For 24 Hour AssistanceCall 1-877-566-9466
Part No. 20751 Rev-012014-08
Pocket Guide Series 3 software
(Delivery System)
Automated Pre-Use ProcedureIntegrated Pneumatic Backup INOMAX® Delivery
Transport Regulator/Cap AssemblyOxygen Dilution Chart
INOMAX Cylinder Duration ChartCircuit Connection DiagramsChanging INOMAX Cylinders
High Calibration Connection DiagramsINOmax DSIR Disposable Adapters
2Part No. 20751 Rev-012014-08
IMPORTANT: This guide is provided as a convenience and for general information only. Do not use this product without clearly and thoroughly understanding the most recent revision of the INOmax DSIR® Operation Manual. The Operation
contained in this guide.
3 Part No. 20751 Rev-012014-08
Mallinckrodt, the “M” brand mark and the Mallinckrodt Pharmaceuticals logo are trademark of a Mallinckrodt company. Other brands are trademarks of a Mallinckrodt company or their respective owner. ©2016 Mallinckrodt
4Part No. 20751 Rev-012014-08
Contents
Automated Pre-Use Checkout ......................................................................5Integrated Pneumatic Backup INOMAX Delivery .........................................8Transport Regulator/Cap Assembly ............................................................10Oxygen Dilution Chart ................................................................................14INOMAX Cylinder 88-size Duration Chart ..................................................15INOMAX Cylinder D-size Duration Chart....................................................16Connection to Various Breathing Systems .................................................17 Acutronics Medical Systems AG Fabian +nCPAP Evolution ............... 18 Acutronics Medical Systems AG Fabian HFO ..................................... 19 A-Plus Medical Babi-Plus Bubble CPAP.............................................. 20 Bagging Systems While Using the Injector Module ..............................22 Bunnell Life Pulse High Frequency Ventilator Circuit ...........................24 Connecting INOmax DSIR Sample Tee to the Bunnell Life Pulse Circuit ....................................................................................28 Connecting INOmax DSIR Injector Module to the Bunnell Life Pulse Circuit ....................................................................................28 Circle Anesthesia System .....................................................................29
Löwenstein Leoni-plus Ventilator ..........................................................30
...........................................31 ....................32
........................33 Hamilton Arabella Nasal CPAP .............................................................34 ICU Ventilator Circuit ............................................................................35 Sensormedics 3100A/B High Frequency Oscillatory Ventilator with a Filtered Circuit .......................................................................36 Sensormedics 3100A/B High Frequency Oscillatory Ventilator with a Rigid or Flexible Circuit .........................................................38 SLE Life Support SLE5000 ...................................................................39 Spontaneous Breathing Patient on a Mask Circuit ...............................40 Spontaneous Breathing Patient on a Nasal Cannula ...........................41
............................42 Transport Ventilator Circuit ...................................................................43 Vapotherm 2000i ..................................................................................46 Vapotherm Precision Flow ....................................................................47 Viasys Infant Flow CPAP System; Cardinal Airlife nCPAP System...............48 Viasys Infant Flow SiPAP .....................................................................49INOblender Circuit Connection ...................................................................50
...............................51Changing INOMAX Cylinders .....................................................................52High Calibration Connection Diagrams ......................................................56INOmax DSIR Disposable Adapters ............................................................58
5 Part No. 20751 Rev-012014-08
Automated Pre-Use Checkout1. Turn INOmax DSIR ON, verify speaker function. Note: A low range calibration automatically starts
following the self test.
A Pre-Use wizard will be displayed on the main screen, which will provide step-by-step instructions to complete the automated Pre-Use procedure.
Pre-Use Checkout
- Pressing the NEXT button initiates the Pre-Use wizard.- Pressing the CANCEL button exits the Pre-Use
wizard. If you cancel the pre-use wizard, the manual pre-use checkout procedure can be found on the INOmax DSIR Plus Pre-Use card, or in the INOmax DSIR Operation Manual, section 9/Appendix.
