Comments on sterilization.
Wim Renders
“Knowledge is the best medical device” (Jan Hoborn)
The road to the goal is the goal itself!
You may never:
• deviate from routine practice without proof that the new technique is an improvement,
• think that the grass is greener on the other side.
Evolutie ingrepen operatiekwartier - AZ Sint-Jan AV
13.560 13.76514.280
16.249
18.00217.107
15.154
0
2.000
4.000
6.000
8.000
10.000
12.000
14.000
16.000
18.000
20.000
2000 2001 2002 2003 2004 2005 2006
Jaartal
Aan
tal
Keep it simple !
Thyroiedectomy:• Farabeuf: 2 cfu• Needle holder: 6 cfu
Thrombectomy:• Suction tube: 14 cfu• Wound spreader: 3 cfu• Forceps: 40 cfu
Rinsing
Ultrasonic unit
Ultrason bath is:
• emptied in the evening and dried out. • The sides and the lid are sprayed with
alcohol 70°, and some alcohol is poured down the drain.
RKI recommendations:
• Detergents pH > 10• Disinfection 5 min 90°C• Sterilisation: 5’ 134°C
Disinfection parameters:• Ao = 600 sec (80°C , Z = 10°C): uncritical • Ao = 3000 sec : semi-critical & critical
Suction tubes are:
• first rinsed with cold water and • than treated with hydrogen peroxide 3
%, • next: normal procedure.
Thyroiedectomy: BeforeAfter• Farabeuf: 2 cfu 10 cfu• Needle holder: 6 cfu 6 cfu
Thrombectomy:• Suction tube: 14 cfu 6 cfu• Wound spreader: 3 cfu 6 cfu• Pincet: 40 cfu 7 cfu
Bowie Dick test
DPCD