controlling relative humidity and condensation in a co2 incubator
TRANSCRIPT
![Page 1: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/1.jpg)
Controlling Relative Humidity and Condensation in a CO2 Incubator
![Page 2: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/2.jpg)
Introduction
Deborah ThibodeauxScience Director
For more than 60 years, Baker has helped advance science, discovery and clinical care by pioneering innovative solutions for air containment and controlled environments. Our solutions include:• Biological safety cabinets• Clean benches• Fume hoods• CO2 Incubators
• Controlled-oxygen workstations• Anaerobic chambers• Media conditioning systems
Cheri GaudetMarketing Specialist
![Page 3: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/3.jpg)
Manufacturer Claim
“Our CO2 incubator provides you with the optimal conditions for culture
growth!”
As scientists, it is our duty to investigate the truth of such claims.
![Page 4: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/4.jpg)
Define “Optimal”
• 37°C – 5% CO2 – 95% RH– “Optimal” conditions for many, but not all,
applications– “Optimal” is contextual
• “Optimal” depends upon the requirements of the system being studied– Psychrotrophic / thermophilic bacteria have different
temperature requirements
• “Optimal” depends upon the aim of your research– Healthy vs. stressed systems (e.g., to mimic
dehydration, manipulation of RH is required)
![Page 5: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/5.jpg)
What You Will Learn
How to tell if your CO2 incubator provides active humidity control
How to prevent condensation, even at high humidity
Whether you can trust that your CO2 incubator really provides 95% RH
![Page 6: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/6.jpg)
DOES YOUR CO2 INCUBATOR PROVIDE ACTIVE HUMIDITY CONTROL?
![Page 7: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/7.jpg)
Types of CO2 Incubator Controls
• Open Loop Controller–Without programmable RH–With programmable RH
• Closed Loop Controller (programmable RH is standard)
Do any of these provideactive humidity control?
![Page 8: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/8.jpg)
Open Loop ControllerWithout Programmable RH
Control Programmi
ng (ONE set point)
Heat Water Pan
Humidity Delivered
to Chamber
![Page 9: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/9.jpg)
× One humidity set point determined by manufacturer, not user
× Microprocessor uses mathematical formula based on static (not real-world) conditions
× Uncertainty – is system delivering promised RH?
Control Programming (ONE set
point)
Heat Water Pan
Humidity Delivered
to Chamber
Open Loop ControllerWithout Programmable RH
![Page 10: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/10.jpg)
Does it provide active humidity control?
× No control over relative humidity whatsoever
Open Loop ControllerWithout Programmable RH
![Page 11: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/11.jpg)
“Dial in” to set point
Heat Water Pan
Humidity
Delivered to
Chamber
Open Loop ControllerWith Programmable RH
![Page 12: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/12.jpg)
User-defined humidity set point
×Microprocessor uses mathematical formula based on static (not real-world) conditions
×Uncertainty – is system delivering promised RH?
“Dial in” to set point
Heat Water Pan
Humidity
Delivered to
Chamber
Open Loop ControllerWith Programmable RH
![Page 13: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/13.jpg)
Does it provide active humidity control?
×Only gives the illusion of control
Open Loop ControllerWith Programmable RH
![Page 14: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/14.jpg)
Open Loop Controllers –The Problem
No System Feedback
![Page 15: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/15.jpg)
Closed Loop Controller
Heat water pan
Humidity delivered to
chamber
RH sensor samples
atmosphere
RH measurement informs
control algorithm
Control algorithm adjusts temp of heating
elements
“Dial in” to set point
![Page 16: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/16.jpg)
Closed Loop Controller
Heat water pan
Humidity delivere
d to chamber
RH sensor
samples atmosph
ere
RH measure
ment informs control algorith
m
Control algorith
m adjusts temp of heating element
s
“Dial in” to set point
User-defined humidity set point
Control algorithm adapts to real-world conditions
Certainty – system maintains RH to set point
![Page 17: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/17.jpg)
Closed Loop Controller
Does it provide active humidity control?
