Laser Visible red, class 2, 690 nm
Standoff and depth of FOV 170 ± 30 mm
Width of FOV user selectable 24, 60, 124 mm
Line per second (max) 53 Hz
Temperature range of sensorfor declared accuracy 15 ~ 32 °C
Sensor size L x W x H 134 x 72 x 60,5 mm
Sensor weight 382 g
Probing Error (*) MPEPF(OT) 60 µm
TemperatureCompensation
PerformanceEnvironment
Optional CLIMA CLIMA CLIMA
Standard: 18 ÷ 22 °C
Option CLIMA: 16 ÷ 26 °C
Standard: 18 ÷ 22 °C
CLIMA: 16 ÷ 26 °C
Standard: 18 ÷ 22 °C
CLIMA: 16 ÷ 26 °C
Probe Configurationfor performance test
• HP-S-X3C/X5: stylus length 60 mm, tip diameter 5 mm • HP-S-X1C: stylus length 20 mm, tip diameter 5 mm
• HP-S-X1S: without extension, stylus length 50 mm, tip diameter 5 mm• HP-S-X1H: with an extension of 100 mm, stylus length 20 mm, tip diameter 5 mm
• TP200: Standard force module, stylus length 10 mm, tip diameter 4 mm• HP-T: Standard force, stylus length 10 mm, tip diameter 4 mm
• HP-TM: Standard Force Module, stylus length 10 mm, tip diameter 4 mm
Performance data are valid if the following
specifications are met
Standard Temperature Range 18 ÷ 22 °C
Max. air temperature variation 1 °C/h - 2 °C/24h
Max. Spatial Gradient 1 °C/m
CLIMA Temperature Range 16 ÷ 26 °C
Max. air temperature variation 1 °C/h - 5 °C/24h
Max. Spatial Gradient 1 °C/m
Max. Scanning Speed
300 mm/s 300 mm/s 500 mm/s
(*) According to VDI 2617 Part 6.2, October 2005 edition
HP-L-1.6T Laser ProbeTechnical Characteristics
Technical Characteristics
X: Micromachined anodized light alloy extrusion
Y: Integral dovetail guideways, machined into the table
Z: Micromachined anodized light alloy extrusion
Mechanical Frame:
Sliding System: Air bearings on all axes
Measuring System: METALLUR® linear scales. System Resolution: 0.039 μm
Surface Plate: Material: Granite; Part Locking: threaded inserts M8 x 1.25; Flatness: according to DIN 876/III
Temperature Compensation: CLIMA: Multisensor Temperature Compensation Technology, 16 ÷ 26 °C (optional on Classic models)
Ram Counterbalance: Pneumatic, adjustable
Air Supply: Minimum air supply pressure: 0.5 MPa
Air consumption: 150 Nl/min
Environment: Metrological specification temperature range: Standard: 18 ÷ 22 °C and CLIMA: 16 ÷ 26 °C
Operating Temperature: 10 ÷ 40 °C
Relative humidity: 20% - 90 % non condensing
Probe Heads and Sensors: HH-A, HH-AS8, HP-S-X1C, HP-S-X3C, HP-S-X5
TP200, HP-T, HP-TM, HP-S-X1, HP-L-10.6T
Classic Performance Advantage Performance Advantage
DEA GLOBAL Silver 09.xx.08 DEA GLOBAL Silver 12.xx.10
516 516 866 433 866
1000
3 DEA GLOBAL SILVER 09.xx.08 - 12.xx.10 | DS CMM 036/01 - EN - 07/14
Performance Verification
Dynamics:
Max. 3D Speed(mm/s)
Max. 3D Accel.(mm/s2) 1700
17001400 (with HP-S-X3C)
43001400 (with HP-S-X3C/X5)
43001400 (with HP-S-X3C/X5)
5 gauges have to be measured 3 times with one probing at each end, in 7 different directions. All measuring results must be within E0.
E0,MPE : maximum permissible error of length measurement
E150,MPE: maximum permissible error of length
R0,MPL: maximum permissible limit of the repeatability range
5 length gauges have to be measured 3 times in the YZ- or XZ plane with opposite styli, mounted 150 mm off the Z spindle axis.
Extreme value of the repeatabi-lity range of the length measu-rement error, calculated by three repeated measurements on each size for a total of 35 values. The 35 repeatability range results must be within the specified R0,MPL.
