© 2010-2011 Masimo Corporation
Probe Request
Probe Response
Authentication Request
Authentication Response
Association Request
Association Response
CLIENT ACCESS POINT
Let’s Get Associated!
![Page 2: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/2.jpg)
© 2010-2011 Masimo Corporation
Episodic vs Continuous Wireless Connectivity
Are devices ready for real time?
FDA 510(k) pending
Brian Long, CWNADirector, Field Systems Operations
![Page 3: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/3.jpg)
© 2010-2011 Masimo Corporation
Presentation Scope
> Wireless Demands & Challenges in Hospital Environments
> Real-time, continuous Patient Data
> Network Requirements
> Testing Approach
> Real World Results
> Recommendations & Conclusions
![Page 4: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/4.jpg)
© 2010-2011 Masimo Corporation
Hospital Wireless Business Case Demands
EPISODIC CONTINUOUS
Voice Waveforms
VideoReal-time, Continuous
Parametric Data
EMR Applications
PCA Drug Library Updates
Real-time Location Systems
Guest Network
Many more!
![Page 5: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/5.jpg)
© 2010-2011 Masimo Corporation
“Healthcare: The Most Challenging WiFi Environment?”Marcus Burton, CWNE #78
![Page 6: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/6.jpg)
© 2010-2011 Masimo Corporation
Masimo Patient SafetyNet™
Connect patients with caregivers quickly, easily, and accurately
Accurate, reliable
wireless
measurements at
the bedside
Actionable alarm
notification to
qualified
clinicians
Rapid response
initiated only
when needed
Earlier and more
appropriate clinical
intervention
![Page 7: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/7.jpg)
© 2010-2011 Masimo Corporation
IT Network Components
> Choice of bedside devices
> Low data rate streaming data
> Hardwired Ethernet (802.3)
> WLAN (802.11b/g/n or a/n)
> Encryption: WEP / WPA / WPA2
> Authentication: PSK / 802.1X
> HP Server
> HP TouchScreen
> Installed in Data Center
> Remote VPN Support
> Supports 80 Bedside Devices
> Supports 4 Separate Clinical
Assignment Stations
> Remote notification choices
> Emergin interface option
> Dedicated or shared
notification devices
> Support IP Phones, Pagers
![Page 8: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/8.jpg)
© 2010-2011 Masimo Corporation
Coverage
Latency
Packet Loss
Availability
Clinician Experience
Testing Approach
24h
r N
etw
ork
Perf
orm
an
ce T
est
![Page 9: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/9.jpg)
© 2010-2011 Masimo Corporation
Requirements
COVERAGE
• Primary -67dB
• Secondary -72dB
LATENCY
• Wired <30ms
• Wireless <100ms
PACKET LOSS
• < 2% for Total Testing Time
AVAILABILITY
• => 99.9% for Total Testing TimeS
CLINICIAN EXPERIENCE
• Network/Application Disconnect Report
![Page 10: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/10.jpg)
© 2010-2011 Masimo Corporation
Initial Wireless Site Survey – Primary Coverage
![Page 11: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/11.jpg)
© 2010-2011 Masimo Corporation
Initial Wireless Site Survey – Secondary Coverage
![Page 12: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/12.jpg)
© 2010-2011 Masimo Corporation
Network/Application Disconnect ReportIn
str
um
ents
![Page 13: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/13.jpg)
© 2010-2011 Masimo Corporation
Wireless Site Survey [Primary]
Network Remediation
![Page 14: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/14.jpg)
© 2010-2011 Masimo Corporation
Wireless Site Survey [Secondary] –
Network Remediation
![Page 15: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/15.jpg)
© 2010-2011 Masimo Corporation
Network/Application Disconnect ReportIn
str
um
ents
![Page 16: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/16.jpg)
© 2010-2011 Masimo Corporation
Dartmouth Study
![Page 17: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/17.jpg)
© 2010-2011 Masimo Corporation
Clinically Proven to Help
Clinicians Improve Outcomes
Taenzer A et al. Anesthesiology, 2010.
![Page 18: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/18.jpg)
© 2010-2011 Masimo Corporation
Patient SafetyNet Reliability, Patient Compliance,
and Alarm Frequency
Ten month study, covering 36 bed unit, 2,841 patients and
9,978 monitored days and over 1 billion data points
Wireless System Uptime Reliability 99.9995%
Patient Compliance 98.2%
Average Alarms per patient per day 4
Taenzer A et al. Anesthesiology, 2010.
![Page 19: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/19.jpg)
© 2010-2011 Masimo Corporation
Patient SafetyNet Paging Protocol
Page Initial
Primary
Esc. Primary
Secondary
Re-Esc.
Primary
Secondary
Supervisor
![Page 20: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/20.jpg)
© 2010-2011 Masimo Corporation
Identify Network Business Objectives
![Page 21: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/21.jpg)
© 2010-2011 Masimo Corporation
Mitigate Risks w/ ISO-80001 In Mind
![Page 22: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/22.jpg)
© 2010-2011 Masimo Corporation
Recommendation – Invest in Wireless Education
![Page 23: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/23.jpg)
© 2010-2011 Masimo Corporation
![Page 24: CLIENT ACCESS POINT - tcbi.org](https://reader035.vdocument.in/reader035/viewer/2022072806/62e115a2e2e3683d171e353a/html5/thumbnails/24.jpg)
© 2010-2011 Masimo Corporation