The Pre-Use wizard can also be initiated by entering the menu screen and selecting the Pre-Use Checkout button.
6Part No. 20751 Rev-012014-08
Pre-Use Checkout
7 Part No. 20751 Rev-012014-08
Pre-Use Checkout
(Intentionally left blank)
8Part No. 20751 Rev-012014-08
Note: Use the integrated pneumatic backup function only for a short time, until a replacement delivery system can be obtained. The INOblender can also be used as a backup.
If the main delivery system fails, the INOmax DSIR has an integrated pneumatic backup delivery function that allows the patient to remain connected to the ventilator. Backup NO delivery does not rely on the operation of the main system. The INOmax DSIR backup function: · Uses a pneumatic on/off switch and a restrictor built into the delivery side of the system.
WARNING:When using the integrated pneumatic backup with
dose will be approximately 40 ppm. Breathing circuit
greater than 40 ppm.
The table below indicates the nominal concentrations
Integrated Pneumatic Backup INOMAX Delivery
10 20
NO Concentration 40 27 20 13 10
9 Part No. 20751 Rev-012014-08
The main screen: · Indicates that backup delivery is on and that the set dose is turned off.
· Displays the estimated dose that the patient should be
· Displays the NO concentration table Note: If the injector module is not functioning, the estimated
backup dose graphic will be inactive.
Backup delivery mode OFF.
Integrated Pneumatic Backup INOMAX Delivery
10Part No. 20751 Rev-012014-08
Cap Assembly
Figure 1 Figure 2
WARNING:• A new INOMAX cylinder and regulator must be purged
before use to ensure the patient does not receive an excess level of NO2.
IR and the INOMAX cylinder for more than one hour will result in interruption of INOMAX delivery.
Caution: When using the Transport Regulator/Cap Assembly (PN 10022) ensure the cap is in place on the cylinder and the infrared cable is connected to the infrared connector port on the back of the INOmax DSIR.
Note: Check the INOMAX cylinder for the correct product identity labels, cylinder concentration and expiration date. Ensure the INOMAX gas cylinder has more than 500 psig.
Step OneNote: Ensure the white plastic tip is in place on the regulator
connector and not chipped or cracked (see Figure 2).
Connect a high pressure regulator to an INOMAX cylinder and tighten
INOMAX cylinder (see Figure 1).
11 Part No. 20751 Rev-012014-08
Cap Assembly
Figure 3
Figure 4
Step TwoConnect the INOMAX regulator hose to one of the INOMAX inlets on the back of the INOmax DSIR (see Figure 3).
Step ThreeConnect the Infrared cable from the Transport Regulator/Cap Assembly to the back of the INOmax DSIR (see Figure 4).
12Part No. 20751 Rev-012014-08
iButton
The electrical cord exits the cap directly above the iButton keyway
Figure 5 Figure 6
Figure 7 Figure 8
Cap Assembly
Step FourPlace the Cap Assembly over the INOmeter (see Figure 5).Note: Be sure to align the keyway inside the Cap Assembly with the
iButton on the INOmeter (see Figure 5 and 6).
Step FiveGrasp the Cap Assembly to open cylinder valve (see Figure 7 and 8).
13 Part No. 20751 Rev-012014-08
Figure 9
Cap Assembly
Final Set-up DiagramThe following diagram and photo illustrates all of the components connected.
Additional InformationCommunication will take place between the INOmax DSIR and the INOmeter after the boot up phase of the INOmax DSIR is complete.
Note: Cylinder icons are not visible and the dose control button will remain inactive until the INOmax DSIR recognizes an INOMAX cylinder.
Note: When using the Transport Regulator/Cap Assembly only one cylinder will be displayed (see Figure 9).
