DATA SHEET 1
Gamma Scintillation Detector
Counter:Source: 137Cs (1Ci)
Voltage: 1000V
Channel (LL) Counts Channel Counts0.2 1954 3.4 230.3 2354 3.5 130.4 19520.5 16030.6 1681
0.7 17430.8 22700.9 26141.0 23231.1 19371.2 15791.3 14591.4 13631.5 13051.6 12561.7 1270
1.8 13321.9 14172.0 11822.1 8292.2 4482.3 2702.4 2202.5 2072.6 2712.7 8082.8 39602.9 80973.0 58373.1 14083.2 1633.3 31
-
8/3/2019 Data Exp. 6 Ijatto
2/10
Graph 1
0
500
1000
1500
2000
2500
3000
3500
4000
4500
5000
5500
6000
6500
7000
7500
8000
8500
9000
0 0.5 1 1.5 2 2.5 3 3.5 4
Count/sample
Channel number
-
8/3/2019 Data Exp. 6 Ijatto
3/10
DATA SHEET 2
Gamma Scintillation Detector
Counter:
Source: 60Co
Voltage: 1000V
Channel (LL) Counts Channel Counts0.2 536 3.4 4210.3 803 3.5 4600.4 658 3.6 4530.5 549 3.7 4890.6 549 3.8 4750.7 568 3.9 4820.8 592 4.0 442
0.9 729 4.1 4231.0 888 4.2 4151.1 844 4.3 3541.2 676 4.4 3231.3 643 4.5 3211.4 563 4.6 3201.5 513 4.7 2901.6 473 4.8 3031.7 429 4.9 5311.8 422 5.0 9391.9 423 5.1 1159
2.0 389 5.2 7422.1 369 5.3 3082.2 407 5.4 1292.3 413 5.5 2122.4 427 5.6 4912.5 373 5.7 8072.6 372 5.8 8832.7 419 5.9 5112.8 387 6.0 2162.9 368 6.1 85
3.0 424 6.2 333.1 364 6.3 243.2 3943.3 426
-
8/3/2019 Data Exp. 6 Ijatto
4/10
Graph 2
0
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
0 1 2 3 4 5 6 7
Count/channel
Channel number
-
8/3/2019 Data Exp. 6 Ijatto
5/10
DATA SHEET 3
Gamma Scintillation Detector
Counter:
Source: unknown
Voltage: 1000V
Channel (LL) Counts Channel Counts0.2 4577 3.4 16580.3 4874 3.5 63230.4 4679 3.6 146390.5 4265 3.7 147130.6 4309 3.8 70620.7 3859 3.9 12970.8 4231 4.0 227
0.9 5970 4.1 911.0 5643 4.2 781.1 4799 4.3 1021.2 3995 4.4 1031.3 3507 4.5 881.4 3339 4.6 731.5 3143 4.7 681.6 30591.7 29291.8 29271.9 2871
2.0 28572.1 28812.2 30612.3 30142.4 29562.5 31892.6 32462.7 26682.8 18572.9 1138
3.0 7263.1 5813.2 4923.3 679
-
8/3/2019 Data Exp. 6 Ijatto
6/10
Graph 3
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
11000
12000
13000
14000
15000
16000
17000
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
Count/Channel
Channel Number
-
8/3/2019 Data Exp. 6 Ijatto
7/10
Energy Calibration Curve
Event (KeV) Disc. Setting (LL)(V)
Photopeak 661.6 2.9
Photopeak 1173.2 5.1
Photopeak 1332.5 5.8
Graph 4
0
200
400
600
800
1000
1200
1400
0 1 2 3 4 5 6 7
Energy(eV)
Channel Number
3.7
830
-
8/3/2019 Data Exp. 6 Ijatto
8/10
From data sheet 3 (unknown source), we know that the energy peak is at 3.7 (channel number).
When we plot the point at energy calibration curve of the spectrometer, we got the energy of the
unknown source which is 830 eV.
Calculation of Energy Resolution of NaI(TI) Detector
Using formula (1) from graph 1 the resolution, R is
= 0.25
= 8.62%
From graph 2 there are two resolutions because of two peaks
= 0.35 = 0.30
= 6.86% = 5.17%
From graph 3 the resolution, R is
= 0.30
= 8.12%
-
8/3/2019 Data Exp. 6 Ijatto
9/10
Discussion
In this experiment we will study the radioactive decay of a nucleus by detecting gamma rays
emitted consequent to the decay. Gamma ray detection is a slightly complicated, multi-step
process: the gamma ray enters a NaI(Tl) scintillator crystal where it produces a rapidly moving
free electron that, in turn, loses its energy by excitation of the ions in its path as it travels through
the crystal. This excitation energy is given off in various ways, one them being emission of
visible light (fluorescence). Thus a single high energy gamma ray entering the scintillator
produces a flash of low energy photons. These photons are directed to the photosensitive surface
of a photomultiplier tube, where they eject electrons via the photoelectric effect. The electrons
are collected in the photomultiplier and amplified to yield a current pulse, which is converted to
a voltage pulse whose height is proportional to the number of photoelectrons and is thus
proportional to the number of photons reaching the tube, which in turn is proportional to the
initial energy of the fast electron.
