comparison of commercial nano lc columns for fast ... · comparison of commercial nano lc columns...

18
1 Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells Tore Vehus 1* , Kristina Erikstad Sæterdal 1 , Stefan Krauss 2 , Elsa Lundanes 1 , Steven Ray Wilson 1 1 Department of Chemistry, University of Oslo, Post Box 1033 Blindern, NO-0315 Oslo, Norway, 2 Unit for Cell Signaling, SFI-CAST Biomedical Innovation Center, Oslo University Hospital, Rikshospitalet, NO-0027 Oslo. *Corresponding author: Tore Vehus [email protected] +47 97 51 81 13 Abstract Targeted proteomics through nanoLC-MS/MS has the potential to measure several proteins in a complex sample in reasonable analysis time. However, to be a contender to existing methods; WB and ELISA, sample preparation has to be minimized, sensitivity high and analysis time short. We compared four commercially available nanoLC columns with various morphologies (Chromolith® (silica monolith), PepMap™ (porous particles), Accucore™ (solid core particles), and PepSwift™ (organic monolith)) for determination of Wnt/beta-catenin related pathway proteins in unfractionated/undepleted cell samples with gradients of 30 minutes. We found that the solid core particle packed column outperformed the other columns in terms of peak capacity (~190 at 50% peak height in 30 minutes). The average LC retention time shift was also least for the two particle based columns (<1 minute), whereas the monoliths were more prone to retention time instability when sample complexity increased. This will affect selectivity of identifications in a complex background, and effectively reduce assay robustness. Of the particle-packed columns, the solid core particle column had the best retention time repeatability in complex samples (1 % RSD). Only small difference in sensitivity was observed, even if the flow-rate was altered in the range of 200 – 600 nL/min. With the solid core particle packed column, Wnt/beta-catenin signal pathway members beta-catenin and GSK3beta were easily identified in non-fractionated samples and could be measured precisely (below 10 % variation) in complex backgrounds using SILAC labeled cell lines as internal standard which corrects for each sample preparation step and LC-MS variation. However, less abundant pathway proteins such as AXIN2 or TNKS2 could not be detected, perhaps calling for more narrow LC columns for fast trace analysis.

Upload: trinhtuyen

Post on 21-Jun-2018

228 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

1

Comparison of commercial nano LC columns for fast, targeted

proteomics of cancer cells

Tore Vehus1*, Kristina Erikstad Sæterdal1, Stefan Krauss2, Elsa Lundanes1, Steven

Ray Wilson1

1Department of Chemistry, University of Oslo, Post Box 1033 Blindern, NO-0315 Oslo, Norway, 2Unit for

Cell Signaling, SFI-CAST Biomedical Innovation Center, Oslo University Hospital, Rikshospitalet, NO-0027

Oslo.

*Corresponding author: Tore Vehus [email protected] +47 97 51 81 13

Abstract

Targeted proteomics through nanoLC-MS/MS has the potential to measure several proteins

in a complex sample in reasonable analysis time. However, to be a contender to existing

methods; WB and ELISA, sample preparation has to be minimized, sensitivity high and

analysis time short.

We compared four commercially available nanoLC columns with various morphologies

(Chromolith® (silica monolith), PepMap™ (porous particles), Accucore™ (solid core particles),

and PepSwift™ (organic monolith)) for determination of Wnt/beta-catenin related pathway

proteins in unfractionated/undepleted cell samples with gradients of 30 minutes.

We found that the solid core particle packed column outperformed the other columns in

terms of peak capacity (~190 at 50% peak height in 30 minutes). The average LC retention

time shift was also least for the two particle based columns (<1 minute), whereas the

monoliths were more prone to retention time instability when sample complexity increased.

This will affect selectivity of identifications in a complex background, and effectively reduce

assay robustness. Of the particle-packed columns, the solid core particle column had the

best retention time repeatability in complex samples (1 % RSD). Only small difference in

sensitivity was observed, even if the flow-rate was altered in the range of 200 – 600 nL/min.

With the solid core particle packed column, Wnt/beta-catenin signal pathway members

beta-catenin and GSK3beta were easily identified in non-fractionated samples and could be

measured precisely (below 10 % variation) in complex backgrounds using SILAC labeled cell

lines as internal standard which corrects for each sample preparation step and LC-MS

variation.

However, less abundant pathway proteins such as AXIN2 or TNKS2 could not be detected,

perhaps calling for more narrow LC columns for fast trace analysis.

Page 2: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

2

Introduction

Nano LC is widely used for comprehensive proteomics to obtain qualitative and quantitative

information of as many proteins as possible. Even though, it is known that in common data-

dependent acquisition, many peptides are not detected due to the vast complexity, dynamic

range and diversity in biological samples (1). Using LC-ESI-MS/MS, typically with

selected/parallel reaction monitoring (SRM or PRM, depending on the instrument), targeted

proteomics aims to selectively identify/quantify specific proteins, preferably within short

analysis times compared to more comprehensive approaches (e.g. 30 minutes instead of 2

hours) (2-4). Therefore, the speed and increased reliability can make targeted approaches

applicable for e.g. clinical settings.

However, the shorter the time used for chromatographing compounds, the greater chance

the target molecules will fail to enter the MS for detection, due to ion suppression by other

(co-eluting) peptides during the electrospray process. This is especially the case if

chromatographed samples are complex (e.g. not pretreated with immunoaffinity-based

isolation). Co-elution can be reduced using high performance LC columns, which produce

narrow peaks and hence increased resolution. A key descriptor of an LC system´s resolution

is its peak capacity (i.e. the number of compounds that can be chromatographically

separated) (5, 6). This number can depend on e.g. the column material/stationary phase,

column length, mobile phase conditions (flow-rate, solvent strength, gradient time etc.),

column packing/polymerization procedure and analytes. In comprehensive proteomics,

correlations between number of identified peptides and peak capacity have been shown (7,

8). Another desired trait of an LC column is a robust/repeatable retention time, which can

support compound identification and allow for narrow SRM/PRM scheduling. Compared to

comprehensive approaches, retention time robustness is arguably more focused upon in

targeted proteomics.

Nano LC columns employed in proteomics are typically defined as having inner diameters

(IDs) ranging from 10 to 100 µm, with mobile phase flow-rates operating between 10 and

500 nL/min. The narrow IDs and low flow-rates can significantly enhance sensitivity/reduce

ion suppression by decreasing radial dilution during chromatography and allowing

production of smaller electrospray droplets, respectively (9, 10).