Proceed with the INOmax DSIR Pre-Use Checkout(see page 6)
14Part No. 20751 Rev-012014-08
Oxygen Dilution Chart
For delivery with 800 ppm cylinder of
Caution FiO2 less than 21%
Please Note:The calculations on this chart have been determined based on an 800 ppm cylinder of INOMAX (nitric oxide) for inhalation.This chart is representative of a range of doses available on the INOmax DSIR and doses higher than 20 ppm are not intended as the recommended therapeutic dose.Calculations are considered estimates and may vary under clinical conditions. All numbers have been rounded to the nearest hundredth.
0.21
0.20
0.20
0.19
0.40
0.39
0.38
0.36
0.79
0.78
0.76
0.72
.21 .40 .60 .80
Set FiO2
10
20
40
80
1.00
0.99
0.98
0.90
Actual FiO2
!
!
!
INO
MA
X D
ose
15 Part No. 20751 Rev-012014-08
For an 88-Size 800 ppm Cylinder Concentration*
* All calculations in the table above are based on a full cylinder, 138 bar (2000 psig), 1963 liter “88” cylinder, with a cylinder change at 14 bar (200 psig). The
a cylinder conversion factor of 14.2 liters per bar/0.98 liters per psig.
Calculations are considered estimates and may vary under clinical circumstances. For more information, call 1-877-KNOW-INO (1-877-566-9466).
INOMAX Cylinder 88-Size
Duration Chart
This chart is representative of a range of doses available on the INOmax DSIR and doses higher than 20 ppm are not intended as the recommended therapeutic dose.
5 39 Days 19.5 Days 9.8 Days 4.9 Days
10 19.4 Days 9.7 Days 4.8 Days 2.4 Days
20 9.6 Days 4.8 Days 2.4 Days 1.2 Days
40 4.7 Days 2.3 Days 1.2 Days 14 Hours
80 2.2 Days 1.1 Days 13.3 Hours 6.6 Hours
16Part No. 20751 Rev-012014-08
Duration Chart
* All calculations in the table above are based on a full cylinder, 138 bar (2000 psig), 353 liter “D” cylinder, with a cylinder change at 14 bar (200 psig). The
and a cylinder conversion factor of 2.6 liters per bar/0.18 liters per psig.
Calculations are considered estimates and may vary under clinical circumstances. For more information, call 1-877-KNOW-INO (1-877-566-9466).
This chart is representative of a range of doses available on the INOmax DSIR and doses higher than 20 ppm are not intended as the recommended therapeutic dose.
INOMAX Cylinder D-Size
Typically used in transport
5 L/min 10 L/min 20 L/min 40 L/min
5 7.0 Days 3.5 Days 1.8 Days 21 Hours
10 3.5 Days 1.7 Days 21 Hours 10.5 Hours
20 1.7 Days 20.7 Hours 10.3 Hours 5.2 Hours
40 20 Hours 10 Hours 5 Hours 2.5 Hours
80 9.5 Hours 4.8 Hours 2.4 Hours 1.2 Hours
17 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Proper use of these products depends on careful reading and understanding of labeling and instructions. Please refer to the INOmax DSIR and INOblender operation manuals for guidance. Also
for use.
INOmax DSIR Warnings: • INOmax DSIR subtracts gas from the breathing circuit via the
some ventilators. The trigger sensitivity of the ventilator should be checked after connecting the INOmax DSIR to the breathing circuit.
• Patient disconnect and high-pressure alarms are required for the ventilator.
INOmax DSIR Cautions:• Insert the Injector Module on the dry side of the breathing circuit
• Avoid medications interfering with the gas monitoring system; administer any aerosolized medications distal to the sampling tee.
18Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Acutronics Medical Systems AG Fabian
1. Fabian+ nCPAP Evolution
3. Injector Module Electrical Cable4. INOmax DSIR5. NO/N2 Injector Tube6. Connecting Tube (15 inches)7. Injector Module
8. 22F X 15M Adapter
10. Inspiratory Breathing Circuit Hose11. Gas Sample Tee12. Patient Wye13. Proximal Pressure Tube14. Expiratory Breathing Circuit Hose
9
3
10
2
12
11
1 4
1413
8
7
6
19 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Acutronics Medical Systems AG Fabian HFO
1. Fabian HFO Ventilator
3. Injector Module Electrical Cable4. INOmax DSIR5. NO/N2 Injector Tube6. Injector Module7. T-Connector Assembly, #7209.e8. Connecting Tube (15 inches)
9. 22F X 15M Adapter
11. Inspiratory Breathing Circuit Hose12. Gas Sample Tee13. Patient Wye14. Proximal Pressure Tube15. Expiratory Breathing Circuit Hose
10
3
11
2
13
12
1
9
8
67
4
14
20Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
A-Plus Medical Babi-Plus Bubble CPAP
1. Oxygen Source2. Oxygen Tubing3. Pressure Relief Manifold4. Injector Module5. Temperature Probe6. 90 Degree Sample Port Adapter7. Nasal Prongs8. Babi Plus Bubble PAP Valve
9. Tee Adapter10. Breathing Circuit
12. NO/N2 Injector Tube13. Injector Module Electrical Cable14. INOmax DSIR
9 811 710
14
34
6
2
12
13
1
21 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
(Intentionally left blank)
22Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Bagging Systems While Using the Injector Module
the following steps should be taken for use with the manual resuscitator bags:• Use the smallest bag adequate to deliver the
desired tidal volume.• Oxygen tubing lengths greater than 72 inches
should not be used.
• Use the lowest practical inspired oxygen concentration.
several times to empty residual gas in the bag prior to using the system to ventilate a patient.
23 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1
2 643
10
12
11
9
8
7
1. O2 Flowmeter (wall outlet or cylinder)2. O2 Tubing3. 15M X 4.5 mm Adapter4. 22M/15F X 22M/15F Adapter5. Injector Module6. 15M X 4.5 mm Adapter7. O2 Tubing8. O2 Tubing Sample Tee
10. NO/N2 Injector Tube11. Resuscitator Bag with O2 Reservoir12. Injector Module Electrical Cable
24Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
WARNING:• 2
2
2
2
2 2
22 2
2
2
25 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. O2 Flowmeter 2. Injector Module Electrical Cable3. NO/N2 Injector Tube
5. O2 Tubing6. O2 Tubing Sample Tee
1011 9 81213
7
6
42
3
1
Bagging Systems While Using the Injector Module
8. Pressure Gauge9. 15M X 4.5mm Adapter10. Injector Module11. 22M/15F X 22M/15F Adapter12. 15M X 4.5mm Adapter13. O2 Tubing
26Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
High Frequency Ventilator Circuit
WARNING:• The INOmax DSIR
patient to avoid NO delivery transiently exceeding the
ventilation as soon as the catheter is removed from the airway. This will limit the extent of over delivery above the NO set dose.
Caution:• If the set dose is below 5 ppm and the Servo pressure is 2.0 psig.
• A one-way valve should be placed between the injector module
the injector module if the Life Pulse is either put into Standby or cycled OFF.
• There are higher pressures in the breathing circuit than normal; use only parts provided in disposable package #50248 and tightly secure all connections.
27 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. INOmax DSIR2. Bunnell Life Pulse
5. Conventional Ventilator6. Life Port Adapter7. Endotracheal Tube
8. Sample Tee9. Patient Box10. One-Way Valve11. Injector Module12. NO/N2 Injector Tube13. Injector Module Electrical Cable
1 2 3 4
6
8
11
7
9
14
101213
28Part No. 20751 Rev-012014-08
Connecting INOmax DSIR Injector Module to the
Connecting INOmax DSIR Sample Tee to the
1. Gas Out Tube from Vent2. 15M X 4.5mm I.D. Adapter3. 22M/15F X 22M/15F Adapter4. Injector Module
5. 15M X 4.5mm I.D. Adapter6. 3cm Piece of Green Gas Out Tube7. One-Way Valve
1 2 3 4 6 7 8
Circuit ConnectionDiagrams
1
2 4
1. From Patient Box2. Cut Green tube at midpoint
(approximately six inches from the Life Port Adapter)
3. Patient Gas Sample Line with
4. Insert Sample Tee5. Life Port Adapter6. Endotracheal Tube
63
29 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Circle Anesthesia System
1. Patient Gas Sample Line with
2. Patient Gas Sample Line Input Connection3. INOmax DSIR4. Bellows Assembly5. Ventilator6. Ventilator Drive Gas7. Absorber Expiratory Port
8. Absorber Inspiratory Port9. Absorber10. Injector Module a. Injector Module Input End b. Injector Module Output End11. Inspiratory Tubing12. 22M/15F X 22M/15F Adapter13. Gas Sample Tee14. Patient Wye
14 13 12 11 10b 8 7 610 10a
1 3 4
9
2
than patient minute ventilation to avoid recirculation of gases.
