX (400 µg ml-1
) was appropriately diluted with the
mobile phase to obtain solutions in the concentration range 0.05 - 60 µg ml-1
OFX.
Twenty microlitre of each solution was injected in triplicate onto the column under the
operating chromatographic conditions described above. The least squares method was
used to calculate the slope, intercept and the correlation coefficient (r) of the regression
line. The relation between mean peak area Y (n=3) and concentration, X expressed by the
equation Y = 17718.5476 X + 2820.7403 (r2 = 0.9999), was linear. A plot of log peak
area Vs log concentration was a straight line with the slope of 0.9240 and this coupled
with a high value of the correlation coefficient (r-value >0.999) indicated excellent
linearity between mean peak area and concentration in the range 0.05 - 60.0 µg ml-1
OFX. Related statistical data are presented in Table 2.4.1.
The limit of quantification (LOQ) was determined by establishing the lowest
concentration that can be measured according to ICH recommendations [69], below
57
which the calibration graph is non linear and was found to be 50 ng ml-1
. The limit of
detection (LOD) was determined by establishing the minimum level at which the analyte
can be reliably detected and it was found to be 17 ng ml-1
.
Selectivity
A chromatogram obtained for placebo solution is shown in Figure 2.4.1 and did
not show any interference from the above substances as shown by the retention time
1.86±0.02.
The peak area value resulting from 40 µg ml-1
OFX in synthetic mixture solution
had nearly the same as that obtained for pure OFX solutions of identical concentration.
This unequivocally demonstrated the non-interference of the inactive ingredients in the
assay of OFX. Further, the slope of the calibration plots prepared from the synthetic
mixture solution was about the same as that prepared from pure drug solution. Method
selectivity was checked by comparing the chromatograms obtained for placebo blank
(Figure 2.4.1), pure OFX solution (Figure 2.4.2), synthetic mixture and tablet solution
(Figure 2.4.3). An examination of the chromatograms of the above solutions revealed the
absence of peaks due to additives present in tablet preparations and urine (Figure 2.4.4).
Table 2.4.1 Regression and Sensitivity parameters
Parameters Value
Linearity range, µg ml-1
0.050 – 60.0
Regression (Y* = a + bX)
Slope (b) 17718.55
Intercept (a) -2820.74 Standard deviation of intercept (Sa) 4804.85
Standard deviation of Slope (Sb) 99.54
Correlation co-efficient (r) 0.9999 Limit of detection (LOD, ng ml
-1) 17.0
Limit of quantification (LOQ, ng ml-1) 50.0
Variance (Sa2) 2.31 × 10
7
ntSa
/±
5041.24
ntSb
/±
104.44
*Y = a+bX, where Y is the area and X concentration in µg ml-1.
ntSa
/± =confidence limit for intercept, ntSb
/± =confidence limit for slope.
58
Figure 2.4.1 Typical chromatogram obtained for placebo blank.
Figure 2.4.2 Typical chromatogram obtained for pure OFX solution (20 µg ml
-1).
Figure 2.4.3 Typical chromatogram obtained for OFX from tablet extract (20 µg ml
-1).
59
Figure 2.4.4 Typical chromatogram obtained for OFX from spiked human urine
(20 µg ml-1
)
Accuracy and precision
Method precision was evaluated from the results of seven independent
determinations of OFX at three different concentrations, 20.0, 40.0 and 60.0 µg ml-1
on
the same day. The inter-day and intra-day relative standard deviation (RSD) values for
peak area and retention time for the selected concentration of OFX were less than 1%.
The method accuracy, expressed as relative error (%) was determined by calculating the
percent deviation found between concentrations of OFX injected and concentrations
found from the peak area. This study was performed by taking the same three
concentrations of OFX used for precision estimation. The intra-day and inter-day
accuracy (expressed as %RE) was less than 2% and the values are compiled in Table
2.4.2.
