76 Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0) 881-1 83-0 · Fax: +49 (0) 881-62 53 9 Conductivity Meters Conductivity Meters Handheld conductivity meters Portable conductivity meters Laboratory conductivity meters Conductivity cells Accessories The conductivity is a cumulative para- meter for the ionic concentration of a solution. The more salt, acid or alkali in a solution, the greater its conductivity. The unit of conductivity is S/cm. The scale for aqueous solutions begins with pure water at a conductivity of 0.05 μS/cm (25°C). Naturally occurring waters such as drinking water or surface water have a conductivity in the range 100 - 1000 μS/cm. The upper end of the scale is reached by some bases such as potassium hydroxide solutions with a conductivity of just over 1000 mS/cm. In practice conductivity measurements are used for applications such as moni- toring plants in which ultrapure water is produced and determining the salinity of sea water. Conductivity is measured by making a measurement of the electrical resistance. The simplest kind of measuring cell used consists of two similar electrodes. An alternating voltage applied to one of the electrodes causes the ions in the solution to migrate towards the electrodes. The more ions in the solution, the greater the current which flows between the electrodes. The instrument measures the current and uses Ohm's law to calculate first the conductance of the solution and then - by taken the cell data into account - the conductivity. Electrical Conductivity Further information on this subject can be requested from WTW: Application reports Offprints Literature Feedwater for high-pressure boilers Fully deionized water (ion exchanger) Deionized water Drinking water Brackish water, sea water Industrial process water Concentrated acids and bases Wastewater Surface water μS/cm mS/cm 0,1 1 10 100 1 10 100 1000 Conductivity of aqueous solutions

Upload: trinhthuy

Post on 03-Apr-2019

216 views

Category:

Documents


0 download

TRANSCRIPT

76

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

ConductivityMetersConductivity

MetersHandheld conductivity meters

Portable conductivity meters

Laboratory conductivity meters

Conductivity cells

Accessories

The conductivity is a cumulative para-meter for the ionic concentration of asolution. The more salt, acid or alkali ina solution, the greater its conductivity.The unit of conductivity is S/cm. Thescale for aqueous solutions begins withpure water at a conductivity of0.05 µS/cm (25°C). Naturally occurringwaters such as drinking water or surfacewater have a conductivity in the range100 - 1000 µS/cm. The upper end of thescale is reached by some bases such aspotassium hydroxide solutions with aconductivity of just over 1000 mS/cm.

In practice conductivity measurementsare used for applications such as moni-toring plants in which ultrapure water isproduced and determining the salinity ofsea water.

Conductivity is measured by making ameasurement of the electrical resistance. The simplest kind of measuring cell usedconsists of two similar electrodes. An alternating voltage applied to one of theelectrodes causes the ions in the solution to migrate towards the electrodes. Themore ions in the solution, the greater the current which flows between theelectrodes. The instrument measures the current and uses Ohm's law to calculatefirst the conductance of the solution and then - by taken the cell data into account- the conductivity.

Electrical Conductivity

Further information on this subject canbe requested from WTW:

Application reportsOffprints

Literature

Feedwater for high-pressure boilers

Fully deionized water (ion exchanger)

Deionized water

Drinking water

Brackish water, sea water

Industrial process water

Concentrated acids and bases

Wastewater

Surface water

µS/cm mS/cm0,1 1 10 100 1 10 100 1000

Conductivity of aqueous solutions

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

77

Conductivity Meters

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Handheld conductivity meters

Portable conductivity meters

Laboratory conductivity meters

Model from page

Handheld Quality in a set of three 78conductivity meters - Simple to use

- Proven with memory function- All-rounder with interface

Portable Classic 82conductivity meters robust, shockproof, waterproof, portable,

conductivity measurement at depths down to 100 m

Laboratory inoLab® – top quality in 3 stages 84conductivity meters - Simple and reliable

- Compact and communicative- Flexible and powerful

Conductivity cells 88 Standard cells 89 Special cells 89

Conductivity meter accessories Flow-thru cells 94 Stands and cell holders 94 Calibration and testing agents 95

For documentation Software, printer and accessories 180

Further instruments for conductivity measurement from page

Multi-parameter instruments 96

Conductivity meters for every application

78

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

ConductivityMetersHandheld

Conductivity MetersCond 315i, Cond 330i and Cond 340i

NEW

Handy

Dust- and waterproof

Easy to use

GLP-conform

Low-power systemallows continuousoperation for up to 2500 h

Long usability

Complete in SET

Features

General Information

about handheld conductivity meterscan be found on pages 14 - 17.

Pocket-size precision

Universal conductivity meters with large multifunction displayfor measurement of conductivity, temperature, salinity and totaldissolved solids. With the proven dirt-insensitive Tetracon®

4-electrode system they offer the highest degree of precisionand operating comfort.

