code p008 date: 20.2.2007 1 clarity user training advanced clarity user training
TRANSCRIPT
Code P008 Date: 20.2.20071
Clarity User TrainingAdvanced
Clarity User Training
Code P008 Date: 20.2.20072
For advanced users
User settings and display options
Sequences
Multidetector systems
Export data
Report setup
Batch data reprocessing
Code P008 Date: 20.2.20073
For advanced users
Data archivation
Column Performance
Noise and Drift
Special calculations
Control modules
GLP a CFR21part11
Code P008 Date: 20.2.20074
User settings are stored in individual users desktop *.dsk
files.For this reason they are set in
the instrument window and not in the main window
User settings
Code P008 Date: 20.2.20075
Text Label and Linesfor legends and arrows
Chromatogram/Create Label
Moving and Scaling of
Chromatograms
Math operation
s
Overlay activation
Labels and display options
Code P008 Date: 20.2.20076
Colored peak area in chromatogram
Peaks in calibration can have a color assigned.
When these peaks are identified in chromatogram they will be
automatically colored
Code P008 Date: 20.2.20077
Colored peak area in chromatogram
Highlight currently selected peaks
Simply select a peak in the chromatogram result table.
The result table selection overrides the color defined in calibration
Code P008 Date: 20.2.20078
Colored peak area in chromatogram
Highlight peaks selected in Results table (signal color used)
Use Peak Color set in Calibration file
Code P008 Date: 20.2.20079
Context menu invoked by right mouse click in table
Table customizing options
Code P008 Date: 20.2.200710
Common columns in Summary table
Independent setting of columns Common - for all compounds
Summary – compound specific
Code P008 Date: 20.2.200711
Summary table options
Code P008 Date: 20.2.200712
Context menu invoked by right mouse click in
Chromatogram window
Axis properties setting
Graph properties
setting
Graph display options
Code P008 Date: 20.2.200713
Gradient display options
Setting for individual
signals
Chromatogram display options
Code P008 Date: 20.2.200714
Customizing Toolbars
Reset Allrestores original
settings
Code P008 Date: 20.2.200715
Define a serie of samples to be measured
Essential for working with autosamplers For controled autosamplers the Vial number and Injection
volume are used from this table
Allows modifications during a run for not yet processed samples Simply add samples by filling any field in the last empty row.
Other fields will be copied from previous. Several samples can be added by copying from a table
Allows for import of sample data from text files
Allows reprocessing of last measured sample set
Sequence table
Code P008 Date: 20.2.200716
Start, End vial and Injections number. Several vials and/or
injections defined on single row
File name, automatic
%variables for SV, EV, IV
Automatic recalibration
from sequence
Row status indicator
Use row, add new rows during
sequence run
Tooltip with actual file name
Individual Post-Run
settings for each row
Sequence table window
Code P008 Date: 20.2.200717
Pasive sequenceClarity only expects Start signal,
analysis time is set on autosamplerActive sequence
Clarity sends a Ready signal to autosampler and waits for Start
signalAS control (Active sequence)
Autosampler injects according to SV, EV, IV and
Inj. Volume in sequence table
Editation of template method used on
selected row
Initial value for %n
Sequence options
Code P008 Date: 20.2.200718
Run sequencestarts complete sequence
(resets the status)
Pause sequence
Resume sequencecontinues from last
injection
Status of the sequence. Stored at Pause, Abort, Stop
Can be changed by Edit/Reset Status for selected rows
Repeat Injection
Skip Vial
Stop sequence will stop the sequence run. Repeated Stop will stop the current
analysis
Sequence control
Code P008 Date: 20.2.200719
Select the file to import
Assign the respective fields to be importedOnly Start vial (SV) and file name must be
set (marked in bold)Other fields will be filled by default values
Sequence import
Code P008 Date: 20.2.200720
Export data to other programs to prevent typing errors Simple Copy and Paste (Ctrl C – Ctrl V)
Export Data settings common for manual export from chromatogram window
automatic export from postrun at the end of each run
automatic export by batch reprocess of chromatograms or sequence
Export of chromatograms in common formats
Export data
Code P008 Date: 20.2.200721
Copy a picture to MS Office (including labels) via Clipboard or store it
in *.emf format
Copy a selected part of table using Copy (Ctrl C) and Paste (Ctrl V)
Export data - simplest
Code P008 Date: 20.2.200722
Headers:Column headers and chromatogram info
Full format:Filename, date and time
before each row
blank field = actual chromatogram name + .txt extensionPath can be used
All exported data are in single file
Export time/signal data pairs
(import to other programs – for
example graphics)
Export data:manual - from Chromatogram window
automatic - according to Setting/Postrunbatch - from Analysis/Batch
Export data - options
Code P008 Date: 20.2.200723
Export data - example
Code P008 Date: 20.2.200724
Report style from Report Setup dialog will be used
do not forget to switch off the Overlay mode
export
reprocess of chromatograms according to instrument method
reprocess whole sequence
electronic signature
File typeChromatograms
Calibration standardsSequence Files
Export according to Settings - Export Data
Electronic signature for series of chromatograms
Batch processing
Code P008 Date: 20.2.200725
Assignment of detectors to Clarity Instruments in the System Configuration dialog
All signals stored in a single chromatogram Result table, Integration table and Acquisition
parameters are individual for each chromatogram
Common calibration file calibration curves are individual for each signal
Summary table for display of results from all signals
Multidetector systems
Code P008 Date: 20.2.200726
Results are displayed individually, some properties like calibration are
common for the whole chromatogram
Connecting link marks signals from one chromatogram. Click icon to set signal
active
Detector Delay compensates for
interdetector volume, inactive for first signal
Multi signal chromatograms
Code P008 Date: 20.2.200727
Summary table displays results for several chromatograms or
signals simultaneously
In overlay, several chromatograms can be
compared. Maximum number can be set in user options
Doubleclick on legend will set signal active
Scale, move and math
operations are active only in
overlay
Ctrl + click will hide/show the signal
Ctrl + click will lock the tool for repe-
titive use.