Table 2.4.2 Results of intra-day and inter-day accuracy and precision study
OFX
injected,
µg ml-1
Intra-day accuracy and precision Inter-day accuracy and precision
OFX
found* %RE %RSD
** %RSD
***
OFX
found* %RE %RSD
** %RSD
***
20.0 19.70 1.50 0.23 0.10 20.18 0.90 0.58 0.12
40.0 40.32 0.80 0.19 0.07 40.65 1.63 0.44 0.13
60.0 59.93 0.12 0.08 0.10 60.6 0.93 0.48 0.09 *Mean value of seven determinations. **Based on peak area. ***Based on retention time
60
Robustness
To determine the robustness of the method small deliberate changes in the
chromatographic conditions like detection wavelength and column temperature were
made, and the results were compared with those of the optimized chromatographic
conditions. The results indicated that changing the detection wavelength (±1 nm) had
some effect on the chromatographic behavior of OFX. However, the alteration in the
column temperature (±1 °C) had no significant effect. The results of this study expressed
as %RSD are summarized in Table 2.4.3.
Application to tablets
The developed and validated method was successfully applied to the assay of
OFX commercial tablets. The results shown in Table 2.4.4 are in good agreement with
those obtained with the BP method [4]. The method involved the titration of OFX in
anhydrous acetic acid with acetous perchloric acid to a potentiometric end point.
Table 2.4.3 Results of robustness study (OFX concentration, 40 µg ml-1
, n = 3)
Chromatogra
phic
condition
Modification
(n = 3)
Peak area precision (n=3) Retention time
precision (n=3)
Mean area ±
SD %RSD
Mean RT ±
SD,
(min)
%RSD
Wavelength
(nm)
253 651854 ±
26142 4.01 1.82±0.01 0.49 254
255
Column
temperature
(°C)
24 711936 ±
1565.9
0.22 2.12±0.03 1.42 25
26
Table 2.4.4 Results of analysis of tablets by the proposed method and statistical
comparison with reference method
Tablet
brand
name**
Nominal
amount, mg
Found* (Percent of label claim ± SD)
Reference method Proposed method
OF-400 400 99.04±1.58 98.56±0.64
t = 0.68
F = 6.09
Zenflox 400 400 101.3±1.26
102.5±0.85
t = 1.79 F = 2.19
*Mean value of five determinations
61
Application to spiked human urine sample
The developed and validated method was successfully applied to determine OFX
in spiked urine sample with satisfactory recovery (Table 2.4.5). Figure 2.4.4 shows the
OFX peak obtained from spiked human urine. The recovery of OFX from urine sample
was measured under the procedure as described above. The recovery for OFX in spiked
human urine analysis was calculated at three concentrations (10.0, 20.0 and 40.0 µg ml-1
).
The recovery for a OFX-spiked human urine samples was in the range of 97.15 –
100.2%.
Recovery studies
To further assess the accuracy and reliability of the method, recovery studies via
standard addition method was performed. To the pre-analyzed tablet powder, pure OFX
was added at three levels and the total was found by the proposed method. Each test was
triplicated. When the test was performed on two different brands of tablets, the percent
recovery of pure OFX was in the range of 94.66 – 103.1 with standard deviation values of
0.36 – 0.84. The results indicated that the method is very accurate and that common
excipients found in tablet preparations did not interfere. The results are compiled in Table
2.4.6.
Table 2.4.5 Results of recovery of OFX in spiked urine sample
Spiked OFX
concentration
(µg ml-1
)
Found ± SD*
% Recovery
10.0
20.0 40.0
10.02±0.04
19.50±0.14 38.90±0.02
100.2
97.84 97.15
*Mean value of five determinations.
62
Table 2.4.6 Results of recovery study by standard addition method
Tablet studied
OFX in
tablet, µg
ml-1
Pure OFX
added,
µg ml-1
Total found,
µg ml-1
Pure OFX
recovered*,
Percent ± SD
OF-400
Zenflox 400
19.71 10.0 28.47 94.66±0.76
19.71 20.0 39.37 98.32±0.36
19.71 30.0 48.38 95.58±0.84
20.50 10.0 30.66 101.6±0.42
20.50 20.0 41.12 103.1±0.74
20.50 30.0 51.31 102.7±0.65 *Mean value of three determinations
63
SECTION 2.5
CONCLUSIONS ON CHAPTER 2-Assessment of the Methods
A comparison of performance characteristics of the proposed titrimetric and
spectrophotometric methods with those of the existing methods is presented in Table
2.6.1.
Table 2.5.1 Comparison of performance characteristics of the proposed methods with the
existing methods A. Titrimetry
Sl.
No.
Reagent Titration conditions
Range,
mg
Remarks
Ref.
No.
1. Anhydrous
acetic acid & perchloric
acid
Non-aqueous potentiometric
titration of OFX in anhydrous acetic acid with perchloric acid
- Fairly large quantities of
0.300 g of OFX for each titration.