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

79

IP 66IP 67

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Handheld Conductivity Meters

Cond 315i, Cond 330i, Cond 340ihandy · waterproof · GLP-conform

Cond 315i:

The simple conductivity meter for meeting higherdemands. With only 5 keys for error-free measurementwith parallel temperature measurement, automatictemperature compensation, 4-electrode measuringtechnology

Cond 330i:

Robust and waterproof handheld conductivity meterfor battery operation; with parallel temperaturedisplay, integrated datalogger, GLP supportingfunctions, automatic and manual temperaturecompensation with linear temperature function andnon-linear function for ultrapure water and naturalwaters according to DIN 38 404 and EN 27 888. The temperature compensation can be switched off;either 20 °C or 25 °C can be selected as the referencetemperature

Cond 340i:

As Cond 330i, but with additional analog and digitalRS 232 output and optional mains power supply

YearWarranty3

Cond 315i

80

Model Cond 315i Cond 330i Cond 340iMeasuring range/ Conductivity 0.0 µS/cm … 500 mS/cm 0.0 µS/cm … 500 mS/cmresolution with 5 ranges in 5 ranges or AutoRange

in AutoRange 0.00 µS/cm ... 19.99 µS/cm (for K=0.1 cm-1)0.00 µS/cm ... 19.99 µS/cm 0.000 µS/cm ... 1.999 µS/cm (for K=0.01 cm-1) (for K=0.1 cm-1)

Temperature -5.0 °C ... +105.0 °C -5.0 °C … +105.0 °CSalinity 0.0 … 70.0 0.0 … 70.0 TDS – 0 … 1999 mg/l Resistivity 0.00 ... 1999 MΩcm 0.000 ... 1999 MΩcm

Accuracy Conductivity ±0,5 % of value ±0.5 % of value (±1 digit) Temperature ±0.1 K ±0.1 K

Ref. temperature Tref 25 °C 20 or 25 °C, selectable

Cell constant Fixed 0.475 cm-1, 0.1 cm-1 0.475 cm-1 0.1 cm-1 and 0.01 cm-1 fixed and0.25 ... 2.5 cm-1 and 0.09 ... 0.11 cm-1 adjustable

Temperature Automatic, can be switched off Automatic, can be switched offcompensation

Temperature- • Non-linear function for natural waters (nLF) • Non-linear function for natural waters (nLF) coefficient to EN 27 888 (DIN 38 404) and to EN 27 888 (DIN 38 404) and ultrapure water function

ultrapure water function • Linear compensation from 0.01 ... 2.99%/K• No compensation • No compensation

Display LCD display 60 x 45 mm visible area, simultaneous display of measurement, temperature and special characters

Data storage – 500 data sets (measurement, temperature, date/time, identity no.); time-controlled storage at 7 intervals 5 s ... 60 min, non-volatile data storage

Data output Via display Via display or interface

Interface – Self-configuring analog output 0 ... 2 V corresponding to display, or RS 232 bi-directional interface

Operating timeMains operation – Continuous operation from mains supply

Battery operation = 500 ... 2500 h* with 4 x 1.5 V batteries with AutoSwitchOff after 1 h (except logger operation)*depending on conductivity

Protection type IP 66 and IP 67 to IEC 529

Dimensions 172 x 80 x 37 mm (H x B x D)

Weight Approx. 300 g

Certified to CE, TÜV/GS, UL and/or cETLus, CUL

Warranty 3 years for the instrument

Technical Data

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

SET Equipment

Each handheld conductivity meter is available in a completeequipment set

SET equipmentfor Cond 315i, Cond 330i and Cond 340i

• Professional case with built-in measuring set-upand separate compartment for measuring cell and accessories

• Conductivity meter with batteries • Measuring cell as chosen• KCl solution 0.01 mol/l for system checks• Stand and measuring beaker• Plug-in power supply (only Cond 340i)• Operating instructions

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

81

Handheld Conductivity Meters

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Please refer to page 88 for conductivity cells and page 94 ff. for other accessories

Handheld Conductivity Meters Order No.

Cond 315i Robust and waterproof battery-operated handheld conductivity meter 2C10-000

Cond 330i Robust and waterproof battery-operated handheld conductivity meter with datalogger 2C20-000

Cond 340i Robust and waterproof battery-operated handheld conductivity meter with datalogger 2C30-000and serial interface

Cond 340i Robust and waterproof handheld conductivity meter with datalogger and serial interface, 2C30-010for mains operation

Ordering Information

Handheld Conductivity Meters SETs Order No.

Cond 315i Handheld conductivity meter, including conductivity cell, professional case and accessories 2C10-001

Cond 330i Handheld conductivity meter, including conductivity cell, professional case and accessories 2C20-001

Cond 340i Handheld conductivity meter, including conductivity cell, professional case and accessories 2C30-001

Cond 340i Handheld conductivity meter with mains power supply, including conductivity cell, professional case and accessories2C30-011

Conductivity cells – please select oneTetraCon® 325 4-electrode conductivity measuring cellTetraCon® 325-3 4-electrode conductivity measuring cell with 3 m cableLR 325/01 Ultrapure water conductivity cell

SET ordering example Cond 330i + TetraCon® 325-3 conductivity cell 2C20-001

Note: Other conductivity cells in SET and instrument versions for 120 VAC/60 Hz are available on request

Ordering Information SETs

1

2

4

2

Conductivity measuring kit according to USP 24, with stainlesssteel flow-thru vessel for pharmaceuticalwater (see page 95).

