cleanfit cpa871 operating instructions - endress+hauser...•regulations for explosion protection...
TRANSCRIPT
Products Solutions Services
Operating InstructionsCleanfit CPA871Flexible retractable process assembly for water,wastewater, chemical industry and heavy industry
BA01323C/07/EN/03.1671324465
Table of contents
Endress+Hauser 3
Table of contents
1 Document information . . . . . . . . . . . . . . 41.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41.2 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . 41.3 Symbols on the device . . . . . . . . . . . . . . . . . . . 4
2 Basic safety instructions . . . . . . . . . . . . 52.1 Requirements for the personnel . . . . . . . . . . . . 52.2 Designated use . . . . . . . . . . . . . . . . . . . . . . . . 52.3 Occupational safety . . . . . . . . . . . . . . . . . . . . . 52.4 Operational safety . . . . . . . . . . . . . . . . . . . . . . 62.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . 6
3 Incoming acceptance and productidentification . . . . . . . . . . . . . . . . . . . . . . . 7
3.1 Incoming acceptance . . . . . . . . . . . . . . . . . . . . 73.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . 73.3 Product identification . . . . . . . . . . . . . . . . . . . . 83.4 Certificates and approvals . . . . . . . . . . . . . . . . 9
4 Installation . . . . . . . . . . . . . . . . . . . . . . . 104.1 Installation conditions . . . . . . . . . . . . . . . . . . 104.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 174.3 Installing the sensor . . . . . . . . . . . . . . . . . . . 224.4 Post-installation check . . . . . . . . . . . . . . . . . . 26
5 Operation options . . . . . . . . . . . . . . . . . 275.1 Commissioning . . . . . . . . . . . . . . . . . . . . . . . 275.2 Operating elements . . . . . . . . . . . . . . . . . . . . 285.3 Manual operation . . . . . . . . . . . . . . . . . . . . . 295.4 Pneumatic operation . . . . . . . . . . . . . . . . . . . 29
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . 316.1 Maintenance intervals . . . . . . . . . . . . . . . . . . 316.2 Cleaning the assembly . . . . . . . . . . . . . . . . . . 326.3 Cleaning the sensor . . . . . . . . . . . . . . . . . . . . 326.4 Cleaning agent . . . . . . . . . . . . . . . . . . . . . . . 336.5 Replacing seals . . . . . . . . . . . . . . . . . . . . . . . 34
7 Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . 447.1 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 447.2 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 447.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
8 Accessories . . . . . . . . . . . . . . . . . . . . . . . 458.1 Installation material for rinse connections . . . 468.2 Cleaning systems . . . . . . . . . . . . . . . . . . . . . . 468.3 Flow vessels . . . . . . . . . . . . . . . . . . . . . . . . . 478.4 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
9 Technical data . . . . . . . . . . . . . . . . . . . . 51
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Document information Cleanfit CPA871
4 Endress+Hauser
1 Document information
1.1 Warnings
Structure of information Meaning
LDANGERCauses (/consequences)Consequences of non-compliance(if applicable)‣ Corrective action
This symbol alerts you to a dangerous situation.Failure to avoid the dangerous situation will result in a fatal or seriousinjury.
LWARNINGCauses (/consequences)Consequences of non-compliance(if applicable)‣ Corrective action
This symbol alerts you to a dangerous situation.Failure to avoid the dangerous situation can result in a fatal or seriousinjury.
LCAUTIONCauses (/consequences)Consequences of non-compliance(if applicable)‣ Corrective action
This symbol alerts you to a dangerous situation.Failure to avoid this situation can result in minor or more serious injuries.
NOTICECause/situationConsequences of non-compliance(if applicable)‣ Action/note
This symbol alerts you to situations which may result in damage toproperty.
1.2 Symbols used
Symbol Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
1.3 Symbols on the device
Symbol Meaning
Reference to device documentation
Cleanfit CPA871 Basic safety instructions
Endress+Hauser 5
2 Basic safety instructions
2.1 Requirements for the personnel• Installation, commissioning, operation and maintenance of the measuring system may
be carried out only by specially trained technical personnel.• The technical personnel must be authorized by the plant operator to carry out the
specified activities.• The electrical connection may be performed only by an electrical technician.• The technical personnel must have read and understood these Operating Instructions
and must follow the instructions contained therein.• Measuring point faults may be repaired only by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided may only be carried outdirectly by the manufacturer or by the service organization.
2.2 Designated useThe CleanfitCPA871 retractable assembly, which can be manually or pneumaticallyoperated, is designed for the installation of sensors in vessels and pipes.
Thanks to its design, it can be used in pressurized systems (see "Technical data").
Use of the device for any purpose other than that described, poses a threat to the safety ofpeople and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3 Occupational safetyAs the user, you are responsible for complying with the following safety conditions:• Installation guidelines• Local standards and regulations• Regulations for explosion protection
Basic safety instructions Cleanfit CPA871
6 Endress+Hauser
2.4 Operational safety1. Before commissioning the entire measuring point, verify that all connections are
correct. Ensure that electrical cables and hose connections are undamaged.
2. Do not operate damaged products, and safeguard them to ensure that they are notoperated inadvertently. Label the damaged product as defective.
3. If faults cannot be rectified:Take the products out of operation and safeguard them to ensure that they are notoperated inadvertently.
2.5 Product safety
2.5.1 State of the artThe product is designed to meet state-of-the-art safety requirements, has been tested, andleft the factory in a condition in which it is safe to operate. The relevant regulations andEuropean standards have been observed.
Cleanfit CPA871 Incoming acceptance and product identification
Endress+Hauser 7
3 Incoming acceptance and productidentification
3.1 Incoming acceptance1. Verify that the packaging is undamaged.
Notify your supplier of any damage to the packaging.Keep the damaged packaging until the matter has been settled.
2. Verify that the contents are undamaged. Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3. Check the delivery for completeness. Check it against the delivery papers and your order.
4. Pack the product for storage and transportation in such a way that it is protectedagainst impact and moisture. The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
3.2 Scope of deliveryThe scope of delivery comprises:• Assembly in the version ordered• Operating Instructions
Incoming acceptance and product identification Cleanfit CPA871
8 Endress+Hauser
3.3 Product identification
3.3.1 NameplateThe nameplate provides you with the following information on your device:• Manufacturer identification• Order code• Extended order code• Serial number• Ambient and process conditions• Safety information and warnings
Compare the data on the nameplate with your order.
3.3.2 Product identificationThe order code and serial number of your product can be found in the following locations:• On the nameplate• In the delivery papers
Obtaining information on the product1. Go to the product page for your product on the Internet.
2. At the bottom of the page, select the "Online Tools" link followed by "Check yourdevice features". An additional window opens.
3. Enter the order code from the nameplate into the search field, and then select "Showdetails". You will receive information on each feature (selected option) of the order code.