30Part No. 20751 Rev-012014-08
1. Patient wye2. Dräger Babylog VN500 / Leoni-plus Ventilator3. Ventilator Expiratory Port4. Ventilator Inspiratory Port5. Patient Gas Sample Line Input Connection6. INOmax DSIR7. NO/N2 Injector Tube Front Panel Connection8. Injector Module Electrical Cable Front Panel
Connection
9. Injector Module10. One-Way Valve
14. Patient Gas Sample Line
15. Gas Sample Tee
1 2 6
9
8
1011
3 4
7
121314
Circuit ConnectionDiagrams
31 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. Oxygen Source2. Oxygen Tubing3. Bubble CPAP Pressure Manifold4. 22F X 15M Adapter5. 22M/15F X 22M/15F Adapter6. Injector Module7. Temperature Probe8. Nasal Prong Infant Interface9. Bubble CPAP Generator
10. F/P Inline Infant Nebulizer Kit (RT010) Adapter11. Breathing Circuit
13. NO/N2 Injector Tube14. Injector Module Electrical Cable15. INOmax DSIR
10 912
8
11
3
4
2
7
13
14
116
6
4
Fisher & Paykel Bubble CPAP
32Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Fisher & Paykel Infant Circuit Nasal Cannula
2. INOmax DSIR3. Oxygen Source4. Oxygen Tubing5. 22F X 15M Adapter6. Injector Module7. Pressure Relief Manifold8. 22M/15F X 22M/15F Adapter
9. Injector Module Electrical Cable10. NO/N2 Injector Tube
12. Breathing Circuit13. Temperature Probe14. Gas Sample Tee15. Nasal Cannula
1
12 11
10
1314
2
6
79
4
3
8
33 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
2. INOmax DSIR3. Oxygen Source4. Breathing Circuit Hose5. Injector Module6. Injector Module Electrical Cable7. NO/N2 Injector Tube8. 22F X 15M Adapter
10. Breathing Circuit11. Temperature Probe12. Gas Sample Tee13. 22M/15F X 22M/15F Adapter14. 22 mm ID X 22 mm ID Cuff Adapter
1214 11
21
74
6
3
8
910
16
17
13
34Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Hamilton Arabella Nasal CPAP
1. Arabella
3. INOmax DSIR4. NO/N2 Injector Tube5. Injector Module Electrical Cable6. Injector Module7. 22F X 15M Adapter
9. Heated Delivery Circuit10. Temperature Probe11. Universal Generator12. Arabella Sample Tee13. 90 Degree Sample Port Adapter
8 7
31 2
9
1311
6 410
12
35 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
ICU Ventilator Circuit
1. Patient Wye
3. Ventilator4. Ventilator Expiratory Port5. Ventilator Inspiratory Port6. Patient Gas Sample Line Input Connection7. INOmax DSIR8. NO/N2 Injector Tube Front Panel Connection9. Injector Module Electrical Cable Front Panel Connection10. 22M/15F X 22M/15F Adapter11. Injector Module Electrical Cable Connection12. Injector Module NO/N2 Injector Tube Connection13. 22F X 15M Adapter
17. Gas Sample Tee
a. Injector Moduleb. 22F Inletc. 22M / 15F Outlet
b
ac
1 3 7
11
9
1213
4 6
8
1617
2
14
10
36Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
a. Injector Moduleb. 22F Inletc. 22M / 15F Outlet
b
a
c
Oscillatory Ventilator with a Filtered Circuit
1 2 4
6
9
10
11
12
11
13
1718
14
7
16
3
8
37 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. Sensormedics 3100A/B Ventilator2. Ventilator Outlet3. 22M Adapter4. Injector Module5. Injector Module Electrical Cable
Connection6. INOmax DSIR7. NO/N2 Injector Tube8. 8 mm Tubing X 15M Adapter9. One-Way Valve
10. Paw Limit Valve Control11. Filter
14. Bias Flow Tube
16. 90 Degree Sample Port Adapter17. Dump Valve Control18. Paw Control Valve
WARNING: Omission of the one-way valve may result in high NO delivery.
38Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Ventilator with a Rigid or Flexible Circuit
WARNING: Omission of the one-way valve may result in high NO delivery.
1. Sensormedics 3100A/B Ventilator2. Ventilator Outlet3. 22M Adapter4. Injector Module5. INOmax DSIR6. NO/N2 Injector Tube Connection7. Injector Module Electrical Cable Connection8. One-Way Valve
11. Patient Gas Sample
12. 90 Degree Sample Port Adapter13. Bias Flow Tube
a. Injector Moduleb. 22F Inletc. 22M / 15F Outlet
b
a
c
1 2 4
6
7
8
9
10
11
13
12
3
39 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. Patient Wye
3. SLE50004. INOmax DSIR5. NO/N2 Injector Tube6. Injector Module Electrical Cable
7. Injector Module8. 22F X 15M Adapter
10. Inspiratory Breathing Circuit Hose11. Expiratory Breathing Circuit Hose12. Gas Sample Tee
1 2 4
7
6
3
91012 811
Note: • Validated for use outside of the United States. • A one-way valve is not required for use during high frequency
ventilation mode.
40Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Spontaneous Breathing Patient on a Mask Circuit
1. O2 Tubing2. 15M X 4.5 mm Adapter3. 22M/15F X 22M/15F Adapter4. Breathing Circuit Tee5. Breathing Circuit Bag6. Injector Module7. Breathing Circuit Hose8. Gas Sample Tee9. 22M/15F X 22M/15F Adapter
10. One-Way Valve11. Sealed Face Mask12. One-Way Valve
14. NO/N2 Injector Tube15. INOmax DSIR16. Injector Module Electrical Cable17. O2 Flowmeter (wall outlet or cylinder)
17
32 1061 7 12114 98
1314
16
41 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Spontaneous Breathing Patient on a Nasal Cannula
The INOmax DSIR can be used with nasal cannula to deliver
rate as low as 2 L/min.