82

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Robust

Shockproof

Waterproof

Portable

Reliable

Standard conductivitymeasurement and con-ductivity measurementat depths down to 100 m

Features

WTW portable conductivity meters

The WTW conductivity meters of the ProfiLine LF 197 series are both hose-proof (IP 66) and submersible (IP 67).

They convince because of their high degree of operator comfort and foolproofoperator guidance via the large multifunctional display. They are equipped with acarrying/support handle and carrying strap as standard.

The portable conductivity meters are available in 2 versions:

ConductivityMeters

Portable Conductivity Meters

ProfiLine LF 197 and ProfiLine LF 197-S

IP 66IP 67

YearWarranty3

ProfiLine LF 197:

ProfiLine conductivity meter, extremelyrobust, hose-proof (IP 66) and submersi-ble (IP 67), for connection of TetraCon®

325 4-electrode measuring cells anddepth armature TA 197 LF with up to100 m cable, with large multifunctionaldisplay for temperature and measure-ment, automatic temperature compen-sation, selectable reference temperatureof 20 °C or 25 °C, built-in memory, TDSand salinity measurements, RS 232 digital

output, integrated carrying and supporthandle, built-in electrode storage sleeve,for rechargeable battery and mainsoperation, and including plug-in powersupply and carrying strap

ProfiLine LF 197-S:

As ProfiLine LF 197 but with 2 analogoutputs corresponding to display forconductivity and temperature values

General Information

about portable conductivity meterscan be found on pages 18- 19.

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

83Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Portable Conductivity Meters

Conductivity depth armature TA 197 LF with built-in temperature sensor, up to100 m cable with waterproof plug (IP 67), pressure-resistant steel armoring(material VA 1.4571) with screw-off protective hood, pressure-resistant to max.10 bar, fits into 2" boreholes

Accessories for conductivity measurement at depths down to 100 m

Model ProfiLine LF 197 ProfiLine LF 197-SMeasuring range/ Conductivity 0.0 µS/cm ... 500 mS/cm in 5 ranges, additional 0.00 .. 19.99 µS/cm (at K=0.1 cm-1)resolution Temperature -5.0 ... 99.9 °C/ ±0.1 K

Salinity 0.0 ... 70.00TDS 0 ... 1999 mg/l

Accuracy Conductivity ±0.5% of value(± 1 digit) Temperature ±0.1 K

Ref. temperature 20 °C or 25 °C, selectable

Cell constant 0.1 cm-1 fixed and 0.45 ... 1.30 cm-1, adjustable

Temperature Automatic or manualcompensation

Temperature • Non-linear function for natural waters to EN 27 888 (DIN 38 404)coefficient • Linear compensation from 0.50 ... 3.00 %/K

• No compensation

Display LCD display 60 x 35 mm visible area, simultaneous display of measurement, temperature, special characters

Data storage Ring memory for 50 pairs of measurements (measurement, temperature) with data protection

Data output Via display or interface

Interface Unidirectional RS 232 interface Unidirectional RS 232 interface2 analog recorder outputs for measurement and temperature output corresponding to display

Operating time Continuous operation from plug-in mains supply; or 150 ... 800 hours with rechargeable batteries (depends on conductivity) with AutoSwitchOff after 1 h

Protection type IP 66 and IP 67 to IEC 529

Dimensions 230 x 185 x 90 mm (B x D x H)

Weight Approx. 1.5 kg

Certified to CE, TÜV/GS, UL, CUL

Warranty 3 years for the instrument

Technical Data

Ordering InformationPortable conductivity meters Order No.

ProfiLine LF 197 Robust, waterproof, submersible conductivity meter 300 350

ProfiLine LF 197-S As ProfiLine LF 197, but additionally with 2 analog recorder outputs 300 351

Note: 120 VAC / 60 Hz versions on request

Depth armatures for conductivity measurements with ProfiLine LF 197 and ProfiLine LF 197-S4-electrode conductivity depth armatures with built-in temperature sensors, cable (length optional) with waterproof plug (IP 67), VA 1.4571 steel armoring and screw-off protective hood, pressure resistant to max. 10 bar, fits into 2” boreholes

Cable length Order No. Cable length Order No.

