Page 1: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

MANUAL

COMMUNICATING WATER LEVEL 77

Im Hölderli 13, CH - 8405 WINTERTHUR (Switzerland)Tel. +41 (0) 52 233 66 66 Fax +41 (0) 52 233 20 53E-Mail: [email protected] Web: www.wylerag.com

WYLER AGInclination measuring system

Page 2: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 1  

Communicating Water Level 77  

with optional depth micrometer and optional connecting hose 

 

Overview 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

   

Measurement Rod 

Measuring Sleeve 

Clamping Screw 

Measuring column 

Millimeter Scale 

Circular Spirit Level 

Port Valve 

Stand 

Clamping Sleeve Thimble Scale 

Connecting Hose 

Page 3: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 2  

Preparing 

1. Alignment oft the measurement object with on board circular spirit level. 

 

 

 

 

 

 

 

 

 

 

2. Connection hose connected with valve 

 

 

   

   

   

   

   

To open the valve turn 

the blue lever 

counterclockwise 90 ° 

   

   

   

   

   

   

 

 

 

   

Page 4: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 3  

3.  Filling of the water.  

 

For longer periods of measuring it is advisable to use distilled water. 

 

‐ Put one measuring column to a low position, e.g. floor. 

 

 

 

 

 

 

 

 

 

‐ Hold the hose higher. Pour the liquid slowly. The liquid has always to rise in the tube. This avoids 

bubbles.  

 

‐ Level up the measuring column. At 

the same time pour the liquid 

slowly into the system. 

 

‐ Level  both measuring columns  to 

the same height.  

 

 

‐ Fill the system until  booth measuring columns are 

              filled 50% with the liquid. 

Page 5: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 4  

 Depth Micrometer 

 

 

 

 

 

 

 

 

Attention: The measuring range of the depth micrometer is 25mm in maximum. 

 

 

Moving of the Measurement Rod: (To adjust the measurement rod  to the current water level) 

 

 

 

 

 

Open the clamping sleeve 

 

 

 

 

 

Open the clamping screw 

 

 

 

 

 

 

 

 

 

 

Page 6: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 5  

 

 

 

 

 

 Move the measuring rod to the  desired 

               position  

 

 

 

 

 

 

 Lock the clamping screw 

 

 

 

 

 

 

 

 

 

Lock the clamping sleeve 

 

 

 

 

 

 

Attention 

Failure to comply with this order may result in serious damage to the thread. 

Page 7: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 6  

Aufstellen der Messzylinder  am Messplatz 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measurement without  depth micrometer 

Measurement  

Reading the column 1 directly on the scale: 68 mm 

Reading the column 1 directly on the scale: 59.5 mm 

 

 

Result: 

Measuring  1 is  8.5 mm higher then measuring 2 

   

   

   

Attention Measurement errors may be caused by the following points: 

• fouling or calcification of the system 

• different air pressure between the measuring stations, caused by draft, air conditioner,  

   or by varying air pressure in two rooms. 

Measuring 1  Measuring 2 

Column 1 Column 2

Page 8: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 7  

Measurement without  depth micrometer 

Measuring  1: 

Set the depth micrometer on the measurement column 

Move the measurement rod close tot he water level. 

Turn the measurement sleeve and move  

 

the  measurement rod lower until   

the water surface 

contacts the needle. 

 

 

 

 

Read out the value 

   

Page 9: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 8  

Read out the value  

The thimble scale shows the result: 

In this case : 7.83 mm 

 

 

 

 

 

 

 

 

 

 

Measuring 2: 

Move the depth micrometer  from measuring 1 to measuring 2  

Repeat the measurement according  measurement1 

 

 

 

Read out the value at the thimble scale: 

21.81 mm 

 

Calculate the difference of measurement 1‐ measurement 2  

Difference: 7.83‐21.81= 13.98 mm 

Point of measurement  2 is 13.98 mm higher then Point of 

measurement  2 

It is advisable to repeat the measurement 2‐3 times and calculate the average.

Page 10: Communicating Water Level 77 - WYLER AG Switzerland · Manual Communicating Water Level 77 page 3 3. Filling of the water. For longer periods of measuring it is advisable to use distilled

 Manual  Communicating Water Level  77  

 

page 9  

Care Instructions 

Drain the system after use. 

Clean the equipment and tubing with cleaning vinegar (max concentration : 8%) 

Allow to dry and then kept in the wooden box.