Yes
![Page 18: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/18.jpg)
HOW CAN YOU PREVENT CONDENSATION, EVEN AT HIGH HUMIDITY?
![Page 19: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/19.jpg)
Is Condensation Completely Avoidable?
×No
Some condensation is unavoidable, such as when cooling the chamber interior
![Page 20: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/20.jpg)
Is Condensation Preventable?
Yes
During normal operation, condensation is preventable, even at high humidity
![Page 21: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/21.jpg)
What Causes Condensation?
Water condenses from the air onto a surface when the temperature of the
surface falls below the dew point
![Page 22: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/22.jpg)
What Causes Condensation in aCO2 Incubator?
• Fluctuation in environmental parameters (e.g., door opens)
• Interior surface temperature gradient changes– Surface temperature gradient: Variance in
temperature across the interior surface, expressed as a range of temperatures
• Portion of the surface temperature gradient falls below dew point
• Condensation occurs
![Page 23: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/23.jpg)
Technologies that Prevent Condensation?
×No prevention above 90% RH×Designated “cold spot” in incubator cooled
to below dew point to “divert” (not prevent) condensate– Does not prevent condensation from occurring
on other surfaces that are at or below the dew point
• Closed-loop controllers provide better control over air temperature & RH, which directly affect dew point, but this is not enough
![Page 24: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/24.jpg)
Muddled Marketing Claims
Incubator suppliers are not being untruthful when representing the level of control they
are able to provide over RH and condensation…
…but they are not telling you the whole story.
The most pervasive claim about CO2 incubators may be the least understood.
![Page 25: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/25.jpg)
DOES YOUR INCUBATOR REALLY PROVIDE 95% RH?
![Page 26: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/26.jpg)
RH Sensor Accuracy
• Closed-loop controllers in CO2 incubators typically use a capacitive RH sensor to provide feedback to system
• Like all instruments, subject to some degree of inaccuracy
• How accurately the incubator control algorithm maintains RH set point directly depends on the accuracy of the RH sensor
![Page 27: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/27.jpg)
RH Sensor Accuracy
Accuracy of capacitive RH sensors depends on:• Temperature• RH• Length of time in continuous operation• Age of instrument
×Most incubator manufacturers do not disclose this information, let alone
account for it in the control algorithm
![Page 28: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/28.jpg)
Typical RH Sensor Accuracy
Conditions RH Sensor Accuracy
RH is between 10-100% @ 25°C ± 1.8%
RH is greater than 80% for 60 hours Additional + 3%
Temperature is between 10°C and 35°C Additional ± 0-1%
Temperature is greater than 35°C Additional ± 1-2%
Yearly Drift Additional < + 0.5%
![Page 29: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/29.jpg)
Real-World Implications
• An RH sensor performs least accurately under the most common conditions
• An incubator in its first year of use, in continuous operation at 37°C and 95% RH may be delivering humidity levels as low as 88.2% and as high as 98.8%.
![Page 30: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/30.jpg)
Real-World Implications
The verdict?× Uncertainty – impossible to tell whether the system
is maintaining anywhere near the RH set point
Unless sensor accuracy is factored in to the control algorithm
![Page 31: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/31.jpg)
TECHNOLOGIES TO LOOK FOR AND INSIST UPON
![Page 32: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/32.jpg)
Humidity & Condensation Control Technologies
Closed-loop control algorithm • Allows user-defined, active humidity control, reducing experimental error and uncertainty
• Helps prevent condensation by providing better parameter control
Two humidity set point ranges1. “Up to 90%” where user can
“dial in” to a desired RH2. “90%-95%” where a “sweet
spot” of >90% RH can be delivered consistently
• Prevents condensation and the risk of contamination that goes along with it
• Significantly reduces uncertainty pertaining to RH accuracy
![Page 33: Controlling Relative Humidity and Condensation in a CO2 Incubator](https://reader031.vdocument.in/reader031/viewer/2022012406/55d1dc74bb61eb5e328b47e5/html5/thumbnails/33.jpg)
Thank You
For more information on this topic, download our white paper:
http://info.bakerco.com/co2-incubator-humidity-control