As the case is with larger bore-LC, nano-columns for reversed phase (RP) LC are by far most

common. Several types of nano RPLC columns are commercially available, e.g. packed with

solid core particles or totally porous particles, silica-based monoliths and organic monoliths

(11-13). Today, totally porous particles for LC are often 2 μm or smaller in diameter, as these

sizes are able to provide higher efficiencies (also at higher speeds) compared to the more

traditional 3.5-5 μm particles (14). However, smaller particles can generate higher

backpressures. Columns packed with solid core particles have comparable speed and

resolution to small porous particles, with lower backpressures, but possibly with lower

loading capacity (15). Monoliths are single piece skeletal structures that are synthesized

Page 3: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

3

within the column, and are mostly organic- or silica-based (16). These structures can also

provide excellent efficiency and increased speeds due to their high permeability (17).

However, these columns are more challenging to prepare reproducibly (18, 19).

All of these column variants can provide high-resolution separations for comprehensive

proteomics (13, 14, 20, 21). But to the authors´ knowledge, the performance of the various

types of nano RPLC columns has not been compared regarding targeted proteomics

experiments (e.g. relatively fast gradients, complex samples and increased focus on LC

robustness/repeatability). We wished to investigate which of four commercially available

columns with the above-mentioned morphologies provided the best chromatographic

performance (peak capacity, peak shape, carry-over, retention time repeatability etc.)

utilizing 30-minute gradients of very complex samples (as opposed to e.g. only assessing

protein standards). Also, we wanted to investigate whether these traits were associated with

success in identifying central proteins of the cancer associated Wnt/beta-catenin-pathway

(22) in complex samples, that also included a universal internal standard solution.

Materials and methods

Sample preparation

Recombinant APC (H00000324-Q01) and AXIN2 (H00008313-Q01) were purchased from

Abnova (Tapei City, Taiwan). Glycogen synthase 3β (GSK3β) were from Life Technologies

(Carlsbad, CA, USA) and beta-catenin (12-537) from Millipore (Billerica, MA, USA). The poly-

ADP-ribosylation polymerization (PARP)-domain of human tankyrase2 (TNKS2) was produced

as described in (23). All amino acid sequences can be found in Supplementary Table 2.

Each standard protein was digested with trypsin (Promega, Madison, WI, USA). Briefly, 10 μg

of each protein was dissolved in 1 mL 8 M urea (Sigma-Aldrich) dissolved in 50 mM Tris-HCl

pH 8.0 (Sigma Aldrich). The samples were reduced in 5 mM dithiothretiol (DTT, Sigma Aldrich)

at 37°C for 30 minutes and alkylated with 15 mM iodoacetamide (IAM) for 15 minutes in the

dark. Trypsin was added to a protein:enzyme ratio of 1:20, and incubated over night at 37°C.

The digested standards were desalted using solid phase extraction (SPE) on RP C18 cartridges

(Bond Elut C18, 100 mg, Agilent, Santa Clara, CA, USA) with water (Millipore) and eluted in 1

mL 80 % acetonitrile (ACN, HiPerSolv CHROMANORM, VWR, Radnor, PA, USA) with 0.1 % FA

(Sigma Aldrich) (v/v) and dried with a SpeedVac (Thermo Fischer Scientific; former Savant,

Waltham, MA, USA). Each standard were reconstituted in 0.1 % (v/v) TFA (Sigma Aldrich) to a

final concentration of 10 μg/mL.

A set of external standard mixtures (ExSMix) containing 1, 0.5, 0.1, 0.05, 0.01, 0.005, 0.001,

0.0005 and 0.0001 μg/mL of each protein standard were prepared by appropriate dilution

with 0.1 % (v/v) TFA.

Page 4: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

4

HCT15 cells (American Type Culture Collection, ATCC, VA, USA) were cultured in RPMI 1640

medium (Life Technologies) supplemented with 10 % fetal bovine serum (FBS, Life

Technologies) and penicillin streptomycin (PS, Life Technologies) and harvested with trypsin

EDTA (Life Technologies) at 80 % confluence. The cells were counted and washed in

phosphate-buffered-saline (PBS, Oslo University Hospital, Oslo, Norway). The proteins were

extracted and digested using the filter-aided sample preparation (FASP) protocol (24). Briefly,

1 million cells were re-suspended in 200 μL lysis buffer, heated for 15 minutes at 70°C and

sonicated for 5 minutes. Debris was removed with centrifugation at 13000 rpm for 10

minutes in a thermostated centrifuge at 20°C (Eppendorf, Hamburg, Germany). The protein

concentration was determined using Bradford Assay (Bradford Quick Start Assay, Bio-Rad,

CA, USA) at absorbance of 595 nm with bovine serum albumin (BSA, Sigma Aldrich) used as

calibration standards. About 100 μg protein was added to 10 kDa 0.5 mL filter devices

(Millipore). The FASP two-step digestion protocol with Trypsin-LysC mix (Promega) (1:20

protein:enzyme ratio) was followed (24). The peptides were desalted using the above-

described SPE-procedure. Peptide concentrations were determined using a NanoDrop2000

instrument (Thermo Fisher Scientific) with absorbance at 205 nm with 31 mg/mL absorption

coefficient. The samples were diluted to a final concentration of 1 mg/mL with 0.1 % (v/v)

TFA.

An internal standard protein solution (ISprot) was prepared by stable isotope labeling of

amino acids in cell culture (SILAC) labeling of HEK293 (ATCC) and HCT15 cells according to

the procedure described by Ong & Mann (25) with 13C615N4-arginine and 13C6-lysine (+10.008

and +6.020 Da, respectively) (Thermo Fisher Scientific) supplemented to RPMI1640 Media

for SILAC acquired from Thermo Fisher Scientific. The labeled cell lines were subsequently

lysed as described above and added to samples prior to protein digestion. Heavy amino acid

incorporation was verified with data-dependent LC-MS/MS analysis of non-labeled cell lines

and labeled cell lines (data not shown).