3
2. Acetic
anhydride and
perchloric
acid
0.1 g of OFX in 275 ml acetic
anhydride titrated against perchloric acid to a
potentiometric end point
detection.
- Employ large amount of
acetic acid and acetic anhydride
4
3 NBS Unreacted NBS titrated
iodometrically
1-8 narrow liner range, use
of hazardous and unstablen reagent
5
4. Ce(IV)
a. Direct titration of
OFX with Ce(IV)SO4
using ferroin as
indicator
b. Oxidation of OFX by
Ce(IV)SO4,and back
titration of residual
Ce(IV)SO4 with iron
(II) ammonium
sulphate in H2SO4
medium
1.5-15.0
Uses stable reagent,
wide range of
determination and
used aqueous medium
Present
method
64
B. Spectrophotometry
Sl.
No. Reagent/s used* Methodology
λλλλmax
(nm)
Linear range
(µg ml-1
)
(ε = l mol-1
cm-1
)
Remarks Ref
1
NBS-
indigocarmine
NBS-metanil
yellow
Increase in absorbance
of blue color measured
Increase in absorbance
of pink color measured
610
530
0.5-5.0
(ε= 5.53 x 104)
0.3-3.0
(ε= 9.24x104)
Use of unstable
and hazardous reagent
5
2.
Ce(IV)-MBTH
Oxidative coupling
reaction product
measured
640
1-10
Uses expensive
reagent, less stable species
measured.
6
3.
Iron(III) nitrate
Ambered colored
complex was measured
370
0-62.5 (A1% = 207)
Buffers used.
7
4.
Iron(III)
chloride/HCl
Yellow complex measured
410 20-160
Less sensitive
8
5
Iron(III)-
ammonium sulphate in acid
medium
Complex formed
measured 420
10-120
(ε=2.51 x 103)
Less sensitive 9
6.
a) Bromophenol blue
Chloroform extractable
yellow colored 1:1ion-
pair complex was measured
410 5-25
(ε= 1.04 x 104) Require close pH
control and involve extraction
step
10 b) Bromothymol
blue 415
2-15 (ε= 2.01 x 10
4)
c)Bromocresol
purple 410
2-20 (ε= 1.04 x 10
4)
7.
a) Tropaeolin 000
(TP 000)
Red chloroform extractable ion-pair
complex measured
485
2.5-30 (ε= 8.24 x 10
3)
Require close pH
control and involve
extraction step
11
b) Supracene Violet 3B (SV
3B)
Chloroform extractable
ion-pair complex
measured
575
2.5-25 (ε= 1.09 x 10
4)
8.
Citric acid-acetic
anhydride Pink colored
chromogen measured 552
5-55 (ε= 6.04 x 10
3)
Boiling for 20 min required.
12
65
9.
a) Ce(IV)-p-TD
Coloured product
formed between Ce(IV)-
p-TD measured
525 0.0 - 120
(ε= 2.4 x 103)
Very simple,
sensitive and
precise. No
heating /extraction
involved. Free
from critical
experimental
variables. stable coloured species
measured
This
work
b) Ce(IV)-o-DA
Coloured product formed between Ce(IV)-
o-DA measured
470 0.0 - 4.0
(ε= 5.99 x 104)
10.
a) Bromothymol
blue
b) Bromophenol
blue
In both the methods,
resulting yellow colored
ion-pair complexes were
measured.
410
a) 1.25-20 µg ml-1
(ε=1.74x 104)
b) 1.0-16 µg ml-1
(ε =2.18 x 104)
Highly sensitive
with wide linear
dynamic ranges,
no heating or
extraction step, no
pH-adjustment,
single step
reaction
This
work
C. High performance liquid chromatography
Sl.
No.
Chromatographic conditions Detector Linear
range,
µg ml-1
Remarks Ref. No.
1. A column containing octadecyl silane
chemically bonded to porous silica
particles (Waters Spherisorb, 5 microm ODS 1, 4.6 x 150 mm) with
0.05 M phosphate buffer-acetonitrile
(65:35, v/v), and the pH is adjusted to
2.7 with orthophosphoric acid as mobile phase
UV detection at
210 nm
24-120
Less
sensitive,
used propylparabe
n (POP) as
the internal
standard
14
2. C18 column containing mobile phase used was a combination of
acetonitrile:0.25M potassium
dihydrogen phosphate buffer (80:20)
with 0.5% v/v of triethylamine and the pH was adjusted to 2.5 by adding
orthophosphoric acid.