Cleanfit CPA871 Incoming acceptance and product identification
Endress+Hauser 9
3.4 Certificates and approvalsRL 94/9/EC (ATEX)The assembly does not fall within the scope of the directive. However, if conditions for safeuse are adhered to, it may be deployed in the hazardous area.
CE/PEDThe CPA871 assembly has been manufactured according to good engineering practice inaccordance with Article 3, Paragraph 3 of the Pressure Equipment Directive 97/23/EC andtherefore is not required to bear the CE label.
Installation Cleanfit CPA871
10 Endress+Hauser
4 Installation
4.1 Installation conditions
4.1.1 OrientationThe assembly is designed for installation on tanks and pipes. Suitable process connectionsmust be available for this.
NOTICEFrost damage to the assembly‣ If used outdoors, ensure that water cannot penetrate the drive.
The assembly is designed in such a way that there are no restrictions with regard to theorientation.
The sensor that is used can restrict the orientation.
Cleanfit CPA871 Installation
Endress+Hauser 11
4.1.2 Dimensions
Short version
XP
XM
/XS
= 3
98/4
34 (
15.6
7/1
7.0
9)
158 (
6.2
2)
61
(2.40)
117
(4.61)
XA
63
(2.4
8)
A0023894
1 Pneumatic drive, short version, dimensions inmm (inch)
61(2.40)
XA
61(2.40)
107(4.21)
42 (1.65)
63
(2.4
8)
XP
155 (
6.1
0)X
M/X
S =
398/4
34 (
15.6
7/1
7.0
9)
A0023897
2 Manual drive, short version, dimensions inmm (inch)
XMXSXPXA
Assembly in measuring positionAssembly in service positionHeight of particular process connection (see table below)Necessary mounting distance for sensor replacement
The mounting distance XA is 280 mm (11.02") for 120 mm sensorsThe mounting distance XA is 408 mm (15.94") for 225 mm sensors
Installation Cleanfit CPA871
12 Endress+Hauser
Long version
61(2.40)
XP
200 (
7.8
7)
107
(4.21)
63
XM
/XS
= 4
40/5
18 (
17.3
2/2
0.3
9)
XA
(2.4
8)
A0023895
3 Pneumatic drive, long version, dimensions inmm (inch)
61
(2.40)
61
(2.40)
107
(4.21)
42 (1.65)
63
(2.4
8)
XP
197 (
7.7
6)
XM
/XS
= 4
40/5
18 (
17.3
2/2
0.3
9)
XA
A0023898
4 Manual drive, long version, dimensions inmm (inch)
XMXSXPXA
Assembly in measuring positionAssembly in service positionHeight of particular process connection (see table below)Necessary mounting distance for sensor replacement
The mounting distance XA is 360 mm (14.17") for 225 mm sensors
Cleanfit CPA871 Installation
Endress+Hauser 13
Immersion chamber version
61(2.40)
117
(4.61)
158 (
6.2
2)
63
XP
XM
/XS
= 3
98/4
34 (
15.6
7/1
7.0
9)
XA
(2.4
8)
A0023896
5 Immersion chamber version with pneumatic drive,dimensions in mm (inch)
XP
XM
/XS
= 3
98/4
34 (
15.6
7/1
7.0
9)
155 (
6.1
0)
61(2.40)
107(4.21)
42 (1.65)
63
(2.4
8)
61(2.40)X
A
A0023899
6 Immersion chamber version with manualdrive, dimensions in mm (inch)
XMXSXPXA
Assembly in measuring positionAssembly in service positionHeight of particular process connection (see table below)Necessary mounting distance for sensor replacement
The mounting distance XA is 280 mm (11.02") for 225 mm sensorsThe mounting distance XA is 570 mm (22.44") for 360 mm sensors
Installation Cleanfit CPA871
14 Endress+Hauser
Process connection height
Process connection Height XP in mm (inch)
CB Clamp 2"ISO2852, ASME BPE-2012
A0024100
16 (0.63)
CC Clamp 2½"ISO2852, ASME BPE-2012
A0024101
16 (0.63)
FA Flange DN 40, EN1092-1
A0024102
18 (0.71)
FB Flange DN 50, EN1092-1
A0024103
18 (0.71)
FC Flange DN 80, EN1092-1
A0024104
20 (0.79)
FD Flange 2" 150 lbs, ASME B16.5
A0024105
19.1 (0.75)
FE flange 3" 150 lbs, ASME B16.5
A0024106
23.8 (0.94)
FF 10K50, JIS B2220
A0024107
16 (0.63)
FG 10K80, JIS B2220
A0024108
18 (0.71)
MA Dairy fitting DN 50 DIN 11851
A0024109
15.5 (0.61)
MB Dairy fitting DN 65 DIN 11851
A0024110
15.5 (0.61)
HB Thread NPT 1½"
A0024111
40.5 (1.57)
Cleanfit CPA871 Installation
Endress+Hauser 15
4.1.3 Immersion depths
X1
X2
X1
X2
X1
X2
1 2 3
A0023893
7 Immersion depths
1 Short stroke (36 mm)2 Long stroke (78 mm)3 Immersion versionImmersion depths in mm (inch)
Versions
Process connection 1 2 3
CB Clamp ISO2852ASME BPE-2012 2"
X1X2
14.9 (0.59)34.2(1.35)
61.0 (2.40)75.7 (2.98)
119.9 (4.72)134.6 (5.30)
CC Clamp ISO2852ASME BPE-2012 2½"
X1X2
14.9 (0.59)34.2(1.35)
61.0 (2.40)75.7 (2.98)
119.9 (4.72)134.6 (5.30)
FA Flange DN 40EN1092-1
X1X2
14.9 (0.59)34.2(1.35)
61.0 (2.40)75.7 (2.98)
119.9 (4.72)134.6 (5.30)
FB Flange DN 50EN1092-1
X1X2
14.9 (0.59)34.2(1.35)
61.0 (2.40)75.7 (2.98)
119.9 (4.72)134.6 (5.30)
FC Flange DN 80EN1092-1
X1X2
12.9 (0.51)32.2(1.27)
59.0 (2.32)73.7 (2.90)
117.9 (4.64)132.6 (5.22)
FD Flange 2" 150 lbsASME B16.5
X1X2
13.8 (0.54)33.1 (1.30)
59.9 (2.36)74.6 (2.94)
118.8 (4.68)133.5 (5.26)
FE Flange 3" 150 lbsASME B16.5
X1X2
--
--
114.1 (4.49)128.8 (5.07)
FF Flange 10K50JIS B2220
X1X2
14.4 (0.57)33.7 (1.33)
61.3 (2.41)76.0 (2.99)
120.2 (4.73)134.9 (5.31)
FG Flange 10K80JIS B2220
X1X2
14.4 (0.57)33.7 (1.33)
60.5 (2.38)75.2 (2.96)
119.4 (4.70)134.1 (5.28)
HB Thread NPT 1½" X1X2
--
63.0 (2.48)77.7 (3.06)
121.9(4.80)136.6 (5.38)
MA Dairy fittingDN 50 DIN11851
X1X2
15.4 (0.61)34.7 (1.37)
61.5 (2.42)76.2 (3.00)
120.4 (4.74)135.1 (5.32)
MB Dairy fittingDN 65 DIN11851
X1X2
15.4 (0.61)34.7 (1.37)
61.5 (2.42)76.2 (3.00)
120.4 (4.74)135.1 (5.32)
NA Thread ISO228G1¼
X1X2
--
61.5 (2.42)76.2 (3.00)
--
Installation Cleanfit CPA871
16 Endress+Hauser
23
(0.9
1)
61.5
(2.4
2)
76.2
(3.0
0)
A0024447
8 Immersion depth in mm (inch) for process connection NA thread ISO228 G1¼
Cleanfit CPA871 Installation
Endress+Hauser 17
4.2 Installation
4.2.1 Measuring system
1 2
3
4
A0029620
9 Measuring system (example)
1 Cleanfit assembly CPA8712 Measuring cable3 Liquiline CM44x transmitter4 Sensor
Installation Cleanfit CPA871
18 Endress+Hauser
4.2.2 Installing the assembly in the processLWARNING
Risk of injury from high pressure, high temperature or chemical hazards if processmedium escapes.‣ Wear protective gloves, protective goggles and protective clothing.‣ Only mount the assembly if the containers or pipes are empty and unpressurized.
