D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

Upload: dwain-ford

Post on 03-Jan-2016

217 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Photo Detectors in High Energy Physics

Dieter Renker

Page 2: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Outline

Calorimeters:PMT, PIN diode, APD, G-APD

Ring image Čerenkov detectors:Gas detector, Multi anode PMT, Hybrid PMT

Time of flight measurements:PMT, MCP, G-APD

Tracker:Multi anode PMT, VLPC

Various detectors will be presented that involve photo sensors. The problems will be pointed out and the realized solutions which have been found, will be described.

Page 3: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Photomultiplier tubes in calorimetersPMTs can be made to cover large areas with up to 20 inch diameter (SuperKamiokande).

Shown here is the crystal ball with 3 inch tubes for the measurement of + 0+e++

Challenges are:• large area• low light yield (pure CsI)• fast response

+

Page 4: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Photomultiplier tubes in calorimeters

The amplification in the dynodes of a PMT has an extremely low level of noise. Summing over a large number of coincident PMT signals is therefore possible.

Shown here is the 800 liter LXe calorimeter with 800 PMTs in the + e++ experiment. The deposited energy is derived from the sum of all PMT signals and the position of the conversion from the distribution of the individual amplitudes.

100 liter prototype

Page 5: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

PMT – some remarks

PMT’s are a commercial product since 70 years. This is a long time for development and optimisation. Even so the progress during the last years is remarkable: the bulky shape turned into a flat design with very good effective area coverage and PMTs became position sensitive due to segmented (pixelised) anodes. Devices with segmented anodes with 256 pixels are available.In addition the prices came down considerably. Industry reduces the number of parts of the mechanically complicated dynode structure inside the vacuum container (some 40 pieces). To some extent the manual mounting of these parts is replaced by automated processes. The price per readout channel can be below 20 US$.

The quantum efficiency is relative low (~25%) but new photo cathodes can reach 40 to 50% (GaAsP and Ultrabialkali).

PMT’s with small transit time spread are available and allow a time resolution smaller than 100 ps.

PMT’s are very sensitive to magnetic fields. Only some proximity focused types tolerate axial fields.

Page 6: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

PIN photodiodes

General purpose detectors need magneticfields for the measurement of the momentum of charged particles. The PMTs have to be replaced by solid state devices.

CLEO pioneered the use of CsI(Tl) crystals and PIN photodiodes in an electromagnetic calorimeter (7800 Crystals and 4 diodes/crystal).

The QE of PIN photodiodes matches the emission wavelength (550 nm) of CsI(Tl) better than PMT’s. It is ~80%.

Consequently the energy resolution is very good: <2% for 1GeV ’s.

The PIN photodiode is a very successfuldevice – all B-factories use them and L3, GLAST …

Page 7: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

PIN photodiodes – problems

• PIN photodiodes have no gain. The operation is very stable but they need a charge sensitive amplifier which makes the signal rise time slow and introduces noise to the system (CPIN ~80 pF/cm2). Calorimeters made of materials with low light yield (pure CsI in KTeV and Čerenkov calorimeters with lead glass) cannot use PIN photodiodes.

• The full thickness of the PIN photodiodes (300 m) is sensitive. Charged particles (e.g. e+ and e-) which leak out at the rear end of the crystals and pass the diode produce an unwanted addition to the signal. A MIP creates some 100 electron-hole pairs per micron in silicon. This makes 30.000 electron-hole pairs which fake ~6 MeV additional energy in a CsI(Tl) calorimeter (Nuclear Counter Effect).

Page 8: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

PIN photodiodes – nuclear counter effect

High energy

Low energy

s

Each dot stands for an energy deposition of more than 10 keV 80 GeV e- beam in a 18 cm long PbWO4 crystal

Page 9: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Basic APD Structure (CMS version)

Photo-conversion electrons from the thin p-layer induce avalanche amplification at the p-n junction.

Electrons created by ionising particles traversing the bulk are not amplified.

deff ~ 6 m

50 times smaller than in a PIN diode.

Page 10: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

APDs in the CMS ECAL

PbWO4 crystal

36 supermodules with 1700 crystals each

2 APD’s/crystal

122.400 APD’s In the endcaps vacuum phototriodes are used because of the very high radiation levels.

Page 11: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

APD Impact on Energy Resolution

ECAL energy resolution:

CMS design goal : a ~ 3%, b ~ 0.5%, c ~ 200 MeV

APD contributions to:

a: photo statistics (area, QE) and avalanche fluctuations (excess noise factor)

b: stability (gain sensitivity to voltage and temperature variation,

aging and radiation damage) c: noise (low capacitance, serial resistance and dark current)

E

cb

E

a

EE

Page 12: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Sampling Calorimeter of KLEO

The KLEO calorimeter is made of 200 layers of scintillating fibers embedded in a lead absorber. For the readout fine mesh PMT’s are used on both ends of the 430 cm long fibers.

An excellent energy resolution has been achieved:

The time resolution of

allows the determination of K00’s vertices from the photon arrival time.

This is almost a homogeneous calorimeter because of the relative small amount of absorber material

E(GeV)5%/E

ps 50E(GeV)ps/ 54t

Page 13: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Sampling electromagnetic calorimeters – LHCb

PMT: Hamamatsu R7899-20

The Shashlik electromagnetic calorimeter of LHCb is a stack of scintillator- and lead tiles with wavelength shifting fibers perpendicular to the tiles and with PMT readout.

