DESpec spectrographs. Jennifer Marshall Darren DePoy Texas A&M University. Prototype design: VIRUS clone. 10 fiber-fed unit spectrographs, 400 fibers each Wavelength range 550-950 nm in one arm Resolution at 950 nm = 3167 Uses 2 DECam CCDs in each arm Based on VIRUS design. VIRUS. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: DESpec  spectrographs

DESpec spectrographs

Jennifer MarshallDarren DePoy

Texas A&M University

Page 2: DESpec  spectrographs

Prototype design: VIRUS clone

• 10 fiber-fed unit spectrographs, 400 fibers each

• Wavelength range 550-950 nm in one arm

• Resolution at 950 nm = 3167• Uses 2 DECam CCDs in each arm• Based on VIRUS design

Page 3: DESpec  spectrographs

VIRUS

• The first highly-replicated instrument in optical astronomy

• 150+ channel fiber-fed Integral Field Spectrograph placing >33,000 1.5” dia fibers on sky

• 350-550 nm coverage and R~700

Page 4: DESpec  spectrographs

VIRUS spectrographs

• Simple design– Single reflection spherical collimator– Schmidt camera

• Two lenses + one spherical mirror– VPH grating

• High throughput

Unit spectrographs packaged in pairs

Page 5: DESpec  spectrographs

Texas A&M’s role in HETDEX

• Participate in optical and mechanical design of VIRUS

• Fabrication and procurement of VIRUS components

• Assemble VIRUS unit spectrographs• Optically align instruments in lab• Ship to McDonald

Page 6: DESpec  spectrographs

HETDEX+VIRUS specs

• Wavelength: 350 – 550 nm• Resolution: R~700• Integration time: t=20 minute• Fiber diameter: 1.5” on sky• Sensitivity

– Line flux limit 3.5e-17 – Continuum detection gAB~22 mag

Page 7: DESpec  spectrographs

Flexibility of VIRUS design

• VIRUS design is readily adaptable to other fiber-fed spectrograph systems– Easy to change resolution, wavelength

range, etc. with simple redesigns• Has already been used as basis of new

spectrograph design– LRS2, a moderate resolution red-optimized

spectrograph for HET

Page 8: DESpec  spectrographs

DESpec as VIRUS clone

• Relatively straightforward redesign of VIRUS can produce DESpec– Change grating– Reoptimize coatings– Refractive camera?

Page 9: DESpec  spectrographs

Prototype design: VIRUS clone

• 10 fiber-fed unit spectrographs, 400 fibers each

• Wavelength range 550-950 nm in one arm

• Resolution at 950 nm = 3167• Uses 2 DECam CCDs in each arm• Based on VIRUS design

Page 10: DESpec  spectrographs

Alternate design: two arms

• 10 fiber-fed unit spectrographs, 400 fibers each• Increased wavelength range• Two arms, blue (500-760) and red (760-1050)• Different resolution in each arm

– 625 nm, R~1923– 950 nm, R~3276

• Uses 2 DECam CCDs in each arm• Significant design modification from VIRUS

– Similar optical layout to GMACS

Page 11: DESpec  spectrographs

GMACS

• Wide-field, multi-object optical spectrograph for GMT

• Four quadrants with two arms (red and blue) each– One quadrant could be

modified to become DESpec unit spectrographs

Page 12: DESpec  spectrographs

How to decide

• Need science input to provide instrument requirements:– Wavelength range– Resolution– Density of targets/number of fibers– Fiber size on sky

Page 13: DESpec  spectrographs

Work required to design DESpec as VIRUS clone

• Science input for instrument requirements

• New optical design for camera• Mechanical redesign of camera• Mechanical design of instrument

mounting scheme on telescope• Cooling system redesign

Page 14: DESpec  spectrographs

Work required to design DESpec as VIRUS clone

• We would need about 2 years of engineering effort for redesign

• A&M could assemble and test spectrographs in ~2 years– Lots of experience from VIRUS!

• These are estimates; will require more careful schedule/planning

Page 15: DESpec  spectrographs

Work required to design DESpec two-arm design

• More optical and mechanical design work required– Increases cost

• May need non-DECam CCDs for blue channel– Increases cost

Page 16: DESpec  spectrographs

Summary

• VIRUS design could be easily and relatively cheaply adapted to DESpec spectrographs– Two-arm re-design is more involved but

possible• Would need ~10 spectrographs• 3-4 years of effort in redesign and

assembly

Page 17: DESpec  spectrographs

Optimal Spectral Resolution

Jennifer MarshallDarren DePoy

Steven VillanuevaTexas A&M University

Page 18: DESpec  spectrographs

What is the “best” spectral resolution (λ/Δλ)?

• Science objectives set broad constraints• Various considerations suggest low resolution

– Easier optics– Smaller CCD format– Cheaper spectrographs

• Low means R=1000-1500– 200-300 km/sec

• Night sky emission lines are bright in the red– Suggest resolution should be higher– Isolates lines and allows for more “clean” pixels– What does “higher” mean?

Page 19: DESpec  spectrographs

Low resolution red spectra compromised by night sky emission lines

Page 20: DESpec  spectrographs

Fewer compromised pixels at higher resolution

Page 21: DESpec  spectrographs

Much less of a problem at bluer wavelengths

Page 22: DESpec  spectrographs

Lower resolution in “blue” not substantially compromised

Page 23: DESpec  spectrographs

Fraction of “uncontaminated” pixels (SNR > 0.9 relative to no night sky emission lines)

Page 24: DESpec  spectrographs

SNR per pixel versus resolution

Page 25: DESpec  spectrographs

SNR per pixel versus resolution

Page 26: DESpec  spectrographs

Conclusions

• Red spectra require relatively high resolution– R > 2500– Optimization is soft

• Blue spectra can be lower resolution– R > 500