Prior to installation, check the flange seal between the flanges.
1. Move the assembly to the service position (the triangle position marking is visible(→ 10).
2. Secure the assembly to the tank or piping using the process connection.
3. Follow the instructions in the next section to connect pipes for compressed air andrinse water (for the relevant assembly version).
A0023307
10 Position markings (service position)
Cleanfit CPA871 Installation
Endress+Hauser 19
4.2.3 Pneumatic connection for automatic operationPrerequisites:• Air pressure 4 to 7 bar (58 to 102 psi)• Compressed air quality in accordance with ISO 8573-1:2001
Quality class 3.3.3 or 3.4.3• Solids class 3 (max. 5 μm, max. 5 mg/m3, contamination with particles)• Water content for temperatures ≥ 15 °C: class 4 pressure condensation point 3 °C or
lower• Water content for temperatures of 5 to 15 °C: class 3 pressure condensation point -20 °C
or lower• Oil content class 3 (max. 1 mg/m3)• Air temperature: 5 °C or higher• No continuous air consumption• Minimum nominal diameter of air pipes: 2 mm (0.08 ")
Connection: Push-in fitting M5, hose 4/2 mm OD/ID (adapter for 6/4 mm OD/IDenclosed)
Damage to seals due to excessive air pressure!
If the air pressure can increase to more than 7 bar (102 psi) (even short pressure surges),a pressure-reducing valve must be installed upstream.
1
2
3
4
5
6
7
8
A0029614
11 Assembly with pneumatic drive (withoutprotection cap)
1 Rinse connection2 Automatic limit position lock, process3 Connection for limit position switch4 Automatic limit position lock, service5 Fastening ring for protection cap6 Pneumatic connection (move to
measuring position)7 Pneumatic connection (move to service
position)8 Rinse connection
Use a pneumatic change-over valve (4/2-way valve) to move the assembly. Connectboth inputs. If you connect only one input (e.g. for test purposes), the piston will beblocked, as the sensor guide moves before the limit position lock is deactivated.
Installation Cleanfit CPA871
20 Endress+Hauser
4.2.4 Rinse connectionsThe service chamber connections of the CPA871 retractable assembly make it possible toclean the chamber and the sensor with water or a cleaning solution at a pressure of6 bar (87 psi) max..
Seals can be damaged if the water pressure is too high.
Install an pressure-reducing valve upstream if there is a possibility that the waterpressure will increase to more than 6 bar (87 psi) (including any short pressuresurges).
4.2.5 Assignment of rinse connectionsIn the standard version, the inlet and outlet of the service chamber can be assigned asdesired.
In the immersion chamber version, the inlet and outlet of the service chamber are fixed.The outlet of the service chamber is located under the leakage borehole. The leakageborehole is sealed with an M5 screw.
1
2
3
4
A0029621
12 Connection of service chamber in the immersion chamber version
1 Service chamber2 Service chamber inlet3 Leakage hole4 Service chamber outlet
4.2.6 Connecting the limit position switchesWith limit position detection, you can notify a system located downstream (transmitter,switching amplifier, output interface terminal) whether the assembly is in themeasurement or service position(in the case of a manual drive, only the measurementposition is queried).
The assembly can be ordered directly with limit position detection, or it can be retrofittedat a later stage.
Switching element function: NAMUR NC contact (inductive)Switching distance: 1.5 mm (0.06 ")Nominal voltage: 8 VSwitching frequency: 0 to 5000 HzHousing material: Stainless steel
Cleanfit CPA871 Installation
Endress+Hauser 21
BN
BN
BU
BU
L+
L+
L-
L-
A
B
C D E
A0017831
13 Inductive limit position switches
A Limit position switch, service positionB Limit position switch, measuring positionC Connector, M12, solder side (inside of assembly)D CodingE Connector, Pin side (outside of assembly)
BN
BK
BU
WH
1
2
3
4 +
+
-
-
A0022163
14 Connecting cable for limit position switch on transmitter, switching amplifier, output interface terminaletc.
1 "Measuring" position2 "Measuring" position3 "Service" position4 "Service" position
Only pins 1 and 2 are assigned for manually activated assemblies with one switch(measuring position).
Signal table for limit position switches
Position of assembly Limit position switch for "measuring"position
Limit position switch for "service"position
Measurement Active LOW (≥ 3 mA) Active LOW (≥ 3 mA)
Service Active HIGH (≤ 1 mA) Active HIGH (≤ 1 mA)
Installation Cleanfit CPA871
22 Endress+Hauser
4.3 Installing the sensor
4.3.1 Preparing the sensor and assembly
1
A0030154
15 Installing the sensor
1 Thrust collar with O-ring
1. Remove the protection cap from the sensor. Ensure that the O-ring and thrust collar(→ 15, item 1) are present.
2. Immerse the sensor shaft in water. This makes for easier installation.
3. Move the assembly to the service position.
4.3.2 Installing and removing sensors
1
2
A B
1
2
A0030155
16 Sensor installation options
1 Sensor adapter2 Retraction pipeA Sensor adapter is on top of the retraction pipeB Sensor adapter is below the retraction pipe (not visible)Depending on the assembly version, the sensor adapter is either visible (→ 16, A) orinstalled inside the retraction pipe where it is not visible (B). As a result, the proceduresfor installing and removing the sensors differ as follows:
Cleanfit CPA871 Installation
Endress+Hauser 23
Installing and removing sensors if the sensor adapter is visible (pos. A)
1
2
3
4
A0030156
17 Installing the sensor
1 Open-ended wrench (AF 17/19 mm)2 Protection cover3 Dummy plug4 Sensor
Gel and KCl sensors can be installed in this version.
