Regulatory Inspections• Opening meeting with inspection team and laboratory
management– Review last audit and corrective actions– Agree on final agenda
• Investigate documentation in an office room– Review organizational structure, tasks and
responsibilities– High level documentation, e.g., quality plan– Review selected procedures– Review selected records
• Walk through the laboratory– Observe and interview people at work– Inspect one or more systems previously discussed
• Exit meeting with management• Follow-up activities
Slide 2
Copyright
Inspection Documentation• 483 Form Inspectional Observation
– only deviations listed– shows the view of the inspection team– written for and discussed during the inspection exit
meeting • Establishment Inspection Report (EIR)
– very detailed (20-40 pages)– more like an inspection protocol
• Warning letter– With significant deviations– Reviewed by FDA centers
www.fdawarningletter.com
Slide 3
Copyright
Training
• Failure to adequately establish procedures for identifying training needs and ensure all personnel are trained to adequately perform their assigned responsibilities and the training is documented (W-228)
• Your firm fails to document on the job training. (W-228)
• Your firm failed to list second shift quality personnel, their positions, and to whom they report within the corporate quality structure. (W-228)
www.fdawarningletter.com
Slide 4
Copyright
Supplier Assessment
• There is no assurance that your firm establishes the reliability of the supplier's analyses through appropriate validation of the supplier’s test results at appropriate intervals (W-245)
• There are no incoming component specifications for acceptance and no supplier quality agreements (W-254)
www.fdawarningletter.com
Slide 5
Copyright
Equipment Qualification
• Failure to have an adequate performance qualification (calibration) program for the QC laboratory instruments.
• Your HPLC calibration lacks a carry over test (sample injection residual test), sample energy (intensity of light source), and lamp use hours determination.
• You fail to challenge the analytical balances for minimum weight, measurement for uncertainty, and drift value.
• In addition, you do not calibrate the Karl Fisher syringe used during xxx API water content analysis.) (W-246)
Slide 6
Copyright
Method Validation and Transfer• The accuracy, sensitivity, specificity, and
reproducibility of test methods have not been established and documented (W-187)
• Failure of your quality control unit/laboratory to ensure your analytical methods used to evaluate the stability of your APIs are validated to be stability indicating. (W-243)
• Methods that were validated at one facility and transferred to xxx site are being used without a methods transfer or revalidation protocol. (W-186)
Slide 7
Copyright
System Suitability Testing
• The suitability of all testing methods is not verified under actual conditions of use.- The HPLC configuration tubing at the time of method validation may not be similar to actual ones used possibly affecting the resolution. (W-241)
• The SOP requirement for the assay analysis of xxx was not followed in that the HPLC system suitability test was only performed weekly per firm SOP, instead of the actual time of testing (133)
www.fdawarningletter.com
Slide 8
Copyright
Incomplete Laboratory Records
• Failure to have laboratory records that include complete data derived from all tests necessary to assure compliance with established specifications and standards.(W-214)
• Laboratory records fail to include the initials or signature of the person who performs each laboratory test [21 CFR § 211.194(a)(7)].
• Specifically, laboratory analysis records for analyses performed on HPLC xxx and xxx do not indicate which analyst performed the injections. (W-171)
Slide 9
Copyright
Consequences of non-Compliance
• 483 form inspection reports• Warning letters• Shipment hold• Product recall• Seizure• Import alerts and denials• Consent decree• Debarment
(personnel excluded from certain positions)
Slide 10
Copyright
Consequences of Non-Compliance
• FDA Press release on Jan 25, 2012• Prevents temporary import for products from two sites in India• Causes pay cut for company executives and directors• Ranbaxy to hire consultant with expertise in data integrity• Possible delay of generic versions of blockbuster drugs (e.g. Lipitor) with uncalculatabled costs
30 products were banned from the US market•GMP violations•Inadequate testing•Falsified data•Data integrity issues
Slide 11
Copyright
Responding to FDA Inspectional Deviations• During the inspection, before the inspection observation has
been drafted• In response to the 483 inspection observations
– Not legally required but recommended to reply within 15 days
• In response to warning letters– Legally required: 15 days
• Through other communications with the FDA
Slide 12
Copyright
Responding to FDA/CFDA Deviations • Exit meeting: Fully understand the content of each deviation • Respond in time in writing (15 business days)• Address each item separately• Mention that you understand and accept the deviation• If some deviations are already fixed, provide documented evidence• For others give a reason for the deviation, write how you will correct
the deviation– Who, how, when
• Attach preliminary documentation for the corrections, if available • Commit to train all affected personnel • Write how you will prevent the same of similar problem re-occurring
Slide 13
Copyright
Prepare Your Organization for FDA Inspections
• Design GMP compliance through learning from warning letters
• Develop SOP for FDA inspections• Develop Checklist • Train management and staff• Conduct ‘FDA inspection like’ internal audits• Learn how to effectively to respond
inspection reports
Slide 14
Copyright
Response to 483 too late
• We acknowledge your written response, dated September 23, 2010,, to the Form FDA 483. • However, because this responses were received more than 15 business days after the
Form FDA 483 was issued; this response have not been considered. • We plan to evaluate your additional responses to the Form FDA 483, along with any other
written material provided, as a direct response to this Warning Letter
www.fdawarningletter.com (253)
483 received on August 26, 2010
Slide 15
Copyright
Inadequate Response to 483 Deviation•For example, the employees of your firm, including a member of your quality control staff, admitted to our investigators that they were unaware of and were not trained to follow your SOP for handling deviations. There were at least two instances in which an OOS investigation was not conducted.