WARNING: Do not use the INOmax DSIR backup mode with
1. O2 Flowmeter2. O2 Tubing3. 15M x 4.5 mm Adapter4. 22M/15F x 22M/15F Adapter5. 300 mm of 22 mm Hose6. Injector Module
4 432
1
6 34 7 8
1110
9
12
7. O2 Tubing Sample Tee8. Patient Nasal Cannula9. Patient Gas Sample Line10. INOmax DSIR11. Injector Module Electrical Cable12. NO/N2 Injector Tube
42Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
10
17
14
31 2
11
18
912
4
6
78
16
13
2. INOmax DSIR3. Injector Module4. System Pressure Relief Valve5. Air/Oxygen Blender or Oxygen Blender6. Oxygen Tubing7. Temperature Probe (Short Cable)8. Angled 22 mm Connector9. Patient Circuit
10. Temperature Probe Connector11. Second Temperature Probe Connector12. Comfort Flo Cannula13. Injector Module Electrical Cable14. NO/N2 Injector Tube15. 22F X 15M Adapter
17. Temperature Probe (Long Cable)18. 90 Degree Sample Port Adapter
43 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Transport Ventilator Diagram
1. Patient Wye2. Expiratory Breathing Circuit Hose3. Patient Gas Sample Line with
4. Ventilator Expiratory Valve5. Ventilator6. INOmax DSIR
1
2
9
8
10
11
3
4
7
1213
6
7. Ventilator Inspiratory Port8. 22M/15F X 22M/15F Adapter 9. Injector Module Electrical Cable10. NO/N2 Injector Tube11. Injector Module12. Inspiratory Breathing Circuit Hose13. Gas Sample Tee
44Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. PEEP/Exhalation Valve2. Patient Wye3. Circuit Hose4. Patient Gas Sample Linewith
5. Ventilator6. INOmax DSIR
1 2
9
8
1011
3 4
7
1213
6
7. Ventilator Inspiratory Port8. 22M/15F X 22M/15F Adapter 9. Injector Module Electrical Cable10. NO/N2 Injector Tube11. Injector Module12. Inspiratory Breathing Circuit Hose13. Gas Sample Tee
45 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
(Intentionally left blank)
46Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. INOmax DSIR2. O2 Flowmeter3. O2 Tubing4. 15M x 4.5mm Adapter5. 22M/15F x 22M/15F Adapter6. 300mm of 22mm Hose7. 22M/15F x 22M/15F Adapter8. Injector Module
10
1
14
12
9
8
16
11
7
6
4
3
2
139. 15M x 4.5mm Adapter10. Vapotherm 2000i11. Patient Delivery Tube12. O2 Tubing Sample Tee13. Patient Cannula14. Patient Gas Sample Line
15. NO/N2 Injector Tube16. Injector Module Electrical Cable
Vapotherm 2000i
47 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. Patient Gas Sample Line
2. INOmax DSIR3. Precision Flow Unit4. Injector Module5. Patient Delivery Tube6. Oxygen Tubing Sample Tee7. Patient Cannula8. Injector Module Electrical Cable9. NO/N2 Injector Tube
31 2
6
9
7
Connection to the Vapotherm Precision Flow
4
8
• The INOmax DSIR adds NO/N2proportion to the NO setting (up to 10% at 80 ppm) and subtracts gas
0.23 L/min.
Vapotherm Precision Flow. It is recommended that after an NO setting
• Follow all manufacturer instructions for connection to the Vapotherm Precision Flow.
48Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
1. INOmax DSIR2. Heated Delivery Circuit3. Infant Flow System4. Infant Flow Generator5. Sample Tee6. Temperature Probe
9. 22F X 15M Adapter10. Injector Module11. NO/N2 Injector Tube12. Injector Module Electrical Cable
4
1 3
12
7 610 8911
2
49 Part No. 20751 Rev-012014-08
Circuit ConnectionDiagrams
Viasys Infant Flow SiPAP
1. INOmax DSIR2. Abdominal Respiratory Sensor3. Transducer Interface4. Infant Flow SiPAP5. Infant Flow Generator6. Sample Tee7. Temperature Probe
8. Heated Delivery Circuit
10. 22F X 15M Adapter11. Injector Module12. Injector Module Electrical Cable13. NO/N2 Injector Tube
8 7 611 9
1 3 4
10
2
13
12
14
50Part No. 20751 Rev-012014-08
INOblender Circuit Connection Diagram
INOblender Warnings:• The purge procedure must be followed to help ensure NO2 is
before the manual resuscitator bag is connected to the patient. The manual bag should be squeezed continuously during use to avoid NO2 building up in the bag. If the bag is not squeezed
removed from the patient and the purge procedure performed before continuing.
• Persons using this device should be trained on and experienced in the use of this device to assure effective administration of INOMAX and to avoid injury to the patient or
dioxide or other reaction products.
INOblender Cautions:• Refer to the manufacturer’s procedures for using the resuscitation
2 until the NO pressure gauge reads zero, then turn the O2
Note: Connections to various ventilators as well as their corresponding disposable circuits, are unique to each manufacturer. Please
instructions for use for guidance.