Install the sensor as follows:
1. Remove the protection cap (→ 17, item 2) (this is possible only if the assembly isin the service position).
2. Remove the yellow dummy plug (pos. 3).
3. Using the open-ended wrench (item 1), screw in the sensor (item 4) in the place ofthe dummy plug and hand-tighten (3 Nm (2.2 lbf ft)).
4. Attach the open-ended wrench back into the protection cap.
5. Put the protection cap on the assembly. When doing so, guide the measuring cablethrough the cable run (top of protection cap).
Always install the protection cap before moving the assembly to the measuringposition. The protection cap cannot be removed in the measuring position. Thisprevents the sensor from being removed.
Installation Cleanfit CPA871
24 Endress+Hauser
Installing and removing sensors if the sensor adapter is not visible (pos. B)
3
45
2
1
A0030157
18 Installing the sensor
1 Socket wrench (AF 17/19 mm)2 Protection cover3 Dummy plug (protection cap)4 Sensor5 Retraction pipe
Gel sensors can be installed in this version. To install KCl sensors, you will need a "Gel- KCl adapter".
Install the sensor as follows:
1. Remove the protection cap (→ 18, item 2) (this is possible only if the assembly isin the service position).
2. Unscrew and remove the retraction pipe (pos. 5) (in an anticlockwise direction).
3. Remove the yellow dummy plug (pos. 3).
4. Using the open-ended wrench (item 1), screw in the sensor (item 4) in the place ofthe dummy plug and hand-tighten (3 Nm (2.2 lbf ft)).
5. Screw the retraction pipe back in.
6. Attach the open-ended wrench back into the protection cap.
7. Put the protection cap on the assembly. When doing so, guide the measuring cablethrough the cable run (top of protection cap).
Always install the protection cap before moving the assembly to the measuringposition. The protection cap cannot be removed in the measuring position. Thisprevents the sensor from being removed.
Cleanfit CPA871 Installation
Endress+Hauser 25
Installation of 360 mm gel and KCL sensors with "Gel - KCl adapter"
1
2
3
5
6
7
A0030158
19 Sensor installation, Part 1
1 Open-ended wrench (AF 17/19 mm)2 Protection cover3 Dummy plug (protection cap)5 Retraction pipe6 Gel - KCl adapter7 Lock nut
Gel sensors can be installed in this version. To install KCl sensors, you will need a "Gel- KCl adapter".
Install the sensor as follows:
1. Remove the protection cap (→ 19, item 2) (this is possible only if the assembly isin the service position).
2. Unscrew and remove the retraction pipe (pos. 5) (in an anticlockwise direction).
3. Turn the lock nut (pos. 7) on the "Gel - KCl adapter" (pos. 6) as far as it will go in anupward direction.
4. Remove the yellow dummy plug (pos. 3).
5. Screw in the "Gel - KCl adapter" (item 6) in the place of the dummy plug and hand-tighten (3 NM (2.2 lbf ft)).
6. Hand-tighten the lock nut in a clockwise direction, and then use an open-endedwrench (AF 24 mm) to tighten it by ¼ turn.
7. Screw the retraction pipe back in.
8. Use the open-ended wrench (item 1) to screw in the sensor (→ 20, item 4) andhand-tighten (3 Nm (2.2 lbf ft)).
9. Attach the open-ended wrench back into the protection cap.
Installation Cleanfit CPA871
26 Endress+Hauser
10. Put the protection cap on the assembly. When doing so, guide the measuring cablethrough the cable run (top of protection cap).
4
1
2
A0030159
20 Sensor installation, Part 2
1 Open-ended wrench2 Protection cover4 360 mm gel or KCl sensor
Always install the protection cap before moving the assembly to the measuringposition. The protection cap cannot be removed in the measuring position. Thisprevents the sensor from being removed.
4.4 Post-installation check• After installation, check the transmitter for damage.• Check whether the transmitter is protected against moisture and direct sunlight (e.g. by
the weather protection cover).
Cleanfit CPA871 Operation options
Endress+Hauser 27
5 Operation options
5.1 CommissioningPrior to initial commissioning, ensure that:• all seals are correctly seated (on the assembly and on the process connection)• the sensor is correctly installed and connected• the water connection at the rinse connections is correct (if present).
LWARNINGRisk of injury from high pressure, high temperature or chemical hazards if processmedium escapes.‣ Before subjecting the assembly to process pressure, verify that all connections are
sealed!
Note that when moving the assembly, an open connection exists for a short periodbetween the process and service chamber.
Operation options Cleanfit CPA871
28 Endress+Hauser
5.2 Operating elements
A0023307
21 Position markings (service position)
Assembly with pneumatic driveThe assembly with pneumatic drive does not have any operating elements.
Assembly with manual drive
3
5
7
A0030305
22 Operating elements
3 Manual drive5 Unlocking button (measuring position)7 Unlocking button (service position)
Cleanfit CPA871 Operation options
Endress+Hauser 29
5.3 Manual operation
OUT - IN
A
B
A0030330
23 Direction of rotation
A Unlocking button (service position)B Unlocking button (measuring position)
Moving the assembly from the service position to the measuring position1. Press the unlocking button (A).
2. Rotate the drive in a clockwise direction, so that the sensor holder enters the process(this is possible only if a sensor is installed).
3. Rotate the drive until the lock engages.
Moving the assembly from the measuring position to the service position1. Press the unlocking button (B).
2. Rotate the drive in an anti-clockwise direction as far as it will go (service position).
3. Perform the required service activities.
5.4 Pneumatic operationThe assembly can be moved only if a sensor is installed.
The operation of the pneumatic version depends on the controller in question. TheOperating Instructions can be found in the manual for the controller.
Use a pneumatic change-over valve (4/2-way valve) to move the assembly. Connectboth inputs. If you connect only one input (e.g. for test purposes), the piston will beblocked, as the sensor guide moves before the limit position lock is deactivated.
Operation options Cleanfit CPA871
30 Endress+Hauser
5.4.1 Moving the assembly if the compressed air supply fails
1
2
A0030306
24 Failure of compressed air supply
1 Limit position lock for service position2 Limit position lock for measuring position
LCAUTIONRisk of injury due to high medium pressure‣ Depressurize the system
If the compressed air supply fails, you can still move the assembly manually. Proceed asfollows here:
1. Use an open-ended wrench (AF 17 mm) to unscrew and remove both limit positionlocks (items 1 and 2), and also remove the internal component. If the internalcomponent is stuck (if an attempt was made to move the assembly withoutdeactivating the limit position lock), move the assembly manually to the respectivelimit position.
