Philippe Gros – [email protected], Mickaël Frotin – [email protected] Circulation and purification of gas in the sealed HARPO TPC Characterization of the gas with cosmic ray data Selection of high momentum cosmic tracks straight tracks uniform <dE/dx> Distribution of charge vs drift time corrected for angular effect Gain from height Drift velocity from length Attenuation from slope relative value combined with threshold effects Charge vs drift time distribution of cosmic rays for a 1h run Evolution of the distribution for 3 runs, at 1.5 month interval The increasing exponential attenuation is clearly visible The HARPO TPC 30 cm cubic TPC Electron amplification by Micromegas+2GEM Gas: Ar/iC 4 H 10 95/5% up to 5bar 2x288 strips readout (x&y), 1mm pitch Readout electronics based on AFTER chips (511 time bins up to 100 MHz) Trigger: 6 scintillators + micromegas signal and timing (+ laser pulse when available) Micromegas GEM Field cage Scintillators Readout strips + Measurement Results Unchanged gas in sealed vessel for 5 month: February to July 2015 CAPS started after 5 month (July 2015) Gas evolution from cosmic ray data The attenuation increases due to electron capture - probably O 2 from leaks The drift velocity decreases - probably H 2 O from outgassing The original properties are recovered after a couple of weeks of circulation 0.5L gas samples => ~30mbar drops start CAPS Gas properties evaluated from cosmic ray data. Extraction of gas samples for analysis caused pressure drops. The voltages were adjusted to keep a similar gain. Pressure [mbar] Ar:iC4H10 (95:5) 2100mbar Gas samples were analysed in a high resolution (R=2800) direct injection mass spectrometer (THERMO MAT 271) at LRMO Before starting of CAPS: 620ppm N 2 , 180ppm O 2 , 190ppm CO 120ppm CO 2 , 5.10% isobutane After one month running 890ppm N 2 , <20ppm O 2 , 250ppm CO 160ppm CO 2 , 4.42% isobutane The H 2 O content cannot be measured in samples (needs online device) Gas contamination analysis and discussion Example of high resolution mass spectrum. 12 C 16 O, 14 N and 12 C 2 1 H 2 can be separated After 1 month of purification, no oxygen is left in the gas. Full spectrum from the mass spectrometer before and after purification O2 disappeared completely The HARPO TPC provided good cosmic data for several month in fully sealed mode Equipped with a simple lightweight purification system, the full performance was recovered in only a few weeks => Could run continuously without degradation A loss of isobutane during purification was observed and needs further investigation Further long time study (>1year) should be done in continuous running mode (CAPS and HV on) Conclusions and discussion 4th Conference on Micro-Pattern Gaseous Detectors, 2015-10-12 D. Bernard, M. Frotin, Y. Geerebaert, Ph. Gros , I. Semeniouk LLR, Ecole Polytechnique, CNRS/IN2P3, 91128 Palaiseau, France D. Attié, A. Delbart, P. Magnier IRFU, CEA Saclay, 91191 Gif-sur-Yvette, France V. Dauvois. D. Durand, S. Legand DEN/DANS/DPC/SECR/LRMO, CEA Saclay, 91191 Gif-sur-Yvette, France Circulation And Purification System (CAPS) Sealed TPC - 100 litres vessel Closed circulation circuit - Small turbine for gas circulation - Oxisorb® filter for O 2 and H 2 O removal - Flowmeter All parts of the circuit can be isolated - Testing for leaks - Sample extraction Thoroughly tested for leaks - in vacuum - under pressure Light and sealed turbine for circulation with flows of about one litre per hour Patent FR 15 50987 (2015/02/09) Immediately after opening CAPS After one week of circulation Oxisorb® filter for O 2 and H 2 O Gas Flow

Upload: vuongdung

Post on 10-Apr-2018

215 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Circulation and purification of gas in the sealed HARPO TPCllr.in2p3.fr/~dbernard/polar/2015-10-10_MPGD2015_poster_CAPS.pdf · Philippe Gros – philippe.gros@llr.in2p3.fr, Mickaël