A precision sphere has to be measured with 25 probings. PFTU is the range of all radii.PFTU = Rmax - Rmin = sphere form. Calibration sphere in the middle of the volume.
PFTU,MPE : maximum permissible single stylus form error
MPETHP/t: maximum permissible scanning probing error
A precision sphere has to be scanned with 4 defined lines. THP is the range of all radii.THP = Rmax - Rmin = sphere form, scanning. Calibration sphere in the middle of the volume.
HP-S-X3C
HH-A, HH-AS/HP-S-X1S, HP-S-1H
HH-A, HH-AS/TP200
HH-A, HH-AS/HP-T, HP-TM
HP-S-X1C
HH-A, HH-AS/HP-S-X1S, HP-S-1H
HH-A, HH-AS/HP-T, HP-TM, TP200
E0,MPE/ E150,MPE PFTU,MPE MPETHP/t
GLOBAL Silver Classic
GLOBAL Silver Performance
E0,MPE/ E150,MPE PFTU,MPE MPETHP/t
RO,MPL
SY1 SY2
PY
LY
Z 80
0LZ 2
946
Y
LX 1450
PH
700
DZ
1 94
6
230
DZ
117
LX1 1340
X 800
860160
6066
156 D1
275
140
175
175
D1
15614
010
0 175
175
8080
111
HH-A HH-AS8
09.15.08
09.12.0809.20.08
80
HP-S-X1C HP-S-X3C
80
101
09.12.08
09.15.08
09.20.08
2165
2465
2965
1300
1500
1800
1650
1850
2250
Y LY
455
540
680
1200
1500
2000
1910
2210
2710
PY SY1 SY2 D1
1000
1130
1350
DEA GLOBAL Silver Classic and Performance 09.12.08, 09.15.08, 09.20.08Specifications according to ISO 10360-2:2009, ISO 10360-5:2010 and ISO 10360-4:2000
4 DEA GLOBAL SILVER 09.xx.08 - 12.xx.10 | DS CMM 036/01 - EN - 07/14
Models
Strokes
(mm)
Overall Dimen-
sions(mm)
Supports
(mm)
Max. Part
Weight(kg)
Surface Plate
(mm)
CMMWeight
(kg)
100.6
24.4
50.6
ThdInsert
(mm)
32.4
56.4
68.5
92.5
1.8 + L/333
1.8 + L/333
1.9 + L/333
2.1 + L/333
1.8
1.8
1.9
2.0
2.9/45
2.9/45
–
–
101.
5
Probe Configuration
Standard Temperature Range18 ÷ 22 °C
CLIMA Temperature Range16 ÷ 26 °C
2.2 + L/300
2.2 + L/300
2.7 + L/300
2.2
2.2
2.7
3.9/68
3.9/68
–
2.1 + L/250
2.1 + L/250
2.1 + L/250
2.4 + L/250
1.8
1.8
1.9
2.0
2.5 + L/250
2.5 + L/250
2.8 + L/250
2.2
2.2
2.7
2.9/45
2.9/45
–
–
3.9/68
3.9/68
–
1.7
1.7
2.1
2.1
2.0
2.0
2.7
75.4
NOTE: E0, MPE/E150, MPE specifications are to be formally understood as E0,MPE* /E150,MPE* for the case where non-normal CTE material calibrated test lengths are used.
NOTE: ISO 10360-2 test with maximum part weight performed as an option upon request only.