UV detection at 320 nm
5 -25
Less sensitive,
narrow linear
range
15
3. Phenomenex C18 column with 0.24%
sodium lauryl sulphate: acetonitrile:
acetic acid (pH-4.0) (58:40:02) as the
mobile phase.
UV detection at
295 nm
160 - 240
Less
sensitive,
narrow linear
range
16
66
4. Kromasil C8, 5µ, 15 cm × 4.6 mm id
column with mobile phase consisting of 0.5%v/v Triethylamine buffer of
pH 3.0 and Acetonitrile in the ratio of
73: 27
PDA detection at
303 nm
10-50
Less
sensitive, narrow linear
range
15
5. C-18 column (RP-18, 5µ) coupled
with a guard column of same material, in isocratic mode with
mobile phase mixture of Acetonitrile:
Water: Tri ethylamine (25:75:1)
UV detection at
300 nm
NA NA
16
6. Intersil C8-3 (5 µm, 3.9 × 150
mm i.d.,) column at ambient
temperature with mobile phase
consisting of potassium
dihydrogen phosphate buffer
(pH 7.0) and methanol (35:65,
v/v)
UV-detection
at 254 nm
0.05 - 60 Highly
sensitive,
wide linear
dynamic
range and
uses no
internal
standard
This
work
Two titrimetric, four spectrophotometric and one HPLC methods were developed
and optimized for the determination of ofloxacin. The titrimetric and spectrophotometric
procedures are based on well established and characterized redox and complexation
reactions. For redox reaction Ce(IV)SO4 has been used as an oxidizing agent in both titrimetry
and spectrophotometry.
A close examination of the performance characteristics of the existing titrimetric
and spectrophotometric methods summarized in Table 2.6.1 reveals that most of the
reported methods are suffer from one or the other disadvantage. Only two titrimetric
methods have been reported for the assay of OFX. Both the methods used perchloric acid
as titrant [3, 4]. In USP method, a fairly large quantity of OFX is required for each
titration and used non-aqueous medium in both the methods [3, 4]. The proposed
titrimetric methods employed Ce(IV)SO4 which is a very stable reagent and the methods
are simple and applicable over a wide range of determination (1.5-15 mg).
The author used Ce(IV)SO4 and ion-pair complex forming reagents like
bromothymol blue and bromophenol blue for the spectrophotometric assay of OFX in
pharmaceuticals. The Ce(IV)SO4 has earlier been used for the assay of OFX based on the
oxidative coupling reaction with MBTH [5]; however MBTH is a very expensive reagent
and the reaction product is less stable. Ion-pair complex formation reactions have also
67
been applied for the assay of OFX using same dyes bromothymol blue and bromophenol
blue. However, the reported methods suffer from certain disadvantages like extraction of
ion-pair complex in chloroform and close pH control. In contrast the proposed methods
are found to be free from such unwelcome experimental steps. Among the proposed
methods, the method employing Ce(IV)SO4 – o-DA is the most sensitive method with ε
value of 5.99 × 104 l mol
-1 cm
-1 and the methods using Ce (IV)SO4 – p-TD quantifies
OFX concentration over a widest linear dynamic range (0.0-120 µg ml-1). The use of
strong oxidizing agent like Ce(IV)SO4 is certainly a non-selective reagent, yet the
reaction is found to be selective under the optimized experimental conditions and offers
quantification of OFX in wide range.
The specificity of the proposed HPLC method is higher than the titrimetric and
spectrophotometric methods proposed so far. Some of the remarkable advantage of the
proposed HPLC method are- it does not use an internal standard, short retention time and
run time and flow rate (1.0 ml min-1
), makes the method attractive since these features
help in saving cost and time of analysis.
The developed methods have been validated according to the current ICH guidelines.
All the methods were optimized for maximum sensitivity, demonstrated being robust and
rugged, and fairly accurate and precise. The accuracy and precision of all the methods are
evaluated in terms of % RE and % RSD, respectively. All the titrimetric and
spectrophotometric methods are fairly accurate and precise. Upon comparing the results
of the proposed methods with those of the reference method using Student’s t-test and F-
test, all calculated values were found below the tabulated t- and F- value. The results of
the method robustness and ruggedness, both express in terms of %RSD, were each less
than 4.64 % for all the proposed methods.
