1 Always Thinking Ahead. CT Basics Principles of Spiral CT Dose

Upload: others

Post on 03-Jun-2020

4 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

1

Always Thinking Ahead.

CT BasicsPrinciples of

Spiral CTDose

Page 2: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

2

Always Thinking Ahead.

Who invented CT ?1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections

Sir Godfrey Hounsfield worked for the Central Research Labs. of EMI, Ltd in England fInvented 1st clinically CT scanner in 1971Nobel Prize Winner 1979

Sir Godfrey Hounsfield

Nobel Prize - Medicine

(1919 – 2004)

Page 3: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

3

Always Thinking Ahead.

Before the Era of CT……...

Structures are superimposed2D image ( 2 planes )

Cannot differentiate subtle tissue densities

What is wrong with this image?

Page 4: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

4

Always Thinking Ahead.

How is a CT scan made ?

Cross-sectional imaging technique

Uses x-rays from x-ray tube

Detector Converts x-rays to light

Light converted to electronic signal

Electronic signal constructed to make image in a computer

Density differentiation – grayscale / colourimages

Page 5: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

5

Always Thinking Ahead.

Inside a CT - Sensation 64

Tube

Collimator

Generator

Detector

Page 6: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

6

Always Thinking Ahead.

CT : From x-ray to detailed image

Image Recon. Computer

Diagnostic images

Page 7: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

7

Always Thinking Ahead.

The FIRST Siemens CT Scanner

Scan time 7 minutes, Image Matrix 80x80 Pixel, Scan field 25 cm, spatial resolution 1.3 mm (4lp/cm)Head scanner

SIRETOM (1974)

Page 8: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

8

Always Thinking Ahead.

Image acquisition – the “slice”

X-ray beam is narrowed by an collimator

Only one thin part of the body will be radiated

Page 9: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

9

Always Thinking Ahead.

Image acquisition: the “Voxel”The slice will be partitioned mathematically into small Volume elements - “voxel”.

The attenuation wiIl be measured in the voxels as a constant value

The image element in a plane is a “Pixel”

Page 10: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

10

Always Thinking Ahead.

Concept of X-ray Attenuation

Incident X-ray Transmitted ray

SCATTERED XSCATTERED X--RAYSRAYS

BODYTISSUE

X-ray beam passing through the body is attenuated (loses its energy) by:AbsorptionScattering

The degree of attenuation is measurable with CT and is expressed in HU (Hounsfield) unitHU gives an indication of tissue density (CT Density)

Page 11: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

11

Always Thinking Ahead.

Hounsfield Unit Scale

Water

Mamma

Air

Bone

Spleen

Fat

Pancreas

Lung

Kidneys

AdrenalGland

Blood

Heart

Liver

Intestine

Tumor

Bladder

3000

60

40

0

-100

-200

-900-1000

Rule of thumb: The CT value of water is 0 and air -1000. The relative values of the other tissues are calculated relative to that of water

Rule of thumb: The CT value of water is 0 and air -1000. The relative values of the other tissues are calculated relative to that of water

This is the CT number or Hounsfield unit (HU)

Page 12: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

12

Always Thinking Ahead.

Headline

Page 13: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

13

Always Thinking Ahead.

CT then…………CT now…………..

SIRETOM 1974 Spirit 2004

Page 14: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

14

Always Thinking Ahead.

CT Basics: Sequential scanningIn 1970’s , CT worked only using a sequential scan

One Scan - First rotation ( 2 seconds per rotation !! )• Rotation of tube and detector around 360° + Radiation

exposure • After each scan a table movement was performed

The tube and detector system moved back to the initial position

• Because of cables !Table movesSecond rotation ( 2 seconds per rotation !! )Table moves etc etc

Problem:Long examination timeThe lung could not be scanned in one breath hold

Page 15: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

15

Always Thinking Ahead.

Conventional sequential CT...

I.S.D* I.S.D*Sequence-Scan:

- long cycle time

•Inter scan delay- time between 1st slice to 2nd slice-Gantry rotation, gantry stop, -patient moves, -gantry rotates, gantry stop etc.

Please note that conventional sequential CT still used today for some head scans ( however, gantry does not require to return anticlockwise to start position due to slip rings )

Page 16: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

16

Always Thinking Ahead.

CT Basics

Mis-registration due to different levels of respiration between slices

deep Inspiration shallow Inspiration

Lungs – like a sponge !

Page 17: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

17

Always Thinking Ahead.

1990: The Spiral CTEra………….

Page 18: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

18

Always Thinking Ahead.

Conventional CT versus SpiralSpiral--CT ...CT ...

3D-Volume 2D-Slice

no I.S.D with I.S.D.*

Page 19: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

19

Always Thinking Ahead.