E0,MPE/ E150,MPE PFTU,MPE MPETHP/t E0,MPE/ E150,MPE PFTU,MPE MPETHP/t
RO,MPL
SY1
PH
SY2
PY
LY
Z
LZ
Y
111
HH-A
HH-AS8 HP-S-X3C
88
200
D1
267
146
175
175
D1196
146
175
175
09.15.08 09.12.0809.20.08
LX
DZ
1
PZ11
7
LX1
X
910200
5625
200
2148
6970
88
8888
HP-S-X5
196
09.12.08
09.15.08
09.20.08
900
900
900
1538
1538
1538
1457
1457
1457
2430
2730
3230
3066
3066
3091
936
936
936
1300
1500
1800
2286
2567
3316
X Y Z(a) LX LX1 LY LZ
700
700
725
PH
550
600
700
1100
1200
1450
1200
1500
2000
800
800
800
2030
2330
2830
PY DZ1 SY1 SY2
76
51
126
D1
250
250
275
PZ
101
HP-S-X3C/X5
HH-A, HH-AS/HP-S-X1S, HP-S-1H
HH-A, HH-AS/TP200
HH-A, HH-AS/HP-T, HP-TM
DEA GLOBAL Silver Advantage 09.12.08, 09.15.08, 09.20.08Specifications according to ISO 10360-2:2009, ISO 10360-5:2010 and ISO 10360-4:2000
5 DEA GLOBAL SILVER 09.xx.08 - 12.xx.10 | DS CMM 036/01 - EN - 07/14
Models
Strokes
(mm)
Overall Dimensions
(mm)
Supports
(mm)Max. Part
Weight(kg)
Surface Plate
(mm)
Day-lights(mm)
CMMWeight
(kg)
Thd Insert(mm)
68.5
92.5
32.4
56.4
75.4
91.5
(a) With HP-S-X5 probe head, Z stroke = 665 mmNOTE: ISO 10360-2 test with maximum part weight performed as an option upon request only.
Standard Temperature Range18 ÷ 22 °C
CLIMA Temperature Range16 ÷ 26 °C
1.4 + L/333
1.4 + L/333
1.9 + L/333
2.1 + L/333
1.4
1.4
1.9
2.0
2.5/45
2.5/45
–
–
1.7 + L/250
1.7 + L/250
2.1 + L/250
2.4 + L/250
1.4
1.4
1.9
2.0
2.5/45
2.5/45
–
–
1.2
1.2
2.1
2.1
Probe Configuration
191.
5
NOTE: E0, MPE/E150, MPE specifications are to be formally understood as E0,MPE* /E150,MPE* for the case where non-normal CTE material calibrated test lengths are used.
SY1
PH
SY2
PY
LY
Z
LZ
Y
88
111
HH-A
88
HH-AS8 HP-S-X3C
88
101
273D1
146
175
175
D1273
146
175
175
12.30.1012.15.1012.22.10
LX
DZ
1PZ
150
1757
X
1210200
5625
200 200
70 69
LZ1
12.15.10
12.22.10
12.30.10
1200
1200
1200
1838
1838
1838
2880
3580
4380
3431
3406
3431
2307
2282
2307
1173
1173
1173
1800
2250
2250
3792
5679
7587
X Y Z LX LY LZ LZ1
625
600
625
PH
650
800
1000
1300
1630
1980
1500
2200
3000
1000
1000
1000
2480
3180
3980
PY DZ1 SY1 SY2
149
149
24
D1
290
350
380
PZ
E0,MPE/ E150,MPE PFTU,MPE MPETHP/t E0,MPE/ E150,MPE PFTU,MPE MPETHP/t
RO,MPL
HP-S-X3C
HH-A, HH-AS/HP-S-X1S, HP-S-1H
HH-A, HH-AS/TP200
HH-A, HH-AS/HP-T, HP-TM
DEA GLOBAL Silver Performance 12.15.10, 12.22.10, 12.30.10Specifications according to ISO 10360-2:2009, ISO 10360-5:2010 and ISO 10360-4:2000
68.5
92.5
32.4
56.4
75.4
66.5
66.5
7 DEA GLOBAL SILVER 09.xx.08 - 12.xx.10 | DS CMM 036/01 - EN - 07/14
Models
Strokes
(mm)
Overall Dimensions
(mm)
Supports
(mm)Max. Part
Weight(kg)
Surface Plate
(mm)
Day-lights(mm)
CMMWeight
(kg)
Thd Insert(mm)
Probe Configuration
Standard Temperature Range18 ÷ 22 °C
CLIMA Temperature Range16 ÷ 26 °C
2.4 + L/333
2.4 + L/333
2.5 + L/333
2.7 + L/333
2.4
2.4
2.5
2.7
4.7/45
4.7/45
–
–
2.8 + L/200
2.8 + L/200
2.8 + L/200
3.1 + L/200
2.4
2.4
2.5
2.7
4.7/45
4.7/45
–
–
1.9
1.9
2.7
2.7
NOTE: E0, MPE/E150, MPE specifications are to be formally understood as E0,MPE* /E150,MPE* for the case where non-normal CTE material calibrated test lengths are used.