68
REFERENCES
1. K. Sat, U. Matsura, M. Inone, T. Ueno, Y. Osada, H. Ogawa, M. Mitshuhashi,
Antimicrob. Agents Chemother,. 22 (1982) 548.
2. N. Ichihara, H. Tachizawa, M. Tsumura, T. Une, K. Sato, J. Antimicrob.
Chemother., 32 (1984) 118.
3. United States Pharmacopoeia, National Formulary XVII, Rockville, USA
Convention, 22 (1990) 912.
4. British Pharmacopeia. Her Majesty’s Stationery office, London, 3 (2003) 1357.
5. K. B. Vinay, H. D. Revanasiddappa, O. Z. Devi, P. J. Ramesh, K. Basavaiah, Braz.
J. Pharm. Sci., 47 (2011) 251.
6. C. S. P. Sastry, K. Rama Rao, D. Siva Prasad, Indian Drugs 32 (1995) 172.
7. J. E. Chukwuenweniwe, O. A. Solomon, O. A. John, J. Antimicrob. Chemother., 39
(1997) 639.
8. S. C. Mathur, Y. Kumar, N. Murugesan, Indian Drugs 29 (1992) 376.
9. K. B. Vinay, H. D. Revanasiddappa, M. R. Divya, N. Rajendraprasad, Ecl.
Quim.,34 (2009) 65.
10. Y. M. Issa, F. M. Abdel-Gawad, M. A. Abou, Anal. Lett., 30 (1997) 2071.
11. C. S. P. Sastry, K. Rama Rao, D. Siva Prasad, Talanta, 42 (1995) 311.
12. P. U. Patel, B. N. Suhagia, M. M. Patel, Ind. Pharm., 6 (2007) 59.
13. K. B. Vinay, H. D. Revanasiddappa, M. R. Divya, N. Rajendraprasad Ecletica
Quimica, 34 (2009) 65.
14. M. S. Ali, M. Ghori, A. Saeed, J. Chromatogr. Sci., 40 (2002) 429.
15. R. R. Kalta, R. Sharma, S. C. Chaturvedi, Indian J. Pharm. Sci., 70 (2008) 491.
16. R. Siva Kumar, P. Kumar Nallasivan, S Saravanakumar, C. S. Kandasamy, R.
Venkatnarayanan, Asian J. Research Chem., 2 (2009) 43-45.
17. J. Dharuman, M. Vasudevan, K. N. Somasekaran, B. Dhandapania, P. D. Ghodea,
M. Thiagarajana, Int. J. PharmTech Res., 1 (2009) 121.
18. R. Singh, M. Maithani, S. K. Saraf, S. Saraf, R. C. Gupta, Eurasian J. Anal.
Chem., 4 (2009)161.
19. R. Singh, M. Maithani, S. K. Sarai, S. Saraf, R. C. Gupta, Eurasian J. Anal. Chem.,
4 (2009) 161.
69
20. A. J. N. Groenveld, J. R. B. Brouwers, Pharm. World Sci., 8 (1986) 79.
21. T. Ohkubo, M. Kudo, K. Sugawara, Anal. Sci., 7 (1991) 741.
22. G. M. Auten, L. C. Preheim, M. Sookpranee, M. J. Bittner, T. Sookpranee, A.
Vibhagool, Antimicrob. Agents Chemother., 35 (1991) 2558.
23. M. Amini, Kh. Abdi, M. Darabi, A. Shafiee, J. Appl. Sci., 5 (2005) 1655.
24. M. A. Gracia, C. Solans, A. Calvo, M. Royo, E. Hernandez, R. Rey, M. A.
Bregante, Chrmatographia, 55 (2002) 431.
25. L. Hing-Biu, T. E. Peart, M. L. Svoboda, J. Chromatogr. A, 1139 (2007) 45.
26. B. Shao, X. Sun, J. Zhang, J. Hu, H. Dong, Y. Yang, J. Chromatogr. A, 1182 (2008)
77.
27. A. Le Coguic, R. Bidault, R. Farinotti, A. Dauphin, J. Chromatogr., 434 (1988)
320.