TA 197 LF-25 25 m 301 980 TA 197 LF-60 60 m 301 982

TA 197 LF-40 40 m 301 981 TA 197 LF-100 100 m 301 983

Please refer to page 88 for conductivity cells and page 94 ff. for other accessories

inoLab® Cond Level 2:compact and communicative

Standard laboratory conductivity meterwith large multifunctional display, paralleltemperature display and automatic tem-perature compensation. An integral mea-surement store with logger function anda real-time clock allow QA-conform dataacquisition. The optional built-in printerallows data printout on thermal paperaccepted for use in official documents.20 °C or 25 °C can be set as referencetemperature. Both TDS and salinity mea-surements are possible as well as conduc-tivity and temperature. The ability to setdifferent cell constants means that otherspecial conductivity cells can be con-nected as well as the TetraCon® 325 4-electrode conductivity cell and LR 325/01ultrapure water conductivity cell (see page 93).84

Thinkinnovative

Seeinnovative

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

ConductivityMeters

Performance-orientedinstrument range

Simplified operatorconvenience

Mains-independent

Suitable as bench andwall-mounted models

Country-specific units

ISO-conform documentation

Application-orienteddisplays

Features

LaboratoryConductivity Meters

inoLab® Cond Level 1, inoLab® Cond Level 2, inoLab® Cond Level 3

inoLab® Cond Level 1: simple and reliable

Routine laboratory conductivity meterwith large multifunctional display, paralleltemperature display and automatic tem-perature compensation. 20 °C or 25 °C can be set as referencetemperature. Both TDS and salinity mea-surements are possible as well as conduc-tivity and temperature. The ability toset different cell constants means thatother special conductivity cells can beconnected as well as the TetraCon® 325 4-electrode conductivity cell and LR325/01 ultrapure water conductivity cell(see page 93).

inoLab® Level 2

inoLab® Level 1

MultiCal®

automatic calibration

AutoReadautomatic driftcontrol for stableand reproduciblemeasurements

Feed and on/off keyskeys for control-ling the built-inprinter

General Information

about inoLab® laboratory instruments can be found on pages 20 - 25.

YearWarrantyIP 43 3

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

85

Touchinnovative

Terminal Level 3

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Laboratory Conductivity Meters

inoLab® Cond Level 3 with Teminal Level 3 or PC software: flexible and powerful

High-performance laboratory conductivity meter with graphics display and digitalrecorder function, parallel display of temperature and automatic temperature com-pensation. An integral measurement store with logger function and a real-timeclock allow QA-conform data acquisition. The optional built-in printer allows dataprintout on thermal paper accepted for use in official documents.

20 °C or 25 °C can be set as reference temperature. Both TDS and salinity measure-ments are possible as well as conductivity and temperature. The ability to set differentcell constants means that other special conductivity cells can be connected as wellas the TetraCon® 325 4-electrode conductivity cell and LR 325/01 ultrapure waterconductivity cell (see page 93).

An PC keyboard interface allows an external keyboard or a barcode reader to beconnected.

Features

Built-in digital plotter

Graphic evaluation possible

Connection of bar-codereader or PC keyboard

Automatic memory fun-ction if bar-code readeris used

Selectable instrumentlanguage

Extended GLP functions

Limit input with acoustic alarm

Meets all the require-ments of USP 24

Free-of-charge softwaredownloads for inoLab®

pilot or terminal

inoLab® Level 3

AutoReadautomatic drift control forstable and reproduciblemeasurements

RUN/ENTERtriggers the measuring process

YearWarrantyIP 43 3

86

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Modelle Cond Level 1 Cond Level 2 Cond Level 3Range/ Conductivity 0.0 µS/cm ... 500 mS/cmResolution in 5 measuring ranges or AutoRange

additionally for K = 0.1 cm-1: 0.00 µS/cm ... 19.99 µS/cmK = 0.01 cm-1: 0.000 µS/cm ... 1.999 µS/cm

Temperature -5.0 ... +99.9 °CSalinity 0.0 ... 70.0TDS: 0 ... 1999 mg/lResistivity: 0.000 ... 1999 MΩcm

Accuracy Conductivity ±0.5% of value(±1 digit) Temperature ±0.1 K

Reference 20 °C or 25 °C, selectabletemperature

Cell constants 0.475 cm-1, 0.1 cm-1 and 0.01 cm-1 fixed, freely selectable 0.25 ... 2.5 cm-1 and 0.09 ... 0.11 cm-1

Temperature Automatic, manual or switched offcompensation

Temperature • Non-linear function for natural water to EN 27 888 (DIN 38 404)coefficient • Linear compensation from 0.001 ... 2.999%/K

• No compensation

Calibration With 0.01 mol KCl

Real-time clock – Non-volatile

Data memory – 200 records 2000 records(measurement, temperature) (measurement, temperature)date/time, ident. no., time-controlled date/time, ident. no., time-controlledstorage interval 5 s ... 60 min storage interval 1 s ... 60 days

Data output Display Display or interface Display, PC or interface

Analog interface – Self-configuring as analog output corresponding to display or asRS 232 bidirectional interface

Digital interface – RS 232 bidirectional interface

Operating time/ 2500 h with 4 x 1.5 V standard batteries With terminal: universal plug-in power Energy supply or mains supply supply 230 VAC or 120 VAC

Without terminal: via PC interface

Protection type IEC 529 / IP 43 without built-in printer

Dimensions 230 x 210 x 70 300 x 250 x 70 300 x 250 x 70(B x D x H in mm)