Treatment of cell lines with G007-LK

The colon carcinoma cells were seeded (100,000 cells/well) in 6-well plates (Nunc™ Cell-

Culture Treated Multidishes, Thermo Fisher Scientific). RPMI1640 were used for HCT15 and

COLO320DM (ATCC) cells with incubation in 5 % CO2, and Leibowitz L-15 medium (Thermo

Fisher Scientific) for SW480 (ATCC) cells with incubation in 0 % CO2. After 24 hours, the

medium was removed and the tankyrase inhibitor G007-LK (23) (dissolved in

dimethylsulfoxide (DMSO, Sigma-Aldrich)) was added to a final concentration of 1 μM in the

cells’ respective medium. An equal volume of DMSO was added as negative control. After 24

hours of incubation the cells were harvested and processed as described above.

Three biological replicates were made and analyzed, with exception of COLO320DM and

SW480 cells, were treated cells were analyzed in duplicates.

LC instrumentation

Page 5: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

5

A NanoLC1000 pump from Thermo Fisher Scientific was used in this study. The mobile phase

A (MP A) contained 0.1 % (v/v) FA in H2O (Optima® LC/MS, Fisher Scientific, part of Thermo

Fisher Scientific) and mobile phase B (MP B) contained 0.1 % (v/v) formic acid (FA) in ACN.

The pre- and analytical columns were coupled through a stainless steel T-piece (Valco, VICI

AG International, Schenkon, Switzerland). Each gradient of 30 minutes was followed with a

linear increase to 95 % MP B for 10 minutes and a 10 to 15 minute hold at 95 % MP B. Each

pre- and analytical column was equilibrated with at least 6 column volumes.

The Acclaim® PepMap RSLC (PepMap™) 75 μm x 20 mm and 50 μm x 150 mm particle

packed pre- and analytical columns, the PepSwift® 200 μm x 5 mm and 100 μm x 250 mm

monolithic poly-styrene divinylbenzene (PS-DVB) pre- and analytical columns, the Accucore®

75 μm x 150 mm solid core particle packed column were from Thermo Fisher Scientific and

the 100 μm x 50 mm and 50 μm x 150 mm Chromolith® Caprod® silica monolithic pre- and

analytical columns were from Merck-Millipore.

MS and MS/MS-parameters

The electrospray voltage was set to 1.8 kV for the 2 μm ID stainless steel emitter (Thermo

Fisher Scientific), the 5 and 8 μm ID New Objective Emitters (New Objective, Woburn, MO,

USA). The Accucore® and PepSwift® were connected to the MS through PicoTip ESI emitters

fitted for the column flow-rate used. The PepMap™ and Chromolith® columns were

connected to stainless steel emitters.

A Q Exactive™ Hybrid Quadrupole-Orbitrap mass spectrometer (Thermo Fisher Scientific)

was used for the entire study. Two main methods were used, full-MS with subsequent data-

dependent MS/MS (ddMSMS) and targeted-MS/MS with selected ions from the proteotypic

peptides chosen. In full-MS, the resolution was set to 140,000 @ m/z 200, automatic gain

control (AGC) to 1,000,000, maximum inject time to 100 ms and scan range m/z 350-1850.

The 10 most intense ions were selected and fragmented using normalized collision energy

(NCE) of 25 % and the MS/MS scans were acquired with; resolution of 70,000, AGC target

value of 100,000, and maximum inject time of 500 ms. Dynamic exclusion was enabled for

20 seconds, and charge states of 1 and >7 were rejected for fragmentation.

For targeted MS/MS, each target was monitored in a retention time window of +/- 4 minutes

relative to the retention time determined by ddMSMS, and either operated in single or

duplexing mode. For single ion fragmentation, the maximum injection time was set to 500

ms, with an AGC target value of 100,000, and a resolution of 140,000. The isolation width

was set to m/z 4.0 and the NCE to 25 eV.

In duplex mode, the resolution was decreased to 35,000, the maximum inject time lowered

to 100 ms and the AGC target value kept unaltered. Each target was fragmented with an

isolation of m/z 2.0 and NCE of 25 eV.

Peptide identification and proteotypic peptide selection

Page 6: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

6

Using the Skyline software (v2.5) (26), a theoretical tryptic digest of each protein standard

was made, and together with data-dependent LC-MS/MS runs for each column, a set of

proteotypic peptides was selected for each protein (Table 1). The peptides selected were

checked for uniqueness against the Uniprot database (27) and only peptides selectively

representing the proteins of interest were chosen as proteotypic peptides. Peptides

containing cysteine, methionine, serine, tyrosine and threonine were avoided if other

possibilities were available.

Data processing

All chromatograms were analyzed with Xcalibur software (v2.1, Thermo Fisher Scientific) and

Proteome Discoverer (v1.4, Thermo Fisher Scientific) with the sequest algorithm were used

for searching against the protein sequences found in Supplementary Table 1. For

identification in ddMSMS, maximum two missed cleavages, maximum 10 ppm precursor and

0.6 Da mass tolerance, respectively, were allowed, combined with a false discovery rate of

0.01. Carbaiodomethylation was set as a static modification and oxidation of methionine as

dynamic. Targeted MS/MS data acquired at resolutions of 140,000 and 35,000 were

extracted using 7 and 25 ppm, respectively. Peak widths, heights and areas were measured

manually, and peak capacity (nc) at half peak height was calculated according to Equation 1,

𝒏𝒄 =𝒕𝑹,𝒍𝒂𝒔𝒕−𝒕𝑹,𝒇𝒊𝒓𝒔𝒕

𝒘𝟎.𝟓̅̅ ̅̅ ̅̅ Equation 1

where 𝒘𝟎.𝟓̅̅ ̅̅ ̅̅ is average peak width at half peak height, tR,first and tR,last are the retention times

for the first and last eluting peak, respectively.