2. Move the assembly to the desired position.
3. Screw the limit position lock back in.
Cleanfit CPA871 Maintenance
Endress+Hauser 31
6 MaintenanceLWARNING
Risk of injury if medium escapes‣ Before each maintenance task, ensure that the process pipe is empty and rinsed.‣ Move the assembly to the "Service" position.
The assembly drive is maintenance-free. It is not possible to carry out maintenance orrepair work on the drive.
6.1 Maintenance intervalsThe specified intervals serve as a guide. For harsh process or ambient conditions, it isrecommended that the intervals be reduced accordingly. Cleaning intervals for thesensor and assembly are dependent on the medium.
Interval Maintenance measures
Monthlyor after 500 strokes (whatever comesfirst)
Check the following for leaks:• rinse lines• process connection• compressed air hoses (pneumatic version).
Does medium escape from the leakage hole when the assembly ismoving? This may indicate that the internal O-rings in the servicechamber are defective.
‣ Check the leakage hole in the service chamber.
1. Remove the sensor and check the sensor for deposits.
2. If deposits are present, check the cleaning cycle (cleaningagents, temperature, duration, flow volume).
When process pressure is applied and the rinse lines are detached,there should be no discharge of medium from the assembly's rinseconnections.
‣ Check for defective process seal(s).
Biannuallyor after 5000 strokes (whichever comesfirst)
In addition:
‣ Replace the seals that are in contact with medium.
Biannuallyor after 5000 strokes (whichever comesfirst)
1. Check the locking mechanism is not blocked.
2. Remove the sensor. The contact surface of the sensor in the assembly is
spring-loaded and must be free to move.
Possible cause of failure: Contamination inside the drive, e.g. causedby a broken sensor.
Maintenance Cleanfit CPA871
32 Endress+Hauser
6.2 Cleaning the assemblyTo ensure stable and reliable measurements, the assembly and the sensor must be cleanedregularly. The frequency and intensity of the cleaning process depend on the medium.
6.2.1 Manually operated assemblyAll parts in contact with the medium, such as the sensor and the sensor guide, must becleaned regularly. To do so, remove the sensor 1).• Remove light dirt and fouling with suitable cleaning solutions. (→ 33)• High levels of contamination are removed using a soft brush and a suitable cleaning
agent.• For very persistent dirt, soak the parts in a cleaning solution. Then clean the parts with a
brush.
A typical example of a cleaning interval would be 6 months in the case of drinkingwater.
6.2.2 Pneumatically controlled assemblyRegular, pneumatically controlled cleaning is possible using the rinse water connectionand the appropriate equipment.
6.3 Cleaning the sensor‣ Clean the ORP electrodes mechanically only and always use water. Never clean with
chemical cleaning agents. Such cleaning agents cause a potential to build up at theelectrode which takes a few hours to dissipate. The potential causes errors in themeasurement.
‣ Do not use abrasive cleaners. These can cause irreparable damage to the sensor.‣ Where necessary, perform a new calibration following the cleaning process.
You must clean the sensor:• Before every calibration• Regularly during operation• Before returning it for repairs
You can remove the sensor and clean it manually, or you can clean it in automatic mode 2)
using the rinse water connection.
1) same as sensor installation, but in reverse order2) only if the assembly is fitted out accordingly
Cleanfit CPA871 Maintenance
Endress+Hauser 33
6.4 Cleaning agentThe choice of cleaning agent depends on the degree and type of contamination. The mostcommon types of contamination and the appropriate cleaning agents can be found in thefollowing table.
Type of soiling Cleaning agent
Greases and oils Hot water or tempered, surfactant-containing (alkaline)media 1) or water-soluble, organic solvents (e.g.ethanol)
Limescale deposits, metalhydroxide buildup, lyophobicbiological buildup
Approx. 3% hydrochloric acid
Sulfide deposits Mixture of 3% hydrochloric acid and thiocarbamide(commercially available)
Protein buildup Mixture of 3% hydrochloric acid and pepsin(commercially available)
Fibers, suspended substances Pressurized water, possibly surface-active agents
Light biological buildup Pressurized water
1) Do not use with the Tophit ISFET sensor! Instead, use commercially available acid cleaners for the foodindustry (e.g. P3-horolith CIP, P3-horolith FL, P3-oxonia active).
NOTICEHealth hazard due to solvents‣ Do not use any halogen-containing, organic solvents or acetone. These solvents may
destroy plastic components of the sensor and are also suspected carcinogens (e.g.chloroform).
Maintenance Cleanfit CPA871
34 Endress+Hauser
6.5 Replacing sealsTo replace the seals in the assembly, you must interrupt the process and remove theassembly completely.
LCAUTIONRisk of injury due to residual medium and elevated temperatures‣ When handling parts that are in contact with the medium, protect yourself from
residual medium and elevated temperatures. Wear protective goggles and safetygloves.
‣ Clean the assembly before replacing the seals. (→ 32)
Preparation:1. Interrupt the process. Pay attention to residual medium, residual pressure as well as
elevated temperatures.
2. Move the assembly to the service position.
3. Completely detach the assembly from the process connection.
4. Clean the assembly. (→ 32)
Cleanfit CPA871 Maintenance
Endress+Hauser 35
6.5.1 Standard version
Seal replacement in the process connection
1 1
A0030290
25 Replacing seals, Part 1
1 Fixing screws
1. Release the four securing screws (item 1).
2
3
A0030291
26 Replacing seals, Part 2
2 Process connection3 O-ring in process connection
2. Remove the process connection (pos. 2).
3. Remove the o-ring (pos. 3) from the process connection (sealing disk).
4. Apply a thin layer of grease to the new o-ring (e.g. Klüber Paraliq GTE 703).
5. Insert the o-ring into the process connection.
Maintenance Cleanfit CPA871
36 Endress+Hauser
Seal replacement in the rinse connection
4
5
4
5
A0030292
27 Replacing seals, Part 3
4 Lock nut5 Rinse connection adapter
1. Release the lock nuts (pos. 4) using an open-ended wrench or socket wrench (AF 19mm, in protection cap).
2. Unscrew and remove the two rinse connection adapters (pos. 5) using an open-endedwrench or socket wrench (AF 17 mm, in protection cap).
10 10
A0030315
28 Replacing seals, Part 3
10 O-rings, rinse connection adapter
3. Remove the O-rings indicated (item 10).
4. Apply a thin layer of grease to the new o-rings (e.g. Klüber Paraliq GTE 703) andinsert the o-rings into the corresponding grooves.
