CHAPTER 5a:PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY
Department of ChemistryFaculty of Science
Universiti Teknologi Malaysia
ANALYTICAL CHEMISTRY
SSC 1293
![Page 2: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/2.jpg)
The word “chromatography” originated from two Greek words, chroma which means “color” and graphy which means “writing”.
Chromatography was founded in 1906 by a Russian botanist, Micheal Tswett, when he separated plant extracts on a column packed with finely divided calcium carbonate.
Chromatography
![Page 3: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/3.jpg)
Chromatography is a method to separate two or more compounds in a mixture based on the differences in the property of each individual substance.
The properties include polarity, solubility, ionic strength, and size.
Chromatography is today one of the most useful analytical methods for separation, identification, and quantitation of chemical compounds.
Purpose of Chromatography
![Page 4: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/4.jpg)
In chromatography, the compounds are physically separated by distributing themselves between two phases: (a) a stationary phase(b) a mobile phase which flows continuously across the stationary phase.
Principles
Injector
Detector was located after the column or stationary phase
Mobile phase flow
Injected or applied sample
Packing material
![Page 5: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/5.jpg)
The stationary phase can be (a) a solid packed into a column(b) a solid coating the surface of a flat, plane material (c) a liquid supported on a solid (d) a liquid supported on the inside wall of an open tube
The mobile phase can be a gas, a liquid or a supercritical fluid.
Principles
![Page 6: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/6.jpg)
The mixture to be analysed is introduced onto the stationary phase and mobile phase carries the components through it.
Each individual analyte interacts with the two phases in a different manner.
Principles (cont.)
Injector
Detector
Mobile phase flow
t1 t2 t3Time
Interaction with stationary phase
Most
Least
![Page 7: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/7.jpg)
Because analytes differ in their affinity for the stationary phase vs the mobile phase, each analyte exhibits different migration and elution patterns and thus a mixture of analytes can be separated and quantified.
The tracing of the output signal vs time or mobile phase volume is called a chromatogram.
Principles (cont.)
Time
Signal
![Page 8: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/8.jpg)
Classification of elution chromatographic techniques
![Page 9: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/9.jpg)
Adsorption Chromatography
Eluent flow
Legend: Active site
Analyte
![Page 10: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/10.jpg)
Partition Chromatography
Solute dissolved in liquid phase coated on surface of solid support
![Page 11: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/11.jpg)
Affinity Chromatography
+
+ +
+
Matrix Ligand Immobilized ligand
Ligand Sample Complex Impurities
Complex Purified sample
![Page 12: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/12.jpg)
Size Exclusion Chromatography
3 2 1Signal
Volume
1 2 3
Pore size
Stationary phase particle
![Page 13: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/13.jpg)
Chromatographic Techniques
Chromatography
Gas LiquidSupercritical Fluid
SolidLiquid Liquid Liquid
IEC
Solid Solid
GLC GSC SFC BPCLSCLLC TLC SEC
GLC Gas-Liquid Chromatography GSC Gas-Solid ChromatographySFC Supercritical Fluid Chromatography LLC Liquid-Liquid ChromatographyTLC Thin Layer chromatography LSC Liquid-solid ChromatographyBPC Bonded Phase Chromatography IEC Ion Exchange ChromatographySEC Size Exclusion Chromatography
Legend
Mobile phase
Stationary phase
![Page 14: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/14.jpg)
General Theory of Chromatography
![Page 15: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/15.jpg)
Properties of a Gaussian peak. is standard deviation, wi is width at the inflection
point, wh is the width at half height, and wb is the width at the baseline intercept.
The plate theory
Tangents to points of inflection
Inflection point
Fraction of peak height
wh = 2.354
wi = 2
wb = 4
1.213
1.000
0.882
0.607
0.500
0.134
0
![Page 16: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/16.jpg)
Schematic diagram showing important parameters in chromatography. to is retention time of unretained compound, tR1 and tR2 are the retention times for component 1 and component 2, respectively.
Chromatogram
w w
a b10% ketinggian
puncak
to
t
t R2
R1
Suntikan
w1/2
1 2
Injection10% of peak height
![Page 17: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/17.jpg)
Chromatographic definitions
Retention time, tr - time elapsed between injection of sample and emergence of peak maximum
Mobile phase time, tm - time required for an injected solvent molecule or any other unretained compound to traverse the column
- also known as to (column dead volume or column void volume)
Adjusted retention time, t’r - difference between analyte retention time and column void volume, given by t’r = tr - to
![Page 18: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/18.jpg)
Chromatographic definitions
Capacity factor, k’ - A measure of how long a sample is retained with respect to the dead volume. Given by
k’ = (tr- to)/to
Column efficiency, N - A measure of the broadening of the sample peak as it passes through the column. Given by
N = 16 (tr’ /wb)2 or N = 5.54 (tr’ /wh)2
where wb is the width of peak at base peak
wh is the width of peak at half height
![Page 19: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/19.jpg)
Chromatographic definitions
Height equivalent to a theoretical plate (HETP), H - used to compare column efficiency of different lengths, L.