Inadequate Response•While your response proposes to provide training to all employees, it does not provide documentation that you have initiated the training and does not specify a timeframe for completion.
• In addition, your response fails to specifically address how the proposed training will ensure that all employees will be trained in SOPs that are relevant to their job functions..
www.fdawarningletter.com (253)
Slide 16
Copyright
Inadequate Response to 483 Deviation from inspection in Sept Oct 2010•Failure to calibrate, in accordance with written procedures and an established schedule, weighing equipment critical for ensuring the quality of APIs. .Inadequate Response• In your response, your firm provided certification calibration records from your calibration services contractor to address actual use ranges of the balances. •Your response, however, is inadequate because you do not include any information regarding a calibration program to ensure future calibration of balances on a routine basis.
Slide 17
Copyright
Deviation: HPLC for Testing of xx not Qualified
Qualify that
specific HPLC
Details How to qualify?
OtherLab
Equip-ment?
Time frame?Training
Other HPLCs in the lab?
Corrective Action
Other labs?Global
Approach?
Impact on other tested
products?
Preventive Actions
Slide 18
Copyright
Responding to 483s/Warning Letters
• Devitions– Failure to adequately establish procedures for identifying training
needs and ensure all personnel are trained to adequately perform their assigned responsibilities and the training is documented (W-228)
• Root cause (assumed): – No document on how to handle training
• Suggestions for corrective action to correct the existing violation– Develop and implement procedure on how to handle training, including
documentation, responsibilities, second shift worker• Suggestions for preventive actions to prevent reoccurrence of the
same or similar problem at the same or similar location – Develop a comprehensive training plan and SOPs that cover all aspects
of people qualification, including documentation and assessment. • Documentation: Corrective and preventive action plans with owners and
deliverables, time schedule
Slide 19
Copyright
Workshop Overview
1. Small groups of attendees review most frequently cited FDA Warning letters/483’s related to laboratory controls
2. Groups discuss and document a. The root cause (reason for the problem, if it is not obvious, make something up, but only one!!!)b: Suggestions for corrective actions, must be in line with the suggested root causec: Suggestions for prevention actionsd. Suggestions for documents to be provided to FDA
3. Groups present the outcome to the entire audience (In english or Japanese)
Slide 20
Copyright
Free Resources• Agilent Primers
– Analytical Instrument Qualification and System Validation (Chinese)– Validation of Analytical Methods– Good Laboratory Practice and Good Manufacturing Practice– Understanding and Implementing ISO/IEC 17025– Compliance for BioPharmaceutical Laboratories– Qualification and Validation for Supercritical Fluid Chromatography– Elemental Impurity Analysis in Regulated Pharmaceutical Laboratories – „Quality by design“ for Quality Control Laboratories: Learning from FDA Warning LettersOther Resources– Tutorials (method validation, computer validation, GLP)– References to FDA Warning letters and 483s– Free Labcompliance Newsletter
www.labcompliance.com/agilent (available until September 30, 2014
Slide 21
Ludwig HuberLabcompliance Sponsored by
Thank YouI would like to thank • All attendees for your attention• Agilent Technologies China for invitation and
organization
Slide 22
Check topics and details of 200+ audio/video seminarsAudio: www.labcompliance.com/seminars/audioVideo: www.labcompliance.com/seminars/video