Principles of Spiral CT

Continuous gantry rotation

Continuous table feed ( head or foot direction )

Continuous radiation

Continuous volume acquisition

Table movement

Page 20: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

20

Always Thinking Ahead.

Principles of Spiral CT

30s - 40s

Volume scanning in a single breath holdRaw data ( CT computer data) is created, then image (picture) data is processed using CT computer

Page 21: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

21

Always Thinking Ahead.

Continuous Slices

Principles of Spiral CT

Concurrent Image Reconstruction during the spiral acquisition orRetrospective Image Reconstruction

Page 22: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

22

Always Thinking Ahead.

Principles of Spiral CT

Image Reconstruction at arbitrary slice positionover a limited volume

Page 23: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

23

Always Thinking Ahead.

Principles of Spiral CT

Image Reconstruction with gaps

larger distance (increment) between the slices than the slice thickness

Page 24: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

24

Always Thinking Ahead.

Principles of Spiral CT

Image Reconstruction with overlap

smaller distance (increment) between the slices than the slice thickness

Page 25: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

25

Always Thinking Ahead.

Principles of Spiral CTProblem: Partial Volume Artifact

• Partial Volume Artifact:

Suspicious lesion is covered partially in one slice

• Overlapping axial slicesReduces Partial Volume Artifacts by centering the lesion within the sliceNo additional radiation dose

Page 26: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

26

Always Thinking Ahead.

Spiral CT benefits

Complete Volume coverageFaster data acquisition

More patient comfortData acquisition in one breath holdLess movement artifactsLess contrast mediaLess patient doseHigher patient throughput

Page 27: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

27

Always Thinking Ahead.

New Concept with Single Slice Spiral

Spiral Scanning can be done at variable speedrelative to the collimation

Pitch = Table Feed per RotationCollimation

Examples for Collimation = 5.0 mm

Table feed per rotation = 5.0 mm Pitch 1.0Table feed per rotation = 7.5 mm Pitch 1.5Table feed per rotation = 10.0mm Pitch 2.0

Freely selectable from 1.0 to 2.0 [Emotion (single slice )]

Page 28: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

28

Always Thinking Ahead.

Clinical Benefits of Increased Pitch

30s

More coverage in the same time

Pitch 1

30s

Pitch 2

Page 29: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

29

Always Thinking Ahead.

Pitch & Image Quality

Image plane

Scan direction

1 2 3 4 5 6 7 8 9 10 11 12 Pitch 1

1 2 3 4 5 6 Pitch 2

Scan direction

Data points are stretched with increasing Pitch causingslice profile broadening

Page 30: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

30

Always Thinking Ahead.

Slice Profile (SP)Slice Profile (SP)

Effective slice thickness of an image Effective slice thickness of an image

Slice ProfileSlice ProfileResolutionResolution

Factors influencing Slice ProfileFactors influencing Slice Profile•• CollimationCollimation•• PitchPitch•• Interpolation algorithm (360Interpolation algorithm (360°° or 180or 180°°))

Page 31: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

31

Always Thinking Ahead.

In the worst case scenario………….

Example

5 mm collimation, Pitch 1.0 5.0 mm Slice Profile

5 mm collimation, Pitch 2.0 6.5 mm Slice Profile

Scanning at Pitch 2 leads to decrease in image quality on single slice CT &other vendors MSCT

Page 32: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

32

Always Thinking Ahead.

Clinical Benefits of Increased PitchReduced scan time & dose for the same coverageassuming same slice size

Pitch 1.0 10mm/rot

Pitch 2.020mm/rot

30s 15s

Page 33: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

33

Always Thinking Ahead.

Clinical Benefits of Subsecond Spiral

Reduced scan time & doseImproved resolutionMore coverageImproved Temporal Resolution

Less motion artifacts

New applications – Cardiac CT

30s with 1.0s rotation

22s with 0.8s rotation

40cm 30cm

0.8s Spiral 1.0s Spiral

10mm30s

8mm28s

7mm32s

0.75s SPIRAL 1s SPIRAL

Page 34: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

34

Always Thinking Ahead.

Chest Imaging Protocols ComparisonSequential CT

Slice Imaging

~ 2s per slice

100 -150 ml contrast

8 - 10 mm collimation

Single Spiral CT

Volume Imaging

25s - 40s per volume

50 - 80 ml contrast

3 - 5 mm collimation

Eliminates respiratory mis-registration

Faster with improved spatial resolution

Reduced contrast dosage

Page 35: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

35

Always Thinking Ahead.

Conventional vs. Single-Slice Spiral vs. Multi-Slice Spiral

Multi-slice VolumeSingle Slice Single Volume

no I.S.DI.S.D*

* Inter Scan Delay

Page 36: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

36

Always Thinking Ahead.

