Data Collection (SerialEM)Corey Hecksel
9/13/19
![Page 2: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/2.jpg)
Requirements of Data Collection Software?
• Image Viewer• Must communicate with camera
software
• Low Dose System • View, Focus/Track, Record
•Navigator System
•Optional: Scripting
![Page 3: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/3.jpg)
Data Collection Strategy?
Tomography Single Particle
![Page 4: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/4.jpg)
Data Collection Strategy?
Tomography
•Dose Symmetric
•Grouped Dose Symmetric
•Bi-directional
•Asymmetric Bi-directional
•Continuous rotation?
•And more…
Single Particle
• Image-shift data collection
•Tilted data collection
•Tilt pair
• Focal pairs
•And more…
![Page 5: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/5.jpg)
Benefits of SerialEM • Highly Adaptable -> Easy to Script
• Free to use (pay for support)
• Available across multiple microscopes -> large user community• Microscope: Thermo Fisher, JEOL, Hitachi• Detectors: Gatan, Thermo Fisher, DirectElectron, TVIPS, AMT, EMSIS, JEOL, STEM• Computer: Windows 2000, Windows XP, Windows 7
• Robust and Quick Bug/Feature Response
• Unified data collection software regardless of data collection strategy
![Page 6: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/6.jpg)
![Page 7: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/7.jpg)
Work Flow for Single Particle Data Collection
• Camera Settings
• Location of Regions to Image• Creation of Grid Map• Selection of Squares• Selection of Holes
• Data Collection and Monitoring• Low Dose
•Other Useful Features
![Page 8: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/8.jpg)
Camera Settings
• Camera Selection• Can use multiple cameras
• Imaging Mode• Presets for different modes ->
View/Focus/Record, etc.
• Readout area• What areas of the camera to readout• Binning
• Exposure Time
• Camera Settings• Gain Reference, Continuous vs Single Image,
Shutter Mode, Counting/Linear/Super-Resolution
• Fractions and Saving
![Page 9: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/9.jpg)
Work Flow for Single Particle Data Collection
• Camera Settings
• Location of Regions to Image• Creation of Grid Map• Selection of Squares• Selection of Holes
• Data Collection and Monitoring• Low Dose
•Other Useful Features
![Page 10: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/10.jpg)
Introduction to Navigator
•Used to navigate the grid, save positions, correlate images collected at different mags
• Set up batch positions
![Page 11: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/11.jpg)
Grid Atlas
• Full grid montage: LM 80-150x
• About 15-20 min acquisition time per montage
• Scripts can be used to montage all grids in the autoloader
• Find intact grid squares with good ice -> requires confirmation at higher mags
![Page 12: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/12.jpg)
Grid Square Selection
![Page 13: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/13.jpg)
Setting up Grid of Points
•Optional: Use Polygon tool to select area for hole selection
• Step 1: Place 5 points to define the geometry of the holes
• Step 2: Generate grid of points
• Step 3: Set all points to “Acquire”
![Page 14: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/14.jpg)
![Page 15: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/15.jpg)
Work Flow for Single Particle Data Collection
• Camera Settings
• Location of Regions to Image• Creation of Grid Map• Selection of Squares• Selection of Holes
• Data Collection and Monitoring• Low Dose
•Other Useful Features
![Page 16: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/16.jpg)
What is Low Dose?
•Data collection strategy used to minimize dose for dose sensitive samples
•Three Main States: View, Focus, and Record• View is used to align your target/hole• Focus is used to set/reset your defocus for collection, image shifted from your
Record area
• Record is used for data is collection
![Page 17: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/17.jpg)
Setting Up Low Dose•View
• Camera settings and defocus offset to produce high contrast images
• Choose a mag to give you slightly more than a single hole
• Focus/Trail: • Usually same imaging conditions as Record• Set off-set so that Record and and Focus/Trail do
not overlap
•Record:• Any additional beam shifts should be with respect
to record.
![Page 18: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/18.jpg)
Setting Up Low Dose: View Mode• Save a clean hole as a template for
hole finding• Should have uniform ice• No contamination• Not Empty
•You should align view mag to your grid square images• Find a piece of contamination in the
grid square
• Mark a point• “Align to Item” in view
![Page 19: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/19.jpg)
Setting Up Low Dose: View ModeSquare map View Mode
![Page 20: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/20.jpg)
Setting Up Low Dose: Record
•Choose mag based on imaging goals -> balance target resolution and number of particles
• Set image conditions based on total dose for record• For Talos: Adjust spot size and C2• For Krios: Adjust spot size, C2, and beam “intensity”
•Decide on defocus range• Higher defocus limits resolution but easier to see
• Image Shift Data Collection?
![Page 21: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/21.jpg)
Setting Up Low Dose: View and Record Mode
Record Mag View Mag
![Page 22: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/22.jpg)
Image Shift Data Collection
•Users can choose between:• Multi-shot per hole• Multiple holes• Combination
•Users define a image shift “template”
•Must make sure to calibrate “image shift vs. coma”
![Page 23: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/23.jpg)
Data Collection Scripts
1. Moves to the grid hole
2. Acquires view image and aligns to hole template
3. Focus
4. Acquire image
5. Repeat… many times
![Page 24: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/24.jpg)
Work Flow for Single Particle Data Collection
• Camera Settings
• Location of Regions to Image• Creation of Grid Map• Selection of Squares• Selection of Holes
• Data Collection and Monitoring• Low Dose
•Other Useful Features
![Page 25: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/25.jpg)
Other Useful Features
• SerialEM can measure and correct for astigmatism by fitting Thon rings in the FFT
•Coma can be visualized using a Zemlin Tableau
• SerialEM contains methods for automated coma free alignment
•Auto eucentricity and focus
![Page 26: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/26.jpg)
What about Tomography?
![Page 27: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/27.jpg)
What about Tomography?
•Requires an anchor map to accurately recall the exact position
•RealignToItem allows features to be accurately recalled
![Page 28: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/28.jpg)
What about Tomography?
•Very flexible -> Many options
•Can change values per tilt-series in batch collection mode
•Data collection times vary based on which parameters are used
![Page 29: Data Collection (SerialEM) - Stanford-SLAC Cryo-Electron … · 2019-09-11 · •Data collection strategy used to minimize dose for dose sensitive samples •Three Main States: View,](https://reader035.vdocument.in/reader035/viewer/2022070111/5f6813e3fc719f480f35d1e4/html5/thumbnails/29.jpg)
Common Problems
•Targeting is doing a bad job• Check Z height• View mode does not have enough contrast -> Thick Ice, Dose too low, etc. • Grid is bent• Hole template is bad
•Drift is really high• Turbo pump for autoloader is on• Increase delay after stage move• Carbon is broken• Poor clipping• Environmental Issues