detection of low energy particle eurons-jra4 dlep

4
EURONS PCC & GA meeting, Helsinki 2007 Olof Tengblad, JRA-04 DLEP University of Århus, Denmark H. Fynbo CSIC, IEM, Madrid, Spain O. Tengblad Chalmers University of Technology Göteborg, Sweden T. Nilsson University of York, UK B.R. Fulton ISOLDE-CERN University of Jyväskylä, Finland A. Jokinen Micron Semiconductors UK http:// www.micronsemiconductor.co.uk / Acqiris http:// www.acqiris.com / MESYTEC http://www.mesytec.de Detection of Low Energy Particle EURONS-JRA4 DLEP

Upload: ivan

Post on 19-Jan-2016

40 views

Category:

Documents


4 download

DESCRIPTION

Detection of Low Energy Particle EURONS-JRA4 DLEP. CSIC, IEM, Madrid, Spain O. Tengblad. University of Århus, Denmark H. Fynbo. University of Jyväskylä, Finland A. Jokinen. University of York, UK B.R. Fulton. Chalmers University of Technology Göteborg, Sweden T. Nilsson. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Detection of Low Energy Particle EURONS-JRA4 DLEP

EURONS PCC & GA meeting, Helsinki 2007Olof Tengblad, JRA-04 DLEP

University of Århus, Denmark H. Fynbo

CSIC, IEM, Madrid, Spain O. Tengblad

Chalmers University of TechnologyGöteborg, Sweden T. Nilsson

University of York, UK B.R. Fulton

ISOLDE-CERN

University of Jyväskylä, Finland A. Jokinen

Micron SemiconductorsUK http://www.micronsemiconductor.co.uk/

Acqirishttp://www.acqiris.com/

MESYTEC http://www.mesytec.de

Detection of Low Energy Particle

EURONS-JRA4 DLEP

Page 2: Detection of Low Energy Particle EURONS-JRA4 DLEP

EURONS PCC & GA meeting, Helsinki 2007Olof Tengblad, JRA-04 DLEP

• Year 1– Test of Si materials and methods √– algorithm development for Particle Identification √

• Year 2– Design of prototype detectors, beam tests in Madrid and Århus. √– Fast preamplifier development, multiplexing, √– Prototype preamplifier and detectors √

• Year 3– Prototype detectors and electronics to be tested on-line @ ISOLDE and JYFL.

October 2007

Workplan 3 year running

period

The monolithic detector was delivered, together with the associated multiplexed electronic in August 2007. An experiment IS417 “study of charge particles from exotic Li isotopes” using these detectors and electronics is scheduled for October 2007. •The data has to be analyzed and evaluated.•We need to further map the detector response to particles of different mass at the tandem accelerator.•Prepare the final reportWe thus ask to be allowed for a prolongation of the project for 1 more year. The participants have funds reserved for this action, why no extra funding is requested.

Page 3: Detection of Low Energy Particle EURONS-JRA4 DLEP

EURONS PCC & GA meeting, Helsinki 2007Olof Tengblad, JRA-04 DLEP

Detector delivered August 2007

64 monolithic telescopes 128ch on 5x5cm2

Side view with connectorsArea 64x 7mm2, E 1m, E 400 m

Doped & cut Si-wafer from SGS-THOMSON -Catania, detector assembly by DLEP and industrial partner MESYTEC

Page 4: Detection of Low Energy Particle EURONS-JRA4 DLEP

EURONS PCC & GA meeting, Helsinki 2007Olof Tengblad, JRA-04 DLEP

Complete readout electronic

Delivered August 2007developed for DLEP by MESYTEC64 ch:

PA + TFA + CFD + A + triggerAnalogue & multiplexed

MDI-2 VME sequencer and ADChandles 512 multiplexed channels

4x 16ch cards

In big production; estimated cost

of complete readout chain 70 €/ch

½ liter 256 ch

256 ch

10M samples/bus12 bit ADC

Incorporated into the MIDAS DAQby Vic Pucknel, RAL Daresbury