Cleanfit CPA871 Maintenance
Endress+Hauser 37
Seal replacement in the support housing
1
A0030310
29 Replacing seals, Part 4
1 Fixing screws
1. Release the four securing screws (item 1).
10
8
7
12
11
6
9
10
A0030293
30 Replacing all seals
6789101112
Service chamberSupport housingO-ring, support housingO-rings, bottom of service chamberO-rings, rinse connection adapterO-rings, top of service chamberO-ring, outer service chamber
2. Remove the support housing (pos. 7).
3. Pull the service chamber (pos. 6) out of the support housing.
4. Remove the o-rings indicated.
5. Apply a thin layer of grease to the new o-rings (e.g. Klüber Paraliq GTE 703) andinsert the o-rings into the corresponding grooves.
Maintenance Cleanfit CPA871
38 Endress+Hauser
Assembly
Fit support housing and service chamber together
7
14
13
6
A0030343
31 Assembling the support housing
6 Service chamber7 Support housing13 Positioning pin14 Positioning groove
1. Place the support housing (pos. 7) on a flat surface (the positioning groove (pos. 14)is visible from above).
2. Place the service chamber (pos. 6) on the support housing and let the servicechamber glide into the support housing.
3. Place the positioning pin (item 13) above the corresponding groove and push theservice chamber in the remaining millimeters.
4. Fit the process connection on the support housing and tighten the securing screwswith a torque of 4 Nm.
Tighten the lock nut
5
4
A0030344
32 Assembling the rinse connection
4 Lock nut5 Rinse connection adapters
1. Turn the lock nut (pos. 4) on both of the rinse connection adapters (pos. 5) as far aspossible in the direction of the arrow.
2. Insert the rinse connection adapters with o-rings into the support housing (open-ended wrench or socket wrench AF 17 mm).
3. Tighten the lock nut (AF 19 mm) (in the opposite direction to the arrow).
4. Fit the support housing on the assembly. Pay attention to the positioning pin.Tighten the four securing screws with a torque of 4 Nm.
Cleanfit CPA871 Maintenance
Endress+Hauser 39
6.5.2 Immersion chamber version
Seal replacement in the process connection
1 1
A0030294
33 Replacing seals, Part 1
1. Release the four securing screws (item 1).
2
3
A0030295
34 Replacing seals, Part 2
2 Process connection3 O-ring in process connection
2. Remove the service chamber (pos. 3) together with the process connection (pos. 2).
3. Remove the o-ring (pos. 3) from the process connection (sealing disk).
4. Apply a thin layer of grease to the new o-ring (e.g. Klüber Paraliq GTE 703).
5. Insert the o-ring into the process connection.
Maintenance Cleanfit CPA871
40 Endress+Hauser
Seal replacement in the rinse connection adapter
4
5
4
5
A0030292
35 Replacing seals, Part 3
4 Lock nut5 Rinse connection adapter
1. Release the lock nuts (pos. 4) using a 19 mm open-ended wrench or socket wrench(in protection cap).
2. Unscrew and remove the two rinse connection adapters (pos. 5).
10 10
A0030315
36 Replacing seals, Part 3
10 O-rings, rinse connection adapter
3. Remove the O-rings indicated (item 10).
4. Apply a thin layer of grease to the new o-rings (e.g. Klüber Paraliq GTE 703) andinsert the o-rings into the corresponding grooves.
Cleanfit CPA871 Maintenance
Endress+Hauser 41
Seal replacement in the immersion chamber
7
8
9
10
11
12
13
14
15
16
17
18
19 19
A0030329
37 Replacing all seals
78910111213141516171819
Support housingImmersion chamber - top partImmersion chamber - middle partImmersion chamber - bottom partFixing screwsO-ring, outer service chamberO-ring, top of service chamberO-ring, top part of inner service chamberO-ring, top part of inner service chamberMolded seal (ensure correct orientation)O-ring, top of service chamberO-rings, bottom of service chamberRinse chamber inlet and outlet
1. Remove the support housing (item 7) with the immersion chamber (items 8 to 10).
2. Pull the immersion chamber out of the support housing.
3. Remove the top part of the immersion chamber (pos. 8).
4. Release the three screws (pos. 11) and remove the bottom part of the immersionchamber.
5. Remove the o-rings and the moulded seal (pos. 12 to 18).
6. Apply a thin layer of grease to the new o-rings (e.g. Klüber Paraliq GTE 703) andinsert the o-rings into the corresponding grooves.
7. Insert the molded seal in such a way that both openings with the stop ridges (item19) are positioned above the rinse chamber inlet and outlet.
Maintenance Cleanfit CPA871
42 Endress+Hauser
AssemblyIn the immersion chamber version, the inlet and outlet of the service chamber are fixed.
When assembling the immersion tube, please ensure that the leakage borehole (pos.11), the service chamber outlet (pos. 12) and the marking at the bottom of theimmersion chamber (pos. 16) are all in one line.
Fit the components of the immersion chamber together.
8
9
10
11
12
13
15
16
A0030368
38 Assembling the immersion chamber
89101112131516
Immersion chamber - top partImmersion chamber - middle partImmersion chamber - bottom partLeakage holeRinse connection - outlet of service chamberPositioning pinPositioning pinPosition marking
1. Fit the bottom part (item 10) and middle part (item 9) of the immersion chambertogether. Ensure correct orientation!
2. Screw both parts securely together using the three securing screws (item 11).
3. Attach the top part (item 8) of the immersion chamber.
Cleanfit CPA871 Maintenance
Endress+Hauser 43
7 14
13
A0030347
39 Assembling the support housing and immersion chamber
4 Support housing13 Positioning pin14 Positioning groove
4. Place the support housing (pos. 7) on a flat surface (the positioning groove (pos. 14)is visible from above).
5. Place the immersion chamber on the support housing and let the immersion chamberglide into the support housing.
6. Place the positioning pin (item 13) above the corresponding groove and push theservice chamber in the remaining millimeters.
7. Fit the process connection on the support housing and tighten the securing screwswith a torque of 4 Nm.
5
4
A0030344
40 Assembling the rinse connection
4 Lock nut5 Rinse connection adapter
8. Turn the lock nut (pos. 4) on both of the rinse connection adapters (pos. 5) as far aspossible in the direction of the arrow.
9. Insert the rinse connection adapters with o-rings into the support housing (open-ended wrench or socket wrench AF 17 mm).
10. Tighten the lock nut (AF 19 mm) (in the opposite direction to the arrow).
11. Fit the support housing on the assembly. Pay attention to the positioning pin.Tighten the four securing screws with a torque of 4 Nm.
Repairs Cleanfit CPA871
44 Endress+Hauser
7 RepairsLWARNING
Danger resulting from improper repair‣ Damage to the assembly, which compromises pressure safety, must be repaired only by
authorized and qualified personnel.‣ Damage to the drive can be repaired only at the place of manufacture. Repairs cannot
be carried out onsite.‣ Following each repair and maintenance task, the assembly must be checked for leaks
using appropriate procedures. Following this, the assembly must again comply with thespecifications in the technical data.