28. O. Okazaki, H. Aoki, H. Hakusui, J. Chromatogr., 563 (1991) 313.
29. Y. Hongyuan, R. K. Ho, J. Liq. Chromatogr. Rel. Technol., 30 (2007) 1497.
30. B. Srividya, R. M. Cardoza, P. D. Amin, Indian Drugs, 40 (2003) 41.
31. F. A. El-Yazbi, Spectrosc. Lett., 25 (1992) 279.
32. Q. J. Gong, J. L. Quiao, L. M. Du, C. Dong, Talanta, 53 (2000) 359.
33. O. Ballesteroes, J. L. Vichez, A. Navalon, J. Pharm. Biomed.l Anal., 30 (2002)
1103.
34. P. Gurumurthy, G. Ramachandran, A. K. Hemanthkumar, P. Venkatesan, P.
Chandrasekaran, P. R. Narayana, Indian J. Pharmacol., 30 (1998) 263.
35. E. Michael, El-Kommos, A. Gamal, Saleh, M. Samia, El-Gizawi, A. Mohamad,
Talanta, 60 (2003) 1033.
36. M. Tuncel, Z. Atkosar, Pharmazie, 47 (1992) 642.
37. G. Zhou, J. Pan, Anal. Chim. Acta, 307 (1995) 49.
38. M. Rizk, F. Belal, F. Aly, N. El-Enany, Talanta, 46 (1998) 83.
39. J. Wu, H. Zhao, L. Wei, T. Z. Ai, X. Z. Dong, Fenxi Huaxue, 29 (2001) 1106.
40. A. Tamer, Anal. Chim. Acta, 231 (1990) 129.
41. C. Yang, S. Zhang, Y. Liu, W. Huang, Frontiers of Chemistry in China, 3 (2008)
353.
42. F. A. Aly, S. A. Al-Tamimi, A. A. Alwarthan, Talanta, 53 (2001) 885.
70
43. Y. Rao, Y. Tong, X. R. Zhang, G. A. Lu, W. R. Baeyens, Anal. Chim. Acta, 416
(2000) 227.
44. Y. D. Liang, J. F. Song, X. F. Yang, Anal. Chim. Acta, 510 (2004) 21.
45. L. Yi, H. Zhao, S. Chen, L. Jin, D. Zheng, Z. Wu, Talanta, 61 (2003) 403.
46. B. Li, Z. Zhang, L. Zhao, Ch. Xu, Talanta, 57 (2002) 765.
47. A. A. Elbashir, B. Saad, A. S. M. Ali, M. I. Saleh, H. Y. Aboul-Enein, J. Liq.
Chromatogr. Rel. Technol., 31 (2008) 348.
48. K. L. See, A. A. Elbashir, B. Saad, A. S. M. Ali, H. Y. Aboul-Enein. Biomed.
Chromatogr., 23 (2009) 1283.
49. http://en.wikipedia.org/wiki/Ammonium_cerium(IV)_sulfate.
50. J. Bassett, R. C. Denney, G. H. Jeffery, J. Mendham, “Vogel’s textbook of
quantitative inorganic analysis”, English Language Book Society and Longman,
Essex, 4th
Edn., 1978, pp. 363.
51. K. Basavaiah, P. Nagegowda, B. C. Somashekarand V. Ramakrishna, Science Asia,
32 (2006) 403.
52. K. Basavaiah, U. Chandrashekar and P. Nagegowda, Bulg. Chem. Commun., 35
(2003) 174.
53. L. G. Chatten, R. A. Locock, R. D. Krause, J. Pharm. Sci., 60 (1971) 588.
54. G. G. Rao, G. S. Sastry, Anal. Chim. Acta., 56 (1971) 325.
55. K. Basavaiah, O. Z. Devi, Eclet. Quím., 33 (2008) 21.
56. N. Rajendraprasad, K. Basavaiah, K. Tharpa, K. B. Vinay, Eurasian J. Anal.
Chem., 4 (2009) 191.
57. N. Rajendraprasad and K. Basavaiah, J. Anal. Chem., 65 (2010) 482.
58. K. Basavaiah, V. Ramakrishna, U. R. Anil Kumar, Acta Pharm., 57 (2007) 211.
59. K. Basavaiah, P. Nagegowda, B. C. Somashekarand V. Ramakrishna, Science Asia,
32 (2006) 403.
60. K. Basavaiah, U. Chandrashekar, H. C. Prameela, Turkish J. Chem., 27 (2003) 591.
61. K. Basavaiah, U. Chandrashekar and P. Nagegowda, Bulg. Chem. Commun., 35
(2003) 174.