H = L /N
Resolution, R s = degree of separation of two components leaving the column.
Rt t
w ws
R1 R2( )
2
1 2
Continued …
![Page 20: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/20.jpg)
Resolution
d
w1 w2
Data:d = 1.8 cmw1 = 0.8 cmw2 = 0.9 cm
Rs = 2 --------------- = ------------- = 2.1 d 3.6 cm (w1 + w2 ) 1.7 cm
![Page 21: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/21.jpg)
Separation factor or selectivity factor
= ( t r (B) - t o )
( t r (A) - t o )=
t’r (B)
t’r (A)
=k’(B)
k’(A)
or
![Page 22: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/22.jpg)
Separation factor or selectivity factor
t’r (B) 2.8 min = ________ = _________ = 1.4
t’r (A) 2.0 min
1 2 30 4 time (min)
InjectionInjection
Air peakAir peak
t’r (B) t’r (A)
![Page 23: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/23.jpg)
Band broadening
It is due to finite rate at which several mass transfer process occur during migration of solute down a column.
Non-equilibrium theory - movement of solute through the column treated as a random walk.
Random walk - progress of a molecule through column is a succession of random stops and starts about a mean equilibrium concentration or band centre.
![Page 24: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/24.jpg)
Band broadening
When solute desorbed from stationary phase and transferred to mobile phase, it moves more rapidly than the band centre.
Three factors contributing:-Eddy diffusion (A)Longitudinal molecular diffusion (B)Rate of mass transfer (C)
![Page 25: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/25.jpg)
Typical pathways for two solute molecules during elution. The distance travelled by molecule 2 is greater than that travelled by molecule 1. Thus, molecule 2 arrieves at B later than molecule 1.
Eddy diffusion
(1)
(2)
A B
![Page 26: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/26.jpg)
Eddy diffusion
Effect of different path lengths due to irregular flow of molecules through packed particles in a column
Each molecule see different paths causing the solute molecules to arrive at the column outlet at different times
Independent of mobile phase velocity
![Page 27: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/27.jpg)
Band broadening due to longitudinal diffusion in mobile phase (A) initial band and (B) Diffusion of band with time.
Longitudinal diffusionA B
Profil kepekatan jalurA B
Arah laluan fasa bergerak
Lebar jalur
Profile of band conc.
Band width
Direction of mobile phase
![Page 28: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/28.jpg)
Longitudinal diffusion
Results from tendency of solutes to diffuse from concentration centre of a band to the more dilute regions on either sides.
Occur primarily in the mobile phase
Longitudinal diffusion inversely proportional to mobile phase velocity
Higher velocity provides less time for diffusion to occur: Means no band broadening
![Page 29: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/29.jpg)
Illustration of the influence of local nonequilibrium on band broadening.
Non-equilibrium
Mobile phase
Stationary phase
Interface
Flow
Equilibrium concentrationActual concentration
Legend:
![Page 30: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/30.jpg)
Concentration profile of analyte at the interface between the stationary phase and the mobile phase: (a) ideal condition before longitudinal movement of mobile phase takes place. (b) Actual condition after longitudinal movement of the mobile phase.
Influence of local nonequilibrium on band broadening
(a)
(b)
Analyte concentration
Mobile phase
Stationary phase
Mobile phase
Stationary phase
New equilibrium
Analyte concentration
![Page 31: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/31.jpg)
Resistance to mass transfer
Finite time required for the solutes to transfer in and out of the stationary phase
Kinetic lag in attaining equilibrium between two phases
Broadening occur when either of these rates is slow
Broadening depend on diffusion rate of analyte (time dependent)
Broadening worsen with increasing mobile phase flow rate
![Page 32: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/32.jpg)
Plot of HETP against flow velocity, showing the contributions of A (eddy diffusion), B (longitudinal diffusion) and C (interface mass transfer).
Van Deemter plot
Flow velocity
C
B
(Hmin)
C
HETP
Van Deemter curve
H AB
uCu
![Page 33: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/33.jpg)
Van Deemter’s Equation
H AB
uCu
Where;(A) Eddy diffusion(B) Longitudinal molecular (C) Diffusion rate of mass transfer
![Page 34: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/34.jpg)
Effect of N and on resolution
to
to
to
to
Poor resolution
Good resolution due
to column efficiencyGood resolution
due to column selectivityPoor resolution despite adequate column efficiency and selectivity (low k’)
![Page 35: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/35.jpg)
Resolution, selectivity, and effeciency
(a)
(b)
(c)
Initial profile
Increased selectivity, N unchanged
Selectivity unchanged, N increased
Good resolution
Good resolution
Poor resolution
![Page 36: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/36.jpg)
Qualitative Analysis
![Page 37: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/37.jpg)
Retention time comparison
By using identical column and operating conditions, the identity of the analyte can be determined by comparing the retention time with the standard compound.