Replace all other damaged components immediately.
7.1 Spare partsFor more detailed information on spare parts kits, please refer to the "Spare Part FindingTool" on the Internet:
www.endress.com/spareparts_consumables
7.2 ReturnThe product must be returned if repairs or a factory calibration are required, or if thewrong product was ordered or delivered. As an ISO-certified company and also due to legalregulations, Endress+Hauser is obliged to follow certain procedures when handling anyreturned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return proceduresand conditions at www.endress.com/support/return-material.
7.3 DisposalThe device contains electronic components and must therefore be disposed of inaccordance with regulations on the disposal of electronic waste.
Observe the local regulations.
Cleanfit CPA871 Accessories
Endress+Hauser 45
8 AccessoriesThe following are the most important accessories available at the time thisdocumentation was issued. For accessories not listed here, please contact your serviceor sales office.
The following accessories can be ordered via the product structure (see orderinginformation):• Weld-in adapter G1¼, straight, 35 mm, 1.4435 (AISI 316 L), safety nozzle• Weld-in adapter G1¼, angled, 35 mm, 1.4435 (AISI 316 L), safety nozzle
G 1¼
12
Ø 42
Ø 25
Ø 25
12
Ø 42
G 1¼
15°
35
(1
.38
)
(1.65)
(0.98)
(0.4
7)
(1.65)
(0.98)
(0.4
7)
35
(1
.38
)
20
(0.7
9)
23.5
(0.9
3)
35
(1.3
8)
20
(0.7
9)
23.5
(0.9
3)
35
(1.3
8)
A0028744
41 Weld-in adapter (safety nozzle), dimensions in mm (inch)
• Dummy plug G1¼, 1.4435 (AISI 316 L), FPM - FDA• Sensor dummy 120 mm, 1.4435 (AISI 316 L), Ra = 0.38 µm• Sensor dummy 225 mm, 1.4435 (AISI 316 L), Ra = 0.38 µm• Sensor dummy 360 mm, 1.4435 (AISI 316 L), Ra = 0.38 µm• Kit, seals for non-wetted parts• Kit, FKM seals, G1¼, wetted parts• Kit, FKM seals, immersion chamber version, wetted parts• Kit, seal, wetted, EPDM• Kit, seal, wetted, FKM• Kit, seal, FFKM, basic, wetted• Cable, plug-in, limit switch, M12, 5 m• Cable, plug-in, limit switch, M12, 10 m• Tool in case for installation/removal
Accessories Cleanfit CPA871
46 Endress+Hauser
8.1 Installation material for rinse connectionsFilter set CPC310, CVC400• Water filter (dirt trap) 100 µm, complete, incl. angle bracket• Order No. 71031661
Pressure reducer kit• Complete, incl. manometer and angle bracket• Order No. 51505755
Hose connection set G¼, DN 12• 1.4404 (AISI 316L) 2 x• Order No. 51502808
Hose connection set G¼, DN 12• PVDF (2 x)• Order No. 50090491
8.2 Cleaning systemsAir-Trol 500• Control unit for Cleanfit retractable assemblies• Order No. 50051994
Technical Information TI00038C/07/EN
Cleanfit Control CYC25• Converts electrical signals into pneumatic signals to control pneumatically-operated
retractable assemblies or pumps in conjunction with Liquiline CM44x• Wide range of control options• Product Configurator on the product page: www.endress.com/cyc25
Technical Information TI01231C
Cleanfit CPA871 Accessories
Endress+Hauser 47
8.3 Flow vesselsFlow vessel in different materials• Coupling: 2 x G¾ internal thread• Material: stainless steel 1.4404 (AISI 316 L), order no. TSP 71268971• Material: PVDF, order no. TSP 71268972• Material: PEEK, order no. TSP 71268973
G¾
G¾
68 (
2.6
8)
Ø 79.5 (3.13)
A0026651
42 Dimensions in mm (inch)
Flow vessel made of stainless steel• Flange DN25 PN16• Material: stainless steel 1.4404 (AISI 316 L)• Order no. TSP 71276097
160 (6.30)
60 (2.36)
Ø 1
15 (
4.5
3)
A0026652
43 Dimensions in mm (inch)
Accessories Cleanfit CPA871
48 Endress+Hauser
8.4 Sensors
8.4.1 pH sensorsOrbisint CPS11D / CPS11• pH electrode for process technology• Optional SIL version for connecting to SIL transmitter• With dirt-repellent PTFE diaphragm• Product Configurator on the product page: www.endress.com/cps11d or
www.endress.com/cps11
Technical Information TI00028C
Memosens CPS31D• pH electrode with gel-filled reference system with ceramic diaphragm• Product Configurator on the product page: www.endress.com/cps31d
Technical Information TI00030C
Ceraliquid CPS41D / CPS41• pH electrode with ceramic junction and KCl liquid electrolyte• Product Configurator on the product page: www.endress.com/cps41d or
www.endress.com/cps41
Technical Information TI00079C
Ceragel CPS71D / CPS71• pH electrode with reference system including ion trap• Product Configurator on the product page: www.endress.com/cps71d or
www.endress.com/cps71
Technical Information TI00245C
Orbipore CPS91D / CPS91• pH electrode with open aperture for media with high dirt load• Product Configurator on the product page: www.endress.com/cps91d or
www.endress.com/cps91
Technical Information TI00375C
8.4.2 ORP sensorsOrbisint CPS12D / CPS12• ORP sensor for process technology• Product Configurator on the product page: www.endress.com/cps12d or
www.endress.com/cps12
Technical Information TI00367C
Ceraliquid CPS42D / CPS42• ORP electrode with ceramic junction and KCl liquid electrolyte• Product Configurator on the product page: www.endress.com/cps42d or
www.endress.com/cps42
Technical Information TI00373C
Ceragel CPS72D / CPS72• ORP electrode with reference system including ion trap• Product Configurator on the product page: www.endress.com/cps72d or
www.endress.com/cps72
Technical Information TI00374C
Cleanfit CPA871 Accessories
Endress+Hauser 49
8.4.3 pH ISFET sensorsTophit CPS441D / CPS441• Sterilizable ISFET sensor for low-conductivity media• Liquid KCl electrolyte• Product Configurator on the product page: www.endress.com/cps441d or
www.endress.com/cps441
Technical Information TI00352C
Tophit CPS471D / CPS471• Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process
engineering• Water treatment and biotechnology• Product Configurator on the product page: www.endress.com/cps471d or
www.endress.com/cps471
Technical Information TI00283C
Tophit CPS491D / CPS491• ISFET sensor with open aperture for media with high dirt load• Product Configurator on the product page: www.endress.com/cps491d or
www.endress.com/cps491
Technical Information TI00377C
8.4.4 pH and ORP combined sensorsMemosens CPS16D• Combined pH/ORP sensor for process technology• With dirt-repellent PTFE diaphragm• With Memosens technology• Product Configurator on the product page: www.endress.com/cps16D
Technical Information TI00503C
Memosens CPS76D• Combined pH/ORP sensor for process technology• Hygienic and sterile applications• With Memosens technology• Product Configurator on the product page: www.endress.com/cps76d