When a radioactive source is placed near the scintillator, the photomultiplier produces a series of
pulses, each corresponding to the decay of a single nucleus. The amplitude of each pulse is
related to the energy of the electron freed by the gamma ray. These pulses are studied using
either a single- or multi-channel analyzer. A single channel analyzer (SCA) counts on the
number of voltage pulses whose height falls within a given (adjustable) window of values, while
a multi-channel analyzer (MCA) sorts the pulses according to height and the counts the number
in each window to give a spectral (energy) distribution of the fast electrons. Figure 4 shows a
typical MCA spectrum. In order to relate this spectrum to the nuclear decay, we need to
understand how gamma rays interact with matter.When entering a crystal, gamma rays produce fast charged electrons by three different processes:
the photoelectric effect, the Compton effect (Compton scattering) and pair production. It is these
fast electrons, which give rise to scintillations, not the gamma ray. The observed spectral
distribution will thus depend on the detailed interaction process of the gamma rays in the crystal.
Consider a beam of mono-energetic gamma rays striking the scintillator. For our purposes the
most important energy loss mechanism is the photoelectric effect. When a gamma ray strikes an
ion in the crystal, it is absorbed and all of its energy is transferred to one of the bound electrons,
which is freed and moves rapidly through the crystal. Since the energy of the gamma ray
(typically about 0.5 MeV) is much greater than the binding energy of the electron of the ion
(typically 10 to 100 eV), the energy of the freed electron can be considered equal to that of the
incoming gamma. (Especially since the energy resolution of the detector is only about 10%.)
Thus the photoelectric effect results in a peak, called the photopeak, in the photomultiplier
spectrum at an energy equal to that of the incoming gamma ray.
-
8/3/2019 Data Exp. 6 Ijatto
10/10
From graph 1 we got the peak value is at 2.9 (channel value) while from the graph 2 we got the
peak value at 5.1 and 5.8. When we draw a graph of vs discriminator setting using this value
we got a linear graph. This is known as the energy calibration curve of the spectrometer (see
figure 6). We use this curve to find the gamma ray energy of an unknown source by reassuring
is spectrum under the same condition. From the graph 3 which is using unknown source, we got
the peak value at 3.7 (channel number). When we plot this value at graph 4, we got the energy is
about 830 eV. According to the unknown sources peak energy is830 keV and based on the
graph 4 it is the Mn54 and its half-life is 312.2
A NaI(Tl) detector has an energy resolution of only about 10%. When a beam of mono-
energetic gamma rays strikes the scintillator, there is a fluctuation from gamma ray to gamma ray
in the height of the voltage pulse from the photomultiplier, which shows up as a broadening of
the photopeak. The pulse height variation is chiefly due to statistical fluctuations in the number
of electrons emitted at the cathode of the photomultiplier when a flash of photons arrives from
the scintillator, but is also due to the occasional escape of electrons, X-rays or gamma rays fromthe crystal, all of which depend on how large the NaI crystal is. The fractional full width of a
peak at half its maximum height (FWHM) is a convenient measure of the resolution of the
instrument. From the calculation of the energy resolution, R at each graph we got the percent is
below than 10% which means the ability to resolve two peaks that are fairly close together in
energy is high for graph 1 and 3 and low for graph 2.
Reference
AN34: Experiments in Nuclear Science, 3rd Ed., EG & G Ortec. Experiment 1, Basic
identifications in Electronic Measurement Systems, pp 1-7; Experiment 3, Gamma Spectroscopy
Using NaI, pp 15-24; Linear and Logic Signals in EG & G Ortec NIM Instruments, pp159-160
D.W. Preston & E.R. Dietz: The Art of Experimental Physics, Appendix B Counting and
Sorting Particles: The Scintillation Counter, pp376-385
general theory
Serway, Moses and Moyer: Modern Physics, pp 431-432 (scintillator/photomultiplier), 389-
392 (beta and gamma decay)
Beiser: Concepts of Modern Physics, pp 443, 471-473 (gamma decay)
Halliday: Introductory Nuclear Physics
Siegbahn: Alpha, Beta and Gamma Ray Spectroscopy