Successive Approximation
(Potentiometric A/D conversion-
continued )
-
7/28/2019 Converters Pp Ppp
3/14
Voltage to Time A/D converter (Ramp type)
-
7/28/2019 Converters Pp Ppp
4/14
Figure: 26.111 Voltage to Time A/D converter (Ramp type)
-
7/28/2019 Converters Pp Ppp
5/14
Voltage to Time A/D converter (Ramp type)
-
7/28/2019 Converters Pp Ppp
6/14
-
7/28/2019 Converters Pp Ppp
7/14
-
7/28/2019 Converters Pp Ppp
8/14
Voltage to Frequency Converter (Integrating Type)
-
7/28/2019 Converters Pp Ppp
9/14
Dual slope integration A/D converter
-
7/28/2019 Converters Pp Ppp
10/14
-
7/28/2019 Converters Pp Ppp
11/14
A classical dual-slope converter is shown at the drawing
A current, proportional to the input voltage, charges a capacitor for a fixed time
interval Tcharge. At the end of this interval the device resets its counter and applies anopposite-polarity (negative) reference voltage to the integrator input. With this opposite-
polarity signal applied the cap is discharged by a constant current until the voltage at the
output of the integrator reaches zero again. The timeTdischarge is proportional to the input
voltage level and used to enable a counter. The final count provides the digital output,
corresponding to the input level.
-
7/28/2019 Converters Pp Ppp
12/14
Note that even the clock frequency does not have to have high stability, because both
ramp-up and ramp down time are measured with the same clock.If the clock slows down
10%, the initial ramp will go 10% higher than normal, requiring 10% longer ramp down
time resulting in the same final count. Only the discharge current produced by precise
Vref has to be of high stability.
Integrating ADCs are extremely slow devices with low input bandwidths. But their ability
to reject high-frequency noise and fixed low frequencies such as 50Hz or 60Hz makes
them useful in noisy industrial environments and applications . Provide 10-18 bit
resolution. A coversion time for a medium speed 12 bit integrating ADC is about 20mS.
This type of ADC is most commonnly used in multi-meters.
-
7/28/2019 Converters Pp Ppp
13/14
Sample & Hold operation
-
7/28/2019 Converters Pp Ppp
14/14
Purpose The sample and hold circuits are essentially used in linear systems. In some kinds of
analog-to-digital converters, the input is often compared to a voltage generated
internally from a digital-to-analog converter (D-A-C). The circuit tries a series ofvalues and stops converting once the voltages are "the same" within some defined
error margin. If the input value was permitted to change during this comparison
process, the resulting conversion would be inaccurate and possibly completely
unrelated to the true input value. Such successive approximation converters will
often incorporate internal sample and hold circuitry. In addition, sample and hold
circuits are often used when multiple samples need to be measured at the sametime. Each value is sampled and held, using a common sample clock.