For statistical analysis, F-test and two-sided student t-test were used. LC-MS/MS raw data is

available on the following link http://www.mn.uio.no/kjemi/supplementary-

data/bach/comparison-of-commercial-nano-lc-columns-for-fast-/

Results

Framework and plan of study

The RP columns investigated were; Chromolith® CapRod (C18 silica monolith), PepMap™ (C18

porous particles), Accucore™ (C18 solid core particles) and PepSwift™ (PS-DVB organic

monolith). The columns had to be compatible with a (commonly used) nano EASY-nLC 1000

pump (Thermo Fisher Scientific), the pump available to us at the time of study (and not

equipped with a column oven); this excluded the testing of instrument specific units as the

Agilent CHiP systems and Waters nanoAcquity column systems. The columns had some

variations in dimensions (50-100 µm IDs and 15-25 cm length); identical dimensions were

not commercially available for all columns. The linear separation gradient was set to 30

minutes (=tG), as this was considered to be an acceptable compromise between speed and

risk of ion suppression in tryptic digest of un-fractioned human cell lysate (containing

between 10,000 – 30,000 protein species). For each column, the gradient was adjusted so

Page 7: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

7

that the last eluting peptide of interest eluted at tG, allowing for a fair comparison between

the chromatographic properties. A standard mobile phase consisting of water, 0.1 % FA, and

ACN were used for all columns. Proteins studied were AXIN2, beta-catenin, GSK3beta and

TNKS2. Additionally, the C-terminal of APC was included, and served as a negative control for

protein identification in the APC-mutated colon carcinoma cell lines. Although this is a

limited number of analytes, these proteins represent a central selection of Wnt-related

proteins and relevant range (for signaling pathway analysis) of endogenous concentrations,

where the low-abundant concentrations may not be detected in ddMSMS) (28). Also, these

analytes are also available for use as external standards (vital for several of the

investigations performed here, but also for “tier 1” level targeted proteomics in general (29)).

The proteotypic peptides generated from these proteins had molar mass of 0.8-1.9 kDa, and

GRAVY index values of -2.32 - 1.24, which are arguably representative for tryptic peptides in

general. The amounts of complex samples loaded onto each system were between 0.5-1 μg,

amounts which are common in proteomics experiments (30, 31) and well below the pre-

column capacities reported by the manufacturers. The LC-MS/MS system set-up was

considered “healthy” as 40 proteins/separation minute could be identified in comprehensive

mode (data not shown).

The main steps of the study were: Optimization of peak capacity for each column and

evaluation of other basic LC traits; Evaluation of repeatability/robustness by comparing

system behavior/performance between that of an external standard mix of the targeted

proteins (“ExSMix”) and lysates spiked with the standards; Investigation of which target

peptides (endogenously present at trace –to moderate levels) were possible to detect in an

unspiked sample; Evaluation of relative quantification by LC-MS/MS with isotopically labeled

internal standards and comparison with established Western blotting (WB) methods.

Columns/hardware were installed by personnel with considerable experience in low nano

flow-rate instrumentation. Just one column per column type were tested; however, columns

used for experiments performed satisfactorily with respect to the manufacturers´

specifications upon installment (or were replaced if they did not; this was the case for one

column), and batch-to-batch issues were not explored.

Chromatographic performance

To compare the nano LC columns´ performance for 30 minute gradients, the flow-rate and

gradient composition were optimized for each column with regards to peak capacity

according to the procedure described by Wang et al. (6). The definition used for peak

capacity is found in Eq.1 (Materials and Methods) A key point is to fully exploit the

separation window, ensuring that the most hydrophobic analytes elute at the end of the

gradient (tfinal peak = tG). The sample was the ExSMix (see Materials and Methods).

Table 1 shows the peak capacity and peak asymmetry for each column set-up, as well as the

enabling solvent conditions. (Table 1 here) The highest peak capacity was obtained with the

solid core particle packed column set-up, with an average peak capacity close to 190 (Figure

1); approximately 1.5 times larger compared to the second-best performing column (the

Page 8: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

8

silica monolith, peak capacity = 130) with our conditions. The solid core particle packed

based column also provided the least peak tailing and the narrowest peaks of the columns

included in this study, with an average asymmetry of 1.1 and a base peak width of 10

seconds. (Figure 1 here) The two monolithic columns tested displayed more peak tailing

compared to the particle-based columns (1.4-1.5 vs 1.1-1.3).

Table 1 - Optimal flow-rate, gradient composition, peak capacity, peak width and asymmetry based on data-

dependent MS/MS scans of separated 1 ng ExSMix from recombinant APC, AXIN2, beta-catenin, GSK3beta and

TNKS2 for the four column set-ups. Retention time RSD values are based on 3 technical replicates.

Column

Precolumn Analytical column Precolumn/analytical

column material

Optimal flow-rate

(nL/min/ linear

velocity in

cm/second

Gradient end

(% (v/v) ACN)

Retention

time

variation

(RSD, %)

Peak

capacity

at half peak

height

Peak width

at half peak

height

(sec)

Asymmetry

ID

(μm)

Length

(mm)

ID

(μm)

Length

(mm)

Chromolith®C

apRod® 100 50 50 150

C18 silica monolithic/

C18 silica monolithic 200/ 0.8 36 0.3 130 19.8 1.4

PepMap™ 75 20 50 150 C18, 3µm, 100Å particles/

C18, 2µm, 100Å particles 200/ 0.9 36 0.4 106 16.2 1.1

Accucore™ 75 20 75 150

C18, 3µm, 100Å particles/

C18, 2.6µm, 80 Å solid core

particles

300/ 0.7 36 0.2 189 10.2 1.1

PepSwift™ 200 50 100 250 PS-DVB monolithic/

PS-DVB monolithic 500/ 0.5 20 1.0 100 16.2 1.5

Page 9: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

9

Figure 1 – A : EIC of peptides from 1 ng standard tryptic mixture chromatographed using the 75 µm x 15 cm

Accucore™ solid core particle packed column set-up. B : EIC of HSSPGTVAAR with peak widths at 10 % and 50 %

peak height and overall peak capacity nC, average peak width ( 𝑤0.5̅̅ ̅̅ ̅ ) and average asymmetry factor (𝐴𝑆̅̅ ̅).

Supplementary Figures 1 – 4 demonstrate each columns’ response in peak capacity and

asymmetry to flow-rate, showing that the silica monolith could be operated in a larger span

of flow-rates without loss in chromatographic performance (i.e. peak capacity of 130 +/- 5

from 200 – 500 nL/min), in contrast to the other columns which had more a more narrow

range of optimal flow-rates. The optimal linear gradient composition was the same (36 %

solvent B at tG ≈ 30 minutes) for the columns with C18 stationary phases. For the less

hydrophobic PS-DVB column, the end gradient composition needed to be significantly lower

Page 10: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

10

(20 % solvent B at tG) to ensure tfinal ≈ tG. With the flow-rates investigated, no major effects

on signal intensity were observed (2-fold changes or less, Supplementary Figures 5 and 6).