Multi-Slice CT Scanning

MultiMulti--Slice CTSlice CT

Multiple slices are acquired for each rotation.

With the currently available multi-slice scanners, either 2, 4*, 6, 10, 16, 24**, 40 or 64 slices can be acquired per rotation.

• Refurbished Volume Zoom / Sensation 4

** Sensation Open

Page 37: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

37

Always Thinking Ahead.

Multi-slice Detector Designs

Single slice CT’s have one detector row,while multi-slice CT’s have multiple detector rows.

Emotion 6

Sensation 64

Page 38: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

38

Always Thinking Ahead.

Multi-Detector Adaptive Array Design

z-axis

Sca

n-FO

V

3 38x.52 22x1 2x1

Emotion 6 adaptive array detector

Page 39: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

39

Always Thinking Ahead.

Adaptive Array Design – S16

4 x 1.5 z-axis4 x 1.516 x .75

Sca

n-FO

V

Page 40: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

40

Always Thinking Ahead.

Detector Configuration Sensation 164x1.5mm 16x0.75mm 4x1.5mm

16x0.75mm

16x1.5mm

Detail Mode

Routine Mode

Page 41: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

41

Always Thinking Ahead.

Detector Configuration Sensation 64

32x0.6mm 4x1.2mm4x1.2mm

64 x 0.6mm Routine Mode

24 x 1.2mm Volume Mode

28.8mm z-coverage

Page 42: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

42

Always Thinking Ahead.

Slice Collimation & Width...

For all Siemens MSCT´s a combination of Collimatedslice and reconstructed slice thickness have to be selected

The slice width is the true width of a Siemens CT reconstructed slice ( unlike other vendor’s MSCT systems who have slice broadening )

Slice width can only be thicker than slice collimation…cannot be thinner

The slice thickness can be changed retrospectively

Page 43: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

43

Always Thinking Ahead.

Slice collimation...

. Single Slice CT

Slice collimation is the slice thickness collimated by the tube collimator determining the Z coverage per rotation.

Multi Slice CT

Slice collimation is the total slice thickness collimated by the tube collimator and then divided by the number of active detector channels.

Page 44: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

44

Always Thinking Ahead.

Two different definitions for Pitch

= Pitch 1,518mm per rotation

16 rows x 0.75 mm Pitch Factor:

Pitch Factor : table movement per rotation divided by the total slice collimationVolume Pitch : table movement per rotation divided by the slice collimation

e.g. Sensation 16 Abdomen scan

= Pitch 2418mm per rotation

0.75 mm Volume PitchFactor:

Page 45: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

45

Always Thinking Ahead.

Pitch...In multi-slice spiral with the Emotion Duo / 6 and the Sensation, you select the slice collimation together with the slice width, and the slice width is not affected by the pitch or the algorithm.

No slice BroadeningWhat you ask for is what

you get!

Page 46: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

46

Always Thinking Ahead.

Clinical Applications of Spiral CT

Routine Volume Acquisition (in a breath hold)

Routine Neuro Spiral Imaging

Multiphasic Spiral Contrast Studies

CT Angiography's

Musculoskeletal 3D Imaging

Page 47: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

47

Always Thinking Ahead.

Routine Applications

Bone and lungs

Soft tissue

Page 48: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

48

Always Thinking Ahead.

Bone & Lungs

“Black & White” - High Contrast

Page 49: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

49

Always Thinking Ahead.

Bone and lungsBig differences in the tissue densitiesHigh contrast resolution is neededHigh images noise – less dose is neededNo differentiation of small tissue density differences

Inner ear Lung

Page 50: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

50

Always Thinking Ahead.

MTF: Catphan HC-insert with kernels H30s, U95u

H30s , FOV = 200 mm , 7lp /cm U95u ,FOV = 100 mm , 20lp /cm

Page 51: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

51

Always Thinking Ahead.

Soft Tissue - Low Contrast Resolution

Possibility to visualize different tissue densities, which are close togetherNo image noise – you need a higher dose

Head Abdomen/Liver

Page 52: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

52

Always Thinking Ahead.

Low Contrast Resolution

To visualize small density differences, a good Low Contrast resolution is needed Specification of Low Contrast Resolution

Phantom: CATPHAN (16mm)Object size: 3mmcontrast diff.: 3HUDose at the surface: 19.7 mGy at 100/104 mAsTechnic: 1.0 sec; 10mm, 120kv

Page 53: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

53

Always Thinking Ahead.

Low-contrast: Emotion (B30s,130 kV, 10 mm)

73 mAs 225 mAs20 cm Catphan

FOV = 250 mm

Page 54: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

54

Always Thinking Ahead.

LC

HC

Routine Mix of CT Examination

Only 10% are High Contrast Studies

Page 55: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

55

Always Thinking Ahead.