Overlay
Code P008 Date: 20.2.200728
Report Setup - logo
Code P008 Date: 20.2.200729
Report Setup – tiled signals
Code P008 Date: 20.2.200730
To reduce the data size compressed archive *.dgz files should be used
*.dgz archives always replace all data (files deleted from project will be during archiving deleted also from the archive).
It is advantageous to archive whole projects – should be considered when planning project structure
Archiving and backup
Code P008 Date: 20.2.200731
Archive and Restore for copying or moving projects Selected file types Without compressing = original format and structure compressed *.dgz archives = single file
Archive data
Code P008 Date: 20.2.200732
Separation parameters
Values used for Capacity ratio and Efficiency calculation
Calculation type
Performance tab (hidden in
default layout)
Advanced tab in the Method Setup (displays
template method
Performance parameters
Code P008 Date: 20.2.200733
Actual noise monitoring in the Data Acquisition window
Noise monitoring
Code P008 Date: 20.2.200734
Drift and Noise displayed for selected intervals in
Result table header
Drift and Noise can be used for in User column calculations
Noise and Drift
Code P008 Date: 20.2.200735
ISTD calculations
Normalized % calculations
Groups calibration of peak group like single compound example - isomers (xylen)
Scale factor recalculation of result to other units
Uncalibrated response quantification of unidentified peaks using Uncalibrated
response factor
Injection volume Injected volume correction for manual injection in GC
Calculations
Code P008 Date: 20.2.200736
ISTD calculations are performed:a) Amount for ISTD is set in both calibration and sampleb) both are zero
ISTD compound is calibrated
ISTD amount in sample is given
Only one ISTD compound is allowed in calibration. If the conditions are not
met, no results will be calculated
ISTD peak can be hidden in the Result table
ISTD calculations
Code P008 Date: 20.2.200737
If Sample amount is non-zero, it will replace the sum in the Amount column and serve as a base for Amount
% calculation
Total line is calculated as a
sum of all groups + sum of peaks not belonging to
any group. Groups are
calibrated, thus peaks should not be both calibrated and in a group to
get correct Amount% values
Amount % gives the normalized % if all
integrated peaks are calibrated and Sample
amount is zero
Normalized % calculations
Code P008 Date: 20.2.200738
1st stepAdding peaks to a group in the integration table
2nd stepSetting a Group in Calibration table.
The compound name and amount is set as for
ordinary peaks
Groups
Code P008 Date: 20.2.200739
Scale factor and Units after scaling:Multiplies values in Amount column and
changes the units in column header.Method specific parameter, multiplies also
the entered Sample Amount value
Uncal. Response:Used for quantification of
unidentified compounds. For universal response detectors
(ELSD, FID, RI)
Dilution:Multiplies values in Amount column.
Sample specific parameter (entered from Sequence or Single Analysis window).
Scale Factor and Uncal.Response
Code P008 Date: 20.2.200740
Default Injection Volume must be set first in calibration and Injection Volume set in standards before calibrating
Supposes linear detector response
Can lead to incorrect results, if used improperly
Setting Default
Injection Volume in calibration
Setting Injection Volume for sample and
standard
Injection Volume
Code P008 Date: 20.2.200741
Command Calibration/Show History displays
data used for selected level
Deviation of point from calibration curve in %
Residual sum of deviation squares
Calibration
Code P008 Date: 20.2.200742
Calibration Retention (Kovats) indexes
LOD and LOQ limitsColumns for Retention (Kovats) indexes are
hidden in default settings.
Use setup columns to disdplay them
Amounts lower then specified LOD or LOQ limits are marked in Peak type column
Retention indexes, LOD, LOQ
Code P008 Date: 20.2.200743
For selected instruments the direct control is available. The advantages are
The instrument parameters are stored in a method
It is not necessary to enter the parameters twice, especially useful for autosamplers
The parameters are recorded in the chromatogram
Control Modules
Code P008 Date: 20.2.200744
Control modules enable full control of connected devices like gradient programe and
other settings. The used settings are stored within a
chromatogram.
Control modules (GC)
Code P008 Date: 20.2.200745
Control modules (LC, AS)
Code P008 Date: 20.2.200746
Option to Send/Confirm/Do not send method to instruments after method change
Send Method options
Confirmation dialog
Control Modules
Code P008 Date: 20.2.200747
Control modules overview
Code P008 Date: 20.2.200748
Other devices can be controlled by A/D board digital outputs: gradient or detector program start detector autozero valve switching fraction collector control
Synchronization in Active Sequence Dig. output set in System Configuration, not in the Event table
Additional control
Code P008 Date: 20.2.200749
Set events in the Event Table tab in the Method Setup
Named events
Digital event input type
Output event can trigger run/sequence commands: Stop Abort Skip vial Shutdown Repeat injection
Event table
Code P008 Date: 20.2.200750
Chromatogram history All saved versions are stored in the chromatogram
Versions can be opened (including calibration)
Audit trails Station, Chromatogram, Sequence, Calibration
User access Can be restricted by User Accounts settings
GLP a CFR21 part 11
Code P008 Date: 20.2.200751
Electronic signatures based on User Accounts
based on Certificates
Validation Installation qualification - standard feature
Performance qualification - by optional Validation kit
Operation qualification - SST module for statistical evaluation and pre-set limits control
GLP a CFR21 part 11