Weight Approx. 1.3 kg Approx. 1.6 kg Approx. 1.6 kg

Certified to CE, TÜV/GS, UL, CUL

Warranty 3 years for the instrument

Technical Data for laboratory conductivity meters

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

87Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

inoLab® laboratory conductometers Order No.

inoLab® Cond Level 1 Simple and reliable conductometer for routine work 1C10-010

inoLab® Cond Level 2 Compact precision conductometer with serial interface 1C20-010

inoLab® Cond Level 2 Compact precision conductometer with additional 1C21-010built-in printer for GLP-conform documentation

inoLab® Cond Level 3 Flexible and powerful precision – the intelligent conductivity measuring place 1C30-010consisting of multifunction box and universal terminal, with PC software/connection cable

inoLab® Cond Level 3 The intelligent conductivity measuring place with additional built-in printer 1C31-010

inoLab® Cond Level 3 The integrated conductivity measuring place with PC software/connection cable; without printer/terminal 1C32-000

Ordering Information for inoLab® conductivity meters

inoLab® Cond SETs Order No.

inoLab® Cond Level 1 Including cell, passive multifunction box and accessories 1C10-011

inoLab® Cond Level 2 Including cell, passive multifunction box and accessories 1C20-011

inoLab® Cond Level 2 Including built-in printer, cell, passive multifunction box and accessories 1C21-011

inoLab® Cond Level 3 Including terminal, cell and accessories 1C30-011

inoLab® Cond Level 3 With built-in printer, including terminal, cell and accessories 1C31-011

inoLab® Cond Level 3 Without printer/terminal, with PC software/connection cable, cell and accessories 1C32-001

Conductivity cells – please select oneTetraCon® 325 4-electrode conductivity measuring cellTetraCon® 325-3 4-electrode conductivity measuring cell with 3 m cableLR 325/01 Ultrapure water conductivity cell

SET ordering example inoLab® Cond Level 1 + conductivity measuring cell LR 325/01 1C10-011

Note: Other measuring cells in SET and instrument versions for 120 VAC/60 Hz are available on request

Ordering Information for inoLab® conductivity meters SETs – “ready to go”

124

4

Please refer to page 88 for conductivity cells and page 94 ff. for other accessories

Laboratory Conductivity Meters

88

TetraCon®

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

ConductivityMetersConductivity cells

TetraCon® conductivity cells

For more than 50 years WTW has been one of the leading manufacturers worldwideof precision conductivity meters and cells.

The TetraCon® 4-electrode system is the perfect result of a logical, practice-orientedfurther development and sets a new standard for professional conductivitymeasurements.

In comparison with conventional measuring cells with 2 electrodes, the TetraCon®

conductivity cell offers numerous technical advantages:

• Highest degree of precision and linearity by optimized cell geometry

• Extremely large measuring range with just one cell

• Long-term cell constant stability with high-quality abrasion-resistant graphiteelectrodes

• With built-in temperature sensor as standard

• Smallest immersion depth possible

• Ideal sample exchange – no dead volume

• No measuring errors even with very dirty electrode – contact resistance on theelectrode surface is automatically compensated

• No measuring errors from cable influences

• No measuring errors from primary or secondary polarization effects

• No measuring errors from alteration to the electrical marginal field during contactwith base or side walls

• No danger of breakage owing to robust epoxy casting technique

Special conductivity depth armatures forProfiLine LF 197 and ProfiLine LF 197-Scan be found on page 83

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

89

Conductivity Cells

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Special conductivity cellTetraCon® 325/Pt

4-electrode cell made of glass

with built-in temperature

sensor. For universal use, even

in partially aqueous or highly

aggressive media.

Electrode material platinum

Shaft length 120 mm

Cell constant K approx. 1 cm-1

Diameter 20 mm

Cable length 1.5 m

Measuring range

1 µS/cm ... 1 S/cm*

Special conductivity cellTetraCon® 325/S

4-electrode cell with built-in

temperature sensor for measure-

ments in pastes, creams and

emulsions

Electrode material graphite

Shaft material epoxy

Shaft length 120 mm

Cell constant K = 0.491 cm-1

Diameter 15.3 mm

Cable length 1.5 m

Measuring range

1 µS/cm ... 2 S/cm*

Ultrapure water con-ductivity cell LR 325/01

With built-in temperature sensor

and flow-thru vessel D 01/T for

the measurement of small

conductivities (<1 µS/cm), e.g.

boiler feed water or deionized

water.

Electrode material V4A steel

Shaft material V4A steel

Shaft length 120 mm

Cell constant K = 0.1 cm-1

Diameter 12 mm

Cable length 1.5 m

Min./max. immersion depth

35/110 mm

and flow-thru measurements

Measuring range

0.001 µS/cm ... 300 µS/cm

Standard conductivity cellTetraCon® 325

4-electrode cell with built-in

temperature sensor

Electrode material graphite

Shaft material epoxy

Shaft length 120 mm

Cell constant K = 0.475 cm-1

Diameter 15.3 mm

Cable length 1.5 m

Measuring range

1 µS/cm ... 2 S/cm*

Temperature range 0 ... 90 °C

* Measuring range depends on particular instrument

90

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Dialysis conductivity cellTetraCon® DU/TH

As TetraCon‚ DU/T, but for

dialyzers with connectors

to DIN 58 352 (Hansen

connectors).