3D Imaging - Abbreviations

MPR – Multi Planar ReconstructionsTransaxial images are combined to volume. The volume can be reformatted to secondary images in selected planes ( sagittal, coronal or oblique)

SSD – Shaded Surface DisplaySurface images of tissue structure are created out of the volumedataset. A three dimensional object is calculated from voxels, whose threshold values are within a specific density range

VRT – Volume Rendering TechniquePossibility to render different tissues, which have a different densities, as a 3D object in different colors and with differentbrightness and opacity

MIP – Maximum Intensity ProjectionPossibility to render different tissues, which have high density and show them as a 3D object in different grey scales

Page 56: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

56

Always Thinking Ahead.

Reconstructions Orientations

axialcoronal sagittal

Page 57: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

57

Always Thinking Ahead.

Axial Image1 mm, Pitch 2, Emotion

Sagittal Reformat

Coronal Reformat

High-Resolution Ankle Study

Axial Imaging with Multiplanar Results

Page 58: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

58

Always Thinking Ahead.

3D Trauma Imaging – Fractures

Spine (VRT)Spine (MPR)

Spine (SSD)

Page 59: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

59

Always Thinking Ahead.

Osteosacroma Left Pelvis2 mm, Pitch 1.5, Emotion

Musculo-Skeletal 3D Imaging (VRT)

Page 60: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

60

Always Thinking Ahead.

Why use an injector with contrast media for an examination ?

Images with contrast media injection helps by the diagnosis and by the differential diagnosis

Multiphase Liver examination

No injection With injection

Page 61: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

61

Always Thinking Ahead.

Arterial Venous

Liver Study with Contrast Medium

Arterial Phase Venous Phase

Page 62: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

62

Always Thinking Ahead.

4-phase Liver Hemangioma (Emotion)1 Pre-contrast2 Arterial3 Portal-venous4 Delayed phase

1 2

3 4

Page 63: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

63

Always Thinking Ahead.

Contrast media injection for CT Angio´s

CT AngiographyVisualization of vessels without catheter

Angio study of Kidney (MIP)

Leg – femoral runoff (MIP)

Page 64: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

64

Always Thinking Ahead.

Advanced Contrast StudiesVRT ( InSpace )

Page 65: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

65

Always Thinking Ahead.2 mm, Pitch 2, Emotion MIP

Neuro CTA – Carotid Arteries

2 x 1mm, Emotion Duo VRT

Page 66: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

66

Always Thinking Ahead. 2 x 1mm, Emotion Duo VRT

Abdominal CT Angiography – Aortic Stent

Page 67: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

67

Always Thinking Ahead.

Headline

Courtesy of University of Erlangen, Department of Radiology and Institute of Medical Physics

SOMATOM Sensation 64VRT( InSpace)6 sec for 350 mm64 x 0.6mm (2x32)Resolution 0.4 mmRotation 0.37 sec120 kV / 150 mAs

Page 68: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

68

Always Thinking Ahead.

Headline

Courtesy of University of Erlangen, Department of Radiology and Institute of Medical Physics

SOMATOM Sensation 64MIP

6 sec for 350 mm64 x 0.6mm (2x32)Resolution 0.4 mmRotation 0.37 sec120 kV / 150 mAs

Page 69: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

69

Always Thinking Ahead.

Headline

3mm

3mm

Heart Specimen Study

SOMATOM Sensation 64

VRT

Stent 3 mm

Slice thick. 0.6 mm

Resolution 0.4 mm

Excellent visualization of 3 mm stent

Significant reduction of stent struts “blooming” artifacts

Page 70: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

70

Always Thinking Ahead.

HeadlineSOMATOM Sensation 64VRT(InSpace)

33 sec for 1570 mm64 x 0.6mm (2x32)Resolution 0.4 mmRotation 0.5sec120 kV / 148mAs

Courtesy of University of Erlangen and University of Tübingen

Page 71: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

71

Always Thinking Ahead.

HeadlineSOMATOM Sensation 64VRT

33 sec for 1570 mm64 x 0.6mm (2x32)Resolution 0.4 mmRotation 0.5sec120 kV / 148mAs

Courtesy of University of Erlangen and University of Tübingen

Page 72: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

72

Always Thinking Ahead.

Real-time Color VRT with InSpace 4DEmotion 6 – Blocked Venous Stent

Courtesy Klinikum Nord,Nürnberg,Germany

Page 73: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

73

Always Thinking Ahead.

Real-time Color VRT with InSpace 4D

Page 74: CT Basics Principles of Spiral CT Dose - …dei-s2.dei.uminho.pt/outraslic/lebiom/seim/CTBasics_mod.pdfPrinciples of Spiral CT 30s - 40s Volume scanning in a single breath hold Raw

74

Always Thinking Ahead. Syngo Colonography