The correlation between concentration (1 – 0.001 ng/µL) and peak area (not corrected with

an internal standard (IS), and measured with proteotypic peptides with a range of

hydrophobicity, see below) was R2>0.97, implying that the trapping efficiency of each pre-

column was satisfactory (Supplementary Figures 7-11 and Supplementary Table 3). The

average RSDs for the tR in the ExSMix were 0.3, 0.2, 0.4, and 1.0 % for the Chromolith®,

Accucore™, PepMap™, and PepSwift™, respectively. Carry-over was negligible (< limit of

detection, LOD) for all columns regarding the analytes and concentrations investigated in

this study.

Chromatography of unfractionated cell lysates in targeted MS/MS mode

Using the optimized LC conditions and data-dependent MS/MS acquisition, 17 proteotypic

peptides (Table 2) of proteins associated with the beta-catenin degradasome were selected

for further targeted MS/MS studies. Table 2 – Peptide identification for each chromatographic column investigated. 1) is 0.5 ng ExSMix, 2) is 0.5 ng

ExSMix spiked into 1 µg HCT15 tryptic cell lysate and 3) is 1 μg HCT15 tryptic cell lysate (all in triplicates). Green

equals identified, red not identified and grey not observed with ddMSMS. MS/MS extraction was done with 7

ppm mass accuracy. Minimum 3 transitions were required for positive identification and not more than 0.5

minutes shift in tR between 2 and 3 were allowed for positive identification. The effect of sample

complexity-related retention time robustness (important for e.g. MS/MS scheduling/aiding

selectivity) was examined by comparing retention times of the simple ExSMix proteotypic

peptides with that of the ExSMix spiked to tryptic HCT15 cell lysate (serving as a complex

matrix). The median change in retention time of the peptides in the ExSMix and the ExSMix

spiked cell sample was significantly larger for the monoliths (-2.2 and 1.7 min, for the organic

and silica monolith, respectively) compared to the particle packed columns (-0.4 and -1.0

min, for the totally porous and solid core particle packed columns, respectively). The average

retention time RSDs of complex samples were 0.8 (1.9), 0.6 (1.0), 2.6 (2.8), and 1.3 (2.3) %

for the Chromolith®, Accucore™, PepMap™, and PepSwift™, respectively (largest variation in

parenthesis). (Table 2 here) Figure 2 shows extracted ion chromatograms (EIC) of the

representative beta-catenin peptide LLNDEDQVVVNK in each sample chromatographed on

the four column set-ups. The peak widths were not changed as the sample complexity

increased (between ExSMix and ExSMix added to the cell lysate), indicating no sign of

column overload.

Page 11: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

11

Table 2 – Peptide identification for each chromatographic column investigated. 1) is 0.5 ng ExSMix, 2) is 0.5 ng

ExSMix spiked into 1 µg HCT15 tryptic cell lysate and 3) is 1 μg HCT15 tryptic cell lysate (all in triplicates).

Green equals identified, red not identified and grey not observed with ddMSMS. MS/MS extraction was done

with 7 ppm mass accuracy. Minimum 3 transitions were required for positive identification and not more than

0.5 minutes shift in tR between 2 and 3 were allowed for positive identification.

Column Chromolith® PepMap™ Accucore™ PepSwift™

Protein Peptide 1 2 3 1 2 3 1 2 3 1 2 3

APC

HSGSYLVTSV

HSSPSGTVAAR

VTPFNYNPSPR

AXIN2

AQSLTLGHFK

ILGKVER

ILGKVERID

Beta-

catenin

LLNDEDQVVVNK

HAVVNLINYQDDAELATR

NEGVATYAAAVLFR

ATVGLIR

GSK3beta

DIKPQNLLLDPDTAVLK

DSSGTGHFTSGVR

LLEYTPTAR

VIGNGSFGVVYQAK

TNKS2

DGGHAGGIFNR

EVSEENHNHANER

SFLQFSAMK

ID count

11 5 6 3 6 5 9 5

Median shift in retention time (min)

between ExSMix and ExSMix in 1 μg

HCT15 cell lysate

-1.7 -0.5 -1.2 -2.2

Average retention time variation (RSD

(%)) in lysates (2 and 3) 0.8 2.6 1.0 2.5

Page 12: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

12

Figure 2 – EIC of LLNDEDQVVVNK corresponding to beta-catenin in 0.5 ng ExSMix, 0.5 ng ExSMix spiked in 1 µg

tryptic peptide extract from HCT15, and 1 µg tryptic peptide extract from HCT15 for each of the

chromatographic systems investigated. (Chromatographic conditions and MS-settings as described in Materials

and Methods).

The proteotypic peptides were subsequently searched for in an unspiked cell sample. For

identification minimum three intense/descriptive MS/MS transitions were required. An

additional criterion was that the retention time variation was maximum ±0.5 minutes

relative to that of the ExSMix in spiked sample (matrix matching, see Figure 3) to ensure very

confident identification.

Figure 3 – EIC of the high-abundance peptide (HAVVNLINYQDDAELATR – beta-catenin), with transitions from

m/z = 1021.51898 to 1) m/z = 775.3958, 2) m/z = 1181.545, 3) m/z = 1295.588 in spiked and unspiked tryptic

digested HCT15 extract. Grey background equals positive identification. Samples were chromatographed on the

Accucore™ column set-up with parameters as described in Materials and Methods.

Page 13: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

13

Beta-catenin (downstream target of Wnt/signaling) and GSK3beta (a kinase crucial for N-

terminal phosphorylation of beta-catenin that leads to its degradation) (~60 ng/μg and ~0.8

ng/μg sample), respectively) were clearly identifiable with all the columns with the above-

mentioned criteria. With the criteria employed the two other trace proteins (AXIN2, TNKS2)

were however not identified. Notably, using just two intense/descriptive MS/MS transitions

would cause false positives in our assay; For example, the C-terminal HSSPSGTVAAR peptide

of intact APC (not present in HCT15 colon cancer cells, due to a premature stop codon (32))

was falsely identified with otherwise same criteria (data not shown).

Quantification with isotopically labeled internal standard – impact on LC-MS/MS

performance

For quantification in complex samples, isotopically labeled ISs are often added, either as

peptides or proteins (e.g. AQUA peptides (33) or stable isotope labeling by amino acids in

cell culture (SILAC) (25)). In this study, SILAC was used to label two cell lines which were

pooled and used as an internal standard protein solution (ISprot) in following experiments

(preparation and workflow described in Materials and Methods and Supplementary Figure

12). In contrast to e.g. spiking with single IS peptides, the labeled mix is a “universal” IS, also

provides correction for protein digestion, SPE clean-up and MS-response. Also the approach

is simpler than producing recombinant labeled IS proteins (34). The use of complex ISprot

affects the LC-MS/MS performance in two ways; sample load and MS/MS complexity.