Adapter cable KKDU 325, length

1 m, is necessary for the

connection.

Conductivity measuringcell LDM

For small sample volumes with

injection syringe for sampling

Cell constant

K approx. 150 cm-1

Conductivity flow-thru cellLDM/S

For column chromatography or

minute sample volumes.

Cell constant K approx. 15 cm-1

Adapter cable K-LDM with

built-in 30 kΩ fixed resistor is

necessary for the connection.

Volume approx. 4 ml

Temperature 0 ... 90 °C

Conductivity flow-thru cellTetraCon® DU/T

4-electrode cell with built-in

temperature sensor for standard

applications in laboratories and

production. Connection via

10 mm dia. hose nozzles.

Electrode material graphite

Shaft material epoxy

Length 155 mm

Filling volume 7 ml

Cell constant K = 0.778 cm-1

Measuring range

1 µS/cm ... 2 S/cm*

Adapter cable KKDU 325, length

1 m, is necessary for the

connection.

Trace conductivity cell LR 325/001

With built-in temperature sensor

and flow-thru vessel for the

measurement of conductivity

and resistance in boiler feed

water or deionized water as well

as in partially aqueous and non-

aqueous media.

Flow-thru vessel V4A steel

Electrode material V4A steel

Shaft material V4A steel

Shaft length 120 mm

Cell constant K = 0.01 cm-1

Diameter 20 mm

Cable length 1.5 m

Min./max. immersion depth

35/110 mm

and flow-thru measurements

Measuring range

0.0001 µS/cm ... 30 µS/cm

* Measuring range depends on particular instrument

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

91

LTA 1 2 2 2 2 2 2 1 2 2

LTA 1 1 2 2 2 2 2 2 1 2 2

LTA 01 1 2 2 2 2 2 2 1 2 2

LTA 10 1 2 1

LTA 100 1 2 2 2 2 2 1 2 2

LTA/S 1 2 2 2 2 2 2 1 2 2

LTA/KS 1 2 2 2 2 2 2 1 2 2

LR 01/T

LR 001/T

LDM 3

LDM/S 3

KLE 1

KLE 1/T

KLE 315

TetraCon® 96

TetraCon® 96-1,5

TetraCon® 96-A-

TetraCon® 96 A-4

TetraCon® 96-T-100 m

TetraCon® 325

TA 197 LF

TetraCon® L

TetraCon® 325/Pt

TetraCon® DU/T 4 5 5 5 5 4 5 4 4 5 4 4 5 5

TetraCon® DU/TH 4 5 5 5 5 4 5 4 4 5 4 4 5 5

KDU 1/T 5 5 5 5 4 5 4 4 5 4 4 5 5

KDU 1/TH 5 5 5 5 4 5 4 4 5 4 4 5 5

LR 325/01

LR 325/001 *

TetraCon® 325/S

1

2

3

4

5

*

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Selection Guide

Adapter (possible conversion with cell constants) is required:Adapter cable KLF 10Adapter cable K/LTA together with temperature sensor TFK 325 or TFK 150Adapter cable K/LDMConnection cable KKDUConnection cable KKDU 325not for Cond 315i

Selection Guide

LF 9

0

LF 9

5

LF 9

6

LF 3

15

LF 3

18

LF 3

20/3

23/3

25

LF 3

30/3

40A

Co

nd

315

i/33

0i/3

40i

LF 5

37

LF 5

38

LF 5

39

LF 3

000

Mul

tiLa

540

Mul

tiLa

P4/P

5

LF 1

96

LF 1

97

LF 5

97

Measuring cell ino

Lab

®C

on

d,

pH

/Co

nd

, M

ulti

Mul

tiLi

ne®

P4,

Mul

ti 3

40i

Mul

tiLi

ne®

P3 p

H/L

F,

pH

/Co

nd

340

i

92

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Standard conductivity cells Order No.

KLE 315 2-electrode conductivity cell with 1.5 m cable and waterproof plug 301 965

TetraCon® 325 4-electrode conductivity cell with built-in temperature sensor, K = 0.475 cm-1 301 960with waterproof plug (IP 68) and 1.5 m cable

TetraCon® 325-3 As TetraCon®‚ 325, but cable length 3 m 301 970

LTA 1 Glass/platinum 2-electrode conductivity cell, K approx. 1 cm-1 301 310(adapter cable K/LTA required)

LTA 100 Glass/platinum 2-electrode conductivity cell with adjustable cell constant, K = 1 cm-1 301 200(adapter cable K/LTA required)

TetraCon® 96 4-electrode conductivity cell with built-in temperature sensor, 301 244K = 0.609 cm-1, 1 m cable

TetraCon® L Glass/platinum 4-electrode conductivity cell with built-in temperature sensor, 301 255K approx. 1.0 cm-1, for universal use, even in partially aqueous or highly aggressive inorganic media.