51 Part No. 20751 Rev-012014-08
INOblender Circuit Connection Diagram
INOblender Connection to the Fisher & Paykel Neopuff Resuscitator
1. Oxygen Source2. Neopuff3. T-Piece Circuit (with Duckbill Port)4. Patient Connection5. Temperature Probe
8. Oxygen Tubing9. INOblender10. INOMAX Inlet
7 6
32
4
Oxygen Source
INOMAX Inlet
4
9
8
10
1
52Part No. 20751 Rev-012014-08
Note: Ensure the white plastic tip is in place.
A. Attach a regulator to an INOMAX cylinder with greater than
Changing INOMAXCylinders
WARNING: • A new INOMAX cylinder and regulator must be purged
before use to ensure the patient does not receive an excess level of NO2.
IR and the INOMAX cylinder for more than one hour will result in interruption of INOMAX delivery.
Caution: • Replace an INOMAX cylinder when its pressure is
less than 200 psig.• When using the Transport Regulator/Cap Assembly (PN 10022)
ensure the cap is fully seated and in place on the INOmeter and the infrared cable is connected and latched to the infrared connector port on the back of the INOmax DSIR.
Changing INOMAX Cylinders
53 Part No. 20751 Rev-012014-08
C. Purge the high pressure hose.
D. Connect the pressure hose.
Changing INOMAXCylinders
B. Perform high pressure leak test.
54Part No. 20751 Rev-012014-08
Changing INOMAXCylinders
may activate the “Two Cylinders Open” alarm until the empty
Note: If using the INOmax DSIR Transport Regulator/Cap Assembly, transfer the cap from the exhausted INOMAX cylinder to the new INOMAX cylinder at this time; the “Cylinder Not Detected” alarm may occur.
F. Close the cylinder valve on the empty cylinder and remove the hose from the back of the INOmax DSIR.
G. Depressurize and remove the regulator from the empty cylinder.
55 Part No. 20751 Rev-012014-08
Changing INOMAXCylinders
(Intentionally left blank)
56Part No. 20751 Rev-012014-08
High CalibrationConnection Diagrams
Connection Diagram for NO and NO2 High Range Calibration
1. Cylinder Cap2. Regulator Seal3. Regulator4. Tubing Adapter5. Attach Tube Kit6. Turn valve counter-clockwise to
7
To INOmax DSIR
6a. If the pressure is in the red or black zone (0-25 psig) select another INOcal cylinder.
7. Attach tube kit to sample line8. Calibrate sensor9. Remove sample line from tube kit
11. One-way Valve
6
6a
57 Part No. 20751 Rev-012014-08
High CalibrationConnection Diagrams
Calibration Setup for O2 High Range Calibration
1. 100% O2 Source2. O2 Tubing3. 15M x 4.5mm I.D. Adapter4. Gas Sample Tee
1
2
3 4
58Part No. 20751 Rev-012014-08
INOmax DSIRDisposable Adapters
INOmax DSIR Patient Circuit Disposables(Note: Graphics not actual size)
Adapter, Adapter, 22F X 15M15M Fits 4.5mm ID Tubing
Adapter, 22M/15F X 22M/15F Adapter, Cuff, 22mm ID X 22mm ID
Adapter, Gas Sample Tee Adapter, 90 degree Sample Port
Bunnell Life Pulse Disposable Disk Filter, 0.5 micronAdapters Convenience Pack
Neonatal Tubing, NO/N2 Injector Tube 10mm (2 pieces)
59 Part No. 20751 Rev-012014-08
One-way Valve, 22F X 22M Patient Gas Sample Line with
Pediatric Extension, Sample Tee, O2 Tubing15 mm (6 inches)
Sensormedics 3100A/B Water Separator CartridgeFiltered Circuit Disposable Adapters Convenience Pack
INOmax DSIRDisposable Adapters
60Part No. 20751 Rev-012014-08