NOTE: ISO 10360-2 test with maximum part weight performed as an option upon request only.
SY1
PH
SY2PY
LY
Z
LZ
Y
88
111
88
88
101
273D1
146
175
175
D1273
146
175
175
12.30.1012.15.1012.22.10
LX
DZ
1
PZ15
0
1757
X
1210200
5625
200 200
70 69
LZ1
12.15.10
12.22.10
12.30.10
1200
1200
1200
1838
1838
1838
2880
3580
4380
3431
3406
3431
2307
2282
2307
1173
1173
1173
1800
2250
2250
3792
5679
7587
X Y Z (a) LX LY LZ LZ1
625
600
625
PH
650
800
1000
1300
1630
1980
1500
2200
3000
1000
1000
1000
2480
3180
3980
PY DZ1 SY1 SY2
149
149
24
D1
290
350
380
PZ
HH-A
HH-AS8 HP-S-X3C
HP-S-X5
88
E0,MPE/ E150,MPE PFTU,MPE MPETHP/t E0,MPE/ E150,MPE PFTU,MPE MPETHP/t
RO,MPL
HP-S-X3C/H5
HH-A, HH-AS/HP-S-X1S, HP-S-1H
HH-A, HH-AS/TP200
HH-A, HH-AS/HP-T, HP-TM
DEA GLOBAL Silver Advantage 12.15.10, 12.22.10, 12.30.10Specifications according to ISO 10360-2:2009, ISO 10360-5:2010 and ISO 10360-4:2000
68.5
92.5
32.4
56.4
75.4
66.5
66.5
191.
5
(a) With HP-S-X5 probe head, Z stroke =905 mm, NOTE: ISO 10360-2 test with maximum part weight performed as an option upon request only.
8 DEA GLOBAL SILVER 09.xx.08 - 12.xx.10 | DS CMM 036/01 - EN - 07/14
Models
Strokes
(mm)
Overall Dimensions
(mm)
Supports
(mm)Max. Part
Weight(kg)
Surface Plate
(mm)
Day-lights(mm)
CMMWeight
(kg)
Thd Insert(mm)
Probe Configuration
Standard Temperature Range18 ÷ 22 °C
CLIMA Temperature Range16 ÷ 26 °C
2.1 + L/333
2.1 + L/333
2.5 + L/333
2.7 + L/333
1.8
1.8
2.2
2.6
3.1/45
3.1/45
–
–
2.5 + L/200
2.5 + L/200
2.8 + L/200
3.1 + L/200
1.8
1.8
2.2
2.6
3.1/45
3.1/45
–
–
1.7
1.7
2.7
2.7
NOTE: E0, MPE/E150, MPE specifications are to be formally understood as E0,MPE* /E150,MPE* for the case where non-normal CTE material calibrated test lengths are used.
7731
79
PORTABLE MEASURING ARMSLASER TRACKERS & STATIONS
WHITE LIGHT SCANNERS
BRIDGE CMMS
ULTRA HIGH ACCURACY CMMSGANTRY CMMS MULTISENSOR & OPTICAL SYSTEMS
HORIZONTAL ARM CMMS
PRECISION MEASURING INSTRUMENTS
SOFTWARE SOLUTIONS
Hexagon Metrology offers a comprehensive range of products and services for all industrial metrology applications in sectors such as automotive, aerospace, energy and medical. We sup-port our customers with actionable measurement information along the complete life cycle of a product – from development and design to production, assembly and final inspection.
With more than 20 production facilities and 70 Precision Centers for service and demonstrations, and a network of over 100 distribution partners on five continents, we empower our customers to fully control their manufacturing processes, enhancing the quality of products and increasing efficiency in manufacturing plants around the world.
For more information, visit www.hexagonmetrology.com
Hexagon Metrology is part of Hexagon (Nordic exchange: HEXA B). Hexagon is a leading global provider of design, measurement and visualisation technologies that enable customers to design, measure and position objects, and process and present data.
Learn more at www.hexagon.com
© 2014 Hexagon Metrology. Part of Hexagon
All rights reserved. Due to continuing product development, Hexagon Metrology reserves the right to change product specifications without prior notice.
Printed in Germany. July 2014
SENSORS