• high radiation environment

• good energy resolution

• fast response

• high dynamic range

Page 14: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Hadron Calorimeters

Hadron calorimeters are almost always built as sampling devices because to some extent they allow to make the response to electrons equal to the response to hadrons (e/h=1, compensation) which is crucial for the energy resolution. The best example is the ZEUS calorimeter with uranium- and scintillator-plates.

The energy sampled is typically a few percent of the total incident energy. A small number of photons have to be detected with good signal to noise ratio.

Up to now all sampling hadron calorimeters which are based on the detection of light from plastic scintillators use PMT’s.

Page 15: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Sampling Calorimeters – CMS HCAL with hybrid PMT’s

19 x 5.4mm 73 x 2.68mm

CMS diode design

CMS HCAL is made of brass, scintillator tiles, wavelength shifting fibers and proximity focused hybrid PMT’s.

• needs to work in a 4 T magnetic field• with large dynamic range,• high position sensitivity and• low crosstalk

Page 16: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Calorimeters with SiPM readout for ILC and T2K

Minical for the ILC:

11 layers of 3x3 plastic scintillator tiles (50x50x5 mm3) with 2 mm Fe in between. Readout with WLS fibers and SiPM’s.

Calibration with light from a LED (shaded area) and with MIP’s from 90Sr. <N> = 25 p.e.

Spectra (data and MC) of the 11 layers expressed in number of MIP’s for a 3 GeV incident e+ beam

V. Andreev et al.,NIM A 540 (2005) 368

Page 17: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Ring Image Čerenkov counters

RICH counter: measure photon impact point on the photon detector surface.Needed is detection of single photons with • large area coverage• good spatial resolution• high efficiency and good signal-to-noise ratio

Special requirements depend on the specific features of individual RICH counter:• Operation in (high) magnetic field• High rate capability• Very high spatial resolution• Excellent timing (time-of-arrival information and background reduction)

Page 18: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Rich in CLEO

CLEO Rich – the classical example• Available space limited• Operation in a magnetic field• Low material budget – in front of the calorimeter 13% of a radiation length

Solution:The photon detector is a wire chamber filled with methan and TEA (triethylamine) which has a QE of 30% at 150 nm.

Page 19: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Gas Photo Multiplier with CsI photo-cathodes

Alice: proximity focussing CsI RICH

• liquid C6F14 radiator• MWPC• cathode pads coated with CsI• total area 11 m2

similar in HADES, COMPASS, J-LAB …

Hadron Blind Detector of PHENIX

• triple Gas Electron Multiplier (GEM)• high QE• low sensitivity to charged particles• fast response (1.6 ns for single photons)

No ion and photon feedback

Page 20: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

DIRC: Detection of Internally Reflected Čerenkov Light

e-

e+

Quartz Barbox

Standoff box

Compensating coil

Support tube (Al)

Assembly flange

BABAR Rich requirements:

CsI(Tl) needs to detect photons down to 20 MeV small radiation length < 20%

small radial size required

/K separation at 4 GeV/c is 6.5 mrad

3 separation requires 2.2 mrad resolutionPMT’s: ETL 9125

Page 21: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

HERA-B RICH with multianode PMT’s

Multianode PMTs:R5900-M16 and R5900-M4

Requirements:

Rates ~1MHzLong term stabilityHigh QE over ~3m2

Gas based detectors could not be used because of the high rate environment

Page 22: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

RICH for LHCb with hybrid PMT‘s

• Large area (2.8m2) with high active area fraction• Fast compared to the 25 ns bunch crossing time • Have to operate in a small magnetic field• Granularity 2.5x2.5mm2

hybrid PMT with 5x demagnification

the anode is a pixel detector with 8192 channels organized in 1024 super-pixels of 500 x 500 m2 size.

Page 23: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Time of flight for particle identification

The plastic scintillators (3 m long, ~4x6 cm2) with fine mesh PMT’s on both sides of the TOF detector provide a time resolution of 100 ps and allow a /K separation up to 1.2 GeV.

The Aerogel Cerenkov Counter (ACC) extends the /K separation to 3.5 GeV. It also uses fine mesh PMT’s.

Page 24: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Tracking with scintillating fibers and VLPC‘s in DØ

• 80000 fibers 2.5 and 1.7 m long

• single photon detection capability

• 80% QE

• operated at 6°K

• gain ~50.000

• position resolution 120 m

Page 25: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Target Tracker of Opera

locate the lead/emulsion brick where a neutrino interaction occurred

Total area 3000 m2

Few photo electrons: 6 to 7 p.e. per MIP

64-channel Hamamatsu H7546 PMT

6700 x 2.6 x 1 cm3

Page 26: D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007 Photo Detectors in High Energy Physics Dieter Renker

D. Renker, PSI International Workshop on new photon-detectors, Kobe, 6/27/2007

Summary

A large variety of techniques of photo sensors has been developed and realized in present detectors for high energy physics. There is an almost infinite number of alternative designs which have been proposed and partly tested in prototypes.

The working horse is still the PMT when weak light flashes need to be detected.

New types of photo sensors have always quickly been adopted in high energy physics experiments. The new Geiger-mode avalanche photodiodes will for sure have a heavy impact on the design of future detectors.