![Page 1: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/1.jpg)
![Page 2: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/2.jpg)
CT scan in
Radiotherapy
Planning
Ashraf Hamed Hassouna
Professor of Radiation Oncology
NCI, Cairo University
![Page 3: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/3.jpg)
Contents
Introduction
CT in RT planning
CT in IGRT
CT in ART
![Page 4: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/4.jpg)
Role of CT scan in RT planning & Treatment:
Patient contour.
Delineation of target volumes.
Delineation of OARs.
CT index and dose calculation.
RT plan evaluation and optimization.
Adaptive RT (ART).
Image Guided RT (IGRT).
![Page 5: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/5.jpg)
Simulation
![Page 6: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/6.jpg)
CT Simulator
![Page 7: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/7.jpg)
Immobilization
![Page 8: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/8.jpg)
![Page 9: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/9.jpg)
(1) Sidewall containing oblique and horizontal copper wires for CT-based measurement
of longitudinal stereotactic coordinate. (2) Longitudinal stereotactic scale. (3)
Stereotactic arc for lateral and AP coordinates. (4) Arc and scaled screw for diaphragm
control. (5) Level control. (6,7) SBF attached laser system (leg and trunk) for assistance
at patient repositioning. (8) Vacuum pillow.
ELEKTA stereotactic body frame (SBF)
![Page 10: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/10.jpg)
Stereotactic body frame
![Page 11: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/11.jpg)
CT number or Hounsfield units
The density of a structure is represented by CT
number or Hounsfield units (HU).
HU range from +1000 (bright white) to −1000
(dark black).
Dense bone (+1000), water (0), air (−1000).
![Page 12: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/12.jpg)
CT Simulator vs Conventional Simulator
Image patient anatomy and gross tumor,
slice by slice. These data can be processed
to view images in any plane or in 3D.
CT numbers can be correlated with tissue
density, allowing heterogeneity corrections
in treatment planning.
![Page 13: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/13.jpg)
CT Simulator vs Conventional CT
Accessories (e.g., flat table, laser light for positioning,
immobilization, and image registration devices, etc.):
accurately reproduce treatment conditions.
Wide aperture (e.g., 80 cm diameter): provide flexibility in
patient positioning for a variety of treatment setups.
CT image data set, with precise localization of patient
anatomy and tissue density information, is not only useful in
generating an accurate treatment plan but also providing a
reference for setting up treatment plan parameters.
![Page 14: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/14.jpg)
Image fusion with CT
MRI
PET
Angiography
Others
![Page 15: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/15.jpg)
Good CT Simulation
Positioning and fixation.
Complete patient contour.
Upper and lower tissue margin.
Contrast.
CT slice thickness.
Artifacts.
![Page 16: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/16.jpg)
CT Simulation can be used for:
2D
3DCRT
IMRT
SRS and SBRT
Brachytherapy
![Page 17: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/17.jpg)
Bone Metastases
![Page 18: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/18.jpg)
![Page 19: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/19.jpg)
![Page 20: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/20.jpg)
![Page 21: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/21.jpg)
Brain Metastases
![Page 22: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/22.jpg)
![Page 23: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/23.jpg)
![Page 24: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/24.jpg)
![Page 25: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/25.jpg)
![Page 26: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/26.jpg)
![Page 27: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/27.jpg)
Breast Cancer
![Page 28: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/28.jpg)
Wh
ole
Bre
ast
• CTV: based on wired breast tissue, limited 5 mm from skin and ant to pectoralis.
• PTVeval: CTV + 7mm, excluding heart, taken 5 mm from skin and ant to ribs/lung (include pectoralis).
![Page 29: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/29.jpg)
Lum
pec
tom
y
• GTV: seroma and clips.
• CTV: GTV + 1 cm, excluding pectoralis and taken 5 mm from skin.
• PTVeval: CTV + 7mm, excluding pectoralis, lung and taken 5 mm from skin.
![Page 30: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/30.jpg)
Accelerated Partial Breast
Irradiation
EBRT
![Page 31: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/31.jpg)
APBI
Multicatheter Interstitial Brachytherapy
![Page 32: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/32.jpg)
Mammosite
34Gy/10f/5d at 1 cm depth
![Page 33: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/33.jpg)
Brain Tumors
![Page 34: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/34.jpg)
![Page 35: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/35.jpg)
![Page 36: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/36.jpg)
![Page 37: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/37.jpg)
![Page 38: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/38.jpg)
![Page 39: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/39.jpg)
![Page 40: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/40.jpg)
![Page 41: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/41.jpg)
Head & Neck Tumors
![Page 42: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/42.jpg)
![Page 43: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/43.jpg)
![Page 44: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/44.jpg)
![Page 45: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/45.jpg)
![Page 46: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/46.jpg)
![Page 47: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/47.jpg)
![Page 48: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/48.jpg)
LN Contouring
![Page 49: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/49.jpg)
2D
![Page 50: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/50.jpg)
IMRT
![Page 51: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/51.jpg)
Cancer Cervix
![Page 52: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/52.jpg)
![Page 53: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/53.jpg)
![Page 54: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/54.jpg)
![Page 55: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/55.jpg)
![Page 56: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/56.jpg)
![Page 57: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/57.jpg)
![Page 58: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/58.jpg)
![Page 59: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/59.jpg)
![Page 60: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/60.jpg)
![Page 61: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/61.jpg)
![Page 62: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/62.jpg)
CRT
IMRT
![Page 63: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/63.jpg)
IMRT
![Page 64: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/64.jpg)
Brachytherapy
![Page 65: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/65.jpg)
![Page 66: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/66.jpg)
![Page 67: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/67.jpg)
![Page 68: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/68.jpg)
![Page 69: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/69.jpg)
![Page 70: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/70.jpg)
CT
artifacts
![Page 71: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/71.jpg)
Fletcher-Suit-Delclos-style Applicator Set
CT & MR Compatible
![Page 72: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/72.jpg)
SBRT
![Page 73: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/73.jpg)
Axial slice of HCC SBRT plan showing a conformal isodose distribution. The red and
blue solid lines represent GTV and PTV. The HCC was treated with 42 Gy in six
fractions (pink line), which tightly surrounds the PTV, with a rapid dose falloff.
![Page 74: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/74.jpg)
Typical RT plan with GTV and PTV delineated
by red and blue lines respectively
33Gy/6F
HCC with PVT
45Gy/5F
Recurrent HCC
after TACE
![Page 75: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/75.jpg)
Image Guided RadioTherapy
(IGRT)
![Page 76: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/76.jpg)
Portal Imaging
DRR EPI
![Page 77: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/77.jpg)
CBCT
![Page 78: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/78.jpg)
![Page 79: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/79.jpg)
Adaptive RadioTherapy
(ART)
![Page 80: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/80.jpg)
LN metastasis from NPC on CT before
RT
After 36 Gy/18 f LN shifted 1 cm medially
(body weight loss and tumor shrinkage)
![Page 81: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/81.jpg)
Dose distributions on CT-1
and the same plan
transferred onto CT-2. Most
of both parotid glands is
included in the 70 % dose
level (blue color).
Shrinkage of the neck
diameter and the LN is
evident on CT-2 obtained
after 38Gy/19f for NPC
patient.
CT-1 CT-2
![Page 82: CT scan in Radiotherapy Planning · accurately reproduce treatment conditions. Wide aperture (e.g., 80 cm diameter): provide flexibility in patient positioning for a variety of treatment](https://reader036.vdocument.in/reader036/viewer/2022071018/5fd2454e6f72d33fdb752e9c/html5/thumbnails/82.jpg)