62. M. Kuchavski, T. Szumilo, Chem. Anal. (Warsaw), 28 (1983) 727.
63. C. S. P. Sastry, M. Aruna, A. R. M. Rao, Talanta, 35 (1988) 23.
71
64. C. S. P. Sastry T. N. V. Prasad, B. S. Sastry, E. V. Rao, Analyst (London), 113
(1988) 255.
65. M. S. Raghu, K. Basavaiah, Chem. Sci. J., CSJ-68 (2012).
66. P. J. Ramesh, K. Basavaiah, C. M. Xavier, K. N. Prashanth, M. S. Raghu, K. B.
Vinay, ISRN Anal. Chem, Volume 2012 (2012), Article ID 818405, 8 pages.
67. A. M. A. Sameer, K. Basavaiah, H. D. Revanasiddappa, K. B. Vinay, Proc. Indian
Natn. SCi Acad, 77 (2011) 233.
68. J. Basset, R. C. Denney, G. H. Jeffery, J. Mendham, “Vogel’s Quantitative
Inorganic Analysis”, English Language Book Society and Longman, Essex, 4th
Edn,
1978, p.360.
69. International Conference On Harmonisation of Technical Requirements for
Registration of Pharmaceuticals for Human Use, ICH Harmonised Tripartite
Guideline, Validation of Analytical Procedures: Text and Methodology Q2(R 1),
Complementary Guideline on Methodology dated 06 November 1996, incorporated
in November 2005, London.
70. A. K. Batalian, S.M. Polyakova, Zhur. Khim., 14 (1963) 178.
71. Z. Marczenko, Translator editor: Mary Masson, Separation and Spectrophotometric
Determination of elements, Ellis Horwood limited publishers. p. 222 (1982).
72. http://en.wikipedia.org/wiki/Ion-association.
73. IUPAC Compendium of Chemical Terminology, 2nd
Edn., 1997, available at:
http://www.iupac.org/goldbook/I03231.pdf
74. K. A. Connors. A textbook of pharmaceutical analysis, 3rd
Edn., Wiley India (P.)
Ltd., New Delhi, 2007.
75. Remington: the science and practice of pharmacy, 21st Edn., Vol. I, Lippincott
Williams & Wilkins, USA, 2006.
76. T. Higuchi, E. Roubal, Division of Analytical Chemistry, ACS, Abstracts of Papers,
149th
Meeting, 28 B (April 1965).
77. S. M. Al-Ghannam. J. Pharm. Biomed. Anal. 40, 2006, 151.
78. D. H. Manjunatha, S. M. T. Shaikh, K. Harikrishna, R. Sudhirkumar, P. B.
Kandagal, J. Seetharamappa. Ecl. Quim. 33, 2008, 37.
79. K. Kovacs-Hadady, I. Fabian. J. Pharm. Biomed. Anal. 16, 1998, 733.
72
80. N. Rajendraprasad, K. Basavaiah, K. B. Vinay. J. Preclin. Clin. Res. 4, 2010, 24.
81. K. B. Vinay, H. D. Revanasiddappa, N. Rajendraprasad, P. J. Ramesh, Cijo. M.
Xavier and K. Basavaiah, Drug Test. Anal., 4 (2012) 116.
82. M. S. Raghu, K. Basavaiah, Proc. Natl. Acad. Sci., India, Sect. A Phys. Sci. 82
(2012) 187.
83. A. M. A. Sameer, K. Basavaiah, Chem. Ind. Chem.l Eng. Quart., 18 (2012)339.
84. O. Z. Devi, K. Basavaiah, K. B. Vinay, J. Chil. Chem. Soc., 56 (2011) 715.
85. K. N. Prashanth, K. Basavaiah, Arab. J. Chem., article in press,
http://dx.doi.org/10.1016/j.arabjc.2011.10.006,
86. K. Harikrishna, B. S. Nagaralli, J. Seetharamappa, J. Food Drug Anal., 16 (2008)
11.
87. A. S. Amin, A. A. El-Fetouh, R. Gouda, F. Z. El-Sheikh, Spectrochim. Acta A, 67
(2007) 1306.
88. http://en.wikipedia.org/wiki/Chromatography.
89. http://www.forumsci.co.il/HPLC/WEBPharm_Review/HPLC_pharma_Stages.html.