In all experiments, the sample load (in mass) was kept below 1 μg. As the ISprot was added to

an approximate ratio of 1:1 to the sample, this means a lower load of endogenous peptides,

making detection of the low-abundant proteins less achievable.

When comparing the columns, the resolution was kept at 140,000 to have the highest

possible selectivity. This setting combined with an injection time of 500 ms lead to a MS

cycling speed of 14.4 seconds. As the MS/MS complexity doubled, the MS/MS was set to

fragment both labeled and non-labeled peptides simultaneously (duplexing). Therefore, the

resolution and inject time were lowered to reach a higher scan rate (2.4 seconds cycling

speed) (Supplementary Figure 13A). Using a sample containing ExSMix spiked into the ISprot,

no significant difference (p = 0.6) between quantification based on peak area or fragment

intensity at a given time was observed (Supplementary Figure 13B-D). Hence, fragment

intensity was used to measure response in further quantification experiments. Practically,

this allows reliable quantification with fewer points per peak (i.e below 8 points for correct

area measurement, as recommended in (35)).

To see whether the solid core column would allow detecting changes in proteotypic peptide

levels of beta-catenin and GSK3beta, we selected the colon carcinoma cell lines HCT15,

COLO320 and SW480 cells. Each were treated with a selective tankyrase inhibitor (G007-LK

(23)) and subjected to quantification with LC-MS/MS and WB. Beta-catenin (the subject of

the AXIN2/tankyrase2/APC/GSKbeta-containing destruction complex) could be relatively

quantified with excellent precision (RSD 8 %) in these solutions with double complexity

Page 14: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

14

(sample +ISprot) well within 20 minutes: Reduction of beta-catenin following treatment with

the selective tankyrase inhibitor G007-LK was observed in SW480 cells and correlated with

results obtained with an established WB protocol (see Figure 4, Supplementary Figure 15 for

raw WB and (36)).

Figure 4 – A : EICs of LLNDEDQVVVNK (blue color, endogenous) and LLNDEDQVVVNk (red color, internal

standard) in DMSO and in 1 µM G007-LK treated SW480 cells, respectively, analyzed by LC-MS/MS on the

Accucore™ column using standard gradient conditions and MS/MS detection in duplexing mode with resolution

at 35,000 and maximum injection time at 100 ms. B – WB of total beta-catenin and actin in DMSO and 1 µM

G007-LK treated SW480 cells (Procedure in Supplementary Materials and Methods). C – Relative quantification

based on LC-MS analysis and WB analysis (Procedure in Materials and Methods and Supplementary Figure 12).

For levels of GSK3beta and beta-catenin in the other cell lines analyzed, see Supplementary

Figure 14.

Discussion

Chromatographic performance

Using 30 minute solvent gradients (for efficient analysis), the solid core particle packed

nano-column provided the highest peak capacity and hence resolution, with the silica

monolith at a clear 2nd place. Similar results has been observed with larger-bore columns

Page 15: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

15

(37), but it was not given that this would be the case for nano-scale columns, as successful

packing/polymerization can be dependent on column diameter. For example, solid core

particles packed in 2.1 mm ID columns have previously not been able to match the efficiency

of comparable 4.6 mm ID columns (38), while monolith columns are easier to prepare in

capillary/nano-format (39).

Insignificant differences were observed between the four columns regarding sensitivity.

Others have shown clear increases in sensitivity with reduced flow-rates in nano-LC (40). In

our hands, with the flow-rates and column dimensions used (200-500 nL/min, 50-100 μm ID),

the signal intensity was in effect not dramatically affected, which may be due to recent

improvements in the needle geometries and interface constructions (9, 41).

The column´s retention time robustness (of importance for strict identification and MS/MS

scheduling) differed significantly, with the particle packed columns having the least retention

time shift as function of sample complexity, with the core shell column having the best

within-complex sample repeatability. The retention time robustness of the solid core particle

packed column was somewhat surprising considering its lower surface. We believe this

factor was less important due to the use of a full porous particle packed pre-column. In

summary: there are considerable differences in the performance of commercial nano LC

columns, and the solid core particle packed column was considered to best within the

framework of this study (which was meant to reflect targeted proteomics applications of

complex samples in general).The other morphologies may of course excel under other

circumstances. For example, the PS-DVB material is very suitable for larger molecules, e.g.

intact proteins (42) or perhaps larger peptides from lys C protease. In addition, the PS-DVB

phase is excellent for very narrow nano-LC (open tubular format), with performance

matching or surpassing the columns investigated here (43, 44). Open tubular LC columns are

however not yet commercially available.

Identification and quantification in complex samples

Targeted LC-MS/MS is an alternative to common approaches such as WB and enzyme-linked

immunosorbant assays (ELISA) (2) , which are far more used in e.g. clinical settings. It is

claimed that selectivity is higher with LC-MS/MS-based measurements, but this relies on

three major factors; 1) high retention time repeatability, 2) use of proper internal standards

and 3) good MS/MS-spectral quality (45, 46). Our results show that 1) is obtainable with

nano LC columns (and not just larger ID columns), but differs between products. Regarding

2), using a SILAC mixture as a “universal” internal standard is advantageous as it corrects for

variations in all steps of sample treatment (e.g. trypsination and ion suppression). Although

this is a relatively simple approach, it adds to complexity. As a result, faster scan times were

necessary, requiring a reduction of MS resolution. This points to 1) being of added

importance, as retention time repeatability thus becomes increasingly important for

compound identification (to compensate for lowered mass accuracy). Regarding 3), it was

clear that using just two MS/MS transitions to attempt identifications could result in false

Page 16: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

16

positives. Hence, we believe that three MS/MS transitions is an absolute minimum for

identification, which also has been used in the study by Abbatiello et al. (47).

Beta-catenin and GSK3beta could be quantified with corresponding ISprot normalization, with

variation less than 10 % (well within the FDA guidelines limit of 20 % RSD (48)), which points

to nano LC-MS/MS as a potentially high precision tool for fast analysis of complex samples.

However, the difficulty of measuring lower abundant analytes call for increased sensitivity.