Technical Information TI00506C
Memosens CPS96D• Combined pH/ORP sensor for chemical processes• With poison-resistant reference with ion trap• With Memosens technology• Product Configurator on the product page: www.endress.com/cps96d
Technical Information TI00507C
8.4.5 Conductivity sensorsMemosens CLS82D• Four-electrode sensor• With Memosens technology• Product Configurator on the product page: www.endress.com/cls82d
Technical Information TI01188C
Accessories Cleanfit CPA871
50 Endress+Hauser
8.4.6 Oxygen sensorsOxymax COS22D / COS22• Sterilizable sensor for dissolved oxygen• With Memosens technology or as an analog sensor• Product Configurator on the product page: www.endress.com/cos22d or
www.endress.com/cos22
Technical Information TI00446C
Cleanfit CPA871 Technical data
Endress+Hauser 51
9 Technical data
9.1 Environment
Ambient temperaturerange
-10 to +70 °C (+10 to +160 °F)
Storage temperature -10 to +70 °C (+10 to +160 °F)
9.2 Process
Process temperature -10 to +140 °C (14 to 284 °F) for all materials except PVDF and conductive PVDF
-10 to +100 / 90 °C (14 to 212 / 194 °F) for PVDF and conductive PVDF materials
Process pressure forpneumatic drive
Materials: 1.4404, Alloy C22, PEEKBasic and immersion chamber version: 16 bar (232 psi) up to 140 °C (284 °F)Materials: PVDF, conductive PVDFBasic version: 16 bar (232 psi) up to 100 °C (212 °F)Immersion chamber version: 4 bar (58 psi) to 90 °C (194 °F)
The service life of the seals is reduced if process temperatures are constantly high or ifSIP is used. The other process conditions may also reduce the service life of the seals.
Process pressure formanual drive
Materials: 1.4404, Alloy C22, PEEKBasic and immersion chamber version: 8 bar (116 psi) up to 140 °C (284 °F)Materials: PVDF, conductive PVDFBasic version: 8 bar (116 psi) up to 100 °C (212 °F)Immersion chamber version: 4 bar (58 psi) to 90 °C (194 °F)
The service life of the seals is reduced if process temperatures are constantly high or ifSIP is used. The other process conditions may also reduce the service life of the seals.
Pressure-temperatureratings for pneumatic drive p [bar]
T[°C]20 806040
T[°F]
p [psi]
0
32
0 0
16232
100 120 140
68 104 140 176 212 248 284
A0029355-EN
44 Pressure temperature ratings for basic and immersion chamber version for materials 1.4404, Alloy C22and PEEK
Technical data Cleanfit CPA871
52 Endress+Hauser
p [bar]
T[°C]20 806040
T[°F]
p [psi]
0
32
0 0
16232
100 120 140
68 104 140 176 212 248 284
8116
A0029358-EN
45 Pressure temperature ratings for basic version for materials PVDF and conductive PVDF
Pressure-temperatureratings for manual drive p [bar]
T[°C]20 806040
T[°F]
p [psi]
0
32
0 0
16232
100 120 140
68 104 140 176 212 248 284
8116
A0029356-EN
46 Pressure temperature ratings for basic and for immersion chamber version for materials 1.4404, AlloyC22 and PEEK
p [bar]
T[°C]20 806040
T[°F]
p [psi]
0
32
0 0
16232
100 120 140
68 104 140 176 212 248 284
8116
A0029357-EN
47 Pressure temperature ratings for basic version for materials PVDF and conductive PVDF
Cleanfit CPA871 Technical data
Endress+Hauser 53
Pressure temperatureratings for immersionchamber version (PVDF)
p [bar]
T[°C]20 806040
T[°F]
p [psi]
0
32
0 0100 120 140
68 104 140 176 212 248 284
8116
232 16
1
2
458
3
A0029359-EN
48 Pressure temperature ratings for immersion chamber version for materials PVDF and conductive PVDF
1 Pneumatic drive2 Manual drive3 Pneumatic and manual drive
Technical data Cleanfit CPA871
54 Endress+Hauser
9.3 Mechanical construction
Design, dimensions → Section "Installation"
Weight Depends on version:Pneumatic drive: 3.8 to 6 kg (8.4 to 13.2 lbs) depending on versionManual drive: 3 to 4.5 kg (6.6 to 9.9 lbs) depending on version
Materials In contact with medium
Seals: EPDM/FPM (Viton)/FFKM
Immersion tube, processconnection, service chamber:
Stainless steel 1.4404 (AISI 316L) Ra < 0.76/PEEK/Alloy C22 Ra <0.76/PVDF/conductive PVDF
Rinse connections: Stainless steel 1.4404 (AISI 316L) and Alloy C22
Not in contact with medium
Manual drive: Stainless steel 1.4301 (AISI 304) or 1.4404 (AISI 316L), plastics PPSCF15, PBT, PP
Pneumatic drive: Stainless steel 1.4301 (AISI 304) or 1.4404 (AISI 316L), plastics PBT,PP
Sensors Short version Gel sensors, ISFET 120 mmGel sensors, ISFET 225 mmKCl sensors 225 mm
Long version Gel sensors, ISFET 225 mmGel sensors, ISFET 360 mm
Immersion chamber version Gel sensors, ISFET 225 mmKCl sensors 360 mm
Limit position switches Switching element function: NAMUR NC contact (inductive)Switching distance: 1,5 mm (0.06 ")Nominal voltage: 8 VSwitching frequency: 0 to 5000 HzHousing material: Stainless steel
Index
Endress+Hauser 55
Index
AAccessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
CCertificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Cleaning agent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Connection
Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
DDesignated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
FFailure of compressed air supply . . . . . . . . . . . . . . . . 30
IImmersion depths . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . 7Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 17Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . 10Installing the sensor . . . . . . . . . . . . . . . . . . . . . . . . . 22
LLimit position switches . . . . . . . . . . . . . . . . . . . . . . . 20
MMaintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . 31Manual operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
NNameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
OO-rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . 28Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Pneumatic: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
PPneumatic connection . . . . . . . . . . . . . . . . . . . . . . . . 19Pneumatic operation . . . . . . . . . . . . . . . . . . . . . . . . . 29Post-installation check . . . . . . . . . . . . . . . . . . . . . . . 26Product identification . . . . . . . . . . . . . . . . . . . . . . . . . 7
RRepairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Rinse connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
SSafety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
TTechnical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
UUse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
WWarnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
www.addresses.endress.com
*71324465*71324465