The the absence of a peak at the same retention time as that of a standard under identical conditions suggests that either the respective compound is absent in the sample or the compound is present at a concentration level below the detection limit of the method.
![Page 38: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/38.jpg)
Kovats’ retention index (RI)
I zt t
t t
100 100
log log
log log
R(x) R(z)
R(z+1) R(z)
' '
' '
I zt t
t t
100 100
R(x) R(z)
R(z+1) R(z)
' '
' '
For linear temperature-programmed GC
For isothermal GC Plots of log tR’ vs C-no. gives a straight line
Plots of tR’ vs C-no. gives a straight line
![Page 39: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/39.jpg)
Kovats RI
The Kovats retention index. The log tR' values of a series of n-
alkanes area plotted against (100 x carbon number). Compound x (log tR' = 0.6) has a retention index value of 680.
200 400 600 800 1000
0.2
0.4
0.6
0.8
1.0
0
log t'R
Indeks penahananRetention Index
![Page 40: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/40.jpg)
Quantitative Analysis
![Page 41: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/41.jpg)
Quantitative Analysis: Peak area
Measurement of peak area: (a) By triangulation. Peak area area of the triangle = (h' x w)/2. (b) Peak area area of the rectangle = h x w1/2.
Gerak
Garis tangen
w
h'
Masa atau isipadu
w
h
Masa atau isipadu
1/2
(a) (b)
balas
![Page 42: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/42.jpg)
Response factor
XAmount
X AreaPeak x F
Response factor of compound X = FX
De
tect
or
resp
on
se
X
0 5 10 min
S
![Page 43: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/43.jpg)
Response factor
Response factor of compound X relative to that of compound S:
x
s
Xof X/Amountof areaPeak S of S/Amount of areaPeak
F
FRF
100 x peaks all of area Total
X of areaPeak %mixturein X%
A
If the response factors of all compounds in the sample are the same,
![Page 44: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/44.jpg)
External standard
Mx Molarity of compound X,
Ms Molarity of standard S,
Fx Response factor of compound X,
Fs Response factor of standard S,
V(x) Volume of injected sample X,
V(s) Volume of injected standard S,
Ax Peak area of sample X,
As Peak area of standard S.
MF A V
F AVMx
s x (s)
x s (x)s
![Page 45: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/45.jpg)
External standard
External standard poses a problem with reproducibility.
•Injection – the volume of injection may vary from one injection to another injection.
•Detection – the sensitivity of the detector may vary from time to time.
![Page 46: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/46.jpg)
Internal standard
MF A N
F AVxs x s
x s x
Mx Molarity of compound X,
Ms Molarity of standard S,
Fx Response factor of compound X,
Fs Response factor of standard S,
V(x) Volume of injected sample X,
Ax Peak area of sample X,
As Peak area of standard S.
N(s) Amount of internal standard S,
![Page 47: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/47.jpg)
Criteria for internal standards
Characteristics of ideal internal standards
(a) It is well separated from the components in the analyzed mixture.
(b) Its response factor is similar to that of the analyte.
(c) Its retention time is close to the that of the analyte to minimize errors in peak area measurement.
(d) Its concentration should be similar to that of the analyte.
![Page 48: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/48.jpg)
Example 1. External standard
Consider the following data of a typical quantitative gas chromatographic analysis where a compound, X, is used as the external standard.
• An injection (1 L) of a mixture containing 10, 12, and 13 ppm of X, Y, and Z, respectively, gave respective peak areas of 515, 748 and 939 arbitrary units.
• An injection (2 L) of the sample containing compounds X, Y, and Z gave peak areas of 232, 657, and 984 arbitrary units, respectively.
• Calculate the concentrations of compounds X, Y and Z.
![Page 49: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/49.jpg)
Example 1. Solution
![Page 50: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/50.jpg)
Example 2. Internal standard
In a chromatographic analysis, a mixture containing 0.0567 M of a compound, X, and 0.0402 mM of a standard, S, gave peak areas of Ax = 498 and AS = 526 arbitrary units.
An aliquot (10.0 mL) of the unknown was added with 10.00 mL of 0.102 M S and the mixture was diluted to 25 mL in a volumetric flask. The mixture gave a chromatogram with Ax = 678 and AS =
588 arbitrary units.
• Calculate the concentration of X in the unknown.
![Page 51: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/51.jpg)
Example 2. (Cont.)
De
tec
tor
resp
on
se
X
0 5 10 min
S Determination of the relative detector response for the standard Add a known amount of the standard compound into a known quantity (volume or weight) of sample, Vx, Measure the peak area, and calculate the concentration of the analyte using the proper equation.
![Page 52: CHAPTER 5a: PRINCIPLES AND BASIC THEORY OF CHROMATOGRAPHY Department of Chemistry Faculty of Science Universiti Teknologi Malaysia ANALYTICAL CHEMISTRY](https://reader036.vdocument.in/reader036/viewer/2022062515/56649f585503460f94c7e40b/html5/thumbnails/52.jpg)
Example 2. Solution