One potential path is further reducing the ID of the nano LC column. For example, 10 μm ID

open tubular columns have recently been demonstrated to identify axin1, in a hyphenated

set-up including on-line enzymatic digestion (49). However, such equipment is not

commercially available, and is rather complex. Perhaps compromises would be possible for

commercial implementation, e.g. 30 μm packed columns (50), and use of new core-shell

particles (http://www.sciencedirect.com/science/article/pii/S0021967314007304).

Several other nano-LC columns are available, and due to practical reasons all of these were

not investigated in this study. However, our results may also serve as a source for

comparison to other columns in “real sample” analysis (and not just assessment with clean

standards).

References

1. Michalski A, Cox J, Mann M. More than 100,000 detectable peptide species elute in single shotgun proteomics runs but the majority is inaccessible to data-dependent LC-MS/MS. J Proteome Res. 2011;10(4):1785-93. 2. Picotti P, Aebersold R. Selected reaction monitoring-based proteomics: workflows, potential, pitfalls and future directions. Nat Meth. 2012;9(6):555-66. 3. Liebler DC, Zimmerman LJ. Targeted Quantitation of Proteins by Mass Spectrometry. Biochemistry. 2013;52(22):3797-806. 4. Chen Y, Gruidl M, Remily-Wood E, Liu RZ, Eschrich S, Lloyd M, et al. Quantification of beta-catenin signaling components in colon cancer cell lines, tissue sections, and microdissected tumor cells using reaction monitoring mass spectrometry. J Proteome Res. 2010;9(8):4215-27. 5. Neue UD. Theory of peak capacity in gradient elution. J Chromatogr A. 2005;1079(1–2):153-61. 6. Wang X, Stoll DR, Schellinger AP, Carr PW. Peak Capacity Optimization of Peptide Separations in Reversed-Phase Gradient Elution Chromatography:  Fixed Column Format. Anal Chem. 2006;78(10):3406-16. 7. Kocher T, Swart R, Mechtler K. Ultra-high-pressure RPLC hyphenated to an LTQ-Orbitrap Velos reveals a linear relation between peak capacity and number of identified peptides. Anal Chem. 2011;83(7):2699-704. 8. Hsieh EJ, Bereman MS, Durand S, Valaskovic GA, MacCoss MJ. Effects of Column and Gradient Lengths on Peak Capacity and Peptide Identification in nanoflow LC-MS/MS of Complex Proteomic Samples. J Am Soc Mass Spectrom. 2013;24(1):148-53. 9. Reschke BR, Timperman AT. A study of electrospray ionization emitters with differing geometries with respect to flow rate and electrospray voltage. J Am Soc Mass Spectrom. 2011;22(12):2115-24. 10. Schmidt A, Karas M, Dulcks T. Effect of different solution flow rates on analyte ion signals in nano-ESI MS, or: when does ESI turn into nano-ESI? J Am Soc Mass Spectrom. 2003;14(5):492-500. 11. Marchetti N, Cavazzini A, Gritti F, Guiochon G. Gradient elution separation and peak capacity of columns packed with porous shell particles. J Chromatogr A. 2007;1163(1–2):203-11. 12. Unger KK, Skudas R, Schulte MM. Particle packed columns and monolithic columns in high-performance liquid chromatography-comparison and critical appraisal. J Chromatogr A. 2008;1184(1–2):393-415. 13. Eeltink S, Dolman S, Detobel F, Swart R, Ursem M, Schoenmakers PJ. High-efficiency liquid chromatography–mass spectrometry separations with 50 mm, 250 mm, and 1 m long polymer-based

Page 17: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

17

monolithic capillary columns for the characterization of complex proteolytic digests. J Chromatogr A. 2010;1217(43):6610-5. 14. Novakova L, Vaast A, Stassen C, Broeckhoven K, De Pra M, Swart R, et al. High-resolution peptide separations using nano-LC at ultra-high pressure. J Sep Sci. 2013;36(7):1192-9. 15. Fekete S, Fekete J. The Potential of Shell Particles in Fast Liquid Chromatography. Ultra-High Performance Liquid Chromatography and its Applications: John Wiley & Sons, Inc.; 2013. p. 133-67. 16. Rozenbrand J, van Bennekom WP. Silica-based and organic monolithic capillary columns for LC: Recent trends in proteomics. J Sep Sci. 2011;34(16-17):1934-44. 17. Dolman S, Eeltink S, Vaast A, Pelzing M. Investigation of carryover of peptides in nano-liquid chromatography/mass spectrometry using packed and monolithic capillary columns. J Chromatogr B Analyt Technol Biomed Life Sci. 2013;912:56-63. 18. Guiochon G. Monolithic columns in high-performance liquid chromatography. J Chromatogr A. 2007;1168(1-2):101-68 19. Nunez O, Nakanishi K, Tanaka N. Preparation of monolithic silica columns for high-performance liquid chromatography. J Chromatogr A. 2008;1191(1-2):231-52. 20. Zhou F, Lu Y, Ficarro SB, Webber JT, Marto JA. Nanoflow Low Pressure High Peak Capacity Single Dimension LC-MS/MS Platform for High-Throughput, In-Depth Analysis of Mammalian Proteomes. Anal Chem. 2012;84(11):5133-9. 21. Rogeberg M, Wilson SR, Malerod H, Lundanes E, Tanaka N, Greibrokk T. High efficiency, high temperature separations on silica based monolithic columns. J Chromatogr A. 2011;1218(41):7281-8. 22. MacDonald BT, Tamai K, He X. Wnt/beta-catenin signaling: components, mechanisms, and diseases. Dev Cell. 2009;17(1):9-26. 23. Voronkov A, Holsworth DD, Waaler J, Wilson SR, Ekblad B, Perdreau-Dahl H, et al. Structural basis and SAR for G007-LK, a lead stage 1,2,4-triazole based specific tankyrase 1/2 inhibitor. J Med Chem. 2013;56(7):3012-23. 24. Wisniewski JR, Zougman A, Nagaraj N, Mann M. Universal sample preparation method for proteome analysis. Nat Meth. 2009;6(5):359-62. 25. Ong SE, Mann M. A practical recipe for stable isotope labeling by amino acids in cell culture (SILAC). Nat Protoc. 2006;1(6):2650-60. 26. MacLean B, Tomazela DM, Shulman N, Chambers M, Finney GL, Frewen B, et al. Skyline: an open source document editor for creating and analyzing targeted proteomics experiments. Bioinformatics. 2010;26(7):966-8. 27. Consortium TU. Activities at the Universal Protein Resource (UniProt). Nucleic Acids Res. 2014;42(D1):D191-D8. 28. Wang M, Weiss M, Simonovic M, Haertinger G, Schrimpf SP, Hengartner MO, et al. PaxDb, a Database of Protein Abundance Averages Across All Three Domains of Life. Mol Cell Proteomics. 2012;11(8):492-500. 29. Carr SA, Abbatiello SE, Ackermann BL, Borchers C, Domon B, Deutsch EW, et al. Targeted Peptide Measurements in Biology and Medicine: Best Practices for Mass Spectrometry-based Assay Development Using a Fit-for-Purpose Approach. Mol Cell Proteomics. 2014;13(3):907-17. 30. Peterson AC, Russell JD, Bailey DJ, Westphall MS, Coon JJ. Parallel reaction monitoring for high resolution and high mass accuracy quantitative, targeted proteomics. Mol Cell Proteomics. 2012;11(11):1475-88. 31. Sun L, Zhu G, Dovichi NJ. Comparison of the LTQ-Orbitrap Velos and the Q-Exactive for proteomic analysis of 1-1000 ng RAW 264.7 cell lysate digests. Rapid Commun Mass Spectrom. 2013;27(1):157-62. 32. Ikediobi ON, Davies H, Bignell G, Edkins S, Stevens C, O'Meara S, et al. Mutation analysis of 24 known cancer genes in the NCI-60 cell line set. Mol Cancer Ther. 2006;5(11):2606-12. 33. Gerber SA, Rush J, Stemman O, Kirschner MW, Gygi SP. Absolute quantification of proteins and phosphoproteins from cell lysates by tandem MS. PNAS. 2003;100(12):6940-5. 34. Edfors F, Bostrom T, Forsstrom B, Zeiler M, Johansson H, Lundberg E, et al. Immunoproteomics using polyclonal antibodies and stable isotope-labeled affinity-purified recombinant proteins. Mol Cell Proteomics. 2014;13(6):1611-24. 35. Lange V, Picotti P, Domon B, Aebersold R. Selected reaction monitoring for quantitative proteomics: a tutorial. Mol Syst Biol. 2008;4:222-35. 36. Lau T, Chan E, Callow M, Waaler J, Boggs J, Blake RA, et al. A Novel Tankyrase Small-Molecule Inhibitor Suppresses APC Mutation–Driven Colorectal Tumor Growth. Cancer Res. 2013;73(10):3132-44.