Philippe Gros – [email protected], Mickaël Frotin – [email protected]

Circulation and purification of gas in the sealed HARPO TPC

Characterization of the gaswith cosmic ray data

Selection of high momentum cosmic tracks

● straight tracks● uniform <dE/dx>

Distribution of charge vs drift time● corrected for angular effect● Gain from height● Drift velocity from length● Attenuation from slope

● relative value● combined with threshold effects

Charge vs drift time distribution of cosmic rays for a 1h run

Evolution of the distribution for 3 runs, at 1.5 month intervalThe increasing exponential attenuation is clearly visible

The HARPO TPC

• 30 cm cubic TPC

• Electron amplification by Micromegas+2GEM

• Gas: Ar/iC4H10 95/5% up to 5bar

• 2x288 strips readout (x&y), 1mm pitch

• Readout electronics based on AFTER chips(511 time bins up to 100 MHz)

• Trigger: 6 scintillators + micromegas signal and timing (+ laser pulse when available) MicromegasGEM

Field cage

ScintillatorsReadout strips

+

Measurement Results

● Unchanged gas in sealed vessel for 5 month: February to July 2015

● CAPS started after 5 month (July 2015)

Gas evolution from cosmic ray data

● The attenuation increases due to electron capture - probably O2 from leaks

● The drift velocity decreases

- probably H2O from outgassing

● The original properties are recovered after a couple of weeks of circulation

0.5L gas samples => ~30mbar dropsstart CAPS

Gas properties evaluated from cosmic ray data. Extraction of gas samples for analysis caused pressure drops. The voltages were adjusted to keep a similar gain.

Pressure

[mbar]

Ar:iC4H10 (95:5) 2100mbar Gas samples were analysed in a high resolution (R=2800) direct injection mass spectrometer (THERMO MAT 271) at LRMO

● Before starting of CAPS:

620ppm N2, 180ppm O2, 190ppm CO 120ppm CO2, 5.10% isobutane

● After one month running

890ppm N2, <20ppm O2, 250ppm CO 160ppm CO2, 4.42% isobutane

● The H2O content cannot be measured in samples (needs online device)

Gas contamination analysis and discussion

Example of high resolution mass spectrum. 12C16O, 14N and 12C

21H

2 can be separated

After 1 month of purification, no oxygen is left in the gas.

Full spectrum from the mass spectrometer before and after purification

O2 disappeared completely

● The HARPO TPC provided good cosmic data for several month in fully sealed mode

● Equipped with a simple lightweight purification system, the full performance was recovered in only a few weeks

=> Could run continuously without degradation

● A loss of isobutane during purification was observed and needs further investigation

● Further long time study (>1year) should be done in continuous running mode (CAPS and HV on)

Conclusions and discussion

4th Conference on Micro-Pattern Gaseous Detectors, 2015-10-12

D. Bernard, M. Frotin, Y. Geerebaert, Ph. Gros, I. SemenioukLLR, Ecole Polytechnique, CNRS/IN2P3, 91128 Palaiseau, France

D. Attié, A. Delbart, P. MagnierIRFU, CEA Saclay, 91191 Gif-sur-Yvette, France

V. Dauvois. D. Durand, S. LegandDEN/DANS/DPC/SECR/LRMO, CEA Saclay, 91191 Gif-sur-Yvette, France

Circulation And Purification System(CAPS)

• Sealed TPC

- 100 litres vessel

• Closed circulation circuit

- Small turbine for gas circulation

- Oxisorb® filter for O2 and H2O removal

- Flowmeter

• All parts of the circuit can be isolated

- Testing for leaks

- Sample extraction

• Thoroughly tested for leaks

- in vacuum

- under pressure

Light and sealed turbine for circulation with flows of about one litre per hour

Patent FR 15 50987 (2015/02/09)

Immediately after opening CAPS

After one week of circulation

Oxisorb® filter for O2 and H2O

Gas Flow