TetraCon® 325/Pt Glass/platinum 4-electrode conductivity cell with built-in temperature sensor; 301 964can also be used in partially aqueous or highly aggressive solutions, K approx. 1 cm-1

For conductivity measurements at depths down to 20 mTetraCon® 325-6 As TetraCon® 325, but cable length 6 m 301 971

TetraCon® 325-10 As TetraCon® 325, but cable length 10 m 301 972

TetraCon® 325-15 As TetraCon® 325, but cable length 15 m 301 973

TetraCon® 325-20 As TetraCon® 325, but cable length 20 m 301 974

Suitable accessories

A 325/K Plastic armoring with protective hood 903 830

A 325/S Steel armoring with stainless steel protective hood, material no. 1.4571 903 831

For conductivity measurements at depths down to 100 mTA 197 LF-25 4-electrode conductivity depth cell with built-in temperature sensor, 25 m 301 980

cable with waterproof plug (IP 68), pressure resistant steel armoring andscrew-off protective hood, fits into 2” boreholes

TA 197 LF-40 As TA 197 LF-25, but with 40 m cable 301 981

TA 197 LF-60 As TA 197 LF-25, but with 60 m cable 301 982

TA 197 LF-100 As TA 197 LF-25, but with 100 m cable 301 983

Ordering Information

Ordering Information Flow-thru conductivity cells Order No.

TetraCon® DU/T Flow-thru cell with built-in temperature sensor for all standard 301 252applications in laboratories and production, connection via 10 mm dia. hose nozzles, K = 0.778 cm-1, without cable

TetraCon® DU/TH Flow-thru cell with built-in temperature sensor for checking dialyzers, 301 254connection via DIN 58 352 “Hansen connectors”, K = 0.778 cm-1, without cable

LDM/S Conductivity flow-thru cell for column chromatography or minute sample 301 750volumes, K = approx. 15 cm-1

KDU 1/T Conductivity flow-thru cell with built-in temperature sensor, for all 301 530standard applications in laboratories and production, connection via 10 mm dia. hose nozzles, K approx. 0. 8 cm-1, without cable

KDU 1/TH Flow-thru cell with built-in temperature sensor for checking dialyzers, 301 535connection via DIN 58 352 “Hansen connectors”, K approx. 0. 8 cm-1, without cable

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

93

1 2 3

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Conductivity Cells

Ordering Information Conductivity cells for special applications

Pure and ultrapure water measurements

LR 325/01 Ultrapure water cell with built-in temperature sensor, K = 0.100 cm-1, 301 961with flow-thru vessel D 01/T and waterproof plug

LTA 01 Glass/platinum 2-electrode conductivity cell for ultrapure water, K approx. 0.1 cm-1 301 410

LR 01/T Ultrapure water cell with built-in temperature sensor, with flow-thru vessel 302 520D 01/T as standard, K = 0.100 cm-1, with 1.5 m cable and waterproof plug

LR 001/T Trace conductivity cell with built-in temperature sensor and flow-thru vessel 302 530for the measurement of conductivity in boiler feed water or deionized water as well as in partially aqueous and non-aqueous media, K = 0.0100 cm-1, with 1.5 m cable

LR 325/001 Trace conductivity cell with built-in temperature sensor and flow-thru vessel for 301 962the measurement of conductivity in boiler feed water or deionized water as well as inpartially aqueous and non-aqueous media, K = 0.0100 cm-1, cable with waterproof plug

Conductivity measurements in pastes and emulsions

TetraCon® 325/S 4-electrode conductivity cell for pastes and emulsions with built-in 301 602temperature sensor, K = 0.491 cm-1, with waterproof plug and 1.5 m cable

LTA/KS 2-electrode conductivity cell for pastes, K = 0.901 cm-1 301 603

Special applications

LDM Conductivity measuring cell for small sample volumes with injection syringe 301 700for sampling, K = approx. 150 cm-1, without cable

LTA 10 Glass/platinum 2-electrode conductivity cell for extremely high conductivity values, K approx. 10 cm-1 301 112

LTA Glass/platinum 2-electrode conductivity cell with special length, 301 100suitable for measurements in test tubes (13 mm dia.), K = approx. 1 cm-1

Connection cables and accessoriesKLF 10 Adapter cable for connection of 2-electrode conductivity cell and TFK 530 303 224

K/LTA Adapter cable for connection of 2-electrode conductivity cell and TFK 325 301 429or TFK 150

K-LDM Adapter cable with built-in 30 kohm fixed resistor, length 1 m 301 430

KKDU Connection cable with 1 m cable 301 532

KKDU 325 Connection cable with waterproof plug and 1 m cable 301 963

TH 340 Telescopic extension for TetraCon® 325 ..., 1.5 ... 3 m, with sensor holder 402 802

Protective armoring

for safe on-site use

1 SM 325 Shock-absorbing, rubber protective armoring with folding support and sensor cable clamping device

2 TG/Oxi-LF Sleeve set, suitable for SM 325 protective armoring, consisting ofelectrode sleeve, holding device and additional carrying strap forfield use. Can also be used for storing the sensor.