Page 18: Comparison of commercial nano LC columns for fast ... · Comparison of commercial nano LC columns for fast, targeted proteomics of cancer cells ... SFI-CAST Biomedical Innovation

18

37. Beneito-Cambra M, Herrero-Martínez JM, Ramis-Ramos G, Lindner W, Lämmerhofer M. Comparison of monolithic and microparticulate columns for reversed-phase liquid chromatography of tryptic digests of industrial enzymes in cleaning products. J Chromatogr A. 2011;1218(41):7275-80. 38. Gritti F, Guiochon G. Repeatability of the efficiency of columns packed with sub-3μm core-shell particles: Part II. 2.7 μm Halo-ES-Peptide-C18 particles in 4.6mm and 2.1mm×100mm column formats. J Chromatogr A. 2012;1252(0):31-44. 39. Ishizuka N, Kobayashi H, Minakuchi H, Nakanishi K, Hirao K, Hosoya K, et al. Monolithic silica columns for high-efficiency separations by high-performance liquid chromatography. J Chromatogr A. 2002;960(1–2):85-96. 40. Shen Y, Zhao R, Berger SJ, Anderson GA, Rodriguez N, Smith RD. High-Efficiency Nanoscale Liquid Chromatography Coupled On-Line with Mass Spectrometry Using Nanoelectrospray Ionization for Proteomics. Anal Chem. 2002;74(16):4235-49. 41. Liu Q, Cobb JS, Johnson JL, Wang Q, Agar JN. Performance comparisons of nano-LC systems, electrospray sources and LC-MS-MS platforms. J Chromatogr Sci. 2014;52(2):120-7. 42. Rogeberg M, Wilson SR, Greibrokk T, Lundanes E. Separation of intact proteins on porous layer open tubular (PLOT) columns. J Chromatogr A. 2010;1217(17):2782-6. 43. Rogeberg M, Vehus T, Grutle L, Greibrokk T, Wilson SR, Lundanes E. Separation optimization of long porous-layer open-tubular columns for nano-LC-MS of limited proteomic samples. J Sep Sci. 2013;36(17):2838-47. 44. Luo Q, Yue G, Valaskovic GA, Gu Y, Wu SL, Karger BL. On-line 1D and 2D porous layer open tubular/LC-ESI-MS using 10-microm-i.d. poly(styrene-divinylbenzene) columns for ultrasensitive proteomic analysis. Anal Chem. 2007;79(16):6174-81. 45. van den Broek I, Romijn FPHTM, Smit NPM, van der Laarse A, Drijfhout JW, van der Burgt YEM, et al. Quantifying Protein Measurands by Peptide Measurements: Where Do Errors Arise? J Proteome Res. 2014. 46. Gallien S, Duriez E, Demeure K, Domon B. Selectivity of LC-MS/MS analysis: Implication for proteomics experiments. J Proteomics. 2013;81(0):148-58. 47. Abbatiello SE, Mani DR, Schilling B, Maclean B, Zimmerman LJ, Feng X, et al. Design, implementation and multisite evaluation of a system suitability protocol for the quantitative assessment of instrument performance in liquid chromatography-multiple reaction monitoring-MS (LC-MRM-MS). Mol Cell Proteomics. 2013;12(9):2623-39. 48. Administration Fad. Guidance for Industry Bioanalytical Method Validation. Rockville, MD: US Deparment of Health and Human Services, Food and Drug Administration, Center for Drug Evaluation and Re. 2001:25. 49. Hustoft HK, Vehus T, Brandtzaeg OK, Krauss S, Greibrokk T, Wilson SR, et al. Open Tubular Lab-On-Column/Mass Spectrometry for Targeted Proteomics of Nanogram Sample Amounts. PLoS ONE. 2014;9(9):e106881. 50. Kocher T, Pichler P, Pra MD, Rieux L, Swart R, Mechtler K. Development and performance evaluation of an ultra-low flow nano liquid chromatography-tandem mass spectrometry set-up. Proteomics. 2014.