3 FM 325/Oxi-LF Field armoring, specially designed for rough use on-site and inindustry, extremely robust and shock-resistant thanks to protective armoring with integrated electrode sleeve, carryinghandle and additional carrying strap with holding device, sensorcable clamping device and folding support for laboratory measurements

Ordering InformationProtective armoring Order No.

SM 325 Shock-absorbing, rubber protective armoring 902 818

TG/Oxi-LF Sleeve set, suitable for protective armoring SM 325 902 812

FM 325/Oxi-LF Field armoring, specially for rough conditions 902 813

94

Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

ConductivityMeters

page Protective armoring 93 Flow-thru vessels 94 Stands and cell holders 94 Calibration and testing agents 95

Ordering Information

Flow-thru vessels

Conductivity Meter Accessories

for LTA 1, LTA, LTA 01 and TFK 530 Order No.

D 530 Flow-thru vessel of transparent PVC, 108 060suitable for conductivity cells and temperature sensors,I.D. 44 mm, V*=97 ml

for TetraCon® 325

D 201 Flow-thru vessel of transparent PVC, 203 730I.D. 18 mm, V*=13 ml

for TetraCon® 96-1,5, TetraCon® 96 and KLE 1

D 1/T Flow-thru vessel, glass 302 730I.D. 24 mm, V*=36 ml

for LR 01/T and LTA 01

D 01/T Flow-thru vessel, glass 302 750I.D. 18 mm, V*=17 ml

for LTA 100

D 100 Flow-thru vessel 302 700I.D. 27 mm, V*=40 ml

V*: filling volume without sensor

Stand STH 650

Benchtop stand with flexible electrode holder for measuring cells and temperature sensors

Stand and cell holder

Benchtop stand and cell holder Order No.

STH 650 Benchtop stand with flexible electrode holder 109 809for pH electrodes, ion sensitive electrodes, reference electrodes, temperature sensors, oxygen sensors and TetraCon® 325 conductivity cells

Ordering Information

Software and printers

Information is given onpages 180 - 183.

Gen

eral

Fe

atur

esSa

mp

lePr

epar

atio

nLa

bo

rato

ry&

Fiel

d

Res

pir

atio

nB

OD

Mul

ti-

Para

met

erC

on

-d

ucti

vity

Oxy

gen

(D.O

.)IS

EO

RP

pH

Co

lon

yC

oun

ter

Pho

to-

met

ryTu

rbid

ity

Soft

war

e,Pr

inte

rsC

usto

mer

Serv

ice

95Wissenschaftlich-Technische Werkstätten GmbH & Co. KG · Dr.-Karl-Slevogt-Straße 1 · 82362 Weilheim · Germany · Tel.: +49 (0)881-183-0 · Fax: +49 (0)881-62539

Conductivity Meter Accessories

Kit for measuring the con-ductivity according to USP 24

This kit contains LR 325/01 Ultrapure water cell, D01/T Flow-thru vessel made of glass (USP-KIT 1) or stainless steel (USP-KIT 2), NIST traceable 15 µS standard with accuracy ±2% and 6R/SET/LabTesting set

Calibration standard 100 µS/cm

Shelf life 2 years, NIST traceable with accuracy ±3%

Calibration standard 15 µS/cm

Shelf life 1 year, NIST traceable with accuracy ±2%

Calibration and testing agents

Ordering Information Kit for measuring the conductivity according to USP 24 Order No.

USP Kit 1 Kit for measuring the conductivity according to USP 24, 300 569consisting of LR 325/01 Ultrapure water cell, D01/T Glass flow-thru vessel, NIST traceable 15 µS standard with accuracy ±2% and 6R/SET/LabTesting set

USP Kit 2 As USP Kit 1, but flow-thru vessel made of 300 568stainless steel instead of D01/T

Calibration agentsKS 100µS Calibration standard 100 µS/cm, shelf life 2 years, 300 578

NIST traceable with accuracy ±3% (300 ml)

KS 15µS Calibration standard 15 µS/cm, shelf life 1 year, 300 579NIST traceable with accuracy ±2% (300 ml)

EP/SET Calibration and platinization set (6 x 50 ml bottles calibration 300 570and control standard, KCl 0.01 mol/l, 30 ml platinizing solution, 1 calibration vessel), only for platinized cells

E/SET Calibration set (6 x 50 ml bottles calibration and control 300 572standard, KCl 0.01 mol/l)

Testing agents for conductivity meters6R/SET/LAB 1 Testing set for checking the conductivity meters according 302 870

to ISO 9000. Allows the measured values to be traced back to national standards.

Only for LF 3000, LF 539, LF 537 and MultiLab® P4/P56R/SET/LAB 2 Testing set for checking the conductivity meters according 302 872

to ISO 9000. Allows the measured values to be traced back to national standards.

NEW